1
Instructions for Use... 6450 SeriesV2 Mask™ ORO-NASAL SINGLE PATIENT USE Face Mask and Headgear for Custom Adapter & Valve Applications. Large, Medium, Small, Extra Small, Petite Intended Use: The Hans Rudolph 6450 series V2 Mask face piece is designed for use with the Hans Rudolph Two-Way Non-Rebreathing Valves (NRBV) and other mask adapters designed for respiratory circuits with spontaneously breathing patients where the application requires a valve to separate the patient’s inspiratory and expiratory flows or a special connector to mate directly to a respiratory flow meter or other device. The typical mouthpiece and nose clip is replaced by this face mask assembly and headgear. The masks and valves are supplied clean, non sterile. Typical applications are for pulmonary function testing, exercise testing and respiratory research. Environment of Use: This mask series is for use in hospitals and other clinical settings by individuals that have at least minimal instructions or training on the use of the mask as well as the device to which the mask are intended to connect. Product Classification: This mask and headgear are classified as disposable for single patient limited use. Sold clean non-sterile. There is no validated disinfection process. Mask and Headgear Service Life: Mask and headgear assembly may be cleaned and re-used on a single-patient basis for several test sessions as long as the fit continues to be leak free and comfortable. Operational Temperature and Humidity Ranges for the Mask: Temperature range: 5-40 C Humidity range: 0-95% RH Cautions: 1. Federal Law restricts this device to sale by or on the order of a physician. 2. Patients with facial hair, especially beards, may experience mask leakage which may require shaving the hair. 3. Do not use these masks and valves in a life support system such as a ventilator circuit. Apply and Fitting Mask 1. Mask sizes: L (Large), M (Medium), S (Small), XS (Extra Small), P (Petite) 2. Headgear sizes: L (Large), M (Medium), S (Small) 3. Determine mask size of patient with a Mask Sizing Gauge (fig 1) provided to measure the patient’s face and aid in selecting the best mask size. For sizing the patient’s facial muscles should be relaxed and jaw closed. Fit the patient’s chin in the chin cup section of the mask and slightly press the mask on the face. The top of the mask should be slightly below the nasal root depression (where the nose meets the forehead). TM MAKERS OF RESPIRATORY VALVES SINCE 1938 HANS RUDOLPH, inc. 8325 COLE PARKWAY, SHAWNEE, KANSAS 66227 U.S.A. TEL: (913) 422-7788 FAX: (913) 422-3337 U.S.A. & CANADA: (800) 456-6695 E-mail: [email protected] www.rudolphkc.com Recommended Headgear Sizes for each Mask Size Mask L M S XS P Headgear L M M S S Figure 1 6450 V2 Mask™ COMPONENTS FACE PIECE HEADGEAR STRAP ATTACHMENT SLOTS STRUCTURAL-BRACES (2) V2 HEADGEAR TRI-GLIDE QUICK-RELEASE STRAP CLIPS (4) MASK ADAPTER TWO-WAY NRBV T or Y SHAPE CUSTOM ADAPTER Fitting the Mask The mask and headgear is supplied fully assembled. 1. Place the mask over the patient’s nose and mouth. The patient’s chin should fit into the chin cup portion of the mask with the top of the mask sealing area on the bridge of the nose. 2. (a) Slide the headgear over the patient’s head. It may be easier for some patients if one of the lower headgear quick-release strap clips is disconnected before putting the headgear on the patient (b) Reconnect the headgear strap clip after the headgear is placed on. The bottom straps should be positioned below the ears and the top straps above the ears and below the eyes. 3. (a&b) Adjust the tension of the headgear Velcro™ straps by pulling back slightly on the straps. Tighten the lower two straps then the top two to achieve a comfortable fit. (c) Finally, adjust the two straps attached by the tri-glide connection at the crown of the head by pulling each in opposing directions until reaching the required tension and then reconnecting the Velcro™. Overtightening can cause leaks. 4. Connect the mask to the patient circuit by following the equipment manufacturer’s operation instructions. 5. For functional test hold palm of hand over outlet of exhalation port or mask adapter. Create a slight positive pressure inside the mask by exhaling gently. If you detect a leak around the mask sealing area, reposition the mask and or adjust the strap tension to eliminate the leak. If the leak continues regardless of your adjustments try another mask size. 1 2a 2b 3c 3a 3b A C B MASK SIZING GAUGE PETITE (P) EXTRA-SMALL (XS) SMALL (S) MEDIUM (M) LARGE (L) HANS RUDOLPH, inc. IF BETWEEN SIZES, WE RECOMMEND YOU USE THE SMALLER SIZE. www.rudolphkc.com MASK SIZING GAUGE P/N 691143 Characteristic Mask face Piece Sizes Large (L) Medium (M) Small (S) Extra Small (XS) Petite (P) A (inches) 5.9 5.6 5.2 4.8 4.6 B (inches) 3.7 3.4 3.2 3.1 3.0 C (inches) 4.2 4.2 4.0 3.9 3.9 Deadspace (ml) 143 125 99 88 78 Weight (gm) 106 96 84 74 70 Disassembly for Cleaning 1. Detach mask from the valve assembly or adapter. 2. Separate the headgear from the mask. Mask 1. Soak for 5 minutes in warm water with a mild detergent (neutral pH) 2. Hand wash with a sponge or soft brush. 3. Rinse with warm tap water. Place in a bath of warm water and agitate for two minutes. 4. Rinse in clean tap water for a least 1 minute. 5. Allow to air dry or dry with a clean, lint free cloth. Headgear 1. Submerge the headgear in warm soapy tap water and gently rub all the areas. 2. Rinse in clean tap water for 1 minute or until all signs of the soap are removed. 3. Air dry. 4. Precaution: Do not use bleach, chlorine or alcohol based solutions to clean any of the mask and headgear components. These solutions can damage this product. Direct sunlight exposure with the mask and headgear components can cause deterioration and reduce product life. Safety Information Safety or technical information regarding this product can be obtained from Hans Rudolph, inc. Phone: 913-422-7788 Fax: 913-422-3337 E-mail: [email protected] PRINTED IN U.S.A. 691303-1 2013-07-15 Rev. C US Pat No’s D573,709 & 8,136,523 and EC Req’d Design No. 000780556-0001 ISO 9001 / ISO 13485 0086 © 2013 Hans Rudolph, inc. Specifications subject to change without notice M.Devices Group Marlborough House Riding Street Southport PR8 1EW, UK P: 44 1704 544 944 F: 44 1704 544 050
Transcript
Page 1: Instructions for Use - Hans Rudolph 0713 C.pdf · The Hans Rudolph 6450 series V2 Mask face piece is designed for use with the Hans Rudolph Two-Way Non-Rebreathing Valves (NRBV) and

