Outstanding biocidal action Optimised corrosion inhibitor system Suitable for the disinfection of endoscopes, ultra-sound probes and other reusable, heat sensitive medical devices Good environmental and operator safety profiles High Level Disinfectant for use with Medical Devices
Transcript
Page 1: High Level Disinfectant for use with Medical Devices...Organism/Disease Strain Test Protocol Contact Time Log Reduction Aspergillus niger (spores) ATCC 16404 EN 13624 60 minutes* >4

• Outstanding biocidal action

• Optimised corrosion inhibitor system

• Suitable for the disinfection of endoscopes, ultra-sound probes and other reusable, heat sensitive medical devices

• Goodenvironmentalandoperatorsafetyprofiles

High Level Disinfectant foruse with Medical Devices

Page 2: High Level Disinfectant for use with Medical Devices...Organism/Disease Strain Test Protocol Contact Time Log Reduction Aspergillus niger (spores) ATCC 16404 EN 13624 60 minutes* >4

BiocidalEfficacy

Proven Broad Spectrum Efficacy

Independentlyprovenhighlyeffectiveagainst viruses, mycobacteria, bacteria and bacterial spores, Rely+On™ PeraSafe™ provides rapid broad spectrum decontamination against disease-causing pathogens of concern to human health, such as, HIV, Hepatitis B & C, Mycobacteria (Tuberculosis),Pseudomonas aeruginosa, Clostridiumdifficile,MRSAandCandidaalbicans.

Rely+On™ PeraSafe™ has been independently evaluated in the United Kingdom by the Hospital Infection Research Laboratory(HIRL).

TheHIRLevaluationfindingsdemonstratethat, in 10 minutes or less, Rely+On™ PeraSafe™ reduces colony forming units of >5 log10 for Bacillus subtilis var niger spores (surface test), and >5 log10 for Mycobacteriumterrae.TheseresultsclearlydemonstrateRely+On™ PeraSafe™’s rapid biocidal action and suitabilityforuseinmedicaldevicedisinfection.

Medical Device Compatibility

Rely+On™ PeraSafe™ is compatible with endoscopes,ultra-sound probes, surgical instruments and other commonreusable,heatsensitivemedicaldevices. Anoptimisedcorrosioninhibitorsystem,alliedtoa neutralpHsolution,optimisesefficacyandcompatibility.

Specifically developed to offer rapid broad spectrum decontamination against viruses, mycobacteria, bacteria, yeasts and bacterial spores. Rely+On™ PeraSafe™, with its optimised corrosion inhibitor system and near neutral pH, is a suitable solution for endoscope, ultra-sound probe and other reusable, heat sensitive medical device disinfection.

Rely+On™ PeraSafe™ powder generates peracetic acid in-situwhendissolvedinwater.TheneutralpHsolution combinesthebiocidalefficacyofperaceticacidwith excellent materials compatibility to allow repeated, rapid decontamination of pathogenic micro-organisms from heatlabile,reusablemedicaldevicesandinstruments.

Certification

Rely+On™ PeraSafe™ complies with the Medical DevicesDirectiveasassessedbyBSIProductServices(NotifiedBodyNo.0086)(CECertificateNo.01455).

Rely+OnTM

PeraSafeTM

Page 3: High Level Disinfectant for use with Medical Devices...Organism/Disease Strain Test Protocol Contact Time Log Reduction Aspergillus niger (spores) ATCC 16404 EN 13624 60 minutes* >4

Operator Safety

Rely+On™PeraSafe™isnotclassifiedasasensitiser inboththepowderedform,andthein-usedilutions.

Environmental Profile

Rely+On™ PeraSafe™ has been formulated to includeingredients that have been carefully selected for theirabilitytodegradenaturallywithintheenvironment.

Lifespan of Disinfectant Solution

Maximum24hoursfromactivationand/or20 immersions,whicheveristheleast.Forfungicidal activityclaims,solutionsaretobeusedwithin8hours ofpreparation.Unusedorinactivesolutionsmaybe disposed of via the normal waste water discharge channels*.

Rely+On™ PeraSafe™ is available in the following presentations:

Pack Size Volume of disinfectant solution

16.2gram(sachet) 1 litre

81gram(bottle) 5 litres

162gram(bottle) 10 litres

810gram(bottle) 50 litres

Step 1Fillacontainerwith theappropriatevolume oflukewarm tapwater(30-35°C).

