©2013 Waters Corporation 1
Exploring the Challenges of the
Chromatographic Separation
of Polar Compounds
©2013 Waters Corporation 2
Topics for “Meet the Experts” Today
Goal of “Meet the Experts” series
Overview of the key principles and practical aspects of
separation chemistry as a function of polarity
Comparison of chromatographic methods for retention of polar
compounds
Conclusions
©2013 Waters Corporation 3
Goal of Today’s Presentation
Obtain at least one idea that can make a positive change in
your workflow this week
Provide pertinent and timely scientific information to aide you in
finding possible solutions for challenges you are facing today or
might face in this year
Provide opportunity to see what other scientists are doing for
the analysis of compounds that are similar to yours
©2013 Waters Corporation 4
What is a Polar Molecule?
General chemistry definition:
– A molecule whose centers of positive and negative charges do not coincide
– The degree of polarity is measured by the dipole moment of the molecule
Dipole moment is the product of the charge at either end of the dipole times the distance between the charges
– The unequal sharing of electrons within a bond results in a separation of positive and negative electric charge.
Polarity is dependent on the electronegativity difference between molecular atoms and compound asymmetry
©2013 Waters Corporation 5
Challenges in Work Flow Processes
Number of analyses is often increasing due to increasing concern of
time to market, brand equity (effective and safe) and gov’t regulations
Costs per sample is increasing due to increasing costs of consumables,
resources necessary for lab (facilities, human, etc.)
Prepare standards and samples
Data acquisition and processing
Data compilation and management
©2013 Waters Corporation 6
Sample Preparation is a Major Source of Laboratory Costs
Sample preparation is the most often cited area of improvement to save time and operating costs
Most sample preparation involves being in an organic phase
– Liquid/Liquid, PPT, Soxhlet, Distillation, Evaporation and Reconstitution
Many matrices will respond best to organic phases (gels, blisters, ointments, synthesis solvents, etc.)
image from dyapharma.com image from sefetec.net image from tasnee.com
©2013 Waters Corporation 7
Typical Sample Preparation Choices
Soxhlet
Extraction
Distillation
Evaporation
Macro Micro
Liquid / Liquid
Drying Grinding SFE
©2013 Waters Corporation 8
Topics for “Meet the Experts” Today
Goal of “Meet the Experts” series
Overview of the key principles and practical aspects of
separation chemistry as a function of polarity
Comparison of chromatographic methods for retention of polar
compounds
Conclusions
©2013 Waters Corporation 9
What Is a Good Chromatographic Separation Method? When Done?
Depends on the goals of the separation
– Resolution of all the analytes in a sample?
– Resolution of key analytes?
– Quantification of single analyte?
– Speed of separation?
– Other?
Use consistent, objective criteria for evaluating chromatographic
separation performance
– Individual peak attributes
– Relative peak attributes
©2013 Waters Corporation 10
Goals Today are a Combination of Business and Science
“Minimize the analysis time (or another overall cost function)
while meeting or exceeding the necessary effective resolution
around every peak of interest.
– Here, the target function is analysis time, and our optimization
goal is its minimization.”*
*Thomas L. Chester, Maximizing the Speed of Separations for Industrial Problems, J of Chrom A, 1261 (2012) 69– 77
©2013 Waters Corporation 11
Goals Today are a Combination of Business and Science
“Minimize the analysis time (or another overall cost function)
while meeting or exceeding the necessary effective resolution
around every peak of interest.
– Here, the target function is analysis time, and our optimization
goal is its minimization.”*
or
“Maximize the resolution of the least-resolved peak of interest,
relative to its specification, without exceeding an analysis time
limit.
– Here the optimization goal and target are the maximization of the
least-resolved peak of interest compared to its specification.”*
*Thomas L. Chester, Maximizing the Speed of Separations for Industrial Problems, J of Chrom A, 1261 (2012) 69– 77
©2013 Waters Corporation 12
Desirable Information is Critical in Separation Chemistry
Chemical properties [functional groups] – Ionizable species, polarity, pKa, molecular weight
Sample solubility – Log P
Number of compounds present – How many components are you trying to separate?
