Gyrolab CDs Product Information Sheet D0012493/D Rapid assay development – Ready to use CDs Save time and minimize sample consumption – Integrated, nanoliter-scale assay workflows Full flexibility in assay design High quality results in less than 1 hour Immunoassays from discovery to the clinic More information from less sample The Gyrolab™ Bioaffy CD range offers an open, nanoliter-scale immunoassay. A broad dynamic range ensures suitability for pharmacokinetic/pharmacodynamic studies and biomarker monitoring as well as product quantification and impurity testing during bioprocess development and production. Analytes measured from picogram to milligram The broad dynamic range minimizes the need for dilutions and repeats, providing a scaleable assay format that does not need to be re-developed. Effects on sample integrity are minimized by lowering the minimum required dilution (MRD). Full pharmacokinetic profiles from three individual mice following i.v. administration of a therapeutic antibody (Bioaffy 200). Data courtesy of Astra Zeneca. The broad dynamic range is well suited when testing for impurities such as HCPs or protein A. 0 500 300 400 200 100 Time (h) 3 mg/kg i.v. Mouse 1 Mouse 2 Mouse 3 Pharmacokinetics: one mouse/onePK 100000 1000 1000 Rat lgG1 (ng/ml) 1 10 1000 10000 100 Concentration (ng/mL) Impurity testing: HCP quantification 100 10 1 Response Bioaffy 200 Bioaffy 20HC
Transcript
Page 1: Gyrolab CDs - Future Science · olunsistov Cme nt e delivery and flow control -hydrophobic barriers stop liquid flow Overflow channels for excess liquid and reproducible volume definition

Gyrolab CDs

Product Information Sheet D0012493/D

Rapid assay development

– Ready to use CDs

Save time and minimize sample consumption

– Integrated, nanoliter-scale assay workflows

Full flexibility in assay design

High quality results in less than 1 hour

Immunoassays from discovery to the clinic

More information from less sample

The Gyrolab™ Bioaffy CD range offers an open, nanoliter-scale immunoassay. A broad dynamic range ensures suitability for pharmacokinetic/pharmacodynamic studies and biomarker monitoring as well as product quantification and impurity testing during bioprocess development and production.

Analytes measured from picogram to milligram

The broad dynamic range minimizes the need for dilutions and repeats, providing a scaleable assay format that does not need to be re-developed. Effects on sample integrity are minimized by lowering the minimum required dilution (MRD).

Full pharmacokinetic profiles from three individual micefollowing i.v. administration of a therapeutic antibody (Bioaffy 200).Data courtesy of Astra Zeneca.

The broad dynamic range is well suited when testing for impurities such as HCPs or protein A.

0 500300 400200100

Time (h)

3 mg/kg i.v.

Mouse 1Mouse 2Mouse 3

Pharmacokinetics: one mouse/onePK100000

1000

1000

Rat

lgG

1 (n

g/m

l)

1 10 1000 10000100

Concentration (ng/mL)

Impurity testing: HCP quantification

100

10

1

Resp

onse

Bioaffy 200Bioaffy 20HC

Page 2: Gyrolab CDs - Future Science · olunsistov Cme nt e delivery and flow control -hydrophobic barriers stop liquid flow Overflow channels for excess liquid and reproducible volume definition

2 D0012493/D

Gyrolab Bioaffy CD

QC Samples Mean(μg/L)

Inter-CD CV (%)*

Relative error (%)

Total error (%)

500 504.21 7.7 0.8 8.6

375 351.98 7.0 -6.1 13.1

250 242.79 3.7 -11.7 15.4

5 4.84 4.5 -3.1 7.6

2 1.94 6.9 -2.8 9.7

* Inter-CD CV% calculated using ANOVA.

Intra-CD Gyrolab Bioaffy 200: in-house validation using a clinical PK assay of a commercially available biotherapeutic recombinant antibody – 5 QC samples in quadruplicate from a single run.

Integrated, nanoliter-scale workflows

Samples are processed simultaneously in parallel, with every step controlled through an automated method. This illustration shows key steps in a sandwich assay and examples of assay formats.

Save time and reduce consumption

Working at nanoliter scale reduces overall processing time and minimizes consumption of sample and reagent. Precise control over assay workflow ensures reproducibility. Faster reactions eliminate lengthy incubation steps found in conventional assay formats.

Single and Gyroplex assays

Users have the option of running single or Gyroplex™ (multiple analyte) assays. Gyroplex assays enable, for example, up to five biomarkers to be monitored in a single run using Bioaffy 1000 CDs. Each CD can include a standard curve and duplicate samples for a specific biomarker. As little as 50 μL of sample (10 μL per duplicate) is consumed.

Bind capture reagent to affinity column

Prepare samples in Rexxip buffer

Design run in Gyrolab Control.

Evaluate data in Gyrolab Evaluator

Load sample

Load detection reagent

Read fluorescence

Generic PKSandwich

Indirect antibody Bridging assays

One microstructure – one data point – no cross-talk.

Common channel for assay reagents

Volume definition area (200 nL)

Affinity-capture column (15 nL)

Individual inlet for sample addition

Overflow channel for excess liquid – ensures reproducible filling of sample

Volume definition chamber – accurate sample volumes (20, 200 or 1000 nL)

Hydrophobic barriers stop liquid flow – ensures consistent volume delivery and facilitate parallel processing

Resp

onse

Concentration (ng/mL)

Spiked human serum, serial dilutions in Rexxip A 3 CD batches: 5 standard curves per CD, one run1 CD: x2 standards, 4 blanks, 5 x 4 QC samples

1 10 100 1000

1

10

100

Reproducibility and precision

Low inter-CD variation of assay on Gyrolab Bioaffy 200: in-house validation using a clinical PK assay of a commercially available biotherapeutic recombinant antibody.

