Technical Specification FLUXUS® G70x TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01 1 FLUXUS FLUXUS Measurement with transducers mounted by Variofix L G704, G705 G709 Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features Precise bi-directional and highly dynamic flow measure- ment with the non-intrusive clamp-on technology High precision at fast and slow flow rates, high tem- perature and zero point stability Automatic loading of calibration data and transducer detection for a fast and easy set-up (less than 5 min), providing precise and long-term stable results User-friendly design Transducers available for a wide range of inner pipe diameters ( ) and fluid temperatures ( ) approved transducers for hazardous areas available Measurement is unaffected by gas density, viscosity, composition, dust, humidity, temperature or pressure Applications Designed for industrial use in harsh environments, in gas processing and natural gas extraction, chemical industry and in the petroleum industry. Practical applications: Measurement on natural gas pipelines and in natural gas storage installations Measurement of synthesized gas and injection gas Measurement for the gas supply industry 6...1600 mm -40...+200 °C ATEX, IECEx, FM Class 1 Div. 2
Transcript
Page 1: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

Technical Specification

FLUXUS® G70x

TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01 1

FLUXUS

FLUXUS

Measurement with transducers mounted by Variofix L

G704, G705

G709

Gas Ultrasonic Flowmeter for Permanent Installation

Designed for wall mounting or installation in 19'' rack systems

Features

• Precise bi-directional and highly dynamic flow measure-ment with the non-intrusive clamp-on technology

• High precision at fast and slow flow rates, high tem-perature and zero point stability

• Automatic loading of calibration data and transducer detection for a fast and easy set-up (less than 5 min), providing precise and long-term stable results

• User-friendly design

• Transducers available for a wide range of inner pipe diameters ( ) and fluid temperatures ( )

• approved transducers for hazardous areas available

• Measurement is unaffected by gas density, viscosity, composition, dust, humidity, temperature or pressure

Applications

Designed for industrial use in harsh environments, in gas processing and natural gas extraction, chemical industry and in the petroleum industry. Practical applications:

• Measurement on natural gas pipelines and in natural gas storage installations

• Measurement of synthesized gas and injection gas

• Measurement for the gas supply industry

6...1600 mm-40...+200 °C

ATEX, IECEx, FM Class 1 Div. 2

Natalie Lundie
HP Stamp
Page 2: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

FLUXUS® G70x Technical Specification

Table of Contents

TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-012

Function ........................................................................................................................................................... 3Measurement Principle ..................................................................................................................................... 3Calculation of Volumetric Flow Rate ................................................................................................................. 3Number of Sound Paths.................................................................................................................................... 4Typical Measurement Setup ............................................................................................................................. 5Standard Volumetric Flow Rate ........................................................................................................................ 5

Flow Transmitter ............................................................................................................................................. 6Technical Data .................................................................................................................................................. 6Dimensions ....................................................................................................................................................... 92 " Pipe Mounting Kit (optional)....................................................................................................................... 11Terminal Assignment ...................................................................................................................................... 12

Transducers................................................................................................................................................... 14Transducer Selection ...................................................................................................................................... 14Transducer Order Code .................................................................................................................................. 17Technical Data ................................................................................................................................................ 18

Transducer Mounting Fixture ...................................................................................................................... 39

Coupling Materials for Transducers............................................................................................................ 41

Damping Mats (optional) .............................................................................................................................. 42

Connection Systems..................................................................................................................................... 44Transducer Cable............................................................................................................................................ 45

Junction Box ................................................................................................................................................. 46Technical Data ................................................................................................................................................ 46Dimensions ..................................................................................................................................................... 462 " Pipe Mounting Kit (optional)....................................................................................................................... 47Terminal Assignment ...................................................................................................................................... 47

Clamp-on Temperature Probe (optional) .................................................................................................... 49

Page 3: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01 3

Technical Specification FLUXUS® G70x

Function

Measurement Principle

In order to measure the flow of a medium in a pipe, ultrasonic signals are used, employing the transit time dif-ference principle. Ultrasonic signals are emitted by a transducer installed on the pipe and received by a sec-ond transducer. These signals are emitted alternately in the flow direction and against it.

As the medium in which the signals propagate is flowing, the transit time of the ultrasonic signals in the flowdirection is shorter than against the flow direction.

The transit time difference, ∆t, is measured and allows the flowmeter to determine the average flow velocityalong the propagation path of the ultrasonic signals. A flow profile correction is then performed in order to ob-tain the area averaged flow velocity, which is proportional to the volumetric flow rate.

Two integrated microprocessors control the entire measuring process. This allows the flowmeter to removedisturbance signals, and to check each received ultrasonic wave for its validity which reduces noise.

Calculation of Volumetric Flow Rate

= kRe . A . ka . ∆t/(2 . tfl)

where

Path of the ultrasonic signal Transit time difference ∆t

- volumetric flow ratekRe - fluid mechanics calibration factorA - cross-sectional pipe areaka - acoustical calibration factor∆t - transit time differencetfl - transit time in the medium

Page 4: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

4 TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01

FLUXUS® G70x Technical Specification

Number of Sound Paths

The number of sound paths is the number of transits of the ultrasonic signal through the medium in the pipe.Depending on the number of sound paths, the following methods of installation exist:

• reflection arrangementThe number of sound paths is even. Both of the transducers are mounted on the same side of the pipe. Correct positioning of the transducers is easier.

• diagonal arrangementThe number of sound paths is odd. Both of the transducers are mounted on opposite sides of the pipe. In the case of a high signal attenuation by the medium, pipe and coatings, diagonal arrangement with 1 sound path will be used.

The preferred method of installation depends on the application. While increasing the number of sound pathsincreases the accuracy of the measurement, signal attenuation increases as well. The optimum number ofsound paths for the parameters of the application will be determined automatically by the transmitter.

As the transducers can be mounted with the transducer mounting fixture in reflection arrangement or diagonalarrangement, the number of sound paths can be adjusted optimally for the application..

a - transducer distance

Reflection arrangement, number of sound paths: 2

Diagonal arrangement, number of sound paths: 3

Diagonal arrangement, number of sound paths: 1 Diagonal arrangement, number of sound paths: 1,negative transducer distance

a

a

a > 0 a < 0

Page 5: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01 5

Technical Specification FLUXUS® G70x

Typical Measurement Setup

Standard Volumetric Flow Rate

The standard volumetric flow rate can be selected as physical quantity to be measured. It will be calculated in-ternally by:

N = . p/pN . TN/T . 1/K

where

The operational pressure p and the operational temperature T of the medium will be entered directly as fixedvalues into the transmitter.

or:

If inputs are installed (optional), pressure and temperature can be measured by the customer and fed in thetransmitter.

The gas compressibility coefficient K of the gas is entered in the transmitter:

• as fixed value or

• as approximation according to e.g. AGA8 or GERG

Example of a measurement setup in reflection arrangement with connection of the inputsto an external process pressure and process temperature measurement for standard volumetric flow rate calculation

N - standard volumetric flow rate- operating volumetric flow rate

pN - standard pressure (absolute value)p - operating pressure (absolute value)TN - standard temperature in KT - operating temperature in KK - compressibility coefficient of the gas: ratio of the compressibility factors of the gas at operating

conditions and at standard conditions Z/ZN

P T

transducers damping mat

transmitter

RS232

outputs inputs

power supply

e.g. external pressure sensor

RS485(optional)

temperature probe

V· V·

Page 6: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

6 TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01

FLUXUS® G70x Technical Specification

Flow Transmitter

Technical Data

FLUXUS G704**-NNG704**-A2

G705**-NNG705**-A2

G709**-NN

design standard field device field devicewith stainless steel housing

19 " module

measurementmeasurement principle transit time difference correlation principleflow velocity 0.01...35 m/s, depending on pipe diameterrepeatability 0.15 % of reading ±0.01 m/smedium all acoustically conductive gases,

e.g. nitrogen, air, oxygen, hydrogen, argon, helium, ethylene, propanetemperature compensation corresponding to the recommendations in ANSI/ASME MFC-5.1-2011

accuracyvolumetric flow rate ± 1...3 % of reading ±0.01 m/s depending on application

± 0.5 % of reading ±0.01 m/s with field calibrationflow transmitterpower supply 100...230 V/50...60 Hz or

20...32 V DCpower consumption < 15 Wnumber of flow measuring channels

1, optional: 2

signal attenuation 0...100 s, adjustablemeasuring cycle (1 channel) 100...1000 Hzresponse time 1 s (1 channel), option: 70 mshousing material aluminum, powder coated stainless steel 316L (1.4404) aluminumdegree of protection accord-ing to IEC/EN 60529

IP65 IP66 IP20

dimensions see dimensional drawing 42HP x 3U(without back panel)

see dimensional drawingweight 2.8 kg 4.76 kg 1.7 kgfixation wall mounting, optional: 2 " pipe mounting 19 " rack mountingambient temperature -40...+60 °C

(-40...-20 °C without operation of the display)-20...+60 °C

display 2 x 16 characters, dot matrix, backlightmenu language English, German, French, Dutch, Spanishexplosion protection

ATEX/IECEx

transmitter G704**-A2 G705**-A2 -zone 2 2 -marking

Ex nA nC ic IIC T4 GcTa -40...+60 °CEx tb IIIC T 120 °C DbTa -40...+50 °C

II3G Ex nA nC ic IIC T4 GcII3D Ex tc IIIC T 100 °C DcTa -20...+60 °C

-

certification ATEX IBExU11ATEX1015 - -certification IECEx IECEx IBE 11.0008 - -type of protection gas: non sparking

dust: protection by enclosuregas: non sparkingdust: protection by enclosure

-

intrinsic safety parameters Um = 250 V (power supply 100...230 V AC)Um = 36 V (power supply 20...32 V DC)

Um = 250 V (power supply 100...230 V AC)Um = 36 V (power supply 20...32 V DC)

-

II3GII2D

0637

Page 7: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01 7

Technical Specification FLUXUS® G70x

measuring functionsphysical quantities operating volumetric flow rate, standard volumetric flow rate, mass flow rate, flow velocitytotalizer volume, masscalculation functions average, difference, sum (2 measuring channels necessary)diagnostic functions sound speed, signal amplitude, SNR, SCNR,

standard deviation of amplitudes and transit timesdata loggerloggable values all physical quantities, totalized values and diagnostic valuescapacity > 100 000 measured valuesSD card, removable (optional)loggable values all physical quantities and totalized values -capacity min. 2 GB -communicationinterface - process integration (optional): RS485 (emitter) or Modbus RTU or

HART or BACnet MS/TP or SD card

- diagnosis: RS2321

- process integration (optional): RS485 (emitter) or Modbus RTU or HART or BACnet MS/TP

- diagnosis: RS232serial data kit (optional)software (all Windows™ versions)

-FluxData: download of measurement data, graphical presentation,conversion to other formats (e.g. for Excel™)

-FluxDiag (optional): online diagnostics and report generation

-FluxKoef: creating medium data sets

-FluxSubstanceLoader: upload of medium data setsсable RS2321 RS232adapter RS232 - USB1 RS232 - USBoutputs (optional)

The outputs are galvanically isolated from the transmitter.number on request

switchable current output (nonEx)All switchable current outputs are switched to active or passive mode at the same time.

