+ All Categories
Home > Documents > EYC 2018 brochure-en-0-3 · testing report DAQ & RS-485 HMI Dew Point Mirror RH System 973 Humidity...

EYC 2018 brochure-en-0-3 · testing report DAQ & RS-485 HMI Dew Point Mirror RH System 973 Humidity...

Date post: 01-Aug-2020
Category:
Upload: others
View: 0 times
Download: 0 times
Share this document with a friend
12
• Calibration Lab – ILAC ( ISO / IEC 17025 ) • Factory Automatic QC Inspection System ─ According to IEC 61298 ISO 9001 www.eyc-tech.com Temperature / Humidity / Dew Point / Pressure / Flow / Air Velocity / CO . CO 2 . PM2.5 / Signal Transmitter OEM / ODM Service & Distributor Wanted Measuring Specialist
Transcript
Page 1: EYC 2018 brochure-en-0-3 · testing report DAQ & RS-485 HMI Dew Point Mirror RH System 973 Humidity Generator Thunder 3900 with Air Velocity ( 120 m/s ) Device under test (DUT) EYC

• Calibration Lab – ILAC ( ISO / IEC 17025 )

• Factory Automatic QC Inspection System ─ According to IEC 61298 – ISO 9001

www.eyc-tech.com

Temperature / Humidity / Dew Point / Pressure / Flow / Air Velocity / CO . CO2 . PM2.5 / Signal Transmitter

OEM / ODM Service & Distributor WantedMeasuring Specialist

Page 2: EYC 2018 brochure-en-0-3 · testing report DAQ & RS-485 HMI Dew Point Mirror RH System 973 Humidity Generator Thunder 3900 with Air Velocity ( 120 m/s ) Device under test (DUT) EYC

www.eyc-tech.com

• Measuring Range : 0.2 ... 100 bar• Electric Connection : DIN43650• Accuracy : ± 0.1 % of F.S.• Connection : Internal thread 1/4-18NPT• Output : 4 ... 20mA• Signal Connection : 2-wire• Application : Petrochemical industry / Chemical industry

Industrial Grade Differential Pressure Transmitterr

P064

• Sensor Type : Piezoresistive diaphragm• Measuring Range : 0 ... 20 bar• Output : 4 … 20mA• Accuracy : ± 0.5 % of F.S.• Signal Connection : 2-wire• Housing : SUS 316L• Application : Mechanical equipment / Water plant / Air equipment / Chiller unit

Differential Pressure Transmitterr

P063

• Measuring pressure for liquid• Relative Pressure : 0.1 ... 400 bar• Accuracy ( at 20 °C ± 5 °C ) : ± 0.5 % of F.S.• Housing material : SUS 316• Isolated diaphragm material : SUS 316L• Application : Process equipment / Food and Pharmaceutical Industry / Machine devices

Pressure Transmitter

P045

Piezoresistive Pressure Transmitter

• Sensor Type : Piezoresistive diaphragm• Measuring Range : Absolute pressure:0 ... 10 bar Relative pressure:0 … 40 bar• Output : 4 … 20 mA• Signal Connection : 2-wire• Accuracy : ± 0.3 % of F.S.• Application : Instrument manufacturing / Machine devices / Water plant / Air devices

P043

Air Quality LED Monitor

• Monitor PM2.5, CO2, temperature & humidity via RS-485 at the same time. • Display Type : High Brightness LED• Accuracy : ± 0.4 ℃ ; ± 3 %RH ( at 20 ... 90 %RH ) ; ± 40 PPM ± 3 % of Reading ; ± 10 μg/m³ ± 5% of Reading• Application : Industrial Process / Plant / IDC / Clean room / Storage / Hospital / Office building / Agriculture