Page 3: Gyrolab CDs - Future Science · olunsistov Cme nt e delivery and flow control -hydrophobic barriers stop liquid flow Overflow channels for excess liquid and reproducible volume definition

3D0012493/D

Gyrolab Mixing CD

ADA controls 1 2 3 4

Concentration (ng/mL) 1000 200 100 50

Intra-CD CV (%) Run 1 3.7 5.6 10.3 15.1

Intra-CD CV (%) Run 2 4.7 5.3 5.2 17.4

Intra-CD CV (%) Run 3 6.2 22.1 10.0 2.8

Inter-CD CV (5) 5.0 19.1 11.7 21.2

The data demonstrates high precision within and between runs for quality control samples in 10% human matrix.

Samples loaded into mixing chambers automatically

Prepare samples in Rexxip ADA

Evaluate data in Gyrolab

ADA Software

Acidic buffer added – mixed and incubated

Drug:ADA complexes dissociate

ADA complexed with free drug

Neutralization buffer and labeled reagents added - mixed and incubated

Labeled drug:ADA complexes form

Mixture passed through affinity –capture column

Fluorescence measured

Labeled drug:ADA complexes bind streptavidin-coated Dynospheres

Automated sample pretreatment and immunoassay at an integrated, nanoliter-scale

The illustration shows key steps in a bridging homogeneous assay format used to measure Anti-Drug Antibodies (ADA). Gyrolab Mixing CD enables 48 samples to be processed simultaneously in parallel, with every step controlled through an automated method.

The detection of ADAs is a prime example of integrating sample pretreatment into the nanoliter-scale immunoassay workflow of Gyrolab Mixing CD. Addition of a mixing chamber allows for greater flexibility in assay workflows requiring additional incubations for equilibration, acid treatment or enzymatic reactions to occur. These additional assay steps typically add complexity and time to existing assays, while the Gyrolab Mixing CD not only integrates sample pretreatment, but also shortens assay and incubation times with its nanoliter-scale workflows.

Compatible with many matrices

The microfluidic, flow-through format, coupled with short incubation times achieved by working at nanoliter scale, minimizes the risk of matrix interference.

Gyrolab CDs can be used with matrices such as human or animal serum, plasma, sputum, cerebrospinal fluid and cell culture supernatant.

Inlet for wash solutions or capture reagents for specific assay formats

Inlet for samples/reagents/buffers

Volume definition (200 nL)

Volume definition (200 nL)

Affinity-capture column (15 nL) packed with streptavidin-coated Dynospheres™

Consistent volume delivery and flow control -hydrophobic barriers stop liquid flow

Overflow channels for excess liquid and reproducible volume definition

Mixing chamber for incubation steps

One microstructure – one data point – no cross-talk.

PK assay in 5% and 50% human serum shows minimal matrix interference.

Resp

onse

Concentration (ng/mL)1 10 100 1000

1

10

100 5% serum50% serum

Reproducibility and precision

Parallel processing with precise control of mixing and incubation times ensures consistent assay performance.

Page 4: Gyrolab CDs - Future Science · olunsistov Cme nt e delivery and flow control -hydrophobic barriers stop liquid flow Overflow channels for excess liquid and reproducible volume definition

www.gyros.com

Gyros, Gyrolab, Bioaffy, Rexxip and Gyros logo are trademarks of Gyros Group. All other trademarks are the property of their respective owners. Products and technologies from Gyros are covered by one or more patents and/or proprietary intellectual property rights. All infringements are prohibited and will be prosecuted. Please contact Gyros AB for further details. Products are for research use only. Not for use in diagnostic procedures. © Gyros AB 2013 D

0012

493/

D

Technical information

Storage of Gyrolab CDs

Refrigerate at + 4 °C to +8°C, unopened package.

Shelf life (unopened package): Minimum 12 months after delivery.

Storage (opened package)*: CDs must be used within one week of opening. Return partially used CDs to original CD box and pouch.

Re-seal. Store dark, dry and at room temperature.* Partially-used CDs can be run again. To guarantee optimal performance, Gyros recommends using a new CD for each run.

Chemical compatibility of Gyrolab CDs

Streptavidin-coated particles and surfaces within CD microstructures stable in buffers pH 2-10.

Compatible with the following solutions: 0.2% Tween 20, 40% methanol, 30% dimethyl sulfoxide (DMSO), 30% 2-propanol, 0.01% Triton X-100, 0.1% SDS.

Bioaffy CD selection and specifications

Several parameters influence the choice of Bioaffy CD such as the required dynamic range, analyte concentration, quality of reagents and assay format. For most applications, the Gyrolab Bioaffy 200 CD is used as a starting point.

Ordering information

Product number Product name

P0004424 Bioaffy 20 HC

P0004180 Bioaffy 200

P0004253 Bioaffy 1000

P0020026 Gyrolab Mixing CD

Sample volume( nL) Microstructures Lower Limit of Quantification (LLOQ)

Column packing

Bioaffy 1000 1000 96 Up to 5 times lower LLOQ than Bioaffy 200

Streptavidin-coated Dynospheres ®

Bioaffy 200 200 112 Assay dependent Streptavidin-coated Dynospheres

Bioaffy 20 HC 20 112 Assay dependent Streptavidin-coated TSK-GEL ® particles

Gyrolab Mixing CD 200 48 Assay dependent Streptavidin-coated Dynospheres

Specifications relevant for CD selection. Running one assay on a single CD takes approximately 60 minutes.

Typical analyte concentration and dynamic range for Gyrolab Bioaffy CDs.

Concentration (M) 10-12 10-9 10-6

Bioaffy1000

Bioaffy200

Bioaffy20HC


Recommended