+ All Categories
Home > Documents > Infection Control in Dentistry -Smile4Dr-1

Infection Control in Dentistry -Smile4Dr-1

Date post: 06-Jul-2018
Category:
Upload: savyasachin
View: 218 times
Download: 0 times
Share this document with a friend

of 29

Transcript
  • 8/17/2019 Infection Control in Dentistry -Smile4Dr-1

    1/29

    http://localhost/var/www/apps/conversion/tmp/scratch_5/dentalbooks-drbassam.blogspot.com

  • 8/17/2019 Infection Control in Dentistry -Smile4Dr-1

    2/29

     3-dentalbooks-drbassam.blogspot.com 

    Advicesheet- dentalbooks-drbassam.blogspot.com

    Infection control in dentistry A12 

    dentalbooks-drbassam.blogspot.com

    contents page Introduction 5 

    Routine procedures 5 

    Patient perception 5 

     Acceptance of patients 5 

    Confidentiality 6 

    The infected dental health care worker 6 

    Infection control in dentistry 7 

    Infection control policy 7 

    Training in infection control 7 Surgery design 7 

    Decontamination area 8 

    Choice of equipment 8 

    Single use (disposable) items 9 

    Decontamination of instruments and equipment 9 

    New reusable instruments 10 

    The decontamination process 10 

    Pre-sterilisation cleaning 10 

    Washer-disinfectors 11 

    Ultrasonic cleaners 12 

    Manual cleaning 12 Cleaning dental handpieces 13 

    Inspection and function testing 13 

    Sterilisation 13 

    Dental handpieces 14 

    Checks, tests and record keeping 14 

    Instrument packaging and storage 15 

    Decontamination of impressions, prostheses and appliances 15 

    Surface decontamination 16 

    Cleaning protocols 16 

    Water supplies 16 

    Dental unit water lines 17 

    Disposal of waste 17 Blood spillages 18 

  • 8/17/2019 Infection Control in Dentistry -Smile4Dr-1

    3/29

     4-dentalbooks-drbassam.blogspot.com 

    contents page 

    Personal protection 18 

    Immunisation 18 

    Hand protection 19 

    Hand hygiene 19 

    Gloves 20 

    Latex allergy 20 

    Eye protection and face masks 20 

    Surgery clothing 21 

    Removing PPE 21 

     Aerosol and saliva/blood splatter 21 

    Inoculation injuries 21 

    Emerging infections 22 

    Transmissible Spongiform Encephalopathies 22 

    Meticillin-resistant Staphylococcus aureus (MRSA) 23 

    Tuberculosis 23 

    Herpes simplex 23 

    Pandemic flu 23 

     Appendices  

     Appendix 1 

    HTM 01-05: Decontamination in primary dental care practices 24 

    Essential quality requirements 24 

    Essential infection control policies 25 

    Best practice 25 

    Example layouts for decontamination areas 26 

     Appendix 2 

    Decontamination in Scotland 28 

  • 8/17/2019 Infection Control in Dentistry -Smile4Dr-1

    4/29

     5-dentalbooks-drbassam.blogspot.com 

    In April 2009, the Department of Health (England) published HTM 01-05:

    Decontamination in primary care dental practice which describes essential and

    best practice requirements. A summary of these requirements is given in

     Appendix 1 and the full version is available on the Department’s website

    (www.dh.gov.uk). This advice sheet incorporates the requirements of the

    Department’s guidance. Models of the various policies and protocols that

    practices are required to have in place are available on the BDA’s website at

    www.bda.org/infectioncontrol. 

    In Scotland, a number of organisations produce guidance on decontamination

    in dental practice, some of which is still being developed. A summary of the

    advice produced for practices in Scotland is given in Appendix 2. 

    Introduction 

    Routine procedures 

    Patient perception 

    Acceptance of patients 

    Dental practices have a responsibility to adopt safe systems of working with

    respect to cross-infection control and decontamination. Those working in

    primary dental care must ensure quality decontamination processes are in

    place. These can be provided by using modern local decontamination

    equipment and quality facilities. External decontamination facilities forreprocessing can be used where they meet the needs of the practice.  

    Implementing safe and realistic infection control procedures requires the fullcompliance of the whole dental team. Every practice must have acomprehensive written infection control policy which identifies the infection

    control procedures to be followed. These procedures should be regularlymonitored during clinical sessions and routinely audited. All members of the

    dental team must understand and practise these procedures; regular discussionat practice meetings is recommended. 

     A thorough medical history should be obtained for all patients at the first visitand updated regularly. Medical history questionnaires alongside direct

    questioning and discussion between the dentist and the patient arerecommended. Discussions should be conducted in an environment that

    permits the disclosure of sensitive and confidential personal information.Medical history information should be retained as part of the patient's dentalrecords. 

    The medical history and examination may not identify asymptomatic carriers of

    infectious disease and standard precautions must be adopted. This means thatthe same infection control procedures must be used for all patients.  

     All dental professionals have a duty of care to their patients to ensure adequate

    infection control procedures are followed. Failure to employ adequate methodsof cross-infection control may call into question a practitioner’s fitness to

    practice. 

     As a result of frequent media coverage, the public is now far more aware of the

    need for dentists to practise good infection control. Displaying an infection

    control statement may be appropriate in your practice to help allay patient

    anxiety and gain their confidence. It may encourage them to ask questions, so

    never be too busy to give an answer. Ensure all the members of your practice

    team are confident and competent to answer patients' queries or know who to

    refer to when necessary. 

    Whilst a health professional has the right to accept or refuse to treat a patient, it

    is important that the dental profession accepts the responsibility of providing

    dental treatment to all members of the community. Dental clinicians have ageneral obligation to provide care to those in need and this should extend to

    infected patients who should be offered the same high standard of care 

    available to any other patient. 

    http://www.bda.org/infectioncontrolhttp://www.bda.org/infectioncontrolhttp://www.bda.org/infectioncontrolhttp://www.bda.org/infectioncontrol

  • 8/17/2019 Infection Control in Dentistry -Smile4Dr-1

    5/29

     6-dentalbooks-drbassam.blogspot.com 

    Those with human immunodeficiency viruses (HIV), who are otherwise well,

    and carriers of the hepatitis viruses may be treated routinely in primary dental

    care settings. In the absence of an inoculation injury (which includes blood-

    contaminated splashes to the eyes or mouth), the evidence indicates that the

    risk of infection to a dental health care worker during the dental treatment of

    HIV-infected individuals is negligible. HIV-infected individuals need a high

    standard of dental care when they are asymptomatic to minimise dental 

    problems. If they subsequently develop Acquired Immune Deficiency Syndrome

    (AIDS), the associated medical problems may make it appropriate for them to

    be referred for specialist dental advice and care. 

    It is unethical to refuse dental care to those patients with a potentially infectious

    disease on the grounds that it could expose the dental clinician to personal risk.It is also illogical as many undiagnosed carriers of infectious diseases passundetected through practices and clinics every day. If patients are refused

    treatment because they are known carriers of an infectious disease, they maynot report their conditions honestly or abandon seeking treatment; both results

    are unacceptable. Those who reveal that they are infected are providingprivileged information. 

     All information disclosed by a patient in the course of medical history taking,

    consultation and treatment is confidential. No part of the information obtainedshould ever be disclosed to any third party, including relatives, without the

    patient's permission. Dentists are responsible for the security of informationgiven by patients, whether it is written on record cards or held on computer. All

    members of the dental team should be aware of the duty of strict confidentialityand seek to ensure it at all times. Practices should have a confidentiality policyin place and contracts of employment for dental staff should include a 

    statement on the need to maintain confidentiality. 

     All health care workers have an overriding ethical and legal duty to protect thehealth and safety of their patients and those who carry out exposure-prone

    procedures should be immune to or non-infectious for hepatitis B. A dentalclinician who believes he or she may be infected with a blood borne virus, TB

    or other infection has an ethical responsibility to obtain medical advice,including any necessary testing. If a clinician is infected, further medical adviceand counselling must be sought. Changes to clinical practice may be required

    and may include ceasing or restricting practice, the exclusion of exposure-prone procedures or other modifications. An infected clinician must not rely on

    his/her own assessment of the possible risks to their patients. Failure to obtainappropriate advice or act upon the advice given would almost certainly lead tothe practitioner’s fitness to practice being questioned. 

    Exposure-prone procedures are those invasive procedures where there is a risk

    that injury to the worker may result in exposure of the patient's open tissues to

    the blood of the worker. These include procedures where the worker's glovedhands may be in contact with sharp instruments, needle tips and sharp tissues(spicules of bone or teeth) inside a patient's open body cavity, wound or  confined anatomical space where the hands or fingertips may not be completely

    visible at all times. 

