+ All Categories
Home > Documents > Highly Selective and Sensitive Thiocyanate PVC Membrane ...Highly Selective and Sensitive...

Highly Selective and Sensitive Thiocyanate PVC Membrane ...Highly Selective and Sensitive...

Date post: 02-Feb-2021
Category:
Upload: others
View: 3 times
Download: 0 times
Share this document with a friend
18
Int. J. Electrochem. Sci., 6 (2011) 3178 - 3195 International Journal of ELECTROCHEMICAL SCIENCE www.electrochemsci.org Highly Selective and Sensitive Thiocyanate PVC Membrane Electrodes Based on Modified Zeolite ZSM-5 Atefeh Badri * , and Pouya Pouladsaz Department of Chemistry, Islamic Azad University, Shahreza Branch, P.O. Box 311-86145, Shahreza, Isfahan, Iran * E-mail: [email protected] Received: 12 June 2011 / Accepted: 11 July 2011 / Published: 1 August 2011 A highly selective membrane electrode based on surfactant modified zeolite is presented. The proposed electrode shows very good selectivity for thiocyanate ions over a wide variety of common inorganic and organic anions. The sensor displays a near Nernstian slope of −58.9 ± 0.9 mV per decade. The working concentration range of the electrode is 1.00×10 -6 - 1.10×10 -1 M with a detection limit of 8.33×10 7 M . The response time of the sensor in whole concentration ranges is very short (<10 s). The response of the sensor is independent on the pH range of 2.010.1. The best performance was obtained with a membrane composition of 25.5% PVC, 64.5% dioctyl phthalate, 10% modified zeolite ZSM-5. It was successfully applied to direct determination of thiocyanate in water samples and as an indicator electrode for titration of thiocyanate ions with AgNO3 solution. Keywords: Thiocyanate, Surfactant modified zeolite ZSM-5 (SMZ), potentiometry, Ion selective membraine electrode (ISE), Poly vinyl chloride (PVC) 1. INTRODUCTION A very interesting development in the field of ion-selective electrodes (ISEs) is the construction of electrodes that can chemically recognize specific anions (chemo selectivity) and offer potentiometric responses that differ from the conventional anion-exchanger-based membranes. The quick determination of trace quantities of ionic species by simple methods is of special interest in analytical chemistry. Construction and application of ion-selective electrode as a potentiometric sensor offers interesting advantages such as simplicity, relatively fast response, low cost, wide linear dynamic range and ease of preparation and procedures. These characteristics have inevitably led to the preparation of numerous sensors for several ionic species, and the list of available electrodes has grown substantially over the past years [1].
Transcript
  • Int. J. Electrochem. Sci., 6 (2011) 3178 - 3195

    International Journal of

    ELECTROCHEMICAL SCIENCE

    www.electrochemsci.org

    Highly Selective and Sensitive Thiocyanate PVC Membrane

    Electrodes Based on Modified Zeolite ZSM-5

    Atefeh Badri*, and Pouya Pouladsaz

    Department of Chemistry, Islamic Azad University, Shahreza Branch, P.O. Box 311-86145, Shahreza,

    Isfahan, Iran *E-mail: [email protected]

    Received: 12 June 2011 / Accepted: 11 July 2011 / Published: 1 August 2011

    A highly selective membrane electrode based on surfactant modified zeolite is presented. The

    proposed electrode shows very good selectivity for thiocyanate ions over a wide variety of common

    inorganic and organic anions. The sensor displays a near Nernstian slope of −58.9 ± 0.9 mV per

    decade. The working concentration range of the electrode is 1.00×10-6- 1.10×10-1 M with a detection

    limit of 8.33×10−7 M . The response time of the sensor in whole concentration ranges is very short

    (

  • Int. J. Electrochem. Sci., Vol. 6, 2011

    3179

    Thiocyanate finds many industrial applications and though not as toxic as cyanide, it is harmful

    to aquatic life. Its determination at low levels in water and industrial effluents is therefore important.

    Thiocyanate sensors are potentially useful for detecting this compound in biological samples (e.g.

    urine and saliva), where elevated levels of thiocyanate correlate with excessive cigarette smoking. If

    the content of thiocyanate ion is a little higher in the body than normal, the protein dialysis will be

    affected and it may even result in coma. Due to the urgent need for selective potentiometric

    determination of trace amounts thiocyanate ion, especially in food, biological and water samples, many

    coordination compounds have been employed as an ionophore in the construction of ISEs thiocyanate

    ion [2–10].

    Recent decade scientific researches introduced several PVC based sensors for different anions,

    such as I− [11–13], SCN- [3, 4, 9], SO4 2− [14–19], ClO4

    − [20], NO2− [21, 22].

    In this work, we report the use of a surfactant-modified zeolite ZSM-5 as an excellent

    ionophore for the preparation of a highly selective and sensitive polymeric membrane sensor for

    thiocyanate ion. The proposed sensor resulted in selectivity coefficients superior to those reported for

    other thiocyanate-selective electrodes.

    2. EXPERIMENTAL

    2.1. Reagents and chemicals

    Sodium silicate used was the product of Merck and its composition is SiO2 (25.2-28.5%), Na2O

    (7.5-8.5%), H2O (63%). Chloridric acid (37%), sulfuric acid (98%), sodium hydroxide (98%),

    tetrapropylammonium bromide (TPABr 98%), Aluminum sulfate eighteen hydrate (>99.3%), sodium

    chloride (99.5%) were supplied from Merck.

