Installation and operating instructions VS 300 S / VS 600 / VS 900 combi-suction unit 2008/04
Transcript
Page 1: Installation and operating instructions VS 300 S / VS 600 ...€¦ · The combi-suction unit is a suction unit with an integrated separation unit. Separation within the treatment

Installation and operating instructionsVS 300 S / VS 600 / VS 900

combi-suction unit

2008/04

Page 2: Installation and operating instructions VS 300 S / VS 600 ...€¦ · The combi-suction unit is a suction unit with an integrated separation unit. Separation within the treatment

2

Page 3: Installation and operating instructions VS 300 S / VS 600 ...€¦ · The combi-suction unit is a suction unit with an integrated separation unit. Separation within the treatment

3

Contents

Important Information

1. Notes ......................................................... 41.1 Test of conformity .............................. 41.2 General Notes .................................... 41.3 General Safety Notes ........................ 41.4 Notes concerning

Medical Appliances ........................... 51.5 Using Peripheral Devices .................. 51.6 Safety notes concerning electric

current ................................................ 51.7 Warnings and Symbols ..................... 5

2. Product information ................................ 62.1 Correct Usage ................................... 62.2 Incorrect Usage ................................. 62.3 Product Description ........................... 6

3. Delivery Contents .................................... 63.1 Suction unit VS 300 S ........................ 63.2 Suction Unit VS 600 ........................... 73.3 Suction Unit VS 900 ........................... 7

4. Technical Data ........................................... 84.1 Suction Unit VS 300 ........................... 84.2 Suction Unit VS 600 ........................... 94.3 Suction Unit VS 900 ......................... 104.4 Environmental conditions ................ 11

5. Functional description .......................... 116. Functional description .......................... 12

Installation

7. Set-up ..................................................... 137.1 Set-up location ................................. 137.2 Alternative set-ups ........................... 137.3 Securing the suction unit ................. 137.4 Fitting a pressure

compensation hose ......................... 137.5 Rinsing Unit ..................................... 137.6 Plumbing materials .......................... 147.7 Hose materials ................................. 147.8 Laying hoses and pipes .................. 14

8. Connections ........................................... 158.1 VS 300 connections ......................... 158.2 VS 600 connections ......................... 168.3 VS 900 connections ......................... 17

9. Electrical Connections ......................... 189.1 Connection requirements ................ 189.2 Control unit (VS 600 + VS 900) ...... 189.3 Motor terminal box connections

(VS 600 + VS 900 ............................ 189.4 Connection of VS 300 S to

Control Unit ...................................... 1910. Commissioning ..................................... 19

Care

11. Cleaning and Disinfection of theSuction Unit ........................................... 20

12. Maintenance ........................................... 20

Disposal

13. Appliance disposal ................................ 20

Trouble-shooting

14. Tips for Technicians .............................. 21

Page 4: Installation and operating instructions VS 300 S / VS 600 ...€¦ · The combi-suction unit is a suction unit with an integrated separation unit. Separation within the treatment

4

ImportantInformation

1. Notes1.1 Test of conformityThis product was tested for conformity to theGuidelines 93/42/EWG of the European Unionand has been found to satisfy all criteria ofthese guidelines.

1.2 General Notes• These Installation and Operating Instructions

form an integral part of the unit. They mustbe kept close to the unit at all times. Preciseobservance of these instructions is a pre-condition for use of the unit for the intendedpurpose and for its correct operation. Newpersonnel must be made aware of thecontents, and they should be passed on tofuture operating staff.

• Safety for the operator as well as trouble-free operation of the unit are only ensured ifuse is made of original equipment parts.Moreover, use may only be made of thoseaccessories that are specified in thetechnical documentation or that have beenexpressly approved and released by DürrDental for the intended purpose.

• Dürr Dental cannot guarantee for the safetyor proper functioning of this unit in the casewhere parts or accessories are used whichare not supplied by Dürr Dental.

• Dürr Dental are only responsible for theequipment with regard to safety, reliabilityand proper functioning where assembly,resettings, changes or modifications,extensions and repairs have been carriedout by Dürr Dental or an agency authorizedby Dürr Dental and if the equipment is usedin conformity with the Installation andOperating Instructions.

