FUNDAMENTALS OF TWO-DIMENSIONAL X-RAY DIFFRACTION (XRD 2 ) Baoping Bob He, Uwe Preckwinkel and Kingsley L. Smith Bruker Analytical X-ray Systems Madison, Wisconsin, USA ABSTRACT Two-dimensional x-ray diffraction (XRD 2 ) refers to x-ray diffraction applications with two- dimensional (2D) detector and corresponding data reduction and analysis. The two-dimensional diffraction pattern contains far more information than a one-dimensional profile collected with the conventional diffractometer. In order to take the advantages of the 2D detector, new approaches are necessary to configure the two-dimensional x-ray diffraction system and to analyze the 2D diffraction data. The present paper discusses some fundamentals about two-dimensional x-ray diffraction, such as geometry convention, diffraction data interpretation, point beam optics, and advantages of XRD 2 in various applications. 1. INTRODUCTION In the field of x-ray powder diffraction, data collection and analysis have been based mainly on one-dimensional diffraction (1D) profiles measured with scanning point detectors or linear position-sensitive detectors (PSD). Therefore, almost all the x-ray powder diffraction applications, such as phase identification, texture (orientation), residual stress, crystallite size, percent crystallinity, lattice dimensions and structure refinement (Rietveld), are developed in accord with the 1D profile collected by conventional diffractometers. [1] In recent years, usage of two-dimensional (2D) detectors has dramatically increased due to the advances in detector technology, point beam x-ray optics, and computing power [2,3]. Although a 2D image contains far more information than a 1D profile, the advantages of a 2D detector can not be fully taken if the data interpretation and analysis methods are simply inherited from the conventional diffraction theory. XRD 2 is a new technique in the field of x-ray diffraction (XRD), which is not simply a diffractometer with a two-dimensional (2D) detector. In addition to the 2D detector technology, it involves 2D image processing and 2D diffraction pattern manipulation and interpretation. Because of the unique nature of the data collected with a 2D detector, a completely new concept and new approach are necessary to configure XRD 2 system and to understand and analyze the 2D diffraction data. In addition, the new theory should also be consistent with the conventional theory so that the 2D data can also be used for conventional applications. An XRD 2 system is a diffraction system with the capability of acquiring diffraction pattern in 2D space simultaneously and analyzing 2D diffraction data accordingly. A two-dimensional x-ray diffraction system consists of at least one two-dimensional detector, x-ray source, x-ray optics, sample positioning stage, sample alignment and monitoring device as well as corresponding computer control and data reduction and analysis software. Figure 1 is an example of an XRD 2 system with five major components. Copyright(c)JCPDS-International Centre for Diffraction Data 2000,Advances in X-ray Analysis,Vol.43 273 ISSN 1097-0002
Transcript
Page 1: FUNDAMENTALS OF TWO-DIMENSIONAL X-RAY …icdd.com/resources/axa/vol43/v43_037.pdf · advances in detector technology, point beam x-ray optics, and computing power [2,3]. Although

FUNDAMENTALS OF TWO-DIMENSIONAL X-RAY DIFFRACTION(XRD2)

Baoping Bob He, Uwe Preckwinkel and Kingsley L. SmithBruker Analytical X-ray Systems

Madison, Wisconsin, USA

ABSTRACT

Two-dimensional x-ray diffraction (XRD2) refers to x-ray diffraction applications with two-dimensional (2D) detector and corresponding data reduction and analysis. The two-dimensionaldiffraction pattern contains far more information than a one-dimensional profile collected with theconventional diffractometer. In order to take the advantages of the 2D detector, new approachesare necessary to configure the two-dimensional x-ray diffraction system and to analyze the 2Ddiffraction data. The present paper discusses some fundamentals about two-dimensional x-raydiffraction, such as geometry convention, diffraction data interpretation, point beam optics, andadvantages of XRD2 in various applications.

1. INTRODUCTION

In the field of x-ray powder diffraction, data collection and analysis have been based mainly onone-dimensional diffraction (1D) profiles measured with scanning point detectors or linearposition-sensitive detectors (PSD). Therefore, almost all the x-ray powder diffraction applications,such as phase identification, texture (orientation), residual stress, crystallite size, percentcrystallinity, lattice dimensions and structure refinement (Rietveld), are developed in accord withthe 1D profile collected by conventional diffractometers. [1]

In recent years, usage of two-dimensional (2D) detectors has dramatically increased due to theadvances in detector technology, point beam x-ray optics, and computing power [2,3]. Although a2D image contains far more information than a 1D profile, the advantages of a 2D detector can notbe fully taken if the data interpretation and analysis methods are simply inherited from theconventional diffraction theory. XRD2 is a new technique in the field of x-ray diffraction (XRD),which is not simply a diffractometer with a two-dimensional (2D) detector. In addition to the 2Ddetector technology, it involves 2D image processing and 2D diffraction pattern manipulation andinterpretation. Because of the unique nature of the data collected with a 2D detector, a completelynew concept and new approach are necessary to configure XRD2 system and to understand andanalyze the 2D diffraction data. In addition, the new theory should also be consistent with theconventional theory so that the 2D data can also be used for conventional applications.

