FLOW CALIBRATORS MODEL 4048/4148 USER’S GUIDE P/N 6010051, REVISION A JULY 2016 Model 4048 Model 4148
Transcript

FLOW CALIBRATORS MODEL 4048/4148

USER’S GUIDE

P/N 6010051, REVISION A

JULY 2016

Model 4048

Model 4148

i

Copyright TSI Incorporated / 2016 / All rights reserved.

Address TSI Incorporated, 500 Cardigan Road, Shoreview, MN 55126 USA

W A R N I N G

TSI flow calibrators employ a heated platinum sensor. They

should not be used in the presence of flammable or explosive

gases or mixtures.

C a u t i o n

TSI flow calibrators are not medical devices under FDA 510k

and in no situation should they be utilized for human

respiration measurements.

LIMITATION OF WARRANTY AND LIABILITY (effective February 2015) (For country-specific terms and conditions outside of the USA, please visit www.tsi.com.)

Seller warrants the goods, excluding software, sold hereunder, under normal use and service as described in the operator's manual, to be free from defects in workmanship and material for 12 months, or if less, the length of time specified in the operator's manual, from the date of shipment to the customer. This warranty period is inclusive of any statutory warranty. This limited warranty is subject to the following exclusions and exceptions:

a. Hot-wire or hot-film sensors used with research anemometers, and certain other components when indicated in specifications, are warranted for 90 days from the date of shipment;

b. Pumps are warranted for hours of operation as set forth in product or operator’s manuals;

c. Parts repaired or replaced as a result of repair services are warranted to be free from defects in workmanship and material, under normal use, for 90 days from the date of shipment;

d. Seller does not provide any warranty on finished goods manufactured by others or on any fuses, batteries or other consumable materials. Only the original manufacturer's warranty applies;

e. This warranty does not cover calibration requirements, and seller warrants only that the instrument or product is properly calibrated at the time of its manufacture. Instruments returned for calibration are not covered by this warranty;

f. This warranty is VOID if the instrument is opened by anyone other than a factory authorized service center with the one exception where requirements set forth in the manual allow an operator to replace consumables or perform recommended cleaning;

g. This warranty is VOID if the product has been misused, neglected, subjected to accidental or intentional damage, or is not properly installed, maintained, or cleaned according to the requirements of the manual. Unless specifically authorized in a separate writing by Seller, Seller makes no warranty with respect to, and shall have no liability in connection with, goods which are incorporated into other products or equipment, or which are modified by any person other than Seller.

The foregoing is IN LIEU OF all other warranties and is subject to the LIMITATIONS stated herein. NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE. WITH RESPECT TO SELLER’S BREACH OF THE IMPLIED WARRANTY AGAINST INFRINGEMENT, SAID WARRANTY IS LIMITED TO CLAIMS OF DIRECT INFRINGEMENT AND EXCLUDES CLAIMS OF CONTRIBUTORY OR INDUCED INFRINGEMENTS. BUYER’S

ii

EXCLUSIVE REMEDY SHALL BE THE RETURN OF THE PURCHASE PRICE DISCOUNTED FOR REASONABLE WEAR AND TEAR OR AT SELLER’S OPTION REPLACEMENT OF THE GOODS WITH NON-INFRINGING GOODS.

TO THE EXTENT PERMITTED BY LAW, THE EXCLUSIVE REMEDY OF THE USER OR BUYER, AND THE LIMIT OF SELLER'S LIABILITY FOR ANY AND ALL LOSSES, INJURIES, OR DAMAGES CONCERNING THE GOODS (INCLUDING CLAIMS BASED ON CONTRACT, NEGLIGENCE, TORT, STRICT LIABILITY OR OTHERWISE) SHALL BE THE RETURN OF GOODS TO SELLER AND THE REFUND OF THE PURCHASE PRICE, OR, AT THE OPTION OF SELLER, THE REPAIR OR REPLACEMENT OF THE GOODS. IN THE CASE OF SOFTWARE, SELLER WILL REPAIR OR REPLACE DEFECTIVE SOFTWARE OR IF UNABLE TO DO SO, WILL REFUND THE PURCHASE PRICE OF THE SOFTWARE. IN NO EVENT SHALL SELLER BE LIABLE FOR LOST PROFITS, BUSINESS INTERRUPTION, OR ANY SPECIAL, INDIRECT, CONSEQUENTIAL OR INCIDENTAL DAMAGES. SELLER SHALL NOT BE RESPONSIBLE FOR INSTALLATION, DISMANTLING OR REINSTALLATION COSTS OR CHARGES. No Action, regardless of form, may be brought against Seller more than 12 months after a cause of action has accrued. The goods returned under warranty to Seller's factory shall be at Buyer's risk of loss, and will be returned, if at all, at Seller's risk of loss.

Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY AND LIABILITY, which contains the complete and exclusive limited warranty of Seller. This LIMITATION OF WARRANTY AND LIABILITY may not be amended, modified or its terms waived, except by writing signed by an Officer of Seller.

Service Policy Knowing that inoperative or defective instruments are as detrimental to TSI as they are to our customers, our service policy is designed to give prompt attention to any problems. If any malfunction is discovered, please contact your nearest sales office or representative, or call TSI's Customer Service department at (800) 874-2811 (USA) or (001 651) 490-2811 (International) or visit www.tsi.com.

Trademarks

TSI and TSI logo are registered trademarks of TSI Incorporated.

iii

CONTENTS CHAPTERS

Chapter 1 ............................................................................................ 5

Unpacking and Parts Identification .................................................. 5

Chapter 2 ............................................................................................ 7

Setting-Up ........................................................................................ 7 Assembly Instructions for the Flow Calibrators ............................ 7 Supplying Power ........................................................................... 8 Connecting Filter and Flow Tubes ................................................ 8

Chapter 3 .......................................................................................... 11

Operation ....................................................................................... 11 Overview ..................................................................................... 11 ON/OFF Switch .......................................................................... 11 Flow Rate Measurement ............................................................ 11 Setting Flow Rates and Calibrating Instruments ........................ 11

Chapter 4 .......................................................................................... 13

Maintenance .................................................................................. 13 Flow Sensor ................................................................................ 13 Re-certification ............................................................................ 13 Cases ......................................................................................... 13 Storage ....................................................................................... 13

Chapter 5 .......................................................................................... 15

Troubleshooting ............................................................................. 15

Technical Contacts ........................................................................ 16

Appendix A ...................................................................................... 17

Specifications................................................................................. 17

iv

(This page intentionally left blank)

5

Chapter 1

Unpacking and Parts Identification

Carefully unpack the instrument and accessories from the shipping

container. Check the individual parts against the list of components in

Table 1. If any parts are missing or damaged, notify TSI immediately.

Table 1-1. List of Components (see figures 1-1 and 1-2)

Qty Item Description Part/ Model

1 Flow Calibrator 4048 or 4148

1 Filter

Model 4048: 22 mm ISO taper

Model 4148: 63 mm HEPA filter with 0.25 in. (6.35 mm) fittings

1602292

1602317

1 Tubing Kit (Model 4048 only)

1 in. OD x ¾ in. ID – 24 in. (25.4 mm OD x 19.05 mm ID – 60.96 cm)

¾ in. x ½ in. ID – 24 in. and 6 in. (19.05 mm x 12.70 mm ID – 60.96 cm and 15.24 cm)

38 in. OD x ¼ in. ID – 6 in.

(9.53 mm OD x 6.35 mm ID – 15.24 cm)

516 in. OD x

316 in. ID – 24 in.

(7.92 mm OD x 4.75 mm ID – 60.96 cm)

3001185

3929420

3929408

3001248

1 Battery Pack 4199

1 Soft Carrying Case 1319289

1 Model 4048/4148 User’s Guide 6010051

1 AC Adapter

115 V, North America, ungrounded

100–240 V, NEMA 5-15 plug, grounded

100–240 V, Europlug, CEE 7/16, grounded

100–240 V, Great Britain, grounded, fused

100–240 V, Australia/NZ

2613033

8918-NA

8918-EC

8918-GB

8918-AT

6 Chapter 3

Figure 1-1. Model 4048 Components

1. Flow Calibrator 5. High-Efficiency Inlet Filter

2. Soft-Sided Carrying Case 6. Tubing Kit (see Table 1-1 for

diameters and length)

3. Universal Power Supply 7. Mounting Lugs (not shown)

4. User’s Guide 8. 6 AA-size Batteries (not shown)

Figure 1-2. Model 4148 Components

1. Flow Calibrator 5. High-Efficiency Inlet Filter

2. Soft-Sided Carrying Case 6. Mounting Lugs (not shown)

3. Universal Power Supply 7. 6 AA-size Batteries (not shown)

4. User’s Guide

Setting-Up 7

Chapter 2

Setting-Up

Assembly Instructions for the Flow Calibrators

The calibrator can be quickly attached to the battery pack by screwing

the mounting lugs into the threaded holes in the base of the calibrator.