DMB05

PM2.5 Air Quality Monitor

• Real-Time Data of PM2.5, CO2, temperature & humidity• Measuring Range : 0 ... 50 ℃ ; 0 ... 100 %RH ; 0 ... 2000 PPM ; 0 ... 500 μg/m³ • Accuracy : ± 0.4 ℃ ; ± 3 %RH ( at 20 ... 90 %RH ) ; ± 40 PPM ± 3 % of Reading ; ± 10 μg/m³ ± 5% of Reading• Application : Indoor air quaility / HVAC / Air ventilation system / Department store

TGP03 / THP03

2018 New Product Release

P01

Page 3: EYC 2018 brochure-en-0-3 · testing report DAQ & RS-485 HMI Dew Point Mirror RH System 973 Humidity Generator Thunder 3900 with Air Velocity ( 120 m/s ) Device under test (DUT) EYC

www.eyc-tech.com

2018 New Product Release

• Resistance : PT 100 Ω / PT 500 Ω / PT 100 Ω• Output : PT 100 Ω / 4 ... 20 mA• Accuracy : A ± ( 0.15 + 0.002 | t | ) B ± ( 0.3 + 0.005 | t | )• Ex d IIC T6 IP68 Ex tD A21 T100 ℃ IP68• Application : Dust environment / Chemical factory

EX temperature Transmitter

EX-R1ATEX

• Input : PT 100 Ω ( 3 wire )• Output : 4 ... 20 mA DC ( 2 wire )• Accuracy : Humidity : ± 2 %RH Temperature : ± 0.2 K at 23 °C• Operating Temperature Range : -40 ... +80 °C• Application : Aerospace Industry / Pharmaceutical Industry / Armament Industry / Metal Surface Grinding

Temperature & Humidity Transmitter

THS14EXATEX• 0 ... 10 bar for level detecting • Output : 4 ... 20 mA • Accuracy : ± 0.5 % of F.S.• Cable length : 3 ... 102 meter ( customize ) • Application : Drinking water equipment / Flow current

L053

Submersible Pressure Transmitter

• Measuring range : Water : 1 ... 200 cm/s Oil : 3 ... 300 cm/s Air : 20 ... 2000 cm/s• Supply voltage : 24 ± 20 %vdc• Current consumption : < 40 mA ( 24vdc )• Operating temperature : -40 ... 85 °C• Application : Ventilation / Hydraulic pressure

Thermal Flow Controller

FTC04

• Bearing : Stainless steel ball bearing / Tungsten arbide / Ceramic journal bearing• Low flow measurement• Accuracy : Better than ±1% of reading ± 0.5% / ±0.2% selectable • Nominal Diameter : DN4 ... DN40 • Application : Semiconductor integrated circuit Petrochemical / Pharmaceutical /

Turbine Flow Meter

FFM04

• Input type : Thermo Differential Pressure Module• Measuring range : 0 ... 500 pa• Accuracy ( at 25 ℃ ) : ± 0.5 % of F.S.• ModBus : RS-485• Physical switch : mbar / Pa / hPa / kPa / mmH2O / mmWS / inH2O / mmHg• Square root function • Application : Monitor differential pressure of Cleanroom / Filter environment / Air duct

Average Flow Measuring Tube + Differential Pressure Transmitter

AFMT+PHM33

P02

Page 4: EYC 2018 brochure-en-0-3 · testing report DAQ & RS-485 HMI Dew Point Mirror RH System 973 Humidity Generator Thunder 3900 with Air Velocity ( 120 m/s ) Device under test (DUT) EYC

www.eyc-tech.com

Temperature / Humidity

• PT100 / NTC10 K ; 0 ... 100 %RH• Output : temp. : PT100 Ω; humidity : 0 ... 10V ; 4 ... 20 mA• Accuracy : Humidity : ± 3 %RH of F.S. ; Temp. : ± 0.3 ℃ • RS-458 ; 0 ... 10V ( 3-wire ) ; 4 ... 20mA ( 2-wire )• Application : HVAC / Machine facility

