2
Horiba PG-250 Portable Multi-Gas Analyser The Horiba PG-250 is a highly reliable and versatile gas analyser for compliance testing of NOx, SO2, CO, CO2, and O2. MCERTS approved portable stack testing Unlike other portable gas analysers that rely upon electrochemical sensors, the Horiba PG-250 utilises the same measurement principles as a permanently installed CEMS. The PG-250 is ideal for CEMS certification testing or for use when stack gas emissions must be monitored periodically at one or multiple stacks. Direct readings on-site The instrument can interface directly with a laptop computer through an RS-232C interface to record measured values. Sample conditioning The Horiba features a built-in sample conditioning system with sample pump,filters, and a thermoelectric cooler. An optional auxiliary thermoelectric pre-cooler (PS-200) is available to condition samples which contain excess water vapour KEYBENEFITS Compact and lightweight housing with a built in carrying case so it is ideal for moving between stacks Large, easy to read front panel displays A single portable analyser capable of measuring five components The ultimate portable MCERTS analyser that can monitor up to 5 separate components NO x SO 2 CO CO 2 O 2 GASESANALYSED
Transcript
Page 1: Horiba pg250 datasheet

Horiba PG-250Portable Multi-Gas AnalyserThe Horiba PG-250 is a highly reliable and versatile gas analyser for compliance testing of NOx, SO2, CO, CO2, and O2.

MCERTS approved portable stack testingUnlike other portable gas analysers that rely upon electrochemical

sensors, the Horiba PG-250 utilises the same measurement principles as a

permanently installed CEMS. The PG-250 is ideal for CEMS certification

testing or for use when stack gas emissions must be monitored

periodically at one or multiple stacks.

Direct readings on-site The instrument can interface directly with a laptop computer through

an RS-232C interface to record measured values.

Sample conditioning The Horiba features a built-in sample conditioning system with

sample pump,filters, and a thermoelectric cooler. An optional auxiliary

thermoelectric pre-cooler (PS-200) is available to condition samples

which contain excess water vapour

KEYBENEFITS• Compact and lightweight housing

with a built in carrying case so it is

ideal for moving between stacks

• Large, easy to read front panel

displays

• A single portable analyser capable

of measuring five components

The ultimate portable MCERTS analyser that can monitor up to 5 separate components

“ “

NOx SO2 CO CO2 O2

GASESANALYSED

Page 2: Horiba pg250 datasheet

TECHNICALSPECIFICATION

COMPONENTS MEASURED • Nitric Oxide (NOx), Sulphur Dioxide (SO2), Carbon Dioxide

(CO2), Carbon Monoxide (CO), and Oxygen (O2)

MEASUREMENT PRINCIPLE• NOx: Chemiluminescence measurement

• SO2, CO2 & O2: Non-Dispersive Infra Red Absorption

• O2: Galvanic cell measurement

REPEATABILITY, LINEARITY & DRIFT• Repeatability: ±0.5% F.S. (Nox:>100 ppm; CO:>1000 ppm

±1.0 F.S. (other components and ranges)

• Linearity: ±2.0% F.S.

• Drift: ±1.0% F.S. per day

±2.0% F.S. per day for SO2

DIMENSIONS• Dimensions: 260 x 260 x 510 mm

10.2 x 10.2” x 20.0””

• Weight: 17Kg (37.5Lbs)

PERFORMANCE • Response time: 45 seconds (moving average from inlet is 10 sec)

• SO2 Response time: Only SO2: 240 seconds with selective

moving average of 10 sec or 30 sec.

• Display: Measured value (3 or 4 digits) active, range, flow

• Sample Gas Flow Rate: Approximately 0.4 SLPM

• Ambien Humidity: <85% R.H

MEASUREMENT RANGES• NO: 0-25/50/100/250/500/1000/2500ppm

• SO2: 0-200/500/1000/3000 ppm

• CO: 0-200/500/1000/2000/5000ppm

• CO2: 0-5/10/20 vol%

• O2: 0-5/10/25 vol%

POWER • Power Requirements: 100-120V ac; 200-240V ac at 50/60

Hz

• Power Consumption: 250 VA/400 VA

11 Ark Royal Way, Lairdside Technology Park, Tranmere, Wirral, CH41 9HTT: +44(0)151 666 8300 F: +44(0)151 666 8329 E: [email protected] W: www.a1-cbiss.com Buy Online: shop.a1-cbiss.com

Cert No. 996ISO 9001

BUHLERCOOLER

PRODUCT OVERVIEW Accurate measurements of gases require gas samples with stable dew points even under harsh ambient conditions

The Buhler Cooler is available for hire through a1-cbiss, it is

typically used in application with Horiba PG-250. The EGK

models from Buhler provide a compressor-type cooling

system connected to a cooling block. The temperature of

the cooling block is regulated by the Buhler Constant

Regulating System.

The cooling block accommodates either a single stream or a

dual stream heat exchanger hence the cooler may serve two

separate sample gas streams

TECHNICAL SPECIFICATION • Cooling Capacity: 320kj/h

• Ambient Temperature: +5...+50 oC

• Dewpoint: approx. 5 oC

• Power Consumption: 290/260 VA Fuse (external) 10 A

• Alarm output: max. 250V, 2 A, 50 VA plug acc. to DIN

43650

• Protection Class: IP 20


Recommended