An XRD2 system is a diffraction system with the capability of acquiring diffraction pattern in 2Dspace simultaneously and analyzing 2D diffraction data accordingly. A two-dimensional x-raydiffraction system consists of at least one two-dimensional detector, x-ray source, x-ray optics,sample positioning stage, sample alignment and monitoring device as well as correspondingcomputer control and data reduction and analysis software. Figure 1 is an example of an XRD2

system with five major components.

Copyright(c)JCPDS-International Centre for Diffraction Data 2000,Advances in X-ray Analysis,Vol.43 273ISSN 1097-0002

Page 2: FUNDAMENTALS OF TWO-DIMENSIONAL X-RAY …icdd.com/resources/axa/vol43/v43_037.pdf · advances in detector technology, point beam x-ray optics, and computing power [2,3]. Although

This document was presented at the Denver X-ray Conference (DXC) on Applications of X-ray Analysis. Sponsored by the International Centre for Diffraction Data (ICDD). This document is provided by ICDD in cooperation with the authors and presenters of the DXC for the express purpose of educating the scientific community. All copyrights for the document are retained by ICDD. Usage is restricted for the purposes of education and scientific research. DXC Website – www.dxcicdd.com

ICDD Website - www.icdd.com

ISSN 1097-0002

Page 3: FUNDAMENTALS OF TWO-DIMENSIONAL X-RAY …icdd.com/resources/axa/vol43/v43_037.pdf · advances in detector technology, point beam x-ray optics, and computing power [2,3]. Although

Figure 1. Five major components in an XRD2 system, an area detector, an x-ray generator,x-ray optics (monochromator and collimator), goniometer and sample stage, andsample alignment and monitoring (laser/video) system.

Figure 2 shows the pattern of diffracted x-rays from a single crystal and from a polycrystallinesample. The diffracted rays from a single crystal point to discrete directions each corresponding toa family of diffraction planes (Figure 2a). The diffraction pattern from a polycrystalline (powder)sample forms a series diffraction cones if large number of crystals oriented randomly in the spaceare covered by the incident x-ray beam (Figure 2b). Each diffraction cone corresponds to thediffraction from the same family of crystalline planes in all the participating grains. The diffractionpatterns from polycrystalline materials will be considered thereafter in the further discussion of thetheory and configuration of XRD2 systems. Polycrystalline materials here refer to single-phase,multi-phase, bulk and thin film materials.

Figure 2. The patterns of diffracted x-rays: (a) from a single crystal and (b) from apolycrystalline sample.

First, we compare the conventional x-ray diffraction (XRD) and two-dimensional x-ray diffraction(XRD2). Figure 3 is a schematic of x-ray diffraction from a powder (polycrystalline) sample. Forsimplicity, it shows only two diffraction cones, one represents forward diffraction (2θ 90°) and one

Copyright(c)JCPDS-International Centre for Diffraction Data 2000,Advances in X-ray Analysis,Vol.43 274ISSN 1097-0002

Page 4: FUNDAMENTALS OF TWO-DIMENSIONAL X-RAY …icdd.com/resources/axa/vol43/v43_037.pdf · advances in detector technology, point beam x-ray optics, and computing power [2,3]. Although

for backward diffraction (2θ>90°). The diffraction measurement in the conventional diffractometeris confined within a plane, here referred to as diffractometer plane. A point detector makes 2θ scanalong a detection circle. If a one-dimensional position sensitive detector (PSD) is used in thediffractometer, it will be mounted on the detection circle. Since the variation of diffraction patternin the direction (Z) perpendicular to the diffractometer plane are not considered in theconventional diffractometer, the x-ray beam is normally extended in Z direction (line focus). Theactual diffraction pattern measured by a conventional diffractometer is an average over a rangedefined by beam size in Z-direction. Since the diffraction data out of the diffractometer plane is notdetected, the corresponding structure information will be either ignored, or measured by variousadditional sample rotations.

Figure 3. Diffraction patterns in 3D space from a powder sample and thediffractometer plane in the conventional diffractometer.

