FUSION 4000 & 6000 USER MANUAL
Transcript

FUSION 4000 & 6000USER MANUAL

2

EU Declaration of Conformity (DoC)....................................................................................................................6

General Description...............................................................................................................................................7

Warranty and Repair Information........................................................................................................................9

Warranty Information................................................................................................................................9

Repair Facilities and Parts........................................................................................................................10

Dead Pixel Policy......................................................................................................................................10

Serial Numbers.........................................................................................................................................10

Calibrations.........................................................................................................................................10

Product Overview..................................................................................................................................................11

Principle of Operation..............................................................................................................................11

Controlled Elements.................................................................................................................................12

Specifications...............................................................................................................................................15

Pump Setup...........................................................................................................................................................17

PoweringOn/OffthePump....................................................................................................................17

Loading a Syringe.....................................................................................................................................18

AdjustingtheSafetyNut/StallRod........................................................................................................21

Pump Operation......................................................................................................................................22

Mode Selection Screen............................................................................................................................24

Pressure&TemperatureSettings(Fusion6000only).......................................................................25

Basic Mode............................................................................................................................................................26

Syringe Selection......................................................................................................................................26

Volume......................................................................................................................................27

Infusion/Withdrawl.................................................................................................................................28

Flow Rate..................................................................................................................................................28

Start-Time Delay......................................................................................................................................29

Step Loop..................................................................................................................................................29

3

Options...........................................................................................................................................................30

Flow-Rate Units........................................................................................................................................30

Options....................................................................................................................................31

Enable Volume Lock.................................................................................................................................31

Multi-Step Mode...................................................................................................................................................32

Setup.......................................................................................................................................32

Syringe Selection.....................................................................................................................................32

Total Steps................................................................................................................................................33

Loop All.....................................................................................................................................................33

Sub-Loops...........................................................................................................................................33

Flow-Rate Units........................................................................................................................................34

Step Parameters...................................................................................................................................................34

Volume.......................................................................................................................................35

Pump Control.......................................................................................................................................................36

System Settings....................................................................................................................................................39

Pump Control by Computer...............................................................................................................................41

Cable Requirements............................................................................................................................................41

Connecting ............................................................................................................................................................41

COM Port Settings....................................................................................................................................42

Pump Commands................................................................................................................................................43

ChemyxPumpControllerPrograms.................................................................................................................48

Pump Maintenance.............................................................................................................................................49

Lubricating thePump.............................................................................................................................49

Lubrication Instructions.........................................................................................................................49

Approved Lubricants...............................................................................................................................49

PumpAccessories................................................................................................................................................50

4

Appendices.................................................................................................................................................52

AppendixA:SyringeVolume/DiameterRefference............................................................................52

AppendixB:AvailableParts....................................................................................................................56

AppendixC:ContactInformation...........................................................................................................57

5

INTELLECTUAL PROPERTYAllIntellectualProperties,asdefinedbelow,ownedbyorwhichisotherwisethepropertyofChemyxInc.oritssuppliersrelatingtotheChemyxsyringepumps,includingbutnotlimitedto,accessories,

partsorsoftwarerelatingthereto(ChemyxSyringePumps),areproprietarytofederalandstatelaws,andinternationaltreatyprovisions.IntellectualPropertyincludesbutisnotlimitedto,inventions

(patentableorunpatentable),patents,tradesecrets,copyrights,software,firmware,computerprograms,andrelateddocumentationandotherworksofauthorship.Moreover,youagreethatyouwill

not,andwillnotattemptto,modify,preparederivativeworksof,reverseengineer,disassembletheChemyxsyringepumps,decompileorotherwiseattempttocreatesourcecodefromtherelated

software/firmware.NotitletoorownershipinIntellectualPropertyistransferredtoyou.AllapplicablerightsoftheIntellectualPropertyshallremainwithChemyxanditssuppliers.

6

Declaration of Conformity

ThePDFversionofthelatestDoCcanbedownloadedatwww.chemyx.com/support

7

General Description

PleasereadthefollowingsafetyprecautionstoensurepersonalsafetyandoperationallongevityoftheChemyxsyringepump.Chemyxisnotresponsiblefortheequipmentifusedinamannernotspecifiedbythemanufacturer;warrantycoverageprovidedbytheequipmentmaybedroppedasaresult.

CHEMYX PRODUCTS ARE NOT APPROVED FOR CLINICAL USE ON HUMANS.

Chemyxisnotresponsiblefortheuseofpowersuppliesoutsidethestatedelectricalspecificationsorfailuretoswitchthepowerconverterfrom220Vto110Vwhileinthe220Venvironmentorvice versa.

Theproductshouldbeproperlygrounded.

Use Proper Power Supply

Ground Product

Safety Information

WarrantycoveragewillbelostifthepumpisopenedwithoutauthorizationfromChemyx.Donottouchanyelectricalconnectorsonorintheproduct.

Eventhoughthepumpcanoperateatextremelyfastspeeds,theusermustdeterminetheproperflowrateforanygivenapplication.Forinstances,pumpingat90mL/minusinga20-gaugeneedlewillcausestallsand/orpotentialburstingofthesyringe.Chemyxisnotresponsibleforanydamagethatmightresultfromsituationssimilartotheexampleabove.

Do Not Open The Pump

Do Not Operate With Suspected Failures

8

Donotplacefingersbetweenthepusherblockandtheendblockwhilethepumpisrunning.

Readalllabelsontheproducttoensureproperusage.

Theuserisresponsibleforwettinggroundglasssyringesandsettingandtighteningthesafetycollar/bar appropriately.

Pinch Hazard

Observe all Warning labels on Product

Chemyx is not Responsible for Syringe Damage

9

Warranty and Repair InformationChemyxprovidesatwo-yearwarrantyfromtheshipmentdateforitspumpsagainstdefectsinmaterialsandworkmanship.Chemyxwillrepairanyproductthatprovesdefectiveduringitsstatedwarrantyperiod.

Theforegoingwarrantywillnotapplytodamageresultingfrom:

YoumustcontactChemyx(call+1281-277-5499orvisitwww.chemyx.com/contact-us)beforereturningaproduct.ChemyxwillissueaReturnAuthorization(RA)numbertoyou.

Return products to:

ChemyxInc.

10905CashRoad|Stafford,TX77477USA.

Warranty Information

Improper or inadequate maintenance or operation.

Unauthorizedmodificationormisuseoftheproduct.

Operationoutsidetheelectricalspecificationsfortheproduct.

Operationoutsidethetemperaturespecificationsfortheproduct.

User-inducedinternalandexternalcontaminationsoftheinstrument.

Failure to use proper surge protection.

Improperproductreturn,packaging,andshipping

Removing serial number from syringe pump

10

Chemyxcanrepairanysyringepumpwithoutmajordamage.YoumustcontactChemyx(call+1281-277-5499orvisitusatwww.chemyx.com/contact-us)beforeshippingaproductforrepair.ChemyxwillissueaReturnAuthorization(RA)numbertoyou.

Return products to:

Chemyx Returns

10905CashRoad|Stafford,TX77477USA

DuringtheLCDMonitormanufacturingprocess,itisnotuncommonforoneormorepixelstobecomefixedinanunchangingstate.Thevisibleresultisafixedpixelthatappearsasanextremelytiny,darkorbrightdot.Inalmosteverycase,thesefixedpixelsarehardtoseeanddonotdetractfromdisplayqualityorusability.Adisplaywiththreetosevenbrightordarkdotsisconsiderednormalandwithinindustrystandards.Ifyourscreendisplaysmorethansevendeadpixelsduringthewarrantyperiod,yoursystemwillqualifyforwarrantyreplacement.PleaseseetheLimitedWarrantysectionfordetailsonhowtoactivatea warranty claim.

Theserialnumberislocatedontheback,topleftcornerorcenterofthepumpunderasmallbarcode.Removaloftheserialnumberlabelwillvoidthewarranty.

Chemyxpumpsarepre-calibrateduponarrivaltoyoursite.Allcalibratedparametersarewithinstatedaccuracyandprecisionspecifications of the pump. Although the pump is highly accurate, syringes of differentmaterials can attribute significantvolumeerrors.Typically,theseerrorscanrangefrom1%forglasssyringesto5%forplasticsyringes.Chemyxisnotresponsiblefor errors generated from syringes.

Repair Facilities and Parts

Dead Pixel Policy

Serial Numbers

Calibrations

11

Product Overview

The Chemyx Fusion product line serve a wide variety of applications. The Chemyx Fusion 4000 is configured with twoindependentlycontrolledsyringepumpchannelswhiletheChemyxFusion6000isahigh-pressuresyringepumpdesignedfordosingviscoussolutions,andsimi-solidmaterialinjectionsagainstpressure.Theunits’flexibilitywithsyringetypesandsizesandanadvancedsoftwareinterfacemakethishigh-pressuresyringepumpsintegrableintoanyoperationorlaboratoryworkflow.Thoroughlydurablebothinsideandout,theChemyxFusionsyringepumpsaredesignedtoprovideyearsofconstantandreliable service.