    The polymeric matrix was prepared from high relative molecular weight polyvinyl chloride

    (PVC). Reagent grade dioctylphthalate (DOP), tetrahydrofuran (THF) and

    hexadecyltrimethylammonium bromide (HDTMABr) were purchased from Aldrich or Fluka and used

    as received. Sodium and potassium salts of all anions were of the highest purity available (all from

    Merck) and used without any further purification. The pH of solutions was adjusted with sodium

    hydroxide or hydrochloric acid solution as appropriate. All other reagents used were of analytical

    reagent grade. De-ionized water was used throughout the experiments. A 0.1M stock solution of

    thiocyanate was prepared by dissolving an appropriate, accurate amount of KSCN. The standard

    1.0×10−7–1.0×10−1M solutions of thiocyanate were prepared daily by sequential dilution of the

    appropriate stock solution with doubly distilled water.

    2.2. Synthesis and modification of ZSM-5 zeolite

    2.4984 g (0.0427 mol) of sodium chloride and 0.5931 g (0.00089 mol) of hydrated aluminum

    sulfate, were dissolved in 10.1200 g (0.5622 mol) of distilled water. 1.8945 g (0.00712 mol)

    tetrapropylammonium bromide, 7.1200 g (0.3955 mol) of distilled water and 1.0883 g (0.01088 mol)

  • Int. J. Electrochem. Sci., Vol. 6, 2011

    3180

    sulfuric acid were added with vigorous stirring. 15.0000 g (0.07125 mol SiO2) of sodium silicate

    solution was subsequently added and the mixture was stirred for 60 min. The pH of the mixture was

    9.38. The mixture was transferred into the autoclave. The autoclave was kept into an oven at 110 °C

    for about 2 h. Then, the temperature was increased to 230 °C and kept at that level for another 5 h. The

    solid phase obtained was filtered out, washed with water several times, dried at 110 °C for about 12-14

    h and subsequently calcined at 540 °C for 3 h [23].

    For preparation of surfactant modified zeolite (SMZ), the Na-ZSM-5 zeolite (2 g) was mixed

    with 100 mL of 50 mM HDTMA solutions, and stirred for 24 h on a magnetic stirrer. The mixture was

    then centrifuged at 5000 rpm for 20 min and the resulting SMZ was dried in air.

    2.3. Characterization of the synthesized zeolite and the SMZ

    The X-ray diffraction patterns of zeolite ZSM-5 was determined using a Bruker diffractometer,

    D8ADVANCE, X-ray tube anode: Cu, wavelength: 1.5406 (Cu Kα radiation), filter: Ni. Fourier

    transformation infrared (FT-IR) spectra of samples, on KBr pellets, were recorded with a Nicolet

    single beam FT-IR (Impact 400D) spectrometer in the range of 400-4000 cm-1. All samples were

    stored in a desiccator until the FT-IR analysis and KBr of spectrometry grade was also heated at 25 ◦C

    to remove the adsorbed water before pressing. Then 1mg dry powdered sample was mixed with 100

    mg KBr, followed by compressing the mixture to pellets with 12 mm diameter and investigating by the

    FT-IR spectrometer.

    2.4. Electrode preparation

    The procedure to prepare the best PVC membrane was to mix thoroughly 25.5 mg of powdered

    PVC, 10.0 mg of SMZ as ionophore and 64.5 mg of plasticizer DOP. Then the mixture was dissolved

    in 3 mL of dry THF. The resulting clear mixture was evaporated slowly until an oily concentrated

    mixture was obtained. A Pyrex tube (5 mm i.d.) was dipped into the mixture for about 10 s so that a

    nontransparent membrane of about 0.3 mm thickness is formed. The tube was then pulled out from the

    mixture and kept at room temperature for about 2 h. The tube was then filled with internal solution

    (1.0×10-2 M KSCN). The electrode was finally conditioned for 16 h by soaking in a 1.0×10-2 M KSCN

    solution. A silver/silver chloride electrode was used as an internal reference electrode. The electrode

    was instantly ready for use.

    2.5. Potentiometric measurements

    The potentiometric measurements were performed with a Metrohm pH meter 691 pH/Ion

    Meter (Metrohm) at 25.0 ± 1C. A SMZ-membrane thiocyanate-selective PVC electrode and a

    standard Ag/AgCl electrode shielded by an intermediate salt-bridge compartment, containing the

    background electrolyte in order to prevent any transfer of potassium ions into the measuring solution,

    were used as an indicator electrode and an external reference electrode, respectively. All potentials

  • Int. J. Electrochem. Sci., Vol. 6, 2011

    3181

    were reported versus this reference electrode. The cells for millivolt measurements were of the

    following type:

    Ag/AgCl, KCl (3M) | internal solution KSCN (1.0 × 10-2 M) |electrode membrane | test solution

    | Ag/AgCl, KCl (3M)

    The potential reading of each solution was recorded when it became stable, and then plotted as

    a logarithmic function of SCN- anion concentration. The activities of anions were based on the

    activity-coefficient data () calculated from the modified form of Debye-Huckel Equation [24], which

    is applicable to any ion and is as follows:

    (1)

    where is the ionic strength and Z is the valency. All measurements were carried out at room

    temperature. The pH measurements were performed using a corning model 125 pH meter equipped

    with a combined electrode.

    3. RESULTS AND DISCUSSION

    3.1. Synthesized zeolite and SMZ characterization

    3.1.1. X-ray diffraction studies

    Synthesized ZSM-5 zeolite was characterized by X-ray diffraction.

    0

    2000

    4000

    6000

    6 10 20 30 40 50 60

    Figure 1. XRD patterns synthesized zeolite ZSM-5.