• These Installation and OperatingInstructions conform to the relevant versionof the equipment and the underlying safetystandards valid at the time of going to press.All switches, processes, trade marks,software programs and appliances named inthis document are registered names.

• Any reprinting of the technicaldocumentation, in whole or in part, is subjectto prior approval of Dürr Dental being givenin writing.

1.3 General Safety NotesThis appliance has been designed andconstructed by Dürr Dental so that correctusage of the appliance is virtually free of anypossible injury or danger. In spite of this, wefeel it is our duty to mention the followingsafety measures in order to prevent anypossible danger.

• When using this appliance all local andrelevant regulations must be observed!Converting or modifying the appliance inany way is strictly prohibited. In such cases,any and all guarantees immediately becomeinvalid. The operation of modifiedappliances can be punishable by law. In theinterests of trouble-free operation theoperator is responsible for observing theseregulations.

• Retain the packaging for possible return ofthe product to the manufacturers. Ensurethat the packaging is kept out of the reach ofchildren. Only the original packagingprovides adequate protection duringtransport of the unit.Should return of the product to themanufacturers be necessary during theguarantee period, Dürr Dental accepts noresponsibility for damage occurring duringtransport where the original packaging wasnot used!

• Before every use the operator must checkthe functional safety and the condition of theappliance.

• The operator must be knowledgeable in theoperation of the appliance.

• The product is not designed to be used inmedical treatment areas where there existsthe danger of explosion. Areas whereexplosions could occur are those whereflammable anesthetic material, skincleansers, oxygen and skin disinfectants arepresent. This appliance is not to be used inareas where the atmosphere could causefire.

Page 5: Installation and operating instructions VS 300 S / VS 600 ...€¦ · The combi-suction unit is a suction unit with an integrated separation unit. Separation within the treatment

5

1.4 Notes concerning MedicalAppliances

• This product is a medical technicalappliance and may only be used by thosepersons whose training and/or experienceguarantees correct usage.

1.5 Using Peripheral Devices• Units may only be connected to the system

or to other units when it has beenestablished that there is no reduction ofsafety for the patient, the operator or theenvironment through such connection.Where it is not absolutely clear from thedocumentation whether safety is reduced bysuch connection, then the operator mustestablish, e.g. by contacting either themanufacturer or an expert, that there is noreduction of safety for the patient, theoperator or the environment through suchconnection.

1.6 Safety notes concerning electriccurrent

• This unit may only be connected to astandard approved earthed electrical socket(VS 600 + VS 900).

• Before connecting the appliance to themains, check that the frequency andvoltages given for the appliance matchthose of the available power supply.

• Before commissioning the appliance allconnections must be checked for possibledamage. Damaged connections, plugs andsockets should be replaced immediately.

• Never touch patients and open sockets ofthe appliance simultaneously.

• All relevant electrical rules and regulationsmust be observed during installation andwhen carrying out any repairs ormaintenance on the appliance.

1.7 Warnings and SymbolsThe operating and installation instructionscontain the following labeling and symbols forespecially important information.

Restrictions and regulationsconcerning the prevention of injuryor damage.

Warning concerning dangerouselectrical voltage.

Special instructions concerningeconomic use of the appliance orother notes.

Obseve notes in supplementarydocumentation

For added safety of operatorsprotective gloves should be wornwhile working on the Suction Units.

Take environmental influences intoconsideration.

Date of manufacture.

Ground connection.

Fuse.

Recycling

~ Single phase AC current.

3~ Three-phase AC current.

3N~ Three-phase AC current with centralconductor.

Page 6: Installation and operating instructions VS 300 S / VS 600 ...€¦ · The combi-suction unit is a suction unit with an integrated separation unit. Separation within the treatment

6

3. Delivery Contents

The items listed under SpecialAccessories are not included in thestandard delivery contents, but can beordered specially.