Instructions for Use...6450 SeriesV2 Mask™ORO-NASAL SINGLE PATIENT USE Face Mask and Headgear for Custom Adapter & Valve Applications.Large, Medium, Small, Extra Small, Petite

Intended Use:The Hans Rudolph 6450 series V2 Mask face piece is designed for use with the Hans Rudolph Two-Way Non-Rebreathing Valves (NRBV) and other mask adapters designed for respiratory circuits with spontaneously breathing patients where the application requires a valve to separate the patient’s inspiratory and expiratory flows or a special connector to mate directly to a respiratory flow meter or other device. The typical mouthpiece and nose clip is replaced by this face mask assembly and headgear. The masks and valves are supplied clean, non sterile. Typical applications are for pulmonary function testing, exercise testing and respiratory research.

Environment of Use:This mask series is for use in hospitals and other clinical settings by individuals that have at least minimal instructions or training on the use of the mask as well as the device to which the mask are intended to connect.

Product Classification:This mask and headgear are classified as disposable for single patient limited use. Sold clean non-sterile. There is no validated disinfection process.

Mask and Headgear Service Life:Mask and headgear assembly may be cleaned and re-used on a single-patient basis for several test sessions as long as the fit continues to be leak free and comfortable.

Operational Temperature and Humidity Ranges for the Mask:Temperature range: 5-40 CHumidity range: 0-95% RH

Cautions:1. Federal Law restricts this device to sale by or on the order of a physician.2. Patients with facial hair, especially beards, may experience mask leakage which may require shaving the hair.3. Do not use these masks and valves in a life support system such as a ventilator circuit.

Apply and Fitting Mask1. Mask sizes: L (Large), M (Medium), S (Small), XS (Extra Small), P (Petite)2. Headgear sizes: L (Large), M (Medium), S (Small)3. Determine mask size of patient with a Mask Sizing Gauge (fig 1) provided to measure the patient’s face and aid in selecting the best mask size. For sizing the patient’s facial muscles should be relaxed and jaw closed. Fit the patient’s chin in the chin cup section of the mask and slightly press the mask on the face. The top of the mask should be slightly below the nasal root depression (where the nose meets the forehead).

