1 HONKON-M40E+ MANUAL BEIJING HONKON TECHNOLOGIES CO., LTD.
Transcript
Page 1: HONKON-M40E+ MANUAL M40E+.pdf · reduction and ease anti-aging. . 2.2 Application: 2.3 Contraindications 1. Customers taking drugs which cause photosensitivity, eg: tetracycline;

1

HONKON-M40E+

MANUAL

BEIJING HONKON TECHNOLOGIES CO., LTD.

Page 2: HONKON-M40E+ MANUAL M40E+.pdf · reduction and ease anti-aging. . 2.2 Application: 2.3 Contraindications 1. Customers taking drugs which cause photosensitivity, eg: tetracycline;

2

INDEX

1. General Description of Structure

1.1 Front View

1.2 Side View

1.3 Back View

2. Treatment theory, application and contraindications

2.1 Treatment Theory

2.2 Application:

2.3 Contraindications

3. Installation

3.1 Installation procedure

3.1.1 Installation of the handle holder

: 3.1.2 Installation of the treatment handle

3.1.3 Installation of the foot pedal

3.1.4 Connecting the power line

3.1.5 Affusing water:

3.1.6 The CPC connector

3.2 Precautions for installation

4. Operation

4.1 Turn on the machine

4.2 Introduction of the interface

4.2.1 The original interface

4.2.2 Operation interface

4.2.3 Warning information

4.2.4 Correct the LCD

Page 3: HONKON-M40E+ MANUAL M40E+.pdf · reduction and ease anti-aging. . 2.2 Application: 2.3 Contraindications 1. Customers taking drugs which cause photosensitivity, eg: tetracycline;

3

4.3 Operation Process

4.3.1 Before the Operation

4.3.2 During the operation

4.3.3 Operation technique

4.3.4 After the operation

4.4 Operating Precautions

4.5 Adverse reflection

4.5.1 RF

4.5.2 IPL

4.6 Maintenance after operation and notice:

5. Daily Maintenance

6. Transport and Storage

7. Possible Failures and Problem-solving

8. Technical parameter

Page 4: HONKON-M40E+ MANUAL M40E+.pdf · reduction and ease anti-aging. . 2.2 Application: 2.3 Contraindications 1. Customers taking drugs which cause photosensitivity, eg: tetracycline;

4

1. General Description of Structure

1.1 Front View

1.2 Side View

1.3 Back View

① Handle and handle holder ② TFT touch LCD ③ Emergency switch ④ Key switch

① Handle holder

② Ventilator

① Water input ② Handle socket ③ Water output ④ Creepage protector ⑤ Foot pedal socket ⑥ Power socket ⑦ Overflow port ⑧ Cathode socket

② ③

Page 5: HONKON-M40E+ MANUAL M40E+.pdf · reduction and ease anti-aging. . 2.2 Application: 2.3 Contraindications 1. Customers taking drugs which cause photosensitivity, eg: tetracycline;

5

2. Treatment theory, application and contraindications 2.1 Treatment Theory E-light: HONKON E-light is a high-intelligent and non-ablative technology that simultaneously harnesses the power of radio frequency (RF) and optical energy (IPL). Ultra optical energy causes the skin to initiate a natural healing process that rebuilds new collagen and removes broken capillaries and pigmented lesions. The radio frequency excites water molecules from which the resultant heat decreases the pore size, tightens the skin, and provides superior wrinkle reduction and smoother skin texture-usually immediately. Cooling RF: Cooling RF is radio frequency which can penetrate epidermis into deep dermis. The radio frequency motivates water molecules to move at high speed, then produces heat and the temperature reaches to 45-65℃.This leads to subsequent contraction of the collagen tissue and makes the loose skin and wrinkle tightened immediately. Moreover, the treatment effect causes collagen re-growth and reconstruction. Meanwhile, it renovates old and damaged collagen; realizes the immediate effect of skin tightening, long-term wrinkle reduction and ease anti-aging.