3.1 Suction unit VS 300 S3.1.1 Contents

Model 7122-01/002Typ 230 V, 1~, 50 Hzwith control unit

Model 7122-02/002Typ 230 V, 1~, 60 Hzwith control unit

Model 7122-05/003Typ 100 V, 1~, 50 - 60 Hzwith control unit

3.1.2 Accessories

Connector set ............................... 7122-001-00Suction hose LW 30, grey ............ 9000-317-27Hose LW 20 .................................. 9000-317-22Hose LW 30, Aluminium ............... 9000-317-37OroCup (not Japan) ..................... 0780-350-00

3.1.3 Special Accessories

Wall mounting ............................... 7130-190-00Housing ........................................ 7122-200-00Exhaust bacterial filter withaccessories .................................. 7120-143-00Rinsing unit ................................... 7100-250-50

2. Product information2.1 Correct UsageThe suction unit has been designed toproduce a vacuum in order to suck up saliva,rinsing water and other fluids which occurduring dental treatment and which aretransported to the waste water system.Installation in medical facilities:

As far as was possible, all requirementsconcerning Medical Products have been takeninto consideration in the design andproduction of these appliances. Thus, theappliance may be installed in any facilityproviding medical treatment.Where the appliance is set up in a medicalclinic or other facility, then the requirementsdemanded under directive 93/42 EWG as wellas all relevant standards must be observed oninstallation.

2.2 Incorrect UsageAny usage above and beyond that explicitlylaid down in the operating instructions isdeemed to be incorrect usage. Themanufacturer accepts no liability for damageor injury resulting from incorrect use. All risk iscarried by the operator.

2.3 Product DescriptionThe combi-suction unit is a suction unit with anintegrated separation unit. Separation withinthe treatment station (e.g. chair) is no longernecessary.The suction unit separates the fluids and solidparticles sucked up during treatment from theair using a two-step separation system ofcyclone separator and separation turbine.The separation turbine is especially effectiveat preventing fluids and blood foam frombeing sucked into the turbine section of thesuction unit.The fluids sucked up are continually beingrotated at high speed and transported to thewaste system; thus, there is no interruption ofsuction due to the system being too full.The suction system is mounted on rubbermountings, which reduce vibration andoperating noise.

Page 7: Installation and operating instructions VS 300 S / VS 600 ...€¦ · The combi-suction unit is a suction unit with an integrated separation unit. Separation within the treatment

7

3.2 Suction Unit VS 6003.2.1 Contents

Model 7128-01/002Type 230 V, 1~, 50 Hzwith control unit

Model 7128-02/002Type 400 V, 3~, 50–60 Hzwith control unit

Model 7128-02/003Type 230 V, 3~, 50-60 Hzwith control unit

Model 7128-05/003Type 200 V, 3~, 50-60 Hzwith control unit

3.2.2 Accessories

Control unitfor type 7128-01/002 .................... 0700-500-50for type 7128-02/002 .................... 0732-100-56for type 7128-02/003 .................... 0732-100-57for type 7128-05/003 .................... 0732-100-57Pipe connector set ....................... 7128-001-00Hose LW 40 .................................. 9000-318-70Hose LW 50.. ........................ 9000-317-002Hose LW 20 .................................. 9000-317-22OroCup ......................................... 0780-350-00

3.2.3 Special accessories

Noise reducing cover ................... 7128-991-00Pressure compensation hose ...... 7112-101-00Wall mounting ............................... 7130-190-00Floor installation console ............. 7130-191-00Bacterial exhaust filter ................. 0732-001-00Fixing plate for exhaust filter ....... 0732-000-06Noise reducer for exhaust ........... 0730-991-00Rinsing Unit .................................. 7100-250-50

3.3 Suction Unit VS 9003.3.1 Contents

Model 7133-01/001Type 230 V, 1~, 50 Hzwithout control unit

Model 7133-02/001Type 400 V, 3~, 50 Hzwithout control unit

Model 7133-01/002Type 230 V, 1~, 50 Hzwith control unit

Model 7133-02/002Type 400 V, 3~, 50 Hzwith control unit

Model 7133-03/002Type 230 V, 3~, 50 Hzwith control unit

Model 7133-05/002Type200 V, 3~, 50-60 Hz230 V, 3~, 60 Hzwith control unit

3.3.2 Accessories

Control unitfor type 7133-01/002 .................... 0732-100-55for type 7133-02/002 .................... 0732-100-56for type 7133-03/002 .................... 0732-100-57for type 7133-05/002 .................... 0732-100-59Pipe connector set ....................... 7133-001-00Hose LW 20 .................................. 9000-317-22Hose LW 50 (0.6m) .................... 9000-317-001Hose LW 50 (1.5m) .................... 9000-317-002OroCup ......................................... 0780-350-00