TM

MAKERS OF RESPIRATORY VALVES SINCE 1938

HANS RUDOLPH, inc.8325 COLE PARKWAY, SHAWNEE, KANSAS 66227 U.S.A.TEL: (913) 422-7788FAX: (913) 422-3337

U.S.A. & CANADA: (800) 456-6695E-mail: [email protected]

www.rudolphkc.com

Recommended Headgear Sizes for each Mask SizeMask L M S XS P

Headgear L M M S S

Figure 1

6450 V2 Mask™COMPONENTS

FACE PIECE

HEADGEAR STRAP ATTACHMENT SLOTS

STRUCTURAL-BRACES (2)

V2 HEADGEAR

TRI-GLIDE

QUICK-RELEASESTRAP CLIPS (4)MASK

ADAPTER

TWO-WAYNRBVT or Y

SHAPE

CUSTOMADAPTER

Fitting the MaskThe mask and headgear is supplied fully assembled.1. Place the mask over the patient’s nose and mouth. The patient’s chin should fit into the chin cup portion of the mask with the top of the mask sealing area on the bridge of the nose.2. (a) Slide the headgear over the patient’s head. It may be easier for some patients if one of the lower headgear quick-release strap clips is disconnected before putting the headgear on the patient (b) Reconnect the headgear strap clip after the headgear is placed on. The bottom straps should be positioned below the ears and the top straps above the ears and below the eyes.3. (a&b) Adjust the tension of the headgear Velcro™ straps by pulling back slightly on the straps. Tighten the lower two straps then the top two to achieve a comfortable fit. (c) Finally, adjust the two straps attached by the tri-glide connection at the crown of the head by pulling each in opposing directions until reaching the required tension and then reconnecting the Velcro™. Overtightening can cause leaks.4. Connect the mask to the patient circuit by following the equipment manufacturer’s operation instructions.5. For functional test hold palm of hand over outlet of exhalation port or mask adapter. Create a slight positive pressure inside the mask by exhaling gently. If you detect a leak around the mask sealing area, reposition the mask and or adjust the strap tension to eliminate the leak. If the leak continues regardless of your adjustments try another mask size.

1 2a 2b 3c3a 3b

▲ ▲▲ ▲

A

C BP E T I T E ( P )

E X T R A - S M A L L ( X S )

S M A L L ( S )

M E D I U M ( M )

L A R G E ( L )

M A S K S I Z I N G

G A U G E

P E T I T E ( P )

E X T R A - S M A L L ( X S )

S M A L L ( S )

M E D I U M ( M )

L A R G E ( L )

TM

H A N S R U D O L P H , i n c .

Directions: Length measurements are used to determine the size of the Hans Rudolph Oro-Nasal mask that will best fit the patient. For measurement: the patient’s facial muscles should be relaxed and jaw closed. Place the Gauge tip to the deepest point of the nasal root depression (sellion) and note the size indication where the bottom of the chin exits the gauge. The gray band widths indicate the size ranges.

This gauge is a reference to aid in sizing and not an exact measurement as other factors may need to be taken into account.

IF BETWEEN SIZES, WE

RECOMMEND YOU USE

THE SMALLER SIZE.

w w w . r u d o l p h k c . c o m

MASK SIZING GAUGEP/N 691143

CharacteristicMask face Piece Sizes

Large (L) Medium (M) Small (S) Extra Small (XS) Petite (P)A (inches) 5.9 5.6 5.2 4.8 4.6

B (inches) 3.7 3.4 3.2 3.1 3.0

C (inches) 4.2 4.2 4.0 3.9 3.9

Deadspace (ml) 143 125 99 88 78

Weight (gm) 106 96 84 74 70

Disassembly for Cleaning1. Detach mask from the valve assembly or adapter.2. Separate the headgear from the mask.

Mask1. Soak for 5 minutes in warm water with a mild detergent (neutral pH)2. Hand wash with a sponge or soft brush.3. Rinse with warm tap water. Place in a bath of warm water and agitate for two minutes.4. Rinse in clean tap water for a least 1 minute.5. Allow to air dry or dry with a clean, lint free cloth.

Headgear1. Submerge the headgear in warm soapy tap water and gently rub all the areas.2. Rinse in clean tap water for 1 minute or until all signs of the soap are removed.3. Air dry.4. Precaution: Do not use bleach, chlorine or alcohol based solutions to clean any of the mask and headgear components. These solutions can damage this product. Direct sunlight exposure with the mask and headgear components can cause deterioration and reduce product life.

Safety InformationSafety or technical information regarding this product can be obtained from Hans Rudolph, inc.Phone: 913-422-7788 Fax: 913-422-3337 E-mail: [email protected]

PRINTED IN U.S.A.691303-1 2013-07-15 Rev. C

US Pat No’s D573,709 & 8,136,523 and EC Req’d Design No. 000780556-0001

ISO 9001 / ISO 134850086 © 2013 Hans Rudolph, inc.

Specifications subject to change without notice