+ All Categories
Home > Documents > FTIR SPECTROSCOPY REFERENCE GUIDE - Agilent · PDF fileFTIR SPECTROSCOPY REFERENCE GUIDE...

FTIR SPECTROSCOPY REFERENCE GUIDE - Agilent · PDF fileFTIR SPECTROSCOPY REFERENCE GUIDE...

Date post: 12-Mar-2018
Category:
Upload: phungdat
View: 250 times
Download: 4 times
Share this document with a friend
1
FTIR SPECTROSCOPY REFERENCE GUIDE INTRODUCTION Infrared spectroscopy is the study of the interactions between infrared electromagnetic energy and matter. The technique of infrared spectroscopy measures the vibrations of molecules, allowing for qualitative and quantitative measurements of samples. A Fourier transform infrared (FTIR) spectrometer is an ideal tool for the identification of unknown organic and inorganic samples whether they exist in the form of a gas, liquid or a solid. ELECTROMAGNETIC SPECTRUM Radiation in the infrared region is commonly referred to in terms of a unit called a wavenumber (ν), rather than wavelength (λ). Wavenumbers are expressed as reciprocal centimeters (cm -1 ) and are the preferred unit as they are directly proportional to energy. A higher wavenumber corresponds to a higher energy. Convert a wavenumber (ν) to a wavelength (λ) by using the following formula: Convert a wavenumber (ν) to Energy (E) by multiplying it by the speed of light (c, in centimeters per second) and Planck's constant (h): SPECTRAL RANGE COVERAGE Spectral range coverage for common sources, beamsplitters, and detectors. The combination of these components generate a spectrometer’s working scan range. First commercial FTIR spectrometer A HISTORY OF COMMERCIAL 'FIRSTS' A portion of the electromagnetic spectrum showing the relationship of the infrared region to other types of radiation (not shown to scale). The relationship between energy, wavenumber, and wavelength is also highlighted. COLLECTING AN FTIR SPECTRUM There are three quick, simple steps involved in obtaining a spectrum of a sample: STEP 1 Record a spectrum with no sample present. (Known as a ‘background’.) STEP 2 Insert the sample into the spectrometer. STEP 3 Record a second spectrum and interpret the data. The key collection parameters are: • The number of scans (background and sample) • The scan range • The spectral resolution. In general increasing the number of scans that are co-added improves the signal-to-noise ratio of the spectrum and can assist in the analysis of weakly absorbing samples. The spectral resolution is user-defined and helps to distinguish closely spaced absorption peaks, and is expressed in wavenumbers. The simplified correlation table on the right allows users to extract structural information from IR spectra. Computer-based search programs are also available for assisting in compound identification. CARY 600 SERIES SPECTROMETER OPTICS ν (cm -1 ) = 1 λ (cm) E (J) = h (J.s) x c (cm.s -1 ) x ν (cm -1 ) High High Gamma Energy Wavenumber (ν) Low Low Short Wavelength (λ) Long (cm -1 ) 4000 400 X-Rays UV Visible Near-IR Mid-IR Far-IR Microwave Radio Beamsplitter storage 50,000 40,000 30,000 20,000 10,000 5,000 0 Wavenumbers (cm -1 ) Sources Deuterium 53,000 – 12,000 Xenon 40,000 – 10,000 Tungsten-Halogen 25,000 – 2,100 Ceramic 9,000 – 29 Mercury Arc 600 – 10 65,000 20,000 30,000 40,000 50,000 60,000 10,000 0 Wavenumbers (cm -1 ) Detectors R446 Photomultiplier Tube 54,000 – 11,500 Silicon 18,000 – 8,600 50,000 10,000 5,000 20,000 30,000 40,000 0 Wavenumbers (cm -1 ) Silicon 18,000 – 8,600 Lead Selenide 16,000 – 2,000 Beamsplitters 50,000 40,000 30,000 20,000 10,000 5,000 0 Wavenumbers (cm -1 ) UV-Vis Quartz 53,000 – 4,000 Near-IR Quartz 20,000 – 2,800 Calcium Fluoride 18,000 – 1,200 Potassium Bromide 11,000 – 375 Cesium Iodide 6,000 – 225 Mylar (6-125 µm) 700 – 10 10,000 2,500 1,250 5,000 7,500 0 Wavenumbers (cm -1 ) *Multiple MCT detectors (narrow, mid & wide band) are available. MCT* 12,000 – 450 DLaTGS 18,000 – 150 Far-IR DLaTGS 700 – 20 Mid-IR detectors Far-IR detectors 1969 First FTIR microscope 1982 First dynamically- aligned step-scan interferometer 1992 1971 First use of an MCT detector in an FTIR 1991 First infinity- corrected FTIR microscope First rapid- scanning FPA chemical imaging 2001 1997 First Mid-IR FPA detector system for chemical imaging Cary 600 series FTIR spectrometers and microscopes 2008 2004 2010 Varian, Inc. acquired Digilab Varian, Inc. acquired by Agilent Reference laser Interferometer Infrared source Detector Sample compartment THE COMPLETE SOLUTION Several different accessories facilitate sample preparation and spectral acquisition, including: • Attenuated Total Reflectance (ATR) • Diffuse Reflectance • Specular Reflectance • Grazing Angle Reflectance • Microscopy and Chemical Imaging • ATR Chemical Imaging (Micro and Macro) INTERPRETING SPECTRA Bond Type of Vibration Wavenumber Range (cm -1 ) C–H Alkane (stretch) 3000 – 2850 –CH 3 (bend) 1450 & 1375 –CH 2 (bend) 1465 Alkene (stretch) 3100 – 3000 (out-of-plane bend) 1000 – 650 Aromatic (stretch) 3150 – 3050 (out-of-plane bend) 900 – 600 Alkyne (stretch) ~ 3300 Aldehyde 2900 – 2700 C=C Alkene 1680 – 1600 Aromatic 1600 & 1475 C^C Alkyne 2250 – 2100 C=O Aldehyde 1740 – 1720 Ketone 1725 – 1705 Carboxylic acid 1725 – 1700 Ester 1750 – 1730 Amide 1680 – 1630 Anhydride 1810 & 1760 Bond Type of Vibration Wavenumber Range (cm -1 ) C–O Alcohols, esters, ethers, 1300 – 1000 carboxylic acid, anhydrides O–H Alcohols, phenols Free 3650 – 3600 H-Bonded 3400 – 3200 Carboxylic acids 3400 – 2400 N–H Primary & secondary amines & amides (stretch) 3500 – 3100 (bend) 1640 – 1550 C–N Amines 1350 – 1000 C=N Imines & oximes 1690 – 1640 C^N Nitriles 2260 – 2240 N=O Nitro (R–NO 2 ) 1550 & 1350 S–H Mercaptans 2550 C–X Halides Fluoride 1400 – 1000 Chloride 785 – 540 Bromide, iodide < 650 • Fiber Optic Probes • PM-IRRAS • Photoacoustic Spectroscopy • TGA-FTIR • GC-FTIR • GPC-FTIR This information is subject to change without notice. © Agilent Technologies, Inc. 2013 Published in Australia, Feburary 1, 2013 K8000-90009 *K8000-90009* *K8000-90009*
Transcript