Figure 4. Comparison of diffraction pattern coverage between point (0D), linear PSD (1D),and area (2D) detectors.

With a 2D detector, the measurable diffraction is no longer limited in the diffractometer plane.Instead, the whole or a large portion of the diffraction rings can be measured simultaneously,depending on the detector’s size and position. Figure 4 shows the diffraction pattern on 2D

Copyright(c)JCPDS-International Centre for Diffraction Data 2000,Advances in X-ray Analysis,Vol.43 275ISSN 1097-0002

Page 5: FUNDAMENTALS OF TWO-DIMENSIONAL X-RAY …icdd.com/resources/axa/vol43/v43_037.pdf · advances in detector technology, point beam x-ray optics, and computing power [2,3]. Although

detector compared with the diffraction measurement range of scintillation detector and PSD. Sincethe diffraction rings are measured, the variations of diffraction intensity in all directions are equallyimportant, the ideal shape of the x-ray beam cross-section for XRD2 is a point (point focus). Inpractice, the beam cross-section can be either circular or square in limited size.

2. GEOMETRY CONVENTIONS IN XRD2 SYSTEM

2.1. Diffraction Cones in Laboratory AxesFigure 5 describes the geometric definition of diffraction cones in the laboratory coordinatessystem, XLYLZL. Analogous to the conventional 3-circle and 4-circle goniometer, the direct x-raybeam propagates along the XL axis, ZL is up, and YL makes up a right handed rectangularcoordinate system. The axis XL is also the rotation axis of the cones. The apex angles of the conesare determined by the 2θ values given by the Bragg equation. The apex angels are twice the 2θvalues for forward reflection (2θ≤90°) and twice the values of 180°-2θ for backward reflection(2θ>90°). The γ angle is the azimuthal angle from the origin at the 6 o’clock direction (-ZL

direction) with right handed rotation axis along the opposite direction of incident beam (-XL

direction). In the previous publications, χ has been used to denote this angle by the presentauthor(s). Since χ has also been used to denote one of the goniometer angles in 4-circleconvention, γ will be used thereafter to represent this angle. The γ angle here is used to define thedirection of the diffracted beam on the cone. The γ angle actually defines a half plane with the XL

axis as the edge, it will be referred to as γ-plane thereafter. Intersections of any diffraction coneswith a γ-plane have the same γ value. The conventional diffractometer plane consists of two γ-planes with one γ=90° plane in the negative YL side and γ=270° plane in the positive YL side. γ and2θ angles forms a kind of spherical coordinate system which covers all the directions from theorigin of sample (goniometer center). The γ-2θ system is fixed in the laboratory systems XLYLZL,which is independent of the sample orientation in the goniometer. This is a very important conceptwhen we deal with the 2D diffraction data.

Figure 5. The geometric definition of diffraction rings in laboratory axes.

2.2. Ideal Detector for Diffraction Pattern in 3D SpaceAn ideal detector to measure the diffraction pattern in 3D space is a detector with sphericaldetecting surface covering all the diffraction directions in 3D space as is shown in Figure 6. The

Copyright(c)JCPDS-International Centre for Diffraction Data 2000,Advances in X-ray Analysis,Vol.43 276ISSN 1097-0002

Page 6: FUNDAMENTALS OF TWO-DIMENSIONAL X-RAY …icdd.com/resources/axa/vol43/v43_037.pdf · advances in detector technology, point beam x-ray optics, and computing power [2,3]. Although

sample is in the center of the sphere. The direction of a diffracted beam is defined by γ (longitude)and 2θ (latitude). The incident x-ray beam points to the center of the sphere through the detectorat 2θ = π. The detector surface covers the whole spherical surface, i.e. 4π in solid angle. The idealdetector should have large dynamic range, small pixel size, and narrow point spread function, aswell as many properties for an “ideal detector”. In practice, such an ideal detector does not exist.However, there are many 2D detector technologies available, including photographic film, CCD,image plate (IP) and multi-wire proportional counter (MWPC). Each technique has its advantagesover the others [2]. The detection surface can be a spherical, cylindrical or flat. The spherical orcylindrical detectors are normally designed for a fixed sample to detector distance, while flatdetector has the flexibility to be used at different sample-to-detector distance so as to choosebetween higher resolution at large distance or higher angular coverage at short distance. Thefollowing discussion on XRD2 geometry will focus on flat 2D detectors.

Figure 6. Schematics of an ideal detector covering 4π solid angle.

