Hd f h Hadrons for cancer therapy at CNAO CNAO Marco Pullia CNAO Foundation CNAO Foundation
Transcript
Page 1: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

H d f hHadrons for cancer therapy at CNAOCNAOMarcoPullia

CNAO FoundationCNAOFoundation

Page 2: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Tumours and radiotherapypy

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 3: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Physics and medicine together since long:diagnosis and therapydiagnosis and therapy

1895

X ray discoveryWilhelm Conrad Röntgen

(1845 – 1923)

X ray discovery

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

( )

(courtesy of U. Amaldi)

Page 4: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Tumours

• Errors incell DNAandnoapoptosisp p• They grow inan uncontrolled wayTh i fil h di i d• They infiltratethesurrounding tissues andcanoriginatemetastasis (malignant)

• When metastatic,only chemotherapy ispossiblepossible

• If localised,surgery orradiotherapy

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 5: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Energy and Efficacygy y

Administered dose1Gy =1J/1Kg(typical dose in radiotherapy 35 X 2 Gy)(typical doseinradiotherapy 35X2Gy)

Howmany cells do I kill?Howmany cells do Ikill?Potential energy (1mfall =10Gy)

(f 5 )Heat (fever 38° =4185Gy)Ionizing radiation (little energy,many damages)

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 6: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Radiation damageg

• Ionization breaks chemical bonds• Freeradicals creation (mainlyhydroxylradical,OH−, and superoxide, O2− . Poison for the cell!)OH ,andsuperoxide,O2 .Poison for thecell!)

• Thetargetis DNA,ionization distribution isrelevantrelevant

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 7: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Cancer therapy100 % of cancer patients

py

Cured Not cured45 % 55 %

Local treatments(surgery, radiotherapy)

40 %

Other treatments(chemotherapy, etc)

5 %40 % 5 %

RT i 40%Surgeryalone

Byradiotherapy

RT in 40% OF CURED

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

22 % 18 % PATIENTS

Page 8: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

General principle of radiation therapy

100

p p py

80

Tumour control

60

ity [%

]

40

prob

abili

20

00 20 40 60 80 100 120

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

dose [Gy]

Page 9: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

General principle of radiation therapy

100

p p py

80

Tumour control

C li i

60

ity [%

]

Complication rate

40

prob

abili

20

00 20 40 60 80 100 120

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

dose [Gy]

Page 10: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

General principle of radiation therapy

100

p p py

80

Tumour control

C li i

60

ity [%

]

Complication rate

40

prob

abili

T t l20

Tumour control without complications

00 20 40 60 80 100 120

p

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

dose [Gy]

Page 11: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

General principle of radiation therapy

100

p p py

80

Tumour control

C li i60

ity [%

]

Complication rate

40

prob

abili

Tumour control

20without complications

00 20 40 60 80 100 120

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

dose [Gy]

Page 12: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Hadron RT proposed by Wilson in 1946p p y

Fi t h d th i th i ti i US (P t )

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

First hadron therapy in the sixties in US (Protons)

Page 13: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Comparison of the depth dose profiles

110 Protons/CarbonInverse dose profileInverse dose profile

p p p

90

100C

dE K( )dx

k z2ZA

1

K( )2 ln

2 511 K( )2 K( )2

I

K( )2

dE K( )

dx

60

70

80

%]

Ph t xII

K( )

40

50

60

Dos

e [% Electrons Photons xeII

0

20

30

40

Tumour

0

10

0 50 100 150 200 250 300

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

0 50 100 150 200 250 300Depth [mm]

Page 14: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Longitudinal ‐ Spread Out Bragg Peak1.2

g p gg

1

0.8

ive

units

]

0.4

0.6

Dos

e [r

elat

i

0.2

D

00 2 4 6 8 10 12 14 16 18 20

Depth [cm]

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Depth [cm]

Page 15: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Macroscopic advantage of hadronsp g

Photons Protons

X rays protons

Rapid fall-off

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 16: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Better dose distribution

9 X beams 1 proton beam9 X beams 1 proton beam

tumor between eyes

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 17: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Lateral radii of elementary beams of electrons and light ions (range of 26 cm) as a function of depth in water

Carbon scattersLess than protonsLess than protons

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014(A. Brahme, S.Rossi et al., NIM B 184 (2001) 569-588)

