+ All Categories
Home > Documents > F. Spurný, K. Brabco vá, I. Jadrníčková

F. Spurný, K. Brabco vá, I. Jadrníčková

Date post: 30-Dec-2015
Category:
Upload: quin-serrano
View: 18 times
Download: 1 times
Share this document with a friend
Description:
To the use of the linear energy spectrometer based on track etch detectors in radiotherapy proton beams, correlation with data measured by means of thermoluminescent detectors. F. Spurný, K. Brabco vá, I. Jadrníčková - PowerPoint PPT Presentation
Popular Tags:
20
To the use of the linear energy spectrometer based on track etch detectors in radiotherapy proton beams, correlation with data measured by means of thermoluminescent detectors F. Spurný, K. Brabcová, I. Jadrníčková Nuclear Physics Institute, Academy of Sciences of the Czech Republic, Na Truhlarce 39/64, Prague, [email protected]
Transcript
Page 1: F. Spurný, K. Brabco vá,  I. Jadrníčková

To the use of the linear energy spectrometer based on track etch detectors in

radiotherapy proton beams, correlation with data measured by means of

thermoluminescent detectors

F. Spurný, K. Brabcová, I. Jadrníčková

Nuclear Physics Institute, Academy of Sciences of the Czech Republic, Na Truhlarce 39/64, Prague, [email protected]

Page 2: F. Spurný, K. Brabco vá,  I. Jadrníčková

Table of contents

• Proton therapy beams at the Dzhelepov Laboratory of Nuclear Problems (DLNP), Joint Institute for Nuclear Research (JINR), Dubna, Russia – history, basic characteristics

• Basic approach to the dosimetry of beams; importance of secondary particles; Methods used

for the dosimetry of the DLNP JINR proton therapy beams

• Results of some dosimetry studies realised by means of active and passive dosimeters along the range, including Bragg peak area.

Page 3: F. Spurný, K. Brabco vá,  I. Jadrníčková

Beams available at the phasotron of the Dzhelepov Laboratory of Nuclear Problems - 1

• Clinical therapy started at Dubna at 1967, interrupted between 1974 and 1987,

• reconstructed to phasotron facility (see Table), six-cabin medical facility at the JINR Laboratory of Nuclear Problems was built:

Energy of accelerated protons (6596) MeV Energy dispersion (3.10.8) MeV Frequency of acceleration 250 Hz Ejected1) beam intensity (fast mode, pulse duration 30 μs) (2-2.5) μA Ejected1) beam intensity (slow mode, beam within 85% of modulation period duration (4 ms)

(1.6-2.0) μA

Ejected beam has a micro-structure: particle bunches of 10 ns duration

follow each other with an interval about 70 ns.

Page 4: F. Spurný, K. Brabco vá,  I. Jadrníčková

Beams available at the phasotron of the Dzhelepov Laboratory of Nuclear Problems - 2

• Four proton medical beams intended for irradiation of surface and deeply lying tumours with broad and narrow proton beams of various energies (between 100 and 660 MeV);

• A medical pion channel for beam therapy using high-intensity negative pion beams with energies between 30 and 80 MeV;

• A neutron channel for medical purposes (the average energy of neutrons in the beam is approximately 350 MeV) for irradiation of large radioresistant tumours.

Page 5: F. Spurný, K. Brabco vá,  I. Jadrníčková

Beams available at the phasotron of the Dzhelepov Laboratory of Nuclear Problems - 3

• Up to now, only proton beams are used, the total number of patients treated at the JINR proton beam is about 460 up to end of 2007 year. Typical dose rates used are of the order of several Gy per minute.

• The main importance has been since 2001 year devoted to the proton three-dimensional conformal radiation therapy and radiosurgery.

• At 2004 year a new proton beam, with the average energy 225 MeV, has been extracted to treat particularly prostate cancer. The ridge filter system permits to form a depth-dose distribution with Spread-Out-Bragg Peak with the width about 70 – 85 mm of water.

