CORESTA 10.23.2018 Frank S. Higgins, E. Kate Thorn, and Mark J. Rusyniak FTIR Chemometrics Applied to E-liquids Quantitative Analysis Altria Client Services l Frank Higgins l 10/23/2018 l 2018 CORESTA l 1
Transcript
Page 1: FTIR Chemometrics Applied to E-liquids Quantitative Analysis...CORESTA 10.23.2018 Frank S. Higgins, E. Kate Thorn, and Mark J. Rusyniak FTIR Chemometrics Applied to E-liquids Quantitative

CORESTA 10.23.2018

Frank S. Higgins, E. Kate Thorn, and Mark J. Rusyniak

FTIR Chemometrics Applied to E-liquids Quantitative Analysis

Altria Client Services l Frank Higgins l 10/23/2018 l 2018 CORESTA l 1

Page 2: FTIR Chemometrics Applied to E-liquids Quantitative Analysis...CORESTA 10.23.2018 Frank S. Higgins, E. Kate Thorn, and Mark J. Rusyniak FTIR Chemometrics Applied to E-liquids Quantitative

Altria Client Services l Frank Higgins l 10/23/2018 l 2018 CORESTA l 2

What is FTIR?

Fourier Transform Infrared (FTIR) spectroscopy• Measures the absorption of infrared light in the 2.5-25µm

wavelength range of the EM spectrum (4000-400cm-1)

Page 3: FTIR Chemometrics Applied to E-liquids Quantitative Analysis...CORESTA 10.23.2018 Frank S. Higgins, E. Kate Thorn, and Mark J. Rusyniak FTIR Chemometrics Applied to E-liquids Quantitative

Infrared Spectrum

Altria Client Services l Frank Higgins l 10/23/2018 l 2018 CORESTA l 3

Absorbance = 0 because

all light hits detector

Peaks appear where sample

absorbs light

Infrared

(light)

Source

Infrared

Detector

A spectrum is a graph of how much infrared light is absorbed by

molecules at each wavenumber of infrared light

Page 4: FTIR Chemometrics Applied to E-liquids Quantitative Analysis...CORESTA 10.23.2018 Frank S. Higgins, E. Kate Thorn, and Mark J. Rusyniak FTIR Chemometrics Applied to E-liquids Quantitative

Altria Client Services l Frank Higgins l 10/23/2018 l 2018 CORESTA l 4

What is FTIR

Advantages

• Specificity - fingerprint technique

• High optical throughput – all wavelengths of light are measured

simultaneously

• High signal to noise

• Fast

• Easy to use

Page 5: FTIR Chemometrics Applied to E-liquids Quantitative Analysis...CORESTA 10.23.2018 Frank S. Higgins, E. Kate Thorn, and Mark J. Rusyniak FTIR Chemometrics Applied to E-liquids Quantitative

FTIR is Trending

Altria Client Services l Frank Higgins l 10/23/2018 l 2018 CORESTA l 5

Advances in optics and chemometrics

• Industries are taking a second look at FTIR

• Replacing or supplementing existing techniques

Organizations accepting FTIR methods

• FDA

• US Pharmacopeia (USP)

• American Oil Chemist Society (AOCS)

• AOAC International

• ASTM International

• International Organization for Standardization (ISO)

Page 6: FTIR Chemometrics Applied to E-liquids Quantitative Analysis...CORESTA 10.23.2018 Frank S. Higgins, E. Kate Thorn, and Mark J. Rusyniak FTIR Chemometrics Applied to E-liquids Quantitative

Altria Client Services l Frank Higgins l 10/23/2018 l 2018 CORESTA l 6

Existing E-liquid Methods

Currently gas chromatography (GC) is the conventional method

to measure the principle components of e-liquids (CORESTA

CRM 84)

• Water, nicotine, propylene glycol (PG), vegetable Glycerin (VG),

menthol, and ethanol

• Ion chromatography (IC) for acids

• Karl Fischer (KF) is now the primary technique for water

FTIR e-liquids method includes all these components in a

single measurement

Page 7: FTIR Chemometrics Applied to E-liquids Quantitative Analysis...CORESTA 10.23.2018 Frank S. Higgins, E. Kate Thorn, and Mark J. Rusyniak FTIR Chemometrics Applied to E-liquids Quantitative