Sample matrix
Detection technique [UV, ELS, RI, FL, MS…etc.] – Based on available equipment or sensitivity requirements of assay
Criteria for success – Concentration range and quantitative requirements
– System suitability
©2013 Waters Corporation 13
Assessing Separation Attributes
Individual peak characteristics — shape, size, time
– Retention time
– Peak width
– Peak height
– Peak area
– Symmetry (USP tailing factor)
0
©2013 Waters Corporation 14
Assessing Separation Attributes
Relative peak characteristics
– Capacity factor
– Selectivity
– Resolution
– Number of peaks
– Column efficiency
0
©2013 Waters Corporation 15
Factors that Influence Chromatographic Resolution
Initial
Increase
Increase k
Increase N
©2013 Waters Corporation 16
Improving Resolution with Complementary Selectivity
1
1
k
k
4
NRs
Maximized in Separations by:
Range of column chemistries Multiple particle substrates Wide usable pH range High retentivity Wide range in selectivity
Maximized in Separations by:
Ultra-low dispersion system Smaller particles Higher pressure capability Well-designed columns
©2013 Waters Corporation 17
Factors that Influence Chromatographic Resolution
Initial
Increase
Increase k
Increase N
Efficiency Selectivity Retentivity
©2013 Waters Corporation 18
Chromatographic Resolution: Impacting Selectivity and Retentivity
Efficiency Selectivity Retentivity
©2013 Waters Corporation 19
Critical Components Chromatography
Define method objectives
Understand analyte properties and intended use of method
Devise initial LC conditions
Develop adequate separation by systematic screening or computer assisted development
Sample preparation procedure
Suitable sample clean-up procedure based on physical and chemical properties of matrix
Standardization [data processing]
Determine method linearity, accuracy and precision
Final method optimization/
robustness testing
Challenge method and identify weak spots
Method validation Prove assay meets
intended use
©2013 Waters Corporation 20
Topics for “Meet the Experts” Today
Goal of “Meet the Experts” series
Overview of the key principles and practical aspects of
separation chemistry as a function of polarity
Comparison of chromatographic methods for retention of
polar compounds
Conclusions
©2013 Waters Corporation 21
Importance of Polarity
Related to the structure of the elements, and any electron charge distribution of that molecule
In order to create chromatographic separations, it helps to understand the polarities of the:
– Sample/analytes
– Mobile phase
– Stationary phase (packing material)
©2013 Waters Corporation 22
Polarity Defines the Modes of Chromatography
Competition between the stationary phase and the mobile phase creates a
separation of compounds in a sample
Typically, the polarity of the mobile phase is OPPOSITE that of stationary phase
©2013 Waters Corporation 23
Comparison of 4 Major Chromatographic Methods for Retention of Polar Compounds
Normal Phase Chromatography (NPLC)
Hydrophilic-Interaction Chromatography (HILIC)
Reversed-Phase Chromatography (RPLC)
Supercritical Fluid Chromatography (SFC)
©2013 Waters Corporation 24
Normal Phase Separation is Excellent for Polar Molecules
Normal Phase
Stationary Phase
Un-bonded
Silica (Polar) Surface
Packing Polarity Polar
Mobile Phase Polarity Non-Polar
Elution Order Most Non-Polar First
Effect of Increasing
Mobile Phase Polarity Reduces
Retention Time
©2013 Waters Corporation 25
Normal Phase LC
Normal phase conditions start with a
POLAR stationary phase and a NON-POLAR mobile phase
©2013 Waters Corporation 26
Normal Phase Benefits
There is a much larger choice of solvents
(compared to reversed-phase) to
manipulate selectivity
Widely applicable to a diverse range of
compounds in both polarity and functional group
Many organic compounds are more soluble in
normal-phase solvents
Ideal for separating positional isomers,
stereoisomers, diastereomers and chiral
compounds
©2013 Waters Corporation 27
Challenges Associated with Normal Phase Chromatography
Polar “contamination” in the stationary phase or mobile phase
will lead to:
– Difficulties in achieving reproducible retention times
– Difficulties in controlling and predicting solvent strength
– Lengthy equilibration times to achieve a stable baseline
– Solvent miscibility and proper mixing issues making gradients
impractical
Increased toxicity, flammability, solvent expense including
purchase and disposal
– Hexane, heptane
– Ethyl acetate, acids
– Chlorinated solvents
©2013 Waters Corporation 28
NPLC is Critical for Chiral Separations
Heightened awareness that enantiomers of racemic compounds
have different pharmacological activities, pharmacokinetics and
pharmacodynamics effects across all more products
Single enantiomers exhibit greater potency with fewer
side effects than most conventional molecules
Rigorous justification will be required for market approval of a
racemate of chiral products including drugs, pesticides, etc.
©2013 Waters Corporation 29
Normal Phase Use by Purpose, Industry, and Function
Purpose:
– Preferred method for sample that have limited water solubility
– Preferred method for separating isomers
Industry:
– Pharmaceutical: 22%
– Academia: 17%
– Chemicals: 15%
– Biotechnology: 11%
– CRO’s: 8%
– Other: 27%
Function:
– Applied R&D: 27%
– QC/QA: 22%
– Method Dev 14%
– Other 37%
SDi: Market Intelligence, Normal Phase HPLC Separations, July 29, 2010
©2013 Waters Corporation 30
If Not Traditional NPLC for Polar Compounds….Then What?
Hydrophilic-Interaction Chromatography (HILIC)
Reversed-Phase Chromatography
Supercritical Fluid Chromatography (SFC)
©2013 Waters Corporation 31
What is HILIC?
HILIC - Hydrophilic Interaction Chromatography
– Term coined in 1990 to distinguish from normal-phase*
HILIC is a variation of normal-phase chromatography without the
disadvantages of using solvents that are not miscible in water
– “Reverse reversed-phase” or “aqueous normal-phase” chromatography
Stationary phase is a POLAR material
– Silica, hybrid, cyano, amino, diol, amide
The mobile phase is highly organic (> 80% ACN) with a smaller
amount of aqueous mobile phase
– Water (or the polar solvent(s)) is the strong, eluting solvent
*Alpert, A. J. J.Chromatogr. 499 (1990) 177-196.