- range 4...20 mA (3.2...22 mA)- accuracy 0.04 % of reading ±3 μA- active output Rext < 350 Ω- passive output Uext = 8...30 V, depending on Rext, Rext < 1 kΩ

current outputcurrent output- range 0/4...20 mA- accuracy 0.1 % of reading ±15 μA- active output Rext < 500 Ω- passive output Uext = 4...24 V, depending on Rext, Rext < 1 kΩcurrent output I1 in HART mode- range 4...20 mA- passive output Uext = 10...24 V

voltage outputrange 0...1 V or 0...10 Vaccuracy 0...1 V: 0.1 % of reading ±1 mV

0...10 V: 0.1 % of reading ±10 mVinternal resistance Ri = 500 Ω

frequency outputrange 0...5 kHzopen collector 24 V/4 mA, Ri = 66.5 Ω

binary outputReed relay 48 V/100 mA, P1...P4: Ri = 22 Ω 48 V/100 mA, P1...P4: Ri = 22 Ωopen collector 24 V/4 mA, P1...P4: Ri = 22 Ω 24 V/4 mA, P1...P4: Ri = 22 Ωoptorelay 26 V/100 mA -binary output as alarm output- functions limit, change of flow direction or error limit, change of flow direction or

errorbinary output as pulse output- pulse value 0.01...1000 units 0.01...1000 units- pulse width optorelay: 1...1000 ms

Reed relay, open collector: 80...1000 ms80...1000 ms

1 G70x**-A2: connection of the interface RS232 outside of explosive atmosphere (housing cover open)

FLUXUS G704**-NNG704**-A2

G705**-NNG705**-A2

G709**-NN

Page 8: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

8 TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01

FLUXUS® G70x Technical Specification

inputs (optional)The inputs are galvanically isolated from the transmitter.

number max. 4, on requesttemperature input

type Pt100/Pt1000connection 4-wirerange -150...+560 °Cresolution 0.01 Kaccuracy ±0.01 % of reading ±0.03 K

current inputaccuracy 0.1 % of reading ±10 μA 0.1 % of reading ±10 μAactive input Ui = 24 V, Ri = 50 Ω, Pi < 0.5 W, not short-circuit proof Ui = 15 V, Ri = 50 Ω, Pi < 0.5 W,

not short-circuit proof- range 0...20 mA 0...20 mApassive input Ri = 50 Ω, Pi < 0.3 W Ri = 50 Ω, Pi < 0.3 W- range -20...+20 mA -20...+20 mA

voltage inputrange 0...1 Vaccuracy 0.1 % of reading ±1 mVinternal resistance Ri = 1 MΩ

binary inputswitching signal 5...30 V, 1 mAfunctions -resetting the measured values

-resetting the totalizers-stopping the totalizers

-activation of the measuring mode for highly dynamic flows

FLUXUS G704**-NNG704**-A2

G705**-NNG705**-A2

G709**-NN

Page 9: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01 9

Technical Specification FLUXUS® G70x

Dimensions

FLUXUS G704

FLUXUS G705

70*

10

Ø 4.5

163

200

287265 fixing holes for wall mounting

thread: 6x M20 x 1.5cable gland: max. 6x M20

* G704**-A2: +10 mm above the display

163

320

cable gland: max. 6x M20 with flat gasket and counter nut

370

255

87

98

Page 10: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

10 TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01

FLUXUS® G70x Technical Specification

FLUXUS G709

in mm

208

129

222

15

123

110

M2.

544

213

129

170

198

Page 11: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01 11

Technical Specification FLUXUS® G70x

2 " Pipe Mounting Kit (optional)

FLUXUS G704

FLUXUS G705

Page 12: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

12 TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01

FLUXUS® G70x Technical Specification

Terminal Assignment

2 The number, type and terminal assignment of the outputs and inputs will be customized.

FLUXUS G704, G705

power supplyterminal strip KL3

terminal connection (AC) connection (DC)PE earth earthN(-) neutral -L(+) phase +

transducersterminal strip KL1

extension cable (transducers *****8*, ****LI*, *****52)transducer cable (transducers *****8*, ****LI*)

extension cable connection system AS(transducers *****Z7, *****C3)

transducer cable (transducers *****52)measuring channel A measuring channel B measuring channel A measuring channel B

terminal connection terminal connection terminal connection

AV signal BV signal X_AV X_BV SMB connector

AVS shield BVS shield X_AR X_BR SMB connector

ARS shield BRS shield X1 X2 AMP-Quick connector1

AR signal BR signal 1 connection system AS

outputs2 RS485, Modbus, BACnet (optional)

terminal strip KL4 terminal strip KL4terminal connection terminal connection

P1+...P4+, P1-...P4- current output, voltage output, frequency output or binary output (Reed relay, open collector)

A+ signal +B- signal -

P5a...P7a, P5b...P7b binary output 101 shield

analog inputs2

terminal strip KL2temperature probe passive current source active current source

terminal direct connection connection with extension cable

connection of an active input

connection of a passive input

T1a...T4a red red not connected not connectedT1A...T4A red/blue gray - +T1b...T4b white/blue blue + not connectedT1B...T4B white white not connected -S1...S4 shield shield not connected not connected

binary inputs2

terminal strip KL4terminal

P1+...P2+, P1-...P2-

! " # ! " $ $ $

% % %

$ % $ % $ % $ %

#

#

#

& & ' & & '

#

( &

(

( &

(

equipotential bonding terminal(FLUXUS G705)

equipotential bonding terminal(FLUXUS G704**-A2)

Page 13: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01 13

Technical Specification FLUXUS® G70x

1 The number, type and terminal assignment of the outputs and inputs will be customized.

FLUXUS G709

power supplyterminal strip KL2

terminal connection (AC) terminal connection (DC)PE earth PE earthN neutral L- -L1 phase L+ +

transducersterminal strip KL6, KL8

extension cable (transducers *****8*, ****LI*, *****52)transducer cable (transducers ****LI*)

extension cable connection system AS(transducers *****Z7, *****C3)

measuring channel A measuring channel B measuring channel A measuring channel B

terminal connection terminal connection terminal connection

AV signal BV signal X_AV X_BV SMB connector

AVS shield BVS shield X_AR X_BR SMB connectorARS shield BRS shield X1 X2 AMP-Quick connectorAR signal BR signal

outputs1 RS485, Modbus, BACnet (optional)

terminal strip KL1 terminal strip KL4terminal connection terminal connection

P1+...P4+, P1-...P4- current output, voltage output, frequency output or binary output 4A+ signal +4B- signal -

P5+...P7+, P5-...P7- binary output (open collector) 43 shieldP5a...P7a, P5b...P7b binary output (Reed relay)

analog inputs1

terminal strip KL3temperature probe passive current source active current source

terminal direct connection connection with extension cable

connection of an active input

connection of a passive input

T1a...T4a red red not connected not connectedT1A...T4A red/blue gray - +T1b...T4b white/blue blue + not connectedT1B...T4B white white not connected -S1...S4 shield shield not connected not connected

binary inputs1

terminal strip KL1terminal

P1+...P2+, P1-...P2-

# # # #

# #

& &

# #

) #

&

&

'

&

&

'

#

#

#

%

%

%

$

$

$

%

$

$

%

$

$

%

%

Page 14: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

14 TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01

FLUXUS® G70x Technical Specification

Transducers

Transducer Selection

Step 1a

Select a Lamb wave transducer:

Step 1bIf the pipe wall thickness is not in the range of the Lamb wave transducers, select a shear wave transducer:

Step 2

inner pipe diameter d dependent on the flow velocity v of the medium in the pipe

The transducers are selected from the characteristics (see next page). Lamb wave transducers are selectedfrom the left column, shear wave transducers from the right column.

Lamb wave transducers: If the values d and v are not in the range, the diagonal arrangement with 1 soundpath may be used, i.e. the same characteristics can be used with doubling the inner pipe diameter. If the val-ues are still not in the range, shear waves transducers regarding the pipe wall thickness have to be selectedin step 1b.

transducer order code

GLG 11 23

GLH 7 15

GLK 4 9

GLM 2 5

GLP 1...3

GLQ 0.5...1

5 10 15 20 25pipe wall thickness [mm]

transducer order code

GSG 14 23

GSK 5 9

GSM 2.5 5

GSP 1.5 3

GSQ 1 1.5

5 10 15 20 25pipe wall thickness [mm]

recommended possible

Page 15: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01 15

Technical Specification FLUXUS® G70x

Lamb wave transducer1 shear wave transducer1

GLG GSG

GLH

GLK GSK

GLM GSM

GLP GSP

GLQ GSQ

1 inner pipe diameter and max. flow velocity for a typical application with natural gas, nitrogen, oxygen in reflection arrangement with2 sound paths (Lamb wave transducers)/1 sound path (shear wave transducers)

0 5 10 15 20 25 30 35vms

200

400

600

800

1000

1200Dmm

0 5 10 15 20 25 30 35vms

200

400

600

800

1000

1200Dmm

0 5 10 15 20 25 30 35vms

200

400

600

800

1000

1200Dmm

0 5 10 15 20 25 30 35vms

200

400

600

800

1000

1200Dmm

0 5 10 15 20 25 30 35vms

200

400

600

800

1000

1200Dmm

0 5 10 15 20 25 30 35vms

20

40

60

80

100

120

140

Dmm

0 5 10 15 20 25 30 35vms

20

40

60

80

100

120

140

Dmm

0 5 10 15 20 25 30 35vms

20

40

60

80

100

120

140

Dmm

0 5 10 15 20 25 30 35vms

20

40

60

80

100

120

140

Dmm

0 5 10 15 20 25 30 35vms

20

40

60

80

100

120

140

Dmm

0 5 10 15 20 25 30 35vms

20

40

60

80

100

120

140

Dmm

Page 16: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

16 TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01

FLUXUS® G70x Technical Specification

Step 3

min. medium pressure

Example

Step 4

for the characters 4...11 of the transducer order code (ambient temperature, explosion protection, connectionsystem, extension cable) see page 17

Step 5

for the technical data of the selected transducer see page 18 et seqq.

Lamb wave transducer shear wave transducertransducer order code

medium pressure1 [bar] transducer order code

medium pressure1 [bar]metal pipe plastic pipe metal pipe plastic pipe

min. min. extended min. min. min. extended min.GLG 15 10 1 GSG 30 20 1

GLH 15 10 1 GSK 30 20 1

GLK 15 (d > 120 mm)10 (d < 120 mm)

10 (d > 120 mm)5 (d < 120 mm)

1 GSM 30 20 1

GLM 10 (d > 60 mm)5 (d < 60 mm)

- 1 GSP 30 20 1

GLP 10 (d > 35 mm)5 (d < 35 mm)

- 1 GSQ 30 20 1

GLQ 10 (d > 15 mm)5 (d < 15 mm)

- 1

1 depending on application, typical absolute value for natural gas, nitrogen, compressed air

d - inner pipe diameter

step1 pipe wall thickness mm 12 12 12 30

selected transducer GLG or GLH GLG or GLH GLG or GLH GS2 inner pipe diameter mm 800 600 800 300

max. flow velocity m/s 15 15 30 15selected transducer GLG GLG or GLH values not in the range

of the characteristics, but by using diagonal arrangement with 1 sound path, the inner pipe diameter in the characteristics is dou-bled: GLG

GSK

3 min. medium pressure bar 17 17 17 35selected transducer GLG GLG or GLH

influence of acoustic noise is reduced with increased transducer frequency, thus recommended:GLH

GLG GSK

Page 17: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01 17

Technical Specification FLUXUS® G70x

Transducer Order Code

1, 2 3 4 5, 6 7, 8 9...11 12, 13 no. of character

tra

nsdu

cer

tra

nsdu

cer

fre

quen

cy

- ambi

ent

tem

pera

ture

expl

osio

n pr

ote

ctio

n

conn

ectio

n sy

stem

- exte

nsio

n ca

ble

/ optio

n

description

GL set of ultrasonic flow transducers for gas measurement, Lamb wave

GS set of ultrasonic flow transducers for gas measurement, shear wave

G 0.2 MHz

H 0.3 MHz (Lamb wave only)

K 0.5 MHz

M 1 MHz

P 2 MHz

Q 4 MHz

N normal temperature range

E extended temperature range (shear wave transducers with trans-ducer frequency M, P, Q)

