    Setup

    Pre-Use Procedures Checklist

    Calibration Connection Diagrams

    Monthly Checkout Checklist

    Circuit Connection Diagrams

    Pocket Guide

  • 7/26/2019 INOvent Pocket Guide

    2/282

    Contents

    Setup

    Setup Checklist 3

    INOvent delivery system, front viexxw 4

    Injector Module 4

    INOvent delivery system, rear view 5

    User controls and indicators on the front 6

    Pre-Use Procedures

    Pre-Use Procedures Checklist 7

    Circuit connections for purge and performance test 7

    Sample tee connection to the Manual NO delivery system 8

    Calibration Connection Diagrams

    Connection diagram for NO and NO2high range calibration 9

    Connection diagram for O2high range calibration 9

    Monthly Checkout

    Monthly System Checkout Checklist 10

    Low pressure system leak test connection diagram 11

    Circuit connections for purge and performance test 12

    Circuit Connection Diagrams

    Neonatal circuit 13

    Pediatric circuit 14

    Adult circuit 15

    Anesthesia system with ventilator 16

    BIO-MED MVP-10 or Infrasonics Infant Star 100 ventilator 17

    Self-filling resuscitator bag 18

    Hyper-inflation bag 19

    Spontaneous breathing circuit 20

    Connectors and Adapters List 21

    Changing Cylinder & Purging Regulator 22

    wImportant: This guide is provided as a convenience and for general

    information only. Never use this product without clearly and

    thoroughly understanding the most recent revision of the

    Operation and Maintenance manual. The manual is the source

    for specific, updated information regarding checklists,

    diagrams, and/or instructions contained in this Guide.

  • 7/26/2019 INOvent Pocket Guide

    3/283

    Setup

    Setup Checklist

    1. Connect supply hoses to INOvent delivery system andtherapy gas.

    2. Check cylinders for correct product identity labels and NO

    concentration.

    3. Ensure both cylinders have adequate supply pressure: not

    less than 200 psig / 14 kPa x 100 (bar).

    4. Empty and clean the fluid trap if necessary.

    5. Connect sample line to the INOvent delivery system.

    6. Connect the injector module tubing and electrical cable to

    both the INOvent delivery system and injector module.

    7. Connect the manual NO delivery system.

    8. Connect the electrical power cord to an emergency-power-

    backed, hospital-grade outlet.

  • 7/26/2019 INOvent Pocket Guide

    4/284

    Setup

    INOvent delivery system, front view

    Electrical cable

    connection to the

    INOvent Delivery

    System

    Inspiratory limb

    breathing circuit;

    from ventilator

    NO/N2injector tube

    connection to theINOvent Delivery System

    Inspiratory limb breathing

    circuit; to humidifier

    Injector Module

    Display andoperating panel

    Manual NO DeliverySystem Activity Indicator

    NO/N2Injector

    Tube connectorHose clip

    Injector Module

    Electrical Cable

    connector

    Fluid

    Trap Bottle

    Patient Gas

    sample line

    connector

    Fluid Trap FilterCartridge

  • 7/26/2019 INOvent Pocket Guide

    5/285

    Setup

    INOvent delivery system, rear view

    NO2Sensor

    NO

    Sensor

    Connector

    to the O2Flowmeter

    NO/N2Input HoseConnectors (2)

    Regulator Low

    Pressure Hose

    with Quick

    Connectors (2)

    Samplegas outlet

    O2

    Sensor Cooling Fanand Filter

    ON/STANDBYswitch

    External

    Ground

    connection

    point

    Power Cord

    connector

    Power Entry

    Module

    Serial Port

    connector

    Nurse Call

    connector

  • 7/26/2019 INOvent Pocket Guide

    6/286

    Setup

    User controls and indicators on the front

    Inspired NO2Measured Value

    Alarm

    message

    areaAlarm

    Silenced

    Countdown

    Timer

    Alarm

    Silence

    button

    Mains

    Power

    light

    Inspired O2Measured Value Alarms

    button

    Setup

    button

    Calibration

    button

    Pause

    Flow

    button

    Set NO

    button

    Control

    Wheel

    NO

    Bargraph

    Alarm

    Historybutton

    Alarm

    History

    indication

    area

    Inspired NO

    Measured

    Value

  • 7/26/2019 INOvent Pocket Guide

    7/287

    Pre-Use Procedures

    Pre-Use Procedures Checklist

    1. Do the Initial connections and leak test.

    2. Do the System purge and performance tests.

    3. Do the Manual NO delivery system purge and

    performance test.