THS17

Temperature & Humidity Transmitter

• 0 ... +50 ℃ ; 0 ... 100 %RH• Output : 4 ... 20 mA / 0 ... 10 V ; Optional output : passive temp. : PT100 / PT1000• Accuracy : Humidity : ±2 % RH ( 20 ~ 80 % RH ) ; Temperature : ± 0.3℃• Application : HVAC process / Air conditioning Environmental ventilation control

CMOS MEMS TechnologyMost Economy

THS130/140

Temperature & Humidity Transmitter

• 0 ... + 50 ℃ ; 0 ... 100 %RH• Analogue output / RS-485• Accuracy : Humidity : ± 3 % RH ( 20 ... 90 % RH ) ; Temperature : ± 0.4 ℃;• Application : HVAC building & HVAC air conditioning control

THR23

Temperature & Humidity Transmitter

• -40 ... +80 ℃ ; 0 ... 100 %RH• Analogue output / RS-485 ( Programmable )• Accuracy : Humidity : ± 2 %RH ;• Temp. : ± 0.15 ℃ ± 0.002 ℃ x Tactual• Application : Outdoor temp. & Humidity monitor

THS307

Temperature & Humidity Transmitter

• - 40 ... + 150 ℃ ; 0 ... 100 %RH ; 10 bar• Analogue output / RS-485 ( Programmable )• Accuracy : Humidity : ± 2 %RH;• Temp. : ± 0.15 ℃ ± 0.002 ℃ x Tactual• Application : Laboratory / Pharmaceutical

THS30X

Temperature & Humidity Transmitter

• -40 ... + 200 ℃ ; 0 ... 100 %RH ; 10 bar• Analogue output / RS-485 ( Programmable )• Accuracy : humidity : ± 1.2 %RH ; Temp. : ± 0.15 ℃ ± 0.002 ℃ x Tactual• Dew point, absolute humidity etc ...• Rugged aluminum alloy, IP67• Application : Semiconductor / Pharmaceutical / Industry

Temperature & Humidity Transmitter

THM80X

P03

Page 5: EYC 2018 brochure-en-0-3 · testing report DAQ & RS-485 HMI Dew Point Mirror RH System 973 Humidity Generator Thunder 3900 with Air Velocity ( 120 m/s ) Device under test (DUT) EYC

www.eyc-tech.com

Temperature / Humidity / Dew Point

• Input : PT 100 Ω • Output : 4 ... 20mA • Accuracy : ± 0.1 % of F.S.• Application : HVAC / Agriculture / Water power

Temperature Transmitter

TP02

• Input : PT 100 Ω• Output : 4 ... 20 mA • Accuracy : ± 0.1 % of F.S.• RTD Head-mounting• Application : Cooling system / Integrated with thermocouple

RTD Temperature Transmitter

TP01

• Input : PT 100 Ω • Output : 4 ... 20 mA , 2 wire• Accuracy : ± 0.5 % of F.S.• Integrated design with anti-condensation function• Application : Food / Pharmaceutical / Machinery manufacturing

Temperature Transmitter

TP04

• -80~+80 ℃ / -80~+60 dp℃• Analogue output / RS-485 ( Programmable )• Accuracy : Dew point:±3 dp ℃±(0.02%F.S./℃); Temp. : ±0.15℃±0.002℃ x Tactual Frost point, absolute humidity etc ...• Application : Industrial process / Pharmaceutical Industry

Dew Point Transmitter

THM86/87

• -100 ... + 60 dp℃ ; 16 bar• Analogue output • RS-485 ( Programmable )• Accuracy : ±2 dp ℃ ± ( 0.02%F.S./℃ )• Application : Adsorption / Freeze dryer / Dehumidifier

High Pressure Dew Point Transmitter

THS88

• 0 ... + 50 ℃ ; 0 ... 100 %RH ; 0 ...2000PPM• Output : 4 ... 20 mA / 0 ...10 V• Accuracy : Humidity : ± 3 % RH ( 20 ... 90 %RH ) ; Temperature : ± 0.4 ℃; CO2 : ± 40 PPM ± 3 % of reading• Application : Monitoring the air quality in indoor environments