A1 ATEX zone 1/IECEx zone 1

A2 ATEX zone 2

F2 FM Class I Div. 2

NN not explosion proof

AS with Amphenol connector (not explosion proof transducers)

TS direct connection or connection via junction box

XXX cable length in m, for max. length of extension cable see page 44

connection system TS:

0 m: without junction box

> 0 m: with junction box JB01 (zone 1), JB02 (ATEX zone 2, FM), JB03 (not explosion proof), JBP2 (transducers IP68, ATEX zone 2), JBP3 (transducers IP68, not explosion proof)

IP68 degree of protection IP68 with connection system TS

OS housing with stainless steel 316 (with connection system TS)

example

GL K - N A1 TS - 030 Lamb wave transducer 0.5 MHz, normal temperature range, ATEX zone 1/IECEx zone 1, connection system TS with junction box JB01 and extension cable 30 m

- - /

Page 18: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

18 TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01

FLUXUS® G70x Technical Specification

Technical Data

Shear Wave Transducers (zone 1)technical type GDG1N81 GDK1N81order code GSG-NA1TS

GSG-NA1TS/OSGSK-NA1TSGSK-NA1TS/OS

transducer frequency MHz 0.2 0.5

medium pressure1

min. extended bar metal pipe: 20 metal pipe: 20min. bar metal pipe: 30

plastic pipe: 1metal pipe: 30plastic pipe: 1

inner pipe diameter d2

min. extended mm 250 70min. recommended mm 380 80max. recommended mm 810 500max. extended mm 1100 720pipe wall thicknessmin. mm 14 5max. mm - -materialhousing PEEK with stainless steel

cap 304 (1.4301), option OS: 316L (1.4404)

PEEK with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)

contact surface PEEK PEEKdegree of protection according to IEC/EN 60529

IP65 IP65

transducer cabletype 1699 1699length m 5 5dimensionslength l mm 129.5 126.5width b mm 51 51height h mm 67 67.5dimensional drawing

ambient temperaturemin. °C -40 -40max. °C +130 +130temperature compensation

x x

explosion protection

ATEX/IECEx

categoryzone

gas: 2G dust: 2D1 21

gas: 2G dust: 2D1 21

explosion protection temperature (pipe surface)min. °C -55 -55max. °C +180 +180marking

Ex e q IIC T6...T3 GbEx tb IIIC TX Db

Ex e q IIC T6...T3 GbEx tb IIIC TX Db

certification ATEX IBExU07ATEX1168 X IBExU07ATEX1168 Xcertification IECEx IECEx IBE 08.0007X IECEx IBE 08.0007Xtype of protection gas: increased safety,

powder fillingdust: protection by enclosure

gas: increased safety, powder fillingdust: protection by enclosure

transducer mounting fixture necessary

x x

1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 shear wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended/max. extended: in diagonal arrangement and for a flow velocity of 15 m/s

l

hb

l

hb

II2GII2D

0637 II2GII2D

0637

Page 19: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01 19

Technical Specification FLUXUS® G70x

technical type GDM2N81 GDP2N81 GDQ2N81order code GSM-NA1TS

GSM-NA1TS/OSGSP-NA1TSGSP-NA1TS/OS

GSQ-NA1TSGSQ-NA1TS/OS

transducer frequency MHz 1 2 4

medium pressure1

min. extended bar metal pipe: 20 metal pipe: 20 metal pipe: 20min. bar metal pipe: 30

plastic pipe: 1metal pipe: 30plastic pipe: 1

metal pipe: 30plastic pipe: 1

inner pipe diameter d2

min. extended mm 30 15 6min. recommended mm 40 20 10max. recommended mm 80 40 20max. extended mm 120 60 30pipe wall thicknessmin. mm 2.5 1.5 1max. mm - - -materialhousing PEEK with stainless steel

cap 304 (1.4301), option OS: 316L (1.4404)

PEEK with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)

PEEK with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)

contact surface PEEK PEEK PEEKdegree of protection according to IEC/EN 60529

IP65 IP65 IP65

transducer cabletype 1699 1699 1699length m 4 4 3dimensionslength l mm 64 64 40width b mm 32 32 22height h mm 40.5 40.5 25.5dimensional drawing

ambient temperaturemin. °C -40 -40 -40max. °C +130 +130 +130temperature compensation

x x x

explosion protection

ATEX/IECEx

categoryzone

gas: 2G dust: 2D1 21

gas: 2G dust: 2D1 21

gas: 2G dust: 2D1 21

explosion protection temperature (pipe surface)min. °C -55 -55 -55max. °C +180 +180 +180marking

Ex e q IIC T6...T3 GbEx tb IIIC TX Db

Ex e q IIC T6...T3 GbEx tb IIIC TX Db

Ex e q IIC T6...T3 GbEx tb IIIC TX Db

certification ATEX IBExU07ATEX1168 X IBExU07ATEX1168 X IBExU07ATEX1168 Xcertification IECEx IECEx IBE 08.0007X IECEx IBE 08.0007X IECEx IBE 08.0007Xtype of protection gas: increased safety,

powder fillingdust: protection by enclosure

gas: increased safety, powder fillingdust: protection by enclosure

gas: increased safety, powder fillingdust: protection by enclosure

transducer mounting fixture necessary

x x x

remark on request1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 shear wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended/max. extended: in diagonal arrangement and for a flow velocity of 15 m/s

l

hb

l

hb

hb

l

II2GII2D

0637 II2GII2D

0637 II2GII2D

0637

Page 20: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

20 TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01

FLUXUS® G70x Technical Specification

Shear Wave Transducers (zone 1, IP68)

technical type GDG1LI1 GDK1LI1 GDM2LI1 GDP2LI1order code GSG-NA1TS/IP68 GSK-NA1TS/IP68 GSM-NA1TS/IP68 GSP-NA1TS/IP68transducer frequency MHz 0.2 0.5 1 2

medium pressure1

min. extended bar metal pipe: 20 metal pipe: 20 metal pipe: 20 metal pipe: 20min. bar metal pipe: 30

plastic pipe: 1metal pipe: 30plastic pipe: 1

metal pipe: 30plastic pipe: 1

metal pipe: 30plastic pipe: 1

inner pipe diameter d2

min. extended mm 250 70 30 15min. recommended mm 380 80 40 20max. recommended mm 810 500 80 40max. extended mm 1100 720 120 60pipe wall thicknessmin. mm 14 5 2.5 1.5max. mm - - - -materialhousing PEEK with stainless steel

cap 316Ti (1.4571)PEEK with stainless steel cap 316Ti (1.4571)

PEEK with stainless steel cap 316Ti (1.4571)

PEEK with stainless steel cap 316Ti (1.4571)

contact surface PEEK PEEK PEEK PEEKdegree of protection according to IEC/EN 60529

IP68 IP68 IP68 IP68

transducer cabletype 2550 2550 2550 2550length m 12 12 12 12dimensionslength l mm 130 130 72 72width b mm 54 54 32 32height h mm 83.5 83.5 46 46dimensional drawing

ambient temperaturemin. °C -40 -40 -40 -40max. °C +100 +100 +100 +100temperature compensation

x x x x

explosion protection

ATEX/IECEx

categoryzone

gas: 2G dust: 2D1 21

gas: 2G dust: 2D1 21

gas: 2G dust: 2D1 21

gas: 2G dust: 2D1 21

explosion protection temperature (pipe surface)min. °C -55 -55 -55 -55max. °C +180 +180 +180 +180marking

Ex q IIC T6...T3 GbEx tb IIIC TX Db

Ex q IIC T6...T3 GbEx tb IIIC TX Db

Ex q IIC T6...T3 GbEx tb IIIC TX Db

Ex q IIC T6...T3 GbEx tb IIIC TX Db

certification ATEX IBExU07ATEX1168 X IBExU07ATEX1168 X IBExU07ATEX1168 X IBExU07ATEX1168 Xcertification IECEx IECEx IBE 08.0007X IECEx IBE 08.0007X IECEx IBE 08.0007X IECEx IBE 08.0007Xtype of protection gas: powder filling

dust: protection by enclosure

gas: powder fillingdust: protection by enclosure

gas: powder fillingdust: protection by enclosure

gas: powder fillingdust: protection by enclosure

transducer mounting fixture necessary

x x x x

1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 shear wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended/max. extended: in diagonal arrangement and for a flow velocity of 15 m/s

3 test conditions: 3 months/2 bar (20 m)/20 °C

3 3 3 3

l

hb

l

hb

l

hb

l

hb

II2GII2D

0637 II2GII2D

0637 II2GII2D

0637 II2GII2D

0637

Page 21: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01 21

Technical Specification FLUXUS® G70x

Shear Wave Transducers (zone 1, extended temperature range)

technical type GDM2E85 GDP2E85 GDQ2E85order code GSM-EA1TS

GSM-EA1TS/OSGSP-EA1TSGSP-EA1TS/OS

GSQ-EA1TSGSQ-EA1TS/OS

transducer frequency MHz 1 2 4

medium pressure1

min. extended bar metal pipe: 20 metal pipe: 20 metal pipe: 20min. bar metal pipe: 30

plastic pipe: 1metal pipe: 30plastic pipe: 1

metal pipe: 30plastic pipe: 1

inner pipe diameter d2

min. extended mm 30 15 6min. recommended mm 40 20 10max. recommended mm 80 40 20max. extended mm 120 60 30pipe wall thicknessmin. mm 2.5 1.5 1max. mm - - -materialhousing PI with stainless steel

cap 304 (1.4301), option OS: 316L (1.4404)

PI with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)

PI with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)

contact surface PI PI PIdegree of protection according to IEC/EN 60529

IP56 IP56 IP56

transducer cabletype 6111 6111 6111length m 4 4 3dimensionslength l mm 64 64 40width b mm 32 32 22height h mm 40.5 40.5 25.5dimensional drawing

ambient temperaturemin. °C -30 -30 -30max. °C +200 +200 +200temperature compensation

x x x

explosion protection

ATEX/IECEx

categoryzone

gas: 2G dust: 3D1 22

gas: 2G dust: 3D1 22

gas: 2G dust: 3D1 22

explosion protection temperature (pipe surface)min. °C -45 -45 -45max. °C +225 +225 +225marking

Ex e q IIC T6...T2 GbEx tb IIIA TX Db

Ex e q IIC T6...T2 GbEx tb IIIA TX Db

Ex e q IIC T6...T2 GbEx tb IIIA TX Db

certification ATEX IBExU07ATEX1168 X IBExU07ATEX1168 X IBExU07ATEX1168 Xcertification IECEx IECEx IBE 08.0007X IECEx IBE 08.0007X IECEx IBE 08.0007Xtype of protection gas: increased safety,

powder fillingdust: protection by enclosure

gas: increased safety, powder fillingdust: protection by enclosure

gas: increased safety, powder fillingdust: protection by enclosure

transducer mounting fixture necessary

x x x

remark on request1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 shear wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended/max. extended: in diagonal arrangement and for a flow velocity of 15 m/s

l

hb

l

hb

hb

l

II2GII2D

0637 II2GII2D

0637 II2GII2D

0637

Page 22: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

22 TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01

FLUXUS® G70x Technical Specification

Shear Wave Transducers (ATEX zone 2, FM or not explosion proof)

technical type GDG1N52 GDK1N52order code GSG-NA2TS, GSG-NA2TS/OS

GSG-NF2TS, GSG-NF2TS/OSGSG-NNNTS, GSG-NNNTS/OS

GSK-NA2TS, GSK-NA2TS/OSGSK-NF2TS, GSK-NF2TS/OSGSK-NNNTS, GSK-NNNTS/OS

transducer frequency MHz 0.2 0.5medium pressure1

min. extended bar metal pipe: 20 metal pipe: 20min. bar metal pipe: 30

plastic pipe: 1metal pipe: 30plastic pipe: 1

inner pipe diameter d2

min. extended mm 250 70min. recommended mm 380 80max. recommended mm 810 500max. extended mm 1100 720pipe wall thicknessmin. mm 14 5max. mm - -materialhousing PEEK with stainless steel cap 304 (1.4301),

option OS: 316L (1.4404)PEEK with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)

contact surface PEEK PEEKdegree of protection according to IEC/EN 60529

IP67 IP67

transducer cabletype 1699 1699length m 5 5dimensionslength l mm 129.5 126.5width b mm 51 51height h mm 67 67.5dimensional drawing

ambient temperaturemin. °C -40 -40max. °C +130 +130temperature compensation

x x

explosion protection

ATEX

order code GSG-NA2TS, GSG-NA2TS/OS GSK-NA2TS, GSK-NA2TS/OScategoryzone

gas: 3G dust: 3D2 22

gas: 3G dust: 3D2 22

explosion protection temperature (pipe surface)min. °C -55 -55max. °C +190 +190marking