Asyringepumpisasmall,positive-displacementpumpusedtograduallytransferprecisevolumesoffluid.AllChemyxFusion-seriessyringepumpsaredrivenviaasteppermotor.Thissteppermotorpreciselyturnsaleadscrewthatisthreadedthroughapusherblock,whichcausesthepusherblocktomove.Whenthepumpisininfusionmode,thepusherblockpushesagainsttheplungerofasecuredsyringe,causingthefluidtobeejectedatanaccurateandpreciserate.

Ifthepumpiscapableofwithdrawal,theplungerofthesyringeisheldbybracketsonthepusherblock.Whenthesteppermotorturnsintheoppositedirection,thepusherblockmovessuchthatthesyringeplungerispulled,thusdrawingfluidintothesyringe.

FortheFusionSeriessyringepumps,thepusherblockmovestotherightforinfusionandtotheleftforwithdrawal.

Principle of Operation

12+1281-277-5499|www.chemyx.com

AllFusionSeriespumpshaveasingleleadscrewthroughwhichthepusherblockisthreaded.Thepusherblockmoveswhentheleadscrewturns.Thetwoguiderodskeepthepusherblockhorizontalandperfectlyperpendiculartotheleadscrew.FortheFusion4000thescrewknobonthepusherblockdisengagesthepusherblockfromtheleadscrew,whichallowsthepusherblocktobeeasilymovedtoanewpositionaslongasthebuttonisheld.Tightingtheknobwilllockthepusherblockbackinplace.AsfortheFusion6000thepusherblockcanonlybemovedbackandforthfromthepumpscreeninthemenusection.

Syringesareplacedonthesyringe-holderblockinthev-shapedgrooves,orchannels,andheldsecurelyusingsyringeclamps.TheFusion4000hastwoindependentsyringechannels,whiletheFusion6000hasonelargesyringechannels.

TheFusion4000hasasafetycollarwhiletheFusion6000hasastallrodtohelpprotecttheintegrityofsyringesbykeepingthepusherblockfrompushingonacompletelydepressedsyringe.

Controlled Elements

Syringe Holder Block

Safety Rod Lead Screw

PusherBlock

Syringe Clamp

SyringeChannelsFusion 6000 pump components

13+1281-277-5499|www.chemyx.com

TheFusionSeriesusesabright,4.25”,resistive,LCDtouchscreenwithanintuitivegraphicaluserinterface(GUI)forsettingupdetailedpumprunmethods.ButtonsandentryfieldswithintheGUIareallselectableusingthetouchscreen.Asaresistive,LCDtouchscreen,theinterfacecanbeoperatedwithanyobject,notjustthetouchofafingertip.Selectingelementsbytouchonlyrequiresasmallamountofpressure;however,pleasedonotusesharpobjectsastheymaydamagethescreen.

TheFusionSeriesalso includesphysicalnavigationcontrols,allowingusers tomanuallymaneuverbetweenselectableGUIelementswithouttouchingthescreen.InputtingnumbersandletterswithintheGUIisdoneusingthephysicalnumerickeypad.Thephysicalpumpcontrolkeysareusedtostart,pause,andstoppumpactivity.

ThepowerLEDandpump-controlLEDsallowuserstoeasilydiscernwhetherthepumpisonandwhetheritiscurrentlyrunning,paused, or stopped.

Navigation Keypad

Front Panel of Pump

Controls

Power LED Pump Controls LCDTouchscreen Numeric Keypad

14+1281-277-5499|www.chemyx.com

TheFusion4000 iscapableofboth infusionandwithdrawlwithtwoindependentlycontrolledpumps.TheFusion4000uses adjustable retainingbraketson thepusherblock andthe syringeholderblock, that keepeach individual syringechannelinplaceduringinfusionandwithdrawl.

InadditiontohavingallofthesamefeaturesandcomponentsoftheFusion4000,theFusion6000iscapableofbothinfusionandwithdrawalonasinglepumpchannel.TheFusion6000also uses an adjustable retaining brackets on the pusherblockand thesyringeholderblock thatkeep thesyringe inplaceduringbothinfusionandwithdrawal.

Bracket Release Knob

Bracket Release Knob

Flange Retaining Bracket

Flange Retaining Bracket

Plunger Retaining Bracket

Plunger Retaining BracketBracket Release Knob

Fusion 4000 Specific Components

Fusion 6000 Specific Components

15

Specifications

Specification Fusion 4000 Fusion 6000

Movement Modes

Infusion/Withdrawal

Infusion/Withdrawal

SyringeSizesW(Min/Max) 0.5μL/140mL 0.5μL/255mL

Flow Rate Range (Smallest Syringe)

0.0001 to 1.08μL/min

0.0001 to 1.5679μL/min

Flow Rate Range (Largest Syringe)

0.0002 mL/minto 128 mL/min

0.0002 mL/minto 128 mL/min

DriveMecha-nism

0.9°steppermotorwith1/128 microstepping

0.9°steppermotorwith1/48 microstepping

Step Resolution 0.012μm/step 0.046μm/step

Nominal Linear Force 29kg(65lbs) 227kg(500lbs)

Accuracy ±<0.35% ±<0.35%

Reproducibility ±<0.05% ±<0.05%

Operation Temp 4°C to 40°C (40°F to 104°F)

4°C to 40°C (40°F to 104°F)

Storage Temp -10°C to70°C (14°Fto158°F)

-10°C to 70°C(14°Fto158°F)

16

Specifications (Continued)

Specification Fusion 4000 Fusion 6000

Relative Humidity 20%-80% 20%-80%

Data Ports RS-232 (DB9 and USB-B), TTL (USB-B)

RS-232 (DB9 and USB-B), TTL (USB-B)

Height 16 cm (6.3 in.) 18 cm (6.9 in.)

Width(Length) 27 cm (10.8 in.) 27 cm (10.8 in.)

Depth 20 cm (7.8 in.) 21 cm (8.2 in.)

Weight 4.5kg(10lb) 5.9kg(13lb)

Screen 4.25inchdiagonalLCD 4.25inchdiagonalLCD

17

Pump SetupToprovidepowertothepump,anIEC60320C13cableshouldbepluggedintothepowersocketonthebackofthepump.Thecableshouldbepluggedintoanappropriatepowersource(110Vor220V)withpropergrounding.Adjustthevoltage-converterswitchonthebackofthepumpsuchthatthevaluedisplayedontheswitchcorrespondswiththevoltageofthepowersupply.

Thepowerswitchforthepumpislocatedontheback,abovethepowersocket.Uponflippingtheswitchtothe“on”position,theblueLEDpowerlightonthefrontofthepumpwillimmediatelyturnon,alongwiththeredLEDlightontheStopkey.Inafewseconds,abeepwillsound,andtheLCDscreenwilldisplayaninitialstartupscreen.

The startup screen displays the version of the firmwarecurrentlyinstalledonthepumpatthetopofthescreen.Thestartupscreenwilldisplayforapproximatelyfivesecondsasindicatedbythecountdownontherightofthestartupscreen.

Afterthestartupscreen,theLCDscreenwilldisplaythelastmodethepumpwasin.

Powering On/Off the Pump

18

TheFusion4000seriescanaccommodateuptotwosyringes,whiletheFusion6000canholdonelargesyringeupto200mL.TheFusion6000canbeexpandedto4channelsusingtheChemyx4-Channel100mLSyringeRackaccessory.ThevolumeoftheFusion4000syringescanvaryfromaminimumof0.5μLuptoamaximumof140mLandtheFusion6000syringecanvaryfrom0.5μLto200mL.Syringesofanymaterial(e.g.,glass,plastic,stainlesssteel,etc.)canbeused;however,someusecasesmayrequiresyringesthataremoreaccurateanddurable.

Loading a Syringe

Fusion 4000

2 31 4

1. Ifnecessary,moveandtightentheinfusionsafetycollar(seeAdjustingtheSafetyCollar).

2. Whileunscrewingtheknobonthepusherblock,theblockwillreleasetoslidethepusherblockallthewaytotheright.

3. Loosenthescrewknobsfortheflangeretainingbracketandadjustthebracketsuchthattheflangeofthesyringebarrelwillfitinbetween.Donottightenthebracket,yet.

4. Pulluponthespring-loadedsyringeclampandplacethesyringeinoneofthechannelsoftheholderblock.Ensurethattheflangeofthesyringebarrelisflushagainsttheedgeoftheholderblock.Slowlylowerthesyringeclamptosecurethesyringe in place.

19

5 76 8

Fusion 4000 (Continued)

5. Loosenthebracketclampknobfortheplungerretainingbracketandadjustthegapsuchthattheplungercapofthesyringemayfit.

6. Unscrewtheknobtoreleasetheblockandslidethepusherblocktotherightuntilitisflushagainstthesyringeplunger.Theplungercapforthesyringeshouldfitinthegapbetweenthebracketandthepusherblock.

