Part No. J32862
Instrument Manager™
Supplemental Specifications for
Laboratory Information System
Export authorized under general license GTDA (General Technical Data Available)
IMPORTANT
The information contained herein is based on the experience and knowledge relating to the subject matter gained by Ortho-Clinical Diagnostics, Inc. prior to publication.
No patent license is granted by the information.
Ortho-Clinical Diagnostics, Inc. reserves the right to change this information without notice, and makes no warranty, express or implied, with respect to the information. The company shall not be liable for any loss or damage, including consequential or special damages resulting from the use of this information, even if loss or damage is caused by its negligence or other fault.
VITROS is a registered trademark of Ortho-Clinical Diagnostics, Inc.enGen is a registered trademark of Ortho-Clinical Diagnostics, Inc.
Ortho Clinical Diagnostics, 2008 - 2017. All rights reserved.
Revision History
Revision Date Section Description
2017-07-10 3.2.1 Added ALT2 and ALTV Test Codes.Added Chloride - "Urine" Fluid Code.Removed “2” from Homocysteine Test Name.Removed HCY LIS Test Code.
3.2.2 Added INS, C-pep, HIV c, NCHEK, TIMP2 and IGFBP Test Codes.
2013-01-10 1.2 Updated references to Data Innovations documents.
3.2.1 Updated Fluid Codes for:• %MB• CKMB• CKMBECI• DGXN• PCP
3.2.2 Updated Fluid Code for TVITD
Test Order Records - updated reference to Data Innovations interface specifications
3.9 Updated default connection names:• Added FS7• Removed Immuno-Assay System• Removed VITROS® 950 Chemistry System• Added 365, 366, 367, 465, 566 and 567
3.12.1 Updated the LIS message
3.14 Updated the Standard Test Codes for Sorting Table
Part No. J32862 Instrument Manager™ Supplemental Specifications for Laboratory Information System iii2017-07-10
2012-06-05 Updated the Title of the document.
Global reformatting of this document.
1.1 Added new Definitions Section
1.2 Updated Introduction Section
2.1 Changed name to Message Options and updated the table.
3.2.1 Updated Test Code Definitions for D%A1CI and DTIBC and added HCY2*.
3.2.2 Added FPSA, %FPSA, TPSA, TVITD, SYPH, and IPTH Test CodesRemoved “International” from the following Test Names:• Anti-HBe• Anti-HIV 1 + 2• CMV IgG• CMV IgM• Free B-hCG• HBeAg• Rubella IgM• Syphilis• Toxoplasma IgG• Toxoplasma IgM• Thyroid Stimulating Hormone 30
3.3 Changed Title of Section
3.13.1 Added fluid Type explanation for AT Test Codes
3.13.2 Added new section for AT Test Code transmission
3.14 Updated Standard Test Codes for Sorting Table
4 Removed the Prefixes and Test with Multiple Result Components(non-VITROS®) Sections. Added 2 different options for connection of Non-OCD Instruments.
Eliminated “Version”. The Revision Date is now the version of the document.
*Product in development. For all assays, availability is subject to regulatory requirements.
Revision Date Section Description
iv Instrument Manager™ Supplemental Specifications for Laboratory Information System Part No. J328622017-07-10
Revision Date Version Description
2011-07-22 2.1 Update to include integration of VITROS® 4600 Chemistry System.
2008-10-31 2.0 Update to include integration of VITROS® 3600 Immunodiagnostic System and VITROS® 5600 Integrated System.
2008-07-18 1.1 Update to Test Code Definitions.
2008-04-30 1.0 First release of document.
List of Revised Pages
Each page in your manual should be at the effective date listed below
Revision Date Page
2017-07-10 All
2013-01-10 All
2012-06-05 All
:
Revision Date Version Page
2011-07-22 2.1 1-1, 3-25
2008-10-31 2.0 All
2008-07-18 1.1 3-4
2008-04-30 1.0 All
Part No. J32862 Instrument Manager™ Supplemental Specifications for Laboratory Information System v2017-07-10
vi Instrument Manager™ Supplemental Specifications for Laboratory Information System Part No. J328622017-07-10
Table of Contents
CHAPTER 1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-11.1 Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-11.2 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-11.3 Scope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
CHAPTER 2 Instrument Manager™. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12.1 Message Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-12.2 Communication Ports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1
CHAPTER 3 VITROS® Systems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13.1 Fluid Code Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-13.2 VITROS® Test Code Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
3.2.1 VITROS® Chemistry Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-23.2.2 VITROS® Immunoassay Tests. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-9
3.3 VITROS® Result Code and Flag Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-133.4 VITROS® Result Values. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-193.5 VITROS® Result Units. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-19
3.5.1 Result Transmission Examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-203.5.2 Impact of Intermediate Result Handling Selection in Result Transmission . . . . . . .3-223.5.3 Impact of Result Prefix Selection in Result Transmission . . . . . . . . . . . . . . . . . . . .3-223.5.4 Impact of Comment Selection in Result Transmission. . . . . . . . . . . . . . . . . . . . . . .3-23
3.6 Result Instrument ID . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-243.7 Result Date/Time. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-243.8 Comments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-243.9 Instrument IDs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-253.10 Operator IDs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-263.11 Laboratory Status Message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-26
3.11.1 Using ASTM. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-263.11.2 Using HL7 2.2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-273.11.3 Using HL7 2.4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-27
3.12 Laboratory Storage Message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-283.12.1 Using ASTM. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-283.12.2 Using HL7 2.2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-293.12.3 Using HL7 2.4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-29
3.13 Aliquoting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-303.13.1 Aliquot Requests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-303.13.2 Example of Transmission using HL7 2.2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-30
3.14 Standard Test Codes for Sorting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-31
Part No. J32862 Instrument Manager™ Supplemental Specifications for Laboratory Information System vii2017-07-10
CHAPTER 4 Non-OCD Instrument. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14.2 Non-OCD Instrument Interface Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
4.2.1 Option 1: Non-OCD Instrument is NOT interfaced to OCD-IM but directly to the LIS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
4.2.2 Option 2: Non-OCD Instrument is interfaced to OCD-IM . . . . . . . . . . . . . . . . . . . 4-1
viii Instrument Manager™ Supplemental Specifications for Laboratory Information System Part No. J328622017-07-10
1 Introduction
1.1 DefinitionsLIS Laboratory Information System
OCD Ortho Clinical Diagnostics
DI Data Innovations
IM Data Innovations Instrument Manager™
OCD-IM Instrument Manager™ Subsystem combined with an OCD standardized configuration file
1.2 IntroductionThe OCD-IM Controller is the interface platform between the instruments and the LIS. It uses an OCD-owned standardized configuration file (test codes mapping, fluid codes mapping, error flags, etc.) as described in this document.
The OCD-IM Controller can be connected to:
• enGen™ Laboratory Automation System track system (used for pre-analytical and post-analytical sample and data management)
• VITROS® Instruments (On-Track and Off-Track)• Non-OCD Instruments (On-Track only, see Chapter 4)
The LIS will not communicate directly with the on-track or off-track VITROS® Systems, but rather communicate with IM software applications developed and owned by DI.
Data Innovations has ASTM and an HL7 protocol documents that describe the specific communication parameters for communicating between Instrument Manager™ and the customer Laboratory Information System (LIS):
• Data Innovations Software Requirements Specifications (SRS) for Host Driver Ortho Clinical Diagnostics ASTM LIS ocdastml SRS-0315
• Data Innovations Software Requirements Specifications (SRS) for Host Driver Ortho Clinical Diagnostics HL7 LIS ocdhl7ml SRS-0314
Part No. J32862 Instrument Manager™ Supplemental Specifications for Laboratory Information System 1–12017-07-10
1.3 ScopeThis document is a supplemental interface specification. It is intended to provide additional information and guidance to the customer and the customer LIS vendor during development of the communication interface to the Instrument Manager™ for Ortho Clinical Diagnostics, Inc. (OCD-IM). The Interface should be tested prior to routine use.