II3G Ex nA IIC T6...T3 GcTa -55...+190 °CII3D Ex tc IIIC TX Dc

II3G Ex nA IIC T6...T3 GcTa -55...+190 °CII3D Ex tc IIIC TX Dc

certification - -type of protection gas: non sparking

dust: protection by enclosuregas: non sparkingdust: protection by enclosure

transducer mounting fixture necessary

x x

FM

order code GSG-NF2TS, GSG-NF2TS/OS GSK-NF2TS, GSK-NF2TS/OSexplosion protection temperaturemin. °C -40 -40max. °C +125 +125marking NI/Cl. I,II,III/Div. 2 /

GP A,B,C,D,E,F,G/Temp. Codes dwg 3860

NI/Cl. I,II,III/Div. 2 /GP A,B,C,D,E,F,G/

Temp. Codes dwg 38601 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 shear wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended/max. extended: in diagonal arrangement and for a flow velocity of 15 m/s

l

hb

l

hb

Page 23: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01 23

Technical Specification FLUXUS® G70x

Shear Wave Transducers (ATEX zone 2, FM or not explosion proof)

technical type GDM2N52 GDP2N52 GDQ2N52order code GSM-NA2TS, GSM-NA2TS/OS

GSM-NF2TS, GSM-NF2TS/OSGSM-NNNTS, GSM-NNNTS/OS

GSP-NA2TS, GSP-NA2TS/OSGSP-NF2TS, GSP-NF2TS/OSGSP-NNNTS, GSP-NNNTS/OS

GSQ-NA2TS, GSQ-NA2TS/OSGSQ-NF2TS, GSQ-NF2TS/OSGSQ-NNNTS, GSQ-NNNTS/OS

transducer frequency MHz 1 2 4medium pressure1

min. extended bar metal pipe: 20 metal pipe: 20 metal pipe: 20min. bar metal pipe: 30

plastic pipe: 1metal pipe: 30plastic pipe: 1

metal pipe: 30plastic pipe: 1

inner pipe diameter d2

min. extended mm 30 15 6min. recommended mm 40 20 10max. recommended mm 80 40 20max. extended mm 120 60 30pipe wall thicknessmin. mm 2.5 1.5 1max. mm - - -materialhousing PEEK with stainless steel cap 304

(1.4301), option OS: 316L (1.4404)

PEEK with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)

PEEK with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)

contact surface PEEK PEEK PEEKdegree of protection according to IEC/EN 60529

IP67 IP67 IP67

transducer cabletype 1699 1699 1699length m 4 4 3dimensionslength l mm 64 64 40width b mm 32 32 22height h mm 40.5 40.5 25.5dimensional drawing

ambient temperaturemin. °C -40 -40 -40max. °C +130 +130 +130temperature compensation

x x x

explosion protection

ATEX

order code GSM-NA2TS, GSM-NA2TS/OS GSP-NA2TS, GSP-NA2TS/OS GSQ-NA2TS, GSQ-NA2TS/OScategoryzone

gas: 3G dust: 3D2 22

gas: 3G dust: 3D2 22

gas: 3G dust: 3D2 22

explosion protection temperature (pipe surface)min. °C -55 -55 -55max. °C +190 +190 +190marking

II3G Ex nA IIC T6...T3 GcTa -55...+190 °CII3D Ex tc IIIC TX Dc

II3G Ex nA IIC T6...T3 GcTa -55...+190 °CII3D Ex tc IIIC TX Dc

II3G Ex nA IIC T6...T3 GcTa -55...+190 °CII3D Ex tc IIIC TX Dc

certification - - -type of protection gas: non sparking

dust: protection by enclosuregas: non sparkingdust: protection by enclosure

gas: non sparkingdust: protection by enclosure

transducer mounting fixture necessary

x x x

FM

order code GSM-NF2TS, GSM-NF2TS/OS GSP-NF2TS, GSP-NF2TS/OS GSQ-NF2TS, GSQ-NF2TS/OSexplosion protection temperaturemin. °C -55 -55 -55max. °C +190 +190 +190marking NI/Cl. I,II,III/Div. 2 /

GP A,B,C,D,E,F,G/Temp. Codes dwg 3860

NI/Cl. I,II,III/Div. 2 /GP A,B,C,D,E,F,G/

Temp. Codes dwg 3860

NI/Cl. I,II,III/Div. 2 /GP A,B,C,D,E,F,G/

Temp. Codes dwg 3860type of protection non incendive non incendive non incendive

remark on request1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 shear wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended/max. extended: in diagonal arrangement and for a flow velocity of 15 m/s

l

hb

l

hb

hb

l

Page 24: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

24 TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01

FLUXUS® G70x Technical Specification

Shear Wave Transducers (ATEX zone 2 or not explosion proof, IP68)

technical type GDG1LI8 GDK1LI8 GDM2LI8 GDP2LI8order code GSG-NA2TS/IP68

GSG-NNNTS/IP68GSK-NA2TS/IP68GSK-NNNTS/IP68

GSM-NA2TS/IP68GSM-NNNTS/IP68

GSP-NA2TS/IP68GSP-NNNTS/IP68

transducer frequency MHz 0.2 0.5 1 2

medium pressure1

min. extended bar metal pipe: 20 metal pipe: 20 metal pipe: 20 metal pipe: 20min. bar metal pipe: 30

plastic pipe: 1metal pipe: 30plastic pipe: 1

metal pipe: 30plastic pipe: 1

metal pipe: 30plastic pipe: 1

inner pipe diameter d2

min. extended mm 250 70 30 15min. recommended mm 380 80 40 20max. recommended mm 810 500 80 40max. extended mm 1100 720 120 60pipe wall thicknessmin. mm 14 5 2.5 1.5max. mm - - - -materialhousing PEEK with stainless steel

cap 316Ti (1.4571)PEEK with stainless steel cap 316Ti (1.4571)

PEEK with stainless steel cap 316Ti (1.4571)

PEEK with stainless steel cap 316Ti (1.4571)

contact surface PEEK PEEK PEEK PEEKdegree of protection according to IEC/EN 60529

IP68 IP68 IP68 IP68

transducer cabletype 2550 2550 2550 2550length m 12 12 12 12dimensionslength l mm 130 130 72 72width b mm 54 54 32 32height h mm 83.5 83.5 46 46dimensional drawing

ambient temperaturemin. °C -40 -40 -40 -40max. °C +100 +100 +100 +100temperature compensation

x x x x

explosion protection

ATEX

order code GSG-NA2TS/IP68 GSK-NA2TS/IP68 GSM-NA2TS/IP68 GSP-NA2TS/IP68categoryzone

gas: 3G dust: 3D2 22

gas: 3G dust: 3D2 22

gas: 3G dust: 3D2 22

gas: 3G dust: 3D2 22

explosion protection temperature (pipe surface)min. °C -40 -40 -40 -40max. °C +90 +90 +90 +90marking

II3G Ex nA IIC T6...T5 GcTa -40...+90 °CII3D Ex tc IIIC TX Dc

II3G Ex nA IIC T6...T5 GcTa -40...+90 °CII3D Ex tc IIIC TX Dc

II3G Ex nA IIC T6...T5 GcTa -40...+90 °CII3D Ex tc IIIC TX Dc

II3G Ex nA IIC T6...T5 GcTa -40...+90 °CII3D Ex tc IIIC TX Dc

certification - - - -type of protection gas: non sparking

dust: protection by enclosure

gas: non sparkingdust: protection by enclosure

gas: non sparkingdust: protection by enclosure

gas: non sparkingdust: protection by enclosure

transducer mounting fixture necessary

x x x x

1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 shear wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended/max. extended: in diagonal arrangement and for a flow velocity of 15 m/s

3 test conditions: 3 months/2 bar (20 m)/20 °C

3 3 3 3

l

hb

l

hb

l

hb

l

hb

Page 25: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01 25

Technical Specification FLUXUS® G70x

Shear Wave Transducers (connection system AS, not explosion proof)

technical type GDG1NZ7 GDK1NZ7order code GSG-NNNAS GSK-NNNAStransducer frequency MHz 0.2 0.5

medium pressure1

min. extended bar metal pipe: 20 metal pipe: 20min. bar metal pipe: 30

plastic pipe: 1metal pipe: 30plastic pipe: 1

inner pipe diameter d2

min. extended mm 250 70min. recommended mm 380 80max. recommended mm 810 500max. extended mm 1100 720pipe wall thicknessmin. mm 14 5max. mm - -materialhousing PEEK with stainless steel

cap 304 (1.4301)PEEK with stainless steel cap 304 (1.4301)

contact surface PEEK PEEKdegree of protection according to IEC/EN 60529

IP67 IP67

transducer cabletype 1699 1699length m 5 5dimensionslength l mm 129.5 126.5width b mm 51 51height h mm 67 67.5dimensional drawing

ambient temperaturemin. °C -40 -40max. °C +130 +130temperature compensation

x x

1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 shear wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended/max. extended: in diagonal arrangement and for a flow velocity of 15 m/s

l

hb

l

hb

Page 26: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

26 TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01

FLUXUS® G70x Technical Specification

Shear Wave Transducers (connection system AS, not explosion proof)

technical type GDM2NZ7 GDP2NZ7 GDQ2NZ7order code GSM-NNNAS GSP-NNNAS GSQ-NNNAStransducer frequency MHz 1 2 4

medium pressure1

min. extended bar metal pipe: 20 metal pipe: 20 metal pipe: 20min. bar metal pipe: 30

plastic pipe: 1metal pipe: 30plastic pipe: 1

metal pipe: 30plastic pipe: 1

inner pipe diameter d2

min. extended mm 30 15 6min. recommended mm 40 20 10max. recommended mm 80 40 20max. extended mm 120 60 30pipe wall thicknessmin. mm 2.5 1.5 1max. mm - - -materialhousing PEEK with stainless steel

cap 304 (1.4301)PEEK with stainless steel cap 304 (1.4301)