     A dental professional who employs a dental nurse who is subsequently found tobe infected with a blood borne virus should seek advice from local occupational

    health services. A risk assessment must be carried out to consider the risk topatients and the need for the nurse to be redeployed within the practice. Therisk assessment must take into account the duties performed by the dental 

    nurse and the likelihood that the infection could be transmitted to a patient oranother member of staff. An infected dental nurse must not undertake exposure

    prone procedures in order to remove, as far as is possible, the risk of transmittinginfection. There may be employment issues that need to be considered andadvice should be sought from the employment advisers at the BDA. 

    Confidentiality 

    The infected dental

    health care worker 

  • 8/17/2019 Infection Control in Dentistry -Smile4Dr-1

    6/29

     7-dentalbooks-drbassam.blogspot.com 

    Infection

    control in

    dentistry 

    Infection control policy 

    Training in infection

    control 

    Surgery design 

    Members of the dental team have a duty to ensure that infection control

    procedures are followed routinely. The mouth carries a large number of

    potentially infective microorganisms; saliva and blood are known vectors of

    infection. Most carriers of latent infection are unaware of their condition and it is

    important, therefore, that the same infection control routine is adopted for all

    patients. 

    The following recommendations for infection control procedures in routine

    primary dental practice are made in light of current knowledge and may be

    subject to revision, as further information becomes available. 

    Every practice must have a comprehensive written infection control policy, which

    is tailored to the routines of the individual practice and regularly updated. It

    should demonstrate that the practice is working to current recommendations for

    all aspects of infection control including personal protection, instrument

    decontamination and equipment maintenance. The policy should be kept readily

    available so that staff can refer to it when necessary.  

    Where a practice does not have a washer-disinfector installed and/or does not

    have separate decontamination facilities, a written assessment of theimprovements needed to incorporate these together with an implementation

    plan (subject to local constraints) should be available. 

     A model infection control policy can be downloaded from the BDA website at

    www.bda.org/infectioncontrol. 

     All dental staff must be aware of the procedures required to prevent the

    transmission of infection and should understand why these are necessary.

    Regular monitoring of the procedures is essential and the infection control

    policy for the practice should be reviewed regularly and updated when

    necessary and at least annually. 

     All new staff must be appropriately trained in infection control procedures prior

    to working in the practice. Training should equip staff to understand: 

    • how infections are transmitted 

    • the practice policy on decontamination and infection control 

    • what personal protection is required and when to use it  

    • what to do in the event of accidents or personal injury.  

    With regard to decontamination procedures, training records should show that

    staff  

    • have been appropriately trained 

    • are competent to decontaminate the reusable dental instruments currently inuse 

    • training is updated for any new instruments introduced into the dental

    practice. 

    Individual records of the training received should be maintained for all staff.

    The layout of the surgery, which should be simple and uncluttered, is an 

    important aspect of infection control. Ideally there should be distinct areas for  

    the operator and the dental nurse, each with a washbasin, which should have

    sensor controlled or elbow/foot-operated mixer taps and dispensers for

    antimicrobial hand wash solutions, liquid soap and alcohol hand rub /gel. The

    operator's area should have access to the turbines, three-in-one syringe, slowhandpiece, bracket table and operating light. The dental nurse's area should

    contain the suction lines, perhaps the three-in-one syringe, curing light, and the

    cabinetry containing dental materials. 

    http://www.bda.org/infectioncontrolhttp://www.bda.org/infectioncontrolhttp://www.bda.org/infectioncontrolhttp://www.bda.org/infectioncontrol

  • 8/17/2019 Infection Control in Dentistry -Smile4Dr-1

    7/29

     8-dentalbooks-drbassam.blogspot.com 

    The surgery should be ‘zoned’ to identify those areas that are likely to be

    contaminated during treatment sessions from those that are unlikely to be

    contaminated. Zoning can help to make the decontamination process more

    efficient; only the contaminated areas need to be cleaned between patients.

     At the end of the clinical session, all surfaces should be cleaned. 

    Work surfaces should be impervious and easy to clean and joins sealed to

    prevent the accumulation of contaminated matter and aid cleaning. Coving

    between the work surface and the wall will aid cleaning. Seek advice from the

    manufacturer on decontamination products compatible with the worksurface. 

    The floor covering should be impervious, smooth and easy to clean; seams

    should be sealed. Coving between the floor and the wall will aid cleaning. 

    There is a clear need to maximise the separation of decontamination activities

    from clinical work and wherever possible, decontamination should take place in

    a room (or rooms) away from the clinical area. Where space and room

    availability allow, dentists should plan for this as a matter of priority. The

    example layouts given in HTM 01-05 Decontamination in primary care dental

    practices (Department of Health, 2009) are reproduced in Appendix 1. 

    Where instruments are reprocessed in the surgery, the reprocessing area

    should be as far from the dental chair as possible. To reduce the risk of

    exposure to aerosol, manual washing, using ultrasonic cleaners without a lid

    and opening decontamination equipment should not take place when the

    patient is in the surgery. 

    The decontamination area should, preferably, comprise a single run of sealed,

    easily cleaned worktops and include: 

    • a separate hand washing sink 

    • a setting down area for dirty instruments 

    • washing and rinsing sinks (or separate bowls within a single sink) adjacent

    to the receiving area 

    • utrasonic cleaner (where used) 

    • a washer-disinfector (where available) 

    • an area with task lighting for instrument inspection and function testing. 

    Where a type B (vacuum) autoclave is used, this area can also be used for

    wrapping instruments prior to sterilisation 

    • autoclave(s) 

    • an area for setting down sterilised instruments where they can be placed

    onto trays for same day use or wrapped for storage (where a type N (non-

    vacuum) or type S autoclave is used) 

    • where possible, air movement should be from clean to dirty areas.  

     A dirty to clean workflow should be maintained throughout the decontamination

    process to minimise the possibility of used instruments coming into contact with

    sterilised instruments. 

     A practice protocol for selecting new equipment will help to ensure that the

    purchase is necessary (and that other devices already present in the practice

    are not suitable), the equipment will achieve what is necessary and, where

    required, can be processed. The protocol will help the practice avoid the

    purchasing items which later prove problematic – for example, where the

    manufacturer recommends decontamination processes that are not available in

    the practice. Some of the aspects to consider are given below. 

    • What you want the equipment to do – will the equipment selected be fit for

    this purpose? Is it compatible with other equipment in the surgery? 

    • How easy will it be to use and maintain? 

    Decontamination area 

    Choice of equipment 

  • 8/17/2019 Infection Control in Dentistry -Smile4Dr-1

    8/29

     9-dentalbooks-drbassam.blogspot.com 

    • Is it CE marked (a mandatory requirement to demonstrate compliance with 

    Medical Devices Regulations)? 

    • Does the instrument need dismantling before cleaning? Are there

    instructions from the manufacturer describing how this can be done? 

    • Does the instrument have a limited life-cycle specified by the manufacturer? 

    • What are the manufacturer’s recommendations for cleaning and will they be

    achievable in practice? Will the instrument withstand automated washer-

    disinfector processes? 

    • When selecting new hand instruments, avoid difficult to clean serrated

    handles and check that hinges are easy to clean. 

    • What cleaning agents are recommended – do they comply with COSHH and

    health and safety requirements? Are these cleaning agents compatible with

    the washer-disinfector, ultrasonic cleaner and instruments already in use in

    the practice? 

    • Check with the manufacturer which cleaning agents are recommended for

    the dental chair covering and work surfaces to ensure that they can be

    regularly decontaminated without deterioration. 

    • Is steam sterilisation (134 – 137OC for three minutes) appropriate for the

    instruments? If another time-temperature range is recommended, can this

    be undertaken? • Select foot controlled equipment whenever possible. 

    • Is training required? Will the manufacturer provide it? 

    • What are the commissioning and validation requirements of the equipment? 

    What are the ongoing costs? 

    • Service response – what is the response time in the event of a breakdown? 

    Single-use (disposable) items 

    Whenever feasible, single-use items should be considered as an alternative to

    processing reusable items. Where instruments and equipment can be

    processed for re-use, manufacturers must provide information on effective

    decontamination procedures. Where an instrument cannot be safely

    decontaminated for re-use, it is described as ‘single-use’ by the manufacturer

    and the packaging will bear the international symbol: 

    Single-use means that a device can be used on a single patient during one

    treatment session and then discarded. It is not intended to be reprocessed and

    used again – even on the same patient at a later session. Anyone who

    decontaminates and reuses a single-use item bears full responsibility for its

    safety and effectiveness. 