©2013 Waters Corporation 32
Benefits of HILIC
Retention of highly polar analytes not retained by reversed-phase – Less interference from non-polar matrix components
Complementary selectivity to reversed-phase – Polar metabolites/impurities/degradants retain more than parent
compound
Enhanced sensitivity in mass spectrometry – High organic mobile phases (> 80% ACN) promotes enhanced ESI-MS
response
– Direct injection of PPT supernatant without dilution
– Facilitates use of lower volume samples
Improved sample throughput – Direct injection of high organic extracts from PPT, LLE or SPE without the
need for dilution or evaporation and reconstitution
©2013 Waters Corporation 33
Expands the Selectivity Range with Polar Compounds
When to Use HILIC:
Need improved retention of
hydrophilic or ionizable
compounds
Need improved MS response
for polar or ionizable
compounds
Need improved sample
throughput for assays using
organic extraction
Reversed-phase
polar non-polar
Compound Index
Normal-phase
ESI-
MS
Re
spo
nse
exce
llen
tp
oo
r
HILIC
©2013 Waters Corporation 34
Influence of Polar Modifier on Retention and Selectivity
10 mM ammonium acetate with 0.02% acetic acid
Analytes:1: methacrylic acid 2: cytosine 3: nortriptyline 4: nicotinic acid
12 3
4
12
3
4
1 23
4
1
34
90:10 ACN:H2O
90:5:5 ACN:H2O:MeOH
90:5:5 ACN:H2O:EtOH
90:5:5 ACN:H2O:IPA
Minutes
0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00
2
Retention increases with decreasing solvent polarity
©2013 Waters Corporation 35
Compounds 1. Nicotinamide 2. Pyridoxine 3. Riboflavin 4. Nicotinic acid 5. Thiamine 6. Ascorbic Acid 7. B12 8. Folic Acid
1
2
3
4
5
6
7 8
Minutes
0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00
Water-Soluble Vitamins Using HILIC Conditions
©2013 Waters Corporation 36
0.00 2.00 4.00 6.00 8.00 10.00 12.00
2
1
3
4
5
Compounds 1. PMPA 2. CMPA 3. MMPA 4. IMPA 5. EMPA 500 ng/mL each
Organophosphonic Acids Using HILIC Conditions
©2013 Waters Corporation 37
If Not Traditional NPLC for Polar Compounds….Then What?
Hydrophilic-Interaction Chromatography (HILIC)
Reversed-Phase Chromatography
Supercritical Fluid Chromatography (SFC)
©2013 Waters Corporation 38
Reversed-Phase Chromatography for Polar Compound Retention
Non-polar stationary phase with >80% aqueous mobile phases
Strengths
– Familiar, well understood technique
– Many stationary phase choices
– Good reproducibility, stable equilibration
– High efficiency
Weaknesses of modern C18 phases designed for improved peak
shape for basic analytes: polar compound retention
– Poor retention of polar analytes on a high coverage, non-polar C18
stationary phase
©2013 Waters Corporation 39
Polar Retention: Why Does Atlantis® T3 Work?
Dominant retention mechanism is reversed-phase (van der
Waals forces – hydrophobic attraction)
– Retention maximized using 100% aqueous mobile phases
– Retention maximized by using reduced C18 coverage
o Polar analytes can “fit” between C18 ligands and interact with
pores of material
o Optimized particle morphology (i.e. pore diameter/volume)
Secondary interactions due to residual silanols that are more
accessible due to reduced C18 coverage
– Cation-exchange interactions
– Hydrogen bonding interactions
©2013 Waters Corporation 40
Minutes
0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00 11.00 12.00 13.00 14.00 15.00 16.00 17.00 18.00 19.00 20.00 21.00
22.00
Scalability to/from UPLC® Technology: T3 Bonding & Endcapping
Minutes
0.00 0.50 1.00 1.50 2.00 2.50 3.00 3.50
4.00
1
2
3
4 5
6
7
ACQUITY UPLC® HSS T3
HPLC Separation
UPLC® Separation
6
5
4
3
2 7
1
Compounds: 1. Norepinephrine 2. Epinephrine 3. Dopamine 4. 3,4-Dihydroxyphenylacetic acid 5. Serotonin (5-HT) 6. 5-Hydroxy-3-indoleacetic acid 7. 4-Hydroxy-3-methoxyphenylacetic acid
(HVA)
Atlantis® T3
T3 bonding and endcapping
present in both HPLC and UPLC
Technology
©2013 Waters Corporation 41
If Not Traditional NPLC for Polar Compounds….Then What?