The use of any assay and fluid combination is subject to limitations and recommendations defined in OCD Instructions for Use documents for each assay.
WARNING: Specification processed represent the default and standard configuration for the OCD-IM. This specification is only applicable to the OCD-IM.
WARNING: The Instrument Manager™ Supplemental Specifications for Laboratory Information System (LIS) document will be voided if the user changes the default mapping as described and/or write rules that impact Instrument Manager™ communication with the LIS.
1–2 Instrument Manager™ Supplemental Specifications for Laboratory Information System Part No. J328622017-07-10
2 Instrument Manager™
2.1 Message OptionsThe configuration of an OCD-IM varies from site to site depending on the LIS capabilities and how results will be processed. The following options are used to augment the results uploaded to the LIS.
Feature Option DescriptionIntermediate Result Handling Send All All results will be sent to the LIS
Send Final Intermediate results will not be sent to the LISResult Prefix Insert < > prefixes VITROS® System results greater or less than the
Reportable Range of the Instrument will be sent to the LIS with the prefix < or > in .
No prefix VITROS® System results greater or less than the Reportable Range of the Analyzer will be sent with the Analyzer flags only. No result prefix will be sent.
Result CommentsDetermines what, if any, information is sent with results that have flags or codes.
Comments Comments will be sent to the LIS to describe result codes and flags, including qualitative and semi-qualitative assay flags.
No comments No comments on error codes shall be sent in the Comments field.
Qualitative Assay Comments
Comments will be sent to the LIS containing text for range units associated with Q, R, S, T, and U, flags associated with qualitative and semi- qualitative assays.
2.2 Communication PortsConnection assignment for the LIS should be selected from the following table utilizing the corresponding port number that meets the site’s current LIS configuration. Individual IP Addresses for the Instrument Manager™ computer will be provided during installation.
Example: The LIS that requires two separate communication lines to send requests and receive results will need to connect on Ports 10011 and 10013. Instrument Manager™ will be in server mode by default.
Connection Mode IP Address Port No.LIS - combined orders/results Server n/a 10010LIS - receiving orders Server n/a 10011LIS - receiving orders (2nd) Server n/a 10012LIS - sending results Server n/a 10013LIS - auto-logging Server n/a 10014
Note: Upon request, serial communication with the LIS can be established through the use of device servers. Contact Ortho Clinical Diagnostics for more information.
Note: Upon request, Instrument Manager can also operate as a client. In this case, the IP Address of the LIS server must be provided.
Part No. J32862 Instrument Manager™ Supplemental Specifications for Laboratory Information System 2–12017-07-10
2–2 Instrument Manager™ Supplemental Specifications for Laboratory Information System Part No. J328622017-07-10
3 VITROS® Systems
3.1 Fluid Code DefinitionsThis section includes the definition of fluid code character strings required to order tests on Instruments through OCD-IM.
The following standard fluid codes are defined in Instrument Manager™:
These codes are mandatory and must be processed as per Data Innovations LIS interface specifications defined in Section 1.1
• ASTM: O (Test Order Record), field 16.1• HL7 V2.2 and V2.4: OBR (Observation Request Segment), field 15.1
Note:
• If the LIS is unable to differentiate between Serum and Plasma fluid types, Serum should be used.• If Serum fluid type is used to identify both Serum and Plasma and a test is required for Plasma only, the system
will assume the specimen fluid type is plasma and will return the result as Plasma.• If a sample includes a request for a Serum only and Plasma only test, the system will decline the entire test
request.
LIS Fluid Code
LIS Fluid Code Definition
Alternate LIS Fluid Code
(request message only)SER Serum 1PLAS Plasma 4CSF Cerebrospinal Fluid 3URI Urine 2BLD Whole Blood 5AMNIO Amniotic Fluid 6
Part No. J32862 Instrument Manager™ Supplemental Specifications for Laboratory Information System 3–12017-07-10
3.2 VITROS® Test Code DefinitionsThis section includes the definition of test code character strings needed to order tests on VITROS® Instruments through Instrument Manager™ regardless of which VITROS® Instrument is used to perform the test.
The following standard test codes are defined in the OCD-IM.
3.2.1 VITROS® Chemistry Tests
Fluid Codes
Test NameLIS Test
Code Serum PlasmaAmniotic
Fluid CSF Urine Blood
%MB %MB SER
% Iron Saturation %SAT SER PLAS
Alpha-1-antitrypsin AAT SER
Acetaminophen ACET SER PLAS
Acid Phosphatase ACP SER PLAS
A/G Ratio AG SER PLAS
Anion Gap AGP SER PLAS
Anion Gap (K+) AGPK SER PLAS
Albumin ALBALB1ALB2ALB3
SER PLAS
Alcohol ALC SER PLAS
Alkaline Phosphatase ALKPALKP1ALKP2ALKP3
SER PLAS
Alanine Aminotransferase ALTALT1ALT2ALT3ALTV
SER PLAS
Ammonia AMON PLAS
Amphetamines AMPH URI
Amylase AMYL SER PLAS URI
Amylase AMYL1AMYL2AMYL3
SER PLAS
ApoA1 APOA1 SER PLAS
ApoB APOB SER PLAS
Anti-streptolysin O ASO SER PLAS
3–2 Instrument Manager™ Supplemental Specifications for Laboratory Information System Part No. J328622017-07-10
Aspartate Aminotransferase
ASTAST1AST2AST3
SER PLAS
Barbiturates BARB URI
Conjugated Bilirubin BCBC1BC2BC3
SER PLAS
Benzodiazepines BENZ URI
Unconjugated Bilirubin BUBU1BU2BU3
SER PLAS
C3 C3 SER PLAS
C4 C4 SER PLAS
Calcium CA SER PLAS URI
Calcium CA1CA2CA3
SER PLAS
Caffeine CAFFN SER PLAS
CHOL/dHDL CH SER PLAS
Cholinesterase CHE SER PLAS
Cholesterol CHOLCHOL1CHOL2CHOL3
SER PLAS
Creatinine Kinase CKCK1CK2CK3
SER PLAS
Creatinine Kinase MB (U/L)
CKMB SER
Chloride CLCL1CL2CL3
SER PLAS URI
Cocaine Metabolite COCM URI
Carbamazepine CRBM SER PLAS
Fluid Codes
Test NameLIS Test
Code Serum PlasmaAmniotic
Fluid CSF Urine Blood
Part No. J32862 Instrument Manager™ Supplemental Specifications for Laboratory Information System 3–32017-07-10
Creatinine CREACREA1CREA2CREA3CREA4CREA5CREA6CREA7CREA8CREA9
CREA10CREA11CREA12CREA13CREA14CREA15
SER PLAS URI
C Reactive Protein CRPCRP1CRP2CRP3
SER PLAS
d%A1c D%A1C BLD
mmA1c (International) D%A1CI BLD
Direct Bilirubin DBILDBIL1DBIL2DBIL3
SER PLAS
Delta Bilirubin DELB SER PLAS
Digoxin DGXN SER
Direct Hemoglobin A1c** DHA1C BLD
Direct Hemoglobin** DHB BLD