Altria Client Services l Frank Higgins l 10/23/2018 l 2018 CORESTA l 7

FTIR Advantages Over GC & KF

FTIR Advantages

• Better PG and VG measurements

• Improved low water measurements

• Low cost of operation

• No consumables

• No expensive gases

• No solvents

• Low instrument maintenance

• Low cost instrumentation

• No dilution or sample preparation

• No post analysis data processing

• Time per sample 2 minutes

Page 8: FTIR Chemometrics Applied to E-liquids Quantitative Analysis...CORESTA 10.23.2018 Frank S. Higgins, E. Kate Thorn, and Mark J. Rusyniak FTIR Chemometrics Applied to E-liquids Quantitative

Altria Client Services l Frank Higgins l 10/23/2018 l 2018 CORESTA l 8

Why FTIR for E-liquids Analysis

Ideal screening technique

• Reduces product development bottle necks

• Can be rolled out to high throughput manufacturing online or at-

line applications

Factory QA/QC• Incoming raw materials (i.e. Glycerol, PG)

• Could be used to quickly identify and possibly prevent out of

spec batches

Page 9: FTIR Chemometrics Applied to E-liquids Quantitative Analysis...CORESTA 10.23.2018 Frank S. Higgins, E. Kate Thorn, and Mark J. Rusyniak FTIR Chemometrics Applied to E-liquids Quantitative

Altria Client Services l Frank Higgins l 10/23/2018 l 2018 CORESTA l 9

Laboratory

Analyzers

Field Portable

Handheld

New FTIR Technology Available

Page 10: FTIR Chemometrics Applied to E-liquids Quantitative Analysis...CORESTA 10.23.2018 Frank S. Higgins, E. Kate Thorn, and Mark J. Rusyniak FTIR Chemometrics Applied to E-liquids Quantitative

Altria Client Services l Frank Higgins l 10/23/2018 l 2018 CORESTA l 10

Agilent Cary 630 FTIR Innovations

• Compact and portable

• Connects to most laptops

• True Michelson Interferometer

ensures full wavenumber

range capabilities (DTGS

Detector)

• High signal to noise –

equivalent performance to full

size FTIR instruments

Page 11: FTIR Chemometrics Applied to E-liquids Quantitative Analysis...CORESTA 10.23.2018 Frank S. Higgins, E. Kate Thorn, and Mark J. Rusyniak FTIR Chemometrics Applied to E-liquids Quantitative

TumblIR Liquids Cell

Altria Client Services l Frank Higgins l 10/23/2018 l 2018 CORESTA l 11

Single Transmission Cell

• Fixed path length liquid transmission cell• Standard 100µm

• Can be factory Set to 30µm

• Liquids only

• Quantitative analysis• 100 ppm to 100%

• Reproducible and easy to use

• Easy to clean

• Ideal for viscous liquids like e-liquids

Page 12: FTIR Chemometrics Applied to E-liquids Quantitative Analysis...CORESTA 10.23.2018 Frank S. Higgins, E. Kate Thorn, and Mark J. Rusyniak FTIR Chemometrics Applied to E-liquids Quantitative

TumblIR Liquids Cell

Altria Client Services l Frank Higgins l 10/23/2018 l 2018 CORESTA l 12

Cell empty - background Cell full – ready for scanning

Page 13: FTIR Chemometrics Applied to E-liquids Quantitative Analysis...CORESTA 10.23.2018 Frank S. Higgins, E. Kate Thorn, and Mark J. Rusyniak FTIR Chemometrics Applied to E-liquids Quantitative

TumblIR Liquids Cell

Cell open - cleaning

Altria Client Services l Frank Higgins l 10/23/2018 l 2018 CORESTA l 13

Page 14: FTIR Chemometrics Applied to E-liquids Quantitative Analysis...CORESTA 10.23.2018 Frank S. Higgins, E. Kate Thorn, and Mark J. Rusyniak FTIR Chemometrics Applied to E-liquids Quantitative

