Hadamard Transform Imaging Paul Holcomb Tasha Nalywajko Melissa Walden
Transcript
Page 1: Hadamard Transform Imaging

Hadamard Transform Imaging

Paul Holcomb

Tasha Nalywajko

Melissa Walden

Page 2: Hadamard Transform Imaging

Problem Definition

• Current 3D imaging systems for brain surgery are too slow and possess too low of a resolution to be effective in an operating room setting

Page 3: Hadamard Transform Imaging

Why is this important?

• 71% mortality rate for diagnosed brain tumors

• Correlation between complete resectioning of tumors and improved prognosis

• Complete resectioning requires knowing the location of the tumor, especially tumor margins

• Imaging in a clinical setting should be fast• Operating room billed by the quarter- or

half hour

Page 5: Hadamard Transform Imaging

Design Criteria

• Must produce an image in real time

• Must accurately reproduce area of interest in the brain

• Must distinguish healthy versus tumor tissue

• Must be small enough to be usable in an operating room setting

• Must interface with operating microscope

Page 6: Hadamard Transform Imaging

Design Objective

Construct imaging system using digital micro-mirror device and Hadamard transform for use with operating microscope in a clinical setting

Page 7: Hadamard Transform Imaging

System Design

Hadamard Transform

• Decreased imaging time

• Increased SNR Hadamard Matrix Definition

Inverse Hadamard Transform

Digital Micro-mirror Device

• Allows use of Hadamard Transform

Page 8: Hadamard Transform Imaging

Fourier vs. Hadamard Imaging

Wuttig and Riesenburg, “Sensitive Hadamard Transform Imaging Spectrometer”

SNR Increase with Hadamard: √n

SNR Increase with S-Matrix: (√n)/2

Page 10: Hadamard Transform Imaging

System Diagram

Decrease image size to fit within 512 x 512 matrix

Magnification:~0.4

Page 11: Hadamard Transform Imaging

System Diagram

Apply Hadamard matrix using

DMD

1

1-1

-1

1

1

-1

-1

Compress image to 160um line

Page 12: Hadamard Transform Imaging

Disperse light spectrally using

spectrograph and collect image

using CCD camera

Apply inverse Hadamard

transform using computer

X

YSpectrum

System Diagram

Page 13: Hadamard Transform Imaging

System Output

Page 14: Hadamard Transform Imaging

Design Timeline

February: Align and test Stage 1; align DMD; align and test

Stage 2

March: Insert, align, and test spectrograph; test system

using reflectance standard to determine SNR; test

system using normal and tumor tissue samples

April: Continue testing and analysis; compile and present

findings at Senior Design Day