PEEK with stainless steel cap 304 (1.4301)

contact surface PEEK PEEK PEEKdegree of protection according to IEC/EN 60529

IP67 IP67 IP67

transducer cabletype 1699 1699 1699length m 4 4 3dimensionslength l mm 64 64 40width b mm 32 32 22height h mm 40.5 40.5 25.5dimensional drawing

ambient temperaturemin. °C -40 -40 -40max. °C +130 +130 +130temperature compensation

x x x

remark on request1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 shear wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended/max. extended: in diagonal arrangement and for a flow velocity of 15 m/s

l

hb

l

hb

hb

l

Page 27: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01 27

Technical Specification FLUXUS® G70x

Shear Wave Transducers (extended temperature range, ATEX zone 2, FM or not explosion proof)

technical type GDM2E52 GDP2E52 GDQ2E52order code GSM-EA2TS, GSM-EA2TS/OS

GSM-EF2TS, GSM-EF2TS/OSGSM-ENNTS, GSM-ENNTS/OS

GSP-EA2TS, GSP-EA2TS/OSGSP-EF2TS, GSP-EF2TS/OSGSP-ENNTS, GSP-ENNTS/OS

GSQ-EA2TS, GSQ-EA2TS/OSGSQ-EF2TS, GSQ-EF2TS/OSGSQ-ENNTS, GSQ-ENNTS/OS

transducer frequency MHz 1 2 4medium pressure1

min. extended bar metal pipe: 20 metal pipe: 20 metal pipe: 20min. bar metal pipe: 30

plastic pipe: 1metal pipe: 30plastic pipe: 1

metal pipe: 30plastic pipe: 1

inner pipe diameter d2

min. extended mm 30 15 6min. recommended mm 40 20 10max. recommended mm 80 40 20max. extended mm 120 60 30pipe wall thicknessmin. mm 2.5 1.5 1max. mm - - -materialhousing PI with stainless steel cap 304

(1.4301), option OS: 316L (1.4404)

PI with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)

PI with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)

contact surface PI PI PIdegree of protection according to IEC/EN 60529

IP56 IP56 IP56

transducer cabletype 6111 6111 6111length m 4 4 3dimensionslength l mm 64 64 40width b mm 32 32 22height h mm 40.5 40.5 25.5dimensional drawing

ambient temperaturemin. °C -30 -30 -30max. °C +200 +200 +200temperature compensation

x x x

explosion protection

ATEX

order code GSM-EA2TS, GSM-EA2TS/OS GSP-EA2TS, GSP-EA2TS/OS GSQ-EA2TS, GSQ-EA2TS/OScategoryzone

gas: 3G dust: 3D2 22

gas: 3G dust: 3D2 22

gas: 3G dust: 3D2 22

explosion protection temperature (pipe surface)min. °C -45 -45 -45max. °C +235 +235 +235marking

II3G Ex nA IIC T6...T2 GcTa -45...+235 °CII3D Ex tc IIIB TX Dc

II3G Ex nA IIC T6...T2 GcTa -45...+235 °CII3D Ex tc IIIB TX Dc

II3G Ex nA IIC T6...T2 GcTa -45...+235 °CII3D Ex tc IIIB TX Dc

certification - - -type of protection gas: non sparking

dust: protection by enclosuregas: non sparkingdust: protection by enclosure

gas: non sparkingdust: protection by enclosure

transducer mounting fixture necessary

x x x

FM

order code GSM-EF2TS, GSM-EF2TS/OS GSP-EF2TS, GSP-EF2TS/OS GSQ-EF2TS, GSQ-EF2TS/OSexplosion protection temperaturemin. °C -45 -45 -45max. °C +235 +235 +235marking NI/Cl. I,II,III/Div. 2 /

GP A,B,C,D,E,F,G/Temp. Codes dwg 3860

NI/Cl. I,II,III/Div. 2 /GP A,B,C,D,E,F,G/

Temp. Codes dwg 3860

NI/Cl. I,II,III/Div. 2 /GP A,B,C,D,E,F,G/

Temp. Codes dwg 3860type of protection non incendive non incendive non incendive

remark on request1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 shear wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended/max. extended: in diagonal arrangement and for a flow velocity of 15 m/s

l

hb

l

hb

hb

l

Page 28: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

28 TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01

FLUXUS® G70x Technical Specification

Shear Wave Transducers (extended temperature range, not explosion proof,connection system AS)

technical type GDM2EZ7 GDP2EZ7 GDQ2EZ7order code GSM-ENNAS GSP-ENNAS GSQ-ENNAStransducer frequency MHz 1 2 4

medium pressure1

min. extended bar metal pipe: 20 metal pipe: 20 metal pipe: 20min. bar metal pipe: 30

plastic pipe: 1metal pipe: 30plastic pipe: 1

metal pipe: 30plastic pipe: 1

inner pipe diameter d2

min. extended mm 30 15 6min. recommended mm 40 20 10max. recommended mm 80 40 20max. extended mm 120 60 30pipe wall thicknessmin. mm 2.5 1.5 1max. mm - - -materialhousing PI with stainless steel

cap 304 (1.4301)PI with stainless steel cap 304 (1.4301)

PI with stainless steel cap 304 (1.4301)

contact surface PI PI PIdegree of protection according to IEC/EN 60529

IP65 IP65 IP65

transducer cabletype 6111 6111 6111length m 4 4 3dimensionslength l mm 64 64 40width b mm 32 32 22height h mm 40.5 40.5 25.5dimensional drawing

ambient temperaturemin. °C -30 -30 -30max. °C +200 +200 +200temperature compensation

x x x

remark on request1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 shear wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended/max. extended: in diagonal arrangement and for a flow velocity of 15 m/s

l

hb

l

hb

hb

l

Page 29: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01 29

Technical Specification FLUXUS® G70x

Lamb Wave Transducers (zone 1)

technical type GRG1N83 GRH1N83 GRK1N83order code GLG-NA1TS

GLG-NA1TS/OSGLH-NA1TSGLH-NA1TS/OS

GLK-NA1TSGLK-NA1TS/OS

transducer frequency MHz 0.2 0.3 0.5

medium pressure1

min. extended bar metal pipe: 10 metal pipe: 10 metal pipe:10 (d > 120 mm)5 (d < 120 mm)

min. bar metal pipe: 15plastic pipe: 1

metal pipe: 15plastic pipe: 1

metal pipe:15 (d > 120 mm)10 (d < 120 mm)plastic pipe: 1

inner pipe diameter d2

min. extended mm 190 120 60min. recommended mm 220 140 80max. recommended mm 900 600 300max. extended mm 1600 1000 500pipe wall thicknessmin. mm 11 7 4max. mm 23 15 9materialhousing PPSU with stainless steel cap 304

(1.4301), option OS: 316L (1.4404)

PPSU with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)

PPSU with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)

contact surface PPSU PPSU PPSUdegree of protection according to IEC/EN 60529

IP65 IP65 IP65

transducer cabletype 1699 1699 1699length m 5 5 5dimensionslength l mm 128.5 128.5 128.5width b mm 51 51 51height h mm 67.5 67.5 67.5dimensional drawing

ambient temperaturemin. °C -40 -40 -40max. °C +170 +170 +170temperature compensation

x x x

explosion protection

ATEX/IECEx

categoryzone

gas: 2G dust: 2D1 21

gas: 2G dust: 2D1 21

gas: 2G dust: 2D1 21

explosion protection temperature (pipe surface)min. °C -55 -55 -55max. °C +140 +140 +140marking

Ex e q IIC T6...T3 GbEx tb IIIC TX Db

Ex e q IIC T6...T3 GbEx tb IIIC TX Db

Ex e q IIC T6...T3 GbEx tb IIIC TX Db

certification ATEX IBExU07ATEX1168 X IBExU07ATEX1168 X IBExU07ATEX1168 Xcertification IECEx IECEx IBE 08.0007X IECEx IBE 08.0007X IECEx IBE 08.0007Xtype of protection gas: increased safety, powder

fillingdust: protection by enclosure

gas: increased safety, powder fillingdust: protection by enclosure

gas: increased safety, powder fillingdust: protection by enclosure

transducer mounting fixture necessary

x x x

1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 Lamb wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended: in reflection arrangement and for a flow velocity of 15 m/spipe diameter max. extended: in diagonal arrangement and for a flow velocity of 25 m/s

l

hb

l

hb

l

hb

II2GII2D

0637 II2GII2D

0637 II2GII2D

0637

Page 30: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

30 TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01

FLUXUS® G70x Technical Specification

technical type GRM1N83 GRP1N83 GRQ1N83order code GLM-NA1TS

GLM-NA1TS/OSGLP-NA1TSGLP-NA1TS/OS

GLQ-NA1TSGLQ-NA1TS/OS

transducer frequency MHz 1 2 4

medium pressure1

min. extended bar - - -min. bar metal pipe:

10 (d > 60 mm)5 (d < 60 mm)plastic pipe: 1

metal pipe:10 (d > 35 mm)5 (d < 35 mm)plastic pipe: 1

metal pipe:10 (d > 15 mm)5 (d < 15 mm)plastic pipe: 1

inner pipe diameter d2

min. extended mm 30 15 7min. recommended mm 40 20 10max. recommended mm 90 50 22max. extended mm 150 70 35pipe wall thicknessmin. mm 2 1 0.5max. mm 5 3 1materialhousing PPSU with stainless steel cap 304

(1.4301), option OS: 316L (1.4404)

PPSU with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)

PPSU with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)

contact surface PPSU PPSU PPSUdegree of protection according to IEC/EN 60529

IP65 IP65 IP65

transducer cabletype 1699 1699 1699length m 4 4 3dimensionslength l mm 74 74 42width b mm 32 32 22height h mm 40.5 40.5 25.5dimensional drawing

ambient temperaturemin. °C -40 -40 -40max. °C +170 +170 +170temperature compensation

x x x

explosion protection

ATEX/IECEx

categoryzone

gas: 2G dust: 2D1 21

gas: 2G dust: 2D1 21

gas: 2G dust: 2D1 21

explosion protection temperature (pipe surface)min. °C -55 -55 -55max. °C +140 +140 +140marking

Ex e q IIC T6...T3 GbEx tb IIIC TX Db

Ex e q IIC T6...T3 GbEx tb IIIC TX Db

Ex e q IIC T6...T3 GbEx tb IIIC TX Db

certification ATEX IBExU07ATEX1168 X IBExU07ATEX1168 X IBExU07ATEX1168 Xcertification IECEx IECEx IBE 08.0007X IECEx IBE 08.0007X IECEx IBE 08.0007Xtype of protection gas: increased safety, powder

fillingdust: protection by enclosure

gas: increased safety, powder fillingdust: protection by enclosure

gas: increased safety, powder fillingdust: protection by enclosure

transducer mounting fixture necessary

x x x

remark on request1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 Lamb wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended: in reflection arrangement and for a flow velocity of 15 m/spipe diameter max. extended: in diagonal arrangement and for a flow velocity of 25 m/s

l

hb

l

hb

l

hb

II2GII2D

0637 II2GII2D

0637 II2GII2D

0637

Page 31: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01 31

Technical Specification FLUXUS® G70x

Lamb Wave Transducers (zone 1, IP68)

technical type GRG1LI3 GRH1LI3 GRK1LI3order code GLG-NA1TS/IP68 GLH-NA1TS/IP68 GLK-NA1TS/IP68transducer frequency MHz 0.2 0.3 0.5

medium pressure1

min. extended bar metal pipe: 10 metal pipe: 10 metal pipe:10 (d > 120 mm)5 (d < 120 mm)

min. bar metal pipe: 15plastic pipe: 1

metal pipe: 15plastic pipe: 1

metal pipe:15 (d > 120 mm)10 (d < 120 mm)plastic pipe: 1

inner pipe diameter d2

min. extended mm 190 120 60min. recommended mm 220 140 80max. recommended mm 900 600 300max. extended mm 1600 1000 500pipe wall thicknessmin. mm 11 7 4max. mm 23 15 9materialhousing PPSU with stainless steel

cap 316Ti (1.4571)PPSU with stainless steel cap 316Ti (1.4571)

PPSU with stainless steel cap 316Ti (1.4571)

contact surface PPSU PPSU PPSUdegree of protection according to IEC/EN 60529

IP68 IP68 IP68

transducer cabletype 2550 2550 2550length m 12 12 12dimensionslength l mm 143.5 143.5 143.5width b mm 54 54 54height h mm 83.5 83.5 83.5dimensional drawing

ambient temperaturemin. °C -40 -40 -40max. °C +100 +100 +100temperature compensation

x x x

explosion protection

ATEX/IECEx

categoryzone

gas: 2G dust: 2D1 21

gas: 2G dust: 2D1 21

gas: 2G dust: 2D1 21

explosion protection temperature (pipe surface)min. °C -55 -55 -55max. °C +140 +140 +140marking