Page 18: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Radiobiological advantage of Cg g

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Schardt & Elsasser, 2010

Page 19: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Warning: RBE depends on

• Biological endpoint

LET• LET

• Particle typeParticle type

• Cell/tissue

• Dose rate

F i i• Fractionation

• etc…

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 20: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Different types of radiationsyp

Distribution of dose andDistribution of dose and of damage (yellow) on the cell nucleus scale (microns) for photons(microns) for photonsand carbon ions

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

(from G. Kraft, Tumor therapy with heavy ions)

Page 21: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Microscopic advantage of C ionsp g

RBE4

3

2

1 10 100 LET

2

1

10 – 20 keV/m =

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

100 – 200 MeV/cm =

20 – 40 eV/(2 nm)

Page 22: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

The The optimaloptimal LETLET

Diagram illustrating whyradiation with a LET of 100 keV/μm has the greatest RBE for cell killing, mutagenesis, or oncogenic trasformationoncogenic trasformation.

Fot this LET, the averageseparation between ionizingevents coincides with the diameter of the DNA doublehelix (i.e. about 2 nm).

Radiation of this quality is mostlikely to produce a doublestrand break from one track fora given absorbed dose.

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014Radiobiology for the radiologist, 7th edition

Page 23: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

3 different cases

‐1LowLET(<20keV/micron)DistancebetweenionizationslargerthanDNAdi t Cl i l di th F ti tidiameter.Classicalradiotherapy;Fractionationveryimportant.

‐2HighLET(50– 200keV/micron)DistancebetweenionizationscomparablewithDNAdiameter.C‐iontherapy;Fractionationlessimportant.

3 Very high LET(> 1000 keV/micron)‐3VeryhighLET(>1000keV/micron)DistancebetweenionizationssmallerthanDNAdiameter; energy in excess in ionizations (overkill).

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

diameter;energyinexcessinionizations(overkill).

Page 24: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Physical and biological dosey g

C li t d t t t l iComplicated treatment planning

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 25: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Beam Deliveryy

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 26: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Beam delivery: passive systemsy p y

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 27: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Passive systems for Carbony

Completely passive system not advisable:

-Smaller scattering implies larger thicknesses and distancesg p gand thus larger energy loss and beam loss which implies largerenergy and current from the accelerator

-Fragmentation of impinging ions which causes more dosedelivered after the tumor and larger production of neutrons.

-The amount of material in the beam line is considerable,leading to an increase in nuclear fragments produced bynuclear interactions with the material of the beam modifiersnuclear interactions with the material of the beam modifiers.These nuclear fragments have lower energies and lead to ahigher LET and thus an increased biological effective dose of

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

the beam already in the entrance region.

Page 28: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Wobblingg

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 29: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Layer stackingy g

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 30: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Active systemsy

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 31: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Scanning Beamg

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

(Found on the web, forgot where... presumably Siemens or HIT)

Page 32: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Active systemsy

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

(Courtesy of E. Pedroni)

Page 33: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Beam position precisionp p

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 34: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

2D

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 35: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Beam position requirementp q

Gaussian beam, FWHM = 10 mm

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014Beam position error ~ 0.1 mm

(M. Donetti)

Page 36: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Beam position errorsp

Tumore   Area irraggiata

Long and medium term stability

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

(large slices, breath synchronization)

Page 37: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

The CNAO Foundation

No profit organisation (Foundation) createdwith the financial law 2001 to build theN ti l C t f H d th d i dNational Center for Hadrontherapy designedby TERA Foundation

Founders:Fondazione Policlinico Ospedale Maggiore- Milano

Institutional Participants:Istituto Nazionale di Fisica Nucleare

Fondazione Istituto Neurologico C. Besta - MilanoFondazione Istituto Nazionale dei Tumori - MilanoIstituto Europeo di Oncologia - MilanoFondazione Policlinico San Matteo - Pavia

Università di MilanoPolitecnico di MilanoUniversità di PaviaComune di PaviaFondazione Policlinico San Matteo - Pavia

Fondazione TERA - NovaraComune di Pavia

Participants:Fondazione Cariplo

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Fondazione Cariplo

Page 38: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

National collaborations

TERA Foundation: final design and high tech specificationsINFN: co-direction HT, technical issues, radiobiology, research, formationUniversity of Milan: medical coordination and formationUniversity of Pavia: technical issues, radiobiology, formationUniversity of Catania: medical physicsUniversity of Catania: medical physicsUniversity of Florence: medical physicsUniversity of Turin: interface beam-patient, TPSPolytechnic of Milan: patient positioning, radioprotection, authorisationsEuropean Institute of Oncology: medical activities, authorisationsS M tt F d ti di l ti iti l i tiSan Matteo Foundation: medical activities, logisticsTown of Pavia: land and authorisationsProvince of Pavia: logistics and authorisation

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

g

Page 39: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

International collaborations

CERN (Geneva): technical issues, PIMMS heritage

GSI (Darmstadt): linac and special components

LPSC (Grenoble): optics, betatron, low-level RF, control system

Med-Austron (Vienna): technical collaboration for MA centre

Roffo Institute (Buenos Aires): medical and research activities

NIRS (Chiba): medical activities, radiobiology, formation

HIT (Heidelberg): research activities

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 40: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 201440

Page 41: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Origins ‐ Historyg y

1990 – U. Amaldi and G. Tosi have the idea of promoting hadrontherapy in Italy

1991 – U. Amaldi and G. Tosi, “Per un centro di teleterapia con adroni”

1991 – ATER experiment at INFN

1992 – TERA Foundation is founded

1996 – PIMMS starts (TERA+CERN+MedAustron+Onkologie2000+GSI)

2000 – 2001 the CNAO foundation is created within the Financial Law

2003 – CNAO gets the project and hires the design group

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 42: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

The CNAO PhasesC O

Phase 0: organisation Years: 2002 - 2004

Phase 1: construction Years: 2005 - 2009

Ph 2 i t tiPhase 2: experimentation Years: 2010 - 2013

Phase 3: running Years : 2014 …

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 43: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

The CNAO accelerator and lines

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 44: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Aim of the center

AIM OF THE PROJECT

To treat deep tumours :

•With ion beams in the range 1 ≤ Z ≤ 6

•With active scanning

•In approximately 3 min/fieldIn approximately 3 min/field

•Dose uniformity ±2.5%

Synchrotron with slow extraction!

Everything safe, proven and/or redundant

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 45: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Design Parameters Ig

Protons (1010/spill)LEBT (*) MEBT SYNC HEBTLEBT (*) MEBT SYNC HEBT

Energy [MeV/u] 0.008 7 7-250 60-250

Imax [A] 1.3×10-3 (0.65, 0.45) 0.7×10-3 5×10-3 7×10-9[ ] ( , ) 7 10

Imin [A] 1.3×10-3 (0.65, 0.45) 70×10-6 0.12×10-3 17×10-12

rms,geo [ mm mrad] 45 1.9 0.67-4.2 0.67-1.43(V)

90,geo [ mm mrad] 180 9.4 3.34-21.2 3.34-7.14 (V)5.0 (H)

Magnetic rigidity [T m] 0.013 (0.026) 0.38 0.38-2.43 0.38-2.43

(p/p)tot ±1.0‰ ±(1.2-2.2)‰ ±(1.2-3.4)‰ ±(0.4-0.6)‰

* (H2+, H3

+)

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 46: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Design Parameters IIg

Carbon (4·108 C/spill)LEBT (C4+) MEBT SYNC HEBTLEBT (C4+) MEBT SYNC HEBT

Energy [MeV/u] 0.008 7 7-400 120-400

Imax [A] 0.15×10-3 0.15×10-3 1.5×10-3 2×10-9[ ] 2 10

Imin [A] 0.15×10-3 15×10-6 28×10-6 4×10-12

rms,geo [ mm mrad] 45 1.9 0.73-6.1 0.73-1.43(V)

90,geo [ mm mrad] 180 9.4 3.66-30.4 3.66-7.14 (V)5.0 (H)

Magnetic rigidity [T m] 0.039 0.76 0.76-6.34 3.25-6.34

(p/p)tot ±1.0‰ ±(1.2-2.0)‰ ±(1.2-2.9)‰ ±(0.4-0.6)‰

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 47: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Facciamo un giro della facilityg y

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 48: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Sources and LEBT

0.008 MeV/u H3+

0.008 MeV/u C4+

I ~ 0.5 mA (H3+)

I ~ 0.2 mA (C4+)

Two ECR sources

Continuous beam

LEBT Chopper

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 49: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