Direct HDL (slide methodology)
DHDL SER PLAS
Direct LDL (MicroTip methodology)
DLDL SER PLAS
Direct TIBC DTIBC SER
Fluid Codes
Test NameLIS Test
Code Serum PlasmaAmniotic
Fluid CSF Urine Blood
3–4 Instrument Manager™ Supplemental Specifications for Laboratory Information System Part No. J328622017-07-10
Enzymatic CO2 ECO2ECO2-1ECO2-2ECO2-3
SER PLAS
Iron FEFE1FE2FE3
SER PLAS
Gentamicin GENTGENTPGENTTGENTR
SER PLAS
Gamma Glutamyltransferase
GGTGGT1GGT2GGT3
SER PLAS
Globulin GLOB SER PLAS
Glucose GLU SER PLAS CSF URI
Glucose GLU1GLU2GLU3GLU4GLU5GLU6GLU7GLU8GLU9
GLU10GLU11GLU12GLU13GLU14GLU15GLU16GLU17GLU18GLU19GLU20GLU21GLU22GLU23GLU24GLU25
SER PLAS
Homocysteine HCY2 SER PLAS
Hemolysis (MicroSensor) HEM SER PLAS CSF
Haptoglobin HPT SER PLAS
High sensitivity C Reactive Protein
HSCRP SER PLAS
Icterus (MicroSensor) ICT SER PLAS CSF
Immunoglobulin A IGA SER PLAS
Fluid Codes
Test NameLIS Test
Code Serum PlasmaAmniotic
Fluid CSF Urine Blood
Part No. J32862 Instrument Manager™ Supplemental Specifications for Laboratory Information System 3–52017-07-10
Immunoglobulin G IGG SER PLAS
Immunoglobulin M IGM SER PLAS
Potassium K SER PLAS URI
Potassium K1K2K3
SER PLAS
Lactate LACLAC1LAC2LAC3
PLAS
Lactate Dehydrogenase LDHLDH1LDH2LDH3
SER PLAS
Low Density Lipoprotein LDL SER PLAS
Lithium LI SER PLAS
Lipase LIPALIPA1LIPA2LIPA3
SER PLAS
Microalbumin MALB URI
Methadone METD URI
Magnesium MG SER PLAS URI
Magnesium MG1MG2MG3
SER PLAS
Sodium NA SER PLAS URI
Sodium NA1NA2NA3
SER PLAS
Neonatal Bilirubin NBIL SER PLAS
Opiates-Low Cutoff OP-LO URI
Opiates-High Cutoff OP-HI URI
Osmolality OSMO SER PLAS
Prealbumin PALB SER PLAS
Phencyclidine PCP URI
Phenobarbital PHBR SER PLAS
Phosphorus PHOS SER PLAS URI
Phosphorus PHOS1PHOS2PHOS3
SER PLAS
Fluid Codes
Test NameLIS Test
Code Serum PlasmaAmniotic
Fluid CSF Urine Blood
3–6 Instrument Manager™ Supplemental Specifications for Laboratory Information System Part No. J328622017-07-10
Phenytoin PHYT SER PLAS
CSF Protein PROT CSF
Rheumatoid Factor RF SER PLAS
Salicylate SALI SER PLAS
Total Bilirubin TBILTBIL1TBIL2TBIL3
SER PLAS
Cannabinoids THC URI
Theophylline THEO SER PLAS
Total Iron Binding Capacity
TIBC SER
Tobramycin TOBRATOBRAPTOBRATTOBRAR
SER PLAS
Total Protein TPTP1TP2TP3
SER PLAS
Transferrin TRFRNTRFRN1TRFRN2TRFRN3
SER PLAS
Triglycerides TRIGTRIG1TRIG2TRIG3
SER PLAS
Turbidity (MicroSensor) TUR SER PLAS CSF
UR/Creatinine Ratio UCR SER PLAS
UDA01 UDA01 SER PLAS AMNIO CSF URI BLD
UDA02 UDA02 SER PLAS AMNIO CSF URI BLD
UDA03 UDA03 SER PLAS AMNIO CSF URI BLD
UDA04 UDA04 SER PLAS AMNIO CSF URI BLD
UDA05 UDA05 SER PLAS AMNIO CSF URI BLD
UDA06 UDA06 SER PLAS AMNIO CSF URI BLD
UDA07 UDA07 SER PLAS AMNIO CSF URI BLD
UDA08 UDA08 SER PLAS AMNIO CSF URI BLD
UDA09 UDA09 SER PLAS AMNIO CSF URI BLD
UDA10 UDA10 SER PLAS AMNIO CSF URI BLD
UDA11 UDA11 SER PLAS AMNIO CSF URI BLD
Fluid Codes
Test NameLIS Test
Code Serum PlasmaAmniotic
Fluid CSF Urine Blood
Part No. J32862 Instrument Manager™ Supplemental Specifications for Laboratory Information System 3–72017-07-10
** See the Instructions for Use for each assay
For all assays, availability is subject to regulatory requirements.
3.2.2 VITROS® Immunoassay Tests
UDA12 UDA12 SER PLAS AMNIO CSF URI BLD
UDA13 UDA13 SER PLAS AMNIO CSF URI BLD
UDA14 UDA14 SER PLAS AMNIO CSF URI BLD
UDA15 UDA15 SER PLAS AMNIO CSF URI BLD
UDA16 UDA16 SER PLAS AMNIO CSF URI BLD
UDA17 UDA17 SER PLAS AMNIO CSF URI BLD
UDA018 UDA18 SER PLAS AMNIO CSF URI BLD
UDA19 UDA19 SER PLAS AMNIO CSF URI BLD
UDA20 UDA20 SER PLAS AMNIO CSF URI BLD
Urine Protein UPRO URI
Urea Nitrogen UREA SER PLAS URI
Urea Nitrogen UREA1UREA2UREA3
SER PLAS
Uric Acid URIC SER PLAS URI
Uric Acid URIC1URIC2URIC3
SER PLAS
Valproic Acid VALP SER PLAS
Vancomycin VANCVANCRVANCPVANCT
SER PLAS
Very Low Density Lipoprotein
VLDL SER PLAS
Fluid Codes
Test Name LIS Code Serum PlasmaAmniotic
Fluid Urine Blood
Alpha-fetoprotein AFPAFP1
SER PLAS AMNIO
Anti-HBc AHBC SER PLAS
Anti-HBe AHBE SER
Fluid Codes
Test NameLIS Test
Code Serum PlasmaAmniotic
Fluid CSF Urine Blood
3–8 Instrument Manager™ Supplemental Specifications for Laboratory Information System Part No. J328622017-07-10
Anti-HBs (U.S.) AHBSQ SER
Anti-HBs (International) AHBS SER PLAS
Anti-HCV AHCV SER PLAS
Anti-HIV 1+2 AHIV SER PLAS
Vitamin B12 B12 SER PLAS
Total B-hCG BHCGBHCG1BHCG2BHCG3BHCG4BHCG5BHCG6BHCG7
SER PLAS
CA 125 II CA125CA125-1
SER PLAS
CA 15-3 CA153CA153-1
SER PLAS
CA 19-9 CA199CA199-1
SER PLAS
Carcinoembryonic Antigen CEACEA1
SER PLAS
Creatinine Kinase MB(ng/mL (ug/L))
CKMBECI SER PLAS
CMV IgG CMVG SER PLAS
CMV IgM CMVM SER PLAS
Cortisol CORT SER PLAS URI
Cortisol CORT1CORT2CORT3CORT4CORT5CORT6CORT7CORT8CORT9
CORT10CORT11CORT12
SER PLAS
C-peptide* C-pep SER PLAS
Estradiol E2E2-1E2-2E2-3E2-4E2-5
SER PLAS
Free B-hCG FBHCG SER
Fluid Codes
Test Name LIS Code Serum PlasmaAmniotic
Fluid Urine Blood
Part No. J32862 Instrument Manager™ Supplemental Specifications for Laboratory Information System 3–92017-07-10
Ferritin FERRFERR1FERR2FERR3FERR4FERR5
SER PLAS
Folate FOLAT SER PLAS BLD
Free PSA FPSA SER PLAS
% free PSA %FPSA SER PLAS
Follicle Stimulating Hormone FSHFSH1FSH2FSH3FSH4FSH5FSH6FSH7FSH8FSH9
FSH10FSH11FSH12
SER PLAS
Free T3 FT3 SER PLAS
Free T3 Index FT31 SER PLAS
Free T3 II FT3II SER PLAS
Free T4 FT4 SER
Free T4 Index FT4I SER PLAS
Anti-HAV IgM HAVM SER PLAS
Anti-HAV Total HAVT SER PLAS
Anti-HBc IgM HBCM SER PLAS
HBsAg Confirmatory HBCON SER PLAS
HBeAg HBE SER
HBsAg HBSAG SER PLAS
HIV Combo HIV c SER PLAS
IGFBP-7 IGFBP SER PLAS
Insulin INS SER PLAS
Intact PTH IPTH SER PLAS
LH/FSH Ratio LF SER PLAS
Fluid Codes
Test Name LIS Code Serum PlasmaAmniotic
Fluid Urine Blood
3–10 Instrument Manager™ Supplemental Specifications for Laboratory Information System Part No. J328622017-07-10
Luteinizing Hormone LHLH1LH2LH3LH4LH5LH6LH7LH8LH9
LH10LH11LH12
SER PLAS
Myoglobin MYOGMYOG1MYOG2MYOG3MYOG4MYOG5
SER PLAS
AKIRisk NCHEK SER PLAS
NT-proBNP NTPBNP SER PLAS
N-Telopeptide NTX URI
Progesterone PROGPROG1PROG2PROG3PROG4PROG5
SER PLAS
Prolactin PROLPROL1PROL2PROL3PROL4PROL5PROL6PROL7PROL8PROL9
PROL10PROL11PROL12
SER PLAS
Prostate Specific Antigen PSA SER PLAS
Rubella IgG RUBG SER PLAS
Rubella IgM RUBM SER PLAS
Syphilis SYPH SER PLAS
TT3/TT4 Ratio T3T4 SER PLAS
T3 Uptake T3U SER PLAS
Fluid Codes
Test Name LIS Code Serum PlasmaAmniotic
Fluid Urine Blood
Part No. J32862 Instrument Manager™ Supplemental Specifications for Laboratory Information System 3–112017-07-10
*Product in development.For all assays, availability is subject to regulatory requirements.