Ex q IIC T6...T3 GbEx tb IIIC TX Db

Ex q IIC T6...T3 GbEx tb IIIC TX Db

Ex q IIC T6...T3 GbEx tb IIIC TX Db

certification ATEX IBExU07ATEX1168 X IBExU07ATEX1168 X IBExU07ATEX1168 Xcertification IECEx IECEx IBE 08.0007X IECEx IBE 08.0007X IECEx IBE 08.0007Xtype of protection gas: powder filling

dust: protection by enclosure

gas: powder fillingdust: protection by enclosure

gas: powder fillingdust: protection by enclosure

transducer mounting fixture necessary

x x x

1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 Lamb wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended: in reflection arrangement and for a flow velocity of 15 m/spipe diameter max. extended: in diagonal arrangement and for a flow velocity of 25 m/s

3 3 3

l

hb

l

hb

l

hb

II2GII2D

0637 II2GII2D

0637 II2GII2D

0637

Page 32: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

32 TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01

FLUXUS® G70x Technical Specification

3 test conditions: 3 months/2 bar (20 m)/20 °C

technical type GRM1LI3 GRP1LI3order code GLM-NA1TS/IP68 GLP-NA1TS/IP68transducer frequency MHz 1 2

medium pressure1

min. extended bar - -min. bar metal pipe:

10 (d > 60 mm)5 (d < 60 mm)plastic pipe: 1

metal pipe:10 (d > 35 mm)5 (d < 35 mm)plastic pipe: 1

inner pipe diameter d2

min. extended mm 30 15min. recommended mm 40 20max. recommended mm 90 50max. extended mm 150 70pipe wall thicknessmin. mm 2 1max. mm 5 3materialhousing PPSU with stainless steel

cap 316Ti (1.4571)PPSU with stainless steel cap 316Ti (1.4571)

contact surface PPSU PPSUdegree of protection according to IEC/EN 60529

IP68 IP68

transducer cabletype 2550 2550length m 12 12dimensionslength l mm 73 73width b mm 31.6 31.6height h mm 46 46dimensional drawing

ambient temperaturemin. °C -40 -40max. °C +100 +100temperature compensation

x x

explosion protection

ATEX/IECEx

categoryzone

gas: 2G dust: 2D1 21

gas: 2G dust: 2D1 21

explosion protection temperature (pipe surface)min. °C -55 -55max. °C +140 +140marking

Ex q IIC T6...T3 GbEx tb IIIC TX Db

Ex q IIC T6...T3 GbEx tb IIIC TX Db

certification ATEX IBExU07ATEX1168 X IBExU07ATEX1168 Xcertification IECEx IECEx IBE 08.0007X IECEx IBE 08.0007Xtype of protection gas: powder filling

dust: protection by enclosure

gas: powder fillingdust: protection by enclosure

1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 Lamb wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended: in reflection arrangement and for a flow velocity of 15 m/spipe diameter max. extended: in diagonal arrangement and for a flow velocity of 25 m/s

3 test conditions: 3 months/2 bar (20 m)/20 °C

technical type GRG1LI3 GRH1LI3 GRK1LI3

3 3

l

hb

l

hb

II2GII2D

0637 II2GII2D

0637

Page 33: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01 33

Technical Specification FLUXUS® G70x

Lamb Wave Transducers (ATEX zone 2, FM or not explosion proof)

technical type GRG1N52 GRH1N52 GRK1N52order code GLG-NA2TS, GLG-NA2TS/OS

GLG-NF2TS, GLG-NF2TS/OSGLG-NNNTS, GLG-NNNTS/OS

GLH-NA2TS, GLH-NA2TS/OSGLH-NF2TS, GLH-NF2TS/OSGLH-NNNTS, GLH-NNNTS/OS

GLK-NA2TS, GLK-NA2TS/OSGLK-NF2TS, GLK-NF2TS/OSGLK-NNNTS, GLK-NNNTS/OS

transducer frequency MHz 0.2 0.3 0.5medium pressure1

min. extended bar metal pipe: 10 metal pipe: 10 metal pipe:10 (d > 120 mm), 5 (d < 120 mm)

min. bar metal pipe: 15plastic pipe: 1

metal pipe: 15plastic pipe: 1

metal pipe:15 (d > 120 mm), 10 (d < 120 mm)plastic pipe: 1

inner pipe diameter d2

min. extended mm 190 120 60min. recommended mm 220 140 80max. recommended mm 900 600 300max. extended mm 1600 1000 500pipe wall thicknessmin. mm 11 7 4max. mm 23 15 9materialhousing PPSU with stainless steel cap 304

(1.4301), option OS: 316L (1.4404)

PPSU with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)

PPSU with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)

contact surface PPSU PPSU PPSUdegree of protection according to IEC/EN 60529

IP67 IP67 IP67

transducer cabletype 1699 1699 1699length m 5 5 5dimensionslength l mm 128.5 128.5 128.5width b mm 51 51 51height h mm 67.5 67.5 67.5dimensional drawing

ambient temperaturemin. °C -40 -40 -40max. °C +170 +170 +170temperature compensation

x x x

explosion protection

ATEX

order code GLG-NA2TS, GLG-NA2TS/OS GLH-NA2TS, GLH-NA2TS/OS GLK-NA2TS, GLK-NA2TS/OScategoryzone

gas: 3G dust: 3D2 22

gas: 3G dust: 3D2 22

gas: 3G dust: 3D2 22

explosion protection temperature (pipe surface)min. °C -55 -55 -55max. °C +150 +150 +150marking

II3G Ex nA IIC T6...T3 GcTa -55...+150 °CII3D Ex tc IIIC TX Dc

II3G Ex nA IIC T6...T3 GcTa -55...+150 °CII3D Ex tc IIIC TX Dc

II3G Ex nA IIC T6...T3 GcTa -55...+150 °CII3D Ex tc IIIC TX Dc

certification - - -type of protection gas: non sparking

dust: protection by enclosuregas: non sparkingdust: protection by enclosure

gas: non sparkingdust: protection by enclosure

transducer mounting fixture necessary

x x x

FM

order code GLG-NF2TS, GLG-NF2TS/OS GLH-NF2TS, GLH-NF2TS/OS GLK-NF2TS, GLK-NF2TS/OSexplosion protection temperaturemin. °C -40 -40 -40max. °C +165 +165 +165marking NI/Cl. I,II,III/Div. 2 /

GP A,B,C,D,E,F,G/Temp. Codes dwg 3860

NI/Cl. I,II,III/Div. 2 /GP A,B,C,D,E,F,G/

Temp. Codes dwg 3860

NI/Cl. I,II,III/Div. 2 /GP A,B,C,D,E,F,G/

Temp. Codes dwg 3860type of protection non incendive non incendive non incendive

1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 Lamb wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended: in reflection arrangement and for a flow velocity of 15 m/spipe diameter max. extended: in diagonal arrangement and for a flow velocity of 25 m/s

l

hb

l

hb

l

hb

Page 34: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

34 TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01

FLUXUS® G70x Technical Specification

technical type GRM1N52 GRP1N52 GRQ1N52order code GLM-NA2TS, GLM-NA2TS/OS

GLM-NF2TS, GLM-NF2TS/OSGLM-NNNTS, GLM-NNNTS/OS

GLP-NA2TS, GLP-NA2TS/OSGLP-NF2TS, GLP-NF2TS/OSGLP-NNNTS, GLP-NNNTS/OS

GLQ-NA2TS, GLQ-NA2TS/OSGLQ-NF2TS, GLQ-NF2TS/OSGLQ-NNNTS, GLQ-NNNTS/OS

transducer frequency MHz 1 2 4medium pressure1

min. extended bar - - -min. bar metal pipe:

10 (d > 60 mm), 5 (d < 60 mm)plastic pipe: 1

metal pipe:10 (d > 35 mm), 5 (d < 35 mm)plastic pipe: 1

metal pipe:10 (d > 15 mm), 5 (d < 15 mm)plastic pipe: 1

inner pipe diameter d2

min. extended mm 30 15 7min. recommended mm 40 20 10max. recommended mm 90 50 22max. extended mm 150 70 35pipe wall thicknessmin. mm 2 1 0.5max. mm 5 3 1materialhousing PPSU with stainless steel cap 304

(1.4301), option OS: 316L (1.4404)

PPSU with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)

PPSU with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)

contact surface PPSU PPSU PPSUdegree of protection according to IEC/EN 60529

IP65 IP65 IP65

transducer cabletype 1699 1699 1699length m 4 4 3dimensionslength l mm 74 74 42width b mm 32 32 22height h mm 40.5 40.5 25.5dimensional drawing

ambient temperaturemin. °C -40 -40 -40max. °C +170 +170 +170temperature compensation

x x x

explosion protection

ATEX

order code GLM-NA2TS, GLM-NA2TS/OS GLP-NA2TS, GLP-NA2TS/OS GLQ-NA2TS, GLQ-NA2TS/OScategoryzone

gas: 3G dust: 3D2 22

gas: 3G dust: 3D2 22

gas: 3G dust: 3D2 22

explosion protection temperature (pipe surface)min. °C -55 -55 -55max. °C +150 +150 +150marking

II3G Ex nA IIC T6...T3 GcTa -55...+150 °CII3D Ex tc IIIC TX Dc

II3G Ex nA IIC T6...T3 GcTa -55...+150 °CII3D Ex tc IIIC TX Dc

II3G Ex nA IIC T6...T3 GcTa -55...+150 °CII3D Ex tc IIIC TX Dc

certification - - -type of protection gas: non sparking

dust: protection by enclosuregas: non sparkingdust: protection by enclosure

gas: non sparkingdust: protection by enclosure

transducer mounting fixture necessary

x x x

FM

order code GLM-NF2TS, GLM-NF2TS/OS GLP-NF2TS, GLP-NF2TS/OS GLQ-NF2TS, GLQ-NF2TS/OSexplosion protection temperaturemin. °C -55 -55 -55max. °C +165 +165 +165marking NI/Cl. I,II,III/Div. 2 /

GP A,B,C,D,E,F,G/Temp. Codes dwg 3860

NI/Cl. I,II,III/Div. 2 /GP A,B,C,D,E,F,G/

Temp. Codes dwg 3860

NI/Cl. I,II,III/Div. 2 /GP A,B,C,D,E,F,G/

Temp. Codes dwg 3860type of protection non incendive non incendive non incendive

remark on request1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 Lamb wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended: in reflection arrangement and for a flow velocity of 15 m/spipe diameter max. extended: in diagonal arrangement and for a flow velocity of 25 m/s

l

hb

l

hb

l

hb

Page 35: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01 35

Technical Specification FLUXUS® G70x

Lamb Wave Transducers (ATEX zone 2 or not explosion proof, IP68)

technical type GRG1LI8 GRH1LI8 GRK1LI8order code GLG-NA2TS/IP68

GLG-NNNTS/IP68GLH-NA2TS/IP68GLH-NNNTS/IP68

GLK-NA2TS/IP68GLK-NNNTS/IP68

transducer frequency MHz 0.2 0.3 0.5

medium pressure1

min. extended bar metal pipe: 10 metal pipe: 10 metal pipe:10 (d > 120 mm),5 (d < 120 mm)

min. bar metal pipe: 15plastic pipe: 1

metal pipe: 15plastic pipe: 1

metal pipe:15 (d > 120 mm),10 (d < 120 mm)plastic pipe: 1

inner pipe diameter d2

min. extended mm 190 120 60min. recommended mm 220 140 80max. recommended mm 900 600 300max. extended mm 1600 1000 500pipe wall thicknessmin. mm 11 7 4max. mm 23 15 9materialhousing PPSU with stainless steel

cap 316Ti (1.4571)PPSU with stainless steel cap 316Ti (1.4571)