Traditional IR Liquid Transmission Cell

Altria Client Services l Frank Higgins l 10/23/2018 l 2018 CORESTA l 14

Filling: Messy Cleaning: Not always this bad

Page 15: FTIR Chemometrics Applied to E-liquids Quantitative Analysis...CORESTA 10.23.2018 Frank S. Higgins, E. Kate Thorn, and Mark J. Rusyniak FTIR Chemometrics Applied to E-liquids Quantitative

Calibrating the method

Altria Client Services l Frank Higgins l 10/23/2018 l 2018 CORESTA l 15

Partial Least Squares (PLS) Chemometrics

• Great for complex mixtures

• Allows multiple regions of the IR spectrum to be used in the calibration

• Preprocessing algorithms can be used to improve the calibration

• Mean centering

• First and second derivatives

• Ideal for calibrations using bands that over lap

• Standard Error of Cross Validation (SECV) is a PLS feature to evaluate

the performance of the calibration.

Large sets of e-liquids are measured

• Known values for each analyte are obtained

• KF, GC, and IC values are used for each sample in the calibration

• Preprocessing and regions are selected to optimize each analyte

Page 16: FTIR Chemometrics Applied to E-liquids Quantitative Analysis...CORESTA 10.23.2018 Frank S. Higgins, E. Kate Thorn, and Mark J. Rusyniak FTIR Chemometrics Applied to E-liquids Quantitative

FTIR PLS Spectral Overlay

Altria Client Services l Frank Higgins l 10/23/2018 l 2018 CORESTA l 16

OH stretch,

water +

PG/VG OH CH stretch,

methyl/methylene

Water, HOH

deformation

C-O

stretch,

PG / VG

Ionized

carboxyl

Nicotine

Wavenumber (cm-1)

Absorb

ance

Page 17: FTIR Chemometrics Applied to E-liquids Quantitative Analysis...CORESTA 10.23.2018 Frank S. Higgins, E. Kate Thorn, and Mark J. Rusyniak FTIR Chemometrics Applied to E-liquids Quantitative

FTIR PLS Spectral Overlay (Fingerprint Region)

Altria Client Services l Frank Higgins l 10/23/2018 l 2018 CORESTA l 17

VG

PG

Nicotine

Nicotine

Acids, ionized carboxyl

Water

Wavenumber (cm-1)

Absorb

ance

Page 18: FTIR Chemometrics Applied to E-liquids Quantitative Analysis...CORESTA 10.23.2018 Frank S. Higgins, E. Kate Thorn, and Mark J. Rusyniak FTIR Chemometrics Applied to E-liquids Quantitative

Water: FTIR PLS Actual vs. Predicted Calibration Plot

Altria Client Services l Frank Higgins l 10/23/2018 l 2018 CORESTA l 18

0

6

12

18

24

30

Pre

dict

ed C

once

ntra

tion

{F:5

C:W

ater

pct

}

6in1

C63

0 W

ater

MY

024

16um

rep

44

-2 6 14 22 30

Actual Concentration {C:Water pct}

2-Actual vs Predicted

1

2

3

5678

910

1112

13

15161719 20

2223

24

25

26

38

39404142

44

4647484950

51

52

54

59

60

63

64

65

6667

68

6970

7172

73

75

76

77

78

79

8182838485

86

87

888990

91929395

96

97

98

100

101

102103104

106107108109110

111112113

114115116

117118119

120121122

123124125

126127128129130131

132133134

135136137

138139140

161162163

164165166

167168169

170171172

173174175

176177178

179180181182183184

185186187188189190

226227228229230231

R² = 0.99955183

Actual Concentration (Water %)

Pre

dic

ted C

on

centr

atio

n (

Wate

r %

)

R2= 0.99955

Page 19: FTIR Chemometrics Applied to E-liquids Quantitative Analysis...CORESTA 10.23.2018 Frank S. Higgins, E. Kate Thorn, and Mark J. Rusyniak FTIR Chemometrics Applied to E-liquids Quantitative

FTIR Method Accuracy and Precision: Low Water

Altria Client Services l Frank Higgins l 10/23/2018 l 2018 CORESTA l 19

Accuracy - Individual aliquots of the same e-liquid sample were measured (N=10). The sample was

measured by validated “gold standard” techniques as shown in the “Actual %” column. The percent

water LOD (3x SD) and LOQ (9x SD) are shown.

