18
INSTALLATION & SOFTWARE MANUAL FEVERSCREEN.COM.AU Automated Screening Technology FEVERSCREEN FEVERSCREEN
Transcript
  • INSTALLATION & SOFTWARE

    MANUAL

    FEVERSCREEN.COM.AU

    Automated Screening

    Technology

    FEVERSCREEN

    FEVERSCREEN

  • FEVERSCREEN

    The space between and immediately around Feverscreen and the individual to be

    screened is critical.

    Feverscreen, IEC and ISO provide a series of recommendations to help ensure best

    possible conditions to achieve more accurate screening results. These include:

    • Approach from 2m away remaining parallel to the terminal at all times

    • Faces must be parallel to unit when screening

    • Eyeglasses and facial covering must not obstruct face during screening

    • Ambient temperature must be controlled

    • Screening, if possible, to be conducted in a controlled indoor environment

    between 18-24°C

    • Ensure screening subject is not sweating at time of screening

    • Ensure screening subject has not recently washed their face

    • The screening unit itself must not be under

    o bright lights

    o air-conditioning

    o drafts

    o heat or cold sources

    • If a high temperature is detected follow up with a clinical device.

    Feverscreen is not a clinical device.

    • Ensure backdrop is Non-Reflective.

    • Avoid conducting screening in spaces which have dark backgrounds

    • Avoid conducting screening in dimly lit environments

    Installations musty comply to the above recommendations and be installed as per

    diagram 1

    What happens in and around this space is critical

    2

  • FEVERSCREEN

    37°C

    Setup the front of the cue to start 2m away from

    Feverscreen Unit

    Install unit as required to allow

    individuals to lean in and place

    face within onscreen outline from screening pt

    Create a screening line

    parallel to Feverscreen

    2.0m

    0.5m -0.75m

    0.0m

    1

    Setup Screening point marker

    approx 0.5-0.75m from Feverscreen

    3

    5

    4

    Setup appropriate social distancing cue markers at 1.5m intervals

    2

    SCREENING POINT

    FEVERSCREEN SCREENING STATION SETUPDIAGRAM 1

    Tilt (Bollard version only) to

    allow individuals to lean in

    appropriately.

    6

    3

  • FEVERSCREEN

    150mmBase Mounting

    Removable

    Cap

    Capping

    plate

    Specifications are subject to change. FeverScreen and Fever Screening Terminal products are sold with a limited 1year warranty. All

    rights reserved. FeverScreen and Fever Screening Terminal logo’s are trademarks or registered trademarks of FST Inc. All other

    trademarks are the property of their respective owner’s.

    13

    4

    2

    1. Remove cap from bollard

    2. Secure bollard to surface (or optional base plate) and run in cables

    3. Secure Feverscreen (Model: FST-021314-AU-B) to cap and attach cabling

    4. Attach cap to bollard

    5. Replace capping plate and power up

    Important Note: Approach and usage must be as per diagram 1. Approach must be

    parallel from 2m away.

    5

    FEVERSCREEN BOLLARD SETUP DIAGRAM 2

    Compatible Model: FST-021314-AU-B

    4

  • FEVERSCREEN

    Specifications are subject to change. FeverScreen and Fever Screening Terminal products are sold with a limited 1year warranty. All

    rights reserved. FeverScreen and Fever Screening Terminal logo’s are trademarks or registered trademarks of FST Inc. All other

    trademarks are the property of their respective owner’s.

    1

    2

    4

    1. Attach supplied wall mount securely to wall

    2. Attach cabling to Feverscreen unit FST-021314-AU-W

    3. Slide Feverscreen unit into wall mount

    4. Secure Feverscreen unit to wall mount with grub screw

    Important Note: Wall mounted version cannot be tilted. Approach and usage must be as

    per diagram 1. Approach must be parallel from 2m away.

    FEVERSCREEN WALL MOUNT SETUP DIAGRAM 3

    Compatible Model: FST-021314-AU-W

    3

    5

  • Input / Output

    NetworkNetwork is required for setting adjustment. Default address

    is 192.168.1.50

    Power 12v DC (2.5mm Tip positive)

    Relay2 pin terminal (N/O or N/C with time adjustment from 0 to

    10 seconds. Adjustable via software)

    I/O 4pin (I/O not available in current release)

    Technical Data - PortsFEVERSCREEN

    Network Power Relay I/O

    6

  • SOFTWARE

    MANUAL

    FEVERSCREEN.COM.AU

    Automated Screening

    Technology

    FEVERSCREEN

    FEVERSCREEN

  • FEVERSCREEN

    2

    SOFTWARE

    v2.3.1

    Automated Temperature Screening

    Technology

    Item Minimum Hardware Requirements

    Processor Intel i3 64-bit (x64) processor or equivalent

    Ram 4MB

    O/S Win10 (recommended)

    HDD (recommended) 64GB or higher recommended for database feature

    HDD (Software

    requirements)

    5MB of available space

    Hardware

  • Software & Settings FEVERSCREEN

    3

    Visitor Scan Function

    IP Address Set the IP address of your unit and press connect. As the first individual scans

    their image will appear in the first tile.

    Scanning As people scan, their image will appear in the respective tile in addition to the

    main tile bottom. The persons reading and STOP, GO or NO MASK result will

    also appear depending on selected options.

    History Panel The last eight people scanned will appear across the top of the history panel

    together with their temperature reading and time they were scanned. Tiles will

    recycle as

    People Counting People counter indicates how many people have scanned across the day.