  • Int. J. Electrochem. Sci., Vol. 6, 2011

    3182

    Figure 1 shows representative XRD patterns of sample synthesized. The measurements were

    conducted in continuous /2 scan refraction mode. The anode was operated at 30 KV and 15 mA the

    2 angles were measured 5–60 at the rate of 2.min-1. The characteristic lines at 2θ of 7.9, 8.8, 23.1,

    23.8 and 24.5 are observed from XRD pattern that show a good agreement with the data of Na- ZSM-5

    zeolite [25]. The crystallinity of the prepared sample was calculated using the ratio of the sum of the

    areas of the most intense peaks for zeolite ZSM-5 sample (in above mentioned 2θ values) to the same

    peaks for the standard (zeolite ZSM-5) and multiplying by 100. This analysis shows that the product

    has a typical zeolite ZSM-5 structure with no amorphous material.

    3.1.2. FT-IR studies

    FT-IR lattice vibration spectra were used to investigate the influence of surfactant on the

    zeolite framework.

    Figure 2. FT-IR patterns for (a) HDTMA (b) synthesized ZSM-5 and (c) SMZ.

  • Int. J. Electrochem. Sci., Vol. 6, 2011

    3183

    Representative spectra of zeolite HDTMA, ZSM-5 and SMZ in the range of 400–4000cm−1 are

    shown in Fig. 2. According to Fig. 2b, the observed frequencies at 3436, 1623, 1098 and 551cm−1

    agrees with the infrared spectral data, which has been obtained for zeolite ZSM-5 by Flanigen et al.

    [26]. The IR pattern of SMZ-ZSM-5 shows characteristic peaks at 2919, 2856, 1478 and 687 cm–1 in

    fig. 3c, which indicate the incorporation of HDTMA into the zeolite structure. There is a slight shift in

    peaks at each wavenumber, however, both of the patterns are nearly the same. Specially, the prescence

    of peaks at 2919, 2856 and 1478 cm-1 in SMZ spectrum (related to C-H, C-C and N-C vibrations in the

    surfactants, respectively), that are absent in the Na- ZSM-5 spectrum, confirm loading of HDTMA on

    zeolite surface.

    3.2. Theoretical considerations of SMZ and electrode behavior

    As previously mentioned, the SMZ can be used for sorption of anionic species. This is based on

    the conceptual model which was illustrated graphically in the previous work [27]. At low

    concentrations, HDTMA forms a monolayer on the zeolite surface with the hydrophobic ends of the

    molecules turned toward the solution. With increasing coverage, a second HDTMA layer is also

    formed. This bilayer structure causes arrangement of positively charged functional groups toward the

    solution, which can serve as sorption sites for anions [28]. Complete formation of the second layer,

    means that sorbed HDTMA concentration is twice of the external cation exchange capacity (ECEC).

    Preliminary experiments showed that when SMZ is mixed with a solution containing

    thiocyanate anion, the concentration of thiocyanate in the remaining solution is decreased. Therefore,

    we were prompted to use the SMZ as an active ingredient for construction of a potentiometric sensor

    for this anion. The sorption mechanism can be shown by the following reaction:

    R4N+ in SMZ + ClO4

    − R4N+ ClO4

    − (2)

    where R4N+ represents the anion-exchange sites of HDTMA surfactant on SMZ. According to

    the above discussions, when proposed electrode is immersed in thiocyanate solutions, anion exchange

    reaction takes place between anions on the electrode surface (Br−) and ClO4− in solution.

    3.3. Optimization of the membrane composition

    In preliminary studies, some experiments regarding the response towards mono and divalent

    anions were carried out. Except with SCN- ion, the E versus pM plots is nonlinear within the

    concentration range of 1.0×10−1 to 1.0×10−7 M, and exhibit hypo-Nernstian slopes (table 1). Our

    preliminary experiments showed that the membrane prepared by using the zeolite ZSM-5 as ion

    sensing proved to be suitable one, and all of optimizations were done on SMZ-PVC electrode. First,

    the amount of SMZ modifier was optimized. The responses of the electrodes to several SMZ

    compositions are presented in Table 2. As can be seen, only the electrode containing 10% SMZ which

    was prepared from Na–ZSM-5 at 50 mM HDTMA, as described in section 2.2, showed Nernstian

  • Int. J. Electrochem. Sci., Vol. 6, 2011

    3184

    behavior. Electrodes which were prepared at 100 and 200 mM HDTMA solution do not show a

    Nernstian behavior.

    Table 1. Potentiometric response of the electrode toward different anions (n=4).

    Anion Slope(mV.decade-1) r2 Linear range (-log C)

    ClO4- 46.7 ± 0.8 0.9809 ± 0.08 5.2-1.1

    ClO3- 36.1 ± 0.9 0.9903 ± 0.08 5.0-3.1

    SO42- 19.2 ± 1.1 0.9393 ± 0.04 4.8-2.1

    SO32- 20.2 ± 0.9 0.9807 ± 0.07 5.1-1.9

    C2O42 17.1 ± 0.9 0.9866 ± 0.04 4.8-1.9

    I- 41.2 ± 0.8 0.9791 ± 0.08 5.1-2.6

    NO3- 18.1 ± 0.8 0.9870 ± 0.008 4.1-2.1

    HPO42- 30.4±1.4 0.9678 ± 0.009 5.1-2.0

    SCN- 60.7 ± 1.6 0.9933 ± 0.005 6.6-1.2

    Br- 34.3 ± 0.9 0.9661±0.004 4.8-2.0

    Table 2. Optimization the modifier amount.