Hydrophilic-Interaction Chromatography (HILIC)
Reversed-Phase Chromatography
Supercritical Fluid Chromatography (SFC)
©2013 Waters Corporation 42
SFC as a Replacement for Normal Phase LC
Normal-Phase LC (NPLC) methods use solvents (aliphatic hydrocarbons and chlorinated solvents) that many laboratories would like to reduce for health, safety, environmental, and cost reasons
Since the principles of SFC are similar to those of NPLC, methods should be able to be converted to SFC
– Reduces solvent usage and disposal
– Lowers the cost per analysis while enhancing green initiatives
SFC offers significant performance advantages over NPLC
– Better reproducibility
– Ability to perform gradient separations
o most NPLC separations are isocratic
– Compatible with mass detection
©2013 Waters Corporation 43
Historical Issues with Analytical SFC
Large system volume
– Prevents the adoption of small particles packing materials
– Prevents high throughput analysis
Pumping stability at low co-solvent percentage
– Limits the range of applications possible
Injection accuracy and compatible format
– Limits quantitative applications
– Limits moving from Normal Phase LC to SFC
Poor detection sensitivity
– Limits are too high QA/QC in GMP environment
©2013 Waters Corporation 44
Normal Phase LC
Normal phase conditions start with a
POLAR stationary phase and a NON-POLAR mobile phase
©2013 Waters Corporation 45
Increasing Selectivity Space: Convergence Chromatography
Stationary Phase
Silica / BEH
2-ethylpyridine
Cyano
Aminopropyl
Diol
Amide
PFP
Phenyl
C18 < C8
©2013 Waters Corporation 46
Increasing Selectivity Space: Convergence Chromatography
Solvent
Pentane, Hexane, Heptane
Xylene
Toluene
Diethyl ether
Dichloromethane
Chloroform
Acetone
Dioxane
THF
MTBE
Ethyl acetate
DMSO
Acetonitrile
Isopropanol
Ethanol
Methanol
Stationary Phase
Silica / BEH
2-ethylpyridine
Cyano
Aminopropyl
Diol
Amide
PFP
Phenyl
C18 < C8
©2013 Waters Corporation 47
SFC Chromatography
Selectivity Space
Unlimited solvent
and stationary phase selectivity
Increasing Selectivity Space: Convergence Chromatography
Solvent
Pentane, Hexane, Heptane
Xylene
Toluene
Diethyl ether
Dichloromethane
Chloroform
Acetone
Dioxane
THF
MTBE
Ethyl acetate
DMSO
Acetonitrile
Isopropanol
Ethanol
Methanol
Stationary Phase
Silica / BEH
2-ethylpyridine
Cyano
Aminopropyl
Diol
Amide
PFP
Phenyl
C18 < C8
Weak
Str
ong
Supercritical CO2
Organic Modifier
©2013 Waters Corporation 48
Genesis of Innovation and Promise of Expanding the Selectivity Factors
Data courtesy of Davy Guillarme, Jean-Luc Veuthey LCAP, University of Geneva, Switzerland
©2013 Waters Corporation 49
Genesis of Innovation and Promise of Expanding the Selectivity Factors
Data courtesy of Davy Guillarme, Jean-Luc Veuthey LCAP, University of Geneva, Switzerland
©2013 Waters Corporation 50
Genesis of Innovation and Promise of Expanding the Selectivity Factors
Data courtesy of Davy Guillarme, Jean-Luc Veuthey LCAP, University of Geneva, Switzerland
©2013 Waters Corporation 51
Genesis of Innovation and Promise of Expanding the Selectivity Factors
Data courtesy of Davy Guillarme, Jean-Luc Veuthey LCAP, University of Geneva, Switzerland
©2013 Waters Corporation 52
Genesis of Innovation and Promise of Expanding the Selectivity Factors
Data courtesy of Davy Guillarme, Jean-Luc Veuthey LCAP, University of Geneva, Switzerland
©2013 Waters Corporation 53
UltraPerformance Convergence Chromatography
Convergence Chromatography is a category of separation science that provides
orthogonal and increased separation power, compared to liquid or gas
chromatography, to solve separation challenges.
UltraPerformance Convergence Chromatography [UPC2] is a holistically
designed chromatographic system that utilizes liquid CO2 as a mobile phase to
leverage the chromatographic principles and selectivity of normal phase
chromatography while providing the ease-of-use of reversed-phase LC.
The ACQUITY UPC2 System is built utilizing proven UPLC Technology to enable
scientists the ability to address routine and complex separation challenges while
delivering reliability, robustness, sensitivity and throughput never before possible
for this analytical technique.
©2013 Waters Corporation 54
Why the Name?
Giddings, J.C. (1965) A critical evaluation of the theory of gas chromatography. In Gas Chromatography. 1964, edited by A. Goldup, p. 3-24. Elsevier, Amsterdam
