HemoSphere Advanced Monitoring Platform Setup guide Parameter tile Navigation bar Main window Power button Status bar Visual alarm indicator Information bar HemoSphere platform
Transcript
Page 1: HemoSphere Advanced Monitoring Platform...HemoSphere advanced monitor screen features Monitoring mode Start/stop moni toring Time-in-Target software Zero and Waveform Intervention

HemoSphere Advanced Monitoring Platform Setup guide

Parameter tile

Navigation bar Main windowPowerbutton

Status bar Visual alarm indicator

Information bar

Hem

oSphere platform

Page 2: HemoSphere Advanced Monitoring Platform...HemoSphere advanced monitor screen features Monitoring mode Start/stop moni toring Time-in-Target software Zero and Waveform Intervention

HemoSphere advanced monitor screen features

Monitoring mode

Start/stop monitoring

Time-in-Target software

Zero and Waveform

Intervention analysis

Continuous blood pressure waveform, available on the following screens:• Graphical trend screen• Main monitoring screen

The settings icon provides access to: • Clinical tools • Settings• Select screens • Help

Silence audible alarm/pause mode

Information bar (located at the top of the screen)

Acumen IQ mode only

Swan-Ganz mode only

Hem

oSph

ere

plat

form

Page 3: HemoSphere Advanced Monitoring Platform...HemoSphere advanced monitor screen features Monitoring mode Start/stop moni toring Time-in-Target software Zero and Waveform Intervention

Clinical actions

Changing parameters1. Touch the displayed parameter label located

inside the parameter tile to change it to a different parameter.

2. The tile configuration menu will show the selected parameter highlighted and other parameters currently being displayed outlined. Available parameters appear on the screen without highlight.

3. Select desired parameter to display.

Changing targets and alarms1. Touch the parameter value on a

parameter tile.2. Adjust target values using arrows or select

value and type in parameter target. 3. Select Target to disable or enable colors

in target ranges.4. To silence parameter alarms, select

Audible Alarm button.

Hem

oSphere clinical actions

Page 4: HemoSphere Advanced Monitoring Platform...HemoSphere advanced monitor screen features Monitoring mode Start/stop moni toring Time-in-Target software Zero and Waveform Intervention

Pause monitoring1. Hold the Silence Alarm button until monitoring

pause confirmation pop-up appears.2. Confirm suspension of monitoring operations.3. To resume monitoring, touch the Monitoring

Pause Exit button.

ZeroingZero with the stopcock off to the patient and open to air:1. Hold down zero button on the pressure

cable to pull up zero screen.2. Hold down zero button until a tone is

heard to complete zeroing. or Zeroing can be completed by selecting the Quick Zero button to pull up the zero screen, then select Zero.