    Electrode Modifier percentage Slope (mV. decade-1) (n=5)

    A 4%(NaZSM-5+100 mM HDTMA) -42.0 ± 1.1

    B 10%( NaZSM-5+100 mM HDTMA) -53.1 ± 0.5

    C 15%( NaZSM-5+100 mM HDTMA) -45.1 ± 0.7

    D 4%( NaZSM-5+200 mM HDTMA) -42.5 ± 0.8

    E 10%( NaZSM-5+200 mM HDTMA) -53.1 ± 0.5

    F 15%( NaZSM-5+200 mM HDTMA) -56.1 ± 0.8

    G 4%( NaZSM-5+50 mM HDTMA) -40.1 ± 1.1

    *H 10%( NaZSM-5+50 mM HDTMA) -58.7 ± 0.3

    I 15%( NaZSM-5+50 mM HDTMA) -41.1 ± 0.8

    Table 3. Composition and potentiometric characterization of thiocyanate selective electrode based on

    SMZ-PVC membrane (n=4).

    No PVC% SMZ% DOP% Avrage Slope

    (mV.decad-1)

    r2 Linear

    range(-log

    C)

    RSD

    1 29 4.0 67 53.4±1.2 0.9910 ± 0.0030 5.5-1.0 3.2

    *2 25.5 10.0 64.5 58.6±0.9 0.9960 ± 0.0190 6.0-1.1 1.7

    3 23.5 15.0 61.5 44.3±1.3 0.9711 ± 0.0200 4.4-1.1 4.3

    4 21 20.0 59 39.2±0.8 0.9701 ± 0.0141 3.9-1.2 2.9

    In this case, a partially formed bilayer may be loaded on the zeolite surface. Consequently, the

    obtained SMZ does not respond appropriately to SCN–. This completely agrees with the results which

  • Int. J. Electrochem. Sci., Vol. 6, 2011

    3185

    were obtained by Wingenfelder et al. [29] for antimonate sorption studies. They found that, in the

    treatments about two times of the ECEC, the zeolite took up over 98% of the initial HDTMA in

    solution. When the applied HDTMA concentrations were over two times as high as the ECEC, the

    zeolite took up 84% of the initial HDTMA concentration.

    The potential response of a sensor is greatly related to the membrane ingredients [30-34]. The

    ratio of membrane ingredients were optimized so that the membranes develop reproducible, noiseless

    and stable potentials. Composition of the membranes, which resulted in the best performance, is shown

    in Table 3 along with their potentiometric characteristics. As it can be seen, electrodes made from

    membranes containing 10% (w/w) zeolite in the presence of 25.5% polymer, 64.5% dioctylphthalate

    were found to be the optimum one in the development of this sensor (membrane no. 2, Table 3). Thus,

    this membrane composition was used for subsequent works.

    3.4. Effect of internal reference solution

    The concentration of the internal solution (KSCN) in the electrode was changed from 1.0×10−2

    to 1.0 ×10−4 M and the emf-pSCN− plots were obtained and the results are shown in Fig. 3.

    Figure 3. Effect of concentration of internal reference solution on the electrode response at various

    thiocyanate concentrations in the range of 1.0×10−1 M-1.0×10-3 M.

    It was found that the variation of the concentration of the internal KSCN solution did not cause

    any significant difference in potential response except for an expected change in the intercept of the

    resulting plots. A 1.0×10−2 M concentration of internal reference solution is quite appropriate for

  • Int. J. Electrochem. Sci., Vol. 6, 2011

    3186

    smooth functioning of the electrode system. Then conditioning time was studied by changing the time

    of being electrode in conditioning situation (Table 4). The obtained results showed the best Nernstian

    behavior was for 16 h.

    Table 4. Conditioning time results for SMZ-PVC membrane electrode (10% modifier; -logC = 6 - 2; )

    n=5).

    r 2 Slope (mV.decade-1) t (h)

    0.9893 ± 0.0377 38.0 ± 0.9 4

    0.9877 ± 0.0438 39.9 ± 1.6 6

    0.9699 ± 0.0069 40.9 ± 0.9 8

    0.9899 ± 0.0066 46.7 ± 0.4 10

    0.9901 ± 0.0038 49.9 ± 0.9 12

    0.9870 ± 0.0080 59.3 ± 0.9 *16

    0.9982 ± 0.0723 52.4 ± 0.9 18

    0.9191 ± 0.0132 53.8 ± 1.2 24

    0.9694 ± 0.0066 55.5 ± 0.9 48

    3.5. Effect of pH

    The influence of pH of the test solution (1.0×10−3 M thiocyanate) on the potential response of

    the SMZ-membrane sensor was tested in the pH range 1.1–13.1 (adjusted with either HCl or NaOH)

    and the results are shown in Fig. 4.

    Figure 4. Influence of pH on the potential response of the thiocyanate-selective electrode to 1.0×10-2

    M SCN-.

  • Int. J. Electrochem. Sci., Vol. 6, 2011

    3187

    As seen, the membrane electrode can be suitably used in the pH range 2.0–10.1. A pH of 6.0

    was adjusted with 0.1 mol L−1 acetic acid-acetate buffer was used for further studies. However, drastic

    potential changes are observed at higher and lower pH values. This is most probably due to the

    simultaneous response of the electrode to SCN− and OH−, at pH > 10.1, and some ionophore

    protonation and consequent diminished responsivity to SCN− and/or simultaneous response to

    thiocyanate and chloride ions, at pH < 2.0.