• Since 3-4 years, mostly head and neck cancers have been treated.

Page 6: F. Spurný, K. Brabco vá,  I. Jadrníčková

Basic approach to the dosimetry of beams

• Basic beam dosimetry and monitoring is assured with ionization chamber (participation NPI), silicon and diamond detectors; these dosimetric data were used to develop the technique of 3D conformal proton radiation therapy and its planning.

• Homogeneity of beams, their profiles and the importance of secondary particles in the proton energy transfer are followed also (participation NPI) by means of thermoluminescent detectors TLD’s and track etch detectors (TED’s) as spectrometers of linear energy transfer (LET)

Page 7: F. Spurný, K. Brabco vá,  I. Jadrníčková

LET spectrometer based on a PADC track-etch detectors

• polyallyldiglycolcarbonate (PADC) – C12H18O7 (CR-39)

– Page 0.5 mm (Page Moulgings Ltd, England),

– Tastrak 0.5 mm (Track Analysis Systems Ltd, Bristol)

• etching in 5 N NaOH at 70oC for 18 hours

• automatic optical image analyzer LUCIA Gtrack parameters => determination of the ratio vT/vB (vT –

etch rate of damaged material, vB – etch rate of

unaffected bulk material) => LET• range covered: ~ 10 to 700 keV/µm; 1 – 100 mSv

Page 8: F. Spurný, K. Brabco vá,  I. Jadrníčková

Calibration• LET = f(V)• irradiation in heavy charged particle beams at

– HIMAC (NIRS Chiba, Japan) in the frame of ICCHIBAN programs – ICCHIBAN 2, 4, 6, and 8;

– NASA Space Radiation Laboratory (Brookhaven National Laboratory) in the frame of ICCHIBAN BNL;

– Nuclotron of the Laboratory of High Energies, JINR, Dubna

• particles: 12C – 84Kr, LET ~ 7.5 – 600 keV/m• detection thresholds: Page ~ 10 keV/m, Tastrak ~ 15

keV/m, i.e. primary protons are not directly registered at the beam entrance

Page 9: F. Spurný, K. Brabco vá,  I. Jadrníčková

Examples of etched tracks

protons – Bragg peak Blind exposure No.3 ICCHIBAN 8

Matrjoshka – 9 months on ISS 16O (20) and 56Fe (402 keV/mm)

Page 10: F. Spurný, K. Brabco vá,  I. Jadrníčková

Microdosimetric distributions of dose in 205 MeV proton beam

Microdosimetric distributions

10

100

1000

10000

1 10 100 1000LET, keV/m

L*D

(L),

rela

tivel

y to

ent

ranc

e do

se

t=0 mm

t=40 mm

t=196 mm

t=233 mm

t= 267 mm

Microdosimetric distributions

0,00

0,03

0,06

0,09

0,12

1 10 100 1000LET, keV/m

L*D

(L),

rela

tivel

y to

poi

nt d

ose

t=0 mm

t=40 mm

t=196 mm

t=233 mm

t=267 mm

Remarks:1. When normalise to entrance dose – relative contribution of high LET

particles increases with the depth2. When normalise to the dose at point – importance of low LET particles

increases (Bragg peak – also primary protons)

Page 11: F. Spurný, K. Brabco vá,  I. Jadrníčková

Dose characteristics

DLET = ∫ (dN / dL) . L . dL

HLET = ∫ (dN / dL) . L . Q(L) .dL

BWED = ∫ (dN / dL) . L . r(L) .dL

RBWE = BWED / D

L – value of linear energy transferdN/dL – number of tracks N in a L interval dLQ(L) – ICRP 60 quality factorr(L) – biological weighting function

• protocols for D and H calculations • errors of doses and dose equivalents

– statistical and systematical uncertainties

BWE r(L) & Quality Factors Q(L)

0

6

12

18

24

30

10 100 1000LET, keV/m

Q(L

)