Testosterone TESTOTESTO1TESTO2TESTO3TESTO4TESTO5
SER PLAS
TIMP-2 TIMP2 SER PLAS
Toxoplasma IgG TOXG SER PLAS
Toxoplasma IgM TOXM SER PLAS
Total PSA II TPSA SER PLAS
Troponin I TROP
TROP1
SER PLAS
Thyroid Stimulating Hormone TSHTSH1TSH2TSH3TSH4TSH5TSH6TSH7TSH8TSH9TSH10TSH11TSH12
SER PLAS
Thyroid Stimulating Hormone 30 TSH30 SER PLAS
Total T3 TT3 SER PLAS
Total T4 TT4 SER PLAS
25-OH Vitamin D Total TVITD SER
Fluid Codes
Test Name LIS Code Serum PlasmaAmniotic
Fluid Urine Blood
3–12 Instrument Manager™ Supplemental Specifications for Laboratory Information System Part No. J328622017-07-10
Multiple LIS Codes
The above tables define additional test names for an assay, for example: GLU, GLU1, GLU2. These tests may be used to order the same test using different test names, for example, multiple samples for glucose tolerance test, multiple samples for peak and trough specimens.
Test Order RecordsThese test codes are included in the test order record and are mandatory and must be processed as per Data Innovations ASTM and HL7 interface specifications to order tests.
For ASTM:• Field 0.12 (Test Order Record/Specimen Action Code) should contain “A”
For HL7 2.2 and 2.4:
• Field ORC.1 (Common Order Segment/Order Control) should contain “XE”• Field OBR.11 (Observation Request Segment/Specimen Action) should contain “A”
Note:Any other value processed in this field is not supported by the enGen™ Laboratory Automation System.
Important: If additional tests are requested for the same sample (add-on tests), only the additional tests should be sent in the message and the appropriate action code (append) must be provided by the LIS as per Data Innovations ASTM or HL7 interface specifications. Tests should not be resent as appended with a new order.
Example 1: ASTM Order Transmission
H|\^&||||||||||P|1|<cr>P|1|1234567890||123-45-6789|LAST^FIRST^M||19011215|F|||||789^DrLast^DrFirst^M|||||||||PF2||F|Fac^Ward^Room^Bed^NS1<cr>O|1|123123123||^^^GLU^\^^^UREA^|R||200612140804||||A||||SER|||||||||<cr>L|1|N<cr>
Example 2: HL7 Order Transmission
MSH|^~\&||||||||||||||<cr>
PID|1||||LAST^FIRST^M||19011215|F||||||||||998877<cr>
ORC|XE|123123123|||||||||1000<cr>
OBR|1|123123123||GLU|||20061114102100|20061114103100|||A||||SER||||^^^^^^P||<cr>
OBR|2|123123123||UREA|||20061114102100|20061114103100||A|||||SER||||^^^^^^P||<cr>
<fs>
Result Records
The same test codes are included in the result record and will also be processed as per Data Innovations ASTM or HL7 interface specifications with every result sent by any VITROS® System connected to Instrument Manager™.
• ASTM: R (Result Record), field 3.4• HL7 V2.2 and V2.4: OBX (Observation Request Segment), field 3
3.3 VITROS® Result Code and Flag DefinitionsThese result codes are included in the result record and will be processed as per Data Innovations ASTM or HL7 interface specifications with every result sent (if applicable) by any VITROS® System connected to Instrument Manager™.• ASTM: R (Result Record), field 7.1• HL7 V2.2: OBX (Observation Request Segment), field 7• HL7 V2.4: OBX (Observation Request Segment), field 8
Part No. J32862 Instrument Manager™ Supplemental Specifications for Laboratory Information System 3–132017-07-10
The following result codes may be associated with a test result when the test results are sent to the LIS from Instrument Manager™.
Result Flags/Code Description
AF Air Filter
AR Adjusted Result
AT Additional Test Result
BP Blank Prediction
CB Cuvette Blank
CC Configuration Conflict
CE Calibration Expired
DD Drop Detector Disabled
DE Drop Error
DL Sample Diluted
DP Substrate Depleted
EA Expired Aliquot
EC Expired Calibrator
ED Edited Result
EM Expired Maintenance
EP Edited Patient Data
ER Math Error
FC Flagged Component
FR Flagged Replicate
HB High Blank
HC High Concentration
HN High Noise
IC Invalid Component
ID Invalid Dilution
II Insufficient Inventory
IR Invalid Reading/Result
IS Insufficient Sample
IT Incubator Temperature
KE Kinetic Error
LS Lot Switch
LT Luminometer Temperature
M1 Category 1 modified values
M2 Category 2 modified values
3–14 Instrument Manager™ Supplemental Specifications for Laboratory Information System Part No. J328622017-07-10
ME Mechanical Error
MN Mean
MW Multiple Windows
NC Not Calibrated
ND No Drop
NF No Fluid
NI No Inventory
NQ No Quality Control
NR No Result
NS No Slide
NT No Tip
NW No Well
OC Dilution (Operator request Concentration)
OD Operator Requested Dilution
OR Outside Reportable Range
OS Outside Spline
PD Pressure Disabled
PF Prediction Failure
PI Potential Interferent
RC Reference Consistency
RD Reflex Diluted
RE Reagent Expired
RP Reflex Processed
RR Recalculated Result
RS Reduced Standard Dilution
SC Spread Check (currently not sent)
SD Standard Dilution
SE Sample Exited/Expired
SP Multiple Spikes
SR Outside Supplementary Range
ST Slide Time Out
TD Test Deleted
TR Trim Error
UC User Calibration
UD Unconfigured Diluent
VS Viscous Sample
Result Flags/Code Description
Part No. J32862 Instrument Manager™ Supplemental Specifications for Laboratory Information System 3–152017-07-10
WD Wetness Disabled
WE Wash Error
WT Well Temperature
ZS Negative Derived Test Result
efd0 No Error
efd1 Above derived test hospital range
efd2 Below derived test hospital range
efd3 Edited derived test result
efd4 Bad derived test component
efd5 No derived test result
efd7 Outside derived test supplement range
efd8 Pre-treated Multiple Sample Derived Test (MSDT)
efdA Above 2 SD from mean
efdB Above 3 SD from mean
efdC Not supported in QC database
efdD Below QC range
efdE Above QC range
efm0 No Error
efm1 Above laboratory’s range
efm2 Below laboratory’s range
efm3 Outside dynamic range
efm4 Above analyzer’s range (the value reported is the maximum limit of the range)
efm5 Below analyzer’s range (the value reported is the minimum limit of the range)
efm6 Prediction failure (reported as No Result)
efm7 Outside Supplementary Range
efmA Control result is more than 2 SDI from baseline interval mean but no more than 3 SDI
efmB Result is more than 3 SDI from baseline interval mean
efmC No baseline interval mean and SDI; or assay is not supported in QC database
efmD Control result below QC range #
efmE Control result above QC range #
efmF Greater than Dynamic Range
efmG Less than Dynamic Range
efmH Greater than Supplementary Range
efmI Less than Supplementary Range
wf0 No warning
wf1 Analyzer-generated warning
Result Flags/Code Description
3–16 Instrument Manager™ Supplemental Specifications for Laboratory Information System Part No. J328622017-07-10
*When one of these flags is associated with a result, a comment that contains the interpreted text may be sent to the LIS with that result, depending on the option selected.