    3.6. Effect of ionic strength on the response of the electrode

    The effect of ionic strength (1.0×10−4 to 3.4×10-3 M NaNO3) on the calibration curve of the

    SMZ-PVC electrodes was investigated and the obtained results are listed in Table 5.

    Table 5. Effect of ionic strength on optimized SMZ-PVC membrane electrode response.

    Concentration of

    NaNO3 (M)

    Slope= x (n=6) (mV.decade-1)

    texp Linear range of thiocyanate

    (M)

    Intercept

    (mV)

    1.0×10-4 -59.1±0.56 0.0.40 1.00×10-6-1.10×10-1 236

    3.0×10-4 -59.5±0.76 1.18 1.00×10-6-1.10×10-1 235

    4.9×10-4 -59.3±0.61 0.73 1.00×10-6-1.10×10-1 233

    1.0×10-3 -59.5±0.66 0.95 1.00×10-6-1.10×10-1 232

    2.0×10-3 -59.7±0.78 1.43 1.00×10-6-8.97×10-2 197

    3.4×10-3 -59.3±0.67 0.33 1.00×10-6-8.94×10-2 190

    Note: t 0.05, 5 =2.57 (Ref. 35)

    The electrode response has a Nernstian slope with linear range of 1.00×10-6-1.10×10-1 of

    thiocyanate anion within the 1.0×10−4 to 1.0×10−3 M of NaNO3. This was indicated by the constant

    Nernstian behavior and the same linearity range obtained at different ionic strength values.

    Nevertheless, within the 2.0×10−3 to 3.4×10−3 M NaNO3, the Nernstian slopes were obtained but the

    linear range decreased to 1.00×10-6-8.97×10-2M of thiocyanate anion. In higher concentrations

    (>1×10−3 M NaNO3, results not showed) it is possible that some of the ion exchange sites in SMZ are

    occupied by NO3− ions. As the result it causes a decrease in electrode response compare to the higher

    concentrations of the thiocyanate anions. The difference in the intercept values at nitrate

    concentrations lower than 1 mM is not significant. The small differences indicate that the sensor is free

    from nitrate interference at concentrations of nitrate lower than 1 mM. But, the intercept values at

    nitrate concentrations higher than 1 mM are more different and nitrate has much more interference.

    The validity of obtained results was performed by statistical methods. By comparing of texp and t0.05,5 =

    2.57 at 95% confidence interval it is concluded that there is a random error in experiments [35].

  • Int. J. Electrochem. Sci., Vol. 6, 2011

    3188

    3.7. Effect of temperature

    To investigate the thermal stability of the electrode, calibration curves (Ecell versus log aSCN-)

    were constructed at different temperatures covering the range 25–60 ◦C. The electrode exhibited good

    Nernstian behavior in the range of 25–45 ◦C. The results are shown in Table 6.

    Table 6. Effect of temperature on optimized SMZ-PVC electrode behavior.

    Temperature ( C ) Nernstian slope

    (mV.decade-1)

    Mean of exp. slope

    (n=10), mV.decade-1)

    texp

    25 -59.2 -59.6 ± 0.7 1.81

    30 -60.2 -60.0 ± 0.8 0.79

    35 -61.2 -61.5 ± 0.5 1.90

    40 -62.2 -62.6 ± 0.6 2.10

    45 -63.2 -63.6 ± 0.7 1.81

    50 -64.2 -65.1 ± 0.6 4.74

    55 -65.0 -67.2 ± 1.3 5.35

    60 -66.0 -67.9 ± 0.9 6.67

    Note: t 0.05, 9 =2.26 (Ref. 35)

    Experimental slope calculated from Eq. (3) which is applicable to any temperature:

    (3)

    where n is the valancy, R, T and F are the universal gas constant, the absolute temperature and

    the Faraday constant.

    The Nernstian and experimental slope values at each temperature were compared statistically,

    showing conformity of the slopes for 10 replicates. The results prove the validation of the obtained

    results with the expected values in the temperature range of 25–45 ◦C. However, at temperatures higher

    than 45 ◦C the slopes show a significant deviation from the theoretical values. This deviation may be

    related to desorption of the surfactant from SMZ and destruction of the electrode surface.

    3.8. Response time

    The response time of the sensor was defined as t95 for the slope of the calibration curve of

    thiocyanate solution when the thiocyanate concentration was increased from 1.0×10−7 to 1.0×10−1 mol

    L−1 (Fig. 5), where t95 is the time required for the sensor to reach 95% of the steady state (final signal

    that does not change during 60 s) potentiometric value. From the results, the best response time of 5 s

    was recorded for membrane having the optimized conditions. After new solutions were exposed to the

    electrode the response changed rapidly and remained at a constant value before 5 s of the exposure.

  • Int. J. Electrochem. Sci., Vol. 6, 2011

    3189

    The potential generated by the membrane (when it’s remained constant in one solution of thiocyanate)

    remained constant for 5 min, after which it started slow deviating (−0.5% per min).

    Figure 5. Response time of the SMZ-PVC to different concentration of thiocyanate (pH=5.5) in the

    range of 1×10−7 to 6×10−1 M (in sequence of “g” to “a” in figure), (Inset: Static response time

    of the proposed membrane electrode towards SCN- ions)

    The static response time of the membrane electrode thus obtained was

  • Int. J. Electrochem. Sci., Vol. 6, 2011

    3190

    and interfering ions. The potentiometric selectivity coefficients of the proposed electrode, in the

    optimized analytical procedure, are calculated using the following equation [37] and summarized in

    Table 7:

    (4)

    Where is the potentiometric selectivity coefficient, Ei the measured cell potential at

    primary ion activity ai (SCN-), Ej the measured cell potential at interference ion activity aj, n and m are

    the charges of the primary and interference ion, respectively. R, T and F have their common meaning.

