Fires Involving MRI Trailer Falmouth Hospital Training Guideline
Transcript
Page 1: Fires Involving Mri  Trailer

Fires Involving MRI Trailer

Falmouth Hospital

Training Guideline

Page 2: Fires Involving Mri  Trailer

By Capt Michael Small

FFD

Page 3: Fires Involving Mri  Trailer

Purpose:

ID hazards associated with fighting fires in MRI trailers

Page 4: Fires Involving Mri  Trailer

Topics of Discussion Approach Securing Electricity Securing Magnet Identifying “The Line in the Sand”

Page 5: Fires Involving Mri  Trailer

APPROACH

Page 6: Fires Involving Mri  Trailer
Page 7: Fires Involving Mri  Trailer

Fire protection Note location of FD connection There is NO fire protection inside the trailer

itself

Page 8: Fires Involving Mri  Trailer
Page 9: Fires Involving Mri  Trailer

ELECTRICAL SHUTOFF SECURING ELECTRICITY DOES NOT

DE-ENERGIZE THE MAGNET!!!!!!!!

Page 10: Fires Involving Mri  Trailer
Page 11: Fires Involving Mri  Trailer

HEILUM GAS In the event the magnet has been

quenched, the helium gas will be vented into the atmosphere, with an approx 500-1 expansion rate

Only a hazard in vented into a closed space (O2 displacement)

Will produce a HUGE white vapor cloud

Page 12: Fires Involving Mri  Trailer
Page 13: Fires Involving Mri  Trailer
Page 14: Fires Involving Mri  Trailer

QUENCHING Quenching refers to the emergency

dumping of Helium gas to de-energize the magnet

Can only be done from INSIDE the room where the magnet is

CANNOT be done by a firefighter wearing protective gear or anything metallic

Page 15: Fires Involving Mri  Trailer

ELECRICITY SECURING THE ELECTRICITY DOES

NOT DE-ENERGIZE THE MAGNET

Page 16: Fires Involving Mri  Trailer

Topic Two Details about this topic Supporting information and examples How it relates to your audience

Page 17: Fires Involving Mri  Trailer

The “Line in the Sand” No personnel wearing anything metallic can

cross this line Very strong magnetic field emanating 6 ‘ in

all directions from the magnet, 24/7/365 Any firefighting has to be done from beyond

this line

Page 18: Fires Involving Mri  Trailer

What This Means Add a strong statement that summarizes

how you feel or think about this topic Summarize key points you want your

audience to remember

Page 19: Fires Involving Mri  Trailer

Technician area View of working area of trailer Note ordinary combustibles


Recommended