wf2 Operator-induced warning
wf3 Both analyzer and operator warning
ef0 No Error
ef1 Above Reference Range
ef2 Below Reference Range
ef4 Above Reportable Range
ef5 Below dynamic range (the value reported is the minimum limit of the range)
ef6 Prediction failure (with value reported as NO RESULT)
ef7 Above Supplementary Range
ef8 Below Supplementary Range
efA The value of the QC Result is at least 2, but less than 3 Baseline Standard Deviations below the Base Mean(-2SD)
efB The value of the QC Result is at least 2, but less than 3 Baseline Standard Deviations above the Baseline Mean (+2SD)
efC The value of the QC Result is at least 3 Baseline Standard Deviations below the Baseline Mean (-3SD)
efD The value of the QC Result is at least 3 Baseline Standard Deviations above the Baseline Mean (+3SD)
efE The values of two consecutive QC Results are both greater than two Standard Deviations above or below the baseline mean (22s)
efF The values of two consecutive QC Results have a change of at least 4 Standard Deviations (R4s)
efG The values of four consecutive QC Results are all greater than one standard deviation above or below the Baseline Mean (41s)
efH The values of ten consecutive QC results are all above or all below the Baseline Mean (10x)
efJ The Result of this Sample Integrity Test exceeds the threshold for this assay
efQ Range limit 1*
efR Range limit 2*
efS Range limit 3*
efT Range limit 4*
efU Range limit 5*
Result Flags/Code Description
Part No. J32862 Instrument Manager™ Supplemental Specifications for Laboratory Information System 3–172017-07-10
The following hemolysis, icterus, and turbidity flags may be associated with the test results, and sent with the result to the LIS from Instrument Manager™.
Note: For all assays, availability is subject to regulatory requirements.
Result Flags DescriptionHecES Hemolysis - Examine sampleHecME Hemolysis - Mechanical ErrorHecNA Hemolysis - SI Globally not availableHecNR Hemolysis - Not RunHecpi Hemolysis - Potential InterferentHef4 Hemolysis - Above Reportable RangeHef5 Hemolysis - Below Reportable RangeHef6 Hemolysis - Prediction Failure (with value reported as NO RESULT)HefJ Hemolysis - The result of this Sample Integrity Test exceeds the threshold for this assayHefU Hemolysis - H-Index flag on initial undiluted sampleIecES Icterus - Examine sampleIecME Icterus - Mechanical ErrorIecNA Icterus - SI Globally not availableIecNR Icterus - Not RunIecpi Icterus - Potential InterferentIef4 Icterus - Above Reportable RangeIef5 Icterus - Below Reportable RangeIef6 Icterus - Prediction Failure (with value reported as NO RESULT)IefJ Icterus - The result of this Sample Integrity Test exceeds the threshold for this assayIefU Icterus - I-Index flag on initial undiluted sampleTecES Turbidity - Examine sampleTecME Turbidity - Mechanical ErrorTecNA Turbidity - SI Globally not availableTecNR Turbidity - Not RunTecpi Turbidity - Potential InterferentTef4 Turbidity - Above Reportable RangeTef5 Turbidity - Below Reportable RangeTef6 Turbidity - Prediction Failure (with value reported as NO RESULT)TefJ Turbidity - The result of this Sample Integrity Test exceeds the threshold for this assayTefU Turbidity - T-Index Flag on initial undiluted sample
3–18 Instrument Manager™ Supplemental Specifications for Laboratory Information System Part No. J328622017-07-10
3.4 VITROS® Result ValuesResult values will be processed in the result record as per Data Innovations ASTM or HL7 interface specifications:• ASTM: R (Result Record), field 4.1• HL7 V2.2: OBX (Observation Result Segment), field 4• HL7 V2.4: OBX (Observation Result Segment), field 5
Warnings
• When a numeric result is received form a VITROS® System and begins with a decimal point, the enGen™ Laboratory Automation System will add a leading zero, for example, when “.3” is received, the enGen™ Laboratory Automation System will display and send “0.3” to the LIS.
• When the result “No Result” is received from a VITROS® System, the enGen™ Laboratory Automation System will change the result to “NO RESULT”.
• Some values are not always numeric, e.g., “No Result”. The LIS should not perform calculations when a non-numeric result is reported.
• The user has the option in Instrument Manager™ to insert the “<“ and “>” symbols when results are out of range. Should this option be selected, the LIS must decode these symbols in order to properly interpret the results. The LIS should not perform calculations when the “<“ or “>” symbols reported with a result, or if a result is accompanied by an “ef4” or “ef5” flag.
3.5 VITROS® Result UnitsVITROS® Systems can express results in different reporting units. The possible units are defined in the publications referenced in Section 1.2.
These units will be processed as per Data Innovations HL7 interface specifications with every numeric result sent by any VITROS® System connected to Instrument Manager™.
• HL7 V2.2: OBX (Observation Result Segment), field 5• HL7 V2.4: OBX (Observation Result Segment), field 6Non-exhaustive list of units currently available on the system:
• %• %Uptake• g/dL• g/L• IU/L• IU/mL• IU/mL• kIU/L• kU/L• meq/L• meq/L• mg/dL• mg/dL(A)• mg/L• mIU/L• mmol/L• mosm/kg• ng/mL• nkat/L• nmol/L• U/L• ug/dL• ug/L• ug/mL• ukat/L• umol/L
Warning
It is critical that the units defined in the LIS match the ones selected by the customer in the VITROS® Systems.