. 2.2 Application:

2.3 Contraindications 1. Customers taking drugs which cause photosensitivity, eg: tetracycline; 2. Lesions with no firm diagnosis; 3. Those with heart pacemaker or relative devices; 4. Those suffering from serious diabetes, hypertension, heart disease and epilepsy; 5. After using the anticoagulant in two weeks; 6. Cicatricial diathesis; 7. Pregnant women; 8. immediately after or before sun insolating; Attention: During RF treatment, forbid to ware metal ornamentals (necklace, finger ring etc.)

RF: Wrinkle reduction Skin tightening Body shaping Fat dissolving

E-light Hair removal Pigmentation lesions Skin rejuvenation Vascular lesions Acne treatment

Page 6: HONKON-M40E+ MANUAL M40E+.pdf · reduction and ease anti-aging. . 2.2 Application: 2.3 Contraindications 1. Customers taking drugs which cause photosensitivity, eg: tetracycline;

6

3. Installation

3.1 Installation procedure 3.1.1 Installation of the handle holder Take out the handle holder, install it at the corresponding place of the machine. As follow the picture:

3.1.2 Installation of the treatment handle and Filters Before installation, first put down the two seal caps from the connection of treatment handle “Socket”, then press the clip springs as picture shown and plug the right handle upon treatment into the “Socket”. A locking crack will sound then pout the handle in the holder. As follow the picture:

The place of the handle holder

Small handle with three filters (Spot size:8x40mm2 ) HR: 610-1200nm SR:530-1200nm VR: 585-1200nm

Big handle for Hair removal (Spot size: 15x60mm2 )

Cooling RF handle (Monopolar: 3 treatment tips) (Bipolar: 2 treatment tips)

Page 7: HONKON-M40E+ MANUAL M40E+.pdf · reduction and ease anti-aging. . 2.2 Application: 2.3 Contraindications 1. Customers taking drugs which cause photosensitivity, eg: tetracycline;

7

Install Filter 3.1.3 Installation of the foot pedal Plug the pedal switch into the “Foot Panel” with the breach upwards and screw down the nut clockwise. As follow the picture:

3.1.4 Connecting the power line Plug the power line into the “Power Socket”, check it is plugged firmly and then buckle down the metal clip to prevent cord from loosening. As follow the picture:

3.1.5 Installing the cathode patches:

Connect the lines in the cathode patches and plug the other end into the “Cathode Socket” When do treatment

with Monopolar handle , please use one or two pieces of Cathode patches to contact with patient’s back or leg

or hands .

(Breach)

Page 8: HONKON-M40E+ MANUAL M40E+.pdf · reduction and ease anti-aging. . 2.2 Application: 2.3 Contraindications 1. Customers taking drugs which cause photosensitivity, eg: tetracycline;

8

3.1.5 Affusing water:

1. First put down the clip spring at the “Overflow”(as the picture), then plug the connector of the CPC pipe into the “Overflow”. In order to protect the overflowed water to make the floor wet, suggest to put a water container under the end of the overflow pipe.

Page 9: HONKON-M40E+ MANUAL M40E+.pdf · reduction and ease anti-aging. . 2.2 Application: 2.3 Contraindications 1. Customers taking drugs which cause photosensitivity, eg: tetracycline;

9

2. Connect the affusion pipe in the “Refill” and infuse about 1800ml pure of distilled water until there is water leak from the “Overflow”.

3. Switch on the machine: First turn the emergency switch clockwise until it is flip, and switch on the creepage protector. Then insert the key into “Key Switch” and turn it clockwise 90°to start the machine.