The lugs should then be placed over the receiving holes in the top of

the battery pack and snapped gently into place. Insert the connector

from the battery pack to the receptacle in the back of the calibrator.

The calibrator is now ready to use.

Figure 2-1. Flow Calibrator Assembly

8 Chapter 3

Supplying Power

The calibrator can be powered in one of two ways: through the power

jack using the supplied 4199 battery pack or using the AC adapter.

The DC power input connector is shown below along with the power

requirements.

Power Supply: 7.5 VDC ± 1.5 V, 300 mA maximum

Connecting Filter and Flow Tubes

The Model 4048/4148 has a thermal sensor exposed to the air flow

that must be protected from foreign matter and particles. TSI supplies

a filter that should be connected to the inlet of the calibrator; however,

any filter will work as long as it has a minimum efficiency of 99.9%.

C a u t i o n

Always use a filter on the inlet of the calibrator. Failure

to filter the air flow may change the calibration and/or

permanently damage the sensor.

Note: Flow direction is identified by the large arrow printed on the bottom side of the calibrator and on the bottom of the battery pack.

Attach the filter to the inlet of the calibrator using supplied tubing

and/or adapters. Connecting a tube to the outlet of the calibrator will

create back pressure. See Appendix A for calibrator accuracy

specifications when operating at various pressures. In general,

minimize back pressure on the calibrator by using shorter lengths of

tubing to maintain highest level of accuracy.

Setting-Up 9

Figure 2-2. Typical Setup Configurations

10 Chapter 3

(This page intentionally left blank)

11

Chapter 3

Operation

Overview

The Model 4048/4148 Flow Calibrator measures volumetric flow rate

inside the flow tube. All measurements made by the Models

4048/4148 are NIST traceable.

ON/OFF Switch

Slide the switch to the ON position. The LCD displays volumetric flow

in units of liters per minute (L/min).

Flow Rate Measurement

Flow rate data will be displayed on the Model 4048/4148 through the

LCD.

Volumetric flow is displayed in units of liters per minute (L/min).

Setting Flow Rates and Calibrating Instruments

The Model 4048/4148 Flow Calibrator may be used to measure or set

flow rates on Condensation Particle Counters or other instruments

which draw an active flow sample. Be sure to follow all instrument

manufacturer’s flow setting and calibration procedures.

1. Start up the Model 4048/4148 Flow Calibrator and let it warm up

(1 minute for Model 4048 or 10 minutes for Model 4148) prior to

setting any flow rates or conducting any calibrations.

2. Connect the outlet of the Model 4048/4148 Flow Calibrator to the

inlet of the active sampling instrument.

NOTE: Please contact TSI Technical Support at 800-874-2811 for instructions on connecting Model 4048 to TSI Filter Testers.

3. Follow the instrument manufacturer’s flow setting and/or

calibration procedures. The volumetric flow in units of liters per

minute (L/min) will be continuously displayed on the LCD of the

calibrator.

12 Chapter 3

(This page intentionally left blank)

13

Chapter 4

Maintenance

Flow Sensor

Periodically inspect the flow sensor by looking into the outlet of the

calibrator. Remove dust, particles and fibers from the sensor with

clean, dry compressed air. The flow sensor may break if touched.

Never run liquids through the calibrator and never touch the

sensor with a brush. Dust or other deposits on the flow sensor will

degrade the Model 4048/4148’s flow accuracy.

C a u t i o n

The calibrator must be switched off for cleaning. Only

use clean, dry, compressed air when attempting to

remove contamination from the sensor.

Re-certification

To maintain a high degree of confidence in the measurements made

by the Model 4048/4148, TSI recommends that the instrument be

returned to TSI every 12 months for re-certification. The calibrator will

be recalibrated and returned with a certificate of calibration to US

National Institute of Standards Technology (NIST) traceable

standards.

Cases

If the instrument case or storage case needs cleaning, wipe it off with

a soft cloth dipped in isopropyl alcohol or mild detergent. Never

submerge the calibrator or allow liquids to enter the flow tube.