Part No. J32862 Instrument Manager™ Supplemental Specifications for Laboratory Information System 3–192017-07-10
3.5.1 Result Transmission Examples
Example 3: ASTM result transmission
H|\^&|||DI||||||v8.00.0004|P|1|20061128030116
P|1|1234567890||123-45-6789|LAST^FIRST^M||19011215|F|||||789^DrLast^DrFirst^M|||||||||PF2||F|Fac^Ward^Room^Bed^NS1
O|1|123123123||^^^GLU^1.0\^^^CA^1.0\^^^HEM^1.0\^^^ICT^1.0\^^^TUR^1.0|R||||||A||||SER|||||||||FS1
R|1|^^^GLU|111|||ef1||V|||||FS1
R|2|^^^CA|10.0|||ef0||V|||||FS1
R|3|^^^HEM|<15|||ef5||V|||||FS1
R|4|^^^ICT|<2|||ef5||V|||||FS1
R|5|^^^TUR|<20|||ef5||V|||||FS1
L|1|N
Example 4: HL7 result transmission
MSH|^~\&|IM||||20061213075844||ORU^R01|5043483|P|2.2|||AL|NE
PID|1||1234567890||LAST,FIRST||19431111|M|||^^^^^||||||||
ORC|RE|123123123|||||^^^^^R|||||
OBR|1|123123123|||R||20061213040000||^||||||SER^^^|||FS1|^^^^^^P|36^1^|||||||^^^^^R|||||||||||||||||
OBX|1|NM|GLU|86|mg/dL||ef0|||V|||20061213061258||||FS1|20061213061258
OBX|2|NM|CL|94|mmol/L||ef0|||V|||20061213060931||||FS1|20061213060931
OBX|3|NM|K|5.8|mmol/L||ef0|||V|||20061213060950||||FS1|20061213060950
OBX|4|NM|NA|141|mmol/L||ef0|||V|||20061213060941||||FS1|20061213060941
OBX|5|NM|ECO2|26|mmol/L||ef0|||V|||20061213061239||||FS1|20061213061239
OBX|6|NM|PHOS|9.4|mg/dL||ef0|||V|||20061213061308||||FS1|20061213061308
OBX|7|NM|CA|8.8|mg/dL||ef0|||V|||20061213061327||||FS1|20061213061327
OBX|8|NM|MG|2.6|mg/dL||ef0|||V|||20061213061336||||FS1|20061213061336
OBX|9|NM|AGP|21|mmol/L||ef0|||V|||20061213061239||||FS1|20061213061239
OBX|10|NM|HEM|15|||ef5|||V|||20061213060815||||FS1|20061213060815
OBX|11|NM|ICT|2|||ef5|||V|||20061213060815||||FS1|20061213060815
OBX|12|NM|TUR|20|||ef5|||V|||20061213060815||||FS1|20061213060815
Example 5: ASTM Result transmission with sample integrity flag (HefJ)
R|1|^^^IBIL|0.9|||HefJ\ef0||V|||||FS1
Example 6: ASTM Result transmission with sample integrity flag (HefJ) and Flagged Component (FC)
R|1|^^^DBIL|0.8|||HefJ\FC\ef0||V|||||FS1
3–20 Instrument Manager™ Supplemental Specifications for Laboratory Information System Part No. J328622017-07-10
Example 7: ASTM Result transmission of a result record that contains a NO RESULT
R|1|^^^GLU|NO RESULT|||HecNR\IecNR\TecNR\ME\OD\ef6||V|||||FS2
Note:The LIS should not perform numeric calculations on results that are not numeric (e.g. NO RESULT)
Example 8: ASTM Result transmission for microsensor data only (VITROS® 5,1 FS Chemistry System)
H|\^&|||DI||||||v8.00.0004|P|1|20061128011507
P|1|1234567890|||LAST^FIRST|||U|||||||||||||||||
O|1|123123123|^A#^1|^^^HEM^1.0\^^^ICT^1.0\^^^TUR^1.0|R||||||||||SER|||||||||FS2
R|1|^^^HEM|33|||ef0||V|||||FS2
R|2|^^^ICT|<2|||ef5||V|||||FS2
R|3|^^^TUR|<20|||ef5||V|||||FS2
L|1|N
Note: If the VITROS® 5,1 FS Chemistry System has processed a sample multiple times, it is possible microsensor data may be sent multiple times.
Part No. J32862 Instrument Manager™ Supplemental Specifications for Laboratory Information System 3–212017-07-10
3.5.2 Impact of Intermediate Result Handling Selection in Result Transmission
This section only displays ASTM examples, however the same principle will be applicable for HL7 transmission.
Example 9: When "Send All" option is selected
First GLU result, > 625 (neat)
R|1|^^^GLU|625|||HefJ\ef4||V|||||FS2
Note: The “HefJ” flag indicates a potential interference with the result because the sample is hemolized.
Second GLU result, > 1250 (1st dilution)
R|1|^^^GLU|1250|||HefU\ef4||V|||||FS2
Note: The “HefU” flag indicates a potential interference with the result because the sample was found hemolized when undiluted.
Third GLU result, 1500 (2nd dilution)
R|1|^^^GLU|1500|||HefU\ef0||V|||||FS2
Note: The ” HefU” flag indicates a potential interference with the result because the sample was found hemolized when undiluted.
Example 10: When "Send Final" option is selected
R|1|^^^GLU|1500|||HefU\ef0||V|||||FS2
Note:
• In this mode, only the final result is sent, the intermediate results are suppressed and not sent to the LIS.• The “HefU” flag indicates a potential interference with the result because the sample was found hemolized
when undiluted.• When intermediate results are suppressed, the “HefU” flag is the only way for the LIS to know that a potential
interference was detected when the sample was tested undiluted.
3.5.3 Impact of Result Prefix Selection in Result Transmission
This section only presents ASTM examples, however these examples are also applicable for HL7 transmission. This option relates to results that are either above or below the reportable range of the Instrument.
Example 11: When "Insert < > prefixes" is selected, result above reportable range
R|1|^^^CRP|>9.0|||OR\ef4||V|||||FS2
Example 12: When "Insert < > prefixes" is selected, result below reportable range
R|1|^^^LI|<0.2|||OR\ef5||V|||||FS2
Example 13: When "No prefix" is selected, result above reportable range
R|1|^^^CRP|9.0|||OR\ef4||V|||||FS2
Note: The LIS must take into account that the presence of ef4 that indicates the result is above the reportable range.
3–22 Instrument Manager™ Supplemental Specifications for Laboratory Information System Part No. J328622017-07-10
Example 14: When "No prefix" is selected, result below reportable range
R|1|^^^LI|0.2|||OR\ef5||V|||||FS2
Note: The LIS must take into account that the presence of ef5 that indicates the result is below the reportable range.
3.5.4 Impact of Comment Selection in Result Transmission
This section only displays ASTM examples, however these examples are also applicable for HL7 transmission.
Example 15: When “Interpret” is selected, ECi result
R|1|^^^AHCV|0.2|||efQ||V|||||EC1
C|1|Negative
Example 16: When "Interpret" is selected, DAT result
R|1|^^^BARB|2.2|||efR||V|||||FS2
C|1|Reactive
Example 17: When "Interpret" is selected, DAT result above reportable range
R|1|^^^BARB||||OR\efR||V|||||FS2
C|1|Reactive
Note: When a DAT test result is above or below the analyzer's reportable range, the code OR is sent, as well as the “efR” flag. In that case, the result field is empty.
Example 18: When "No Comment" is selected
R|1|^^^AHCV|0.2|||efQ||V|||||EC1
Note: The LIS must translate the “efQ” flag into the correct interpretation.
Example 19: When "No Comment" is selected, DAT result
R|1|^^^BARB|2.2|||efR||V|||||FS2
Note:The LIS must translate the “efR” flag into the correct interpretation.
Example 20: When "No Comment" is selected, DAT result above reportable range
R|1|^^^BARB||||OR\efR||V|||||FS2
Note:• The LIS must translate the “efR” flag into the correct interpretation.• When a DAT test result is above or below the analyzer's reportable range, the code OR is sent, as well as the
“efR” flag. In that case, the result field is empty
Part No. J32862 Instrument Manager™ Supplemental Specifications for Laboratory Information System 3–232017-07-10
3.6 Result Instrument IDInstrument Manager™ returns to the LIS with every test result, an instrument identification associated with the test performed. This instrument identification follows each individual specific result in the same record.
Instrument Manager™ specifies that the “Test Instrument ID” field contains the specific instrument on which the test was performed. The record where the results are located depends on the LIS specific interface. For example, if the test was performed in the HL7 interface, the Instrument ID is located in the OBX record and if the test was performed in the ASTM interface the Instrument ID is located in the R record. Refer to Section 3.5.1 for further examples.