0

1

2

3

4

5

r(L

)

ICRP26

ICRP60

r(L)

Radiation quality change

1

1,04

1,08

1,12

1,16

1,2

1,24

1,28

0 50 100 150 200 250Residual proton energy, MeV

BW

E

1

1,6

2,2

2,8

3,4

4

4,6

5,2

Q

BWE-150

BWE-205

Q(60)-150

Q(60)-205

Page 12: F. Spurný, K. Brabco vá,  I. Jadrníčková

Thermoluminescent detectors

Materials of TLDs

• Al2O3:C, detection threshold (DT) for photons: ~1 µSv;

• Czech alumophosphate TL glasses (Al-P glass), DT for photons: ~ 10 µSv;

• CaSO4:Dy, DT for photons: ~ 1 µSv

• MCP700, LiF:Mg,Cu,P; DT for photons: ~ 1 µSv

Interpretation procedure – calibration performed in 60Co photon etalon beams in Kair ; readings transformed to Dtissue

Relative response (RR) to other radiationsRR defined relatively to the dose absorbed in tissue from studied radiation X and reference radiation :

RR = (TLreading/Dtissue)X/(TLreading/Dtissue)

i.e. as the ratio of TL signals after the exposure to the same dose in tissue due to a particle and due to 60Co photons

Page 13: F. Spurný, K. Brabco vá,  I. Jadrníčková

Relative response of TLD as a function of LET

Page 14: F. Spurný, K. Brabco vá,  I. Jadrníčková

Dosimetry in 145 MeV proton beam

• The detectors were positioned orthogonal to the beam direction;

• The absorbed doses in tissue Dpic were measured with precise ionisation chamber (PIC).

• To monitor the JINR beam, an ionisation chamber of the clinical dosimeter was used.

• The residual uncertainty in protons energy caused by protons range straggling ranged as Eaver ~ (2525) MeV for the Bragg peak region.

Depth Dose Distribution

0,0

0,5

1,0

1,5

2,0

2,5

100 105 110 115 120 125 130 135 140 145 150 155 160 165 170 175

t, mm of Water

Rela

tive D

ose

• Dpic were equal to (180±5) mGy at the beam entrance,

• at the Bragg peak region the doses were (180±5) mGy for luminescent detectors, (4.25±0.12) mGy for TED.

Page 15: F. Spurný, K. Brabco vá,  I. Jadrníčková

Results - TLD’sLD Dose in tissue, mGy Dose in tissue, mGy

145 MeV Bragg peak region MCP7 240 ± 12 153 ± 5

Al2O3:C 160 ± 22 89 ± 8 Al-P glass 180 ± 13 199 ± 7 CaSO4:Dy 166 ± 25 172 ± 12

145 MeV protons• The average value is, when uncertainties of individual values are not

considered, equal to (184±27) mGy, close to the reference value. • Nevertheless, the response MCP materials is clearly higher than for

other materials.

Bragg peak region• Spread of values is large, due to the different dependence of RR on

the LET of particle transferring energy. • For Al2O3:C as for MCP the RR decreases with LET faster than for

other two materials, therefore for average LET in Bragg region (~ 10 keV/μm) the decrease is sensible.

• For other two TLD’s RR’s are close to 1.00, their readings are close to the reference one.

Page 16: F. Spurný, K. Brabco vá,  I. Jadrníčková

Track etch detectors as spectrometers of LET – distributions of D, and H in LET

Protons - 0,5 mm Tastrak

0,00

0,05

0,10

0,15

0,20

0,25

100 1000 10000LET [MeV.cm2.g-1]

L*H

145 MeV (first layer)

145 MeV (second layer)

Bragg peak (first layer)

Bragg peak (secondlayer)

Page - protons

0,00

0,04

0,08

0,12

0,16

100 1000 10000

LET [MeV.cm2.g-1]

L*H

145 MeV (first layer) 145 MeV (second layer)

Bragg peak (first layer) Bragg peak (second layer)

Microdosimetric distributions of HLET, as measured in

proton beam with Tastrak PADC LET spectrometer. Microdosimetric distributions of HLET, as measured in

proton beam with Page PADC LET spectrometer.