In this article Dr. Giddings stated “One of the most interesting features of ultra high pressure gas chromatography would be convergence with classical liquid chromatography.”
Prof. Calvin Giddings (1930-1996)
©2013 Waters Corporation 55
Harnessing the Promise of Smaller Particles
2.1 x 150, 1.7µm Flow = 1.4 mL/min
Caffeine, Carbamazepine, Uracil, Hydrocortisone, Prednisolone, and Sulfanilamide
2.1 x 150, 5µm Flow = 1.4 mL/min
AU
0.00
0.05
0.10
0.15
0.20
0.25
0.30
Minutes
0.00 0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00
NUSP(Sulfan) = 19,809
NUSP(Sulfan) = 6,561
AU
0.00
0.05
0.10
0.15
0.20
0.25
0.30
0.00 0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00
1
2 6
4
5
3
3X improvement in efficiency 1.7X increase in sensitivity 1.7X increase in resolution
©2013 Waters Corporation 56
ACQUITY UPC2 Binary Solvent Manager: Controlled Gradient Solvent Delivery
0.5% difference in programmed solvent composition (even below 5%) results in controlled retention shift and minimized baseline noise
n=10
RT %RSD: <0.15
Area %RSD: <0.67
n=10
RT %RSD: <0.34
Area %RSD: <0.69
1.0 - 20% B Gradient
1-C
oum
arin -
0.7
88
2-F
lavon
e -
1.0
72
3-C
aff
ein
e -
1.2
09
4-T
hym
ine -
1.5
59
5-P
apaveri
ne -
1.6
41
AU
0.00
0.02
0.04
0.06
0.08
0.10
0.12
0.14
0.16
0.18
0.20
0.22
0.24
Minutes 0.00 0.10 0.20 0.30 0.40 0.50 0.60 0.70 0.80 0.90 1.00 1.10 1.20 1.30 1.40 1.50 1.60 1.70 1.80 1.90 2.00
1-C
oum
arin -
0.8
40
2-F
lavon
e -
1.1
16
3-C
aff
ein
e -
1.2
47
4-T
hym
ine -
1.5
84
5-P
apaveri
ne -
1.6
64
AU
0.00
0.02
0.04
0.06
0.08
0.10
0.12
0.14
0.16
0.18
0.20
0.22
0.24
Minutes 0.00 0.10 0.20 0.30 0.40 0.50 0.60 0.70 0.80 0.90 1.00 1.10 1.20 1.30 1.40 1.50 1.60 1.70 1.80 1.90 2.00
1.5 - 20% B Gradient
©2013 Waters Corporation 57
UPC2 Provides Opportunity for Expanding the Selectivity Space
: CSH PFP
AU
0.000
0.012
0.024
0.036
0.048
: HSS C18 SB
AU
0.000
0.012
0.024
0.036
0.048
: BEH HILIC
AU
0.000
0.012
0.024
0.036
0.048
: 2-EP
AU
0.000
0.012
0.024
0.036
0.048
Minutes
0.00 0.60 1.20 1.80 2.40 3.00 3.60 4.20 4.80 5.40 6.00
ACQUITY UPC2 Hybrid 2-EP 1.7 µm
ACQUITY UPC2 Hybrid 1.7 µm
G A D(1,2)
C
H F
B
E
G A D
C
H F
B
E
G A
D C
H F
B E
G A
D C
H
F*
B
E
ACQUITY UPC2 CSH Fluoro-Phenyl 1.7 µm
ACQUITY UPC2 HSS C18 SB 1.7 µm
AP
I
AP
I
AP
I
AP
I
©2013 Waters Corporation 58
Orthogonal to RPLC Metoclopramide Related Substances
ACQUITY UPC2
Reversed-Phase
AU
0.000
0.013
0.026
0.039
0.052
Minutes
0.00 1.20 2.40 3.60 4.80 6.00 7.20 8.40 9.60 10.80
1 2
3 4
5 6
8 9
AU
-0.003
0.000
0.003
0.006
0.009
Minutes
0.00 1.10 2.20 3.30 4.40 5.50 6.60 7.70 8.80 9.90
2
Metoclopramide
Metoclopramide
12 minutes
12 minutes
©2013 Waters Corporation 59
4.6 x 250 mm silica NPLC column (L3)
Hexane / Dichloromethane / glacial acetic acid
2.0 mL/min
Replacing NPLC with UPC2:
Anthralin USP Drug Substance Assay
Normal Phase HPLC
Cost approx: $0.92 per run
An
thra
lin
- 2
.212
IST
D -
5.0
55
AU
0.00
0.02
0.04
0.06
0.08
0.10
0.12
0.14
0.16
0.18
0.20
0.22
0.24
Minutes
0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00
10.0
©2013 Waters Corporation 60
4.6 x 250 mm silica NPLC column (L3)
Hexane / Dichloromethane / glacial acetic acid
2.0 mL/min
Viridis 2-EP 4.6 x 150 mm
CO2 / MeOH / glacial acetic acid
3.5 mL/min
Replacing NPLC with UPC2:
Anthralin USP Drug Substance Assay
SFC
Cost approx: $0.92 per run Cost approx: $0.04 per run
Suitability requirements met with NO change to
sample and/or standard
preparation
An
thra
lin
- 2
.212
IST
D -
5.0
55
AU
0.00
0.02
0.04
0.06
0.08
0.10
0.12
0.14
0.16
0.18
0.20
0.22
0.24
Minutes
0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00
An
thra
lin
- 1
.595
IST
D -
2.1
35
AU
0.000
0.010
0.020
0.030
0.040
0.050
0.060
0.070
0.080
Minutes
0.00 1.00 2.00 3.00 4.00 5.00 6.00
10.0 6.0
Normal Phase HPLC
©2013 Waters Corporation 61
Replacing NPLC with UPC2: Low Level Impurity Analyses by UPC2
Compound RT %Area S/N
Unk. Impurity 6.24 0.006 2.9
Unk. Impurity Not Found --- ---
Unk. Impurity 10.86 0.01 2.7
Unk. Impurity Not Found --- ---
Unk. Impurity 20.85 0.018 3
Unk. Impurity 26.63 0.021 3.2
Estradiol 30.86 99.87 ---
Main Impurity 36.81 0.077 9.2
6.23
7
10.8
55
20.8
50
26.6
32
30.8
56
35.8
19
AU
-0.0004
-0.0002
0.0000
0.0002
0.0004
0.0006
0.0008
0.0010
0.0012
0.0014
0.0016
0.0018
0.0020
Minutes
0.00 5.00 10.00 15.00 20.00 25.00 30.00 35.00 40.00 45.00 50.00
Normal Phase HPLC: Cost per run ~ $5.89
4.6 250 mm silica column
2,2,4-trimethylpentane / n-butyl chloride / MeOH, 2.0 mL/min