Note: The HL7 driver will also send the Instrument ID in the OBR segment. It is not recommended to collect the Instrument ID from that segment since a specimen might be associated with results obtained from different instruments. LIS should collect the Instrument ID from the OBX segment as indicated above.
3.7 Result Date/TimeResult date/time is provided as per Data Innovations HL7 interface specifications with every result sent by any VITROS® System connected to Instrument Manager™:
• ASTM: R (Result Record), field 13• HL7 V2.2: OBX (Observation Result Segment), field 13• HL7 V2.4: OBX (Observation Result Segment), field 14
3.8 CommentsComments attached to a test result will be processed per Data Innovations ASTM and HL7 specifications.
• ASTM: C (Comment Record), field 4• HL7 V2.2 and V2.4: NTE (Note and Comments Segment), field 3
3–24 Instrument Manager™ Supplemental Specifications for Laboratory Information System Part No. J328622017-07-10
3.9 Instrument IDsInstrument IDs are provided as per Data Innovations ASTM or HL7 interface specifications with every result sent by any VITROS® System connected to Instrument Manager™:
• ASTM: R (Result Record), field 14• HL7 V2.2: OBX (Observation Result Segment), field 17• HL7 V2.4: OBX (Observation Result Segment), field 18
The default connection names are as follows, the names can be changed upon customer request:
VITROS® 5,1 FS Chemistry Systemn (on-track and off-track)• FS1• FS2• FS3• FS4• FS5• FS6• FS7VITROS® ECi/ECiQ Immunodiagnostic System (off-track)• EC1• EC2• EC3• EC4VITROS® 250/350 Chemistry System (off-track)• 251• 252VITROS® 3600 Immunodiagnostic System (on-track and off-track)• 361• 362• 363• 364• 365• 366• 367VITROS® 4600 Chemistry System (on-track)• 461• 462• 463• 464• 465VITROS® 5600 Integrated System (on-track and off-track)• 561• 562• 563• 564• 565• 566• 567
Part No. J32862 Instrument Manager™ Supplemental Specifications for Laboratory Information System 3–252017-07-10
3.10 Operator IDs
Operator IDs are provided as per Data Innovations HL7 interface specifications with every result sent by any VITROS® System connected to Instrument Manager™:
• ASTM: R (Result Record), field 11.4• HL7 V2.2: OBX (Observation Result Segment), field 15• HL7 V2.4: OBX (Observation Result Segment), field 16
3.11 Laboratory Status Message
This message is sent to the LIS when the sample is put in any entry module of the track. The format of the message will depend on the configuration of Instrument Manager™.
Note: This is only available with the enGen™ Laboratory Automation System.
3.11.1 Using ASTM
The enGen™ Laboratory Automation System will send 1 or 2 messages. If the ASTM Robotics driver is used, the following message is sent to the LIS:
<stx>1H|\^&|||||||||v8.00.0004|P|1|20080502102115<cr>P|1|||||||<cr>O|1|2011|||R|||||||||20080502102115|||||||||| TCA IN<cr>C|1|I| TCA IN|I<cr>L|1|N<cr><etx>BD<cr>
<lf>
In addition, the enGen™ Laboratory Automation System will send the following result message regardless of the ASTM driver used. It contains the test code “ARRIV” and the result “1”.
<stx>1H|\^&|||||||||v8.00.0004|P|1|20080502101451<cr>
P|1|||||||<cr>
O|1|2011|^^|^^^ARRIV|R|||||||||20080502101421|SER||||||||| TCA IN<cr>
R|1|^^^ARRIV|1|||||F||||| TCA IN<cr>
L|1|N<cr>
<etx>EF<cr>
<lf>
In these examples, “2011” is the ID of the sample processed on the track. “TCA IN” is the name of the connection used in Instrument Manager™ to connect the track.
3–26 Instrument Manager™ Supplemental Specifications for Laboratory Information System Part No. J328622017-07-10
3.11.2 Using HL7 2.2
By default the enGen™ Laboratory Automation System will send 2 messages. The first message is the actual laboratory status message, as defined in the HL7 2.2 specifications:
<vt>MSH|^~\&|IM||||20080501165832||ORM^O01|18|P|2.2|||AL|NE<cr>
ORC|RE|101|||||^^^^^|||||<cr>
<fs><cr>
The second message is a result message that contains the test code “ARRIV”and the result “1”.
<vt>MSH|^~\&|IM||||20080501165559||ORU^R01|10|P|2.2|||AL|NE<cr>
ORC|RE|101|||||^^^^^R|||||Last^First^M<cr>
OBR|1|101|||R||20071225141503||^||||||SER^^^|Last^First^M|| TCA IN|^^^^^^P|^^|||||||^^^^^R|||||||||||||||||<cr>
OBX|1|NM|ARRIV|1||||||||||||| TCA IN<cr>
<fs><cr>
In these examples, “101” is the ID of the sample processed on the track. “TCA IN” is the name of the connection used in Instrument Manager™ to connect the track.
Note: It is possible multiple messages are sent to the LIS.
3.11.3 Using HL7 2.4
By default the enGen™ Laboratory Automation System will send 2 messages. The first message is the actual laboratory status message, as defined in the HL7 2.4 specifications:
<vt>MSH|^~\&|IM||||20080501170306||SSU^U03|27|P|2.4|||AL|NE<cr>
EQU| TCA IN|20080501170306<cr>
SAC|||1051||||20080501170306|I<cr>
<fs><cr>
The second message is a result message that contains the test code “ARRIV” and the result “1”.
<vt>MSH|^~\&|IM||||20080501173803||ORU^R01|63|P|2.4|||AL|NE<cr>
ORC|RE|1051|||||^^^^^R|||||ID^Last^First^<cr>
OBR|1|1051|||R||20080501173735||^|^^^^^^|||||SER^^^^^^P|ID^Last^First^|| TCA IN||^^|||||||^^^^^R|||||||||||||||||<cr>
OBX|1|NM|ARRIV||1||||||||||||| TCA IN<cr>
<fs><cr>
In these examples, “1051” is the ID of the sample processed on the track. “TCA IN” is the name of the connection used in Instrument Manager™ to connect the track.
Note: It is possible multiple messages are sent to the LIS.
Part No. J32862 Instrument Manager™ Supplemental Specifications for Laboratory Information System 3–272017-07-10
3.12 Laboratory Storage Message
This message is sent to the LIS when the sample is sent to an exit module on the track. The format of the message will depend on the configuration of Instrument Manager™.
Note: This is only available with the enGen™ Laboratory Automation System.