    Where instruments are difficult to clean, single-use alternatives (if available)

    should be considered. In dentistry, this includes, but is not limited to, matrix

    bands, saliva ejectors, aspirator tips and three-in-one tips. Endodontic reamersand files must be treated as single-use (regardless of the manufacturer’s

    recommendation) to reduce the risk of prion transmission in dentistry.  

    Decontamination

    of instruments

    and equipment 

     All instruments contaminated with oral and other body fluids must be thoroughly

    cleaned and sterilised after use. The decontamination process (also known as

    reprocessing) includes pre-sterilisation cleaning, disinfection, inspection,

    sterilisation and storage. Manufacturers are required to provide instructions for

    the decontamination of their equipment – these instructions should be followed.

    It is worth checking with the manufacturer prior to purchase that the equipment

    can be used for the purpose intended and decontaminated by the methods used in the practice. 

  • 8/17/2019 Infection Control in Dentistry -Smile4Dr-1

    9/29

    10 - dentalbooks-drbassam.blogspot.com 

    New dental instruments should be fully decontaminated before use. Identify

    instruments that can withstand automated cleaning processes (washer-

    disinfectors and ultrasonic cleaners) and those which require manual cleaning.

    Some instruments may require dismantling before cleaning and sterilising. It is

    important to follow the manufacturer’s instructions, especially if the new

    equipment is unfamiliar to those responsible for its reprocessing. 

     A systematic approach to the decontamination of instruments after use can helpto ensure that dirty instruments are segregated from clean. The diagram

    summarises how the individual stages ideally link together to complete thedecontamination process. 

    New reusable instruments 

    The decontamination

    process 

    With a washer- disinfector (WD) 

    With or without an ultrasonic cleaner

    manually 

    Source: HTM 01-05 Decontamination in primary care dental practices (Department of Health, 2009)  

     After sterilisation, instruments for immediate use (ie on the same day) can be

    put onto individual covered trays. At the end of each patient treatment, all

    instruments on the tray (used and unused) must be regarded as contaminated

    and reprocessed. At the end of the day, unused trays of instruments should be

    reprocessed before use. Keeping to a minimum the instruments put onto trays

    at the start of the day will reduce the decontamination workload.  

     A practice protocol should describe the safe procedures for transferring

    contaminated instruments to the decontamination area and for transferring

    sterilised instruments to the treatment or storage area. 

    Instruments for use at a later date should be wrapped to prevent

    recontamination during storage: 

    • non-vacuum autoclave – instruments should be dried after sterilisation,

    wrapped/ packaged and used within 21 days 

    • vacuum autoclaves – pre-wrapped/packaged sterilised instruments should

    be used within 30 days 

    Effective cleaning of instruments before sterilisation will reduce the risk of

    transmission of infectious agents. Wherever possible, cleaning should be

    undertaken using an automated and validated washer-disinfector in preference

    to manual cleaning; a washer-disinfector includes a disinfection stage thatrenders instruments safe for handling and inspection. Manual cleaning should

    be considered where the manufacturer’s instructions specify the device is not

    compatible with automated processes. 

    Pre-sterilisation cleaning 

  • 8/17/2019 Infection Control in Dentistry -Smile4Dr-1

    10/29

    11 - dentalbooks-drbassam.blogspot.com 

    Validation means that a process has been verified, tested and documented

    and is consistently reproducible. A summary of the validation, testing and

    maintenance requirements is available on the BDA website at

    www.bda.org/infectioncontrol. A validated washer-disinfector (and, if possible,

    ultrasonic cleaner) demonstrates that instruments and equipment are reliably

    and consistently cleaned. 

    Instruments cleaned as soon as possible after use may be more easily cleaned

    than those left for a number of hours before reprocessing. Blood, saline and

    iodine are corrosive to stainless steel instruments and will cause pitting and 

    then rusting if remaining on instruments for any length of time. Where a delay is

    anticipated, instruments should be kept moist by immersion in water or an

    enzymatic cleaner (following the manufacturer’s recommendations for use) or

    the use of a foam spray intended to maintain a moist or humid environment.

    Long periods of wet storage should be avoided, however. 

    Dental materials (especially cements) can harden on instruments so should be

    removed from instruments as soon as possible after use to allow effective

    cleaning. 

    Where recommended by the manufacturer, instruments and equipment that

    consist of more than one component should be dismantled to allow each part to

    be adequately cleaned. Members of the dental team should be trained to  

    ensure competence in dismantling, cleaning, sterilising and reassembling

    instruments and equipment. 

    Washer-disinfectors 

    Washer-disinfectors offer the best option for the control and reproducibility of

    cleaning with a process that can be validated. Dentists should plan, where

    possible, to install a validated washer-disinfector to remove the need for manual

    cleaning. There are a number of different models that meet current

    requirements. The size, model and type chosen should be considered against

    the workload and throughput requirements, together with the availability of

    space. 

     A typical washer-disinfector cycle includes five stages: 

    1. Flush - removes gross contamination using a water temperature of less that  

    45OC 

    2. Wash – removes remaining soil using detergents specified by the

    manufacturer  

    3. Rinse(s) – removes detergents 

    4. Thermal disinfection – temperature raised for required time: 80OC for 10

    minutes or 90OC for 1 minute, for example 

    5. Drying – heated air removes residual moisture. 

    The manufacturer’s instructions for use should be followed, including

    recommendations for water quality/type, detergents and/or disinfectants and

    instrument loading. Staff must be trained how to use it and how to perform daily

    tests. Records of training must be maintained. 

    Washer-disinfectors must be loaded correctly to ensure effective cleaning. This

    involves: 

    • not overloading instrument carriers or overlapping instruments 

    • opening instrument hinges and joints fully 

    attaching instruments requiring irrigation to the irrigation system correctly,ensuring filters are in place if required (eg for handpieces). 

    http://www.bda.org/infectioncontrolhttp://www.bda.org/infectioncontrolhttp://www.bda.org/infectioncontrol

  • 8/17/2019 Infection Control in Dentistry -Smile4Dr-1

    11/29

    12 - dentalbooks-drbassam.blogspot.com 

    Washer-disinfector logbooks and records should include cycle parameters and

    details of routine testing and maintenance. Automated data-loggers or

    interfaced small computer-based recording systems can be used, provided the

    records are kept securely and replicated (to guard against fading). Records

    should be kept for at least two years. 

    Ultrasonic cleaners 

    Evidence supports the use of ultrasonic cleaners as an effective means of

    cleaning dental instruments and reduces contact with contaminated

    instruments. The cleaner must be maintained according to manufacturer’s

    recommendations with quarterly testing to ensure that it is fully functional. The

    results of all tests should be recorded. 

     After use, instruments should be immersed briefly in cold water (with detergent)

    to remove visible soiling, taking care to avoid inoculation injuries. A container

    with a sealing lid is recommended. 

    The manufacturer’s instructions for operating the ultrasonic cleaner should be

    followed. 

    Place instruments in a suspended basket and not on the floor of the cleaner

    (avoid overloading and overlapping) and fully immerse in the cleaning solution.

    Joints and hinges should be fully opened and instruments disassembled where

    appropriate before immersion. Set the timer, close the lid and do not open until

    the cycle is complete. Drain the basket of instruments and rinse using clean

    fresh reverse osmosis (RO) or distilled water to remove residual soil and

    detergent. Instruments to be wrapped and sterilised in a vacuum autoclave

    must be dried first using a disposable non-linting cloth. 

    The water/fluid must be changed at the end of the clinical session and more

    frequently if it becomes heavily contaminated. At the end of each day, the

    ultrasonic cleaner must be emptied, cleaned and left dry. 

    Manual cleaning 

    Compared with other cleaning methods, manual cleaning carries a greater risk

    of inoculation injury. It is however, important for practices to have the facilities,

    documented procedures and trained staff to carry out manual cleaning when

    other methods are not appropriate or available. 

    Manual cleaning, although simple to set up, is difficult to validate as it is not

    possible to ensure that it is carried out effectively each time. Where manual

    cleaning is necessary, the parameters should be controlled as much as

    possible to reduce variability in cleaning. A written procedure should be

    available and followed routinely. A model protocol can be downloaded from the

    BDA website at www.bda.org/infectioncontrol. 

     A dirty-to-clean workflow should be maintained throughout. Two sinks are

    needed - one for cleaning and one for rinsing with separate areas for setting

    down dirty and clean instruments. If there is only sufficient space for one setting

    down area, the surface should be cleaned with a water-detergent solution

    between stages. 