USP Method: Chromatographic Purity
of Estradiol
50.0
©2013 Waters Corporation 62
Replacing NPLC with UPC2: Low Level Impurity Analyses by UPC2
Compound RT %Area S/N
Unk. Impurity 6.24 0.006 2.9
Unk. Impurity Not Found --- ---
Unk. Impurity 10.86 0.01 2.7
Unk. Impurity Not Found --- ---
Unk. Impurity 20.85 0.018 3
Unk. Impurity 26.63 0.021 3.2
Estradiol 30.86 99.87 ---
Main Impurity 36.81 0.077 9.2
Compound RT %Area S/N
Unk. Impurity 2.26 0.012 3.4
Unk. Impurity 2.59 0.004 1.9
Unk. Impurity 3.34 0.01 3.1
Unk. Impurity 5.66 0.006 1.7
Unk. Impurity 6.15 0.016 5.5
Unk. Impurity 8.13 0.013 3.1
Estradiol 8.81 99.89 ---
Main Impurity 9.99 0.046 16
6.23
7
10.8
55
20.8
50
26.6
32
30.8
56
35.8
19
AU
-0.0004
-0.0002
0.0000
0.0002
0.0004
0.0006
0.0008
0.0010
0.0012
0.0014
0.0016
0.0018
0.0020
Minutes
0.00 5.00 10.00 15.00 20.00 25.00 30.00 35.00 40.00 45.00 50.00
UPC² : Cost per run ~ $0.05
Normal Phase HPLC: Cost per run ~ $5.89
4.6 250 mm silica column
2,2,4-trimethylpentane / n-butyl chloride / MeOH, 2.0 mL/min
2.1 x 150 mm ACQUITY UPC² BEH, 1.7 µm CO2 / MeOH
USP Method: Chromatographic Purity
of Estradiol
Transferred UPC² Method
UPC2 Benefits: Additional Impurities Detected and Higher
Sensitivity
50.0
14.0
©2013 Waters Corporation 63
Chiral Separation in a Validated Method Provides Competitive Advantage and Brand Equity
Chiral screening
Chiral method development
– MS and UV detection
Chiral inversion studies
Enantiomeric excess
Pesticides
Drugs of Abuse
Beta-blockers
Binol
Warfarin
Benzyl Mandelate (enantiomeric excess)
Carprofen (chiral method development)
Pantoprazole and Oxfendazole (chiral
method development with MS)
Clenbuterol
Phenylalanine methyl esters
Flurbiprofen
Cyclometalated Iridium (III) Complexes
Fast Chiral Separations
©2013 Waters Corporation 64
AU
0.00
0.12
0.24
0.36
0.48
Minutes
0.00 0.10 0.20 0.30 0.40 0.50 0.60 0.70 0.80 0.90 1.00
AU
0.00
0.30
Minutes0.00 2.00 4.00 6.00 8.00 10.00 12.00 14.00
Key advantages of moving to UPC2
– Results that are equal to or better
– 30X reduction in analysis time
– Nearly 75X reduction in solvent
• 135 µL of MeOH vs 10 mL of hexane/ethanol
Applicability of UPC2: Fast Chiral Screening
Providing meaningful impact to scientists from
discovery to QC based on the reducing of non-value
adding steps in analytical workflow process
Reduces the time consuming solvent mixing and
sample preparation so can reallocate resources to
other value adding analytical work
Increases the column lifetime so can reallocate
consumable budget
Reduces the cost of solvent investments of purchase
and removal
Reduces the complexity of instrument multi-method
use so can reduce the capital investments, or increase
value added human resources
Fast Chiral Screening UPC²
NPLC 15 min
0.5 min
©2013 Waters Corporation 65
Positional Isomers of DMBA
2.1
34
2.2
45
2.3
42
2.5
52
2.6
82
3.4
30
AU
0.00
0.10
0.20
0.30
0.40
0.50
0.60
0.70
0.80
0.90
Minutes0.00 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40 2.60 2.80 3.00 3.20 3.40 3.60 3.80
2,5
2,33,5
2,4
3,4
2,6
Mixture of 6 positional isomers of DMBA Sample: Each at 0.2 mg/mL in isopropanol (IPA) Column: 3.0 x 100 mm, 1.7 µm ACQUITY UPLC BEH125 Solvent: CO2 / MeOH with 0.2% formic acid
©2013 Waters Corporation 66
Increase in Resolution Power Provides Savings in Columns, Solvent and Time
AU
0.00
0.10
0.20
0.30
0.40
0.50
Minutes
0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00
Time0.00 2.00 4.00 6.00 8.00 10.00 12.00 14.00 16.00 18.00 20.00 22.00 24.00 26.00 28.00 30.00 32.00 34.00
AU
0.0
2.5e-2
5.0e-2
7.5e-2
1.0e-1
1.25e-1
1.5e-1
1.75e-1
2.0e-1
2.25e-1
2.5e-1
2.75e-1
UPC2
IC + OJ-H
Traditional SFC
2 x IC + 2 x OJ-H
Baseline resolution of all isomers was achieved in less than 10 min using UPC² even with three chiral centers Compared to the chiral HPLC methods, the UPC2
method offers a better resolution and a much shorter run time The UPC2 method also eliminated the need for toxic hexane often used in NPLC methods
©2013 Waters Corporation 67
Simplicity in Workflow
Simplify the workflow with UPC2
– Any simplification in the whole workflow, from initial sample
collection and sample preparation to analysis, generally has the
greatest business impact in any market segment
Combining multiple techniques (LC and GC into CC)
Combining multiple methods (NPLC and RPLC into CC)
Reducing sample prep and analysis times
– Directly injecting organic solvent extracts (LLE, SPE, etc.)
