+ All Categories
Home > Documents > infant Radiant waRMeRs Model MatRixdigitronicsmedical.com/wp-content/uploads/2017/07/... ·  ·...

infant Radiant waRMeRs Model MatRixdigitronicsmedical.com/wp-content/uploads/2017/07/... ·  ·...

Date post: 21-Mar-2018
Category:
Upload: dodieu
View: 214 times
Download: 1 times
Share this document with a friend
2
MECHANICAL SPECIFICATIONS SC Wheels: 4 with break Height (cm) 192 192 190 Width (cm) 70 70 71,5 Depth (cm) 91 91 94 Warmer weight with accessories (kg) 60 60 100 SC BED NB bed height (since floor) 100 100 100 Bed weight capacity 10 10 10 Width (cm) 32 42 61 Depth (cm) 61 70 72 Electric height adjustment Electric height adjustment SC – Foldable acrylic bassinet walls R1 – Removable crib R2 – Foldable acrylic bassinet walls Heater rotation (180º) R1–Removablecribwithoptionalre- versePhototherapyModelMedphoto6 VOLTAGE REQUIREMENTS Voltage automatic selection 100 - 140V~ 205 - 245V~ Current 5,0A 3,0A Network frequency 50//60Hz Power consumption 1400 VA Heating element Quartz-coated electric resistance Power: 580 Watts Lamp Halogen lamp with dichroic reflector – 12V/20W Temperature control (Automatic mode) From 25ºC – 38°C, with 0.1°C increments Power heating control (manual mode) 0 – 100%, with 10% increments Temperature Indication From 20ºC – 45°C, with ± 0,1°C accuracy Maximum admissible error ± 0,2°C ± 0,3°C ISO 13458 European Authorized Representative Center Obelis S.A. www.obelis.net 0 4 7 0 INFANT RADIANT WARMERS MODEL MATRIX NEONATAL RODMIDIA The information contained herein is subject to alterations without previous notice. Further informations: www.olidef.com.br +55 (16) 3919.9350 [email protected] – Ribeirão Preto/SP – Brazil
Transcript

Mechanical specifications R¹ R² scWheels: 4 with break Height (cm) 192 192 190

Width (cm) 70 70 71,5

Depth (cm) 91 91 94

Warmer weight with accessories (kg) 60 60 100

R¹ R² scBED NB bed height (since floor) 100 100 100

Bed weight capacity 10 10 10

Width (cm) 32 42 61

Depth (cm) 61 70 72

Electric height adjustment Electric height adjustment SC – Foldable acrylic bassinet walls

R1 – Removable crib

R2 – Foldable acrylic bassinet walls

Heater rotation (180º)

R1 – Removable crib with optional re-verse Phototherapy Model Medphoto 6

Voltage RequiReMentsVoltage automatic selection

100 -140V~

205 -245V~

Current 5,0A 3,0A

Network frequency

50//60Hz

Power consumption

1400 VA

Heating element

Quartz-coated electric

resistancePower: 580

Watts

LampHalogen lamp with dichroic

reflector – 12V/20W

Temperature control (Automatic mode)

From 25ºC – 38°C, with

0.1°C increments

Power heating control (manual mode)

0 – 100%, with 10%

increments

Temperature Indication

From 20ºC – 45°C, with ±

0,1°C accuracy

Maximum admissible error

± 0,2°C ± 0,3°C

iso13458

european authorizedRepresentative center

obelis s.a.www.obelis.net

0470

infant Radiant waRMeRs Model MatRix

NEONATAL

RODM

IDIA

The

info

rmat

ion

cont

aine

d he

rein

is s

ubje

ct to

alte

ratio

ns w

ithou

t pre

viou

s no

tice.

Further informations: www.olidef.com.br – +55 (16) 3919.9350 – [email protected] – Ribeirão Preto/SP – Brazil

technical especificationsControl unit MicroprocessadoTemperature adjustment range 25,0ºC - 38,0ºCSkin temperature measurement 20,0ºC - 45,0ºCHeat power (10% increments) 0 - 100%Display resolution 0,1ºCThermostatic precision 0,2ºCBrazilian ANVISA registration number 10227180023Standards NBR IEC 60.601-1

NBR IEC 60.601-1-2NBR IEC 60.601-2-21

Voltage 110/220V automático

geneRal chaRacteRistic R¹ R² scRadiant warmer system Radiant warmer systemHeater rotation (180º): access for X-rays devices

contRol systeMs R¹ R² scSafety electronic systemManual heater modeSkin mode (servo–controlled)

contRol Module R¹ R² scAlphanumeric LCD Skin temperature probeDigital power indicatorApgar timerRechargeable internal batteryAlarme silence buttonKeyboard blockSimultaneous temperature display¹Standby indicator LED

alaRMsPower failureManual mode warning each 10 min.

High skin temperature + 1ºC

High temperature safety 15 min.¹

Low skin temperature - 1ºC

Heating failureSensor failureSystem failureHeater out of positionSensor disconnected from the patient skinSensor disconnectAlarm silence 10 min.

¹ Manual mode: power adjust proportional heater indicator and Apgar timer Automatic mode: temperature adjust, skin temperature, proportional heater indication/Apgar timer

¹Available jars capacity: 2 L; 2,5 L; 3 L; 5 L – ²With flowmeter, suction unit with vacuum gauge and humidifier – ³ With flowmeter, suction unit and humidifier

BaBy´s coMpaRtMent R¹ R² scRemovable crib Foldable acrylic bassinet wallsX-rays transparent bedX-ray cassete trayMattress (impermeable/non-allergic/non-toxic)

acessóRios R¹ R² scVentilator tube support Temperature sensor coverTransport handlesCabinetWeighing scale (external display)Instrument trayIntermediate tray (45 x 60 cm)Autoclavable mattress cover Oxygen hood (three sizes)Cylindres/cylinders holderPhotoherapy model Maxiphoto halogenReverse phototherapy model Medphoto 6Collection jar with lid¹Large drawer (up to 2 units)Drawer set (up to 2 units)X-rays transparent bedBumpersManual resuscitatorRuler for installation of accessoriesTriple ruler AR²Triple ruler O2²Triple ruler O2³Régua tripla mista – Ar – O2 – Vacuômetro Continous bed tilting systemElectrical height adjustment Hydraulic bed tilting systemIV poleHigh adjustable IV poleMonitor trayPower outlets (4 or 8)

¹ in the maximum power

R1: with removable acrylic crib R2: with flat bed and foldable acrylic walls Standard Optional accessory N/A – Not available

linha neonatal

Newborns prematures babies have greater difficulties in maintaining their body temperature than full-term babies and adults because of their large body surface in relations to their volume. The use of the thermal therapy has been demonstrated as being very useful to increase the survival rate in low-weight babies. Several devices have been developed to actively warm and maintain appropriate temperature in neonates and infants, such as radiant warmers and hot air incubators. The neonatal infant radiant warmers models Matrix R e SC, are recommended for delivery rooms, nurseries and intensive care units and provide a warm and stable microenvironment for small babies, who need additional care in this critical period in the beginning of their lives. These are high-technology equipment whose resources may include control systems of temperature, along with factors such as quality, effectiveness and user – friendly operation, all specially designed to fully meet the user´s expectations, therefore providing greater comfort for the small patients.

Infant radiant warmers model Matrix

R2 - with flat bed and foldable acrylic walls

R1 - with removable acrylic crib

SC

The

phot

o is

onl

y ilu

stra

tive.


Recommended