+ All Categories
Home > Documents > FDA – Industry Communication AdvaMed 2012 - Silverman.pdf · FDA-Industry Communication ......

FDA – Industry Communication AdvaMed 2012 - Silverman.pdf · FDA-Industry Communication ......

Date post: 11-Apr-2018
Category:
Upload: doancong
View: 219 times
Download: 3 times
Share this document with a friend
18
1 FDA-Industry Communication How to Get the Most Out of Your Courtship with FDA -An FDA Reviewer’s Perspective Society For Clinical Trials May 20, 2013 Phyllis M. Silverman, M.S. Biostatistics Team Leader FDA/CDRH
Transcript
Page 1: FDA – Industry Communication AdvaMed 2012 - Silverman.pdf · FDA-Industry Communication ... •Modular PMA’s •Q-Sub (formally Pre-IDE), Pre-PMA’s ... •Appropriateness of

1

FDA-Industry Communication

How to Get the Most Out of Your Courtship with FDA

-An FDA Reviewer’s Perspective

Society For Clinical Trials May 20, 2013

Phyllis M. Silverman, M.S. Biostatistics Team Leader

FDA/CDRH

Page 2: FDA – Industry Communication AdvaMed 2012 - Silverman.pdf · FDA-Industry Communication ... •Modular PMA’s •Q-Sub (formally Pre-IDE), Pre-PMA’s ... •Appropriateness of

Menu

• Venues of Communication

• Be Prepared to Discuss………..

• Top Ten Communication Crimes

2

Page 3: FDA – Industry Communication AdvaMed 2012 - Silverman.pdf · FDA-Industry Communication ... •Modular PMA’s •Q-Sub (formally Pre-IDE), Pre-PMA’s ... •Appropriateness of

Venues of Communication Non Verbal

• IDE’s, 510(K)’s, PMA’s, HDE’s

• Modular PMA’s

• Q-Sub (formally Pre-IDE), Pre-PMA’s

• Email correspondence

• eCopy

• Electronic Data Submission

3

Page 4: FDA – Industry Communication AdvaMed 2012 - Silverman.pdf · FDA-Industry Communication ... •Modular PMA’s •Q-Sub (formally Pre-IDE), Pre-PMA’s ... •Appropriateness of

4

Venues of Communication (cont’d)

Verbal

• Face to face meetings

– After pre-IDE review

– During IDE review cycle

– During PMA review cycle

• Teleconferences

Always have someone take minutes at meetings and telcons

Page 5: FDA – Industry Communication AdvaMed 2012 - Silverman.pdf · FDA-Industry Communication ... •Modular PMA’s •Q-Sub (formally Pre-IDE), Pre-PMA’s ... •Appropriateness of

5

Tips for Pre-IDE and IDE Meetings

• Submit protocol to FDA in advance

• Make it clear if your study is a feasibility or pivotal study and if the claim will be for superiority or NI

• Prepare an agenda or list of specific questions for FDA sent in advance

• Bring the necessary experts to the meeting, such as the statistical consultant

• Let FDA know if you are bring a

statistician so we can invite ours

Page 6: FDA – Industry Communication AdvaMed 2012 - Silverman.pdf · FDA-Industry Communication ... •Modular PMA’s •Q-Sub (formally Pre-IDE), Pre-PMA’s ... •Appropriateness of

For an IDE Pivotal Study Be prepared to discuss………..

• Sample size justification, assumptions, and rationale

• Hypotheses to be tested

• Appropriateness of the control

• Statistical methods that will be used

• Primary analysis cohort

6

Page 7: FDA – Industry Communication AdvaMed 2012 - Silverman.pdf · FDA-Industry Communication ... •Modular PMA’s •Q-Sub (formally Pre-IDE), Pre-PMA’s ... •Appropriateness of

Be prepared………..

• Important covariates and how they will be handled

• Randomization scheme

• Futility boundaries

• For Non-Inferiority claim: What advantage does the new device have that could justify a slightly lower success rate?

7

Page 8: FDA – Industry Communication AdvaMed 2012 - Silverman.pdf · FDA-Industry Communication ... •Modular PMA’s •Q-Sub (formally Pre-IDE), Pre-PMA’s ... •Appropriateness of

Have a Plan For………..