©2013 Waters Corporation 68
Separation of Compounds from Matrix Interferences (Bioanalysis)
Clopidogrel (RPLC)
Clopidogrel (UPC2)
Interfering Phospholipids (RPLC)
Interfering Phospholipids (UPC2)
©2013 Waters Corporation 69
Combining Multiple Techniques for Lipid Analysis
Gas Chromatography Liquid Chromatography Convergence Chromatography
Free fatty acids are typically derivatized to form the methyl esters (FAMEs) Analysis time 30 min
Analyzed by both HILIC and RP HILIC separates lipid classes by polar head group RP separates based on acyl chain length and number of double bonds
Single methodology to separate complex lipids by class Faster baseline separation of lipids based on chain length and number of double bonds
©2013 Waters Corporation 70
UPC2 Analysis of a Mouse Heart Extract
PC
SM LPC
PE
TAG
TAG: Triacylglycerides PE: Phosphotidylethanolamine PC: Phosphotidylcholine SM: Sphynogomyelin LPC: Lysophosphotidylcholine
ACQUITY UPC2 BEH column 5-50% B
©2013 Waters Corporation 71
Time0.60 0.70 0.80 0.90 1.00 1.10 1.20 1.30 1.40 1.50 1.60 1.70 1.80 1.90 2.00 2.10 2.20 2.30 2.40 2.50 2.60 2.70 2.80 2.90 3.00 3.10
%
0
100
0.60 0.70 0.80 0.90 1.00 1.10 1.20 1.30 1.40 1.50 1.60 1.70 1.80 1.90 2.00 2.10 2.20 2.30 2.40 2.50 2.60 2.70 2.80 2.90 3.00 3.10
%
0
100
8:0
14:0
16:0
18:0
20:0
22:0
24:0
10:0
12:0
ESI negative mode Free Fatty Acids (FFA) containing 8-24 acyl chain
ESI positive mode Triacylglycerols (TG) and Cholesterol esters (CE)
1
7
2
8 3
4 5
6 11
10
13 12
9
Peak Lipid Species
1 15:0/15:0/15:0 TG
2 18:3(∆9,12,15Cis)/18:3(∆9,12,15Cis)/18:3(∆9,12,15Cis) TG
3 16:0/16:0/16:0 TG
4 18:2(∆9,12Cis)/18:2(∆9,12Cis)/18:2(∆9,12Cis) TG
5 18:1(∆9Tr)/18:1(∆9Tr)/18:1(∆9Tr) TG
6 17:0/17:0/17:0 TG
7 18:1(∆9Tr)/18:1(∆9Tr)/18:1(∆9Tr) TG
8 18:3 CE
9 18:2 CE
10 17:0 CE 18:1 CE
11 18:0/18:0/18:0 TG
12 18:0 CE
13 19:0 CE
Neutral Lipids Based on Chain Length and Number of Double Bonds
ACQUITY UPC2 HSS C18 SB column 1-10% B
©2013 Waters Corporation 72
Official AOAC Method
Reducing Sample Prep and Analysis Time for -carotene Analysis
Dissolve/Hydrolyze
Extract
Dilute
Filter
LC Analysis
Modified AOAC Method
30 min
120 min
30 min
20 samples ~12.5 hrs
Dissolve/Extract
Filter
LC Analysis
30 min
20 samples ~10 hrs
UPC2 Method
Dissolve/Extract
Filter
UPC² Analysis
2 min
20 samples ~ 40 minutes
0.00 0.10 0.20 0.30 0.40 0.50 0.60 0.70 0.80 0.90 1.00 1.10 1.20 1.30 1.40 1.50
AU
0.0
1.0e-1
2.0e-1
3.0e-1
0.00 0.10 0.20 0.30 0.40 0.50 0.60 0.70 0.80 0.90 1.00 1.10 1.20 1.30 1.40 1.50
AU
0.0
1.0e-1
2.0e-1
3.0e-1
0.00 0.10 0.20 0.30 0.40 0.50 0.60 0.70 0.80 0.90 1.00 1.10 1.20 1.30 1.40 1.50
AU
0.0
5.0e-2
1.0e-1
1.5e-1
2.0e-1
2.5e-1
3.0e-1
0.91
0.91
0.91
0.77 1.31
-carotene standard
-carotene capsule
Carotenoids Mix
6 Replicates Peak Area Retention
Time (min)
Average 10326 0.91
RSD% 0.34 0
Label Claim: 15 mg/capsule
Assay
#1
Assay
#2
Assay
#3 Average RSD%
15.13 15.39 15.24 15.25 0.84%
©2013 Waters Corporation 73
Streamlining Sample Preparation of Creams and Ointments
Analysis of main API and impurities from creams and ointments involved a
tedious sample prep for QC assay
o Extraction from cream with organic solvent
o Back extraction into aqueous for RPLC analysis
Impact of UPC² in workflow
– Inject organic sample extraction directly
– Condense the sample preparation overhead
– Provide excellent reproducibility (peak area and retention time) and sensitivity
Imp 1 0.35%
Imp 2 0.11%
API 99.6%
*Standard sample of API & impurities
n = 5 injections Area %RSD < 1.0%
©2013 Waters Corporation 74
Monitoring Unstable Intermediates
Compound D sensitive to hydrolysis,
reverts to C
Could not use a single technique
– NPLC 90 min
– GC 18 min could not monitor B
Measure starting materials,
products and impurities
Analysis 8.5 min
Excellent sensitivity
Cpd A Cmpnd B
Cmpnd C Cmpnd D
Cmpnd E Cmpnd F
API
UPC2
©2013 Waters Corporation 75
EPA Method 8330B
C18, 4.6x250 mm
Cyano or phenyl hexyl column. 4.6x250 mm
• Typically analyzed by HPLC or GC • Long analysis times using HPLC or GC • Thermally labile compounds like tetryl
can not be done by GC • Inadequate baseline separation for
some of the compounds using official methods like EPA 8830B, requiring use of 2 orthogonal columns
©2013 Waters Corporation 76
UPC2 for 14 Explosives
Eliminates need for two analyses and two systems running one method each
Eliminates toxic solvents and concerns from waste disposal
Reduce time from 26 minutes (12 + 14) to 4 minutes with fast screening