Precision – The same aliquot of E-Liquid 1 measured consecutively (N=9).

E-Liquid 1 Average (%) St.Dev. (%) %RSD

Actual %

(GC, KF, IC)

Accuracy (%

Difference) LOD % LOQ %

KF St.Dev.

(%) KF %RSD

Water (%) 0.147 0.010 7.080 0.144 1.883 0.031 0.094 0.032 20.071

Propylene Glycol (%) 37.692 0.102 0.270 37.853 -0.427

Glycerol (%) 57.504 0.181 0.314 57.598 -0.163

Nicotine (wt %) 2.486 0.015 0.607 2.461 1.028

Menthol (%) 0.000 0.000 NA 0.000 NA

EtOH (%) 0.617 0.010 1.646 0.614 0.456

Acids (%) 0.000 0.000 NA 0.000 NA

E-Liquid 1 Average (%) St.Dev. (%) %RSD

Water (%) 0.157 0.002 1.474

Propylene Glycol (%) 37.627 0.039 0.104

Glycerol (%) 57.419 0.067 0.116

Nicotine (wt %) 2.489 0.016 0.637

Menthol (%) 0.000 0.000 NA

EtOH (%) 0.607 0.004 0.717

Acids (%) 0.000 0.000 NA

Page 20: FTIR Chemometrics Applied to E-liquids Quantitative Analysis...CORESTA 10.23.2018 Frank S. Higgins, E. Kate Thorn, and Mark J. Rusyniak FTIR Chemometrics Applied to E-liquids Quantitative

FTIR Method Repeatability Test: High Water

Altria Client Services l Frank Higgins l 10/23/2018 l 2018 CORESTA l 20

Individual Aliquots = Same E-liquid 2 sample with replicate ~0.25mL aliquots measured.

Same Aliquot = Same E-liquid 2 sample and aliquot measured 5 consecutive times.

E-liquid 2 Air Exposure Experiment

Individual Aliquots

E-Liquid 2

Rep1

E-Liquid 2

Rep2

E-Liquid 2

Rep3

E-Liquid 2

Rep4

E-Liquid 2

Rep5 Average Std. Dev. %RSD

KF Average

(%)

KF St.

Dev. (%) KF %RSD

Water (%) 15.590 15.640 15.540 15.600 15.610 15.596 0.036 0.234 15.583 0.107 0.682

Same Aliquot

E-Liquid 2

Rep1

E-Liquid 2

Rep2

E-Liquid 2

Rep3

E-Liquid 2

Rep4

E-Liquid 2

Rep5 Average Std. Dev. %RSD

Water (%) 15.590 15.600 15.590 15.580 15.590 15.590 0.007 0.045

Same Aliquot

E-Liquid 2

Avg

E-Liquid 2

30 Mins

E-Liquid 2

3hours

Water (%) 15.60 12.36 11.96

Propylene Glycol (%) 23.98 24.78 24.82

Glycerol (%) 55.89 57.80 58.22

Nicotine (wt %) 3.64 3.84 3.86

Menthol (%) 0.00 0.00 0.00

EtOH (%) 0.84 0.52 0.39

Acids (%) 0.74 0.87 0.88

Page 21: FTIR Chemometrics Applied to E-liquids Quantitative Analysis...CORESTA 10.23.2018 Frank S. Higgins, E. Kate Thorn, and Mark J. Rusyniak FTIR Chemometrics Applied to E-liquids Quantitative

FTIR Method Accuracy and Precision: Propylene Glycol

The lowest PG calibration sample (E-liquid 3) was measured to

determine LOD and LOQ (N=6).

Altria Client Services l Frank Higgins l 10/23/2018 l 2018 CORESTA l 21

E-Liquid 3Average

(%)

St.Dev.