    Circuit connections for purge and performance test

    NO/N2Injector Tube

    Injector Module

    Electrical Cable

    O2Flowmeter

    O2Tubing

    15M x

    4.5 mm

    Adapter

    22M/15F x

    22M/15F

    Adapter

    Injector

    Module

    300 mm of

    22 mm hose

    Sample

    Tee

    Patient Gas

    sample line

  • 7/26/2019 INOvent Pocket Guide

    8/288

    Pre-Use Procedures

    Sample tee connection to the

    Manual NO delivery system

    NO/O2outlet

    O2inlet O2supply hose from

    the O2flowmeter

    Oxygen tubing

    (user supplied)

    15M x 4.5 mm

    Adapter

    Patient Gas

    sample line

    Sample

    Tee

    NO/N2input

    hoses fromthe regulators

    Pressure

    CompensatedO2flowmeter

    (user supplied),

    shown set to OF

    (no float activity

  • 7/26/2019 INOvent Pocket Guide

    9/289

    Calibration Connection Diagrams

    Connection diagram for NO and

    NO2high range calibration

    Patient Gas

    sample line

    Calibration gas

    cylinder

    pressure gauge

    NO or N2calibration

    gas cylinder

    Cylinder

    ON/OFF

    control

    Vent to the atmosphere or

    scavenge - Do not occlude

    One-way

    valve

    Connection diagram for O2high range calibration

    100% O2 Flowmeter Sample TeeTo the

    Atmosphere

    O2to theSample System

    INOvent delivery

    system Sample Line

  • 7/26/2019 INOvent Pocket Guide

    10/2810

    Monthly Checkout

    Monthly System Checkout Checklist

    The monthly system checkout consists of:

    1. Initial connections and leak tests.

    2. Purge and system alarms tests.

    3. Calibration and monitoring alarms.

    4. INOvent delivery system performance.

    5. Manual NO delivery system purge and performance.

  • 7/26/2019 INOvent Pocket Guide

    11/2811

    Monthly Checkout

    Low pressure system leak test connection diagram

    NO/O2outlet

    O2inletInjector

    Module

    Electrical

    CableO2tubing

    15M x

    4.5 mm

    Adapter

    NO/N2input

    hoses from

    the regulators

    Pressure

    Compensated

    O2flowmeter

    Injector

    Module

    Squeeze

    Bulb

    O2supply

    hose from the

    O2flowmeter

    22M/15F x

    22M/15F

    Adapter

    NO/N2Injector Tube

  • 7/26/2019 INOvent Pocket Guide

    12/2812

    Monthly Checkout

    Circuit connections for purge and performance test

    NO/N2Injector Tube

    Injector Module

    Electrical Cable

    O2Flowmeter

    O2Tubing

    15M x

    4.5 mm

    Adapter

    22M/15F x

    22M/15F

    Adapter

    Injector

    Module

    300 mm of

    22 mm hose

    Sample

    Tee

    Patient Gas

    sample line

  • 7/26/2019 INOvent Pocket Guide

    13/2813

    Patient Gas

    sample line

    9 mm I.D.

    15 mm I.D.

    to 3/8 in.

    Tubing

    Adapter

    22M /15F x

    15M Adapter

    with gas

    sample tee

    15 mm O.D. to

    3/8 in. Tubing

    AdapterHumidifier

    22F x 15MAdapter

    To

    INOvent

    delivery

    system

    22M/15F x22M/15F

    Adapter

    Ventilator

    Inspiratory port

  • 7/26/2019 INOvent Pocket Guide

    14/2814

    Circuit Connection Diagrams

    Pediatric circuit

    Patient Gas

    sample line

    15 mm,

    6 inch

    Inspiratory

    Extention

    22M/15F x

    15M Adapter

    with gas

    sample tee

    Humidifier

    22F x 15MAdapter

    To

    INOvent

    delivery

    system

    22M/15F x22M/15F

    Adapter

    Ventilator

    Inspiratory port

  • 7/26/2019 INOvent Pocket Guide

    15/2815

    Circuit Connection Diagrams

    Adult circuit

    Patient Gas

    sample line

    22M/15F x

    22M/15F

    Adapter

    22M/15F x

    15M Adapter

    with gas

    sample tee

    Humidifier

    22F x 15M

    Adapter

    To

    INOvent

    delivery

    system

    22M/15F x

    22M/15F

    Adapter

    Ventilator

    Inspiratory port

  • 7/26/2019 INOvent Pocket Guide

    16/2816

    Circuit Connection Diagrams

    Anesthesia system with ventilator

    Patient Gas sample

    line connection

    (Sample Inlet port)

    Patient Gas

    sample line

    Patient

    wye

    Sample

    Tee

    22M/15F x

    22M/15FAdapter

    Expiratory

    port

    Inspiratory

    port

    Connections

    on absorber

    Injector

    Module

    INOvent

    delivery

    system

    Hose

  • 7/26/2019 INOvent Pocket Guide

    17/2817

    Circuit Connection Diagrams

    BIO-MED MVP-10 or Infrasonics

    Infant Star 100 ventilator

    Ventilator

    Expiratory

    Valve

    Ventilator INOvent

    delivery

    system

    Ventilator

    Inspiratory

    port

    22M/15F x

    22M/15F

    Adapter

    Injector

    Module

    Electrical

    Cable

    NO/N2

    Injector

    Tube

    Injector

    Module

    Inspiratory

    Breathing

    Circuit Hose

    Expiratory

    Breathing

    Circuit Hose

    Sample

    Tee

    Patient

    wye

    Patient Gas

    sample line

  • 7/26/2019 INOvent Pocket Guide

    18/2818

    Circuit Connection Diagrams

    Self-filling resuscitator bag

    wCaution: Tubing length downstream of the O2tubing sample tee

    must be minimized. The INOvent monitoring systemoperates at pressures up to 70 cm H2O above ambient

    at the sample tee. Excessive tubing length can result in

    Monitoring Failure alarm from over-pressurization. To

    help ensure proper INOvent delivery system operation,

    test the circuit configuration before using it.