7. Tightenthebracketclampknobfortheplungerretainingbrackettosecurethesyringeplungerinplace.

8. Movetheflangeretainingbracketflushagainsttheflangeofthesyringebarrelandtightenthebracketclampknobstosecurethesyringebarrelflangeinplace.

20

1. If necessary,move and tighten the infusion safety rod(seeAdjustingtheSafetyRod).

2. TomovethepusherblockusetheForwardandReversefucnctionsthatcanbefoundonthemenuscreenofthepump.Movethepusherblockallthewaytotheright

3. Unscrewthesyringeclampandplacethesyringeinoneofthechannelsofthesyringe-holderblock.Ensurethattheflangeofthesyringebarrelisflushagainsttheedgeofthesyringe-holderblock.Slowly lowerandscrewthesyringeclamptosecurethesyringeinplace.

4. Usethemenuforwardandreversefunctionstomovethe

Fusion 6000

2

1

3

5 6 78

pusherblocktotheleftuntilitisflushagainstthesyringeplunger.

5. Loosenthebracketclampknobfortheplungerretainingbracketandadjustthegapsuchthattheplungercapofthesyringemayfit.

6. Usethemenuforwardandreversefunctiontomovethepusherblocktotheleftuntilitisflushagainstthesyringe

plunger.Theplungercapforthesyringeshouldfitinthegapbetweenthebracketandthepusherblock.

7. Tightenthebracketclampknobfortheplungerretainingbrackettosecurethesyringeplungerinplace.

8. Movetheflangeretainingbracketflushagainsttheflangeofthesyringebarrelandtightenthebracketclampknobstosecurethesyringebarrelflangeinplace.

21

TheFusionSeriespumpcangenerateasignificantamountof linearforce.Whilethis isextremelyusefulfordispensingviscousfluidsathigherflowrates,thehigherlinearforceoftheFusionseriescanalsopotentiallydamagethesyringesinuse.Onesourceofpotentialdamageofasyringemayoccurwhentheplungerofthesyringeiscompletedepressed,andthepusherblockisstillpushingagainstthesyringe.Dependingonthematerialofthesyringe(i.e.,glassorplastic)andthethicknessoftheplunger,thepressurefromthepusherblockcouldbend,distort,orjustbreakthesyringe.

Toprotectagainstthispotentialproblem,Fusion4000syringepumpsusesafetycollarsandFusion6000useastallrodtophysicallystopthepusherblockbeforeitcancompletelydepressthesyringeplunger.Whenapusherblockhitsthesafetymechanismandcannolongermoveforward,thepumpwillstoprunning,emitalongbeep,andenterpump-stalledmode.

Adjusting the Safety Nut/ Stall Rod

Adjusting the safety collar requires theuseof thehex key,convenientlystoredonthebackleftofthepump,rightabovethe power switch. Loosening the hex screw on the safetycollarallowsforthecollartobeslidfreelyalongtheguiderodandthensecurelytightenedtosetthepositionofthesafetycollar.Thepositionofthesafetycollarshouldbesetslightlyfurtherfromthesyringe-blockholderthanthetopofsyringeplungerwhenfullydepressed.

TheFusion6000usesastall rod insteadofa collar to limitthedistance thepusherblockmay travel.By loosening thesafetybarclampknobontopofthepusherblock,thesafetybarcanbeslidleftorright.Thesafetybarshouldbesetsuchthatthe lengthofthebarexposedontherightsideofthepusherblock is longer thanthedistancefromthetopofafullydepressedsyringe plungertotheedgeofthesyringe-holderblock.Oncethesafetybarhasbeensettotheproperlength,tighteningthesafetybarclampknobwillkeepthebar from moving.

Fusion 4000

Fusion 6000

Caution:On the Fusion Series, during withdrawal, a syringe plunger can potentially be pulled out from the syringe, which could cause the contents of the syringe to leak out. Users should be especially aware of the length of the plunger and the point where the withdrawing pusher block will exceed that length. There are no safety collars on the Fusion Series to prevent too much withdrawal.

Stall Rod adjustment

Safety collar adjustment

22

TheLCDscreendisplaysagraphicaluserinterface(GUI)fromwhichallthepumpsettingsandrunparametersareentered.Wheneveran itemiscurrentlyselected,thecolorschemeoftheelementtypicallyturnstoadarkershadeofthecolor.Forexample,thelightbluebuttonsturndarkbluewhenselected,andthewhitebackgroundsurroundinganunselectededitabletextfieldturnsbluewhenthefield isselectedandavailable forediting.Also,whenaGUIelement isselected, thetopbluebarwillshowtheactivityperformed(buttonsonly)orshowthelimitstothevaluesthatmaybeenteredfortheselectedfield(editablefieldsonly).

Selecting the itemsthatwillbemodifiedcanbedonetwoways: (1)using theLCDtouchscreenor (2)using thenavigationkeypad.

AlloftheelementsoftheGUIcanbeinteractedwithbytouch.Typicaltouchinteractionsaretappingafingeronabuttonortextfieldorslidingafingertoscroll.Asaresistivetouchscreen,theseinteractionsarenotrestrictedtoonlyunglovedfingertips.Manytypesofobjectscanbeusedtoselectitemsonthescreenwithonlyveryslightpressure,butavoidusingsharpobjects,astheymaypermanentlydamagethescreen.

ElementsoftheGUImayalsobeselectedusingthenavigationkeypadonthefrontoftheinstrumenttomovebetweentheelements.

Toactivate(or“press”)abuttonusingthenavigationkeypad,selectthebuttonelementsuchthatitishighlightedandthenpresstheENTERkeyatthecenterofnavigationkeypad.EditablefieldscanimmediatelybeediteduponselectionwithouttheneedtopresstheENTER key.

Pump Operation

Operating Interface

Limitsforselectedfields

Selected Fields

Unselected Field

Tip:Users may find the navigation keypad easier to use for scrolling through items in a scrollable list, such as the list of syringes in the syringe library.

23

AllofthescreensintheFusionSeriesGUIexistinahierarchy,withtheModeSelectionscreenatthetop.Enteringanewscreenisperformedsimplybyactivating,ortapping,abuttonoriconthatopensthenewscreen.

Mosteditablefieldswillrequiretheusertoenternumbers,whichisdoneusingthenumerickeypadonthefrontoftheinstrument.

Uponfirstselectinganeditablefield,pressinganynumberwillfirstcleartheoldvalueandreplaceitwiththenewvalue.Thisonlyapplieswhenfirstselectingthefield.Onafieldthathasalreadybeenselectedandedited,anyfurthereditswilljustappendthenumbertotheend.Insuchacase,toclearthecurrentvalueinthefield,pressthewhiteCkeyonthenumerickeypad.

Valuesenteredinaneditablefieldareaccepteduponpressingthe ENTER key, selecting a newelement,orpressingtheStart key.Valuesthatexceedthe limitsof theeditablefield(showninthebluebaratthetopofthescreen)areautomaticallycorrectedtobewithinthelimitsuponacceptingthevalue.

Toreturntothepreviousscreen,taporactivatethered“X”buttoninthetoprightofthescreen.Thiswill“close”thecurrentscreen,andreturntothescreenthatwasusedtoentertheclosedscreen.ThiscanalsobeaccomplishedbypressingtheStopkeyonthefrontoftheinstrumentwhenthepumpisnotrunning.

Navigating Between Screens

Entering Numbers

PhysicalNumericKeypad

24

Mode Selection ScreenThe Mode Selection screen is the top levelmenufortheFusionseriespumps.TheFusion4000 screen allows users to select between pumpsandCycleMode.TheFusion6000screenallows users to select flow and pressure andtemperaturecontrolingprofiles.

Fusion 4000 Mode Selection

Fusion 6000 Mode Selection

25

Pressure & Temperature Settings (Fusion 6000 Only)TheTemperature Settings screen allows you tocontrol theoptional syringeheaters for theFusion6000pump.Theoptionsonthescreenallowyoutostarttheheaterwithflow,controltemperature, pre-heat the syringe and thenstartthepumpaswellassettingatimeronthesyringeheater.Whileusing the syringeheatertheonlywaytoturnthehearteroffistounplug,to set temperatre below room temperature or usetheheatertimer.

The heater is started and stopped with everystartandstopofflow,pausingthepumpdoesnot stop the heater. The “Pre-heat then startflow”optionwillpause theflowandwiatuntilthe target temperature is reached then startflow;asepcialstatustextwillbeshownifactive.

The Pressure Settings screen allows you to control the prssure for the syringe for theFusion6000pump.Theoptionsonthescreenallowyou to stopflowat a targetpressureorholdatargetpressure.Whileusingthepressurefunctionalitytheonlywaytoturnthepressuresensor off is to unplug or to set temperatureveryhigh.

When“Stopflowattargetpressure”optionwilldisplay a special stop status text if pressure is reached. The option “Hold target pressureand modulate flow” when selected will slowdownfloaorspeedupflowtotargetflowrateto maintain constant pressure. This will alsodisplay a special status message if active and removal of edit rate.