FTIR SPECTROSCOPY REFERENCE GUIDEINTRODUCTIONInfrared spectroscopy is the study of the interactions between infrared electromagnetic energy and matter. The technique of infrared spectroscopy measures the vibrations of molecules, allowing for qualitative and quantitative measurements of samples. A Fourier transform infrared (FTIR) spectrometer is an ideal tool for the identification of unknown organic and inorganic samples whether they exist in the form of a gas, liquid or a solid.

ELECTROMAGNETIC SPECTRUMRadiation in the infrared region is commonly referred to in terms of a unit called a wavenumber (ν), rather than wavelength (λ). Wavenumbers are expressed as reciprocal centimeters (cm-1) and are the preferred unit as they are directly proportional to energy. A higher wavenumber corresponds to a higher energy.

Convert a wavenumber (ν) to a wavelength (λ) by using the following formula:

Convert a wavenumber (ν) to Energy (E) by multiplying it by the speed of light (c, in centimeters per second) and Planck's constant (h):

SPECTRAL RANGE COVERAGESpectral range coverage for common sources, beamsplitters, and detectors. The combination of these components generate a spectrometer’s working scan range.

First commercial FTIR spectrometer

A HISTORY OF COMMERCIAL 'FIRSTS'

A portion of the electromagnetic spectrum showing the relationship of the infrared region to other types of radiation (not shown to scale). The relationship between energy, wavenumber, and wavelength is also highlighted.

COLLECTING AN FTIR SPECTRUMThere are three quick, simple steps involved in obtaining a spectrum of a sample:

STEP 1 Record a spectrum with no sample present. (Known as a ‘background’.)

STEP 2 Insert the sample into the spectrometer.

STEP 3 Record a second spectrum and interpret the data.

The key collection parameters are:• The number of scans (background and sample)• The scan range• The spectral resolution.

In general increasing the number of scans that are co-added improves the signal-to-noise ratio of the spectrum and can assist in the analysis of weakly absorbing samples. The spectral resolution is user-defined and helps to distinguish closely spaced absorption peaks, and is expressed in wavenumbers.