Creepage protector Emergency switch

Remark :The instrument instantaneous can produce the charging current in the circular telegram, in the partial rooms leaks the protector to be able to have the protection, which belongs to the normal phenomenon, then put the leak the protector to reposition 4. About 5 minutes after starting, please turn off the machine and infuse water again because the treatment handle and the radiating system will store certain water. Make sure the seal caps are removed before using “Overflow” and “Refill”. This machined is equipped with semiconductor cooling system, please avoid using with water. When the machine is used frequently, it is better to change water per 60 days. (How to change the water: 1.open the “Drain” to release all of the water; 2.close “Drain” and infuse 1/3 water of total; 3.start machine for 3-5 minutes then turn off again; 4. drain all of the water again; 5. Close “Drain” and infuse water again.)

Page 10: HONKON-M40E+ MANUAL M40E+.pdf · reduction and ease anti-aging. . 2.2 Application: 2.3 Contraindications 1. Customers taking drugs which cause photosensitivity, eg: tetracycline;

10

3.1.6 The CPC connector

1. Insert the CPC into overflow hole, if there has water in the machine, the water will flow out and the overflow hole is open.

2. Before take down the CPC,press the clip spring and the CPC connector bounce out from the overflow hole, if there has water in the machine , the water can’t flow out , the overflow hole is closed.

3.2 Precautions for installation 1. Place the machine smoothly in a clean room with temperature 5℃-35℃, humidity≤85%, without strong magnetic field etc. 2. Put the machine in a drafty place, avoid sunshine directly. 3. Before using please check that the power supply will match with our specification according to machines’ technical parameters. 4. The handles are equipped with precise optical apparatus, please avoid from impacting or drop. 5. The instrument various ties should peg grafts the arriving, also the attentions do not insert the wrong connection or empty meets; 6. Only pure or distilled water is allowed. Make sure the second effusion is done when use the machine or change the water for the first time.

4. Operation

4.1 Turn on the machine First please switch on the creepage protector, and turn the emergency switch clockwise until it is flip, to

make sure that they are in the position of working. Then plug the power cord into a 3-pin earthed socket and insert the key into “Key Switch” and turn it clockwise 90°to start the machine. The original interface will be shown.

Page 11: HONKON-M40E+ MANUAL M40E+.pdf · reduction and ease anti-aging. . 2.2 Application: 2.3 Contraindications 1. Customers taking drugs which cause photosensitivity, eg: tetracycline;

11

4.2 Introduction of the interface 4.2.1 The original interface (Picture-1) Turn on the machine after 8 seconds, enter the original interface:

(Picture-1) 4.2.2 Operation Screen (Picture-2-3)

Counter clear, this is for clear the machine’s total counter. Input the password by or

Temperature: display the Machine’s inner temperature, when the temperature is on 65 oC, the machie will have a sound of alarm and stop working automatically

Water Cycle: displaying the water cycle of the machine, if the water cycle is not good or without water

cycle, there will have alarm sound and the wheel of water cycle label will stop.

Decrease / Increase: All parameter adjust by the two buttons

Page 12: HONKON-M40E+ MANUAL M40E+.pdf · reduction and ease anti-aging. . 2.2 Application: 2.3 Contraindications 1. Customers taking drugs which cause photosensitivity, eg: tetracycline;

12

E-light function:

the three functions work with normal handle F+E (three filters changeable),

which one is chose , the button color will become from to

which work with Big handle F+E+, when it is chose, the button color will come to IPL Parameter:

IPL Energy, Time, Number “, if which one is chosen, the button color will become from to

“First Pulse, First delay, Sub pulse, Sub delay”, if which one is chose, the button color will become

from to ,

Page 13: HONKON-M40E+ MANUAL M40E+.pdf · reduction and ease anti-aging. . 2.2 Application: 2.3 Contraindications 1. Customers taking drugs which cause photosensitivity, eg: tetracycline;

13

when it is chose , the button color will come to

on this state ,all parameter can be adjust, when it is chose , the button color will come to

on this state, the handle light is on, the machine is ready for work, all parameters can’t be adjust.