LINAC systemy

217 MHz

RFQ0.008‐0.4 MeV/u H3

+

0.008‐0.4 MeV/u C4+

IH0.4‐7 MeV/u H3

+0.4 7 MeV/u H30.4‐7 MeV/u C4+

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 50: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

CNAO RFQ

St tt i tStruttura internaIngresso ioni

F d lik t217 MHz

Four‐rod like typeEnergyrange =8– 400keV/uElectrode length =1.35m,l d l kElectrode voltage =70kVRFpower loss(pulse):about 100kWLowdutycycle:around 0.1%

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

50Uscita ioni

Page 51: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

LINAC

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 52: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

MEBT Layouty

7 MeV p7 MeV/u C6+

I ~ 0.75 mA (p)I ~ 0.12 mA (C6+)

Current selection

Stripping foil

Debuncher

Emittance dilution

(x x’)I j

Match betas

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

(x,x )Inj

Page 53: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Intensity degradery g

4 transmission levels: 100%, 50%, 20%, 10%Keep overall emittance unchanged

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

p g

Page 54: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Multiturn injectionu t tu ject o

SeptumIncoming beam The bumper creates an orbit

Septum bump collapsing in tens of turns

Closed orbit

Bumper Bumper

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 55: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Synchrotrony

7‐250 MeV p7‐400 MeV/u C

I ~ 0.1‐5 mA (p)I ~ 0 03 1 5 mA (C)

Slow extraction

I ~ 0.03‐1.5 mA (C)

Betatron core

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 56: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Resonance

RF cavity

sextupoleD = D’ ≠ 0

BetatronD = D’ = 0 Demanding requirements

on magnet power supplies

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

P inj P – 60 MeV P – 250 MeV C6 inj C6+ – 120 MeV C6+- 400 MeV

B(T m) 0.4 1.1 2.4 0.8 3.3 6.4

Page 57: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Machine Cycley57

Hysteresis cycleBmax

y y

BBextr

Bi jBminBinj

per o

nrn

ping

ratio

n

p

etat

ron

etat

ron

yste

rest

is

cle

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Inj b

umM

ultit

ur

Rf t

rapp

Acc

eler

Rf j

ump

Rf o

ffS

tart

b e

Sto

p be

Sta

rt hy

End

cyc

(Courtesy of C. Biscari)

Page 58: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Treatment execution

5 s Sync1 s

5 i E t ti li5 min Extraction lines

Injection lines 10 s

Extracted beam

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 59: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Extraction possibilities at CNAO

Betatron core

p

Empty bucket

Betatron core

Air core quadrupole

Empty bucket

RF-KO with Schottky Pick-up

Beam shaping with Schottky PU

Additional quad winding

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 60: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Betatron core

Pushes the beami t thagainst the

resonance

= 2.46 Wb

Magnetic screenneededneeded

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 61: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Empty bucketp y

E

Amplitude

Resonance linefor low betatronamplitudes

Amplitude

Resonanceregion

Resonance linefor high betatronamplitudesSense of stack

acceleration

PHASE

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 62: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Air core quadrupoleq p

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 63: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Ripple compensationpp p

I t ti ti 100 (10 kH d t )

100 ms

Integration time 100 us (10 kHz data)

No compensationEmpty bucket+HFRI

Empty bucket Empty bucket + ACQ feedback

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 64: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Extracted beam

Twiss functions at entry (ES in ring)Twiss functions at entry (ES in ring)

x = 5 m x = 0 ‘Free’ parameter.

Ex = 5 mm mrad ‘Unfilled’ ellipse - ‘free’.

z = 7.16 m z = -0.18 Values from ring.

Ez,RMS = 0.7324 to 1.4286 mm mradEz RMS = 0.6679 to 1.4286 mm mrad

Carbon range from ring.Proton range from ring.

0.0008

0.0009

z,RMS g g

Dx = 2.095 m Dx = -0.0393 Determined by extraction.

