Date post: | 16-Dec-2015 |
Category: |
Documents |
Upload: | benjamin-doyle |
View: | 215 times |
Download: | 0 times |
Halo, SLAC, 2014
Towards wire scanner measurements with Large Dynamic Range (> 106)
Pavel Evtushenko,Jefferson Lab
Halo, SLAC, 2014
Outline
Motivation: Why large dynamic range diagnostics?
Experience with existing high average current FEL driver
Large dynamic range transverse beam profile measurements
Wire scanner measurements
experience so far (CEBAF) counting
PMT in analog mode
Signal generation
Halo, SLAC, 2014
Motivation: Why Large Dynamic Range?
there are several applications of electron LINACs under consideration / design that require average beam powerof several MW
these applications also require very high peak beam brightness, comparable to the one at pulse NC LINACs
similar to low average current (NC, pulses) LINACs, with high average current LINAC a diagnostic beam mode must be used
the significant difference is the ratio of beam currents in the diagnostic mode and full current mode
for a high average current LINAC this ration can easily be tens of thousands
One example of an electron LINAC, which have operated with high average current 9 mA, while driving FEL (also high average power) JLab IR/UV Upgrade FEL.
Halo, SLAC, 2014
JLab IR/UV Upgrade: 1.2 MW beam power
Ebeam 135 MeVaverage current 9 mA(135 pC at 74.85 MHz)
Average beam power ~ 1.2 MW !
If lost beam average <P>=1 W possible problem for vacuum concern for the FEL
undulator livetime
25 μJ/pulse in 250–700 nm UV 120 μJ/pulse in 1-10 μm IR
Halo, SLAC, 2014
Lessons from high current FEL operation
when setting this machine up for high current operation, at fist diagnostic beam mode is used, this gives “best” RMS setup, i.e., the setup which optimizes FEL performance and does not show any measurable beam lose (at that current level)
then as average beam current is increased we always found that there is a need to alter transverse match to further reduce beam loss to allow higher current operation
important point is that, such adjustments of the transverse match must be small
there are very small fractions of the beam, which could prevent high current operation, but are not measured when diagnostic beam mode is used
it also appears that, such small fractions of the beam have different Twiss parameters than the core of the beam, i.e., transverse phase space is not described well by a single set of Twiss parameters
Halo, SLAC, 2014
Beam dynamics driven halo generation
Measured: JLab FEL injector, intensity difference of the peak and “halo” is about 300.(YAG:Ce, standard CCD - 57 dB SNR10-bit frame grabber)
Simulations: PARMELA, 3×105 particles; X and Ybeam profile and its projection show the halo aroundthe core of about 3×10-3.Even in idealized system non-linear beamdynamics can lead to formation of halo.
Halo, SLAC, 2014
LINAC’s non equilibrium (non Gaussian) beam
Propagating in drift space …
FODO matching section
This are not beam distributions from a nominal setup, but an experiment that shows complexity of the phase space distribution – no single set of Twiss parameters describes the beam
This is also not a halo. Dynamic range of this measurements is ~ 500, all of this beam later is matched to the FEL’s optical cavity and participates in the FEL interaction
Halo, SLAC, 2014
Beam viewer wire-scanner combination
Must have impedance shield, due to high average I
Two diagnostics at one location
Can use YAG:Ce or OTR viewer with easy switch
Shielded, 3 position viewer design for FEL
Halo, SLAC, 2014
Wire scanner measurements: counting
CEBAF uses wires scanners for transverse beam profile measurements
499 MHz repetition – very good for counting
One of a very few LDR beam profile measurements examples
Due to very low current (5 nA) made with CW beam
Max. counting frequency ~ 10 MHz (not a dedicated hardware)
Coincidence effective to reduce background, but at the expanse of even longer measurements time
With CW beam measurements time of about 15 min.
