StorageTank
UV
O3
OzoneGenerator
HeatExchanger
Treated Water Supply
Pointsof Use
1. Determine O3 concentration in storage tank.
2. Determine O3 concentration is “zero” after ozone destruct loop during production.
3. During loop sanitization, determine O3 concentration at loop return. No production during this time.
O3
O3
TOC
TOCC O3
C
3. Storage and Distribution
Measurement Points ConductivityC O3 Dissolved OzoneTOC TOCTOC
2
Water Purification Technologies
Global Pharmaceutical Water Compliance Conductivity and TOC Requirements
Pharmaceutical Water Preparation
Unit ProcessParameters
Cond/Res Temp TOC pH/ORP O3 Press Level Flow
Media Filters, Depth & Sand Filters • • • •
Softeners • • • •
Carbon (GAC) Filters • • • •
pH Adjust (Acid/Caustic Inj.) *PID Control • *Reverse Osmosis • • • • • •
Ion-Exchange (Mixed-Bed) • • • •
Electrodeionization • • • •
Distillation • • •
Ozone (injection, destruction, sanitization) •
Storage Tank • • • •
Distribution Loop • • • • • •
Pharmacopeia Requirements for TOC
Calibration and Test Requirements for ConductivityCOnDUCTIvITy ELECTROnICS (ResIsTAnCe MeAsuReMenT)
• Calibration required, frequency not specified, determined by owner, annual calibration accepted.
• Temperature measurement circuit to be verified.
• Instrument reports uncompensated conductivity or resistivity.
• Display resolution of 0.1 μs/cm minimum. 1.0 μs/cm resolution is unacceptable.
• Verify performance to ±0.1 μs/cm by replacing sensor with traceable precision (0.1%) resistor. For example: 50 kΩ resistor with 0.1 cm-1 cell constant should display 2.0 ± 0.1 μs/cm.
COnDUCTIvITy SEnSOR
• Cell constant known to ± 2%. • Calibrate sensor in a solution with a stated conductivity (from nIsT or certified solution). • Calibrate sensor in a solution prepared to a specific conductivity (AsTM D1125 standard or ultrapure water). • Calibrate sensor vs. another calibrated traceable sensor – performed in-line or in-lab or return to manufacturer.
• Temperature accurate to ± 2°C.
Pharmacopeia Requirements for Conductivity
Pharmaceutical Water Distribution
Attribute USP EP JP ChP IPSource Water for PW and WFI us, eu, Japan, WHO
drinking waterHuman consumption JP water specification Potable water or
Purified WaterPotable water or Purified Water
PURIFIED WATERProduction Method suitable process suitable process Distillation, ion-exchange,
uF, or combinationDistillation, ion-exchange,
or suitable processDistillation, ion-exchange,
or suitable process
Total Aerobic (cfu/mL) 100 100 100 100 100
Conductivity (μs/cm at 25°C) 1.3 (3 stage) 5.1 (1 stage) 1.3 on-line or 2.1 off-line 5.1 (1 stage) 1.3 (3 stage)
TOC (mg/L) 0.5 0.5 (optional) 0.5 (0.3 for control) 0.5 0.5
nitrates (ppm) 0.2 0.2 Required
Acidity/Alkalinity Required
Ammonium (ppm) 0.2 Required
Oxidizable Substances Required
WATER FOR InJECTIOnProduction Method Distillation or suitable
processDistillation Distillation or RO with uF,
from Purified WaterDistillation Distillation
Total Aerobic (cfu/100 mL) 10 10 10 10 10
Conductivity (μs/cm at 25°C) 1.3 (3 stage) 1.3 (3 stage) 1.3 on-line or 2.1 off-line 1.3 (3 stage) 1.3 (3 stage)
TOC (mg/L) 0.5 0.5 0.5 (0.3 for control) 0.5 0.5
Bacterial Endotoxins (eu/mL) 0.25 0.25 0.25 0.25 0.25
nitrates (ppm) 0.2 0.2 Required
Heavy Metals Required
Acidity/Alkalinity Required
Ammonium (ppm) 0.2
Bulk Purified Water and Water for Injection
www.mt.com/ism-pharma
5000TOCi with IsM• Complete process control and precise
data trending• Plug and Measure functionality• Instant notification of excursion• simplified record keeping for ssT and CAL• Trouble-free maintenance• Robust, reliable design virtually free from
moving parts• Compliant with all pharmacopeias Delivering the power of an analyzer with the convenience of a sensor.
M800• Multi-channel (2 or 4) transmitter with
multi-parameter measurement flexibility• Intelligent sensor Management (IsM)• Plug and Measure simplicity• iMonitor – predictive maintenance• Dynamic Lifetime Indicator • Trending graphics for all sensors • Assured compliance with pharmacopeia
regulations simple menu operation with unparalleled performance and accuracy.
M300• Multi-channel (2) transmitter with
multi-parameter measurement flexibility• Intelligent sensor Management (IsM)• Plug and Measure simplicity• Measurements are available within seconds
of connection• Meets all pharmacopeias’ requirements
for calibration Quick and easy installation thanks to Plug and Measure sensor capabilities.
uniCond® with IsM • Conductivity Factory Calibration data stored
internally in sensor• Plug and Measure functionality• Integral high-performance measuring circuit• Robust digital output signal• Complete calibration for pharmaceutical
compliance Calibration tracked by iMonitor (M800) to ensure compliance.
Ozone with IsM• Rapid, accurate response• easy maintenance with drop-in membrane• Plug and Measure functionality• Positive zero detection• Assured compliance with “no added
substance” rule Confidently monitor complete sanitization and confirm ozone destruction.
Compliance comes easy with METTLER TOLEDO Thornton‘s Intelligent Sensor Management (ISM®)
Pointsof Use
1. Pre-treatment
2. Purification 3. Storage and Distribution
Multi-mediaFilter
RO Unit Ultra FiltrationUnit
Electrodeionization(EDI)
UV
Softener GAC
UV
CartridgeFilter(1-5µ)
BreakTank
Distillationfor WFI
PW or HPW
StorageTank
FeedWater
Water treatment can bedivided into three stages:
1. Pre-treatment
2. Purification
3. Storage and Distribution
O3CC
C
CC
CC
C C
C TOC
TOC
Measurement Points ConductivityC TOC TOCTOC
Calibration and Test Requirements for TOCTOC REqUIREMEnTS
• Instrument must have a Limit of Detection of 0.050 mg C/L (50 ppb TOC).
• Calibrate according to manufacturer’s recommendations.
• Must distinguish inorganic carbon, i.e., CO2, HCO3, from organic carbon.
• Must meet system suitability Test (ssT) periodically
SySTEM SUITABILITy TESTInG (ssT)
• Measure TOC of water used to prepare these solutions, Rw. not to exceed 100 ppb.
• Measure TOC of solution prepared to concentration of 0.50 mg carbon/L (as sucrose), Rs.
• Measure TOC of solution prepared to concentration 0.50 mg carbon/L (as p-benzoquinone), Rss.
• The accepted frequency for testing is upon uV lamp change.
• Response shall be between 85 and 115%.
Response = 100 xRss - Rw
Rs - Rw( )
1
3
The multi-parameter
M800
METTLER TOLEDO Thornton, Inc.36 Middlesex TurnpikeBedford MA 01730Tel: +1-781-301-8600Fax: [email protected]
subject to technical changes© MeTTLeR TOLeDO Thornton, Inc.Printed in usA 58 087 003 05/13
* shaded boxes – no test required