3.3.3 Special accessories

Noise reducing cover ................... 7128-991-00Pressure compensation hose ...... 7130-991-00Wall mounting ............................... 7130-190-00Floor installation console ............. 7130-191-00Bacterial exhaust filter ................. 0732-001-00Fixing plate for exhaust filter ........ 0732-000-06Noise reducer for exhaust ........... 0730-991-00Rinsing unit ................................... 7100-250-50

Page 8: Installation and operating instructions VS 300 S / VS 600 ...€¦ · The combi-suction unit is a suction unit with an integrated separation unit. Separation within the treatment

8

2 3

4. Technical Data4.1 Suction Unit VS 300 SModel 7122 -01 -02 -05

Voltage V 230 230 100

Supply frequency Hz 50 60 50 - 60

Phases 1 1 1

Rated current A 2.9 3.7 8,0 - 10

Residual current A 8.2 9.1 21 - 20,5

Motor protection switch A Motor winding overheat protector 160°C (±5°C)

Power consumption W 580 800 650 - 850

R.P.M. min-1 2750 3100 2810 - 3220

Max. flow volume l/min 4

Air flow l/min see Fig. 3

Max. No. treatment stations 1

Weight kg 12,5

Dimensions see Fig. 2

Noise level* dB(A), ±1.5 63 - 64

Duty cycle % 100

Protection type IP 24

Protection class I

Vacuum connection DürrConnect Spezial (hose ø 30 mm (inner))

Exhaust air connection DürrConnect Spezial (Aluminium hose ø 30 mm (inner))

Waste water connection DürrConnect System (hose 20 mm (inner))

Auxiliary air vent setting mbar –

Protective low voltageProtective low voltageProtective low voltageProtective low voltageProtective low voltage V 24 ~

Power outputPower outputPower outputPower outputPower output VA 4

* According to EN ISO 1680 air noise emissions; measured in soundproof room. Higher values will be achieved in reverberant rooms.

Page 9: Installation and operating instructions VS 300 S / VS 600 ...€¦ · The combi-suction unit is a suction unit with an integrated separation unit. Separation within the treatment

9

4 5

4.2 Suction Unit VS 600Model 7128 -01 -02 -05

Voltage V 230 400 230 200

Supply frequency Hz 50 50 - 60 50 - 60 50 - 60

Phases 1 3 3 3

Rated current A 5,0 1.8 - 2.3 3.1 - 4.1 3.2 - 4.0

Residual current A 22 8 9 31

Motor protection switch A – 2.5 - 4.0 3.5 - 4.5 4

Power consumption W 1100 1000 1420 1420

R.P.M. min-1 2850 2850/3300

Max. flow volume l/min 10

Air flow l/min see Fig. 5

Max. No. treatment stations 3

Weight kg 25

Dimensions see Fig. 4

Noise level* dB(A), ±1.5 63

Duty cycle %ED 100

Protection type IP 44

Protection class I

Vacuum connection ø 40 mm (outer) (DN 40)

Exhaust air connection ø 50 mm (outer)

Waste water connection DürrConnect System

Auxiliary air vent setting mbar 170 (170 hPa)

* According to EN ISO 1680 air noise emissions; measured in soundproof room. Higher values will be achieved in reverberant rooms.

Page 10: Installation and operating instructions VS 300 S / VS 600 ...€¦ · The combi-suction unit is a suction unit with an integrated separation unit. Separation within the treatment

10

76

4.3 Suction Unit VS 900Model 7133 -01 -02 -03 -05

Voltage V 230 400/230 230 200 230

Supply frequency Hz 50 50 50 50-60 60

Phases 1 3 3 3 3

Rated current A 6.5 2.5/4.0 4.0 5.7-6.7 6.7

Residual current A 29 14/24 24 33 34

Motor protection switch A 7.5 2.7/4.6 4.5 6,0-7,0 7,0

Power consumption W 1480 1520 1420 1500-2080 2130

R.P.M. min-1 2770 2820 2820 2810-3200 3310

Max. fluid flow l/min 16

Air flow l/min see Fig. 7

Max. No. treatment stations 5

Weight kg 35

Dimensions see Fig. 6

Noise level* dB(A), ±1.5 64

Duty cycle % 100

Protection type IP 44

Protection class I

Vacuum connection ø 47 mm (outer)

Exhaust air connection ø 50 mm (outer)

Waste water connection DürrConnect System

Auxiliary air vent setting mbar 170 (17 hPa)

* According to EN ISO 1680 air noise emissions; measured in soundproof room. Higher values will be achieved in reverberant rooms..