Press the switch button on handle or foot pedal, the machine is on working state, will change to

IPL Energy: adjustable range 20-50J, one step is1J; Time: adjustable range 3-5S, one step is 1s; Number: adjustable range 1-10T, one step is1T; First pulse duration: adjustable range 0.1-10.0ms, one step is 0.1ms;

First pulse delay: adjustable range 0.1-20.0ms, one step is 0.1ms; Sub-pulse duration: adjustable range 0.1-10.0ms, one step is 0.1ms; Sub-pulse delay: adjustable range 0.1-20.0ms, one step is 0.1ms;

Simmer time: adjustable range 100-990ms, one step is10ms;

RF parameter for E-light function :

“RF Energy, RF time” if which one is chosen, the button color will become from to

.

On this state, Just IPL is working,

On this state, IPL and RF work together.

Page 14: HONKON-M40E+ MANUAL M40E+.pdf · reduction and ease anti-aging. . 2.2 Application: 2.3 Contraindications 1. Customers taking drugs which cause photosensitivity, eg: tetracycline;

14

Cooling Bipolar and Monopolar RF Operation Screen

On this State, All IPL parameters can’t be adjust. Just RF parameters can be adjust (Picture3)

“RF Energy, RF Duration, RF Delay“ if which one is chose, the button color will become from

to

RF Energy: Adjust range 1-50 RF Duration: Adjust range 0.1-20s RF Delay: Adjust range 0.1-20s

On this state, there has no RF output on handle, just cooling

On this state, The Cooling RF is ready for work, press the switch on handle or foot pedal,

The machine is on working state. Treatment function: there are five functions “B-HR(big handle for hair removal) , Hair removal(small handle), Skin rejuvenation, Vascular removal ,Cooling RF(bipolar and monopolar) ” according to the different treatment to chose the different spectrum treatment handle to install.

Page 15: HONKON-M40E+ MANUAL M40E+.pdf · reduction and ease anti-aging. . 2.2 Application: 2.3 Contraindications 1. Customers taking drugs which cause photosensitivity, eg: tetracycline;

15

4.2.3 Warning information 1. The high temperature alarm: When the operating time excessively is long or other reasons creates the instrument internal temperature to surpass 65 , the temperature cannot read, and is accompanied by ℃

continuing to alarm with the sound. Now, suggested to turn off the instrument, and wait the instrument to cool then continue to work again. If the instrument is being at the active status, will be automatic by the active status to return the readiness for action, then wait the temperature to restore and continue to use. 2. Water flow alarm: When there is no the current of water, the water cycle is not unimpeded or the circulation water volume is insufficiency, which can appear to alarm. The water wheel stops rotating and is accompanied by continuously alarm with the sound. If the instrument will be at the treatment condition, will return to the readiness for action, and the current of water restores may enter the normal work condition automatically. This time suggested to the exam current of water circulating normally or not, water cycle quantity is sufficient or not, and making the corresponding processing. If is the fluent switch has the question, please promptly contact with the after-sales service department. 3. The simmer fail to alarm: When simmer is not successful, which can alarm with the sound, and return to the readiness for action . 4. The thermal resistance broken line to alarm: When the DS18B20 installment or the work is not normal, can appear to alarm, the performance alarm is same to the high temperature. Now must contact with the company, inspect DS18B20 whether it is installed or replaced DS18B20. 5. The input password alarm wrongly: First elimination counter must input the password, the password input mistake, can alarm with the sound, and input the password figure value automatic reset, must input again. 6. Not selected any treatment option under the condition (already not simmer IPL, also has not selected the RF function), pressing the button on handle switch or steps on foot pedal switch, can send out reporting to the police with the sound. Now, loosen the handle switch or foot pedal switch, carry on according to the normal operation. 7. Press the button on handle switch or step on foot pedal switch at the same time starting, also can send out reporting to the police the sound, loosens the button and foot pedal can start again. 8. Choose E-energy can work when the skin is not contacted, also can send out reporting to the police with the sound. Shut down the treatment switch, and then contact the skin well to carry on the operation again. 4.2.4 Correct the LCD

Press the touch LCD under the turning off the machine, then turn on the machine, loose the touch LCD when listening to buzz once, wait for 1-2 seconds and then enter orientation starting interface. Orientation starting interface is white screen, there are “+” signed on the top left corner and down right corner, click the top left corner and down right corner in turn to realize the orientation.