Fusion 6000 Temperature Setting

Fusion 6000 Pressure Setting

26

Basic Mode (or Single-Step Mode) is used for applications that do not require advanced automation, variable volumeand rate settings, step looping, or ramping rates.

The Basic Mode screen can be reached by tapping on thewaterdropleticonontheMode Selection screen.

Basic Modeallowsuserstoselectthesyringebeingused,tosetavalidvolumeandrate,andtochoosewhethertoinfuseorwithdraw.

ThepumpcanonlybestartedinBasic ModewhiletheBasic Mode screen is displayed.

Basic Mode

Syringe Selection

Basic Mode Screen

Foreveryapplication,thefirstrequiredstepforsetupistoselecttheinnerdiameter(ID)ofthesyringe(s)beingused.Thisvalueisimportantforthesyringepumptoaccuratelycalculatethetransferrateandthetotaltransferredvolumefortheliquidinthesyringe.Therearetwoapproachestoenteringthisvalue:(1)enteracustominnerdiameteror(2)usethebuilt-insyringelibraryto import an inner diameter.

Enteringacustominnerdiameter(ID)isassimpleastypingintheIDinmillimetersintheSyringeeditablefield.ThemaximumandminimumvaluesfortheIDaredisplayedinthebluebaratthetopofthescreenwhiletheSyringeeditablefieldisselected.Thisvalueisalsolimitedtoonlythreedecimalplaces,whereanyadditionaldecimalsortrailingzeroesaredropped.IfthevalueoftheVolumefieldissmaller/largerthanthemin/maxvolumeallowedfortheinputinnerdiameter,thevalueintheVolumefieldwillbeadjustedtobewithinrange.Otherwise,nochangestothevolumewilloccur.

TappingtheFind SyringebuttontotherightoftheSyringe editablefieldwillopenanewSyringe Library screen. Syringe manufacturers are found in the list on the left. Selecting amanufacturerwill display all of the syringes available fromthatmanufacturerinthelistontheright.Thelistofsyringestypicallyindicatesthevolumeofthesyringefirst,followedbytheinnerdiameterofthesyringe.Bothlistscanbescrolledthroughbydraggingafingeronthescrollbar;however,theeasiestapproachinthisscreenmaybetousethenavigationkeypadtoscrollthroughthelists.Alistofthesyringesinthelibrary can be found in Appendix A.

Syringe Library Screen

27

Whenaparticularsyringeisselectedfromthelistontheright,anImport buttonwillappearnexttotheselecteditem.Tappingthe Importbutton (orpressingtheEnterkeyonthenavigationkeypad)will import the innerdiameterassociatedwiththeselectedsyringeandexittheSyringe Library screen.

Uponimportofasyringefromthesyringelibrary,theSyringeeditablefieldwillbeupdatedtothenewinnerdiameter,andtheVolumeeditablefieldwillbeupdatedtothevolumeoftheimportedsyringe.PleasenotethatthevolumeunitsandvaluesfortheVolume and RatefieldsmaychangetomatchtheunitsofvolumedisplayedforthesyringeintheSyringe Library.

Volume

TheVolumefieldrepresentsthetotalvolumeofliquidthatthepumpwilltransfer(infusionorwithdrawal)beforestopping.Theamountoftimethatthepumpwillrunwiththesettargetvolumecanbedeterminedbydividingthevolumebytheflowrate.

Thevalueenteredcanbeanyintegerordecimalwithinthelimitsforthesyringe.Themin/maxvolumelimitscanbefoundinthebluebaratthetopofthescreenwhentheVolumefieldisselected.Anyenteredvaluebeyondtheselimitswillbeautomaticallyadjustedtotheminimumormaximumlimit,whicheveriscloser.Thetotalnumberofdecimalplacesislimitedtofive.

ThevolumecanbestatedinunitsofmLorμL.TheunitsusedforthevolumearedependentuponthesetunitsofvolumefortheflowrateintheOptions screen.

Tocontinuouslyrunthepumpandusetheentirevolumeofthesyringe,userscansetthevolumetothetotalvolumeofthesyringeortothemaximumlimit.However,usersarestronglyencouragedtosetthesafetycollar/bar(seeAdjustingtheSafetyCollar)sothatthepumpdoesnotpotentiallyruinthesyringe.

Tip:Because changing the values of the syringe ID may possibly change the values and units for volume and rate, selecting a syringe or entering the syringe ID should always be the first step when setting up a run.

Important Note:If more than one syringe will be used at the same time, it is recommended that the all of the syringes be of the same inner diameter. Because Fusion pumps calculate the rate that the pushing block moves based on the input syringe inner diameter and flow rate, syringes with larger or smaller inner diameters will transfer more/less volume than indicated by the GUI.

Important Note:The maximum volume limit for a syringe assumes that the length of the syringe is the length of the pump. Therefore, this maximum volume limit is almost always larger than the actual volume the syringe can hold.

28

Flow RateTheFlow Ratefieldsindicateshowfastthepusherblockwillmove.Theamountoftimethatthepumpwillruncanbedeterminedbydividingthesetvolumebytheflowrate.

In Multi-Step Mode,theflowrateduringthestepcanbestaticorvariable,whichissetbytheratefieldsineachstep.Step 1 representstheflowratewhenthestepstartswhileStep 2representstheflowratewhenthestepends(seeStaticFlowRateandVariableFlowRateformoredetails)andsoforthtoothersteps.

Foreachfield,thevalueenteredcanbeanyintegerordecimalwithinthelimitsforthesyringe.Themin/maxflowratelimitscanbefoundinthebluebaratthetopofthescreenwhentheRatefieldisselected.Anyenteredvaluebeyondtheselimitswillbeautomaticallyadjustedtotheminimumormaximumlimit,whicheveriscloser.Thetotalnumberofdecimalplacesislimitedtofive.

Theratecanbestatedwithfourdifferentunits:mL/min,mL/hr,μL/min,andμL/hr.Adjustingtheunitswillaffectthemin/maxlimitsfortheRatefield.TheunitsforratecanbesetinOptionsduringthemulti-stepsetup.

Caution:Flow rate is often set for the needs of the experiment; however, users should be aware that setting high flow rates, especially for viscous fluids, may require high linear force beyond the limits of the pump which may cause the pump to stall out. Additionally, high flow rates can exert tremendous pressure within a syringe that may cause glass or plastic syringes to burst. In the case of high-pressure experiments, users should consider using Chemyx Stainless-Steel Syringes and the high-pressure Chemyx Fusion 6000 pump.

Infusion/Withdrawal

TheFusionSeriesiscapableofbothinfusionandwithdrawal.Onthesystems,anInfusion/WithdrawaltogglebuttonshouldappeartotherightoftheVolumefield.Thetextandarrowdirectiononthebuttonindicatesthemodethatiscurrentlysetanddirectionthatthepusherblockwillmoveuponstartingtherun.Tappingthebuttonwillswitchbetweenthetwomodes.

29

Step LoopTheStep Loopfielddetermineshowmanyadditionaltimestosequentiallyperformtheparticularstep.Thisisthelowest-levelloopthatiterateswheneverthatparticularstepisreachedandcompleted.

TheStep Loopfieldtakesanyintegerfrom0to99.Avalueof0indicatesthatthemethodwillrunthesteponlyonce.Anyvalueabovezerorepresentsadditionaltimestorunthestep.

InthePump Runningscreen,thenumberofremainingloopsofthecurrentstepappearsontheright,nexttotherepeaticon.

Start-Time DelayTheDelayfieldrepresents theamountof time (inminutesandseconds) todelaystarting thepumpafter thecurrentstepbegins.

Whenenteringthisvaluewiththenumerickeypad,thedecimalkeyisusedtoindicatetheseparationbetweenminutesandseconds.

Toremovethedelay,simplypresstheclearkey(C)orsetthetimetozero.

Example:To set a delay of 1 min and 30 seconds, press the “1” key, then the decimal key, and finally the “3” and “0” keys.

Example:In a multi-step method of five steps (1, 2, 3, 4, 5) where Step 4 has a Step Loop of “2”. The method will run the following steps sequen-tially: 1, 2, 3, 4, 4, 4, 5. When the step loop is set to “2”, Step 4 is repeated twice more. If the entire method also has a Loop All setting of “1”, the method would run the following steps sequentially: 1, 2, 3, 4, 4, 4, 5, 1, 2, 3, 4, 4, 4, 5.

30

Flow-Rate Units

TheFusionSeriesiscapableofbothinfusionandwithdrawal.Onthesystems,anInfusion/Withdrawaltogglebuttonshouldappearwhen theVolumefield is selected.Thebubble form indicates themode that is currentlysetanddirection that thepusherblockwillmoveuponstartingtherun.

Theflow-rateunitsdeterminewhichunitsofvolumeandtimewillbeusedfortheVolume and Ratefields.ItalsodeterminestheunitsforthePriming/Bolus Rate.ThissettinginBasic ModealsochangestheunitsforMulti-Step Mode.