(%)%RSD

Actual %

(GC, KF, IC)

Accuracy (%

Difference)LOD LOQ

Water (%) 15.565 0.023 0.149 15.583 -0.119

Propylene Glycol (%) 23.336 0.020 0.088 23.386 -0.215 0.061 0.184

Glycerol (%) 56.339 0.067 0.118 56.644 -0.538

Nicotine (%) 3.502 0.021 0.614 3.490 0.339

Menthol (%) 0.000 NA NA NA NA

EtOH (%) 0.736 0.009 1.233 0.754 -2.431

Acids (%) 0.706 0.002 0.237 0.723 -2.284

Page 22: FTIR Chemometrics Applied to E-liquids Quantitative Analysis...CORESTA 10.23.2018 Frank S. Higgins, E. Kate Thorn, and Mark J. Rusyniak FTIR Chemometrics Applied to E-liquids Quantitative

FTIR Method Accuracy and Precision: Glycerol

The lowest glycerol calibration sample (E-liquid 4) was measured to

determine LOD and LOQ (N=6).

Altria Client Services l Frank Higgins l 10/23/2018 l 2018 CORESTA l 22

E-Liquid 4Average

(%)

St.Dev.

(%)%RSD

Actual %

(GC, KF, IC)

Accuracy (%

Difference)LOD LOQ

Water (%) 15.060 0.025 0.164 15.000 0.398

Propylene Glycol (%) 66.719 0.100 0.150 67.950 -1.812

Glycerol (%) 17.417 0.016 0.092 17.050 2.153 0.048 0.144

Nicotine (%) 0.000 NA NA NA NA

Menthol (%) 0.000 NA NA NA NA

EtOH text 0.000 NA NA NA NA

Acids (%) 0.000 NA NA NA NA

Page 23: FTIR Chemometrics Applied to E-liquids Quantitative Analysis...CORESTA 10.23.2018 Frank S. Higgins, E. Kate Thorn, and Mark J. Rusyniak FTIR Chemometrics Applied to E-liquids Quantitative

FTIR Method Accuracy and Precision: Nicotine and Menthol

The lowest nicotine and menthol calibration sample (E-liquid M4)

was measured to determine LOD and LOQ (N=6).

Altria Client Services l Frank Higgins l 10/23/2018 l 2018 CORESTA l 23

E-Liquid M4Average

(%)

St.Dev.

(%)%RSD

Actual %

(GC, KF, IC)

Accuracy (%

Difference)LOD LOQ

Water (%) 3.487 0.012 0.334 NA NA

Propylene Glycol (%) 53.147 0.058 0.109 52.993 0.290

Glycerol (%) 40.659 0.051 0.125 41.273 -1.487

Nicotine (%) 0.580 0.021 3.679 0.599 -3.172 0.064 0.192

Menthol (%) 0.542 0.025 4.567 0.481 12.613 0.074 0.223

EtOH (%) 0.169 0.007 4.253 0.208 -18.590

Acids (%) 0.000 NA NA NA NA

Page 24: FTIR Chemometrics Applied to E-liquids Quantitative Analysis...CORESTA 10.23.2018 Frank S. Higgins, E. Kate Thorn, and Mark J. Rusyniak FTIR Chemometrics Applied to E-liquids Quantitative

FTIR Method Accuracy and Precision: Ethanol

The lowest ethanol calibration sample (E-liquid M17) was measured

to determine LOD and LOQ (N=6).

Altria Client Services l Frank Higgins l 10/23/2018 l 2018 CORESTA l 24

E-Liquid M17Average

(%)

St.Dev.

(%)%RSD

Actual %

(GC, KF, IC)

Accuracy (%

Difference)LOD LOQ

Water (%) 11.571 0.028 0.238 11.253 2.825

Propylene Glycol (%) 27.031 0.032 0.117 27.197 -0.609

Glycerol (%) 56.097 0.058 0.104 56.429 -0.588

Nicotine (%) 3.674 0.016 0.439 3.661 0.334

Menthol (%) 0.000 NA NA NA NA

EtOH (%) 0.146 0.006 3.930 0.155 -5.853 0.017 0.051

Acids (%) 0.713 0.003 0.359 0.713 -0.038

Page 25: FTIR Chemometrics Applied to E-liquids Quantitative Analysis...CORESTA 10.23.2018 Frank S. Higgins, E. Kate Thorn, and Mark J. Rusyniak FTIR Chemometrics Applied to E-liquids Quantitative

FTIR Method Accuracy and Precision: Total Acids

The lowest total acids calibration sample (E-liquid M21) was measured

to determine LOD and LOQ (N=6).