Notice: Don’t need to orienta again after the first orientation, need orientation again only when the touch LCD orientation can not work or not right.

Page 16: HONKON-M40E+ MANUAL M40E+.pdf · reduction and ease anti-aging. . 2.2 Application: 2.3 Contraindications 1. Customers taking drugs which cause photosensitivity, eg: tetracycline;

16

4.3 Operation Process 4.3.1 Before the Operation: Consultation →observation →build up the treatment record and take photos for filing →select the treatment handles → clean →clear skin →start the machine→daub the gel→ wear protecting goggle and eye-patches →adjust the parameters →do a testing

Consultation:

Have a complete communicate with the customer, asking about his/her skin problem and condition, his/her endurance to pain, possibility to attend an important social activity and expectation to the effect, learning his/her skin condition and psychology to judge the expected effect after the first treatment. Make sure the customer is without contraindication

Observation:

Observe the customer’s skin character as well as the features, distribution and level of pathological changes according his/her description. Decide the Rx for customer after analyzing the face dimension. Build up the treatment record and take photos for files: After the consultation, write down the relative details on the special customer file and make sure all items are complete, and take a photo with customers’ signature then keep in file. So it will be more convenient to compare the treatment effects and to improve and adjust the following treatment program. Select treatment handles: Select the right treatment handle and heads according to the treatment area and skin condition. Clean: Clean the treatment area with the gentle cleansing lotion. Suggest to do a deep cleaning with our HONKON Oasis machines. Clear skin: Before treatment, remove the hair in the treatment area.( it is best to keep 1-3mm of the hair for hair removal treatment ) Start the machine Plug the power cord into a 3-pin earthed socket and insert the key into “Key Switch” and turn it clockwise 90°to start the machine. When use RW-II, please put the handle lower then the machine itself to make a complete water cycle, then touch the treatment head to feel the cooling effect with your hand. Daub gel: Daub about 2mm gel on the treatment area after cleaning, and could add some more according to the skin condition and requirement. Wear protecting goggle and eye-patches Protect the customer’s eyes with protecting goggle while the operator should wear respirator and medical gloves. (Only to wear the protecting goggle and eye-patches when does IPL treatment, no need to wear when only do RF.)

Page 17: HONKON-M40E+ MANUAL M40E+.pdf · reduction and ease anti-aging. . 2.2 Application: 2.3 Contraindications 1. Customers taking drugs which cause photosensitivity, eg: tetracycline;

17

Adjust the parameters Select the treatment function and adjust the parameters according to the treatment area (face, eyes, body) and skin condition (skin type, lesion area, age and level of photo aging) and the expected effect. Do a testing Choose the inner area of forearm with lower energy: press the switch button on the handle or pedal the switch for treating and make twice flash, then put the treatment head on the skin, then Enquire customer’s feeling and observe the responses of the skin at anytime, and then tepidity will be felt normally, if not, increase the energy until got the normal response. Note: The choice causes the skin and the hair has the obvious change, also does not have adverse reactions of energy parameter as the treatment parameter. If the patient has the hemp pain, must confirm whether the head does cling the skin or the decrease energy parameter.

