Date post: | 16-Oct-2016 |
Category: |
Documents |
Upload: | gordon-wright |
View: | 214 times |
Download: | 0 times |
____________________________ CURRENT REVIEW
Hemodynamic Analysis Could Resolve the PulsatileBlood Flow ControversyGordon Wright, PhD
Department of Biomedical Engineering and Medical Physics, University of Keele, Staffordshire, England
Reasons for fluctuations in the popularity of pulsatilecardiopulmonary bypass are discussed. The advantagesof pulsatile flow have previously been stated in terms ofphysiologic parameters. This has produced a weak theoretic framework that has led to serious misunderstandings of the fundamental mechanical properties of pulsatile flow. Failure to consider these properties in the past
Long before the birth of open heart surgery, mechanicalpumps had been designed, manufactured, and used
to establish circulatory support in isolated organs andwhole animals [1, 2]. By the time technologic developments had proceeded sufficiently to permit the first openheart operation with cardiopulmonary bypass [3], it waswidely believed that pulsatile flow was better than nonpulsatile flow because the body was adapted to it. However, the practice of open heart surgery was commencedand continued for many years with the aid of ripple(nonpulsatile) flow, extracorporeal circulation. There wereprobably three main reasons for this.
First, the case for pulsatile flow rested on classic physiologic theory [1, 4-6], and evidence from studies ofkidney, limb, and total body perfusion suggested that therewere no changes in reactivity, function, or survival timeassociated with nonpulsatile flow, so long as blood flowrates and mean arterial pressures were maintained withinthe normal range [7-9]. Wesolowski's publications duringthe early 1950s were an influential barrier to the adoptionof pulsatile flows for extracorporeal circulation. He and hiscolleagues showed that nonpulsatile right heart bypass indogs was compatible with normal respiratory functionsand associated with no permanent alterations in circulatory dynamics [10]. This was later confirmed by Clarke [11]and Wemple [12] and their colleagues. In the same series ofexperiments, Wesolowski's group produced preliminaryevidence that pulsatile perfusion caused less systemichypotension than did nonpulsatile perfusion, but thispoint was subsequently negated by their finding no significant differences in the hemodynamics, kidney function,vascular tone, recovery rate, blood indices, and organhistologic characteristics [13].
The criticism later made of these results was that Wesolowski's use of high-flow rates probably eliminated thedifferences [14]. However, by this time, operative mortality
Address reprint requests to Dr Wright, Department of Biomedical Engineering and Medical Physics, University of Keele, Staffordshire ST5 5BG,England.
© 1994 by The Society of Thoracic Surgeons
may have allowed modified roller pumps to be acceptedas "pulsatile," even though their pulsatile power outputsare small compared with those of the human heart. Afresh approach is required in which pulsatile blood flowis analyzed in terms of hemodynamic power and impedance parameters.
(Ann Thorae Surg 1994;58:1199-204)
was declining rapidly and standard procedures were becoming established. Mortality rates below 10% were considered satisfactory and postoperative morbidity was becoming less severe, if not less frequent. Current techniquesof extracorporeal circulation were perceived to be successful, so it is understandable that perfusionists and surgeonswere reluctant to make fundamental changes in theirprocedures.
Second, although pulsatile pumps were used for animalperfusions, there was no pulsatile pump for use in patientsbecause of the inherent complexity of available devices andbecause of problems of cost, unreliability, and sterilization.The roller pump had been available for some time [15, 16].In early versions, the machining tolerances were relativelywide and high compression and shear forces were generated by the rollers. This resulted in high levels of hemolysis, but, even so, the roller pump was preferred to any thencurrently available pulsatile pump, or the even morehemolytic multiple-finger pump [17]. An excellent reviewof blood pump technology during that period was presented by Galletti and Brecher [18] and later reproduced inthe American Physiological Society'S Handbook of Physiology [19].
We now know that blood trauma is not primarilydetermined by the type of flow but rather by the mechanical forces applied to the blood by the pumping mechanism. Positive displacement mechanisms, such as pistonactuated ventricles, are inherently less traumatic thanroller pumps, but cam-operated valves can increase thetrauma if they are not carefully designed and set. In anyevent, blood trauma during cardiopulmonary bypass isprimarily due to cardiotomy suction, not to blood pumping [20-22].