    Table 7. Potentiometric selectivity coefficients of various interfering anions (n=4).

    Interfering anion logPot

    ClOK

    4

    ± Interfering anion logPot

    ClOK

    4

    ±

    ClO4- -0.9±0.08 CN- -3.1±0.09

    I- -1.2±0.05 CO32- -2.6±0.06

    IO4- -2.1±0.07 BrO3- -3.0±0.11

    ClO3- -2.8±0.07 IO3- -3.1±0.21

    C2O42- -2.9±0.09 Cl- -3.3±0.11

    NO3- -1.9±0.10 Br- -2.5±0.10

    CH3COO− -3.4±0.07 Salicylate -2.4±0.08

    ClO4- > I- > NO3

    - > IO4- > Salicylate> Br- > CO3

    2-> ClO3-> C2O4

    2- > BrO3->IO3

    - = CN-

    As can be seen, of the different anions tested, the largest sensitivity was obtained for

    thiocyanate, perchlorate, iodide, nitrate and periodate. a poorer response was observed with salicylate

    and the responses to other anions were practically negligible.

    The selectivity sequence follows the Hofmeister Series (the selectivity based solely on

    lipophilicity of anions) [38], (ClO4− > SCN− > I− >NO3

    − >Br− >Cl− >HCO3− >CH3COO

    − >SO42−

    >HPO42−) although some slight deviations might be interpreted because of some preferential

    coordination of this ionophore with certain anions.

    3.10. Response characteristics of the electrode

    Using the optimized composition and conditions described above, the critical response

    characteristics of the electrode were assessed according to IUPAC recommendations [39]. The emf

    response of the electrode to varying concentrations of thiocyanate species indicates a linear range from

    1.00×10-6 to 1.10×10-1 M (r = 0.9966) (Fig. 6). The average slope of calibration curves based on 14

    replicate measurements was −58.9 ± 0.9 mV per decade of SCN− concentration. Comparison of this

    calculated value with theoretical Nersntian value (59.2 mV per decade of SCN− concentration) by

    statistical (t) parameter (tcalc = 1.87 and t 0.05,13 =2.16) shows that the difference between them is due to

  • Int. J. Electrochem. Sci., Vol. 6, 2011

    3191

    random errors. On the other hand, the proposed SMZ membrane electrode is applicable to measure

    SCN− anion in aqueous solutions. The lower and upper detection limits were 8.33×10−7 and 0.13 M,

    respectively.

    Figure 6. Potentiometric response of the proposed electrodes toward thiocyanate.

    Long term stability of the electrode was investigated by taking the response of the optimized

    electrode in a period of 30 days. Potentiometric measurements were performed every day during the

    first week and then every three days for the rest of the period. The electrode retained its full activity

    during this period. The relative standard deviations (RSD) of a single electrode (within electrode

    variation) were determined every 30 s during a period of 0-120s, (n=5). This process was repeated 10

    times. The RSD values were between 0.5 and 1.4%. The small relative standard deviations (0.2–1.0%)

    and hence good repeatability of the electrodes indicates that the PVC membrane is homogeneous.

    The reproducibility of the potentiometric response of the optimized electrode was also

    investigated by carrying the measurements on four independent electrodes with the same composition.

    These measurements were also performed in 10 replicates. The obtained relative standard deviation

    (between electrode variations) was less than 1.5%. The validity of results was investigated by “g”

    value statistical test, g-test [35], as:

    22

    2

    2

    1

    2

    ... ksss

    shighestg

    (5)

    Where is the variance of each class of determinations. The calculated “g” values for each

    case was 0.4793 which are smaller than the critical g value at 95% confidence interval, g0.05,10,4 =

  • Int. J. Electrochem. Sci., Vol. 6, 2011

    3192

    0.5017 (4 class, 10 replicate in each class) [35] indicating that there is no significant difference

    between the behavior of the four independent electrodes and those compositions are homogeneous.

    Table 8. Comparison of some characteristics of the proposed electrode with previous SCN− ISEs.

    Ref. Slope

    (mVdecade−1)

    Working

    pH range

    Linear range (M) DL (M) Ionophore

    Proposed

    electrode

    −58.9 ± 0.9 2.0–10.1 1.00×10-6–1.01×10−1 8.33×10−7 SMZ

    [40] -56.3 2.0-8.0 6.0×10-6–1.0×10−2 5. 0×10−6 Tetra-midepyridine

    [40] -58.3 2.0-8.0 1.0×10-5–1.0×10−2 5. 0×10−6 Hexa-imidepyridine

    [41] -57.0 5.0 9.0×10−7–1.0×10−1 7.0×10−7 [Cu(II)2-BBSTA]

    [42] -56.5 3.5–8.5 1.0×10−6–1.0×10−1 7.0×10−7 [Zn(BDSH)]