4.3.2 During the operation:

Use the cooling RF separately, take the face as the example, treat the cheek from the gap between nose and lip to the front of ear and down to the chine of jaw; treat the neck from the chine of jaw to the mid-neck; presently, suggest that take the cheek, jaw and neck as the whole unit of treatment, from the top of cheekbone and the socket of eye to the gap between nose and lip, outside to the front ear, down to below cheekbone. Cheekbone and 1/3 front neck and also include the chin. At present, there is no unified operation order, below is reference. (Make: when treat the mandible and neck, operate from the top of the skin to avoid sensitive area, treat it again is also important. The energy of temporal part is down, inside the cheek front area should be avoid operating repeat.) Use the IPL or IPL and RF together, there is no obvious order, but it need a spot size cover a spot size 1/3, and according to the color , thickness and the phenomena of patient skin, adjust the parameters and energy in time, avoid burning and no use in the treatment. Treat many times and repeat (especially the skin lesion area) but the principle is that the skin can not be burnt by accumulating much hot energy. The infrared temperature machine is need during the operation. Enquiry: Ask customer about his feeling (e.g.: how do you feel now? Any pain? Feel hot?) Then adjust the parameters upon his responses. Observation During RF treatment, you will note the face lifting, rube faction and gentle dropsy in the treatment area.

Note: in the RF treatment, the effect will be perfect when the temperature of skin surface (“skin temperature” for short) is 42-48℃. You can test the skin temperature with our infrared temperature testing device which is a selective configuration, and you could adjust the treatment parameter according to it. For example, in the first treatment, the skin temperature corresponding to the treatment parameter is 37℃. To achieve ideal effect, operator could increase the energy and at the same time enquire customer’s feeling (parameter adjustment should based strictly on the feeling of customer), test the temperature once you increase the treatment parameters, until it is up to 42-48℃

Page 18: HONKON-M40E+ MANUAL M40E+.pdf · reduction and ease anti-aging. . 2.2 Application: 2.3 Contraindications 1. Customers taking drugs which cause photosensitivity, eg: tetracycline;

18

4.3.3 After the operation:

1. After the treatment is completed, do a face cleaning for the client; 2. Record all of the parameters and the responds during and after treatment. 3. Wash the treatment handle and cap it. 4. Out of the working condition, turn the key switch counter-clockwise to close condition, turn off the power supply, clean the treatment heads, put the handles on the holders and return the machine to its place. (Keep the key in a proper place or to an appointed person accordingly) 5. Suggest doing a deep cleaning with our HONKON Oasis machines.

4.4 Operating Precautions 1. The environment for operation should avoid disturbing from strong magnetic field and keep away from heat, make sure a fit temperature and ventilation. 2. Before turn on the machine, please make sure the emergency switch and creepage protector under the state of working, check if the machine have been filled with water and make sure the connection between treatment handle and cathode patch and make the plugs right to the socket; 3, when starting the machine, please avoid the LCD touching the rigid thing. 4.During using, keep one of the “Overflow” and “Refill” without cap, otherwise it will not work. 5. Make sure the treatment area to be dry, please do not use the depilatory paste, the depilate frost, and avoid plucking hairs, spreading the wax and the electrolysis; 6. Adjust the parameters from the low to the high and step by step, regulate the energy not more than 2J each time; you should regulate while asking and observing, according to the customer’s feeling; 7. Gel should daub equally on the therapy part of the skin, you can add it at any time, but you can not repeat to use it. 8. Under the e-light treatment, it is necessary to make about 2 shots after regulating parameters and before treating.

4.5 Adverse reflection 4.5.1 RF 1. The rate of the bad reaction is very low, include the process and the uncomfortable after treatment, aroused by the mild and mid-degree erythematic, dropsy, blain, nerve block; 2, Front ear and the below mandible may feel hurt sometime, it can be cured by ibuprofen Some time there is 0.5-1cm scar. These can fade 2-4 weeks later normally.