The third reason why pulsatile flow was not favored forextracorporeal circulation may have been that pulsatilepumps generate very high levels of energy in the arterialline. Dislodgement of tubing connectors is not unknownwith the use of high-amplitude pulsatile pumps, and this isa risk that cannot be taken when human lives are involved.
Consequently, pulsatile flow was not adopted for car-
0003-4975/94/$7.00
1200 REVIEW WRIGHTPULSATILE BLOOD FLOW
diopulmonary bypass until the practical problems hadbeen solved. The solution represented a massive compromise. An additional electronic control system was developed so that the familiar roller pump could provide a newkind of pulsatile flow that fluctuated in time but bore littleresemblance to the pulsatile flow generated by the naturalheart, except in terms of the radial artery pressure pulseamplitude [23, 24].
By this time, additional experimental evidence had beenreported, and most of this favored the use of pulsatile flow.It was then necessary to ask whether the much modifiedpulsatile flow generated by roller pumps is really adequateto obtain the benefits attributed to pulsatile flow shown bylaboratory experiments. Taylor's group, working in Glasgow, acquired relevant biochemical and physiologic evidence during human open heart procedures. When theStockert roller pump was operated in its "pulsatile" moderather than its "continuous" mode, they found less stimulation of the renin-angiotensin system and lower peripheral vascular resistance [25]; less depression of the usualstress response to operation as the result of anteriorpituitary secretion of adrenocorticotropic hormone andcortisol release [26]; and less pituitary hypofunction [27]. Ina review of 1,217 patients who underwent open heartprocedures, Taylor [24] claimed a significant reduction inthe number of operative and low-output-related deathswithin the first 24 hours after operation.
Primary Mechanisms of Pulsatile Flow
The Glasgow group attributed the benefits of pulsatileperfusion with the modified roller pump to the concept ofenergy-equivalent pressure, capillary critical closing pressure and microcirculatory patency, and neuroendocrinereflex mechanisms triggered by baroreceptor discharge[24].
Energy-Equivalent PressureThe concept of energy-equivalent pressure was introducedby Shepard and colleagues [28], but has been little usedsince then. Energy-equivalent pressure (EEP) is a way ofexpressing stroke work (joules) in the more familiar unitsof pressure (millimeters of mercury). It is calculated as:
fpqdtEEP=--
f.,where p is pressure and q represents flow.
When the flow is purely nonpulsatile, the integral signscan be removed and energy-equivalent pressure equals themean pressure, but, when the flow is pulsatile, the energyequivalent pressure includes energy within the pressureand flow waveforms. In the normal human heart, thedifference is approximately 10%. Energy-equivalent pressure is therefore not a mechanism, but only a mathematicaldevice to measure variables that are not familiar to clinicians, namely energy and work. Actually, the concept ofenergy is fundamental to an understanding of pulsatile
Ann Thorae Surg1994;58:1199 -204
flow mechanics and should not be circumvented in thismanner, first because it paralyzes further progress towardan understanding of more complex problems in hemodynamics, and second because of the principle that measurement units should not be misused.
Critical Closing PressureThe idea that the microcirculation may collapse at a certaincritical closing pressure was enunciated by Burton in 1954[29]. The only direct experimental evidence that this actually occurs during ripple, or nonpulsatile, perfusion oftissues was supplied by Takeda [30], who described capillary collapse and microcirculatory shunting during rippleflow perfusion. However, there is also some indirect evidence of this. First, diffuse nerve cell changes that arecharacteristic of tissue ischemia were found in the brains ofdogs subjected to total cardiac bypass for 3 hours using aconventional roller pump [31]. Collapsed capillaries wereassociated with these changes. The second piece of evidence comes from unpublished findings from researchperformed in collaboration with Professor C. R. H. Wildevuur at the Surgical Research Centre in Groningen, TheNetherlands. Tissue oxygen pressure histograms [32] wereperformed on the sartorius muscle of dogs undergoingtotal cardiopulmonary bypass using either a conventionalroller pump or a ventricle-type pulsatile pump [33].Preperfusion histograms had a Gaussian distribution, andthis was maintained during pulsatile perfusions with theventricle pump, but, during ripple flow perfusions, thehistograms became markedly left skewed, in that there wasa predominance of low oxygen pressures. However, theconversion from ripple to pulsatile perfusion did notimmediately restore the normal histogram. This took approximately 20 minutes, suggesting the presence of microcirculatory collapse during the period of ripple flow perfusion that was not easily reversed.