for non-Gaussian beams
A. Freyberger, in DIPAC05 proceedings,Measurements made at CEBAF
Halo, SLAC, 2014
PMT current range
Counting can provide LDR, but is really practical only with high (~ 100 MHz) bunch frequency
For smaller bunch frequencies alternative is analog mode - PMT current measurements
Typically average PMT current must be ≤ 100 µA
With low duty cycle beam (100 µs @ 60 Hz) PMT current within the 100 µs can be much higher
PMTs with dark current of a few nA are available (low Q.E. cathode at long wavelength)
For low duty cycle systems like diagnostic mode beam, gated integrator (GI) is a for small signal recovery
For a single GI dynamic range of 107 is very challenging and probably impossible (sub µV noise for 10 V signals)
Halo, SLAC, 2014
Gated Integrator (GI)
Digitize
Integrate Discharge ReadyReady
PMTs with HV at the cathode and anode at the ground potential are used – this results in negative current, which needs to be inverted
A current mirror is used to 1. invert the current and 2. to make multiple “copies” of the PMT current
Two outputs of the current mirror:#1 ~ 100 % of PMT current, #2 ~ 1 % of PMT current
Halo, SLAC, 2014
GI calibration with precision source
Preparing version two with FET transistor based current mirror
The non linearity by itself is not a really a problem if the behavior is reproducible
Calibration is to be used as a look up table
Output of each GI is digitized with 16-bit ADC at 4 MS/s
Output of a GI is available for digitalization during charge integration as well – better than the gate width time resolution
Results of GI calibration with a precision DC current source (Keithley 6221) in the range from 100 pA through 10 mA are shown
RMS noise level ~ 250 µV
Non linearity of the 1 % channel (red) is du to nonlinear operation of the current mirror, too little current for bipolar transistor
Halo, SLAC, 2014
Calibration cross-check
Halo, SLAC, 2014
GI stability
the 1 % variation is attributed to the source stability
GI stability is ~ 10 times better (0.1 %)
Halo, SLAC, 2014
GI + PMT test
PMT driven by a pulsed LED, 100 us “macro pulse”
LED is driven by pulse generator at fixed micro pulse rep. rate of 100 MHz
Width of the LED pulse adjusted from 620 ps down to 380 ps to generate the plot
Halo, SLAC, 2014
PMT dark current measurements
~ 2 nA dark current is at the level of PMT specification (3 nA typical 20 nA max)
measurements with two gates allows to subtract the dark current
then limiting factor is the GI intrinsic noise level – equivalent to ~ 100 pA RMS
Halo, SLAC, 2014
Wire Scanner: analog mode
an alternative to GIs are commercially available Logarithmic Converters
Originally designed for photo diode measurements(fiber optics communications)
Dynamic range of 160 dB and 200 dB
Bandwidth of several MHz but varies dependent of signal level
Shows more complex than GI noise behavior, which needs to be studied further
Calibration of AD8304 log-amp is shows
The calibration was made using the same setup – DC current source and ADC as used for GI evaluation and testing
4 calibration without and 4 with a CM are shown
Halo, SLAC, 2014
Wire Scanner / Cherenkov converter one way to convert E-M shower
e- and e+ to visible photons
“prompt” – much faster than a fast PMT with few ns pulse length
direction sensitive – to reduce background, i.e., insensitive to particles coming from “wrong” direction
all reflective optics – to use wavelength as short as possible (3 reflectors)
output matched to a quartz fiber to transport light to a PMT outside of the accelerator tunnel (background reduction)
thicker converter generated more photons, but limited by multiple Coulomb scattering – beam energy dependent
H20 n=1.333 > sqrt(2); Cherenkov radiation is not trapped in the radiator
Cylindricalreflector #1
Conereflector #2
90˚ off axisparabolicreflector #2
Optical fiber input
Cherenkov radiator
Halo, SLAC, 2014
W-S signal via Cherenkov converter
How many photons would Cherenkov radiator make?
• - 3 mm stainless steel wall;- 50 µm W radiator;- 200 nm – 650 nm wavelength range;- 200 pC;- 50 mm diameter, 125 µm thick radiator at 0.1 rad relative to the beam direction
~ 1.1×105 photons
Halo, SLAC, 2014
That is all folks. Thank you.
Halo, SLAC, 2014
back up
Halo, SLAC, 2014
FEL Injector as an example of #1 (1/6)
downstream ofthe gun
Halo, SLAC, 2014
FEL Injector as an example of #1 (2/6)
upstream of thebuncher cavity
Halo, SLAC, 2014
FEL Injector as an example of #1 (3/6)
downstream of thebuncher cavity
Halo, SLAC, 2014
FEL Injector as an example of #1 (4/6)
upstream of theSRF cavity 1
Halo, SLAC, 2014
FEL Injector as an example of #1 (5/6)
downstream of theSRF cavity 1
Halo, SLAC, 2014
FEL Injector as an example of #1 (6/6)
downstream of theSRF cavity 2