2.1. Diffraction Cones and Conic Sections on 2D DetectorsFigure 7 shows the geometry of a diffraction cone. The incident x-ray beam always lies along therotation axis of the diffraction cone. The whole apex angle of the cone is twice the 2θ value givenby Bragg Law. For a flat 2D detector, the detection surface can be considered as a plane, whichintersects the diffraction cone to form a conic section. D is the distance between the sample andthe detector, and α is the detector swing angle. The conic section takes the different shapes fordifferent α angles. When imaged on-axis (α = 0°) the conic sections appear as circles. When thedetector is at off-axis position (α ≠ 0°), the conic section may be an ellipse, parabola, or hyperbola.For convenience, all kinds of conic sections will be referred to as diffraction rings or Debye ringsalternatively hereafter in this paper. The 2D diffraction image collected in a single exposure will bereferred to as a frame. The frame is normally stored as intensity values on 2D pixels. Thedetermination of the diffracted beam direction involves the conversion of pixel information into theγ-2θ coordinates. In an XRD2 system, γ and 2θ values at each pixel position are given according tothe detector position. The diffraction rings can be displayed in terms of γ and 2θ coordinates,disregarding the actual shape of each diffraction ring.

Copyright(c)JCPDS-International Centre for Diffraction Data 2000,Advances in X-ray Analysis,Vol.43 277ISSN 1097-0002

Page 7: FUNDAMENTALS OF TWO-DIMENSIONAL X-RAY …icdd.com/resources/axa/vol43/v43_037.pdf · advances in detector technology, point beam x-ray optics, and computing power [2,3]. Although

Figure 7. A diffraction cone and the conic section with a 2D detector plane.

2.4. Detector Position in the Laboratory SystemThe detector position is defined by the sample-to-detector distance D and the detector swing angleα. D is the perpendicular distance from the goniometer center to the detection plane and α is aright-handed rotation angle above ZL axis. In the laboratory coordinates XLYLZL, detectors atdifferent positions are shown in Figure 8. The center of the detector 1 is right on the positive sideof XL axis (on-axis), α=0. Both the detector 2 and 3 are rotated away from XL axis with negativeswing angles (α2<0 and α3<0). The swing angle is also called as detector two-theta and denoted by2θD in the previous publications. It is very important to distinguish between the Bragg angle 2θand detector angle α. At a given detector angle α, a range of 2θ values can be measured.

Figure 8. Detector position in the laboratory system XLYLZL: D is the sample-to-detector distance; α is the swing angle of the detector.

Copyright(c)JCPDS-International Centre for Diffraction Data 2000,Advances in X-ray Analysis,Vol.43 278ISSN 1097-0002

Page 8: FUNDAMENTALS OF TWO-DIMENSIONAL X-RAY …icdd.com/resources/axa/vol43/v43_037.pdf · advances in detector technology, point beam x-ray optics, and computing power [2,3]. Although

2.3. Sample Orientation and Position in the Laboratory System

In an XRD2 system, three rotation angles are necessary to define the orientation of a sample in thediffractometer. These three rotation angles can be achieved either by an Eulerian cradle (4-circle)type geometry, a kappa (κ) geometry or other kind of geometry. The 4-circle geometry will bediscussed in this paper. The three angles in 4-circle geometry are ω (omega), χg (goniometer chi)and φ (phi). Since the χ symbol has also been used for the azimuthal angle on the diffraction conesin the previous publications and some software, a subscript g indicates the angle is a goniometerangle. Figure 9(a) shows the relationship between rotation axes (ω, χg, φ) and the laboratorysystem XLYLZL. ω is defined as a right-handed rotation about ZL axis. The ω axis is fixed on thelaboratory coordinates. χg is a left-handed rotation about a horizontal axis. The χg axis makes anangle of ω with XL axis in the XL-YL plane. The χg axis lies on XL when ω is set at zero. φ is a left-handed rotation. The χg angle is also the angle between φ axis and the ZL.

Figure 9(b) shows the relationship among all rotation axes (ω, χg, ψ, φ) and translation axes XYZ.ω is the base rotation, all other rotations and translations are on top of this rotation. The nextrotation above ω is the χg rotation. ψ is also a rotation above a horizontal axis. ψ and χg have thesame axis but different starting positions and rotation directions, and χg = 90°-ψ. The next rotationabove ω and χg(ψ) is φ rotation. The sample translation coordinates XYZ are so defined that,when ω = ψ = φ =0, the relationship to the laboratory axes becomes X= -XL, Y= ZL, and Z= YL.The φ rotation axis is always the same as the Z-axis at any sample orientation.

Figure 9. Sample rotation and translation. (a) Three rotation axes in XLYLZL

coordinates; (b) Rotation axes (ω, χg, ψ, φ) and translation axes XYZ.