Neuroendocrine ReflexesIt was shown by Ead and colleagues [34] that baroreceptorsin the carotid sinuses respond to both mean and pulsatilepressures and that carotid sinus nerve discharge is decreased by damping the carotid artery pulse. The sinusnerve exerts an afferent inhibitory effect on vasomotordischarge, so that reflex vasoconstriction occurs when thesinus pulse is damped. A second mechanism was identified by Many and associates [35-37], who showed thatrenin secretion was increased and renal function wasimpaired when blood flow through the abdominal aortawas depulsed by passing it through a high-compliancechamber.
Thus, it appears that there may be at least two separatemechanisms underlying the vascular responses to nonpulsa tile flow-a loss of vasomotor inhibition by the carotidsinus nerve and renal secretion of renin, with subsequentliberation of the powerful vasoconstrictor angiotensin II. Itis not clear why two mechanisms exist for the samefunction, but there are precedents in other physiologiccontrol systems in which the nervous reflex provides arapid response and the endocrine system provides thesustained response.
Ann Thorae Surg1994;58:1199-204
Other FactorsThere are other factors that have been suggested as primary mechanisms for the improved tissue perfusion observed during pulsatile extracorporeal circulation.
THE TELEOLOGIC ARGUMENT. Even primitive hearts generate pulsatile flow. However, during the course of evolution, both the mean arterial pressure and the pulse amplitude have increased [39]. The argument is that "natureknows best" and should be emulated. This is a weakargument, because the observation of a trend does notnecessarily constitute any experimental evidence of a benefit. However, it may be accepted as circumstantial evidence, if one likes.
HIGHER RATES OF TISSUE FLUID MOVEMENT AND THE FLOW
OF LYMPH. The monumental and diligent research of Parsons and McMaster [39, 40] on the spread of dyes inisolated perfused rabbit ears convincingly demonstratedthat the arterial pulse is important for promoting themovement of tissue fluids, as well as the formation andflow of lymph. There are three mechanisms responsible forthe promotion of lymph flow from the tissues to thelymphatic duct. These are muscular compression, arterialpulsations, and lymphatic contractions. In anesthetizedpatients, muscular contractions should be absent. Withregard to the arterial pulsations, the small lymphaticvessels are wrapped around the arteries and protected bya common sheath. Therefore, arterial contractions aretransmitted to the lymphatics and the flow of lymphtoward the lymphatic duct is ensured by one-way valves.This mechanism depends on pulsatile energy supplied tothe lymphatics, and this is reduced greatly during nonpulsatile extracorporeal circulation. This leaves the lowfrequency lymphatic contractions as the only mechanismto move lymph during nonpulsatile extracorporeal circulation, and we do not know how effective these contractions are.
However, Anabtawi and colleagues [41] showed thatlymph formation and flow were similar during pulsatileand nonpulsatile perfusions, and that, in the intact animal,they are actually related to the mean arterial pressure andlimited by the flow capacity of the thoracic duct.
HIGHER RATES OF METABOLISM. Shepard and Kirklin [42]showed that oxygen consumption is higher during pulsatile perfusions than during nonpulsatile perfusions, andspeculated that the possible mechanisms for this may be,first, that "jiggling of the tissues during pulsatile flow ...may act to change the boundary layer of interstitial fluidaround cell membranes and thus enhance diffusion"; second, that the increased lymph and interstitial fluid flowmay minimize the development of increased diffusiondistances between capillaries and cells; or, third, "thatpulsatile energy may be required to ensure that a normalpercentage of the total number of arterioles in a vascularbed are open at anyone time.... r r
REVIEW WRIGHT 1201PULSATILE BLOOD FLOW
Hemodynamic Energy Transmission
The preceding discussion suggests that there is considerable disagreement about the primary physiologic mechanisms by which pulsatile flow confers benefits in terms oftissue perfusion. In fact, several of these mechanisms maybe involved, but they are not the primary ones, and toconsider them as such may be misleading. As in allbiologic systems, the primary mechanism is energy transduction. In the case of the cardiovascular system, theparticular transduction mechanism is the conversion ofchemical energy into mechanical energy. The lack of appreciation of this fundamental mechanism and of themechanical properties of the vascular bed that facilitate thetransmission of mechanical energy through the vascularbed to the tissues forms the intellectual background thathas resulted in the development of mechanical pumps thatgenerate various types of blood flow that are labeled"pulsatile," but which are not designed to be matched tothe vascular impedance.