• Interim analyses including Type 1 error adjustment if necessary

• Subgroups analyses

• Handling multiplicity

• Minimizing lost to follow-up

• Handling missing data

• Appropriate simulations if Bayesian or adaptive design

8

Page 9: FDA – Industry Communication AdvaMed 2012 - Silverman.pdf · FDA-Industry Communication ... •Modular PMA’s •Q-Sub (formally Pre-IDE), Pre-PMA’s ... •Appropriateness of

9

Pre-PMA and PMA Meetings

• Pre-PMA – Brief FDA on highlights of study results and

anything unusual that might have happened

– Can occur before data analysis is complete

• Possibly even before the blind is broken

– Get FDA feedback on presentation of data tables and summaries

• Show table templates

Page 10: FDA – Industry Communication AdvaMed 2012 - Silverman.pdf · FDA-Industry Communication ... •Modular PMA’s •Q-Sub (formally Pre-IDE), Pre-PMA’s ... •Appropriateness of

10

For PMA Meetings Topics of Discussion……..

• Patient accountability

• Analysis cohorts

• Adjustment for multiplicity

• Clinical meaningfulness

of the results

• Protocol violations

Page 11: FDA – Industry Communication AdvaMed 2012 - Silverman.pdf · FDA-Industry Communication ... •Modular PMA’s •Q-Sub (formally Pre-IDE), Pre-PMA’s ... •Appropriateness of

11

PMA Topics of Discussion……

• Was there a physician or operator

learning curve?

• Patient compliance to treatment regimen

• Loss to Follow-up

• Potential biases

• Success of the masking

Page 12: FDA – Industry Communication AdvaMed 2012 - Silverman.pdf · FDA-Industry Communication ... •Modular PMA’s •Q-Sub (formally Pre-IDE), Pre-PMA’s ... •Appropriateness of

12

More possible topics……….

• Justification of pooling across sites

or countries

• Appropriateness of statistical assumptions

• Sensitivity analysis on missing values

• Robustness of the statistical inference

Page 13: FDA – Industry Communication AdvaMed 2012 - Silverman.pdf · FDA-Industry Communication ... •Modular PMA’s •Q-Sub (formally Pre-IDE), Pre-PMA’s ... •Appropriateness of

13

Top Ten Communication Crimes

1. Spending too much time at a meeting giving an overview of the device or clinical study

2. Spending too much time at a meeting discussing a single issue

3. Failure to address all FDA concerns

Page 14: FDA – Industry Communication AdvaMed 2012 - Silverman.pdf · FDA-Industry Communication ... •Modular PMA’s •Q-Sub (formally Pre-IDE), Pre-PMA’s ... •Appropriateness of

14

Top Ten Crimes

4. Not successfully communicating the study objective and statement of hypotheses

5. Not having someone oversee the clinical report to check for consistency and accuracy

6. Providing voluminous tables of computer printout without summary tables

Page 15: FDA – Industry Communication AdvaMed 2012 - Silverman.pdf · FDA-Industry Communication ... •Modular PMA’s •Q-Sub (formally Pre-IDE), Pre-PMA’s ... •Appropriateness of

15

Top Ten Crimes

7. Having poorly labeled tables without units, sample sizes, or analysis cohort

8. Conveying results in lengthy text rather than using tables or graphs

9. Failure to submit electronic data and software codes

Page 16: FDA – Industry Communication AdvaMed 2012 - Silverman.pdf · FDA-Industry Communication ... •Modular PMA’s •Q-Sub (formally Pre-IDE), Pre-PMA’s ... •Appropriateness of

16

Drum Roll………………..

10. Not coming in early to talk to FDA!

Page 17: FDA – Industry Communication AdvaMed 2012 - Silverman.pdf · FDA-Industry Communication ... •Modular PMA’s •Q-Sub (formally Pre-IDE), Pre-PMA’s ... •Appropriateness of

17

Don’t Hesitate - Communicate!

It’s the best way

to get FDA to say

to your proposal.

Page 18: FDA – Industry Communication AdvaMed 2012 - Silverman.pdf · FDA-Industry Communication ... •Modular PMA’s •Q-Sub (formally Pre-IDE), Pre-PMA’s ... •Appropriateness of

[email protected]

Thank You!

18


Recommended