Page 11: Installation and operating instructions VS 300 S / VS 600 ...€¦ · The combi-suction unit is a suction unit with an integrated separation unit. Separation within the treatment

11

4.4 Environmental conditionsTake environmental influences intoaccount. The appliance must not beused in damp or wet locations.

Storage and transportTemperature: ......................... –10 °C to +60 °CRel. humidity: .................................... max. 95%

OperationTemperature: ........................ +10 °C to +40 °CRel. humidity: .................................... max. 70%

5. Functional descriptionVS 900 as example of principle function

In the suction unit the fluids and solid particleswhich are sucked up are separated from airusing a two-step separation system Thisseparation system consists of a cycloneseparator and a separation turbine.The suction process runs continuously. Theintake mixture of fluid, solid particles and airenters the suction unit via the intakeconnection (D). The coarse filter (B) retains thelarger solid particles. The remaining mixtureproceeds to the cyclone separator (I) and isthen rotated in a spiral action. In this 1st stagethe fluids and any remaining solid particles arethrown against the outer walls of the cycloneseparator by the centrifugal action. Initiallythere is only a "coarse separation“ of thefluids.In the following 2nd stage the separationturbine (J) sets a "fine separation“ into play, inwhich the remaining fluids are now separatedfrom the air which has transported them thusfar.The waste water pump (H) now transports thefluids from the centrifuge together with the finesolid particles to the waste water connection(E) and the central waste water system.The air, now free of any fluids, is transportedto the exhaust air system (C) using thevacuum provided by the action of the turbinewheel (K).The turbinen wheel and the waste water pumpare driven by the motor (L).

In order to separate dental amalgam itis necessary to install an amalgamseparator, e.g. model 7801-07, afterthe waste connection (E).

Should an amalgam separator fromanother manufacturer be fitted, thenthe max. fluid flow rate of thesuction unit must be taken intoaccount.

Page 12: Installation and operating instructions VS 300 S / VS 600 ...€¦ · The combi-suction unit is a suction unit with an integrated separation unit. Separation within the treatment

12

6. Functional description

A Auxiliary air ventB Coarse filterC Exhaust air connectionD Intake sleeveE Waste water connectionF Diaphragm valve

G Exhaust air noise reducerH Waste water pumpI Cyclone separatorJ Separation turbineK Turbine wheelL Motor

B

C

D

K

I

L

J

A

F

G

H

E

Page 13: Installation and operating instructions VS 300 S / VS 600 ...€¦ · The combi-suction unit is a suction unit with an integrated separation unit. Separation within the treatment

13

Installation

7. Set-up7.1 Set-up location• The ambient room temperature must not fall

below 10 °C and should not be allowed torise above 40 °C. The relative humidity mustnot exceed 70%.

• Installation in a purpose-built room, e.g. inboiler room, must be approved before-hand(i.e. observe local regulations).

• Installation in wet rooms is not permissible.

• When installing this unit inside a cabinet orin a machinen room, sufficient air inlet andoutlet openings must be provided; thesemust have a minimum of 120 cm² cross-section.If insufficient ventilation is available then afan must be fitted, this must be able toprovide at least 2 m3/min ventilation flow;additionally, an inlet opening must beprovided for cool air.

7.2 Alternative set-ups• On the same floor as the surgery.

• In a ventilated cabinet (e. g. Dürr PTS 105/195).

• In Dürr-housing (VS 300 S only)as extension to the treatment unit andconnected to electrical floor connection.

• On a lower floor than the surgery.

VS 300 S

When installing the VS 300 S in a cellar orsimilar room, then the unit should be mountedon a platform or on the wall at a height of 30cm above the floor.

7.3 Securing the suction unit• When setting up the suction unit together

with an amalgam separator a floor consoleshould be included.