For optimal design methods to be feasible, some understanding of pulsatile flow hemodynamics is required, andthis must involve further consideration of the mechanicalproperties of both the heart, or pump, and the vascularbed, and of the interactions between them. We need toconsider the heart, or pump, as a complex power sourceand the vascular load as a complex impedance possessingboth resistive and reactive properties. Methods of computing these variables have been described [43-47] and textbooks on the topic are available [48-50], but the methodsare rarely mentioned in the cardiopulmonary bypass literature.
The techniques are based on the Fourier transformationof blood pressure and flow data recorded from the proximal aorta. The techniques have limitations and only provide a starting point [51]. There is much more to belearned, especially about the interactions between the heartand the vascular load, and it may be that nonlinearrelationships and additional dimensions will have to beincorporated to obtain a sufficiently comprehensive mathematical description of all of the properties of the cardiovascular system. Nevertheless, this approach offers realinsights into the pulsatile properties of cardiovascularfunction that cannot be obtained by a purely physiologicapproach.
The basic philosophy is that the cardiovascular system isanalogous to an electrical transmission line, and thussimilar mathematical techniques can be used to describe itsperformance. Then, left ventricular function is defined interms of its total hemodynamic power output consisting ofmean and pulsatile components. The load that the leftventricle works against is the aortic input impedance,which has resistive and reactive properties, and the reactive components are arterial compliance and the inertia ofblood.
These parameters are not independent. Power and impedance interact and the pulsatile properties of the heartare dependent on the reactive properties of the vascularimpedance. In addition, the individual components of leftventricular power output (mean and pulsatile) interact,
1202 REVIEW WRIGHTPULSATILE BLOOD FLOW
and so do the individual components of the vascularimpedance. For instance, increasing the peripheral resistance by the addition of the vasoconstrictor drug methoxamine immediately increases the characteristic resistance,reduces arterial compliance, and reduces inertia [47]. Theseimpedance changes are reversed by administering thevasodilator drug hydralazine. However, to complicate theissue further, arterial reactive changes later ensue to offsetthe immediate changes in the characteristic resistance andcompliance. Evidently, these parameters are under theinfluence of a physiologic control system, and this must berelated to the pulsatile power output of the left ventriclebecause the mean power output is unaffected by thesechanges.
It may be that the role of this control system is to ensurethat the total hemodynamic energy level is maintained at anear constant level, or that the ratio between pulsatile andnonpulsatile energy is maintained. The results of experiments in which the impedance parameters were artificiallyvaried are currently being studied in an attempt to definethese properties.
We have noticed that the initial vascular responses justdescribed also occur in perfused mechanical models, aphenomenon that cannot be adequately explained by theanalogy with an electrical transmission line theory appliedto alternating currents, indicating that the techniques are inneed of further refinement. However, the delayed reactiveresponses do not occur in mechanical models, and thisshows that they are not merely mathematical phenomenabut may have real physiologic significance. If a mechanismexists to compensate for changes in impedance and pulsatile energy levels, this implies that it is important tomaintain the pulsatile component of energy.
It also implies that some arterial receptor mechanismexists to monitor the pulsatile energy content of the blood.This mechanism may be the carotid baroreceptors, whichare actually misnamed because they respond to sinus wallstretch, not to sinus pressure. They are therefore affectedby radial, rather than perpendicular, forces, and respond toboth mean and pulsatile applications of these forces. Thedischarge rate from individual baroreceptor units is increased by sinus pulsations, but additional units are alsorecruited [34,52]. Thus, the sensory mechanism for recognizing arterial pulsations may reside in the selective response of individual baroreceptor units within the carotidsinus.