Dz = 0 Dz = 0

0.0004

0.0005

0.0006

0.0007 Twiss functions at exit (all beam exits)

x = 7.2 m x = 0

According to medical 2 t 27 0

0

0.0001

0.0002

0.0003

0.037 0.039 0.041 0.043 0.045 0.047 0.049 0.051

specifications and earlierchoice of ‘free’parameters.

z = 2 to 27 m z = 0

Dx = 0 Dx = 0

Dz = 0 Dz = 0

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

z z

Page 65: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Beam shapep

Vertical distribution: bell shape/gaussian like

H i l di ib i b f hHorizontal distribution: bar of charge

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

At extraction septum (x y) In the line

Page 66: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Beam at HEBT entrance

Bar of charge

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 67: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Chopperpp

Fast turn on/off for the beam

Intrinsically safeIntrinsically safe

Allows beam qualification

The chopper stops (and starts)The chopper stops (and starts) the beam within 200 s.

U d f i di ti f t d t

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Used for irradiation of separated parts of slices and for synchronization withbreathing.

Page 68: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Chopped beampp

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 69: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Beam position at HEBT endp

B iti t bilit ( t th ) 0 2Beam position repeatability (at the same energy): 0.2 mmBeam position precision (at different energies): 0.3 mm

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 70: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Beam size at isocenter

Protons

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 71: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Dose delivery systemy y

Box 1 Box 21 Integral chamber:• Beam Intensity measure

every 1 s

1 Integral chamber:• Beam Intensity measure every

1 s

2 Strip chambers (X and Y):• Beam position measure every

1 Pixel chamber:• Beam position and dimension

every 1 s 1 s

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

100 s, with 100 m of precision p

measure every 100 s/1 ms, with 200 m of precision

Page 72: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Dose deliveryy

Giordanengo et al.,NIM A613

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 73: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

First scanningsg

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 74: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Artistic use of the beam

Radiochromic film

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014Maggio 2011

Page 75: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Patient Positioning and Verification strategy at CNAOIntegrated robotic, X‐ray and IR localization system

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 76: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Beam measurements

Depth Dose Distributions (mono-en. pencil beams)

3-D motorized water ph.Peakfinder water column

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

p

Page 77: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Different fall-offMeasured Bragg Peaks

Different fall-off

Protons

Carbonionsions

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

12/04/12C. Biscari - LNF, INFN77

Page 78: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

In vitro measurements

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 79: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Mice crypt survival assayRisultati novembre 2010

• 2 beam time sessions

• 3 points in the SOBP

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

• 6 dose levels, 4 mice per position

(Courtesy of B. Vischioni)

Page 80: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Start of medical activitiesFirstpatient with Protonbeam

( b )(September 2011)

First patient with Carbon beam

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Firstpatient with Carbon beam(November 2012)

Page 81: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Patients treated

28 open protocols28openprotocolsMainlytumorsintheheadandneckorsacral

iregionRecentlyadded:prostate,liverandpancreas

344 (246C + 98p) patients treated + 28 under344(246C+98p)patientstreated+28undertreatment

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 82: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

www.cnao.it

CE LabelCE Label

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 83: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Future and R&D

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 84: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Future developmentsp

• Coping with tumor motion• Coping with tumor motion

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 85: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

On‐line imaging“Minimal” choice: breathing synchronisation(already applied in Chiba HIT and CNAO)

External surrogates with correlation models

breathingbreathing

(already applied in Chiba, HIT and CNAO) correlation models

X-rays

Ultrasound, MRI

breathingtrigger

extractiontrigger

breathingtrigger

extractiontrigger

,

Particle radiography

Bmax

B1B2

extrE 1

extrE 2

extrE 3

B3

trigger

Bmax

B1B2

extrE 1

extrE 2

extrE 3

B3

trigger

time

E 1

< 1 s < 0.2 s

E 2 E 3

time

E 1

< 1 s < 0.2 s< 1 s < 0.2 s

E 2 E 3

Interesting also for IMRT: lots of efforts and devices

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014(Review in Riboldi et al, Lancet Oncology 2012) (Courtesy of Medical Intelligence)

Page 86: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Tumour tracking

p+1 or C+6

GSI approach

Energy variation

Transversevariation

4D

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 87: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Future developmentsp

• Real time dose visualization• Real time dosevisualization

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 88: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Dose visualisation: “in beam PET”Pre-collision Post-collision

Projectile fragmentProjectile

Atomic nucleus

12C 11C16O 15O Neutron

Target fragmentAtomic nucleus

of tissueO O

[kG

y]

6 Dose

Dos

e

0

2

4

[10

]

5

0

+Activity

Cou

nts

3 0

0 50 100 150 200Depth [mm]