Altria Client Services l Frank Higgins l 10/23/2018 l 2018 CORESTA l 25

E-Liquid M21Average

(%)

St.Dev.

(%)%RSD

Actual %

(GC, KF, IC)

Accuracy (%

Difference)LOD LOQ

Water (%) 0.919 0.010 1.042 0.889 3.337

Propylene Glycol (%) 36.719 0.050 0.137 37.138 -1.128

Glycerol (%) 56.868 0.082 0.145 57.520 -1.132

Nicotine (%) 2.526 0.014 0.563 2.541 -0.597

Menthol (%) 0.000 NA NA NA NA

EtOH (%) 0.566 0.005 0.960 0.583 -3.033

Acids (%) 0.050 0.001 1.690 0.048 3.452 0.003 0.008

Page 26: FTIR Chemometrics Applied to E-liquids Quantitative Analysis...CORESTA 10.23.2018 Frank S. Higgins, E. Kate Thorn, and Mark J. Rusyniak FTIR Chemometrics Applied to E-liquids Quantitative

FTIR Method Calibration Summary

Altria Client Services l Frank Higgins l 10/23/2018 l 2018 CORESTA l 26

E-Liquids SECV % R2 LOD % LOQ %

Water (%) High 0.156 0.9996 NA NA

Water (%) low 0.094 0.9985 0.031 0.094

Propylene Glycol (%) 0.305 0.9994 0.061 0.184

Glycerol (%) 0.510 0.9990 0.048 0.144

Nicotine (wt %) 0.037 0.9993 0.064 0.192

Menthol (%) 0.052 0.9983 0.074 0.223

EtOH (%) 0.030 0.9950 0.017 0.051

Acids (%) 0.014 0.9974 0.003 0.008

Page 27: FTIR Chemometrics Applied to E-liquids Quantitative Analysis...CORESTA 10.23.2018 Frank S. Higgins, E. Kate Thorn, and Mark J. Rusyniak FTIR Chemometrics Applied to E-liquids Quantitative

Conclusions

Altria Client Services l Frank Higgins l 10/23/2018 l 2018 CORESTA l 27

• Current FTIR technology makes analysis of viscous liquids, such

as e-liquids, faster and easier.

• FTIR methods can provide immediate results for e-liquids

research and manufacturing applications.

• FTIR analysis is a less expensive and easier to maintain

compared to other techniques.

• Chemometric PLS calibration to “gold standard” techniques, such

as GC, KF, and IC, produces accurate and precise measurement

of e-liquids by FTIR.

• FTIR e-liquids results indicate comparable performance to GC,

KF, and IC techniques, and in some cases better performance.

Page 28: FTIR Chemometrics Applied to E-liquids Quantitative Analysis...CORESTA 10.23.2018 Frank S. Higgins, E. Kate Thorn, and Mark J. Rusyniak FTIR Chemometrics Applied to E-liquids Quantitative

Acknowledgments

Altria Client Services l Frank Higgins l 10/23/2018 l 2018 CORESTA l 28

• Special Thanks for the Karl Fischer, GC, and IC data, samples,

and background knowledge of e-liquids formulations

• Robert Ragland

• Sean Platt

• Charnise Jackson

• Xue Luo

• Geniya Prepelitskaya

Page 29: FTIR Chemometrics Applied to E-liquids Quantitative Analysis...CORESTA 10.23.2018 Frank S. Higgins, E. Kate Thorn, and Mark J. Rusyniak FTIR Chemometrics Applied to E-liquids Quantitative

Altria Client Services l Frank Higgins l 10/23/2018 l 2018 CORESTA l 29

For copies of this presentation visit the Altria’s Science Website at

www.altria.com/alcs-science


Recommended