Note:Even if the units have changed, the values within the Volume and Rate fields will not change, which will result in significant changes in the flow rate and volume transferred. Additionally, the min/max limits for these fields will also be changed, so if the current values for volume and rate exceed the new limits, they will be adjusted to be within those set limits.

Volume Settings screen

Options

31

Options

Enable Volume Lock

Primingallowsfortheusertotemporarilyrunthepumptoremovedead/emptyvolumeinthesyringeneedleortubingpriortorunningtheexperiment.

The rateatwhichprimingoccurs is set in thePriming/Bolus Rate field. Thevalueenteredcanbeany integerordecimalwithinthelimitsforthesyringe.Themin/maxflowratelimitscanbefoundintheorangebaratthetopofthescreenwhenthe Priming/Bolus Ratefieldisselected.Anyenteredvaluebeyondtheselimitswillbeautomaticallyadjustedtotheminimumormaximumlimit,whicheveriscloser.Thetotalnumberofdecimalplacesislimitedtofour.

TheunitsforthePriming/Bolus RatearedeterminedbytheFlow-RateUnitssettinginOptions.Adjustingtheunitswillaffectthemin/maxlimitsforthePriming/Bolus Ratefield.

ToenterprimingmodeonFusionpumps,pressandholdtheStartkeywhileinthemainscreenofBasic Mode.

Thisfunctionhasbeendeprecatedandisnolongersupported.

Caution:Flow rate is often set for the needs of the experiment; however, users should be aware that setting high flow rates, especially for viscous fluids, may require high linear force beyond the limits of the pump which may cause the pump to stall out. Additionally, high flow rates can exert tremendous pressure within a syringe that may cause glass or plastic syringes to burst. In the case of high-pressure experiments, users should consider using Chemyx Stainless-Steel Syringes and the high-pressure Chemyx Fusion 6000 pump.

32

Multi-Step Modeisusedformoreadvancedapplicationsthatmayrequirenumeroussteps,multipledifferentvolumeandratesettings, step looping, or ramping rates.

The Multi-Step ModescreencanbereachedbytappingontheMenuiconontheMode Selection screen.

Thepumpcanonlybestarted inMulti-Step Modewhen theMulti-Step Mode screen isshowing.Thisalsoholds true forcomputercontrolofthepumpthroughaserialconnection;thepumpmustbeswitchedtotheMulti-Step Modescreenonthepumpinterfacefirst.

Multi-Step Mode

Setup

Syringe Selection

In Multi-Step Mode, the left side of the screen shows a list ofselectableitems.Thefirst(topmost)iteminthislistisSetup. Tapping SetupwilltakeuserstotheMulti-Step Method Setup screen,wherethe syringe inner diameter (ID), total number of steps, number ofloops,andflow-rateunitscanbeset.

Syringe selection for Multi-Step ModeworksinmuchthesamemannerasseeninBasic Mode.

ChangesinthesyringeIDwillaffectthemin/maxvaluesfortheVolumefieldinallsteps.However,unlikeBasic Mode, importing asyringefromtheSyringeLibrarywillnotchangethevolumesineachsteptothevolumeofthesyringeunlesstheyexceedthevolumelimitsofthenewsyringe.Allvaluesineachstepthatareoutsidethesenewlimitswillbeadjustedtobewithinthelimits.

Multi-Step Mode Setup screen

Tip:Because changing the values of the syringe ID may possibly change the values and units for volume and rate, selecting a syringe or entering the syringe ID should always be the first step when setting up a run.

33

Total Steps

Loop All

Sub-Loops

TheTotal Stepsfielddetermineshowmanystepswilloccurinthemethod.EachstephasindependentvaluesforVolume, Rate, Delay, and Infusion/Withdrawal.Uponenteringthenumberofsteps,thelistontheleftofthescreenwillupdatetoshowallofthesteps.Eachoneofthestepswillhavetheirownsetofparametersthatwillneedtobeset(seeStepParameters).

Aminimumof1andmaximumof99stepscanbeenteredasamethod.

AddingstepsisaccomplishedbyincreasingtheTotal Steps,whichwilladdthenewstepsattheendofthelist.DecreasingtheTotal Stepswillremovestepsfromtheendofthelist.Ineithercase,thevaluesintheStepParametersforeachexistingstepdoesnotchangeupontheadditionordeletionofsteps.

Inthecasethatthenumberofstepsinthelistontheleftsideofthescreenexceedstheavailablespaceonthescreen,abluetrianglearrowwillappearbelowthelistindicatingthatitcanscroll.Thislistcanscrolltwoways:(1)theusercandragtheirfingerupanddownonthelisttoscrollthelistandthentapontheitemofinterest;or(2)theusercanusetheup/downarrowsinthenavigationkeypadtoscrollthroughtheitems.

TheLoop Allfielddetermineshowmanyadditionaltimesthepumpwillperformallofthesteps.Thisisthehighestlevelloopandisthelastlooptoiterateafterallthestepsandsteploopshavecompleted.

Thisfunctionhasbeendeprecatedandisnotsupported.Please leave these fields empty.

TheLoop Allfieldtakesanyintegerfrom0to99.Avalueof0indicatesthatthemethodwillrunthroughtheentiretyofstepsonlyonce.Anyvalueabove0representsadditionaltimestorunallofthesteps.

Inthepumprunningscreen,thecurrentloopnumberandtotalnumberofloopsareindicatedinthebluebaratthetopofthescreen.

Example:In a multi-step method of three steps (1, 2, 3) where the Loop All is set to “2”. The method will run the following steps sequentially: 1, 2, 3, 1, 2, 3, 1, 2, 3. As shown, the method will loop through all three of the steps two additional times.

34

Flow-Rate UnitsTappingtheVolume or Pump Ratebuttonopensascreenthatallowsflow-rateunitstobeset.Theflow-rateunitdetermineswhichunitofvolumeandtimewillbeusedfortheVolume and Ratefieldsforallofthesteps.ThissettinginMulti-StepModealsochangestheunitsforBasicMode.

Every step in themulti-stepmethod will have a selectableentryinthelist.Atthetopofthescreenyouwillfindaplacetoenteryoursyringvsizeaswellasthenumberofstepsthemulti-stepmethod to take. Tapping the stepwill allow youtochange theVolumeandRate.Thearrowson this screenrepresentinfuseorwithdrawforeachstep.

Important Note:Even if the units have changed, the values within the Volume and Rate fields will not change, which will result in significant changes in the flow rate and volume transferred. Additionally, the min/max limits for these fields will also change, so if the current values for volume and rate exceed the new limits, they will be adjusted to be within the limits.

Multi-Step Mode Options screen

Multi Step Program Screen

Step Parameters

35

VolumeTheVolumefieldrepresentsthetotalvolumeofliquidthatthepumpwillattempttotransfer(infusionorwithdrawal)beforecompletingthestep.Theamountoftimethatthepumpwillattempttorunwiththesetvolumecanbedeterminedbydividingthevolumebytheflowrate(aslongastheflowrateisstatic).

Thevalueenteredcanbeanyintegerordecimalwithinthelimitsforthesyringe.Themin/maxvolumelimitscanbefoundinthebluebaratthetopofthescreenwhentheVolumefieldisselected.Anyenteredvaluebeyondtheselimitswillbeautomaticallyadjustedtotheminimumormaximumlimit,whicheveriscloser.Thetotalnumberofdecimalplacesislimitedtofive.

ThevolumecanbestatedinunitsofmLorμL.TheunitsusedforthevolumearedependentuponthesetunitsofvolumefortheflowrateintheOptionsscreenduringthemulti-stepsetup.

Tocontinuouslyrunthepumpandusetheentirevolumeofthesyringe,userscansetthevolumetothetotalvolumeofthesyringeortothemaximumlimit.However,usersarestronglyencouragedtosetthesafetycollar/bar(seeAdjustingtheSafetyCollar)sothatthepumpdoesnotpotentiallyruinthesyringe.

Important Note:Even if the units have changed, the values within the Volume and Rate fields will not change, which will result in significant changes in the flow rate and volume transferred. Additionally, the min/max limits for these fields will also change, so if the current values for volume and rate exceed the new limits, they will be adjusted to be within the limits.

36

Thepump’sactionsarecontrolledbytheStop, Pause, and Start keysonthefront.Inordertostartarun,thescreenshouldbeshowingeithertheBasic Mode (for basic runs) or Multi-Step Mode(formulti-stepruns)mainscreens.PressingtheStartkeyineitherotherthesescreenswillstartarunwiththecurrentlysetparameters.

Amount Functions

Start

PressingtheStart buttononcewillstartanewrunwiththecurrentlydisplayedparametersorresumeapausedrun.ThisbuttonisonlyfunctionalwhenaruniscurrentlypausedorwhenthedisplayisshowingtheBasic Mode or Multi-Step Mode screensthatshowthecurrentparameters.

PressingtheStartbuttonwhileadelaycountdownisoccurringwillimmediatelyskiptheremainingdelaytimeandstarttherun/step.

PressingandholdingtheStartbuttonwillbeginpriming,wherethepumpwillcontinuouslyrunatthePrime/BolusRateaslongasthebuttonisbeingpressed.ThePrime/Bolus Rate is set in Basic Mode > Options. Priming is only available in Basic Mode.