It follows from this discussion that the pulsatile poweroutput of the left ventricle may be matched to the reactiveproperties of the aortic input impedance, thereby optimizing the transmission of mechanical energy from the heartto the tissues. Replacement of the normal aortic inputimpedance by an abnormal load is likely to impair the leftventricular hemodynamic power output. This appears tobe confirmed in the setting of isolated rat hearts ejectinginto a mechanical model of the rat's aortic input impedance. As the model's arterial compliance was reduced, leftventricular hemodynamic power output decreased and, ata critical value of compliance that is approximately 40% ofthe normal value in the circulation, it approached zero. Theconverse of this is that replacement of the left ventricle by
Ann Thorae Surg1994;58:1199-204
a mechanical pump delivering levels of pulsatile powerthat are not matched to the reactive properties of the aorticinput impedance is likely to result in the suboptimaldelivery of energy to the tissues. Otherwise, pulsatilepower transmission to the vascular beds will be less thanoptimal and the benefits of pulsatile flow may not beachieved. This may be the crucial factor.
With this hypothesis in mind, it is instructive to examinethe pulsatile power output of the human left ventricle andto compare this with those of available so-called pulsatilepumps. Sample pressure and flow waveforms generatedby the human heart and various pulsatile pumps areavailable [53-55]. There is a wide range of flow waveformsgenerated by pumps that have been labeled "pulsatile,"and the pulsatile power outputs of these pumps aremarkedly affected by the vascular impedance and by themechanical characteristics of the arterial line [54,55]. Basedon measurements of aortic blood pressure and flow inpatients undergoing open heart surgical procedures, wecomputed left ventricular pulsatile power outputs of 111 ::':::14 mW (standard error of the mean) [56,57]. The pulsatilepower output of the Stockert roller pump, operated underthe most favorable conditions (automatic control set tomaximum, a stiff 12.6-mm [inner diameter] pump tube,and no inclusions in the arterial line) was found to be 12 ::':::1 mW (standard error of the mean), or 10.8% of thepulsatile power output of the human heart, depending onthe impedance of the vascular load. When the pump wasset to prevent negative pressure from developing by adjusting the automatic control to 80% of maximum, thepulsatile power output was even lower, that is, 4.5 mW, or4.05% of the pulsatile power output of the human heart; inthe continuous mode, the pulsatile power output wasapproximately 2 mW, or 1.8% of the pulsatile poweroutput of the human heart. The difference in power outputs between the two modes of operation was thereforeonly 2.7%. In contrast, the Keele ventricle pump generatedpulsatile power outputs higher than those of the humanheart.
Unfortunately, it is the modified roller pump that hasbeen most commonly used to generate "pulsatile" flow forcomparative clinical investigations, and it is probablybecause of the low pulsatile power output of this pump,and some other pumps, combined with various vascularimpedances (particularly arterial compliance), that someinvestigators have been unable to find any differencesbetween "pulsatile" and "continuous" perfusions (seereviews by Mavroudis [58] and Hickey and colleagues[59]). Interest in pulsatile perfusion for open heart surgicalprocedures has declined, probably because the pumps donot generate sufficient pulsatile power to achieve many ofthe theoretic advantages of pulsatile power, but also because of the increased use of arterial line membraneoxygenators, which dampen the pulse to levels below 3mW.
Experimental and clinical benefits that have been attributed to pulsatile flow include improved preservation ofkidney, heart, pancreas, thyroid, and brain function; elimination of the depressed renal function associated withnonpulsatile perfusion; reduced levels of postoperative
Ann Thorac Surg1994;58:1199-204
systemic hypertension; and reduced postoperative mortality. However, many contradictory findings have also beenreported, and these showed that pulsatile flow was nobetter than nonpulsatile flow in some respects. The explanation offered by Hickey and colleagues [59] for thisdilemma was that "different investigators employ differentforms of pulsatile perfusion, only some of which areeffective." This is my view also, but I would define"effective" to mean "matched to the reactive properties ofthe vascular impedance."