The suction unit must be at least20 cm above any amalgam separatorwhich is also installed.

• For mounting on the wall we recommendusing the Dürr wall-mounting unit.

Information concerning fitting can befound in the installation instructionswhich are supplied with the floorconsole or the wall-mounting unit.

7.4 Fitting a pressure compensationhose

• For suction units VS 600 and VS 900 theinstallation of a pressure compensation hoseis necessary where an amalgam separator ispresent.

Due to the high rate of fluid flowthrough these units, a pressurecompensation hose is requiredbetween the suction unit and amalgamseparator and serves as intermediatestorage in cases of spontaneous rushof water.

Information on installation can befound in the instructions supplied withthe pressure compensation hose.

7.5 Rinsing Unit It is strongly recommended that a rinsing unitbe added to the suction unit for surgicaltreatment. This enables a small amount ofwater to be fed into the system during suctionwhich serves to dilute the secretions and allowthem to be transported through the systemmore easily.This rinsing unit should either be integratedinto the treatment station itself or can be setup in the vicinity of the suction unit.

Page 14: Installation and operating instructions VS 300 S / VS 600 ...€¦ · The combi-suction unit is a suction unit with an integrated separation unit. Separation within the treatment

14

7.6 Plumbing materialsOnly the following pipe materials may beused:

Vacuum-sealed HT-drain pipe made frompolypropylene (PP), chlorinated polyvinyl-chloride (PVC-C), unplasticized polyvinyl-chloride (PVC-U) or polyethylene (PEh).

Do not use:Acrylonitrile-Butadiene-Styrene(ABS) or Styrene-Copolymer blends(e.g. SAN+PVC).

7.7 Hose materialsFor waste water and suction connections useonly PVC flexible spiral hoses with integratedspiral or hoses of equivalent type.

Do not use:Hoses which are not resistant todental disinfectants and otherchemicals, or rubber or PVC hoseswith insufficient flexibility.

7.8 Laying hoses and pipesWaste water connections should becompleted in accordance with the localregulations.

The connection between supply andthe suction unit itself should be kept asshort as possible, be straight withoutbends and carried out with the flexiblehose supplied. This will avoid anyvibration being effected in theplumbing system.

Page 15: Installation and operating instructions VS 300 S / VS 600 ...€¦ · The combi-suction unit is a suction unit with an integrated separation unit. Separation within the treatment

15

8. ConnectionsThe connections illustrated here areonly meant as examples which can bevaried according to the conditionspresent in the individual surgery.

8.1 VS 300 connections

1 Connector 30/361a O-Ring

2 Hose clip ø30mm3 Exhaust hose (Aluminium) inner ø30mm4 Angle piece DN 305 O-Ring 30x26 Securing ring7 Plug, external ø36mm8 O-Ring 20x2.09 Securing ring

10 Hose socket ø25mm11 Hose clamp ø28mm12 Suction hose, internal ø30mm13 Double plug14 Waste water hose, internal ø20mm

Page 16: Installation and operating instructions VS 300 S / VS 600 ...€¦ · The combi-suction unit is a suction unit with an integrated separation unit. Separation within the treatment

16

8.2 VS 600 connections

5 O-Ring 30x27 Plug, external ø36mm8 O-Ring 20x2,09 Securing ring

10 Hose socket ø20mm11 Hose clamp ø28mm13 Double plug14 Waste water hose, internal ø20mm20 Elbow DN5021 Hose clamp ø55mm22 Exhaust air hose, internal ø50mm23 Hose plug DN40/5024 Hose clamp ø46mm25 Suction hose ø40mm

Page 17: Installation and operating instructions VS 300 S / VS 600 ...€¦ · The combi-suction unit is a suction unit with an integrated separation unit. Separation within the treatment

17

8.3 VS 900 connections

5 O-Ring 30x27 Plug, external ø36mm8 O-Ring 20x2.09 Securing ring

10 Hose socket ø20mm11 Hose clamp ø28mm13 Double plug14 Waste water hose, internal ø20mm20 Elbow DN5021 Hose clamp ø55mm22 Exhaust air hose, internal ø50mm30 Snap-lock connector, straight31 Snap-lock connector, bent32 Collar seal33 Union nut34 Suction hose, internal ø55

