TO DOWNLOAD A COPY OF THIS POSTER, VISIT WWW.WATERS.COM/POSTERS ©2017 Waters Corporation
Figure 4. Spectral library creation in Empower. Pre-characterized glycan names were input in the Component Tab (left) to create the library. The collected information can be added to the library via the dropdown menu (right).
INTRODUCTION
Glycosylation is one of the most common and complex post-translational modifications on monoclonal antibodies. Due
to its direct correlation to the efficacy of antibody drugs, glycosylation has been recognized as a critical quality attribute
(CQA) of drug product. To ensure the accuracy and consistency of glycosylation profiles, the suitable characterization
and monitoring methods are critical for the product quality throughout the product lifecycle. To this end, methods that
can increase the confidence in CQA assessment are highly desired for product quality control.
As an orthogonal technique, MS detection has been increasingly applied in regulatory environment as an orthogonal
detection technique for its high sensitivity and specificity of measuring product quality attributes. The ACQUITY QDa
mass detector requires minimal knowledge to operate and can be readily adapted to the existing LC-UV or LC/FLR
workflow, making it easy to fit in QC environment.
In this study we present a LC-FLR/MS workflow using MS enabled Spectral Library matching algorithm for improved
productivity and confidence of analysis of glycan profiles.
IMPROVING ROUTINE ANALYSIS OF GLYCAN PROFILES IN A QC ENVIRONMENT USING SPECTRAL LIBRARIES
Brooke Koshel; Ximo Zhang; Robert Birdsall; Joe Fredette; Min Du; Ying Qing Yu Scientific Operations, Waters Corporation, Milford, MA
METHODS
LC conditions
QDa condition
References
1. Exploiting Rapifluor-MS labeling to Monitor Diverse N-Glycan Structures via Fluorescence and
Mass Detection. Waters application note, 720005353EN
2. Rapidly Monitoring Released N-Glycan Profiles During Process Development Using RapiFluor-MS
and the ACQUITY QDa Detector . Waters application note, 720005438EN
RESULTS AND DISCUSSION
CONCLUSION
Spectral Library provides a tool for glycan identification and routine analysis
Added MS detection increases confidence in results compared to optical-based assays
Empower enabled automated data processing and reporting
Readily investigate out of spec results
LC system ACQUITY UPLC H-Class Bio
Detector ACQUITY FLR detector (λexcitation=265 nm, λemission=425 nm)
ACQUITY QDa MS detector (performance model)
LC column ACQUITY Glycan BEH Amide column, 1.7 µm, 130 Å, 2.1 × 150 mm
Column temp. 60 °C
Mobile phase A H2O with 50 mM Ammonium Formate
Mobile phase B ACN
Informatics Empower 3
Mass range 350-1250 m/z
Mode ESI Positive
Capillary voltage 1.5 kV
Cone voltage 10 V
Probe temp 500 °C
Sample rate 2 points/sec
Figure 2, The ACQUITY® QDa Mass Detector. The
QDa detector is a single quadrupole mass detector that can be easily incorporated into an existing LC stack.
Time (min)
Flow (mL/min)
%A %B
Initial 0.400 25.0 75.0
35.00 0.400 46.0 54.0
36.50 0.200 100.0 0.0
39.50 0.200 100.0 0.0
43.10 0.200 25.0 75.0
47.60 0.400 25.0 75.0
55.00 0.400 25.0 75.0
Gradient table
Figure 1, Improved workflow for routine analysis using MS-enabled Spectral Library. With the collected LC-MS data from reference standard, orthogonal information such as name, retention time, and base peak mass, are stored in the Spectral Library and used to match against the samples. Once built, the Spectral Library is universal to all projects in Empower to ensure the consistency of spectral matching. The Spectral Library can be used for routine analysis of glycan profiles to increase confidence in data analysis, provide orthogonal mass spectra confirmation, and support investigations into assay failures.
Figure 7, Spectral Library matching results of the detected five peaks. The results table shows retention time, base peak mass, and matched spectral name of detected peaks. A custom field “Match_Result” was used to compare the glycan identification based on retention time and MS spectra. Results can be exported as report via Empower in an automated fashion.
Workflow for Spectral Library Building and Matching
Orthogonal MS information to support investigations
Start
Figure 5. MS Library Search tab. Parameters can be adjusted for the best matching.
Figure 6. Spectral matching for glycan sample. Using the RFMS glycan standard as reference, Peak 1-5 are the detected glycans from an intact mAb sample using a defined processing method in Empower. Spectral matching was automatically performed with the Empower processing method. The retention time filter can be used to address challenging glycan spectra such as isomers for improved matching ability.
A custom field used as indication of matching results
MS spectra of Glycan isomers
Glycan isomers
Extract MS spectra from auto-integration
Input glycan names in Empower processing method
Retention time filter Helps identify isomers
Spectral Library matching results MS spectra Released glycan from Trastuzumab
Failure of matching due to the additional m/z peak from co-eluted Man5
Open investigation and re-develop method for higher resolution
Figure 8. Investigations supported by orthogonal information. Using released glycan from Trastuzumab as a sample, a 35 min gradient partially resolved the Man5 and A2G1a peak, causing the additional peak (774 m/z) in the MS spectrum of A2G1a, which leads to a mismatch. The orthogonal MS data allows for efficient investigations to identify method failure points and recommend the appropriate course of action.
Collect LC-MS data of reference standard
Matching Results Data acquisition of glycan samples
Enter glycan name and create library Adjust noise threshold
Set library searching parameter
Figure 3. LC/MS data acquisition. The QDa was used as an in-line detector to provide MS information. Retention time and base mass of each peak were obtained via automated integration in Empower. A threshold of relative intensity can be set up to reduce the background noise in Spectral Library.
Report of glycan profiles
XIC
XIC
1) Acquire LC/MS Reference Data
2) Build Spectral Library 3) Run Samples 4) Report