THG03

Temperature & Humidity Transmitter

P04

Page 6: EYC 2018 brochure-en-0-3 · testing report DAQ & RS-485 HMI Dew Point Mirror RH System 973 Humidity Generator Thunder 3900 with Air Velocity ( 120 m/s ) Device under test (DUT) EYC

www.eyc-tech.com

Air velocity / Gas

Gas Monitoring Transmitter

• Input : T / H / CO / HCHO / VOC• Analogue output / RS-485 ( Programmable )• Accuracy : Temp. : ± 0.3 ℃ Humidity : ± 3 % RH CO : ± 3 % of F.S. HCHO : ±5 % of F.S. VOC :±5% of F.S.• Application : Monitor parking lot / Factory Air quality

GTH53

• Measuring media : air• Differential pressure transmitter• Average flow measuring tube length : 2 ... 40” • Operating pressure : max. 10 bar• Operating temperature : 250 ℃• Stainless steel 316• Application : Exhaust gas removal

Average Flow Measuring Tube +Differential Pressure Transmitter

AFMT+PMD33

P05

• 0 ... + 20 m/s;0 ... 10 V

• Accuracy : ± 5 % of F.S.

• Economical type

• Application : HVAC air process /

Clean room / Laminar flow

Hot wire Air Velocity Transmitter

FTS07

• 0 ... +20 m/s• Analogue output• Accuracy : ± 3 % of F.S.• IP54 • Switching analogue output by dip switch• Application : Light industry process / HVAC Building / Factory

FTS14

Air Velocity Transmitter

• 0 ... +40 m/s• Analogue output / RS-485 ( Programmable )• Accuracy : ± 2 % of F.S.• Auto-zero key, IP54 • Switch for physical quantity:[m/s]、[ft/s]...• Application : Light industry process / Pharmaceutical / Paper Industry

Air Velocity Transmitter

FTS34/35

• 0 ... + 120 m/s ; 0 ... + 120 ℃ • Analogue output / RS-485 ( Programmable ) • Accuracy : ± 1.5 % F.S. ; temp. : ± 0.3 ℃ • Rugged aluminum alloy, IP67• Application : Industrial process / Chemical / Pharmaceutical / Iron and steel industry

Thermo Air Velocity Transmitter

FTM84/85

Page 7: EYC 2018 brochure-en-0-3 · testing report DAQ & RS-485 HMI Dew Point Mirror RH System 973 Humidity Generator Thunder 3900 with Air Velocity ( 120 m/s ) Device under test (DUT) EYC

www.eyc-tech.com

Gas / Pressure / Signal

• Input : Current / Voltage / RTD / TC / Potentiometer / Linear Resistance• Local display, Setting by key pad• Analogue output / RS-485 / Frequency / Relay• Isolation : Power supply / Analogue Input & Output / RS-485• Application : Distribution cabinet / Industrial facility

Signal Converter

DPT02

• Input : Current / Voltage / RTD / TC Potentiometer / Linear Resistance / • Analogue output / RS-485 / Relay• Isolation : Power supply / Analogue Input & Output / RS-485• Display range : 5 digits, 7-segmant• Application : Distribution cabinet

Signal Display Monitor

DPM02

• 330° turnable ; the direction of panel is changeable; good monitoring • Intput : 4 ... 20 mA • Accuracy : 0.2 % of F.S.• Electrical connection : DIN 43650 plug / M12 connector• Application : Steel Industry / Water Plants Environmental Engineering

Integrated Signal Indicator

SD05

• Input : CO, 0 ... 500 PPM• Analogue output / RS-485 ( Programmable )• Accuracy : ± 3 % of F.S. • Application : Parking lot / Factory / Building

CO Transmitter

GM33/34

• CO2 : 0 ... 2000 PPM• Analogue output / RS-485 ( Programmable )• Accuracy : ± 40 PPM ± 3 % of reading• Non-dispersive Infrared ( NDIR )• Indoor monitor• Application : HVAC / Air conditioning / Museums / Supermarkets / Data centers / Wharehouses