PPSU with stainless steel cap 316Ti (1.4571)

contact surface PPSU PPSU PPSUdegree of protection according to IEC/EN 60529

IP68 IP68 IP68

transducer cabletype 2550 2550 2550length m 12 12 12dimensionslength l mm 143.5 143.5 143.5width b mm 54 54 54height h mm 83.5 83.5 83.5dimensional drawing

ambient temperaturemin. °C -40 -40 -40max. °C +100 +100 +100temperature compensation

x x x

explosion protection

ATEX

order code GLG-NA2TS/IP68 GLH-NA2TS/IP68 GLK-NA2TS/IP68categoryzone

gas: 3G dust: 3D2 22

gas: 3G dust: 3D2 22

gas: 3G dust: 3D2 22

explosion protection temperature (pipe surface)min. °C -40 -40 -40max. °C +90 +90 +90marking

II3G Ex nA IIC T6...T5 GcTa -40...+90 °CII3D Ex tc IIIC TX Dc

II3G Ex nA IIC T6...T5 GcTa -40...+90 °CII3D Ex tc IIIC TX Dc

II3G Ex nA IIC T6...T5 GcTa -40...+90 °CII3D Ex tc IIIC TX Dc

certification - - -type of protection gas: non sparking

dust: protection by enclosure

gas: non sparkingdust: protection by enclosure

gas: non sparkingdust: protection by enclosure

transducer mounting fixture necessary

x x x

1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 Lamb wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended: in reflection arrangement and for a flow velocity of 15 m/spipe diameter max. extended: in diagonal arrangement and for a flow velocity of 25 m/s

3 test conditions: 3 months/2 bar (20 m)/20 °C

3 3 3

l

hb

l

hb

l

hb

Page 36: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

36 TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01

FLUXUS® G70x Technical Specification

technical type GRM1LI8 GRP1LI8order code GLM-NA2TS/IP68

GLM-NNNTS/IP68GLP-NA2TS/IP68GLP-NNNTS/IP68

transducer frequency MHz 1 2medium pressure1

min. extended bar - -min. bar metal pipe:

10 (d > 60 mm)5 (d < 60 mm)plastic pipe: 1

metal pipe:10 (d > 35 mm)5 (d < 35 mm)plastic pipe: 1

inner pipe diameter d2

min. extended mm 30 15min. recommended mm 40 20max. recommended mm 90 50max. extended mm 150 70pipe wall thicknessmin. mm 2 1max. mm 5 3materialhousing PPSU with stainless steel

cap 316Ti (1.4571)PPSU with stainless steel cap 316Ti (1.4571)

contact surface PPSU PPSUdegree of protection according to IEC/EN 60529

IP68 IP68

transducer cabletype 2550 2550length m 12 12dimensionslength l mm 73 73width b mm 31.6 31.6height h mm 46 46dimensional drawing

ambient temperaturemin. °C -40 -40max. °C +100 +100temperature compensation

x x

explosion protection

ATEX

order code GLM-NA2TS/IP68 GLP-NA2TS/IP68categoryzone

gas: 3G dust: 3D2 22

gas: 3G dust: 3D2 22

explosion protection temperature (pipe surface)min. °C -40 -40max. °C +90 +90marking

II3G Ex nA IIC T6...T5 GcTa -40...+90 °CII3D Ex tc IIIC TX Dc

II3G Ex nA IIC T6...T5 GcTa -40...+90 °CII3D Ex tc IIIC TX Dc

certification - -type of protection gas: non sparking

dust: protection by enclosure

gas: non sparkingdust: protection by enclosure

transducer mounting fixture necessary

x x

1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 Lamb wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended: in reflection arrangement and for a flow velocity of 15 m/spipe diameter max. extended: in diagonal arrangement and for a flow velocity of 25 m/s

3 test conditions: 3 months/2 bar (20 m)/20 °C

3 3

l

hb

l

hb

Page 37: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01 37

Technical Specification FLUXUS® G70x

Lamb Wave Transducers (not explosion proof, connection system AS)

technical type GRG1NC3 GRH1NC3 GRK1NC3order code GLG-NNNAS GLH-NNNAS GLK-NNNAStransducer frequency MHz 0.2 0.3 0.5

medium pressure1

min. extended bar metal pipe: 10 metal pipe: 10 metal pipe:10 (d > 120 mm)5 (d < 120 mm)

min. bar metal pipe: 15plastic pipe: 1

metal pipe: 15plastic pipe: 1

metal pipe:15 (d > 120 mm)10 (d < 120 mm)plastic pipe: 1

inner pipe diameter d2

min. extended mm 190 120 60min. recommended mm 220 140 80max. recommended mm 900 600 300max. extended mm 1600 1000 500pipe wall thicknessmin. mm 11 7 4max. mm 23 15 9materialhousing PPSU with stainless steel

cap 304 (1.4301)PPSU with stainless steel cap 304 (1.4301)

PPSU with stainless steel cap 304 (1.4301)

contact surface PPSU PPSU PPSUdegree of protection according to IEC/EN 60529

IP65 IP65 IP65

transducer cabletype 1699 1699 1699length m 5 5 5dimensionslength l mm 128.5 128.5 128.5width b mm 51 51 51height h mm 67.5 67.5 67.5dimensional drawing

ambient temperaturemin. °C -40 -40 -40max. °C +170 +170 +170temperature compensation

x x x

1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 Lamb wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended: in reflection arrangement and for a flow velocity of 15 m/spipe diameter max. extended: in diagonal arrangement and for a flow velocity of 25 m/s

l

hb

l

hb

l

hb

Page 38: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

38 TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01

FLUXUS® G70x Technical Specification

technical type GRM1NC3 GRP1NC3 GRQ1NC3order code GLM-NNNAS GLP-NNNAS GLQ-NNNAStransducer frequency MHz 1 2 4

medium pressure1

min. extended bar - - -min. bar metal pipe:

10 (d > 60 mm)5 (d < 60 mm)plastic pipe: 1

metal pipe:10 (d > 35 mm)5 (d < 35 mm)plastic pipe: 1

metal pipe:10 (d > 15 mm)5 (d < 15 mm)plastic pipe: 1

inner pipe diameter d2

min. extended mm 30 15 7min. recommended mm 40 20 10max. recommended mm 90 50 22max. extended mm 150 70 35pipe wall thicknessmin. mm 2 1 0.5max. mm 5 3 1materialhousing PPSU with stainless steel

cap 304 (1.4301)PPSU with stainless steel cap 304 (1.4301)

PPSU with stainless steel cap 304 (1.4301)

contact surface PPSU PPSU PPSUdegree of protection according to IEC/EN 60529

IP65 IP65 IP65

transducer cabletype 1699 1699 1699length m 4 4 3dimensionslength l mm 74 74 42width b mm 32 32 22height h mm 40.5 40.5 25.5dimensional drawing

ambient temperaturemin. °C -40 -40 -40max. °C +170 +170 +170temperature compensation

x x x

remark on request1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 Lamb wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended: in reflection arrangement and for a flow velocity of 15 m/spipe diameter max. extended: in diagonal arrangement and for a flow velocity of 25 m/s

l

hb

l

hb

l

hb

Page 39: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01 39

Technical Specification FLUXUS® G70x

Transducer Mounting Fixture

Order Code

1, 2 3 4 5 6 7...9 10, 11 no. of character

tran

sduc

er

mou

ntin

g fix

ture

tran

sduc

er

- mea

sure

men

t ar

rang

emen

t

size

- fixat

ion

oute

r pi

pe

diam

eter

/ optio

n

description

VL Variofix L

VС Variofix C

K transducers with transducer frequency G, H, K

M transducers with transducer frequency M, P

Q transducers with transducer frequency Q

D reflection arrangement or diagonal arrangement

R reflection arrangement

S small

M medium

L large

S tension straps

W welding

N without fixation

002 10...20 mm

004 20...40 mm

T36 40...360 mm

013 10...130 mm

036 130...360 mm

092 360...920 mm

200 920...2000 mm

IP68 degree of protection IP68

OS housing with stainless steel 316

Z special design

example

VL K - D S - S 200 Variofix L and tension straps for transducers with transducer frequency G, H, K

- - /

Page 40: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

40 TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01

FLUXUS® G70x Technical Specification

Variofix L (VLK, VLM, VLQ) material: stainless steel 304 (1.4301), 301 (1.4310), 410 (1.4006)option OS: 316 (1.4571), 316L (1.4404), 17-7PH (1.4568)

inner length:VLK: 348 mm,option IP68: 368 mmVLM: 234 mmVLQ: 176 mm

dimensions:VLK: 423 x 90 x 93 mm,option IP68: 443 x 94 x 105 mmVLM: 309 x 57 x 63 mmVLQ: 247 x 43 x 47 mm

Variofix C (VC) material: stainless steel 304 (1.4301), 301 (1.4310)option OS: 316 (1.4571)

inner length:VCK-*L: 500 mmVCK-*S: 350 mmVCM: 400 mmVCQ: 250 mm

dimensions:VCK-*L: 560 x 122 x 102 mm,option IP68: 560 x 126 x 120 mmVCK-*S: 410 x 122 x 102 mm,option IP68: 410 x 126 x 120 mmVCM: 460 x 96 x 80 mmVCQ: 310 x 85 x 62 mm

Page 41: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01 41

Technical Specification FLUXUS® G70x

Coupling Materials for Transducers

Technical Data

normal temperature range(4th character of transducer order code = N)

extended temperature range(4th character of transducer order code = E)

< 100 °C < 170 °C < 150 °C < 200 °C< 24 h coupling compound type N

or coupling foil type VTcoupling compound type E or coupling foil type VT

coupling compound type E or coupling foil type VT

coupling compound type E or H or coupling foil type VT

long time measurement

coupling foil type VT1 coupling foil type VT2 coupling foil type VT1 coupling foil type VT2

1 < 5 years2 < 6 months

type order code ambient temperature material remark°C

coupling compound type N

990739-1 -30...+130 mineral grease paste

coupling compound type E

990739-2 -30...+200 silicone paste

coupling compound type H

990739-3 -30...+250 fluoropolymer paste

coupling foil type VT 990739-0 -10...+200 fluoroelastomer for transducers with transducer frequency G, H, K

990739-6 for shear wave transducers with transducer frequency M, P

990739-14 for shear wave transducers IP68 and Lambwave transducers with transducer frequency M, P

990739-5 for transducers with transducer frequency Q

Page 42: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

42 TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01

FLUXUS® G70x Technical Specification

Damping Mats (optional)Damping mats will be used for the gas measurement to reduce acoustic noise influences on the measure-ment.

transducer damping matTransducer damping mats will be installed below the transducers.

pipe damping matPipe damping mats will be installed if the sound propagation is disturbed at reflection points (e.g. flange,weld). Depending on the noise, the pipe damping mats will be installed at one or both sides of the transducerdamping mat. If the local conditions are unknown, pipe damping mats should be installed.