Page 19: HONKON-M40E+ MANUAL M40E+.pdf · reduction and ease anti-aging. . 2.2 Application: 2.3 Contraindications 1. Customers taking drugs which cause photosensitivity, eg: tetracycline;

19

4.5.2 IPL 1. It has be possibly the slight scalding hot feeling and slight blushing after treatment the former possibly continues for half an hour to 2 hours, the latter possibly continues for 4-12 hours; regarding the pigment venereal diseases it is possibly appear the pigment spot temporary pigment deepening, generally, it will be scab cicatrisation within 3-10 days; 2. The remarkable dropsy, prompted possibly result for using the photosensitivity medicine. It can be deal with available cerebral cortex steroid hormone oral administration and external use processing; 3. The pigment drops or the pigment escapes loses: it is possibly appears in the persons who is black of the skin color or in the sunlight insulation person. It is important to prevent for this kind of patient, the parameter establishment should be enlarged the pulse width suitably, and enlarge pulse gap, decreases the energy density; 4. Temporary inflammation pigmentation will recover treatment after 3-10 days.

4.6 Maintenance after operation and notice: 1,Make the patient to know the nursing method certainly after the operation. The treatment result 30% depends on treats, but the rest depend on the maintenance, so it is very important whether did achieve best treatment result nursing. If can be coordinated the butterfly to supply the water, to make up the oxygen treatment result to be better. we are recommend to use Italy Noble Choice according to the skin condition after operation, after the operation finished uses super fair elite and ORP moisture products; and coordinates to live at home the use; 2, The attention evades the light and guards against exposes to the sun nursing. 3, Avoiding to direct or indirect hot stimulation within 3 days after treatment, don’t use the irritating face milk. 4, Avoiding using kinds chemical or machinery stimulation on the treat spot within one week after treatment. 5, Don’t use the filling material or injects the treatment with 2 weeks after treatment, taboo pungent, irritant, and some foods which including pigment 6, Patient should consult promptly to the operator if you feel not good, prohibits independently decides, processes voluntarily 7, Spa nursing

Note: Treats using this instrument 3-5 times for a treatment course, after waits all symptoms to vanish for 4-7 days only then to carry on the next treatment.

5. Daily Maintenance 1. Place the machine smoothly in a clean room with temperature 10℃-40℃, humidity≤75%, without strong magnetic field etc. 2. Place an air flowing, far away from radiation and other heat source or easy shock position 3. Need to switch on 3-5 times repeatedly to ensure water cycle well after water affusion. 4. Keep the laser nozzle head clean to ensure energy stable. 5. Clean carefully treatment handle and put storage after operation. 6. Await 5 seconds for cutting off power after stop using. 7. Avoid raising a dust, especially carpet floor, and ensure air vent clear when clean treatment room. 8. Replace water as in instruction, and must be the water kind.. 9. Forbid using flammables cleaning for treatment head and treatment area. 10. Don’t wipe external case using cleaning rag with petrol, alcohol or diluents 11. It is forbidden to open the shell of the machine for repair or maintenance. Please contact our after sales service department when the machine broke down. 12. In order to ensure the machine work normally, please contact directly factory or agent, if you buy or change instrument accessory or parts..

Page 20: HONKON-M40E+ MANUAL M40E+.pdf · reduction and ease anti-aging. . 2.2 Application: 2.3 Contraindications 1. Customers taking drugs which cause photosensitivity, eg: tetracycline;

20

ATTETION:

Customer only need to give daily clean to the apparatus, simple trouble shooting can be taken under the direction of our technician. Other maintenance should be handed over the technician appointed by our company

Must cut off power and pull out the power cable before maintenance.

It is dangerous to maintainer and the apparatus along with power in maintenance. a

Fire prevention

Must follow the instruction or precaution and avoiding damage the apparatus or other object.

6. Transport and Storage

1. Make sure to drain all of the water before transportation. 2.The circumstance of transport and storage will be temperature -20℃-55℃, relative humidity ≤75% and a good air quality. 3. Being a precise appliance, avoid vibrancy and impact during moving and transport, and prohibit inversion in case damaging the system and electron parts of the machine. 4. Do not move or transit this machine in the dusty circumstance without dustproof. 5. If this machine will be not in use for a long time, please cut off the power supply and drain completely the water in the tank and remove the handles and cathode patches and store them separately. 6. If once again put into use, please re-install in accordance with the first installation process.