     Always use detergents specifically made for the manual cleaning of instruments

    and mix with water to the correct concentration and temperature (as specified 

    by the manufacturer). The temperature should not exceed 45OC. Fully submerge

    the items to be cleaned (unless manufacturer recommends otherwise) and

    scrub using long-handled brushes. Drain the cleaning water and rinse itemsusing RO or freshly distilled water. Instruments to be wrapped and sterilised in a

    vacuum autoclave must be dried first using a disposable non- linting cloth.  

    http://www.bda.org/infectioncontrolhttp://www.bda.org/infectioncontrolhttp://www.bda.org/infectioncontrolhttp://www.bda.org/infectioncontrolhttp://www.bda.org/infectioncontrol

  • 8/17/2019 Infection Control in Dentistry -Smile4Dr-1

    12/29

    13 - dentalbooks-drbassam.blogspot.com 

    Cleaning dental handpieces 

    Dental handpieces must be decontaminated after use. 

    • Where the manufacturer confirms that a handpiece can withstand cleaning

    in a washer-disinfector and the washer-disinfector can be adapted to clean

    handpieces, this method is preferred. 

    • Dedicated handpiece-cleaners can be considered where a washer-

    disinfector is not recommended. 

    • Commercial products for decontaminating handpieces can be used where

    the product can be shown to reduce the risk of infection transmission or the

    process can be validated. 

    • The manufacturer’s recommendations for lubrication should be followed. 

    • Separate canisters of lubricant should be used for unclean and cleaned

    handpieces. 

    Inspection and function

    testing 

    Sterilisation 

     After cleaning, instruments should be inspected for cleanliness and checked for

    wear or damage before sterilisation. A magnifying glass with task lighting is

    recommended. 

    •If there is residual contamination, the instrument should be rejected and re-cleaned. 

    • Working parts should move freely and joints should not stick. The occasional

    use of a non-oil-based lubricant may be necessary where hinges are stiff. 

    • The edges of clamping instruments should meet with no overlap or rough

    edges. 

    • The edges of scissors should meet to the tip and move freely across each

    other with no overlap or rough edges. 

    •  All screws on jointed instruments should be tight. 

    Instruments found to be faulty or damaged should be taken out of use. If they

    are to be sent for repair, they should be decontaminated fully (cleaned and

    sterilised) and labelled ‘decontaminated’ before dispatch. Equipment that

    cannot be sterilised must be thoroughly cleaned and disinfected in accordance

    with the manufacturer’s instructions. 

    Saturated steam under pressure delivered at the highest temperature

    compatible with the product is the preferred method for the sterilisation of most

    instruments and devices used in the clinical setting. In dentistry, this is usually a

    temperature of 134-137OC with a holding time of 3 - 3.5 minutes.  

    Three types of autoclaves are suitable for use in dentistry:  

    Type N: passive displacement of air with steam (non-vacuum). Designed for

    unwrapped, non-hollow and non-air retentive instruments 

    Type B (vacuum): designed for hollow, air retentive and packaged loads Type S: designed to reprocess specific loads (determined by the manufacturer). 

    Effective sterilisation requires steam to contact all surfaces of the instrument.

    Instruments must, therefore, be loaded into the chamber to allow free

    circulation of steam. This is particularly important when air removal is passive;

    air remaining in the chamber will impair or prevent the sterilisation process.

     Avoid overloading the autoclave chamber. 

    Water reservoirs should be filled daily using fresh distilled or reverse osmosis

    (RO) water. Autoclave water should be discharged after each cycle but where

    this is not possible, the reservoir must be drained at the end of each working

    session to reduce the likelihood of a build up of toxins in the water supply. Afterthe final use of the day, the chamber should be drained, cleaned and dried and

    left with the door open. 

  • 8/17/2019 Infection Control in Dentistry -Smile4Dr-1

    13/29

    14 - dentalbooks-drbassam.blogspot.com 

    Dental handpieces 

    The internal lumen of dental handpieces makes them difficult to clean although a

    compatible validated washer-disinfector may produce successful cleaning

    results. Furthermore, the presence of lubricant in the lumen means that,

    whichever autoclave is used, handpiece sterility is unlikely. Good cleaning and

    steam sterilisation will, however, result in a beneficial reduction in contamination

    levels and bioburden. 

    Checks and tests and record keeping 

    Before use each day: 

    • Clean the rubber door seal with a clean, damp non-linting cloth 

    • Check the chamber and shelves for cleanliness and debris 

    • Fill the reservoir with freshly distilled or RO water  

    • Turn on the power source. 

    Daily tests and housekeeping tasks should then be carried out and the results

    recorded in the logbook: 

    •Steam penetration test (vacuum autoclaves only) 

    •  Automatic control test (all autoclaves) to demonstrate that the autoclave is

    actually working 

    • Where required, a warm-up cycle before instruments can be processed. 

    Records of regular checks must be maintained to demonstrate compliance. An

    autoclave that fails to meet any of the test requirements should be withdrawn

    from service and advice sought from the manufacturer and/or maintenance

    contract. 

     Autoclaves should be commissioned when first purchased to ensure that they

    are appropriately calibrated and functioning correctly. Validation before use by a

    Competent Person (Decontamination) or service engineer is needed to

    demonstrate that the right conditions for sterilisation are achieved. The

    equipment must be properly maintained according to the manufacturer's

    instructions and periodically examined by a competent person. A summary of  

    the validation, testing and maintenance requirements is available on the BDA 

    website at www.bda.org/infectioncontrol. 

    The parameters should be monitored for each cycle. Printouts and automated

    data loggers or interfaced computer-based recording systems are acceptable

    provided the records are kept securely and replicated. Printouts fade within a

    short time, so require photocopying. Manual records (where no automatic data

    production is available) are acceptable and should document the

    temperature/pressure achieved or an absence of failure. Records should be

    maintained for at least two years. 

    The readings should be compared with the recommended values – if any

    reading is outside its specified limits, the sterilisation cycle must be regarded as

    unsatisfactory, irrespective of the results obtained from chemical indicators and

    the autoclave cycle checked again. If the second cycle is unsatisfactory, the

    autoclave should not be used until the problem has been rectified by an

    appropriately trained engineer. Chemical indicators (TST strips, for example)

    demonstrate only that instruments have been through a sterilisation cycle, not

    that they have been sterilised 

    http://www.bda.org/infectioncontrolhttp://www.bda.org/infectioncontrolhttp://www.bda.org/infectioncontrolhttp://www.bda.org/infectioncontrolhttp://www.bda.org/infectioncontrol

  • 8/17/2019 Infection Control in Dentistry -Smile4Dr-1

    14/29

    15 - dentalbooks-drbassam.blogspot.com 

    Instrument packaging and

    storage 

    Decontamination of

    impressions, prostheses

    and appliances 

    Sterilised instruments must be protected against the possibility of

    recontamination by wrapping or storing in a covered container. The autoclave

    used affects the wrapping and storing options. 

    Type B autoclave (vacuum): dried instruments can be pre-wrapped. Once

    sterilised, the instruments may be stored for up to 30 days.  

    Type N autoclave (displacement): dried instruments can only be wrapped

    after sterilisation using sealed view packs. If trays of instruments are to be

    stored, the entire tray should be placed in a sealed pack. Instruments can be

    stored for up to 21 days. Alternatively, instruments can be covered and used

    within the current session. 

    Type S autoclaves : the manufacturer’s guidance for pre-wrapping should be

    followed although post-sterilisation packing remains an option. 

    Disposable non-linting cloths should be used to dry instruments and disposed

    of after each sterilisation load. 

    The area where sterilised instruments are packaged for storage should be freeof clutter and wiped clean with detergent and alcohol wipes at the start of each

    session. 

    Instruments should be stored in an area dedicated for the purpose and away

    from direct sunlight and water in a secure, dry and cool environment. Where

    this is in the surgery, the storage area should be as far from the dental chair as

    reasonably practicable; a purpose designed storage cabinet that can be easily

    cleaned will be useful. Ideally, air flow should be from clean to dirty areas.

    Where possible, practices should plan to store instruments in a separate

    environment, away from the surgery. 

    Storage systems must ensure easy identification of instruments and monitoring

    of storage times to ensure recommended intervals are not exceeded. The

    packaging should therefore display the use-by date and a system of first-in,  

    first-out introduced. Simple record keeping for infrequently used instruments will

    help to avoid excessive periods of storage resulting in pathogen recolonisation.

    The record should show the date of decontamination and an expiry date.  

    Before use, check that the packaging is intact or the instruments have remained

    covered, the sterilisation indicator confirms that the pack has been sterilised (if  

    a type B autoclave has been used) and visible contamination is absent.  

    The responsibility for ensuring these devices have been cleaned and disinfected

    prior to dispatch to the laboratory lies solely with the dentist. It is good practice

    to agree the cleaning and disinfection process with the laboratory and label thedevice to indicate disinfected status. This removes uncertainty and, for

    impressions, also removes the possibility of repeated disinfection, 

    which may detract from quality. 

    Immediately on removal from the mouth, the device should be rinsed under

    running water to remove saliva, blood and debris. Continue the process until it

    is visibly clean. If an appliance or prosthesis is grossly contaminated, it should

    be cleaned in an ultrasonic bath containing detergent and then rinsed. 

    The device should then be disinfected following the manufacturer's

    recommendations. Products that are suitable for the disinfection of impressions,

    prostheses or appliances are CE marked to demonstrate conformity toEuropean Directives. 

  • 8/17/2019 Infection Control in Dentistry -Smile4Dr-1

    15/29

    16 - dentalbooks-drbassam.blogspot.com 

    There are two methods of disinfection: immersion and dipping. 

    Immersion in disinfectant (following the manufacturer’s recommendations for

    dilution and duration) can be effective but may be compromised by the limited

    working life of the disinfectant, which is affected by the frequency of use and

    the presence of biological debris. 

    Dipping avoids the prolonged immersion that can distort hydrocolloid and

    polyether impression materials. The recommended contact time is still

    necessary, during which the impression must not be allowed to dry out. 

    Following disinfection, the device must be thoroughly rinsed in water before

    packaging to send to the laboratory with a confirmation that it has been

    disinfected. Items received from a laboratory should also be disinfected. 

    Surfaces should be impervious and easily cleanable. Work surfaces and floor

    coverings should be continuous, non-slip and, where possible, without joints. If

    present, joints should be sealed. Coving between the floor and wall will help

    prevent accumulation of dust and dirt. The manufacturer’s advice should be

    sought on the compatibility of detergents and disinfectants with the surface orequipment. 

    The practice should have a written protocol outlining surface- and room-

    cleaning schedules and maintain simple records. Cleaning staff should be

    briefed on cleaning patient care areas and decontamination rooms. 

    Surfaces can be effectively cleaned using commercial bactericidal cleaning

    agents and wipes. Alcohol, although effective against viruses, binds to blood

    protein and stainless steel; it should therefore be avoided. Water with suitable

    detergents is satisfactory, provided the surface is dried after cleaning. Following

    initial deep cleaning of a surface, subsequent use of a wet or dry microfibre

    cloth can achieve satisfactory removal of infectious agents. The mircrofibre can

    then be reprocessed as laundry at the end of each session or when obviously

    contaminated. 

     A strict system of zoning aids and simplifies the cleaning process. In practice,

    this means defining the areas, which will become heavily contaminated during

    operative procedures – worksurfaces, dental chair, curing lamp, inspection light,

    hand controls, spittoons, and aspirator, for example. Light and chair hand

    controls can be protected with disposable impervious coverings and changed

    between patients. If these are not used, the controls must be cleaned effectively

    between patients. 

     At the end of clinical sessions, all work surfaces, including those apparently

    uncontaminated, should be thoroughly cleaned using disposable cloths ormicrofibre materials and should include the taps, drainage points, splashbacks,

    cupboard doors and sinks. Aspirators, drains and spittoons should be cleaned

    at the end of a session according to manufacturers’ instructions.  

    Computer keyboards should be either washable or provided with covers that

    can be easily decontaminated at frequent intervals. 

    Practices should have a written scheme (a course of action) for controlling

    Legionella bacteria in water systems and preventing or controlling the risk. A

    risk assessment will identify potential problems in the system (for example,

    excess storage capacity, temperature distribution problems, low water usage,

    inappropriate materials etc). Further information is available on the BDA’swebsite (www.bda.org). 

    Surface decontamination 

    Cleaning protocols 

    Water supplies 

  • 8/17/2019 Infection Control in Dentistry -Smile4Dr-1

    16/29

    17 - dentalbooks-drbassam.blogspot.com 

    Dental unit water lines (DUWLs) 

    The majority of dental units will harbour biofilm, a source of microbial

    contamination for the water produced by the unit, so the water will not be

    potable (ie of drinking water quality). Contaminated water is a potential hazard

    to both patients and surgery staff and may harbour potentially pathogenic

    organisms such as Legionella spp and Pseudomonas aeruginosa. 

     All water lines should be fitted with anti-retraction valves to help prevent

    contamination of the lines but these valves cannot be relied upon to prevent

    infected material being aspirated back into the system. DUWLs should be

    flushed for at least two minutes at the beginning of the day and for at least  

    20-30 seconds between patients to reduce the microbiological counts in the

    water delivery tube. 

    No currently available single method or device will completely eliminate

    biocontamination of DUWLs or exclude the risk of cross-contamination. The

    manufacturer’s instructions should be followed for the periodic disinfection of

    water lines. Introducing chemical treatments into the dental unit is best achieved

    via a water reservoir (bottled water system), which can be fitted retrospectively, 

    if not fitted at the time of purchase. The water bottles should be removed,flushed with distilled or RO water, left to dry overnight and stored inverted. 

     An effective treatment regime includes an initial purge to remove longstanding

    biofilm followed by a daily maintenance regime to prevent the reformation of

    fresh biofilm. Waterline biofilm reforms rapidly if the unit remains untreated, so

    less frequent intermittent treatments may fail. For surgical procedures, an

    independent system with a sterile irrigant should be used. 

    Manufacturers of dental units should supply guidance for the maintenance of  

    the unit and treatment regimes to ensure water quality. A simple dipslide culture

    test allows practices to assess the effectiveness of the treatment regime. This

    test involves a plastic slide coated with an agar culture medium dipped into a

    sample of water from the dental unit and incubated in a sealed container at

    room temperature for seven days, at which stage bacterial colonies will be

    visible to the naked eye. A guide on counting the bacterial colonies should be

    supplied with the dipslide test. Providers of biofilm treatment systems will often

    offer this service as part of their package. 

    The design of some dental equipment requiring a mains water supply means

    that it is possible for contaminated water to be drawn back through the

    waterlines to the mains water supply (backflow/backsiphonage). This can affect

    the dental unit, wet-line suction pumps, automatic radiographic processors and

    washer-disinfectors. Interrupting the water supply to the surgery by a physical

    break (air gap) will remove the possibility of backflow. Some equipment

    requiring a water supply is now manufactured to incorporate an air gap - checkthis with the manufacturer. 

    Disposal of waste   All waste in the practice must be: 

    • correctly segregated 

    • stored safely and securely on the premises 

    • packaged appropriately for transport 

    • described accurately and fully on the accompanying documentation when

    removed 

    • transferred to an Authorised Person for transport to an authorised waste site  

    • appropriately registered, with necessary records and returns at the practice. 

    Dental practices produce a range of hazardous and non-hazardous waste.

    Further information can be found in the BDA Advice Note Management of

    healthcare waste available at www.bda.org/infectioncontrol. 

    http://www.bda.org/infectioncontrolhttp://www.bda.org/infectioncontrolhttp://www.bda.org/infectioncontrolhttp://www.bda.org/infectioncontrolhttp://www.bda.org/infectioncontrol

  • 8/17/2019 Infection Control in Dentistry -Smile4Dr-1

    17/29

    18 - dentalbooks-drbassam.blogspot.com 

    If blood is spilled, the spillage should be dealt with as soon as possible following

    a protocol that protects against infection. The spilled blood should be completely

    covered either by disposable towels, which are then treated with sodium

    hypochlorite solution or sodium dichloroisocyanurate granules, both

    producing10,000 ppm chlorine. Good ventilation is essential. Allow at least 5

    minutes to elapse before clearing and disposing of towels as clinical waste. The

    dental health care worker dealing with the spillage must wear appropriate

    protective clothing: household gloves, protective eyewear, a disposable apron

    and, in the case of an extensive floor spillage, protective footwear. 

    The employer has a duty of care towards employees to provide a safe place of

    work. It is not sufficient simply to provide personal protective equipment such 

    as gloves and eye protection; the employer must ensure that it is being used in

    the correct manner. It is important that all staff understand the principles of

    personal protection and that compliance is part of their contracts of

    employment. 

     Advice should be sought from the local occupational health department on theappropriate vaccination requirements of all clinical staff. All those involved in

    clinical procedures must be vaccinated against hepatitis B. If an inoculation

    injury is sustained before completion of the course, follow up action, including

    boosters and tests for hepatitis B markers, is essential. The hepatitis B vaccine

    is effective in preventing infection in individuals who produce specific antibodies

    to the hepatitis B surface antigen (anti-HBs). Antibody responses to the 

    hepatitis B vaccine vary widely between individuals. It is preferable to achieve

    anti-HBs levels of above 100mIU/ml, although levels of 10mIU/ml or more are

    generally accepted as enough to protect against infection. Protection against

    infection is maintained even if antibody concentrations at the time of exposure

    have declined. Antibody titres should be checked one to four months after

    completion of a primary course of the vaccine.  