    NO/N2Injector Tube

    Injector Module

    Electrical Cable

    O2Flowmeter

    O2Tubing

    15M x

    4.5 mm

    Adapter

    22M/15F x

    22M/15F

    Adapter

    Injector

    Module

    15M x

    4.5 mm

    Adapter

    O2Tubing

    Patient Gas

    sample line

    O2Tubing

    Sample

    Tee

    Resuscitator bagwith O2reservoir

  • 7/26/2019 INOvent Pocket Guide

    19/2819

    Circuit Connection Diagrams

    Hyper-inflation bag

    NO/N2Injector Tube

    Injector Module

    Electrical Cable

    O2Flowmeter

    O2Tubing

    15M x

    4.5 mm

    Adapter

    22M/15F x

    22M/15F

    Adapter

    Injector

    Module

    15M x

    4.5 mm

    Adapter

    O2Tubing

    Patient Gas

    sample line

    O2Tubing

    Sample

    Tee

    Hyper-

    inflation

    Bag

    Pressure

    Gauge

  • 7/26/2019 INOvent Pocket Guide

    20/2820

    Circuit Connection Diagrams

    Spontaneous breathing circuit

    NO/N2Injector Tube

    Injector Module

    Electrical Cable

    O2Flowmeter

    O2Tubing

    15M x

    4.5 mm

    Adapter22M/15F x

    22M/15F

    Adapter

    Injector

    Module

    22M/15F x

    22M/15F

    Adapter

    Sample

    Tee

    Patient Gas

    sample line

    One-way

    Valves

    Face

    Mask

    Breathing

    Circuit

    Bag

    Breathing

    Circuit

    Tee

    Breathing

    Circuit

    Hose

  • 7/26/2019 INOvent Pocket Guide

    21/2821

    Connectors and Adapters List

    Adapters:15M x 4.5 mm

    22M / 15F x 22M / 15F22F x 15M

    22M /15F x 15M with Sample Tee

    15 mm O.D. to 3/8 in. Tubing Adapter

    15 mm I.D. to 3/8 in. Tubing Adapter

    Others:

    O2Tubing Sample TeeOne-way Valve

    15 mm, 6 inch Inspiratory Extension

    O2Tubing

    9 mm I.D. Hose

    300 mm of 22 mm Hose

    Breathing Circuit Hose

  • 7/26/2019 INOvent Pocket Guide

    22/2822

    Changing Cylinder & Purging Regulator

    wWARNING You MUST purge the regulator assembly

    immediately before using a new NO cylinder.

    wImportant: Replace NO therapy gas cylinder when its

    pressure is < 200 psig / 14 kPa x 100 (bar).

    1. Purge regulator assembly before using new NO therapy gas cylinder:

    a. Determine which regulator assembly is not being used and

    requires purging. Cylinder valve should be closed on

    regulator assembly not being used.

    b. On the regulator assembly requiring purging, disconnect the

    low pressure hose quick-connect fitting from the NO/N2input on the rear of the INOvent delivery system.

    c. Open the cylinder valve on the new NO therapy gas cylinder.

    d. Close the cylinder valve on the new NO therapy gas cylinder.

    f. Insert low pressure hose quick-connect fitting into purge

    manifold:

    g. Firmly push and hold the quick-connect fitting in place while

    the pressure drops to zero on the regulator gauge.

    h. After the pressure drops to zero (about 15 seconds),

    reconnect the low pressure hose quick-connect fitting to the

    NO/N2

    input on the rear of the INOvent delivery system.

    2. Open the cylinder valve on the new cylinder.

    3. Close the cylinder valve on the empty cylinder and label appropriately.

    4. Replace the empty NO therapy gas cylinder. Leave the cylinder valve

    on the replacement cylinder turned off until needed.

    5. Purge the regulator assembly immediately before use.

    Purge

    Manifold Low Pressure Hose

    Quick-Connect Fitting

  • 7/26/2019 INOvent Pocket Guide

    23/2823

    Notes

  • 7/26/2019 INOvent Pocket Guide

    24/2824

    Notes

  • 7/26/2019 INOvent Pocket Guide

    25/2825

    Notes

  • 7/26/2019 INOvent Pocket Guide

    26/2826

    Notes

  • 7/26/2019 INOvent Pocket Guide

    27/2827

    Notes

  • 7/26/2019 INOvent Pocket Guide

    28/28

    wImportant: This guide is provided as a convenience and for general

    information only. Never use this product without clearly and