CO2 Transmitter

GS23

• Input : CO2, 0 ... 10000 PPM• Analogue / RS-485 ( Programmable )• Accuracy : 400 ... 2000 PPM: ± 40 PPM ± 3 % of reading Other : ± 50 PPM ± 3 % of reading• Applicatrion : Parking lot / Greenhouse / Agriculture

CO2 Transmitter

GS43/44

P06

Page 8: EYC 2018 brochure-en-0-3 · testing report DAQ & RS-485 HMI Dew Point Mirror RH System 973 Humidity Generator Thunder 3900 with Air Velocity ( 120 m/s ) Device under test (DUT) EYC

www.eyc-tech.com

Factory Automatic QC Inspection System ( IEC 61298 )

Dew Point ( - 80 dp°C ... + 60 dp°C )

SD05+THS88 Dew Point Transmitter

PC-based & human machine interface Automatic QC inspection sheet printing

Automatic QCtesting report

DAQ & RS-485

HMI

Dew Point MirrorRH System 973

Humidity Generator

Thunder 3900

with

Air Velocity ( 120 m/s )

Device under test (DUT)EYC FTS84/85

DAQ

VFD ( Inverter)

PC

Inspection System

Wind Tunnel

P07

Page 9: EYC 2018 brochure-en-0-3 · testing report DAQ & RS-485 HMI Dew Point Mirror RH System 973 Humidity Generator Thunder 3900 with Air Velocity ( 120 m/s ) Device under test (DUT) EYC

www.eyc-tech.com

Calibration Range

Dew Point System

temp. : - 80 ... + 60 ℃

temp. : + 60 °C ... -80 °C

temp. : 0.14 °C ... 0.97 °C

Temperature System Humidity System

temp. : - 40 ... + 300 ℃Single point & Space calibration temp. : 0 ℃ ... 70 ℃ Humidity : 10 %RH ... 95 %RH

Single point

temp. : 0.11 °C ... 0.17 °C

Humidity : 0.11 %RH ... 0.55 %RH

temp. : 0.017 ... 0.094 ℃

FlukeMetrology Well

-95℃

Dew Point MirrorRH System 473

Dew Point MirrorRH System 473

Humidity GeneratorThnder 2500

Humidity GeneratorThunder 3900

CalibrationRange

CalibrationMeter

Calibration System

SmallestUncertainty

Chino standard Platinum

Thermometer

ILAC Mutual Recognition Arrangement Lab

Temperature & HumidityDew Point Calibration Lab

• International Lab Alliance Recognition ─ TAF ( ISO / IEC 17025 )

• Accreditation no. : 3032

• Fit for Global Application

• Comparable to National Grade Laboratory

• Professional calibration staff

• Wide range of high-precision measurement

♦ Temperature : - 40 ... + 300 °C

♦ Dew point / frost point : - 80 ... + 60 °C

♦ Relative humidity : 10 ... 95 % RH

• Extremely low calibration uncertainty, quickly repeat testing, apply to

the calibration of the temperature & humidity / dew point instrument

International recognition, fit for Global application

▲ TAF Certification

P08

Page 10: EYC 2018 brochure-en-0-3 · testing report DAQ & RS-485 HMI Dew Point Mirror RH System 973 Humidity Generator Thunder 3900 with Air Velocity ( 120 m/s ) Device under test (DUT) EYC

Put Pressure On

Pressure Generator

Expansion Platform0 ... 10 MPa

Additel ADT 681Digital Pressure Gauges

Data record

Check

STD.

