Date post: | 16-Jan-2016 |
Category: |
Documents |
Upload: | delilah-mosley |
View: | 218 times |
Download: | 0 times |
Improvements of the A2 beamline for the
linear polarised beam
Patrik OttInstitut für Kernphysik
Johannes-Gutenberg Universität Mainz
28.02.2011
Content• Motivation
– Characteristics of linear polarised photon beam
– Why do we need these two devices
• Beam Stabilisation– Recent Status– Possible automation?
• Pair-Spectrometer
Therefore collimation is a simple way to increase the relative amount of coherent linear polarised photons in the beam
Coherent peak
Ken Livingsten
Characteristics of linear polarised photon
beam
Roman Leukel, Diss. 2001
• Large incoherent background
• Coherent part is strongly collimated
Colli at 7.5 m pol. 70%
Colli at 5.0 m pol. 58%
Colli at 2.5 m pol. 44%
Problems of the collimation
So far we have no possibility to measure the coherent peak without analysing the data
Cause of the small emission angle, a small deviation of the electron beam can cause the complete loss of the linear polarised photons
Position control
A2-Beamline
• TM110 Scanners
• The stabilisation converges best with the correction magnets SFA2WEDL06 and SFA2WEDL08.
Measuring the position in both cavities as a function of the current in both correction magnets
2
1
2221
1211
2
1
I
I
cc
cc
x
x
2
1
1
2221
1211
2
1 9.0x
x
cc
cc
I
I
With these gradients one can calculate the necessary change of current.
The Hysteresis of the correction magnets cause an unpredictable error. Due to this only 90% of the aberration is compensated. After a few iterations the nominal position is reached without oscillation.
Measuring the position in both cavities as a function of the current in both correction magnets
Cavity1
Cavity2
Magnet 1
Magnet 2
Necessity for convergence:
Different slopes for both magnets in at least one cavity
Lock-in amplifier Cavity output
2.45 GHz
MAMI main oscillator
2.45 GHz – 100 kHz
2.45 GHz
∆φ
• Add a reference Signal of 2.45 GHz – 100 kHz to reduce the frequency
• Measure the output signal at a time defined by the MAMI main oscillator
• To adjust the phase ∆φ one needs to steer the beam highly. This produces a large output signal so that the Lock-in amplifier can find a reproducible reference phase.
100 kHz
Wavelength : 12.2 cm
A difference of the electron path through MAMI of 2 cm
reduces the signal output to one half.
Measured horizontal beam position
Position is measured 25 times before switching to the other orientation. This is done to conserve the HF-Relay.
Measured vertical beam position
Position is measured 25 times before switching to the other orientation. This is done to conserve the HF-Relay.
MAMI-Optimisation
After a MAMI-optimisation the beam position often has a large offset, but the operator can bring back the beam with one button.
• After the first 2 days of beam time, the accelerator is in thermal equilibrium and the position is quite stable.
• If an aberration is detected, it is not clear if the reference phase or the beam position is not right.
• Due to this, there is no automation so far. The operator gets an acoustic signal and adjusts the phase again. If an aberration is still detected he can correct the beam.
• During the phase adjustment we need the blank radiator.• After a MAMI-optimisation the operator can bring back
the beam with one button.
Position Control
Pair-Spectrometer
Experimental setup
Measure the tagger channels in coincident with the Pair-Spectrometer to get the coherent Peak.
Hamamatsu APD
Dark current Signal
• No need of HV
• Is not disturbed by magnetic fields
• easy connection with scintillator
First tests
• Does the setup work with the APD?
• Measure the influence of the current and thefoil-thickness
• Can we find the coincident of both scintillators?
• Find the timing between Pair-Spectrometer and Tagger (Moeller-FPGA)
• Provide an excellent grounding of the electronic to suppress noise.
• Find the correct timing in the FPGA
• Open the beam line to build in longer scintillators.
Pair-Spectrometer
To do list: