+ All Categories
Home > Documents > Fast Screening for Non-Nutritive Sweeteners using …...Waters Corporation 34 Maple Street Milford,...

Fast Screening for Non-Nutritive Sweeteners using …...Waters Corporation 34 Maple Street Milford,...

Date post: 10-Jul-2020
Category:
Upload: others
View: 0 times
Download: 0 times
Share this document with a friend

Click here to load reader

Transcript
  • GOA L

    To rapidly analyze artificial sweeteners in

    beverages and canned fruits using SPE followed

    by the ACQUITY UPLC® H-Class System

    with ELSDs.

    BAC KG ROU N D

    The increase in obesity due to the consumption

    of sugary foods has reached epidemic levels,

    especially in the developed world. Food

    manufacturers are therefore resorting to non-

    caloric sweeteners to reverse this trend. Ideally,

    a blend of sweeteners will mimic the taste

    and sensation of sugar as closely as possible.

    Common sweeteners used today are aspartame,

    saccharin, acesulfame K, and sucralose. Less

    common substances include sodium cyclamate,

    NHDC (neohesperidin dihydrochalcone), and

    neotame. Regulations vary as to the legality

    of certain sweeteners in various parts of

    the world.

    Because of the different regulations, it is

    desirable to have a rapid method that can

    be used to look for the presence of these

    sweeteners. Acesulfame K, saccharin, and

    aspartame, along with caffeine and certain

    preservatives, are routinely analyzed in soft

    drinks using UV detection. However, cyclamate

    and sucralose are UV transparent rendering UV

    detection unfeasible.

    Run multiple samples in under five minutes using the

    ACQUITY UPLC H-Class System coupled with an ELSD.

    Fast Screening for Non-Nutritive Sweeteners using the ACQUITY UPLC H-Class System with ELSD

    Figure 1. ELSD Chromatogram of the UPLC separation of seven sweeteners.

    Minutes0.00 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40 2.60 2.80 3.00 3.20 3.40 3.60 3.80 4.00 4.20 4.40 4.60

    1

    76

    5

    4

    3

    2

    30 L

    SU

    1- AcesulfameK- 57.6 ppm2- Na_Saccharin- 87.6 ppm3- Na_Cyclamate- 172.8 ppm4- Sucralose- 69.6 ppm5- Aspartame- 166.8 ppm6- NHDC- 58.8 ppm7- Neotame- 56.4 ppm

    T H E SO LU T IO N

    Using a Waters® ACQUITY UPLC H-Class System equipped with an

    Evaporative Light Scattering Detector (ELSD), separation of the seven listed

    analytes can be achieved in less than five minutes, as shown in Figure 1.

    This method uses UPLC® HSS T3 Column chemistry and a ternary gradient.

    A calibration curve using a log linear fit is employed for quantification. This

    procedure can also be used for qualitative screening.

  • Waters Corporation34 Maple StreetMilford, MA 01757 U.S.A.T: 1 508 478 2000F: 1 508 872 1990www.waters.com

    For simple beverage samples, filtration and, if

    necessary, degassing to remove carbonation can be

    employed prior to direct injection of the samples.

    For more complicated matrices, a simple cleanup

    using Sep-Pak® C18 Cartridge can be used for

    sample preparation. This procedure is described in

    the European Commission’s validated method for

    the analysis of nine sweeteners (EUR 22727 EN,

    A. Wasik and M. Buchgraber, 2007). The sample,

    canned fruit in this example, was homogenized

    and thoroughly mixed with a buffer solution, then

    centrifuged. The supernatant was loaded onto a

    Sep-Pak C18 Cartridge, washed, and the analytes

    eluted with methanol. Chromatograms from a diet

    soft drink and canned diet pears carried through

    this procedure is shown in Figure 3. Three of the

    sweeteners were identified and quantified in the soft

    drink and two were identified and quantified in the

    canned pear sample.SUMMA RY

    Using the ACQUITY UPLC H-Class System with ELSD detection, baseline

    separation of seven sweeteners can be achieved in less than five minutes. The

    short run time allows many samples to be run either for screening purposes,

    or as a quantitative QC check.

    Minutes0.00 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40 2.60 2.80 3.00 3.20 3.40 3.60 3.80 4.00 4.20 4.40 4.60

    12

    3

    2

    1

    7 LS

    U7 LS

    U

    1-AcesulfameK- 79 ppm2- Sucralose- 59 ppm3- Aspartame- 214 ppm

    1-AcesulfameK- 42 ppm2- Sucralose- 6 ppm

    A: Diet Soft Drink

    B: Canned Diet Pears

    Figure 3. ELSD chromatograms of two samples following SPE cleanup. A: diet soft drink; B: canned diet pears. The calculated concentrations of each of the detected sweeteners is shown (with sample preparation adjustments accounted for in the final value).

    Figure 2. The ACQUITY UPLC H-Class System with ELSD detection.

    Waters, ACQUITY UPLC, UPLC, and Sep-Pak are registered trademarks of Waters Corporation. The Science of What’s Possible is a trademark of Waters Corporation. All other trademarks are the property of their respective owners.

    ©2011 Waters Corporation. Produced in the U.S.A.November 2011 720004166EN AO-PDF


Recommended