However, there is one irritating problem with this conclusion, and this is that some investigators have attributedmajor benefits to pulsatile flow delivered by modifiedroller pumps, which clearly are unable to deliver morethan a small fraction of the pulsatile power delivered bythe human left ventricle. Does this imply that even a smallamount of pulsatile power is sufficient to improve tissueperfusion if it is matched to the vascular impedance, or isthe explanation that the benefits observed were actuallydue to factors other than pulsatile flow? One such possibility is that the flow rates may have been different in thepulsatile and nonpulsatile groups. Usually the method ofrecording the pump flow rate is not declared, so wepresume that it was read from the pump display. This isdifficult to read with any degree of accuracy and theauthors do not say how, or how frequently, the displaywas calibrated.
Conclusion
The use of pulsatile flow for cardiopulmonary bypass inhuman patients was most popular during the 1980s. Morerecently, its use has declined, and there are several possiblereasons for this. First, there has been a lack of understanding of the basic mechanical properties of the left ventricleand major arteries and of the interactions between them.Second, there has not been a satisfactory definition ofpulsatile flow, and this has led to the use of pumps thatdeliver different types of flow for experimental purposes.A third reason for the declining use of pulsatile flow forcardiopulmonary bypass has been the recent developmentof membrane gas-exchange devices that dampen the pulseand that are intolerant to high pressures in the arterial line.The concerted use of modified roller pumps to generateso-called pulsatile flow, even though this bears little resemblance to the pulsatile flow generated by the natural heart,is a fourth reason.
All experimental and clinical investigations should begin with the definition of new terms and the developmentof a suitable theoretic framework. This procedure has notbeen followed in the work on pulsatile flow. Such aframework is available and makes use of the concepts ofhemodynamic power and vascular impedance. Furtherelaboration of the framework is required, but, even in itspresent form, it offers far greater insights into the fundamental mechanisms by which pulsatile flow may improvetissue perfusion than do the hitherto used physiologicexplanations. In particular, it may explain why some flowwaveforms are beneficial, while others are not, based on aknowledge of the pulsatile power output of the pump in
REVIEW WRIGHT 1203PULSATILE BLOOD FLOW
relation to the reactive properties of the vascular impedance. So far, the only pump that is guaranteed to generateoptimal pulsatile power output in the human circulation isthe human heart.
References
1. Hooker DR. A study of the isolated kidney: the influence ofpulse pressure upon renal function. Am J Physiol 1910;27:24-44.
2. Dale HH, Schuster EA. A double perfusion pump. J Physiol(Lond) 1928;64:356.
3. Gibbon JH. Application of a mechanical heart and lungapparatus to cardiac surgery. Minn Med 1954;37:171-85.
4. Ludwig C, Schmidt A. Das Verhalten der Gase, welche mitdem Blut durch den reizbaren Saugethiermuskel Stromen.Leipzig Ber 1868;20:12-72.
5. Gesell RA. On the relation of pulse pressure to renal secretion.Am J Physiol 1913;32:70-93.
6. Carrel A, Lindbergh C. Culture of organs. New York: Hoeber,1938.
7. Selkurt E. Effect of pulse pressure and mean arterial pressuremodification on renal hemodynamics and electrolyte andwater excretion. Circulation 1951;4:541-51.
8. Randal JE, Stacy RW. Pulsatile and steady pressure-flowrelations in the vascular bed of the hindleg of the dog. Am JPhysiol 1956;185:351-4.
9. Wesolowski SA. Extracorporeal circulation: continuous, controlled variation of the frequency, volume and systolic risetime of the pulse. J Appl Physiol 1954;6:809-14.
10. Wesolowski SA, Fisher JH, Welch CS. Perfusion of the pulmonary circulation by nonpulsatile flow. Surgery 1953;33:370-5.
11. Clarke CP, Kahn DR, Dufek JH, Sloan H. The effects ofnonpulsatile blood flow on canine lungs. Ann Thorac Surg1968;6:450-7.
12. Wemple RR, Mockros LF, Lewis FJ. Pulmonary functionduring pulsatile and nonpulsatile right heart bypass. J ThoracCardiovasc Surg 1969;57:190-202.
13. Wesolowski SA, Sauvage LR, Pine RD. Extracorporeal circulation: the role of the pulse in maintenance of the systemiccirculation during heart-lung bypass. Surgery 1955;37:663-82.
14. Ogata T, Ida Y, Nonoyama A, Takeda 1, Sasake H. A comparative study on the effectiveness of pulsatile and non-pulsatileblood flow in extracorporeal circulation. Arch Ipn Chir 1960;29:59-66.