Page 18: Installation and operating instructions VS 300 S / VS 600 ...€¦ · The combi-suction unit is a suction unit with an integrated separation unit. Separation within the treatment

18

9. Electrical ConnectionsThe electrical supply unit must conform to allnational rules and regulations concerningpower supply to surgeries and clinics.Where electrical connection to the mains is viafloor or ceiling an all-polar isolating device (all-polar switch or all-polar power safety switch(fused)) with > 3mm contact opening widthmust be built into the system.Fusing: LS-Switch 16 A, Characteristics B, Cand D according to EN 60898

Electrical connection to the mainpower supply using a shockproofplug or CCE-type plug is notpermitted

9.1 Connection requirements100–110 V / 230 V / 400 V cables(permanently fitted mains supply):

• NYM-J 3 x 1.5 mm² / 5 x 1.5 mm²

100–110 V / 230 V / 400 V cables (mainssupply, flexible):

The connection between control unit andsuction unit or between mains supply socketand suction unit should be of PVC-sleevedcable:H05 VV-F 5G1.5 mm² / 5G1.5 mm²or rubber-sleeved:H05 RN-F 3G1.5 mm² / 5G1.5 mm²,H05 RR-F 3G1.5 mm² / 5G1.5 mm²A cable of cross-section 1 mm² may be usdewhen installing the VS 300 S.

24 V control connection, VS 600 + VS 900

Protective low voltage for:

• Hose holder (manifold)

• Station selector switch

• Spittoon valve

Permanent connection: (N)YM (St)-J 4x1.5mm² sheathed shielded cable.

Flexible connection: PVC-Data cableLiYCY 4x1.0 mm², sheathed shielded cable asused in telecommunications and EDP or light-PVC-control sheathed shielded cable.

24 V Control line for VS 300 S

Flexible cable connection: PVC-Data cableLiYY 3 x 0.5 mm²Order-No. 9000-118-83

9.2 Control unit (VS 600 + VS 900)The suction unit can be connected to a controlunit which is either included in the contents or,if not, may be ordered as a special accessory.The connection plans and circuit diagrams areincluded in the control unit installation andoperating instructions.

Page 19: Installation and operating instructions VS 300 S / VS 600 ...€¦ · The combi-suction unit is a suction unit with an integrated separation unit. Separation within the treatment

19

10. Commissioning• Switch on unit or main surgery power switch.

• Check the setting of the motor protectionswitch (see section 4. Technical Data) andadjust if necessary.

• Check direction of motor rotation (at 3/N/PEAC).

• Check function of unit and seals of allconnections.

• Carry out an electrical safety checkaccording to local rules and regulations andrecord the results, e.g. in the technician'sreport.

• Check that the coarse filters are in position(e.g. in spittoon).

The suction unit must not beoperated without installation ofcoares filters, as large particlessuch as tooth chippings or fillingscan lead to damage.

13

12

9.3 Motor terminal box connections(VS 600 + VS 900

The power supply from the control unit areconnected to the appropriate contact points inthe motor terminal box. Connection plans andthe relevant circuit diagrams can be found inthe Installation and Operating Instructions ofthe Control Unit.

Suction Units VS 600 and VS 900

• 1/N/PE AC 230 V

• 3/N/PE AC 230 V, 3/N/PE AC 400 V

For details concerning the contact bar in themotor terminal box of the suction units VS 600and VS 900, see fig. 13.

9.4 Connection of VS 300 S to ControlUnit

1/N/PE AC 230 V, with Control Unit integratedinto sound-reducing housing, see fig. 14.X1 Power supply connectionX2 Motor connectionX3 Connection to manifold

24VAC / max. 80mAX4 Control signal output 24VAC / max. 20mA

14

Page 20: Installation and operating instructions VS 300 S / VS 600 ...€¦ · The combi-suction unit is a suction unit with an integrated separation unit. Separation within the treatment

20

Care

11. CLEANING ANDDISINFECTION OF THESUCTION UNIT

After every patient treatmentFor reasons of hygiene after every singletreatment a glass of cold water should bedrawn up through the larger and the smallersuction hose - even if only the saliva extractorhas been used.