ThegreenLEDlightontheStartbuttonwillbeonwhilethepumpisrunning.

Pause

Whilethepumpisrunning,pressingthePausebuttonwillpausethecurrentrun.Pausingallowsthepumptocontinuethesamerunfromthepointwhenthepauseoccurred.

TheyellowLEDlightonthePausebuttonwillbeonwhenthecurrentrunhasbeenpaused.

Stop

Duringarun,pressingtheStopbuttonwillimmediatelystoptherun,andreturnthedisplaytotheBasic Mode or Multi-Step Mode screen.

Whileoutsideofanactiverun,pressingtheStopbuttonwillexitthecurrentscreen.Thisresultisthesameasactivatingthe“X”buttoninthetoprightcornerofthescreenbytouchornavigationalkeypad.

TheredLEDlightontheStopbuttonwillbeonwheneverthepumpisneitheractivelyrunningnorpaused.Thelightwillalsobeonforactiverunsthatarecurrentlyinadelay before starting a step.

TheredLEDlightwillblinkifthepumphasstalled.

Uponstartingapumprun,thescreenwillswitchtotheSystem Statusscreen.TherearetwodifferentversionsofthisscreendependinguponwhetherthepumpwasstartedinBasic Mode or Multi-Step Mode.

Pump Control

37

Run Status

The Run Status screen for Basic Modehighlightsthecurrentactivityofthepumpduringarun.

TheElapsed Timefieldindicateshowmuchtimehaspassedsincethepumphasstartedphysicallyrunning,notincludinganystart-time delay. The Infusion Timefieldshowsthecalculatedtotaltimeforthecurrentruntofinish,barringanypauseorstallintheinstrument.

Thetworectangularblocksonthescreenillustratethetransferofvolumeduringtherun.Theleftrectanglerepresentsthevolumeinthesyringewhiletherightrectanglerepresentsthevolumetransferredinoroutofthesyringe.Whilethepumpisrunning,asmallarrowwillappearbetweenthetworectanglesindicatingthedirectionofflow.Dependinguponthedirectionofthearrow,thenumberswithineachrectanglewithchangeasvolumeistransferred.

Thecoloredtextdirectlynexttothevolumerateamountindicatesthecurrentstatusofthepump:green-running;yellow-pausedordelayed;andred-stoppedorstalled.

The Volume,andRatefieldsindicatetheparametersthatweresetforthecurrentlyrunningmethodintheBasic Mode main screen.

TappingtheEdit Ratebuttonallowsuserstochangetheflowrateofthepumpinthemiddleoftherun.

Basic Mode System Status screen

38

Start Options

Start Options

When starting the Fusion Series pump a display of startoptions will be shown. User will be required to select oneoptionbasedon thepump thatneeds tobestarted.Otherfunctionsarealsoavalibletopauseandstopthepumponcethepumphasstarted.

Pause Options

Stop Options

Start Options

Pause Options

39

Thebaudrateistherateatwhichinformationistransferredthroughtheserialconnection.Thisrateisoftenlimitedbythetypeof serial connection used.

TheFusionSeriespumpsprovidetwopossiblebaudrates:9600and38400.

SerialconnectionsusingtheUSB-Bportonthebackofthepumpshouldusethefasterbaudrateof38400.

SerialconnectionsusingtheDB9portonthebackofthepumpwillneedtousetheslowerbaudrateof9600.

See Pump Control by Computer for more information.

Baud Rate

TheSystem SettingsscreencanbereachedbytappingontheMenuiconontheMode Selection screen.

System Settings

System Settings screen

Power Settings screen

TheSystem SettingsscreencanbereachedbytappingonthePowericonontheMode Selection screen.

40

TheFusion4000andFusion6000pumpscangenerateuptoamaximumof65lbsand500lbsoflinearforce,respectively.Somesyringesmadeofmorefragilematerialsmaynotbeabletowithstandthefulllinearforceofthepump.OntheFusionSeriespumps,themotorpower,andthusthelinearforce,canbedecreasedintheSystem Settings screen.

Toadjustthissettingbytouch,tapthesectionofthepercentagetogglethatrepresentsthedesiredmotorpowerpercentage(1000%or50%).

TheLCDdisplaygoestosleepafter1hourafterthelastinteractionwiththepumpcontrols.Whenthedisplaygoestosleep,thepumpisstillon.Itwillcontinuetorunthecurrentmethodandisabletoreceivecommandsviacomputerconnection.

Motor Force

LCD Power-Off Time

Important Note:Even if the units have changed, the values within the Volume and Rate fields will not change, which will result in significant changes in the flow rate and volume transferred. Additionally, the min/max limits for these fields will also change, so if the current values for volume and rate exceed the new limits, they will be adjusted to be within the limits.

41

Pump Control by ComputerTheFusionSeriessyringepumpscanbecontrolledwithanexternalcomputerusinganRS232serialconnection.

Twotypesofconnectionsareavailableforexternalcomputercontrol:DB9andUSB,cablesforeachareavailableforpurchasethroughChemyx(call1-281-277-5499orvisitwww.chemyx.com/contact-us).

Mostcomputeroperatingsystems (e.g.,Windows,macOS,Linux,etc.)havebuilt-in terminalprograms thatcanbeused totypeinthecommandsthatrunthepump.OnWindowsXPandearlier,HyperTerminalcanbeusedtocontroltothepump.OnmacOSandLinux,thescreencommandinTerminalcanalsobeusedtocontrolthepump.Therearealsonumerousthird-partyprogramsforserialconnectionsthatexistforeachoperatingsystemthatmaybeeasiertouseandprovidemorefeatures.Athird-partyprogramwillbenecessarytoconnectwithcomputersrunningWindowsVistaandlater.

For tipsonoperating thebuilt-in terminalprogramsand informationonavailable third-partyprograms,please visitwww.chemyx.com/support.

FortheDB9serialconnection,aMale-to-FemaleDB9cablemustbeused.Donotuseany“nullmodem,”“crossover”,or“crossed-over”cables.Thecablecanbeofanylengthupto50ft;however,thespeedoftheconnectionisoftendetrimentallyaffectedatlengthsover15ft.

ForaUSBserialconnection,aUSB-AtoUSB-Bcableshouldbeused.ThemaximumpossiblelengthofasingleUSBcableis15ft;however,USBcablesshouldbelessthan10ftforoptimumperformance.MultipleUSBcablescanbechainedtogethertomakealongerconnectionaslongasthehubconnectingeachindividualcableisconnectedtoapowersupply.

AUSB-to-DB9serialadaptercanbeusedtoconnectaDB9serialconnectiontotheUSBportonacomputer.

Alloftheseconnectionsrequiretheproperdriverstobeinstalledonthecomputer.Typically,mostcurrentoperatingsystemsalreadycontaindriversfortheUSBserialconnection.TheymayalsoalreadyhavedriversfortheDB9serialconnection.USB-to-DB9adapterstypicallyneedtohaveaUSB-to-serialdriverinstalled.Formoreinformationondownloadingandinstallingdriverspleasevisitwww.chemyx.com/support.

DB9 Serial Cable (Male-to-Female) USB-A to USB-B cable

Cable Requirements

Connecting

42

Regardlessoftheoperatingsystembeingused,theterminalprogramcontrollingtheserialconnectionshouldusethefollowingconnection settings:

ThebaudratesettingfortheserialconnectionmustmatchthebaudratesetinthepumpsettingsGUI(SeeBaudRateinSystemSettings).

Generally,thepumpshouldbesettousethehigherbaudrate(38400)ifaUSB-onlyconnectionisused.However,ifaDB9-onlyorUSB-to-DB9connectionisused,thebaudratemustbesetto9600.

Typically,ifthebaudrateisnotsetcorrectly,therewillbegarbledornoresponsetoanycommands

COM Port Settings

Baud Rate 9600 or 38400

Data Bits 8

Parity None

Stop Bits 1

Flow Control None

43

Regardlessoftheoperatingsystembeingused,theterminalprogramcontrollingtheserialconnectionshouldusethefollowingconnection settings:

Command Description

help Displays a list of available commands.

start Startsthepumpwiththecurrentlysetparameters.Theparametersthatwillrundependuponwhichmode(Basic Mode or Multi-Step Mode)thepumptouchscreenisshowing.

pause Pausesthecurrentpumprun.Theruncanberesumedfromthispointusingthe start com-mand.

stop Endsthecurrentrun.Theruncannotberesumedandcanonlyberestartedfromthebegin-ning.

set diameterX

Setstheinnerdiameter(ID)ofthesyringeinmillimeters.

ThissettingsimultaneouslyaffectsthesyringeIDusedinbothBasic Mode and Multi-Step Modesettings,whichwillalsoaffectthevolumeandrateparametersinbothmodes.

Xcanbeanintegerordecimal(uptothreedecimalplaces)from0.103mmto40.000mm.