    Responders with anti-HBs levels ≥100mIU/ml do not require any further primary

    doses; once a response has been established further assessment of antibody

    levels is not indicated. A single booster dose at around five years after primary

    vaccination is recommended for all health care workers who have contact with

    blood, blood stained fluids and patients' tissues. Pre- and post-testing at the

    time of this booster is not required if the individual responded to the primary

    course of vaccine. 

    Responders with anti-HBs levels of 10 to 100mIU/ml should receive one

    additional dose of vaccine at the time; further assessment of antibody levels is

    not indicated. They should also receive the booster at five years.  

     An antibody level below 10mIU/ml is classified as a non-response to the

    vaccine and testing for markers of current or past infection is required (and to

    exclude the possibility of being a carrier of infection). Those identified as non-

    responders should undergo a repeat course of vaccine, followed by retesting

    one to four months after the second course. Those who still have anti-HBs

    levels below 10mIU/ml and who have no markers of current or past infection,

    will require hepatitis B immunoglobulin for protection if exposed to the virus.  

    Employers must hold documentary evidence to demonstrate that all relevant

    members of the dental team have been immunised and their responses to the

    vaccine checked; post vaccination blood test results will show whether an

    adequate level of immunity has been achieved. The consent of the employeemust be obtained before the occupational health department or the GMP is

    approached. Any information provided is confidential and should be stored

    appropriately. 

    Blood spillages 

    Personal

    protection 

    Immunisation 

  • 8/17/2019 Infection Control in Dentistry -Smile4Dr-1

    18/29

    19 - dentalbooks-drbassam.blogspot.com 

    New staff who are not immunised should undergo a course of vaccination as

    soon as possible. Chairside assisting can begin after the first vaccination as

    long as a risk assessment of their duties has been carried out, and the

    appropriate controls identified have been put in place. 

    Further information is available in BDA Advice Note Immunisation against

    hepatitis B available at www.bda.org./infectioncontrol. 

    Hand protection  Hand care is vital to infection control; lacerated, abraded and cracked skin can

    offer a portal of entry for microorganisms. Clean hands complement the use of

    gloves; neither is a substitute for the other. Training in hand hygiene should be

    included in a staff induction programme and regular update training provided to

    all staff. 

    Hand hygiene 

    There are different levels of hand hygiene depending on the potential for

    contamination of the hands and the process to be undertaken. 

    Social (10-15 seconds) will remove transient microorganisms using plain or

    antimicrobial liquid soap. When: general non-clinical activities, including decontamination. 

    Hygienic (15-30 seconds) will destroy microorganisms and provide a residual

    effect using an antiseptic cleanser or antimicrobial soap from a dispenser.

    When: before wearing gloves to carry out clinical procedures and after contact

    with blood and other body fluids. 

    Surgical scrub (2-3 minutes, ensuring all areas of the hands and forearms

    are covered) will substantially reduce the numbers of resident microorganismsusing an antiseptic hand cleaner (chlorhexidine gluconate 4%, povidone iodine 7.5%). 

    When: oral, periodontal and implant surgery 

     A poster depicting the relevant method(s) should be displayed above everywash-hand basin in the practice. A poster is included in the Department of

    Health’s guidance (HTM 01-05) and is also available on the BDA website atwww.bda.org/infectioncontrol. 

    To reduce the risk of irritation, mild liquid soap should be applied to wet handsand hands washed under running water. Hands should be washed: 

    • Before and after each treatment session 

    • Before and after the removal of PPE 

    • Following the washing of dental instruments 

    • Before contact with sterilised instruments (wrapped and unwrapped) 

    •  After cleaning or maintaining decontamination devices used on dental

    instruments 

    •  At the completion of decontamination work. 

     After washing, hands should be dried thoroughly, using disposable towels, toprevent transfer of microorganisms and prevent skin damage. Hand cream

    (preferably water-based) will help to avoid chapped or cracking skin. A wall-mounted dispenser with disposable cartridges should be used. 

    Fingernails should be kept clean, short and smooth. False nails and nail polishshould not be used. Rings, bracelets and wrist watches should not be worn

    during clinical procedures. If a wedding ring is worn, the skin beneath it shouldbe washed and dried thoroughly. 

    Further information is available in the BDA’s model policy on hand hygiene,

    available on the BDA website at www.bda.org/infectioncontrol. 

    http://www.bda.org./infectioncontrolhttp://www.bda.org./infectioncontrolhttp://www.bda.org./infectioncontrolhttp://www.bda.org./infectioncontrolhttp://www.bda.org/infectioncontrolhttp://www.bda.org/infectioncontrolhttp://www.bda.org/infectioncontrolhttp://www.bda.org/infectioncontrolhttp://www.bda.org/infectioncontrolhttp://www.bda.org/infectioncontrolhttp://www.bda.org/infectioncontrolhttp://www.bda.org/infectioncontrolhttp://www.bda.org/infectioncontrolhttp://www.bda.org./infectioncontrol

  • 8/17/2019 Infection Control in Dentistry -Smile4Dr-1

    19/29

    20 -dentalbooks-drbassam.blogspot.com

    Gloves 

    Gloves must be worn for all clinical procedures and treated as single use items,

    so a new pair of gloves must be used for each patient. It is important that 

    gloves fit properly. Gloves should be put on immediately before contact with the

    patient and removed as soon as clinical treatment is complete. Used gloves

    must be disposed of as clinical waste. 

    There is a variety of gloves available for clinical procedures. Those selected

    should be –  

    • good quality non-sterile medical gloves (to European standard BSEN 455,

    parts 1 and 2, medical gloves for single use), worn for all clinical procedures

    and changed after every patient 

    • well fitting and non-powdered. The powder from gloves can contaminate

    veneers and radiographs, disperse allergenic proteins into the surgery

    atmosphere and interfere with wound healing 

    • low in extractable proteins (

  • 8/17/2019 Infection Control in Dentistry -Smile4Dr-1

    20/29

    21 -dentalbooks-drbassam.blogspot.com

    Surgery clothing 

    Removing PPE 

    Aerosol and saliva/blood

    splatter 

    Inoculation injuries 

     A wide variety of clothing is worn in dental surgeries and in many practices is

    used to reinforce the corporate image. Surgery clothing should not be worn

    outside the practice; adequate changing and storage facilities must be

    provided. Short sleeves allow the forearms to be washed as part of the

    handwashing routine. Long sleeves are more likely to become contaminated

    during clinical sessions and could cause a breach in infection control. 

    Surgery clothing can become contaminated with microorganisms during

    procedures, so freshly laundered uniforms should be worn every day. Machine-

    washing with a suitable detergent at a minimum temperature of 60OC will 

    reduce any potential microbial contamination. 

    Disposable plastic aprons should be worn during decontamination processes.

    Depending on the type of PPE worn, it should be removed in the following  

    order: 

    1. Gloves – ensuring that the gloves end up inside out and that the hands do

    not become contaminated. If contaminated, wash hands thoroughly before

    removing other PPE. 2. Plastic disposable apron – by breaking the neck straps and gathering the

    apron together touching the inside only. 

    3. Face mask – by breaking the straps of lifting over the ears, avoiding

    touching the outer surface of the mask. Never allow mask to hang around

    neck. 

    4. Face and eye protection, taking care not to touch outer surfaces.  

    5. Wash hands thoroughly. 

    Good surgery ventilation and efficient high-volume aspirators, which exhaust

    externally from the premises, will reduce the risk of infection by dispersing and

    eliminating aerosols. High-volume aspirators turned on prior to the handpiece

    will reduce risk from aerosols. External vents should discharge without risk to

    the public or re-circulation into any building. Aspirators and tubing should be

    cleaned and disinfected regularly in accordance with the manufacturer's

    instructions and the system should be flushed through at the end of each

    session with their recommended surfactant/detergent and/or non-foaming

    disinfecting agent. 

    Rubber dam isolation of teeth also offers substantial advantages and should be

    used whenever practicable. It enhances the quality of the operative 

    environment and virtually abolishes saliva/blood splatter. When working without

    rubber dam, the use of high-volume aspiration is essential. 

    Inoculation injuries are the most likely route for transmission of blood borne

    viral infections in dentistry. The definition of an inoculation injury includes allincidents where a contaminated object or substance breaches the integrity of

    the skin or mucous membranes or comes into contact with the eyes. The

    following are typical examples: 

    • sticking or stabbing with a used needle or other instrument 

    • splashes with a contaminated substance to the eye or other open lesion 

    • cuts with contaminated equipment 

    • bites or scratches inflicted by patients. 