Storage

When storing the calibrator, always cover the ends of flow tubes with

the caps provided to prevent dust or other foreign matter from

entering the tube.

14 Chapter 4

(This page intentionally left blank)

15

Chapter 5

Troubleshooting

Table 3 lists the symptoms, possible causes, and recommended

solutions for common problems encountered with Model 4048/4148

Flow Calibrators. If the symptom is not listed, or if none of the

solutions solve the problem, please contact TSI Technical Support at

1-800-874-2811 or 651-490-2811.

Table 3. Troubleshooting

Symptom Possible Causes Corrective Action

No display. Unit not switched on. Switch on the unit.

No power to

instrument.

Check power connection.

Temperature

reads high at

low or zero

flows.

Temperature sensor is

being heated from the

flow sensor.

The temperature value

will track the actual air

temperature once the

flow rate through the

calibrator exceeds

1 L/min.

Flow readings

fluctuate greatly.

The flow is fluctuating. Improve inlet conditions.

Display shows

flows over-range

with no flow

passing through

flow tube.

The sensor may be

damaged or broken.

Return calibrator to TSI

for service.

16 Chapter 5

Technical Contacts

If you have any difficulty installing the Model 4048/4148, or if you

have technical or application questions about this instrument,

contact an applications engineer at TSI Incorporated, (651)

490-2811 or contact [email protected].

If the Model 4048/4148 fails, or if you are returning it for service,

visit our website at http://rma.tsi.com or contact TSI at:

TSI Incorporated

500 Cardigan Road

Shoreview, MN 55126 USA

Phone: +1-800-874-2811 (USA) or +1 (651) 490-2811

E-mail: [email protected]

TSI GmbH Neuköllner Strasse 4 52068 Aachen GERMANY

Telephone: +49 241-52303-0 Fax: +49 241-52303-49 E-mail: [email protected] Web: www.tsiinc.de

TSI Instruments Ltd. Stirling Road Cressex Business Park High Wycombe, Bucks HP12 3ST UNITED KINGDOM

Telephone: +44 (0) 149 4 459200 Fax: +44 (0) 149 4 459700 E-mail: [email protected] Web: www.tsi.com

17

Appendix A

Specifications

Specifications*

Flow Measurement Measurement Range Accuracy Air

Model 4048: 0 to 200 L/min Model 4148: 0 to 20 L/min Model 4048: ±2% of reading or 0.05 std. L/min,

whichever is greater Model 4148: ±2% of reading or 0.005 std. L/min,

whichever is greater at standard conditions (21.1°C and 101.3 kPa) See notes 1 through 5 below.

Instrument Temp. Range Operation, Ambient Storage, Ambient

0 to 50°C -20 to 60°C

Physical Dimensions External Dimensions Tube Adapters (Inlet & Outlet) Weight Flow Body Material

Model 4048: 7.2 in. × 2.5 in. × 2.1 in.

(18.3 cm × 6.4 cm × 5.3 cm) Model 4148: 5 in. × 2 in. × 1.25 in.

(12.7 cm × 5 cm × 3.2 cm) Model 4048: 0.5 inch O.D. straight Model 4148: 0.25 inch O.D. straight Model 4048: 1.2 lbs (0.5 kg) Model 4148: 1.7 lbs (0.8 kg) Polycarbonate

Warm-up Time Model 4048 – 1 minute Model 4148 – 10 minutes

Power Battery pack

7.5 VDC ± 1.5 V, 300 mA maximum

Notes: 1 Accuracy stated at standard conditions of 21.1°C and 101.3 kPa.

2 Accuracy stated with air temperature and flow body temperature within ±10°C of one another.

3 Accuracy stated measuring dry air (less than 10% R.H.).

4 Includes ±0.5% of reading repeatability.

*Specifications subject to change without notice.

18 Appendix A

Figure A-1. Model 4048 Dimensions

Figure A-2. Model 4148 Dimensions

TSI Incorporated – Visit our website www.tsi.com for more information. USA Tel: +1 800 874 2811 UK Tel: +44 149 4 459200 France Tel: +33 1 41 19 21 99 Germany Tel: +49 241 523030

India Tel: +91 80 67877200 China Tel: +86 10 8219 7688 Singapore Tel: +65 6595 6388

P/N 6010051 Rev. A ©2016 TSI Incorporated Printed in U.S.A.