Remarks: • For 145 MeV protons predominates in the H-distribution particles with LET higher than ~ 60 keV/µm, i.e. secondary particles formed through the nuclear interactions of primary protons.• At Bragg peak region, predominate particles with LET below 100 keV/µm, i.e. protons, both secondary ones as well as a part of primary protons with energies below the threshold of PADC Page detector, i.e. ~10 MeV.

Page 17: F. Spurný, K. Brabco vá,  I. Jadrníčková

Track etch detectors as spectrometers of LET – total dosimetric quantities

Remarks: • Doses due to higher LET secondary particles represent at the

beam entrance, i.e. 145 MeV protons, few % of the dose delivered through inelastic collisions of primary protons. The contribution is naturally more important for Page PADC, with lower LET threshold (~ 10 keV/µm, compared with ~ 15 keV/µm for Tastrak).

• At Bragg peak region, the contribution of particles with LET higher than the threshold represents tens of %, These tendencies determine also the values of quality factors. At the entrance they are close to values usual for neutrons, at Bragg peak region are for Page PADC much lower.

Quantity, Ep 145 MeV Bragg peak unit Page 1*) Page 2*) T0.5 Page 1 Page 2 T0.5 DLET,mGy 10.4±1.1 5.52±0.62 4.02±0.34 3.54±0.52 1.48±0.21 1.53±0.20 HLET, mSv 67.4±5.5 55.2±5.8 41.5±3.9 14.9±1.5 10.8±1.4 14.0±1.2 QF 6.5±0.9 10.0±1.6 10.3±1.3 4.2±0.8 7.3±1.4 9.3±1.5 DLET/Dtotal

0.058±0.006 0.031±0.003 0.022±0.002 0.42±0.06 0.17±0.02 0.18±0.02 *) Page 1 – above 10 keV/µm; Page 2 – above 15 keV/µm

Page 18: F. Spurný, K. Brabco vá,  I. Jadrníčková

Dependence of RR as a function of the residual energy in a high energy proton beam

Remarks:1. Negligible corrections needed for in-depth measurements with AlP

glass, resp. CaSO4:Dy,2. Ratio of Al2O3:C to other TLDs readings can give an estimation on

LET

TLD's RR to protons

0,4

0,7

1

1,3

1,6

0 30 60 90 120 150 180 210

Ep, MeV

RR

0,01

0,1

1

10

100

LE

T,

keV

/m

AlP glass

Al2O3:C

CaSO:Dy

LET

Page 19: F. Spurný, K. Brabco vá,  I. Jadrníčková

CONCLUSIONS

• TLD’s and TED based LET spectrometers proved their flexibility for the solution of many different dosimetry tasks.

• The importance of the energy transfer through the secondary higher LET particles has been qualified and quantified in high energy proton beams from the beam entrance up to the Brag beak region

Page 20: F. Spurný, K. Brabco vá,  I. Jadrníčková

Acknowledgements• Some part of studies presented in this contribution has

been realised in the frame of ESA DOBIES project. Authors are much obliged to Filip Vanhavere (SCK MOL, Belgium) for the coordination of the project, and to ESA Prodex Office for financial assistance.

Their thanks belong also to • A.G. Molokanov for the realisation of 145 MeV proton

irradiation• staffs of HIMAC (NIRS, Chiba), NSRL (BNL), and

Nuclotron (JINR, Dubna), and the organizers of the runs: Y. Uchihori, J. Miller, E.R. Benton, A.G. Molokanov, and V.P. Bamblevski†

• M. Gelev (INRNE BAS Sofia) for the fabrication and delivery of CaSO4:Dy TLD materials, and to

• Their colleagues at NPI AS CR for the help during the detectors treatment and development.


Recommended