15. DeBakey ME. Simple continuous-flow blood transfusion instrument. New Orleans Med Soc J 1934;87:386-9.
16. Henry H, Jouvelet P. Appareil a transfusion du sang. BullAcad Med Paris 1934;111:312-9.
17. DeWall RA, Warden HE, Read RC, et a1.A simple, expendableartificial oxygenator for open-heart surgery. Surg Clin NorthAm 1956;36:1025-34.
18. Galletti PM, Brecher GA. Heart-lung bypass: principles andtechniques of extracorporeal circulation. New York: Grune &Stratton, 1962:207-12.
19. Galletti PM, Brecher GA. Artificial blood circulation. In: Hamilton WF, Dow P, eds. Handbook of physiology; section 2,circulation, vol 3. Washington: American Physiological Society, 1965;2123-61.
20. de [ong JCF, ten Duis HJ, Smit Sibinga CT, Wildevuur CRH.Hematologic aspects of cardiotomy suction in cardiac operations. J Thorac Cardiovasc Surg 1980;79:227-36.
21. Wright G. Haematological effects of cardiotomy suction. In:Longmore DB, ed. Towards safer cardiac surgery. Lancaster:MTP Press, 1981;313-23.
22. Wright G. Blood cell trauma. In: Taylor KM, ed. Cardiopulmonary bypass: principles and techniques. London: Chapmanand Hall, 1986:249-76.
23. Taylor KM, Bain WH, Maxted K1, Hutton MH, McNab WY,Caves PK. Comparative studies of pulsatile and non-pulsatileflow during cardiopulmonary bypass. 1. Pulsatile system
1204 REVIEW WRIGHTPULSATILE BLOOD FLOW
employed and its hematological effects. J Thorac CardiovascSurg 1978;75:569-73.
24. Taylor KM. Pulsatile perfusion. In: KM Taylor, ed. Cardiopulmonary bypass: principles and techniques. London: Chapmanand Hall, 1986:85-114.
25. Taylor KM, Morton JJ, Brown JJ, Bain WH, Caves PK. Hypertension and the renin-angiotensin system following openheart surgery. J Thorac Cardiovasc Surg 1977;74:840-5.
26. Taylor KM, Wright GS, Reid JS, et al. Comparative studies ofpulsatile and non-pulsatile flow during cardiopulmonary bypass. II. The effects on adrenal secretion of cortisol. J ThoracCardiovasc Surg 1978;75:574-8.
27. Taylor KM, Wright GS, Bain WH, Caves PK, Beastall GS.Comparative studies of pulsatile and nonpulsatile flow duringcardiopulmonary bypass. III. Response of anterior pituitarygland to thyrotropin-releasing hormone. J Thorac CardiovascSurg 1978;75:579-84.
28. Shepard RB, Simpson De, Sharp JE. Energy equivalent pressure. Arch Surg 1966;93:730-40.
29. Burton AC Physiology and biophysics of the circulation: anintroductory text. Chicago: Year Book, 1954:80-1.
30. Takeda J. Experimental study of peripheral circulation duringextracorporeal circulation, with a special reference to a comparison of pulsatile flow with non-pulsatile flow. Arch [pnChir 1960;29:1407-30.
31. Sanderson JM, Wright G, Sims FW. Brain damage in dogsimmediately following pulsatile and non-pulsatile blood flowsin extracorporeal circulation. Thorax 1972;27:275-86.
32. Pranger RL, Mook PH, Elstrodt JM, Kessler M, Lubbers OW,Wildevuur CRH. Improved tissue perfusion (P02 histograms)in extracorporeal circulation (ECC) using membrane (MO)instead of bubble (BO) oxygenators. J Thorac Cardiovasc Surg1980;79:513-26.
33. Wright G. The hydraulic power outputs of pulsatile andnonpulsatile cardiopulmonary bypass pumps. Perfusion 1988;3:251-62.
34. Ead HW, Green JH, Neil EA. Comparison of the effects ofpulsatile and non-pulsatile blood flow through the carotidsinus on the reflexogenic activity of the sinus baroreceptors inthe cat. J Physiol (Lond) 1952;118:509-19.
35. Many M, Soroff HS, Birtwell We, Giron F, Wise H, DeterlingRA. The physiologic role of pulsatile and nonpulsatile bloodflow. II. Effects on renal function. Arch Surg 1967;95:762-7.
36. Many M, Soroff HS, Birtwell We, Deterling RA. The effects ofbilateral renal artery depulsation on renin levels. Surg Forum1968;19:387-9.
37. Many M, Giron F, Birtwell We, Deterling RA, Soroff HS.Effects of depulsation of renal blood flow upon renal functionand renin secretion. Surgery 1969;66:242-9.
38. Wilkens H, Regelson W, Hoffmeister FS. The physiologicimportance of pulsatile blood flow. N Engl J Med 1962;267:443-6.
39. Parsons RJ, McMaster PD. The effect of the pulse upon theformation and flow of lymph. J Exp Med 1938;68:353-76.
40. McMaster PO, Parsons RJ. The effect of the pulse on the spreadof substances through tissues. J Exp Med 1938;68:377-400.
Ann Thorae Surg1994;58:1199-204
41. Anabtawi IN, Womack CE, Ellison RG. Thoracic duct lymphflow during pulsatile and nonpulsatile extracorporeal circulation. Ann Thorac Surg 1966;2:38-43.
42. Shepard RB, Kirklin JW. Relation of pulsatile flow to oxygenconsumption and other variables during cardiopulmonarybypass. J Thorac Cardiovasc Surg 1969;58:694-702.
43. Milnor WR, Bergel DH, Bargainer JD. Hydraulic power associated with pulmonary blood flow and its relation to heartrate. Circ Res 1966;19:467-80.
44. O'Rourke MF. Steady and pulsatile energy losses in thesystemic circulation under normal conditions and in simulated arterial disease. Cardiovasc Res 1967;1:313-26.
45. Cox RH. Determinants of systemic hydraulic power in unanesthetized dogs. Am J Physiol 1974;226:579-87.
46. Westerhof N, Elzinga G, Sipkema P, van den Bos GC Quantitative analysis of the arterial system and heart by means ofpressure-flow relations. In: Hwang NHC, Normann NA, eds,Cardiovascular flow dynamics and measurements. Baltimore:University Park Press, 1977:403-38.
47. Wright G. Effects of vasoactive drugs upon haemodynamicpower and input impedance in normal rats. Cardiovasc Res1991;25:923-9.
48. McDonald DA. Blood flow in arteries. London: Edward Arnold,1974:351-8.
49. Milnor WR. Hemodynamics. Baltimore: Williams & Wilkins,1982:157-271.
50. Nichols WW, O'Rourke MF. McDonald's blood flow in arteries: theoretical, experimental and clinical principles, 3rd ed.London: Edward Arnold, 1990.
51. Wright G. Engineering and physiological approaches to thestudy of cardiovascular function: science and pseudoscience?Cardiovasc Res 1992;26:215-7.
52. Folkow B, Neil E. Circulation. London: Oxford UniversityPress, 1971:320-39.
53. Sanderson JM, Morton PG, Tolloczko TS, Vennart T, Wright G.The Morton-Keele pump-a hydraulically activated pulsatilepump for use in extracorporeal circulation. Med BioI Eng1973;March:182-90.
54. Wright G. The hydraulic power outputs of pulsatile andnonpulsatile cardiopulmonary bypass pumps. Perfusion 1988;3:251-62.
55. Wright G. Factors affecting the pulsatile hydraulic poweroutput of the Stockert roller pump. Perfusion 1989;4:187-95.
56. Wright G, Sum Ping JTS, Campbell CS, Tobias MA. Computation of haemodynamic power and input impedance in theascending aorta of patients undergoing open heart surgery.Cardiovasc Res 1988;22:179-84.
57. Wright G, Sum Ping JTS, Campbell CS, Tobias MA. Assessment of cardiovascular function in patients undergoing coronary artery bypass grafting. J Thorac Cardiovasc Surg 1988;96:400-7.
58. Mavroudis C To pulse or not to pulse. Ann Thorac Surg1978;25:259-71.
59. Hickey PR, Buckley MJ, Philbin OM. Pulsatile and nonpulsatile cardiopulmonary bypass: review of a counterproductivecontroversy. Ann Thorac Surg 1983;367:20-37.