    STOP: Indicates individual with abnormal

    temperature

    NO MASK: Indicates individual not wearing

    facemask

  • Software & Settings FEVERSCREEN

    Visitor History Function

    Date Range Date Range can be set by user to view respective records

    Info Panel Provides Record #, Device #, Date, Time, Mask Status and Green or Red level

    temp result

    Query After selecting date range user must click query to update results

    Export to Export to allows user to export excel spreadsheet of database. Default location

    or export is Windows Documents

    Pages Allows user to navigate across results

    4

  • Software & Settings FEVERSCREEN

    Settings Function

    IP Address Default IP address is 192.168.1.50

    Capture Interval Time required to successfully take temperature readings between scanning

    intervals. 3seconds should be used as a minimum.

    Picture Quality Picture quality required to perform a scan

    Picture Blur Picture blur level required to perform a scan

    Living Threshold Higher threshold requires satisfaction of additional AI evaluations to be

    considered human.

    Score Threshold Represents the cumulative satisfaction of all above variables as a %

    Recognizing Time How many times a person is scanned before a reading is provided. Use higher

    levels in harsh environments.

    Distance Distance persons are scanned from the unit. Longer distances will decrease

    accuracy.

    Exposure Lighting Exposure Level

    Temperature Threshold Temperature set point. Anything above this point will result in a fail result.

    Temp Offset Allows for a temperature offset in harsh environments. NOTE: incorrectly using

    this feature can result in false readings. Adjusting this feature is the responsibility

    of the user.

    Relay Switching Time Time relay unit will be active

    Closing State Select between Normal (will activate when temperature is normal) or Abnormal

    (will activate when temperature is high)

    OI Out Mode (Beta) N/A

    Temp Detect Mode Select between indoor or outdoor.

    5

  • Software & Settings FEVERSCREEN

    Feature Function

    Set Sound on Scanning Provides verbal notifications on monitoring PC

    Show Temperature for

    calibration

    Allows user to display temperature during visitor scans

    Monitor for Facemask Allows user to monitor for facemasks. When this is ticked field unit will remind

    users to wear facemask. In addition if an individual presents without a facemask

    an amber alert will appear in Visitor Scan with the label ‘No Mask”. If Mask email

    alerts are to be used this setting must be ticked.

    Save Changes Save changes must be clicked after each modification

    6

    Important information: Ports 80 and 8085 must be open for software to communicate

  • Software & Settings FEVERSCREEN

    7

    Time Function

    Time Sync When connected click Sync Now to synchronize time to local PC

  • Software & Settings FEVERSCREEN

    8

    Advanced Function

    Temperature When connected click Reset to calibrate temperature manually

    Reboot Click Reboot to remotely reboot device

    Application data path Allows user to set path for database C:\Feverscreen is default

  • Software & Settings FEVERSCREEN

    9

    Network Function

    IP Address Default IP address is 192.168.1.50

    Subnet Mask Allows setup of subnet mask to client requirements

    Default Gateway Allows setup of default gateway to client requirements

    Primary DNS Server Allows setup of Primary DNS server to client requirements

    Secondary DNS Server Allows setup of Secondary DNS server to client requirements

  • Software & Settings FEVERSCREEN

    10

    Light Function

    Mode Onboard lighting can be set to AUTO or Manual

    Manual Allows user to turn lights on in dimly lit environments to improve scanning quality

    Auto Allows user to set an auto on/off time dependant on ambient light levels

    Brightness Allows user to set the brightness level of the onboard lighting

  • Software & Settings FEVERSCREEN

    11

    Mail Function

    SMTP Server Enter SMTP server details

    Mail Address Enter From: Mail address

    Port Enter required port

    Username Enter username

    Password Enter password

    Send email alert on high

    temperature

    Sends email including picture when an individual is detected with abnormal

    temperature

    Send email alert on not

    wearing masks

    Sends email including picture when an individual is detected not wearing a

    facemask

    Test Email Sends a test email to confirm mail settings

    Save Changes Save changes must be clicked after each modification

  • FEVERSCREEN

    CAUTION

    In order for Feverscreen to maintain accuracy, units must be installed in a stable environment which

    does not fluctuate in temperature. ISO recommends indoor environments between 18-24° Celsius.

    External environments and environments with temperatures outside of this range must be within the

    specification of the unit and not be subject to sudden fluctuations in temperature.

    Feverscreen cannot be installed in environments effected by the below:

    1. In direct sun

    2. In any situation where sun is reflecting off surrounding

    surfaces (Feverscreen utilizes a highly sensitive thermal

    module and can pick up on reflected sun/heat sources

    from over 100m away)

    3. In instances where a persons face is reflecting sun. Both

    the thermal module and facial recognition module require a

    clear fully shaded view of the persons face. Where a

    person presents with a half shaded view of the face AI

    algorithms may be inhibited.

    4. In any situation where sun will influence the temperature of

    an individual or the unit itself

    1. In rain or wet environments

    2. Despite the unit being splash proof, any form of water or

    humidity over the thermal module will impact on the

    accuracy of results

    3. Water, dust or grime covering the lens of the facial

    recognition camera can impact on AI calculations and

    effect temperature readings

    1. Under air conditioning

    2. In areas effected by frost or mildew

    3. In any situation where the temperature of the unit can

    fluctuate significantly due to fluctuations in surrounding air

    temperature

    1. In wind or high flow environments

    2. In any situation where the temperature of the unit can

    fluctuate significantly due to winds or fans causing airflow

    between the unit and individual being screened.


Recommended