Example set diameter 11.73

Returnsthediameter.

setunitsX

Setstheunitsusedfortherateofinfusion/withdrawal.Thisparameteralsosetstheunitsfortransfervolume,wherethetransfervolumeunitwillbethesameasthevolumeunitusedintherate.

ThissettingsimultaneouslychangestheunitsusedinbothBasic Mode and Multi-Step Mode settings.

Xcanbeanintegerfrom0to3.Eachintegerrepresentsaspecificrateunit.

0 = mL/min1=mL/hr2=μL/min3=μL/hr

Example set units 2

Returnstheintegercorrespondingtothesetrateunit.

Pump Commands

44

Command Description

set volume X,X,X,...

Setsthetransfervolume.Thepumpwillrununtilthevolumeinfusedorwithdrawnequalsthisvolume.Thiscommandalsosetswhetherthepumpisinfusingorwithdrawing(Fusion200only)andthenumberofstepsusedinMulti-StepMode.

Xmustbeanintegerordecimal(uptofivedecimalplaces)withinthevolumelimitsforthesyringe(Seereadlimitparametercommand).Anegativesigncanbeaddedinfrontofanyvaluetoindicateaswitchtowithdrawalmode(Fusion200only).

For Basic Mode, only one value is required. Entering only one value will automatically switchthepumptoBasicMode.

Example set volume 25.0

TosetvolumesinMulti-StepMode,thepumpmustbemanuallyswitchedtoMulti-StepModeonthepump.Foreachstep,avaluemustbeentered,separatedbycommasandnospaces.Thenumberofstepswillbeautomaticallyadjustedtothenumberofvaluesentered(i.e.,enteringonlythreevalueswillchangethenumberofstepstothree).

Example set volume 3,-2,4

Returnsthetransfervolume,transferrate,andtransfertime(inBasicMode).

Returnsonlythetransfervolumes(inMulti-StepMode).

settimeX,X,X,...

Setsthetargettimeforapumprun.Thiswilladjustthetransferratesothattherunwilllastfortheinputtime.Thetargetvolumewillremainthesame.ThiscommandalsosetsthenumberofstepsusedinMulti-StepMode.

Xmustbeanintegerordecimal(uptofivedecimalplaces).Theresultingtransferratemustbewithintheratelimitsforthesyringe(Seereadlimitparametercommand).

For Basic Mode, only one value[ is required. Entering only one value will automatically switchthepumptoBasicMode.

Example set time 1.0

TosettargettimesinMulti-StepMode,thepumpmustbemanuallyswitchedtoMulti-StepModeonthepump.Foreachstep,avaluemustbeentered,separatedbycommasandnospaces.Thenumberofstepswillbeautomaticallyadjustedtothenumberofvaluesentered(i.e.,enteringonlythreevalueswillchangethenumberofstepstothree).

Example set time 1.0,0.5,1.3

Returns target time and rate (in Basic Mode).

Returns no values (in Multi-Step Mode).

45

Command Description

setrateX,X,X,...

Setsthetransferrateforinfusion/withdrawal.

Xmustbeanintegerordecimal(uptofivedecimalplaces)withintheratelimitsforthesyringe (See read limit parameter command).

For Basic Mode, only one value is required.

Example set rate 1.0

TosetratesinMulti-StepMode,thepumpmustbemanuallyswitchedtoMulti-StepModeonthepump.Foreachstep,avaluemustbeentered,separatedbycommasandno spaces.

Example set rate 1.0,0.5,1.3

Ifthenumberofvaluesislessthanthenumberofsteps,theremainingstepswillgainthesameratevalueasthelastenteredvalue.Anyextravaluesbeyondthenumberofsteps will be truncated.

VariableratesmayalsobesetinMulti-StepMode.Rate1andRate2foreachstepcanbesetusingabackslash.Eachstepisstillseparatedbyacomma.

Example set rate 0.5/1.0,0.5/1.0,1.3

Returnstheflowrateandtransfertime(inBasicMode).

Returnstheflowrate1andflowrate2foreachstep(inMulti-StepMode).

46

Command Description

setdelayX,X,X,...

Setsthestarttimedelay(inminutes)priortoeachstepofapumprun.Uponstartingtheastep,thepumpwillwaituntilthedelaytimeelapsesbeforestartingtotransfervolume.

Xmustbeanintegerordecimal(uptofivedecimalplaces)withintheratelimitsforthesyringe(Seereadlimitparametercommand).

For Basic Mode, only one value is required. Entering only one value will automatically switchthepumptoBasicMode.

Example set delay 1.0

TosetratesinMulti-StepMode,thepumpmustbemanuallyswitchedtoMulti-StepModeonthepump.ForeachstepofMulti-StepMode,avaluemustbeentered,separated by commas and no spaces.

Example set delay 1.0,0.5,1.3

Ifthenumberofvaluesislessthanthenumberofsteps,theremainingstepswillgainthesamedelayvalueasthelastenteredvalue.Anyextravaluesbeyondthenumberof steps will be truncated.

Returnsthedelaytime(inBasicMode).

Returnsthedelaytimeforeachstep(inMulti-StepMode).

read limit parameter

Returnsthemin/maxvaluesthatcanbesetfortransfervolumeandtransferrate.Thereturnedvaluesaredependentonthesyringeinnerdiameter(ID).Thetransfervolumelimitsaretypicallygreaterthanthemaximumvolumeofthesyringebeingused.

Returns: max rate, min rate, max volume, min volume

dispensed volume Returnsthetransferredvolumeforthecurrentrun/step.Ifnorunhasstarted,returnsthetransferredvolumeforthelastrun/step.

elapsed timeReturnstheelapsedtime(inminutes)forthecurrentrun/step.Ifnorunhasstarted,returnstheelapsedtimeforthelastrun/step.Elapsedtimedoesnotincludeanydelay time.

view parameter

Returnsthecurrentlysetparameters(BasicModeonly).

Returns:rateunits,syringeinnerdiameter,transferrate,primingrate,time(inwholenumberminutes),transfervolume,timedelay(inwholenumberminutes)

47

Command Description

status

Returnsthecurrentstatusofthepump.Thereturnedintegercorrelatestothestatusofthepump.

0: Pump stopped1: Pump running2: Pump paused3: Pump delayed4: Pump stalled

Returns:integerthatrepresentsthecurrentstatusofthepump.

hexw2XXXXXXstart

Sets multiple parameters in one command (Basic Mode only).

Xvariablesinorder:units: see set units commandtransfermode:0-infusion,1-withdrawalsyringe ID: see set diameter commandtransfervolume:seesetvolumecommand;nonegativestransfer rate: see set rate commandstart time delay: see set delay command

ExampleThe following example shows the command using these parameters: units = mL/min; transfer mode = withdrawal; diameter = 23.05 mm; volume = 1.2 mL; rate = 3.0 mL/min; delay = 30 sec (0.5 min) hexw2 0 1 23.04 1.2 3.0 0.5

Astartcommandcanbeaddedafterthesixparameterstoimmediatelystarttherunwiththenewparameters.

Example hexw2 0 1 23.04 1.2 3.0 0.5 start

Allsixparametersdonotneedtobeentered;however,allparametersuptotheparameter(s)ofinterestmustbealsobeincluded.Theotherparametersnotenteredwillremainastheyarecurrentlyset.

Example hexw2 0 1 23.04 2.4

save setting Deprecated command.

48

Command Description

setprimerateX

Setsthepriming/bolusrate(BasicModeonly).

Xmustbeanintegerordecimal(uptofivedecimalplaces)withintheratelimitsforthesyringe(Seereadlimitparametercommand).

Only one value is required.

Example set primerate 1.0

Returnsthepriming/bolusrate(inBasicMode).

restart Powersoffandthenautomaticallypowersonthepump.

Insomecases,dependingonthesoftwareusedtoconnecttothepump,eachcommandmayneedtobefollowedbyalinefeedcharactersuchas“\n”or“\r”toexecute.Ifacommanddoesnotappeartowork,tryadding“\r”or“\n”totheendoftheline.

Whenthepumpdoesnotrecognizeacommand,itwillreturntextsimilartothefollowing:

Exampleset units 1 \r

ExampleCommand not recognized-type in “help”and press enter to see a command list.

In addition to the manual control of the pumps through a terminal, Chemyx provides pump controller programs as anunsupportedopen-sourceapplicationswritteninPython,aLabVIEWpackage,andaMATLABpackage.Moreinformationonthesetoolscanbefoundonthewebsiteatwww.chemyx.com/support.

Chemyx Pump Controller Programs

49

Pumplubricationisthemostimportantpartofregularpumpmaintenance.Properlubricationensureslongservicelife,smoothflowperformance,andprotectionagainstcorrosivelabfumes.Thepumpmaynotbeadequatelycleanedandlubricatedifthefollowingoccur:(1)grindingsoundsarecomingfromtheleadscrew;(2)theblockreleasebuttondecouplesundertheloadbeforethepumpstalls;or(3)thepumpstallswithincreasedfrequency.