    [43] -58.7 3.0–8.0 1.0×10−5–1.0×10−1 4.0×10−6 Rh[(trpy)(bpy)Cl](PF6)2

    [44] -57.6 3.9–9.2 1.0×10−6–1.0×10−1 5.0×10−7 CCT

    [44] -58.7 4.3–9.8 1.0×10−6–1.0×10−1 5.7×10−7 NOBP

    [45] -52.0 2.0–3.0 3.0×10−5–1.0×10−1 2.5×10−5 Calix[4]arene

    Some response characteristics of the proposed electrode compared with other thiocyanate

    selective electrodes and the obtained results are summarized in Table 8. As can be seen from Table 8,

    in some cases, selectivity coefficient and dynamic linear range that were obtained for the proposed

    electrode are superior and comparable to those reported previously for other thiocyanate-selective

    membrane electrodes based on variety of different ionophores. It is important to note that in all cases

    in Table 8, response time, stability and electrode response independent pH range for the proposed

    SMZ-PVC are much better than those reported for other compared electrodes. The SMZ-PVC has

    applicability in a wide pH range (2.0–10.1), for acidic to basic solutions. High stability of the electrode

    is another advantage of proposed electrode that it in turn causes an increasing of precision. In

    investigation of reproducibility and repeatability, the RSD values for our SMZ-PVC (1.3–2.4%) are

    smaller than others (2–5%). This shows the SMZ-PVC has a better precision.

    3.11. Practical application

    The optimized membrane thiocyanate electrode was successfully used for potentiometric

    determination of SCN− via Ag+ titrations. Fig. 7 shows the titration curve of 10mL of a 1.0×10−2 M

    thiocyanate solution titrated with 1×10-2 M of AgNO3. The titration volume was measured at the

    inflection point of the titration curve and a concentration of thiocyanate of 9.7×10-3M was determined.

    The new thiocyanate -selective electrode was applied for the determination of thiocyanate in

    river and tap water samples with satisfactory results. The analysis was performed by using the standard

    addition technique. The results are summarized in Table 9. Good recoveries were obtained in all

    samples.

  • Int. J. Electrochem. Sci., Vol. 6, 2011

    3193

    Figure 7. Potentiometric titration curve for 10 ml of 1.0× 10−2 M SCN- solution with 1.0 ×10−2 M

    AgNO3 using the proposed sensor as an indicator electrode (n=5).

    Table 9. Determination of thiocyanate in drinking and river water.

    Samples Added thiocyanate

    (ml/gμ)

    Foundet thiocyanate

    (ml/gμ)

    Recovery(%)

    Tap Water of Isfahan - 0 -

    1.4 0.21±1.33 95.0

    2.3 0.44±2.15 93.5

    3.0 0.32±2.92 97.3

    5.3 0.45±5.23 98.1

    The River Water of

    Zayandeh Rood in

    Isfahan

    - 0 -

    10.1 0.23±10.16 100.6

    15.3 0.55±15.14 99.3

    20.2 0.35±20.0 1 99.1

    25.4 0.88±25.12 98.9

    4. CONCLUSIONS

    To sum up, we have shown that the use of the modified zeolite by cationic surfactant can be

    used as suitable ionophores in the development of new thiocyanate-selective PVC membrane

    electrodes. These electrodes show a Nernstian response, detection limits of near 8.33×10−7 M and a

    low response time of a few seconds in the presence of thiocyanate. The optimized membrane electrode

    was used as indicator electrode in potentiometric titrations of thiocyante with silver.

  • Int. J. Electrochem. Sci., Vol. 6, 2011

    3194

    ACKNOWLEDGMENT

    We wish to acknowledge the financial support from the Research Council of Islamic Azad University,

    Shahreza Branch.

    References

    1. V. K. Gupta, R. N. Goyal, and R. A. Sharma, Int. J. Electrochem. Sci., 4 (2009) 156 2. S. S. Hassan, M. H. Abou Ghalia, A. G. E. Amr, and A. H. K. Mohamed, Anal. Chim. Acta, 482

    (2003) 9

    3. T. Poursaberi, M. Salavati-Niasari, S. Khodabakhsh, L. Haji-agha Babaei, M. Yousefi, S. Rouhani, and M. R. Ganjali, Anal. Lett., 34 (2001) 2621

    4. M. R. Ganjali, T. Poursaberi, F. Basiripour, M. Salavati-Niasari, M. Yousefi, and M. Shamsipur, J. Anal. Chem., 370 (2001) 1091

    5. M. K. Amini, A. Rafi, M. Ghaedi, M. H. Habibi, and M. M. Zohory, Microchem. J., 75 (2003) 143 6. A.Abbaspour, M. A. Kamyabi, A. R. Esmaeilbeig, and R. Kia, Talanta, 57 (2002) 859 7. M. M. Ardakani, A. A. Ensafi, M. Salavati-Niasari, and S. Mihoseini Chahooki, Anal. Chim. Acta,

    462 (2002) 25

    8. X. Huang, Y. Chai, R. Yuan, X. Wang, and Q. Li, Anal. Sci., 20 (2004) 1185 9. M. R. Ganjali, M. Yousefi, M. Javanbakht, T. Poursaberi, M. Salavati-Niasari, L. Haji-agha

    Babaei, E. Latifi, and M. Shamsipur, Anal. Sci., 18 (2002) 887

    10. J. Y. Dai, Y. Q. Chai, R. Yuan, Y. S. Zhang, Y. Liu, X. Zhong, and D. P. Tang, Chem. Lett., 34 (2005) 62

    11. T. Poursaberi, M. Hosseini, M. Taghizadeh, H. Pirelahi, M. Shamsipur, and M. R. Ganjali, Microchem. J., 72 (2002) 77

    12. M. R. Ganjali, T. Poursaberi, M. Hosseini, M. Salavati-Niasari, M. Yousefi, and M. Shamsipur, Anal. Sci., 18 (2002) 289