The simplified correlation table on the right allows users to extract structural information from IR spectra. Computer-based search programs are also available for assisting in compound identification.

CARY 600 SERIESSPECTROMETER OPTICS

ν (cm-1) = 1λ (cm)

E (J) = h (J.s) x c (cm.s-1) x ν (cm-1)

High

High

Gam

ma

Energy

Wavenumber (ν)

Low

Low

Short Wavelength (λ) Long

(cm-1)

4000 400

X-Ra

ys

UV Visi

ble

Near

-IR

Mid

-IR

Far-

IR

Mic

row

ave

Radi

o

Beamsplitter storage

50,000 40,000 30,000 20,000 10,000 5,000 0Wavenumbers (cm-1)

Sources

Deuterium53,000 – 12,000

Xenon40,000 – 10,000

Tungsten-Halogen25,000 – 2,100

Ceramic9,000 – 29

Mercury Arc600 – 10

65,000 20,00030,00040,00050,00060,000 10,000 0Wavenumbers (cm-1)

Detectors

R446 Photomultiplier Tube 54,000 – 11,500

Silicon18,000 – 8,600

50,000 10,000 5,00020,00030,00040,000 0Wavenumbers (cm-1)

Silicon18,000 – 8,600

Lead Selenide16,000 – 2,000

Beamsplitters50,000 40,000 30,000 20,000 10,000 5,000 0

Wavenumbers (cm-1)

UV-Vis Quartz53,000 – 4,000

Near-IR Quartz20,000 – 2,800

Calcium Fluoride18,000 – 1,200

Potassium Bromide11,000 – 375

Cesium Iodide6,000 – 225

Mylar (6-125 µm)700 – 10

10,000 2,500 1,2505,0007,500 0Wavenumbers (cm-1)

*Multiple MCT detectors (narrow, mid & wide band) are available.

MCT*12,000 – 450

DLaTGS18,000 – 150

Far-IR DLaTGS 700 – 20

Mid-IR

detectorsFar-IR

detectors

1969

First FTIR microscope

1982

First dynamically-aligned step-scan interferometer

1992

1971First use of an MCT detector in an FTIR

1991First infinity-corrected FTIR microscope

First rapid-scanning FPA chemical imaging

2001

1997First Mid-IR FPA detector system for chemical imaging

Cary 600 series FTIR spectrometersand microscopes

2008

2004 2010Varian, Inc. acquired Digilab

Varian, Inc. acquired by Agilent

Reference laser

Interferometer

Infrared source

Detector

Sample compartmentTHE COMPLETE SOLUTION

Several different accessories facilitate sample preparation and spectral acquisition, including: • Attenuated Total Reflectance (ATR)• Diffuse Reflectance• Specular Reflectance• Grazing Angle Reflectance• Microscopy and Chemical Imaging• ATR Chemical Imaging (Micro and Macro)

INTERPRETING SPECTRABond Type of Vibration Wavenumber Range (cm-1)C–H Alkane (stretch) 3000 – 2850 –CH3 (bend) 1450 & 1375

–CH2– (bend) 1465 Alkene (stretch) 3100 – 3000 (out-of-plane bend) 1000 – 650 Aromatic (stretch) 3150 – 3050 (out-of-plane bend) 900 – 600 Alkyne (stretch) ~ 3300 Aldehyde 2900 – 2700C=C Alkene 1680 – 1600 Aromatic 1600 & 1475C^C Alkyne 2250 – 2100C=O Aldehyde 1740 – 1720 Ketone 1725 – 1705 Carboxylic acid 1725 – 1700 Ester 1750 – 1730 Amide 1680 – 1630 Anhydride 1810 & 1760

Bond Type of Vibration Wavenumber Range (cm-1)C–O Alcohols, esters, ethers, 1300 – 1000 carboxylic acid, anhydridesO–H Alcohols, phenols Free 3650 – 3600 H-Bonded 3400 – 3200 Carboxylic acids 3400 – 2400N–H Primary & secondary amines & amides (stretch) 3500 – 3100 (bend) 1640 – 1550C–N Amines 1350 – 1000C=N Imines & oximes 1690 – 1640C^N Nitriles 2260 – 2240N=O Nitro (R–NO2) 1550 & 1350S–H Mercaptans 2550C–X Halides Fluoride 1400 – 1000 Chloride 785 – 540 Bromide, iodide < 650

• Fiber Optic Probes• PM-IRRAS• Photoacoustic Spectroscopy• TGA-FTIR• GC-FTIR• GPC-FTIR

This information is subject to change without notice.© Agilent Technologies, Inc. 2013

Published in Australia, Feburary 1, 2013K8000-90009

*K8000-90009**K8000-90009*

Recommended