In an aligned diffraction system, all three rotation axes and the primary x-ray beam cross at theorigin of XLYLZL coordinates. This cross point is also known as goniometer center or instrumentcenter. X-Y plane is usually the sample surface and Z is the sample surface normal. In a preferredembodiment, XYZ translations are above all the rotations so that the translations will not moveany rotation axis away from the goniometer center. Instead, the XYZ translations bring differentparts of the sample into the goniometer center.

Copyright(c)JCPDS-International Centre for Diffraction Data 2000,Advances in X-ray Analysis,Vol.43 279ISSN 1097-0002

Page 9: FUNDAMENTALS OF TWO-DIMENSIONAL X-RAY …icdd.com/resources/axa/vol43/v43_037.pdf · advances in detector technology, point beam x-ray optics, and computing power [2,3]. Although

3. SUMMARY OF XRD2 APPLICATIONS

Compared to a conventional one-dimensional diffraction system, an XRD2 system has manyadvantages in various applications:• Phase identification (Phase ID) can be done by integration over a selected range of 2θ and γ.

The integrated data gives better intensity and statistics for phase ID and quantitative analysis,especially for those samples with texture, large grain size, or small quantity [3].

• Texture measurement is extremely fast. An XRD2 system collects texture data and backgroundvalues simultaneously for multiple poles and multiple directions. Due to the high measurementspeed, Pole figure can be measured at very fine steps for sharp textures [4,5].

• Stress can be measured using the 2D fundamental equation, which gives the direct relationshipbetween the stress tensor and the diffraction cone distortion. Since the whole or a part of theDebye ring is used for stress calculation, it can measure stress with high sensitivity, high speedand high accuracy. It is very suitable for large grain and textured samples [6,7].

• Percent crystallinity can be measured faster and more accurately with the data analysis overthe 2D pattern, especially for samples with anisotropic distribution of crystalline orientation.

• Small angle x-ray scattering (SAXS) data can be collected at high speed. Anisotropicfeatures from specimens, such as polymers, fibrous materials, single crystals and bio-materials,can be analyzed and displayed in two-dimension. De-smearing correction is not necessary dueto the collimated point x-ray beam. Since one exposure takes all the SAXS information, it iseasy to scan over the sample to map the structure information [1,8].

• Microdiffraction data is collected with speed and accuracy. The 2D detector captures wholeor a large portion of the diffraction rings, so spotty, textured, or weak diffraction data can beintegrated over the selected diffraction rings.

• Thin film samples with a mixture of single crystal, random polycrystalline layers and highlytextured layers can be measured with all the features appearing simultaneously in 2D frames.

REFERENCES[1] Ron Jenkins and Robert L. Snyder, Introduction to X-ray Powder Diffractometry, John Wiley

& Sons, New York, 1996.[2] Philip R. Rudolf and Brian G. Landes, Two-dimensional X-ray Diffraction and Scattering of

Microcrystalline and Polymeric Materials, Spectroscopy, 9(6), pp 22-33, July/August 1994.[3] S. N. Sulyanov, A. N. Popov and D. M. Kheiker, Using a Two-dimensional Detector for X-ray

Powder Diffractometry, J. Appl. Cryst. 27, pp 934-942, 1994.[4] Hans J. Bunge and Helmut Klein, Determination of Quantitative, High-resolution Pole Figures

with the Aea Detector, Z. Metallkd. 87(6), pp 465-475, 1996.[5] K. L. Smith and R. B. Ortega, Use of a Two-dimensional, Position Sensitive Detector for

Collecting Pole Figures, Advances in X-ray Analysis, Vol. 36, pp 641-647, Plenum, 1993.[6] Baoping B. He and Kingsley L. Smith, Fundamental Equation of Strain and Stress

Measurement Using 2D Detectors, Proceedings of 1998 SEM Spring Conference onExperimental and Applied Mechanics, Houston, Texas, USA, 1998.

[7] Baoping B. He, Uwe Preckwinkel and Kingsley L. Smith, Advantages of Using 2D Detectorsfor Residual Stress Measurements, Advances in X-ray Analysis, Vol. 42, Proceedings of the47th Annual Denver X-ray Conference, Colorado Springs, Colorado, USA, 1998.

[8] H. F. Jakob, S. E. Tschegg, P. Fratzl, J. Struct. Biology, Vol. 113. Pp. 13-22, 1994.

Copyright(c)JCPDS-International Centre for Diffraction Data 2000,Advances in X-ray Analysis,Vol.43 280ISSN 1097-0002