QED Environmental Systems Ltd. Cyan Park - Unit 3, Jimmy Hill Way, Coventry, CV2 4QP, UNITED KINGDOM FEATURES 0 - 1000 ppm N 2 O Storage for 1000 readings TWA calculated EH40 occupaonal exposure limits calculated Leak detecon User seable alarms Data download for graphing and reporng The G200 N 2 O analyser is designed to safety check background and breathing zone levels of N 2 O (0-1000ppm) in medical applicaons. SECTOR Medical gas APPLICATIONS Operang theatres Dental pracces Veterinary clinics X-ray departments BENEFITS Accurate verificaon of exposure limit breaches Dual purpose background analyser or personal analyser. Leak detecon for N 2 O storage Highly portable is running efficiently N 2 O GAS ANALYSER | MEDICAL STAFF SAFETY qedenv.com [email protected] +44 (0)333 800 0088 PAGE 1 OF 2 | DS15-ISSUE.11 © Product designs and specificaons are subject to change without noce. User is responsible for determining suitability of product. G 200
Transcript
Page 1: G200 - Keison Products · The G200 N. 2. O analyser is designed to safety check background and breathing zone levels of N. 2. O (0-1000ppm) in medical applications. SECTOR. cal gasMedi.

QED Environmental Systems Ltd.Cyan Park - Unit 3, Jimmy Hill Way, Coventry, CV2 4QP, UNITED KINGDOM

FEATURES

• 0 - 1000 ppm N2O• Storage for 1000 readings• TWA calculated• EH40occupationalexposure

limits calculated• Leakdetection• Usersettablealarms• Data download for graphing

andreporting

The G200 N2O analyser is designed to safety

check background and breathing zone levels of N2O

(0-1000ppm) in medical applications.

SECTOR

Medical gas

APPLICATIONS

• Operatingtheatres

• Dentalpractices

• Veterinary clinics

• X-ray departments

BENEFITS

• Accurateverificationofexposurelimitbreaches

• Dualpurposebackgroundanalyseror personal analyser.

• LeakdetectionforN2O storage• Highlyportableisrunningefficiently

N2O GAS ANALYSER | MEDICAL STAFF SAFETY

qedenv.com [email protected] +44 (0)333 800 0088PAGE1OF2|DS15-ISSUE.11

© Productdesignsandspecificationsaresubjecttochangewithoutnotice.Userisresponsiblefordeterminingsuitabilityofproduct.

G200

Page 2: G200 - Keison Products · The G200 N. 2. O analyser is designed to safety check background and breathing zone levels of N. 2. O (0-1000ppm) in medical applications. SECTOR. cal gasMedi.

POWER SUPPLY

Batterytype Li Ion

Batterylife 12 hours (10 hours with pump)

Batterylifetime 600 cycles

Batterycharger 5Vdcexternalpowersupplyandinternalchargingcircuit

Chargetime Approximately4hoursfromcompletedischarge

Alternativepower 5Vdc power supply

GAS RANGES

Gases measured N2O By custom dual wavelength infra-red cell with reference channel

Range N2O0 - 1,000ppm

0 -10,000ppm(leakdetectionmode,indicationonly)

Typical accuracy* N2O

Resolution:1ppm

±5ppmfor0-100ppmaftercalibration

±1.5%ofrangefor101ppmandabove,aftercalibration

Responsetime,T90 N2O ≤40seconds

*Typical accuracies Alltypicalaccuraciesquotedareaftercalibrationplusaccuracyofcalibrationgasused.

FACILITIES

Visualandaudiblealarm UserselectableN2O and TWA alarms

Communications USB type B mini-connector, HID device class

Data Storage1000 reading sets + 270 events

50 site ID’s and 300 sample point ID’s

PUMP

Flow 100cc / min typically

ENVIRONMENTAL CONDITIONS

Operatingtemperature 0°C to 50°C

Barometric pressure 500to1500mb

Relativehumidity 5% to 95% non condensing

IPrating IP40

PHYSICAL

Weight 500 grams

Size L 165mm, W 100mm, D 55mm

Case material ABS/PolypropylenewithSiliconeRubberInserts

Keys 17ResincappedSiliconerubberkeys

DisplayLiquidcrystaldisplay,128x64pixel

WithRGBLEDback-light

Gas Sample Filters UserreplaceablePTFEwatertrapfilter,G1.10-SodalimeCO2filter

CERTIFICATION

EN50270:2006 Electromagneticcompatibility-Electricalapparatusforthedetectionandmeasurementofcombustiblegases,toxicgasesoroxygen

EN61010-1:2010 Safetyrequirementsforelectricalequipmentformeasurement,control,andlaboratoryuse.Part1:Generalrequirements

Note1:Foroptimumperformanceduringcontinuousmonitoring,a45minutewarm-upperiodisrecommended.

© Productdesignsandspecificationsaresubjecttochangewithoutnotice.Userisresponsiblefordeterminingsuitabilityofproduct.

G200TECHNICAL SPECIFICATIONS


Recommended