7. Possible Failures and Problem-solving

1. Completely No Power Failure Analysis

(1)Power cord was broken; (2)Emergency switch did not be flipped; (3)Emergency switch or key switch were broken; (4)The fuse was burned; (5)The air switch was cut off.

Solution (1)Change a new power cord; (2)Flip up the emergency switch; (3)Change the a new emergency switch or a new key switch; (4)Change a new fuse; (5)Close the air switch.

Page 21: HONKON-M40E+ MANUAL M40E+.pdf · reduction and ease anti-aging. . 2.2 Application: 2.3 Contraindications 1. Customers taking drugs which cause photosensitivity, eg: tetracycline;

21

2. The treatment handles be heated up.

Failure Analysis (1)No water, or no enough water; (2)Air bubbles in the water after affusion caused a poor water cycling. (3)The water tank in the handle was blocked, hence impeded water cycling and heat can not be

taken away. (4)Too long working times, hence too high the water temperature.

Solution (1)Affused enough water; (2) Check whether the pump works or switch on and off the machines more then three times so as

to eject the air in the water; (3)Clean the cavity of handle; (4)Keep a proper working time. If used frequently, please change water. Note: When changing

water, drain 2/3 of water and leave 1/3, and do not affuse to too cold water in case to damage the xenon lamp and filter tube. 3. No cooling of treatment handle.

Failure Analysis (1)Too long working times, hence too high the water temperature

(2)Cooling system was broken. Solution

(1)Turn off the machine for 30 minutes while change the water. (2)Contact our company for repairing. 4. Get a feeling of electric shock during operation.

Failure Analysis (1)The clime is dry. (2)Treatment head wasn’t contact well with skin. (3)Frequency output line was loosen.

Solution (1)Suggest to wear less chemical fiber garment. (2)Contact the treatment head well with skin. (3)Rehandle the loosening parts.

5. Malfunction of switch button in the handle or of the foot pedal. Failure Analysis

(1)Fail connection of accessories with the machine. (2)Damage of the switch button in the handle or of the foot pedal. (3)Some components of the Power controlling board.

Solution (1)Check and reconnect the sockets according to the precautions of installation.

(2)contact with manufacturer for repairing.

Page 22: HONKON-M40E+ MANUAL M40E+.pdf · reduction and ease anti-aging. . 2.2 Application: 2.3 Contraindications 1. Customers taking drugs which cause photosensitivity, eg: tetracycline;

22

8. Technical parameter

Treatment Handles F+E+,F+E Spot Size 8mm*40mm, 15mm*60mm

Wavelength HR 610-1200nm VL 585-1200nm SR 530-1200nm

Energy Density 20-50J/cm2 Discharge Interval 3-5s 1st –pulse width 0.1-10ms Sub-pulse width 0.1-10ms 1st –pulse interval 0.1-20ms 1st –pulse interval 0.1-20ms Pulse Number 1-10 RF Power 1-50 RF Frequency 750±10% KHz RF Duration 0.1-3.0s adjustable

E-light

Cooling -2℃~0℃ Treatment Handpieces RW-V (Cooling Mono-polar & Bi-polar RF)

RF Power 60W

RW-V Tips Mono-polar:10mm, 25mm, 40mm, Bi-polar: 25mm,35mm

RF Working Mode Continuous Pulse RF Time 0.1-20S RF Delay 0.1-20S

RF

Energy 1-50J

Cathode Pads 188x108mm2(Plastic) ,200x138mm2(Stainless Steel)

Cooling System Air, Water, Semi-conductor Power Supply AC 220V 50Hz/ AC 110V 60Hz Dimension (Before packaged) 600mm*500mm*510mm Dimension (After packaged) 740mm*540mm*610mm Net Weight 30KGS Gross Weight 35KGS