Technical Data

type E30R4 E30R3width mm 225 50thickness mm 0.7length (per roll) m 10weight kg/m² 1.015ambient temperature °C -30...+80properties self-adhesive

outer pipe diameter ≥ 900 mm outer pipe diameter < 900 mm

transducer damping matpipe damping mat (optional)

installation length

diagonal arrangement

reflection arrangement

Page 43: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01 43

Technical Specification FLUXUS® G70x

Dimensioning

transducer damping mattransducer

mounting fixtureorder code

type number of layers

transducer damping mat transducer damping mat+

2x pipe damping matmax. instal-lation length number of rolls1

max. instal-lation length number of rolls1

[mm] standard2 extended 2 [mm] standard extendedVarioFix LVLK GLG E30R4 3 890 4 4 1830 9 13

GSG 3 4 4 9 11GLH 2 2 3 4 7GLK 1 1 1 1 2GSK 1 1 1 2 3

VLK-**-****/IP68 GLG E30R4 3 930 5 5 1910 10 14GSG 3 5 5 10 11GLH 2 2 3 5 7GLK 1 1 1 2 2GSK 1 1 2 2 3

VLM GLM E30R3 1 660 1 1 1360 1 2GSM 1 1 1 1 2GLP 1 1 1 1 1GSP 1 1 1 1 1

VLQ GLQ E30R3 1 540 1 1 1120 1 1Variofix CVCK-*LVCK-*L-****/IP68

GLG E30R4 3 1160 6 6 2360 11 15GSG 3 6 6 11 12GLH 2 3 4 5 8GLK 1 1 1 2 2GSK 1 2 2 3 3

VCK-*SVCK-*S-****/IP68

GLG E30R4 3 860 4 4 1760 7 10GSG 3 4 4 7 8GLH 2 2 3 4 5GLK 1 1 1 1 2GSK 1 1 1 2 2

VCM GLM E30R3 1 960 1 2 1960 2 3GSM 1 1 1 2 2GLP 1 1 1 1 1GSP 1 1 1 1 1

VCQ GLQ E30R3 1 660 1 1 1360 1 11 calculation on the base of:

- max. installation length (installation of one transducer mounting fixture per transducer in reflection arrangement) and- max. recommended pipe diameter (standard) or max. extended pipe diameter (extended) (for inner pipe diameter max. recommended and max. extended see Technical Data of the Transducers from page 18)

2 calculation for the number of rolls when both transducers are mounted in one transducer mounting fixture (reflection arrangement) or in diagonal arrangement: number of rolls/2 and round up to the nearest integer

Page 44: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

44 TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01

FLUXUS® G70x Technical Specification

Connection Systemsconnection system TSconnection with extension cable direct connection transducers

technical type

*****8*

****LI*

*****52

connection system ASconnection with extension cable transducers

technical type*****Z7*****C3

transducer frequency(3d character of transducer

order code)

G, H, K M, P Q S

TS

x l x l x l x lcable length m 5 ≤ 300 4 ≤ 300 3 ≤ 90 2 ≤ 40

cable length (option LC) m 9 ≤ 300 - - - - - -cable length (option IP68) m 12 ≤ 300 12 ≤ 300 - - - -

AS

x y l x y l x y l x y lcable length m 2 3 ≤ 100 2 2 ≤ 100 2 1 ≤ 50 1 1 ≤ 20

cable length (option LC) m 2 7 ≤ 100 - - - - - - - - -

x, y - transducer cable lengthl - max. length of extension cable

(only G704, G705)

tran

smitt

er

x

l

JB01

tran

smitt

er

x

tran

smitt

er

x

l

JB01, JBP2, JBP3

tran

smitt

er

x

tran

smitt

er

x

l

JB02, JB03

tran

smitt

er

x

tran

smitt

er

l x y

Page 45: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01 45

Technical Specification FLUXUS® G70x

Transducer Cable

Technical Data

transducer cable extension cabletype 1699 2550 (option IP68) 6111 2551 2615connection system AS TSstandard length m see table above -

max. length m - see table aboveambient temperature °C -55...+200 -40...+100 -100...+225 -25...+80 -40...+70properties longitudinal water

tighthalogen free

fire propagation test according to IEC 60332-1

combustion test according to IEC 60754-2

sheathmaterial stainless steel 304

(1.4301)- stainless steel 304

(1.4301)- -

option OS: 316L (1.4404)

option OS: 316L (1.4404)

outer diameter mm 8 - 8 - -cable jacketmaterial PTFE PUR PFA TPE-O PURouter diameter mm 2.9 5.2 ±0.2 2.7 8 12thickness mm 0.3 0.9 0.5 2color brown gray white black blackshield x x x x x

110

Page 46: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

46 TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01

FLUXUS® G70x Technical Specification

Junction Box

Technical Data

Dimensions

technical type JB01S4E3M JB02 JB03 JBP2 JBP3dimensions see dimensional

drawingsee dimensional drawing

see dimensional drawing

see dimensional drawing

see dimensional drawing

fixation wall mounting, optional: 2 " pipe mounting

wall mounting, optional: 2 " pipe mounting

wall mounting, optional: 2 " pipe mounting

wall mounting, optional: 2 " pipe mounting

wall mounting, optional: 2 " pipe mounting

materialhousing stainless steel 316L

(1.4404)stainless steel 304 (1.4301)option OS: 316L (1.4404)

stainless steel 304 (1.4301)option OS: 316L (1.4404)

stainless steel 316L (1.4404)

stainless steel 316L (1.4404)

gasket silicone silicone silicone silicone siliconedegree of protection according to IEC/EN 60529

IP67 IP67 IP67 IP67 IP67

ambient temperaturemin. °C -40 -40 -40 -40 -40max. °C +80 +80 +80 +80 +80explosion protection

ATEX/IECEx

zone 1 2 - 2 -marking

Ex e mb IIC (T6)...T4 GbEx tb IIIC T 100 °C DbTa -40...+(70)80 °C

II3G Ex nA IIC (T6)...T4 GcII3D Ex tc IIIC T 100 °C DcTa -40...+(70)80 °C

- II3G Ex nA IIC (T6)...T4 GcII3D Ex tc IIIC T 100 °C DcTa -40...+(70)80 °C

-

certification ATEX IBExU06ATEX1161 - - - -certification IECEx IECEx IBE 08.0006 - - - -type of protection gas:

• increased safety• decoupled

network: encapsulation

dust: protection by enclosure

gas: non sparkingdust: protection by enclosure

- gas: non sparkingdust: protection by enclosure

-

in mm

II2GII2D

0637

174

1192

wall mounting holder

70163.5

156

Ø 9

thread: 3x M20 x 1.5cable gland: max. 2x M20

Page 47: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01 47

Technical Specification FLUXUS® G70x

2 " Pipe Mounting Kit (optional)

Terminal Assignment

JB01

JB02, JB03

&

&

'

'

& &

# #

transducers

terminal strip KL1terminal connection

V transducer , signalVS transducer , internal shieldRS transducer , internal shieldR transducer , signalcable gland external shield

extension cable

terminal strip KL2terminal connection

TV signalTVS internal shieldTRS internal shieldTR signalshield terminal external shield

shield terminal

equipotential bonding terminal(at wall mounting holder)

&

#

&

&

'

'

transducers

terminal connectionXV transducer , SMB connectorXR transducer , SMB connectorcable gland external shield

extension cable

terminal strip KL2terminal connection

TV signalTVS internal shieldTRS internal shieldTR signalshield terminal external shield

shield terminal

JB02: equipotential bonding terminal(at wall mounting holder)

Page 48: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

48 TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01

FLUXUS® G70x Technical Specification

JBP2, JBP3

&

&

'

'

&

&

# #

transducers

terminal strip KL1terminal connection

TV transducer , signalTVS transducer , internal shieldTRS transducer , internal shieldTR transducer , signalcable gland external shield

extension cable

terminal strip KL2terminal connection

TV signalTVS internal shieldTRS internal shieldTR signalshield terminal external shield

shield terminal

JBP2: equipotential bonding terminal(at wall mounting holder)

Page 49: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01 49

Technical Specification FLUXUS® G70x

Clamp-on Temperature Probe (optional)

Technical Datatechnical type PT12N PT12N PT12N PT12N PT12F PT12Forder code 670415-1

770415-1670414-1770414-1

770415-1A2 770414-1A2 670415-2 670414-2

design ATEX zone 2 short response timetype Pt100 2x Pt100

matched according to EN 1434-1

Pt100 2x Pt100 matched according to EN 1434-1

Pt100 2x Pt100 matched according to EN 1434-1

connection 4-wire 4-wire 4-wiremeasuring range °C -30...+250 -30...+250 -50...+250

accuracy T ±(0.15 °C + 2 . 10-3 . |T [°C]|)class A

±(0.15 °C + 2 . 10-3 . |T [°C]|)class A

±(0.15 °C + 2 . 10-3 . |T [°C]|)class A

accuracy ∆T - ≤ 0.1 K(3 K < ∆T < 6 K), more corresponding to EN 1434-1

- ≤ 0.1 K(3 K < ∆T < 6 K), more corresponding to EN 1434-1

- ≤ 0.1 K(3 K < ∆T < 6 K), more corresponding to EN 1434-1

response time s 50 50 8housing aluminum aluminum PEEK, stainless steel 304

(1.4301), copperdegree of protection according to IEC/EN 60529

IP66 IP66 IP66

weight (without con-nector)

kg 0.25 0.5 0.25 0.5 0.32 0.64

fixation clamp-on clamp-on clamp-onaccessoriesthermal conductivity paste 200 °C

670415-1: x770415-1: -

670414-1: x770414-1: -

- x

thermal conductivity foil 250 °C

x x x

plastic protection plate, insulation foam

- - x

dimensionslength l mm 15 15 14width b mm 15 15 30height h mm 20 20 27dimensional drawing 670415-1: A

770415-1: C670414-1: A770414-1: C

C B

explosion protection

ATEX

zone - 2 -explosion protection temperaturemin. °C - -30 -max. °C - +250 -marking -

II3G Ex nA IIC T6...T2 GcTa -30...+250 °C

-

extension cableh

l

junction box

C

B

h

l

A

Cextension cable

h

l

Page 50: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

50 TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01

FLUXUS® G70x Technical Specification

Connection

Temperature Probe

Connector

Cable

pin cable of temperature probe extension cable1 white/blue blue2 red/blue gray3, 4, 5 not connected6 red red7 white white8 not connected

cable of temperature probe extension cabletype 4 x 0.25 mm² black or white LIYCY 8 x 0.14 mm² graystandard length m 3 5/10/25max. length m - 200cable jacket PTFE PVC

Junction Boxtechnical type JBT2 JBT3dimensions see dimensional drawing see dimensional drawingfixation wall mounting

optional: 2 " pipe mountingwall mountingoptional: 2 " pipe mounting

materialhousing stainless steel 304 (1.4301) stainless steel 304 (1.4301)gasket silicone siliconedegree of protection according to IEC/EN 60529

IP67 IP67

cable gland max. 2x M12 max. 2x M12ambient temperaturemin. °C -40 -40max. °C +80 +80explosion protection

ATEX

zone 2 -marking

II3G Ex nA IIC (T6)...T4 GcII3D Ex tc IIIC T 100 °C DcTa -40...+(70)80 °C

-

certification - -type of protection gas: non sparking, dust: protection by enclosure -

red/blue red white/blue white

Page 51: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01 51

Technical Specification FLUXUS® G70x

Terminal Assignment

JBT2, JBT3

# #

temperature probe

terminal strip KL1terminal connection

1 red2 red/blue3 white4 white/blue

extension cable

terminal strip KL2terminal connection

1 red2 gray3 white4 blue

shield terminal

JBT2: equipotential bonding terminal(at wall mounting holder)

Page 52: Gas Ultrasonic Flowmeter for Permanent Installation...Gas Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features •

52 TSFLUXUS_G7V1-6-1EN_Leu, 2014-06-01

FLUXUS® G70x Technical Specification

FLEXIM GmbH

Wolfener Str. 36

12681 Berlin

Germany

Tel.: +49 (30) 93 66 76 60

Fax: +49 (30) 93 66 76 80

internet: www.flexim.com

e-mail: [email protected]

Subject to change without notification. Errors excepted.

FLUXUS® is a registered trademark of FLEXIM GmbH.