Please contact Chemyx (call +1 281-277-5499 or visit www.chemyx.com/contact-us) to confirm the safe use of a differentlubricant.

1. Cleancontactsurfaceswithacleancloth.Removeanydebrisontheleadscrewandtheguiderods.

2. Applylubricantalongthelengthoftheleadscrewandtheguiderods.

3. Presstheblock-releasebuttonandslidethepusherblocktotheleftedge.

4. Runthepumpend-to-endwithoutasyringetoworkthelubricantintothecomponents.

Pumplubricationshouldbeperformedeverythreemonths.

Pump Maintenance

• Engine/motor oil

• Silicone oil

• Machineoil

• Toolingoilor“WD-40”

• Lithium-basedgrease

Lubricating the Pump

Lubrication Instructions

Approved Lubricants

50

TheChemyxFootswitchprovidesuserstheabilitytostart,pause,orstopthepumpfromashortdistancewiththetapofafoot.

TheChemyx4-ChannelSyringeRackfortheFusion6000providesuserswithanexpansionofthenumberofsyringesthatcanbeusedsimultaneously.The4channelsyringerackcanaccommodateupto4syringeswithvolumesfrom 10 mL to 100 mL.

TheChemyx Footswitch for Fusion Series pumpshas aUSB-B cable thatplugsintotheTTLportonthebackofthepump.

TousetheFootswitch,thepumpmustbeineithertheBasicModeorMulti-StepModescreens.ThebasiccommandsaregivenbyacertainnumberofclicksoftheFootswitch.

AsingleclickoftheFootswitchwillstartthepumprunorpausethepumprundependingonwhetherthepumpiscurrentlyrunning.

AquickdoubleclickoftheFootswitchwillcompletelystopthepumprun.

Footswitch

Footswitch

Using the Footswitch

Pump Accessories

Footswitch

4-ChannelSyringeRack

4-Channel Syringe Rack (Fusion 6000 Only)

51

Attaching the Rack

Fusion 200

SyringeHolderwith10Channels

Flange Retaining Bracket

Clamp Extensions

Clamp Extensions

Plunger Retaining Bracket

Syringe Clamp

52

Size Diameter

6.00 mL 9.53mm

20.00 mL 19.13 mm

50.00mL 28.6 mm

Size Diameter

0.50mL 4.64 mm

1.00 mL 4.64 mm

2.00 mL 8.93 mm

2.50mL 8.66 mm

3.00 mL 8.93 mm

5.00mL 11.73 mm

10.00 mL 14.68 mm

20.00 mL 19.60 mm

30.00 mL 22.70 mm

50.00mL 28.03 mm

60.00 mL 28.60 mm

Size Diameter

0.50μL 0.10 mm

1.00μL 0.15mm

2.00μL 0.21 mm

5.00μL 0.33 mm

10.00μL 0.48 mm

25.00μL 0.73 mm

50.00μL 1.03 mm

100.00μL 1.46 mm

250.00μL 2.30 mm

500.00μL 3.26 mm

1.00 mL 4.61 mm

2.50mL 7.28 mm

5.00mL 10.30 mm

10.00 mL 14.57mm

25.00mL 23.03 mm

50.00mL 32.57mm

Size Diameter

1.00 mL 4.64 mm

3.00 mL 4.64 mm

5.00mL 12.07 mm

10.00 mL 14.50mm

20.00 mL 19.13 mm

30.00 mL 21.69 mm

60.00 mL 26.72 mm

Size Diameter

1.00 mL 4.80 mm

1.00 mL 6.70 mm

2.00 mL 6.70 mm

2.00 mL 9.20 mm

3.00 mL 10.30 mm

5.00mL 12.20 mm

10.00 mL 15.00mm

20.00 mL 19.00 mm

30.00 mL 22.50mm

50.00mL 25.50mm

100.00 mL 34.00 mm

Chemyx Stainless Steel Hamilton Glass

Hoshi

BD Glass

BD Plastic

AppendicesAppendix A: Syringe Volume/Diameter Reference

53

Size Diameter

1.00 mL 4.66 mm

2.00 mL 6.90 mm

2.50mL 9.10 mm

5.00mL 12.62 mm

10.00 mL 14.34 mm

20.00 mL 19.68 mm

30.00 mL 22.44 mm

50.00mL 28.80 mm

100.00 mL 36.68 mm

Size Diameter

0.25mL 3.45mm

0.50mL 3.45mm

1.00 mL 4.50mm

2.00 mL 8.92 mm

3.00 mL 8.99 mm

5.00mL 11.70 mm

10.00 mL 14.70 mm

20.00 mL 19.58mm

30.00 mL 22.70 mm

50.00mL 29.00 mm

Size Diameter

0.25mL 2.60 mm

0.50mL 3.20 mm

1.00 mL 4.30 mm

2.00 mL 6.30 mm

3.00 mL 7.30 mm

5.00mL 9.50mm

Size Diameter

2.00 mL 9.12 mm

5.00mL 12.34 mm

10.00 mL 14.55mm

20.00 mL 19.86 mm

30.00 mL 23.20 mm

50.00mL 27.60 mm

Size Diameter

1.00 mL 4.71 mm

3.00 mL 9.50mm

5.00mL 12.91 mm

10.00 mL 15.40mm

20.00 mL 19.30 mm

30.00 mL 23.40 mm

50.00mL 29.00 mm

Size Diameter

1.00 mL 4.70 mm

2.50mL 9.70 mm

5.00mL 12.48 mm

10.00 mL 15.89mm

20.00 mL 20.00 mm

30.00 mL 22.50mm

50.00mL 28.90 mm

JMC Air-Tite Plastic Norm-Ject Plastic

Popper & Sons Glass

Ranfac

Natsume

Nipro

54

Size Diameter

25.00μL 0.73 mm

50.00μL 1.03 mm

100.00μL 1.46 mm

250.00μL 2.30 mm

500.00μL 3.26 mm

1.00 mL 4.61 mm

2.50mL 7.28 mm

5.00mL 10.30 mm

10.00 mL 14.57mm

Size Diameter

1.00 mL 4.70 mm

2.00 mL 6.40 mm

3.00 mL 9.30 mm

6.00 mL 13.10 mm

12.00 mL 15.40mm

25.00mL 21.00 mm

30.00 mL 23.00 mm

50.00mL 29.00 mm

Size Diameter

1.00 mL 4.65mm

3.00 mL 8.94 mm

6.00 mL 12.70 mm

12.00 mL 15.90mm

20.00 mL 20.40 mm

35.00mL 23.80 mm

50.00mL 26.60 mm

Size Diameter

1.00 mL 4.73 mm

3.00 mL 9.00 mm

5.00mL 13.04 mm

10.00 mL 15.79mm

20.00 mL 20.18 mm

30.00 mL 23.36 mm

60.00 mL 29.45mm

Size Diameter

1.00 mL 4.73 mm

1.00 mL 6.50mm

3.00 mL 8.65mm

5.00mL 13.00 mm

10.00 mL 15.80mm

20.00 mL 20.15mm

30.00 mL 23.10 mm

50.00mL 29.10 mm

SGE Glass Terumo Japan

TopSherwood Plastic

Terumo

55

Size Diameter

10.00μL 0.46 mm

25.00μL 0.73 mm

50.00μL 1.03 mm

100.00μL 1.46 mm

250.00μL 2.30 mm

500.00μL 3.26 mm

1000.00μL 4.61 mm

Size Diameter

1.00 mL 4.65mm

3.00 mL 8.94 mm

6.00 mL 12.70 mm

12.00 mL 15.90mm

20.00 mL 20.40 mm

35.00mL 23.80 mm

60.00 mL 26.60 mm

Unimetrics

Kendall Monoject Plastic

56

Item Description Part Number

Fusion100Touch 07100

Fusion200Touch 07200

Fusion400Touch 07400

Item Description Part Number

Footswitch(w/USB-Bconnectioncable) 50002

DB9 Serial Cable (Male-to-Female) 50007

USB Serial Cable (USB-B-to-USB-A) 50017

10-ChannelSyringeRack(Fusion200only) 50003

Item Description Part Number

Stainless-Steel Syringe (6 mL) SS6

Stainless-Steel Syringe (20 mL) SS20

Stainless-SteelSyringe(50mL) SS50

O-Ring Replacement 10-pack (6 mL) CP6

O-Ring Replacement 10-pack (20 mL) CP20

O-RingReplacement10-pack(50mL) CP50

Fusion Series Syringe Pumps

Options and Accessories

Stainless-Steel Syringes

Appendix B: Available Parts

57

Chemyx Inc.

10905CashRd.

Stafford,TX77477

Webpage:www.chemyx.com

ContactUs:www.chemyx.com/contact-us

Phone:+1281-277-5499

Fax:+1281-277-5499

OfficeHours:Monday–Friday,8:30am–5pm(CentralTime)

Formorearticlesandtutorialvideos,visit:www.chemyx.com/support

Follow Us

LinkedIn:www.linkedin.com/company/chemyx

Twitter:www.twitter.com/chemyxinc

Appendix C: Contact Information