Additel ADT 918 Pneumatic Pressure

Test Pump

+

Pressure System

Diffuser

Exhaust

Intake

Compressor

Velocity ControlData Record

Record / Control

Air Velocity System

YOKOGAWADiffrential Pressure Transmitter

Pitot tube

Air Velocity / Pressure Lab

│ Capability

│ Capability

Standard Check

0 ... 40 m/s eYc FTS84 Air Velocity Transmitter

40 ... 120 m/s Pitot Tube Differential Pressure Transmitter

0.25 ... 60 m/s

Calibration

Meter

Calibration

Range

Smallest

Uncertainty 0.04 m/s ... 0.90 m/s

Additel ADT 681

Digital Pressure Gauges

Wika Laboratory

Grade Pressure Gauge

Relative pressure 10 ... 7000 kPa

Absolute pressure 20 ... 170 kPa

Relative pressure 0.15 kPa

Absolute pressure 1.5kPa ... 7 kPa

Standard Check

Calibration

Meter

Calibration

Range

Smallest

Uncertainty

www.eyc-tech.comP09

Page 11: EYC 2018 brochure-en-0-3 · testing report DAQ & RS-485 HMI Dew Point Mirror RH System 973 Humidity Generator Thunder 3900 with Air Velocity ( 120 m/s ) Device under test (DUT) EYC

Gas / Flow Lab

│ Capability

│ Capability

Standard Check

Standard CO / CO2eYc CO / CO2

Transmitter

CO : 0 ... 500 μmol/mol(ppm)

CO2 : 0 ... 10000 μmol/mol(ppm)

Calibration

Meter

Calibration

Range

Smallest

Uncertainty

CO : 5.4 ... 23.4 μmol/mol

CO2 : 17 ... 274 μmol/mol

Standard Check

Calibration

Meter

Calibration

Range

Smallest

Uncertainty

YOKOGAWA

Magnetic

Flowmeter

TRICOR

Coriolis

Flow Meter

Flow Rate : 2.4 m3/h ... 30.0 m3/h ( 40 L/min ... 500 L/min)Velocity : 0.2 ... 3 m/s

Flow Rate : U = 0.6 %Velocity : U = 0.07 m/s

Standard CO & CO2

STD.Span Gas

CO / CO2

Regulator

Regulator

Chamber

DUT.

( Device Under Test )

Vogtlin Thermal Mass Flow Meters & Controllers• CO ( 1 L )• CO2 (1 L)• N2 ( 10 L )

eYc

CO / CO2 Transmitter

CHECK

1

Control Value

Control Value

Vent

By Pass

Gas Dilutor

Zero GasN2

Data Record

CO / CO2 / Gas System

Flow System

DUT.

( Device Under Test )

YOKOGAWA

Magnetic Flowmeter

STD.

MonitorTemperature & Pressureby eyc transmitter

Circulation Water Flow Direction

Tube 1.5”

TRICOR

Coriolis Flow Meter

CHECK

eyc Water FlowTransmitter

valve

Water Flow Control

Data Direction

Data Direction

Data Direction

Water Tank

eyc RTD TC & Heater

MonitorTemperature

Heater

Tube 3”

Data Record

www.eyc-tech.com P10

Page 12: EYC 2018 brochure-en-0-3 · testing report DAQ & RS-485 HMI Dew Point Mirror RH System 973 Humidity Generator Thunder 3900 with Air Velocity ( 120 m/s ) Device under test (DUT) EYC

Temperature &Humidity Transmitter

Multifunction Signal Converter

Dew PointTemperatureTransmitter

Air VelocityTransmitter

Multifunction Signal Display Monitor

CO / CO2Transmitter

Differential Pressure Transmitter

Industrial Factory / HVAC building / Smart City

Sensor / Transmitter / Signal converter & monitor

A.I. GatewayGateway

IIoT Hub Cloud Platform Cloud Storage

D.I. & A.I.

D.O. & A.O.

SCADAHDMI / MONITORModbus Master

Data LoggerTerminal

Monitor & Control

D.I. & A.I. D.O. & A.O.

WI-FIModbus TCP

MachineLearning

Edge Computing

RS-485 & WiFi, BT, ZigBee, Sub-G, LTE, 4G, 5G

www.eyc-tech.com

EYC AIoT Monitor Cloud System