    Inoculation injuries must be dealt with promptly and correctly.  

     Allow the wound to bleed and then wash thoroughly with running water. 

  • 8/17/2019 Infection Control in Dentistry -Smile4Dr-1

    21/29

    22 -dentalbooks-drbassam.blogspot.com

    •  Assess the risks associated with the patient and the injury. Where there is

    reason to be concerned about the possible transmission of infection, the

    injured person should seek urgent advice according to the local

    arrangements in place on what follow up action, including serological

    surveillance, is necessary. All practices should have formal links with their

    local occupational health service, so that management of sharps injuries is

    undertaken promptly and according to accepted national protocols. 

    • Contact the occupational health service of the primary care organisation for

    advice on post-exposure prophylaxis. Practices without an NHS contract

    may have to arrange this privately. Every practice should have details of the

    local contact displayed prominently. 

    • When local advice is not available, advice should be obtained from the

    following sources: 

    England: Health Protection Agency Centre for Infections 61 Colindale Avenue, London NW9 5EQ 

    Tel: 020 8200 4400, Email:  [email protected] 

    Scotland: Health Protection Scotland 

    Clifton House, Clifton Place, Glasgow G3 7LN Tel: 0141 300 1100, Email: [email protected] 

    Wales: PHL Cardiff  

    University Hospital of Wales, Heath Park, Cardiff CF14 4XWTel: 02920 742718 

    N Ireland: Director of Public Health at the local Health and Social 

    Services Board 

    • Make a full record of the incident in the accident book, including details of

    who was injured, how the incident occurred, what action was taken, which

    dentists were informed and when and, if known, the name of the patient

    being treated. Both the injured person and the dentist in charge shouldcountersign the record. 

    In dentistry, the risk of acquiring HIV infection following an inoculation injury is

    very low. If, however, the injury is risk-assessed as significant for transmission

    of HIV and the source patient is HIV infected, post exposure prophylaxis (PEP)

    should be commenced as soon as possible after the incident and ideally within

    the hour. PEP involves the use of a short course (four weeks) of treatment with

    anti-retroviral drugs in an attempt to reduce even further the risk of infection  

    with HIV following exposure. Dentists should clarify with their local occupational

    health service the local arrangements for urgent access to PEP before any

    incident occurs. 

    CJD and related conditions raise new infection control questions: 'prions', the

    infectious agents that cause them, are much more difficult to destroy than

    conventional micro-organisms, so optimal decontamination standards need to

    be observed. All instruments must be thoroughly cleaned before autoclaving, in

    order to remove as much matter as possible. Patients with vCJD or CJD, or

    identified as ‘at-risk’ of vCJD for public health purposes, (or their relatives)

    should not be refused routine dental treatment. 

    Guidance on the prevention of transmission is available in Transmissible

    Spongiform Encephalopathy Agents: safe working and the prevention of

    infection produced by the Advisory Committee on Dangerous Pathogens(December 2003) and supplemented by a letter from the Chief Dental Officer

    (February 2005). Both are available on the Department of Health and BDA

    websites. 

    Emerging

    infections 

    Transmissible Spongiform 

    Encephalopathies 

    mailto:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]

  • 8/17/2019 Infection Control in Dentistry -Smile4Dr-1

    22/29

    23 -dentalbooks-drbassam.blogspot.com

    Meticillin-resistant

    Staphylococcus aureus

    (MRSA) 

    Tuberculosis 

    Herpes simplex 

    Pandemic flu 

    No additional infection control precautions are necessary for the dental

    treatment of patients colonised with MRSA. However, members of the dental

    team known to be colonised with MRSA should not undertake or assist with

    invasive procedures. A clinical microbiologist or communicable disease

    physician will be able to provide treatment to eradicate the MRSA colonisation.  

    The incidence of all forms of tuberculosis (TB) is rising and now approximately

    one third of the world's population is infected. The disease is spread by droplets

    or by direct contact and has been transmitted by dental procedures. Although

    Mycobacterium tuberculosis is the usual cause of TB, other species of

    mycobacterium can also cause the disease. The infection control procedures

    described in this document should be adequate protection against transmission

    of TB. Staff infected with TB should seek guidance from their local occupational

    health services. 

    Herpes simplex virus type 1 (HSV-1) is usually associated with infections of the

    lips, mouth and face. It is the most common virus and is usually associated with

    childhood. HSV-1 often causes lesions inside the mouth such as cold sores and

    is transmitted by contact with infected saliva. By adulthood, up to 90% of

    individuals will have antibodies to HSV-1. The herpes virus can reside in thebody for years, appearing only as a cold sore when something provokes it, for

    example, illness, stress, hormonal changes and sun exposure. Individuals

    usually experience a tenderness, tingling or burning before the actual sore

    appears, initially as a blister which subsequently crusts over. 

    During the prodromal stage and active stage with open lesions, the individual is

    shedding large amounts of the virus. Ideally, dental treatment should not be

    undertaken but the decision lies with the individual clinician bearing in mind that  

    • the herpes simplex virus is highly infectious and easily transmitted 

    • manipulation of the facial and oral tissues can exacerbate the condition and

    cause breakdown of the lesion and bleeding 

    • spread of the virus to other areas of the skin can cause significant problems

    (new primary lesions, for example); infection of the eyes is a rare but

    significantly serious complication. 

     A patient requiring urgent dental care should not be denied it but until the

    herpetic lesions are healed, the dental team should take care to prevent the

    spread of the virus. 

    The Department of Health has issued specific guidance for dental practices on

    what to do in the event of pandemic flu, which is available at

    www.dh.gov.uk/en/Publichealth/Flu and summarised on the BDA’s website at

    www.bda.org/infectioncontrol. 

    Influenza is a respiratory illness characterised by rapid onset of a wide range of

    symptoms including fever, cough, headache, sore throat and aching muscles

    and joints. It has an average incubation time of two to three days and people

    are most infectious soon after they develop symptoms. 

    Transmission is through close contact with an infected coughing or sneezing

    person. Hand washing (with soap and water or alcohol handrub) and

    environmental cleaning will deactivate the virus and help control spread through

    contact. 

    The main measures for containing the infection include 

    • standard infection control measures and droplet precautions 

    • a ‘stay at home’ approach for anyone with flu-like symptoms 

    http://www.dh.gov.uk/en/Publichealth/Fluhttp://www.bda.org/infectioncontrolhttp://www.bda.org/infectioncontrolhttp://www.bda.org/infectioncontrolhttp://www.bda.org/infectioncontrolhttp://www.dh.gov.uk/en/Publichealth/Flu

  • 8/17/2019 Infection Control in Dentistry -Smile4Dr-1

    23/29

    24 -dentalbooks-drbassam.blogspot.com

    • separating flu-infected patients from well patients when dental care is

    needed 

    • preventing symptomatic visitors (accompanying well patients, for example) 

    from attending the practice. 

    HTM 01-05: Decontamination in primary dental care practices 

    Published by the Department of Health in 2009, the guidance HTM 01-05 is

    based on a principle of continuous improvement and introduces benchmarks to

    achieve compliance with ‘essential quality requirements’ and ‘best practice’.

    Essential requirements should be in place within 12 months of publication.

    There is no timescale for implementing best practice, but practices should plan

    how to progress towards it where they can, identifying when they might have

    separate decontamination facilities incorporating a washer-disinfector. The

    guidance recognises, however, that it will take some practices longer than

    others to comply and some may never comply fully.  

    Where new practices are commissioned or new premises contemplated, bestpractice requirements should be adopted wherever reasonably practicable. 

    Essential quality requirements 

    Prior to sterilisation, cleaned instruments should be free of visible contaminants

    when inspected. Reprocessing using a validated decontamination process,

    which includes a cleaning and steam sterilisation (using a validated autoclave),

    should provide instruments in a sterilised state at the end of the reprocessing

    cycle. Reprocessed instruments should be stored in a way to prevent

    microbiological recolonisation. Decontamination processes should be audited

    quarterly. 