ISSUES: low statistics;

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Courtesy of GSIISSUES: low statistics;

blood flow dilution; off-line PET logistics

Page 89: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Secondaries emission and reconstructionP t R R di h (PRR)Electronic telescope for the measure of position and residual range of protons; it gives the density map of the traversed volumes; it permits to check in real time the treatment planning assumptions on position and dimensions of the traversed tissues and organs

Proton Range Radiography (PRR)

dimensions of the traversed tissues and organs. Nuclear Scattering Tomography (NST)

Three-dimensional map of the tissues densities obtained by vertex reconstruction of high energy

Interaction Vertex

reconstruction of high energy protons interactions (> 600 MeV).

Interaction Vertex Imaging (IVI)

Density of interactionDensity of interaction vertex reconstruction gives information on the Bragg peak position.

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014(U. Amaldi et al.) PROMPT radiation (Gamma) - Enlight

Page 90: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Future developmentsp

• Treatment Planning System (TPS) improvement• TreatmentPlanningSystem(TPS)improvement– Radiobiology measurement andmodelsS d l l ti ( d ti t t t)– Speed upcalculation (adaptive treatment)

– Self contouringR l ti i i d l l ti– Real time imaging andcalculation

• Improve densitymeasurement inimaging• Biomarkers

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 91: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Future developmentsp

• Proton centers are already commercial• Protoncenters arealready commercialproducts (tens worldwide);Carbon ion centersnot yet really (only 7 worldwide)not yet really (only 7worldwide).

• Cost reduction for treatmentdiffusion

• Singleroom facilitiesg• Next generationofaccelerators• Carbon Ion Gantries• Carbon Ion Gantries

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 92: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

MEVION S250

Superconducting SC

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Superconducting SCDiameter 1.8 m December 19th, 2013-First treatment at

S. Lee Kling Center for Proton Therapy at the Siteman Cancer

Page 93: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Dielectric Wall Accelerator (DWA)( )

Pulsed High-Voltage accelerators (G. Caporaso et al)

built in collaboration with Tomotherapy – Madison (T Mackie)

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

built in collaboration with Tomotherapy – Madison (T. Mackie)

Far into the future

Page 94: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Single room facility by IBANew superconducting

synchrocyclotron

rotation 220°

PROTEUS ONE

30 m 

PROTEUS ONE

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 95: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

TULIP

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

(Courtesy of U. Amaldi)

Page 96: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

The only ion therapy cyclotrony py y

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 97: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Rapid cycling synchrotron (first publication 1999’s, S. Peggs et al.)

Injection linac at 8 MeV/u

Racetrack, FODO in the arcs, D=0 ss

Fast inj+extr, C = 60 m

(from D. Trbojevic et al. IPAC2011)

30 Hz repetition rate (repainting?)

Fast energy changegy g

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 98: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

150 MeV/uTERA cyclinac for C-ions

CABOTO

Linac for Image Guided Hadron THerapy  LIGHT  150‐400 MeV/u

CABOTO  = CArbon BOoster for 

Therapy in OncologyOncology

400 MeV/u

Source EBIS‐ SCCyclotron K 600 ‐ SC 300 Hzy K 600 SC

200tonsLinac [email protected]

16modules

300 Hz

RFpowersystem

16Klystrons(Ppeak =12MW)

Energy is adjusted in 2 ms in the full range by changing the power pulses sent to the accelerating

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Energy is adjusted in 2 ms in the full range by changing the power pulses sent to the accelerating modules

Charge in the spot is adjusted every 2 ms with the computer controlled source

Page 99: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Laser + linac

5 106 t 60 M V @ 10 H5 106 p at 60 MeV @ 10 Hz

Fuchs, Antici et Al, Proc HB2006 Review of proton beams 2006

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Rossi F., Londrillo P., Sinigardi S., Turchetti G., Giove D., De Martinis C.; questa conferenza etPRSTAB 16, 031301 (2013)

Page 100: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Dual‐stage ion FFAG proton FFAG with pCT