Using the larger suction hose to drawup water causes a larger amount of air(~300 l/min) to be sucked up and thisserves to considerably improve thecleaning efficiency.

Before the mid-day break and at theend of the working daythe Suction Unit should be cleaned anddisinfected by drawing up a suitable cleaningand disinfectant agent, e.g. OROTOL Ultra orOROTOL Plus, as recommended by themanufacturer.

Do not use any foaming agent, e.g.household cleaning agent,instrument disinfection agent orabrasive agent.

Do not use any agent containingchlorine or any thinner, e.g.Acetone. These agents can causedamage to the materials. Guaranteeclaims will become null and void.

Further information is contained in theInstructions for Use “Disinfection andCleaning of Suction Units”, order number9000-605-10/.. as well as in “CleaningInstructions for contaminated Suction Units”,order number P007-235-01.

1x weekWhere the local water is very hard werecommend cleaning the unit once a weekbefore a mid-day break with DÜRR MD 555special cleaner for Suction Units.

12. MaintenanceEvery 4 weeks (or every 3 months for VS600 + 900) the filter located in the air intake ofthe suction unit should be checked and, ifnecessary, cleaned. Remove the suction hosefrom the suction unit. If required, extract thefilter from the suction connections and cleanthoroughly.

Wear non-porous safety gloves!

Every 2 years (VS 600 + 900) the auxiliary airvent should be checked and, if necessary,cleaned.

Every 2 years the exhaust bacterial filter (iffitted) should be checked and, if necessary,cleaned or replaced.

The separation unit in the suctionsystem does not keep back germs.Fitting a bacterial filter into the exhaustair vent is, therefore, stronglyrecommended.

The bacterial filter is supplied with amemo sticker, which can be stuck intothe surgery planner to remindpersonnel when to carry out the filterreplacement.

Every 3–4 years the waste water valve mustbe checked by a technician and, if necessary,replaced.

Wear non-porous safety gloves!

Page 21: Installation and operating instructions VS 300 S / VS 600 ...€¦ · The combi-suction unit is a suction unit with an integrated separation unit. Separation within the treatment

21

Disposal

13. Appliance disposalThe machine may be contaminated.Please inform the waste disposalcontractors in order that they can takethe appropriate safety measures.

Non-contaminated plastic parts of thesuction unit can be disposed of forrecycling.

The control units, electronic PCB and othercomponents should be disposed of as electricwaste. The remaining metal parts (e. g. turbinehousing) should be disposed of as metallicwaste.If returning the appliance, e.g. to your localDepot or to Dürr Dental, make sure that allconnections are closed.

Page 22: Installation and operating instructions VS 300 S / VS 600 ...€¦ · The combi-suction unit is a suction unit with an integrated separation unit. Separation within the treatment

22

Trouble-shooting

14. Tips for TechniciansThe following steps concerning trouble-shooting and correction of faults are onlydesigned for our technicians. Repairs areonly to be carried out by qualifiedtechnicians.

Possible cause• No mains supply voltage.

• Under or over voltage.

• Motor protection switch set toolow (see section 4. TechnicalData for values).

• Motor protection switch defect.

• Turbine is blocked due to solidparticles or sticky dirt: Motorprotection switch activated.

• Solid particles in the turbinechamber.

• Diaphragm valve blocked.

• Mechanical movements ofturbine hindered by dirt.

• Coarse filter blocked.

• Leaks in the suction unitplumbing.

Problem1. Suction unit

does not start.

2. Suction unitproduces unusualnoises.

3. Water drops fromair ventconnection.

4. Suction unit powertoo low.

Solution• Check mains supply, fuses in the

control unit and/or on the PCB,and replace if necessary.Check supply voltage.

• Check the supply voltage, ifnecessary inform electrician.

• Check current. Set motorprotection switch to the correctvalue.

• Check motor protection switch;replace, if necessary.

• Dismantle the suction unit andclean the turbine thoroughly.

• Dismantle the suction unit andclean the turbine thoroughly.

• Check the diaphragm valve onthe waste water section andclean or replace as necessary.

• Dismantle the suction unit andclean the turbine thoroughly.

• Coarse filter at entrance shouldbe cleaned.

• Check all pipes, hoses andconnections for leaks andreplace if necessary.


Recommended