©2013 Waters Corporation 77
Combining Multiple LC Methods Into One Fat Soluble Vitamins
Vitamin A Normal phase 12 minutes
Vitamin D3 Normal phase 20 minutes
Vitamin E Normal phase 30 minutes
Vitamin K1 Reversed-phase
12 minutes
β-carotene Normal phase 10 minutes
Lycopene Normal phase 10 minutes
Lutein Reversed-phase
10 minutes
©2013 Waters Corporation 78
Combining Multiple LC Methods Into One Fat Soluble Vitamins
Vitamin A Normal phase 12 minutes
Vitamin D3 Normal phase 20 minutes
Vitamin E Normal phase 30 minutes
Vitamin K1 Reversed-phase
12 minutes
β-carotene Normal phase 10 minutes
Lycopene Normal phase 10 minutes
Lutein Reversed-phase
10 minutes
Simultaneous Analysis of Fat-soluble Vitamins and Carotenoids in 10 minutes
Minutes
0.00 2.00 4.00 6.00 8.00 10.00
AU
Analyte Retention
Time (min) λ (nm)
(1) Vit A Acetate 3.14 345
(2) Vit E Acetate 3.76 263
(3) Vit K1 3.88 263
(4) Vit A Palmitate
4.34 345
(5) Vit E Tocopherol
4.82 263
(6) Lycopene 4.99 456
(7) Vit E Succinate
5.09 263
(8) β-carotene 5.12 456
(9) Vit D3 5.76 263
(10) Lutein 7.21 456
1
2 3
4
5
6
7,8
9 10
©2013 Waters Corporation 79
Improving Resolution Through Expanding Selectivity Space
1
1
k
k
4
NRs
Selectivity [α] and retentivity [k] impacted by: Stationary phase (column selectivity) Organic solvent (eluotropic series) Mobile phase additives (pH and ionic strength)
System efficiency [N] impacted by:
System dispersion Reduction in particle size
Impact on Rs % Improvement
Double N 20-40% Double k 15-20%
Double α > 400%
Convergence
Chromatography Selectivity Space
Unlimited solvent and stationary
phase selectivity
Solvent
Pentane, Hexane, Heptane
Xylene
Toluene
Diethyl ether
Dichloromethane
Chloroform
Acetone
Dioxane
THF
MTBE
Ethyl acetate
DMSO
Acetonitrile
Isopropanol
Ethanol
Methanol
Stationary Phase
Silica / BEH
2-ethylpyridine
Cyano
Aminopropyl
Diol
Amide
PFP
Phenyl
C18 < C8
©2013 Waters Corporation 80
Comparison of Chromatographic Methods for Retention of Polar Compounds
Technique Advantages Challenges
Traditional NPLC
There is a much larger choice of
solvents (compared to reversed-
phase) to manipulate selectivity
Widely applicable to a diverse range of
compounds in both polarity and
functional group
Many organic compounds are more
soluble in normal-phase solvents
Ideal for separating positional
isomers, stereoisomers, diastereomers
and chiral compounds
Long equilibration times Difficult to run gradients Not easily compatible with MS Difficult method development Purity of reagents Chemical modification of column
HILIC
High % organic mobile phases give
higher sensitivity in MS
Eliminate evaporation of SPE eluents
Sample and mobile phase solubility problems
Not well understood Not widely applicable
RPLC
Familiar technique
High efficiency
Rapid equilibration
Wide selection of columns
Dewetting under aqueous conditions Poor retention of polar compounds
Convergence
Chromatography and
UPC2
Familiar technique for purification
High efficiency
Rapid equilibration
Wide selection of columns that are
same as NPLC and RPLC
Uses same control, acquisition and
results sw
Initial learning of key method development factors as they are different from RPLC
Transfer of current methods need cross validation projects before in SOP
©2013 Waters Corporation 81
Supporting of Customers Seeking Theoretical and Practical Insight
Primary Theoretical Contributions
Developing the basis of the theoretical
support of the science
Provides basis of making next actions on
method development and key factors to
the separation
Practical & Business Contributions
Placing into workflow processes and
monitoring the metrics of impact based on
practitioners work
Best practices that are based on others
experiences including hyphenated
techniques (MS, ELSD, etc.)
©2013 Waters Corporation 82
Practitioners UPC2 Applications Expanding from the Customer’s Experiences
http://upc2.waters.com
©2013 Waters Corporation 83
Opportunity to Explore …
Frequent Concerns
Will my compound work with this
technology?
Is there any advantage to hyphenated
detection versus doing currently?
I do not have time to evaluate
technology as I have objectives to meet
in my job
Take the Challenge
Form to submit a sample for feasibility
analysis based on your criteria
Provides quick assessment before
considering more exploration how would
get to future state
Doesn’t interfere with taking care of your
current objectives, but doesn’t limit you
from future proofing your lab
©2013 Waters Corporation 84