1st stage 1st stage: Cyclotron or FFAG

proton FFAG with pCT

1 stage 18– ~250‐330MeVH‐

Fixedorswept‐frequencyRF,DCbeam LowintensityforpCT

g y

Strippingcontrolsextractionenergyandintensityinadditiontosourcemodulation

OR 9‐~70‐90MeVchargetomassratioof½

~6 m

Fixed‐frequencyRF,DCbeamforallions Variableenergyextraction Upstreaminjectorforhigh‐energyring 2nd stage: 70/90 – 430 MeV/nucleon ions

2nd stage(~4mx5‐6mlong) 70/90MeV– 430MeV/nucleon Variableenergyextraction

Variable energy selection:Injection/extraction straight

Adjustable,fastorbitbumpmagnets/extractionseptuminlongstraight DCextractedbeam Variableenergyonscaleoftensofmicroseconds

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Investigatingextractedenergyrange

(Courtesy of C. Johnstone)

Page 101: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Gantries

Conventional RTCarbon Ion GantryB < 6.4 Tm

Proton GantryB < 2 4 TmB < 2.4 Tm

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 102: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Future gantriesg

• Superconducting magnets• Superconducting magnets• FFAG• Mobileisocenter

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 103: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

NCI‐DOE Ion Therapy workshop, Jan 2013py p

• Further R&D and requirements for future• FurtherR&Dandrequirementsforfuturemachines

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Page 104: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Requirements: next‐generation ion therapy*

Multi‐ioncapabilityR d d H Li B C O NRecommended:p,He,Li,B,C,O,NeEssential:p,He,Li,B,C1‐ 30cmfortreatment

60MeV/nucleon– 430MeV/nucleon(forcarbon)TreatmentOptions‐Vary single treatment parameter (e g low vs high LET) in ns

tone

)

Varysingletreatmentparameter(e.g.,lowvs highLET)inclinicaltrialsMulti‐iontreatmentoptionincludingwithinasinglefraction

B f l d h h h d h GTV yof

C. J

ohn

BetterconformaldosewithhighdosetohypoxicGTVAvoiddosetonormaltissuefromfragmentationtail

Hypofractionation withhigherRBEions (Cou

rtesy

yp gImaging:Automaticallyintegrated(20‐ 60cmavailableforimaging†)F ll f i i t h l i i ti i h t f iliti

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Fullscopeofimagingtechnologiesexistinginphotonfacilities*from final report of the joint NCI-DOE Ion Therapy workshop, Jan, 2013†imaging with carbon will be limited to 20 -30 cm

Page 105: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

Requirements: next‐generation ion therapy*

Treatment Monitoring and Adaptation

Requirements: next generation ion therapy

TreatmentMonitoringandAdaptationTargetingandImageGuidance

Withimaging,allmotionmanagementcapabilitiesavailableinphotonfacilitiesincludinggatedbeamdelivery

Pre‐ andintra‐treatmentverificationwithparticlebeamCTandradiography ns

tone

)

g p y Pre‐treatment3Dtargetpositionandrangeverification Simultaneous“real‐time”radiographictargetpositionandintegratedrangeverificationduringtreatment y

of C

. Joh

n

g gPost‐treatmentverificationofdelivereddosewithparticlebeamCT(patientposition)andwithPET(doseconfirmation)

Adaptive Therapy

(Cou

rtesy

AdaptiveTherapyLow‐doseparticle‐beamCTallowsunlimitedscans

Planmodificationusingpre‐treatmentparticle‐beamCT

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Planmodificationusingpost‐treatmentCTorPETimaging

*from final report of the joint NCI-DOE Ion Therapy workshop, Jan, 2013

Page 106: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

R i t t ti i th *Requirements: next‐generation ion therapy* DoseDeliveryRateforTreatment

20 Gy/min/liter has been defined as the minimum “standard” for the ion 20Gy/min/literhasbeendefinedastheminimum standard fortheionaccelerator* Twofields(representdifferenttechnicalspecificationsforbeam):

30cmx30cm(singlelayer field) 10 x 10 x 10 cm3 10x10x10cm3

Requires~40energystepstoevenlycoverindepth;(assumes0.25cm/layer,~2MeVenergystep)

1Gy/sec/liter Based on DNA repair time for single strand break ns

tone

)

BasedonDNArepairtimeforsinglestrandbreak

Hypofractionation 1Gy/sec/liter y

of C

. Joh

n

For20Gy TotalDose 4fractions,5Gy/fraction

1to5‐8sec,orbreath‐holddelivery(1secchallengingforbeammonitoring)