    13. M. R. Ganjali, A. Daftari, F. Mizani, and M. Salavati-Niasari, Bull. Korean Chem. Soc., 24 (2003) 14. M. Shamsipur, M. Yousefi, M. Hosseini, M. R. Ganjali, H. Sharghi, and H. Naemi, Anal. Chem.,

    73 (2001) 2869

    15. M. Shamsipur, M. Yousefi, M. R. Ganjali, T. Poursaberi, and M. Faal-Rastgar, Sensor Actuator B, 82 (2002) 105

    16. M. R. Ganjali, V. Akbar, A. Daftari, P. Norouzi, H. Pirelahi, and A. Mouradzadegan, J. Chin. Chem. Soc., 51 (2004) 309

    17. M. R. Ganjali, M. Ghorbani, A. Daftari, P. Norouzi, H. Pirelahi, and H. D. Dargahani, Bull. Korean Chem. Soc., 25 (2004) 172

    18. M. R. Ganjali, F. Mizani, H. Pirelahi, M. Taghizadeh, and J. Amani, Can. J. Anal. Sci. Spect., 47 (2002) 157

    19. M. R. Ganjali, M. R. Pourjavid, M. shamsipur, T. Poursaberi, M. Rezapour, M. Javanbakht, and H. Sharghi, Anal. Sci., 19 (2003) 995

    20. M. R. Ganjali, M. Yousefi, T. Poursaberi, L. Naji, M. Salavati-Niasari, and M. Shamsipur, Electroanalysis, 18 (2003) 1476

    21. M. Shamsipur, M. Javanbakht, A. R. Hassaninejad, H. Sharghi, M. R. Ganjali, M. F. Mousavi, Electroanalysis, 15 (2003) 1251

    22. M. R. Ganjali, M. Rezapour, M. R. Pourjavid, and M. Salavati-Niasari, Anal. Sci., 19 (2003) 1127 23. M. Ghiaci, F. Seyedeyn-Azad, and R. Kia, Mater. Res. Bull., 39 (2004) 1257 24. R. Buck, and V. Cosofret, Pure Appl. Chem., 65 (1993) 1849 25. H. Robson, Verified Syntheses of Zeolitic Materials, Elsevier, Amsterdam, (2001) 26. E. M. Flanigen, H. Khatami, and H. A. Szymanski, Molecular Sieve Zeolites Advances in

    Chemistry Series, vol. 101, American Chemical Society, Washington, D. C., (1971)

  • Int. J. Electrochem. Sci., Vol. 6, 2011

    3195

    27. A.Nezamzadeh-Ejhieh, and N. Masoudipour, Anal. Chim. Acta, 658 (2010) 68 28. Z. Li, and R. S. Bowman, Environ. Sci. Technol., 31 (1997) 2407 29. U. Wingenfelder, G. Furrer, and R. Schulin, Micropor. Mesopor. Mat., 95 (2006) 265 30. M. R. Ganjali, P. Norouzi, A. Atrian, F. Faridbod, S. Meghdadi, and M. Giahi, Mater. Sci. Eng. C,

    29 (2009) 205

    31. V. K. Gupta, A. K. Jain and G. Maheshwari, Int. J. Electrochem. Sci., 2 (2007) 102 32. M. R. Ganjali, R. Nemati, F. Faridbod, P. Norouzi, and F. Darviche, Int. J. Electrochem. Sci., 3

    (2008) 1288

    33. V. K. Gupta, R. N. Goyal and R. A. Sharma, Int. J. Electrochem. Sci., 4 (2009) 156 34. V. K. Gupta, M. Al Hayat, A. K. Singh and M. K. Pal, Anal. Chim. Acta, 634 (2009) 36 35. R. L. Anderson, Practical Statistics for Analytical Chemists, Van Nostrand Reinhold, New York,

    (1987)

    36. G. Dimeski, Clin. Biochem. Rev. Suppl (i), 29 (2008) 43 37. Y. Umezawa, P. Buhlmann, K. Umezawa, K. Tohda, and S. Amemiya, Pure Appl. Chem., 72

    (2000) 1851

    38. F. Hofmeister, Arch. Exp. Patol. Pharmakol., 24 (1988) 247 39. G. G. Guibault, R. A. Durst, M. S. Frant, H. Freiser, E. H. Hansen, T.S. Light, E. Pungor, G.

    Rechnitz, N. M. Rice, W. Simon, and J. D. R. Thomas, Pure Appl. Chem., 48 (1976) 127

    40. S. S. M. Hassan, M. H. A. Ghalia, A. G. E-Amr, and A. H. K. Mohamed, Anal. Chim. Acta, 482 (2003) 9

    41. R. Yuan, X. L. Wang, L. Xu, Y. Q. Chai, Z. Y. Sun, X. Q. Huang, X. Q. Li, Q. Zhao, and L. Zhou, Electrochem. Commun., 5 (2003) 717

    42. M. Mazloum-Ardakani, A. Sadeghi, and M. Salavati-Niassari, Talanta, 66 (2005) 837 43. M. Shamsipour, S. Ershad, N. Samadi, A. R. Rezvani, and H. Haddadzadeh, Talanta, 65 (2005)

    991

    44. M. R. Ganjali, H. A. Zamani, and F. Malekzadegan, Anal. Chim. Acta, 555 (2006) 336 45. S. Erden, A. Demirel, S. Memon, M. Yılmaz, E. Canel, and E. Kılıç, Sensor Actuator B, 113

    (2006) 290

    © 2011 by ESG (www.electrochemsci.org)

    http://www.electrochemsci.org/

Recommended