Danielle DEIDIER, M.D. Clinique Saint Vincent, Toulon – France The author of this poster has received travel expense reimbursement from B&L. In vivo evaluation and post-operative outcomes with a micro-incision preloaded IOL system. ASCRS 2010. Purpose. - PowerPoint PPT Presentation
Popular Tags:
10
ASCRS 2010 ASCRS 2010 Danielle DEIDIER, M.D. Clinique Saint Vincent, Toulon – France The author of this poster has received travel expense reimbursement from B&L
Transcript
Page 1: In vivo evaluation and post-operative outcomes with a micro-incision preloaded IOL system

ASCRS 2010ASCRS 2010

Danielle DEIDIER, M.D.Clinique Saint Vincent, Toulon – France

The author of this poster has received travel expense reimbursement from B&L

Page 2: In vivo evaluation and post-operative outcomes with a micro-incision preloaded IOL system

Purpose To evaluate the behaviour of the Micro AY 123 preloaded

IOL system during surgery and to report per-operative and post-operative outcomes in comparison to a previous study with the same lens implanted with a conventional micro-incision injector.

ASCRS 2010ASCRS 2010

Page 3: In vivo evaluation and post-operative outcomes with a micro-incision preloaded IOL system

Methods Single centre, one surgeon Prospective study 32 eyes implanted with Micro AY Eyes examined at 1week, 3 months and 12 months Biaxial phaco: 1.3mm incision, enlarged to 1.7mm for IOL

implantation Endothelial cells count, astigmatism, visual acuity, IOL

centration, IOL surface state, anterior segment aspect, posterior capsule aspect and complications

Performance of the micro-incision 123 Premium peloaded system

ASCRS 2010ASCRS 2010

Page 4: In vivo evaluation and post-operative outcomes with a micro-incision preloaded IOL system

Results

0,16

0,720,76

0,37

0,980,93

Preop (n=32) 3 month (n=32) 12 month (n=29)

UCVABCVA

ASCRS 2010ASCRS 2010

Micro AY (PhysIOL)

25% hydrophilic acrylic, 6.15mm optic, 10.75mm overall, aspheric, UV and blue blocker, 360° square edge

Page 5: In vivo evaluation and post-operative outcomes with a micro-incision preloaded IOL system

Results

Preop(n = 32)

3 month(n=32)

12 month(n=29)

Astigmatism 0.66 0.59 0.68

Preop(n = 32)

1 week(n=32)

12 month(n=27)

Endothelial cells count (cell/mm2) 2284 2155 2146

3 month(n=32)

12 month(n=29)

IOL centration Centered Centered

ASCRS 2010ASCRS 2010

Page 6: In vivo evaluation and post-operative outcomes with a micro-incision preloaded IOL system

Results1 year

Micro AY(n = 29)

2 yearsMicroSlim*

(n=52)

Nd:Yag 0 0

Posterior capsule 100% clear90% clear

6% light fibrosis4% fibrosis

* MicroSlim has the same design as Micro AY except a spherical optic and a UV blocker only

ASCRS 2010ASCRS 2010

Page 7: In vivo evaluation and post-operative outcomes with a micro-incision preloaded IOL system

Preloaded system

ASCRS 2010ASCRS 2010

Page 8: In vivo evaluation and post-operative outcomes with a micro-incision preloaded IOL system

Preloaded system

ASCRS 2010ASCRS 2010

The lens is packed in a micro-incision The lens is packed in a micro-incision cartridge. The injector is clipped to the cartridge. The injector is clipped to the cartridge and removed from the cartridge and removed from the container. After having placed container. After having placed viscoelastic behind the lens, the plunger viscoelastic behind the lens, the plunger is advanced slowly. The pressure is is advanced slowly. The pressure is slightly higher in the final stage, when slightly higher in the final stage, when the orange silicone cushion stretches in the orange silicone cushion stretches in the thinnest part of the injection tip. The the thinnest part of the injection tip. The lens unfolds gently and is perfectly intact lens unfolds gently and is perfectly intact after injection.after injection.

Page 9: In vivo evaluation and post-operative outcomes with a micro-incision preloaded IOL system

Experience with Micro AY 123 preloaded system

10 lenses implanted Same incision size and implantation technique as

Micro AY with MicroSet injector (1.7mm) At 3 months:

20/20 BCVAGood lens centrationGood capsular transparency

ASCRS 2010ASCRS 2010

Page 10: In vivo evaluation and post-operative outcomes with a micro-incision preloaded IOL system

ConclusionMicro-incision lenses

Excellent visual outcomesLow level of induced astigmatism Low endothelial cells lossGood stability over timeGood maintenance of capsular bag transparency over

two yearsNo IOL-related complications

Micro-incision preloaded systemA promising alternative for a safer and more

reproducible injection

ASCRS 2010ASCRS 2010