Step 2Prepare the solution by adding 81gramsof Rely+On™ PeraSafe™ powder to every 5litresofwater.

Step 3Stir until thepowder dissolvesto form a bluesolution.Thenleavefor 15 minutesto allow thedisinfectant tofullyactivate.

* Alwayscheckandobservethe local site requirements concerning the disposal of solutions entering waste watertreatmentfacilities.

Easy to Prepare

Page 4: High Level Disinfectant for use with Medical Devices...Organism/Disease Strain Test Protocol Contact Time Log Reduction Aspergillus niger (spores) ATCC 16404 EN 13624 60 minutes* >4

Organism/Disease Strain Test Protocol Contact Time Log Reduction

Aspergillusniger(spores) ATCC16404 EN13624 60minutes* >4log

Aspergillusniger(spores) ATCC16404 EN14562 30 minutes >4log

Candida albicans ATCC10231 EN13624 10 minutes >4log

Candida albicans ATCC10231 EN13624 60minutes* >4log

Candida albicans ATCC10231 EN14562 30 minutes >4log

Organism Strain Test Protocol Contact Time Log Reduction

Adenovirus Type 5 EN14476 5 minutes >4log

Hepatitis B DHBV strain EN14476 10 minutes >4log

Hepatitis C BVD surrogate strain EN14476 10 minutes >4log

HIV HTLV-IIIB USAEPA40part158 5 minutes >4log

Poliovirus Type 1 EN14476 5 minutes >4log

Murine norovirus s99 EN14476 10 minutes >4log

VirucidalEfficacy

Lanxess/PC

SRely+OnT

M P

eraS

afeTM

/FY/12.17/G

ThetablesbelowsummariseindependentefficacydataofRely+On™PeraSafe™againsttheimportantdisease-causingorganismsaffectinghumanhealth.

©2017LANXESS.Rely+On™PeraSafe™,LANXESS™andtheLANXESSLogoaretrademarksofLANXESSDeutschlandGmbHoritsaffiliates.Alltrademarksareregisteredinmanycountriesworldwide.

Organism Strain Test Protocol Contact Time Log Reduction

Mycobacterium terrae ATCC15769 EN14563 10 minutes >4log

Mycobacterium avium ATCC15755 EN14563 10 minutes >4log

Mycobacterium terrae NCTC10856 EN14348 10 minutes >5 log

MycobactericidalEfficacy

Organism Strain Test Protocol Contact Time Log Reduction

Campylobacter jejuni Wild strain EN1276 5 minutes >5 log

Escherichia coli NCTC 12900 EN1276 5 minutes >5 log

Enterococcus hirae NCTC12367 EN13727 10 minutes >5 log

Enterococcus hirae ATCC10541 EN14561 10 minutes >5 log

Klebsiella aerogenes NCTC9528 EN1276 5 minutes >5 log

Staphylococcus aureus MRSANCTC12493 EN1276 5 minutes >5 log

Staphylococcus aureus EMRSAtype15 EN1276 5 minutes >5 log

Staphylococcus aureus NCTC10788 EN13727 10 minutes >5 log

Staphylococcus aureus ATCC6538 EN14561 10 minutes >5 log

Salmonella typhimurium Phagetype104 EN1276 5 minutes >5 log

Pseudomonas aeruginosa NCTC6749 Endoscope in-use trial 4minutes >7log

Pseudomonas aeruginosa NCTC6749 EN13727 10 minutes >5 log

Pseudomonas aeruginosa ATCC15442 EN14561 10 minutes >5 log

BactericidalEfficacy

Organism Strain Test Protocol Contact Time Log Reduction

Bacillus subtilis var niger (spores) NCTC10073 EN13704(modified) 5 minutes >4log

Clostridiumdifficile Stool isolate Surface 10 minutes >6log

BactericidalSporeEfficacy

AntecInternationalLimitedWindham Road, Chilton Industrial EstateSudbury,SuffolkCO102XD,UnitedKingdom

Tel:+44(0)1787377305relyondisinfection@lanxess.comwww.relyondisinfection.com

Fungicidal/YeasticidalEfficacy

FullcopiesoftheindependenttestdatasummarisedaboveareavailableuponrequestfromLanxess.*8hoursolutiondata.**Fungicidalclaimsfortheproductareonlyvalidforsolutionsupto8hoursold,fromthetimeoffirstpreparation.


Recommended