Horiba PG-250Portable Multi-Gas AnalyserThe Horiba PG-250 is a highly reliable and versatile gas analyser for compliance testing of NOx, SO2, CO, CO2, and O2.
MCERTS approved portable stack testingUnlike other portable gas analysers that rely upon electrochemical
sensors, the Horiba PG-250 utilises the same measurement principles as a
permanently installed CEMS. The PG-250 is ideal for CEMS certification
testing or for use when stack gas emissions must be monitored
periodically at one or multiple stacks.
Direct readings on-site The instrument can interface directly with a laptop computer through
an RS-232C interface to record measured values.
Sample conditioning The Horiba features a built-in sample conditioning system with
sample pump,filters, and a thermoelectric cooler. An optional auxiliary
thermoelectric pre-cooler (PS-200) is available to condition samples
which contain excess water vapour
KEYBENEFITS• Compact and lightweight housing
with a built in carrying case so it is
ideal for moving between stacks
• Large, easy to read front panel
displays
• A single portable analyser capable
of measuring five components
The ultimate portable MCERTS analyser that can monitor up to 5 separate components
“ “
NOx SO2 CO CO2 O2
GASESANALYSED
TECHNICALSPECIFICATION
COMPONENTS MEASURED • Nitric Oxide (NOx), Sulphur Dioxide (SO2), Carbon Dioxide
(CO2), Carbon Monoxide (CO), and Oxygen (O2)
MEASUREMENT PRINCIPLE• NOx: Chemiluminescence measurement
• SO2, CO2 & O2: Non-Dispersive Infra Red Absorption
• O2: Galvanic cell measurement
REPEATABILITY, LINEARITY & DRIFT• Repeatability: ±0.5% F.S. (Nox:>100 ppm; CO:>1000 ppm
±1.0 F.S. (other components and ranges)
• Linearity: ±2.0% F.S.
• Drift: ±1.0% F.S. per day
±2.0% F.S. per day for SO2
DIMENSIONS• Dimensions: 260 x 260 x 510 mm
10.2 x 10.2” x 20.0””
• Weight: 17Kg (37.5Lbs)
PERFORMANCE • Response time: 45 seconds (moving average from inlet is 10 sec)
• SO2 Response time: Only SO2: 240 seconds with selective
moving average of 10 sec or 30 sec.
• Display: Measured value (3 or 4 digits) active, range, flow
• Sample Gas Flow Rate: Approximately 0.4 SLPM
• Ambien Humidity: <85% R.H
MEASUREMENT RANGES• NO: 0-25/50/100/250/500/1000/2500ppm
• SO2: 0-200/500/1000/3000 ppm
• CO: 0-200/500/1000/2000/5000ppm
• CO2: 0-5/10/20 vol%
• O2: 0-5/10/25 vol%
POWER • Power Requirements: 100-120V ac; 200-240V ac at 50/60
Hz
• Power Consumption: 250 VA/400 VA
11 Ark Royal Way, Lairdside Technology Park, Tranmere, Wirral, CH41 9HTT: +44(0)151 666 8300 F: +44(0)151 666 8329 E: [email protected] W: www.a1-cbiss.com Buy Online: shop.a1-cbiss.com
Cert No. 996ISO 9001
BUHLERCOOLER
PRODUCT OVERVIEW Accurate measurements of gases require gas samples with stable dew points even under harsh ambient conditions
The Buhler Cooler is available for hire through a1-cbiss, it is
typically used in application with Horiba PG-250. The EGK
models from Buhler provide a compressor-type cooling
system connected to a cooling block. The temperature of
the cooling block is regulated by the Buhler Constant
Regulating System.
The cooling block accommodates either a single stream or a
dual stream heat exchanger hence the cooler may serve two
separate sample gas streams
TECHNICAL SPECIFICATION • Cooling Capacity: 320kj/h
• Ambient Temperature: +5...+50 oC
• Dewpoint: approx. 5 oC
• Power Consumption: 290/260 VA Fuse (external) 10 A
• Alarm output: max. 250V, 2 A, 50 VA plug acc. to DIN
43650
• Protection Class: IP 20