(Cou

rtesy

Radiobiology 5Gy/sec/liter

SingleFraction,20Gy/fraction 4‐8 sec delivery (corresponding timescale if possible)

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

4 8secdelivery(correspondingtimescaleifpossible)

* from final report of the joint NCI-DOE Ion Therapy workshop, Jan, 2013

Page 107: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

R i t t ti i th *Requirements: next‐generation ion therapy*

AdditionalAcceleratorandBeamDeliveryParameters B P ti BeamProperties:

Selectablespotsize:3,5,and10mm(FWHM) Profile characterizedandstable(transverse,energy,preferablyGaussian)

Energy /Range Modulation: Energy/RangeModulation: 2MeV stepsforprotons(~0.25cmstepinrange) 2 MeV/nucleonstepsforcarbon(~0.1cmstepinrange)

100millisec steprate

nsto

ne)

FieldSize:Maximum‐ 40x40cm2,minimum‐ 20x20cm2

Lateraltargetingaccuracy@Braggpeak

yof

C. J

ohn

Protons:±0.5mm Carbon:±0.2mm(needstobestudied)

Doseaccuracy/fraction 2 5% monitored at ≥40 kHz during dose deposition

(Cou

rtesy

2.5%monitoredat≥40kHzduringdosedeposition Real‐timeBeammonitoring

Fastnondestructivemonitoringandfeedback Analysisofpatient‐inducedsecondaries duringtreatment

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

a ys s o pat e t duced seco da es du g t eat e t

* from final report of the joint NCI-DOE Ion Therapy workshop, Jan, 2013

Page 108: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

N t ti i th l t *Next‐generation ion therapy accelerators* DoseDeliveryforTreatment

20 Gy/min/liter has been defined as the minimum “standard” for the ion accelerator* 20Gy/min/literhasbeendefinedastheminimum standard fortheionaccelerator Twofields(representdifferenttechnicalspecificationsforbeam):

30cmx30cm(singlelayer field) 10x10x10cm3

Requires~40energystepstoevenlycoverindepth;(assumes0.25cm/layer,~2MeV energystep) S i R t 5 / (10 / i t t t f th t) ScanningRate:5cm/msec (10cm/msec iscurrentstateoftheart) Energymodulation,100msec/energystep ~109 p/Gy/cm2 (forcarbondividebyratioofRBEs,~3).

For 20 Gy Total Dose nsto

ne)

For20Gy TotalDose NormalFraction:

20treatments,1Gy/fraction,1secdelivery 1012p/secfor30cmx30cm(singlelayer field) 4x1012 p/secforand10x10x10cm3 fied (40layers)

H f ti ti yof

C. J

ohn

Hypofractionation: 4fractions,5‐8Gy/fraction

1secdeliveryincreasesintensitybydosefactor upto2‐3x1013 p/sec

5‐8secdelivery S i t iti l f ti d 1 d li

(Cou

rtesy

Sameintensitiesasnormalfractionand1secdelivery Radiobiology:

SingleFraction,20Gy/fraction,5‐8secdelivery(ifpossible) 2‐4x1012p/secfor30cmx30cm(singlelayer field) 1‐1.6‐4x1013 p/secforand10x10x10cm3 fied (40layers)G. Coutrakon, et. al., Proceedings 1999 PAC

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

, , g

Page 109: HdHadrons cancer therapy CNAO - University of Novi Sadcern-danube-school.uns.ac.rs/assets/talks/Pullia... · 2014-09-21 · Physicsand medicine togethersincelong: diagnosis and therapy

ConclusionsProtontherapy centres are commercial systems (and single

l ti i ) Thi i t t f broom solutions are coming up). This is not true for carbonfacilities yet (space and need for firms involvement).

CNAO is now treating patients with both protons and carbon, butimprovements and R&D are always ongoing.

Improvements of technology in hadrontherapy are not limited toaccelerators but invest a wide spectrum of systems: some moreaccelerators, but invest a wide spectrum of systems: some moreurgent than others.

Collaborations, intercomparisons, networking are key issues forthe success of hadrontherapy and are needed to establishE id B d M di i ( ti t th h t i i )

Marco Pullia – Hadrons for cancer therapy - Danube School on Instrumentation in Elementary Particle & Nuclear Physics – September 8th, 2014

Evidence Based Medicine (patient throughput is an issue) .