    In maintaining and developing decontamination practices, the following should

    be included: 

    • a local infection control policy, updated as necessary 

    • protocol for decontaminating instruments (as part of the infection control

    policy) 

    • storage, preparation and use of decontamination products in line with 

    COSHH Regulations 

    • procedures for management of single-use and re-usable instruments 

    • reprocessing of instrument using dedicated equipment 

    • dedicated handwashing facilities 

    • instrument cleaning using an ultrasonic bath and or manual cleaning and

    inspected to ensure free from visible contamination. Practices should plan

    for the introduction of washer-disinfectors • separation of instrument processing from other clinical work by physical or

    temporal means – the designated area for decontamination may be in or

    adjacent to a clinical room 

    • decontamination equipment should be fit for purpose and validated – it

    should be commissioned, maintained and periodically tested by a 

    Competent Person (decontamination), with records of maintenance kept and

    functioning monitored and recorded 

    • appropriate and controlled disposal of waste 

    • a documented training protocol with individual training records for all staff

    involved with decontamination 

    • an assessment of changes needed to progress to best practice 

    immunisation against hepatitis B for all staff involved in decontamination (and tetanus, if local policies require) 

    Appendix 1 

  • 8/17/2019 Infection Control in Dentistry -Smile4Dr-1

    24/29

    25 -dentalbooks-drbassam.blogspot.com

    • plan to use washer-disinfectors to clean and disinfect handpieces or use

    dedicated cleaning equipment 

    • two dedicated sinks for decontamination (in addition to the dedicated sink(s) 

    for handwashing) 

    • routine (quarterly) audits of infection control requirements (in line with HTM 

    01-05); use of the audit tool produced by the Infection Prevention Society is

    recommended 

    Essential infection control policies 

    Having the correct documentation is key to the essential quality requirements.

     All practices should have an infection control policy together with the policies

    and procedures listed below. Models of these policies and procdures are

    available on the BDA website at www.bda.org/infectioncontrol. 

    • Minimising the risk of blood-borne virus transmission, including needlestick

    injuries (policy) 

    • Decontamination and storage of dental instruments (policy) 

    • Cleaning, disinfection and sterilisation of dental instruments (procedures) 

    • Clinical waste disposal (policy) 

    •Hand hygiene (policy) 

    • Decontamination of new reusable instruments (policy) 

    • Personal protective equipment use (procedures) 

    • Management of dental instruments and equipment in infection control 

    (procedures) 

    • The use, storage and disposal of disinfectants within the practice (procedures) 

    • Spillage procedures (as part of COSHH) 

    • Environmental cleaning and maintenance (policy) 

    • Transfer of contaminated items from the treatment to decontamination area 

    (procedures) 

    •  A documented training scheme with individual training records for all staff

    engaged in decontamination. 

    Systems should be developed to ensure instruments sterilised in a non-vacuum

    autoclave are used immediately or, if wrapped after sterilisation, within 21 days.

    Instruments wrapped and sterilised in a vacuum autoclaves should be used

    within 30 days. 

    Best practice 

    Best practice is concerned with achieving higher standards in infection control

    through improvements in premises and equipment, and changes in practice

    management and the culture in which patients are treated by the dental team.

    The Department of Health has not set timescales for achieving best practice,

    recognising that it will take some practices longer than others and that some

    may never be able to comply fully. Some PCTs, however, are implementing

    local timescales for complying with best practice requirements; NHS practicesshould, therefore, be aware of local policies for achieving best practice. 

    Best practice requirements include: 

    • Installing a modern validated washer-disinfector of adequate capacity to

    remove the need for manual washing. 

    • The use of a decontamination room or rooms to provide more complete

    separation from other work activities, enhancing the distinction between

    clean and dirty workflows. Access should be restricted to those staff

    performing decontamination duties. 

    • Suitable instrument storage away from the surgery to reduce exposure to air

    and possible pathogenic contamination. Systems should ensure instrumentsare easily identified for selection and are used on a first-in, first-out principle

    within the recommended time frames. 

    http://www.bda.org/infectioncontrolhttp://www.bda.org/infectioncontrolhttp://www.bda.org/infectioncontrolhttp://www.bda.org/infectioncontrolhttp://www.bda.org/infectioncontrol

  • 8/17/2019 Infection Control in Dentistry -Smile4Dr-1

    25/29

    26 -dentalbooks-drbassam.blogspot.com

    Example layouts for decontamination areas 

    Single decontamination room, no washer-disinfector  

    Source: HTM 01-05 Decontamination in primary dental practices. Department of Health 2009 

    Single decontamination room, with washer-disinfector  

    Source: HTM 01-05 Decontamination in primary dental practices. Department of Health 2009 

  • 8/17/2019 Infection Control in Dentistry -Smile4Dr-1

    26/29

    27 -dentalbooks-drbassam.blogspot.com

    Two decontamination rooms, with washer-disinfector  

    Source: HTM 01-05 Decontamination in primary dental practices. Department of Health 2009 

  • 8/17/2019 Infection Control in Dentistry -Smile4Dr-1

    27/29

    28 -dentalbooks-drbassam.blogspot.com

    Decontamination in Scotland 

     A number of organisations provide useful guidance on compliance with

    decontamination standards in Scotland. The following guidance on

    decontamination in dentistry can be found at www.scottishdental.org. 

    SDCEP Decontamination into Practice 

    The Scottish Dental Clinical Effectiveness Programme (SDCEP) is an initiative

    of the National Dental Advisory Committee (NDAC) and is supported by the

    Scottish Executive and NHS Education for Scotland. 

    The Programme aims to provide user-friendly, evidence-based guidance for the

    dental profession in Scotland. SDCEP guidance is designed to help the dental

    team provide improved care for patients by bringing together the best available

    information and presenting it in a user-friendly way. 

    Decontamination into practice is a series of documents providng advice on

    instrument decontamination. The first part of the series Cleaning of dental

    instruments was published in March 2007 and can be downloaded from the

    SDCEP website at www.sdcep.org.uk. An introduction to the series andappendices that provide additional advice are also available to download. The

    second part of the series Sterilisation of dental instruments is in development. 

    Health Protection Scotland  

    The Scottish Health Planning Note 13 Part 2 Decontamination facilities: Local

    decontamination units provides guidance on planning and designing an LDU

    using three models: 

    • two room local decontamination unit with ante rooms 

    • two room local decontamination unit 

    • single room local decontamination unit. 

    Planning Note 13 Part 2 is available at www.scottishdental.org. 

    Chief Dental Officer Letters 

    Various CDO letters provide useful information on dental decontamination and

    capital funding for NHS practices. The following are available at

    www.scottishdental.org 

    • Capital funding to support practice improvements in decontamination of

    instruments in dental primary care (CDO letter dated September 2008) 

    • Decontamination of instruments in dental primary care (CDO letter 1 dated 

    2007) • Decontamination for primary care dental services (CDO letter 2 dated 2007) 

    The National Procurement contr act 

    NHS Scotland National Procurement has established a national contract

    (MEK005) for the provision of local decontamination equipment (ie autoclaves,

    washer disinfectors and ultrasonic cleaners). The contract includes the

    provision of equipment and its installation, commissioning, validation, periodic

    testing and maintenance. 

    The decision as to which equipment is available on the new contract, and

    therefore recommended to practitioners, is based on the outcome of a tenderevaluation process. The evaluation process includes a number of factors

    including whether the equipment meets the technical requirements, the level of

    after sales support and the cost. 

    Appendix 2 

    http://www.scottishdental.org/http://www.scottishdental.org/http://www.sdcep.org.uk/http://www.scottishdental.org/http://www.scottishdental.org/http://www.scottishdental.org/http://www.scottishdental.org/http://www.scottishdental.org/http://www.scottishdental.org/http://www.sdcep.org.uk/http://www.scottishdental.org/

  • 8/17/2019 Infection Control in Dentistry -Smile4Dr-1

    28/29

    29 -dentalbooks-drbassam.blogspot.com

    The MEK005 contract is revised and published on an annual basis. The list of

    currently approved decontamination equipment and a guide on how to access

    the contract is available at www.scottishdental.org 

    Dental Decontamination Work ing Group  

    The Scottish Dental Decontamination Working Group was formed in 2008 and

    includes representatives from the Scottish Government, the BDA and members

    of the dental profession. Its remit includes: 

    • fully establishing and quantifying the extent of the issues facing NHS Boards

    and their independent contractors and setting a baseline position 

    • exploring with NHS Boards the risk assessments and service development

    plans which are in place to support compliance with decontamination and

    sterilisation requirements in dental practice 

    • ensuring coherence across the work of Health Protection Scotland, NHS

    Education Scotland, Healthcare Facilities Scotland, National Procurement

    and NHS Boards in the provision of advice and support to NHS Boards and

    practices, and to ensure the development of strategic solutions 

    •developing an action plan which sets realistic and achievable timescalesunderpinned by risk assessment and which reflects the complexities of

    emerging technology and scientific research. 

    http://www.scottishdental.org/http://www.scottishdental.org/http://www.scottishdental.org/

  • 8/17/2019 Infection Control in Dentistry -Smile4Dr-1

    29/29

    1ll)A' dvice 

    British Dental Association 

    64 Wimpole Str eet London W1G 8YS Tel: 020 7563 4563 Fax: 020 7487 5232 

    E-mail: [email protected] www.bda.org © BDA May 2009 

    mailto:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]

Recommended