3.12.1 Using ASTM
The enGen™ Laboratory Automation System will send 1 or 2 messages. If the ASTM Robotics driver is used, the following message is sent to the LIS:
<stx>1H|\^&|||||||||v8.00.0004|P|1|20080502103519<cr>
M|1|EQU|^ TCA IN^RD|20080502|OP|L|N<cr>
M|2|SAC|||2011||||20080502103518|R|Carrier|12345|A^1<cr>
L|1|N<cr>
<etx>FE<cr><lf>
In addition, the enGen™ Laboratory Automation System will send the following result message regardless of the ASTM driver used. It contains 2 results:
• “SRACK” is the ID of the rack• “SPOS” is the position of the sample on the rack
<stx>2H|\^&|||||||||v8.00.0004|P|1|20080502103523<cr>
P|1|||||||<cr>
O|1|2011|^^|^^^SRACK\^^^SPOS|R|||||||||20080502101421|SER||||||||| TCA IN<cr>
R|1|^^^SRACK|12345|||||F||||| TCA IN<cr>
R|2|^^^SPOS|1|||||F||||| TCA IN<cr>
L|1|N<cr>
<etx>FF<cr><lf>
In these examples:
• “2011” is the ID of the sample routed to the exit module• “Carrier” is the type of exit rack used• “12345” is the ID of the rack• “1” is the position of the rack
3–28 Instrument Manager™ Supplemental Specifications for Laboratory Information System Part No. J328622017-07-10
3.12.2 Using HL7 2.2
By default, the enGen™ Laboratory Automation System will send 1 message:
<vt>MSH|^~\&|IM||||20080501175749||ORU^R01|129|P|2.2|||AL|NE<cr>
ORC|RE|1056|||||^^^^^R|||||ID^Last^First^<cr>
OBR|1|1056|||R||20080501175709||^|^^^^^^|||||SER^^^|ID^Last^First^|| TCA IN|^^^^^^P|^^|||||||^^^^^R|||||||||||||||||<cr>
OBX|1|TX|SRACK|12345||||||||||||| TCA IN<cr>
OBX|2|NM|SPOS|5||||||||||||| TCA IN<cr>
<fs><cr>
In this example:
• “1056” is the ID of the sample routed to the exit module• “Carrier” is the type of exit rack used• “12345” is the ID of the rack• “5” is the position of the rack
3.12.3 Using HL7 2.4
The first message is the actual laboratory storage message, as defined in the HL7 2.4 specifications:
<vt>MSH|^~\&|IM||||20080501174507||SSU^U03|79|P|2.4|||AL|NE<cr>
EQU| TCA IN|20080501174507<cr>
SAC|||1052||||20080501174507|R|Carrier|M70RCD1R1|1^A<cr>
<fs><cr>
The second message is a result message that contains 2 results:
• “SRACK” is the ID of the rack• “SPOS” is the position of the sample on the rack
<vt>MSH|^~\&|IM||||20080501174510||ORU^R01|86|P|2.4|||AL|NE<cr>
ORC|RE|1052|||||^^^^^R|||||ID^Last^First^<cr>
OBR|1|1052|||R||20080501174309||^|^^^^^^|||||SER^^^^^^P|ID^Last^First^|| TCA IN||^^|||||||^^^^^R|||||||||||||||||<cr>
OBX|1|TX|SRACK||12345||||||||||||| TCA IN<cr>
OBX|2|NM|SPOS||1||||||||||||| TCA IN<cr>
<fs><cr>
In these examples:
• “1052” is the ID of the sample routed to the exit module• “Carrier” is the type of exit rack used• “12345” is the ID of the rack• “1” is the position of the rack
Part No. J32862 Instrument Manager™ Supplemental Specifications for Laboratory Information System 3–292017-07-10
3.13 Aliquoting
3.13.1 Aliquot Requests
Aliquots are created by the enGen™ Laboratory Automation System based on test requests sent by the LIS. The enGen™ Laboratory Automation System expects the LIS to send one or multiple test requests associated with the primary ID of the sample.
Supported test codes are “AT01” - “AT999”.
• Aliquot test codes from “AT01” to “AT949” are assigned to fluid codes SER and PLAS• Aliquot test codes from “AT950” to “AT999” are assigned to fluid codes URI
The enGen™ Laboratory Automation System will determine how many aliquots and their respective volumes based on a table setup in the enGen™ Laboratory Automation System software.
3.13.2 Example of Transmission using HL7 2.2
MSH|^~\&|Sending Application|Sending Facility|Receiving Application|Receiving Facility|20120206161717||ORM^001|358773|P|2.2<cr>
PID|7||12345678||TEST^^|||M<cr>
PV1|1<cr>
ORC|XE<cr>
OBR|1|1202060888||ALT|R||||||||||SER||||||||||XE<cr>
OBR|2|1202060888||AST|R||||||||||SER||||||||||XE<cr>
OBR|3|1202060888||ALKP|R||||||||||SER||||||||||XE<cr>
OBR|4|1202060888||GGT|R||||||||||SER||||||||||XE<cr>
OBR|5|1202060888||AT05|R||||||||||SER||||||||||XE<cr><fs>
3–30 Instrument Manager™ Supplemental Specifications for Laboratory Information System Part No. J328622017-07-10
3.14 Standard Test Codes for Sorting
The system will route the samples to the following destination based on the test code sent by the LIS.
Note: Applies to enGen™ Laboratory Automation System only.
LIS Code Centrifuge Decapper First Exit
XA1 No No <Dr01R1>
CXA1 Yes No <Dr01R1>
DXA1 No Yes <Dr01R1>
CDXA1 Yes Yes <Dr01R1>
XA2 No No <Dr01R2>
CXA2 Yes No <Dr01R2>
DXA2 No Yes <Dr01R2>
CDXA2 Yes Yes <Dr01R2>
XA3 No No <Dr01R3>
CXA3 Yes No <Dr01R3>
DXA3 No Yes <Dr01R3>
CDXA3 Yes Yes <Dr01R3>
XA4 No No <Dr01R4>
CXA4 Yes No <Dr01R4>
DXA4 No Yes <Dr01R4>
CDXA4 Yes Yes <Dr01R4>
XB1 No No <Dr02R1>
CXB1 Yes No <Dr02R1>
DXB1 No Yes <Dr02R1>
CDXB1 Yes Yes <Dr02R1>
XB2 No No <Dr02R2>
CXB2 Yes No <Dr02R2>
DXB2 No Yes <Dr02R2>
CDXB2 Yes Yes <Dr02R2>
XB3 No No <Dr02R3>
CXB3 Yes No <Dr02R3>
DXB3 No Yes <Dr02R3>
CDXB3 Yes Yes <Dr02R3>
XB4 No No <Dr02R4>
CXB4 Yes No <Dr02R4>
DXB4 No Yes <Dr02R4>
CDXB4 Yes Yes <Dr02R4>
XC1 No No <Dr03R1>
CXC1 Yes No <Dr03R1>
DXC1 No Yes <Dr03R1>
CDXC1 Yes Yes <Dr03R1>
Part No. J32862 Instrument Manager™ Supplemental Specifications for Laboratory Information System 3–312017-07-10
XC2 No No <Dr03R2>
CXC2 Yes No <Dr03R2>
DXC2 No Yes <Dr03R2>
CDXC2 Yes Yes <Dr03R2>
XC3 No No <Dr03R3>
CXC3 Yes No <Dr03R3>
DXC3 No Yes <Dr03R3>
CDXC3 Yes Yes <Dr03R3>
XC4 No No <Dr03R4>
CXC4 Yes No <Dr03R4>
DXC4 No Yes <Dr03R4>
CDXC4 Yes Yes <Dr03R4>
XD1 No No <Dr04R1>
CXD1 Yes No <Dr04R1>
DXD1 No Yes <Dr04R1>
CDXD1 Yes Yes <Dr04R1>
XD2 No No <Dr04R2>
CXD2 Yes No <Dr04R2>
DXD2 No Yes <Dr04R2>
CDXD2 Yes Yes <Dr04R2>
XD3 No No <Dr04R3>
CXD3 Yes No <Dr04R3>
DXD3 No Yes <Dr04R3>
CDXD3 Yes Yes <Dr04R3>
XD4 No No <Dr04R4>
CXD4 Yes No <Dr04R4>
DXD4 No Yes <Dr04R4>
CDXD4 Yes Yes <Dr04R4>
XE1 No No <Dr05R1>
CXE1 Yes No <Dr05R1>
DXE1 No Yes <Dr05R1>
CDXE1 Yes Yes <Dr05R1>
XE2 No No <Dr05R2>
CXE2 Yes No <Dr05R2>
DXE2 No Yes <Dr05R2>
CDXE2 Yes Yes <Dr05R2>
XE3 No No <Dr05R3>
CXE3 Yes No <Dr05R3>
DXE3 No Yes <Dr05R3>
CDXE3 Yes Yes <Dr05R3>
XE4 No No <Dr05R4>
LIS Code Centrifuge Decapper First Exit
3–32 Instrument Manager™ Supplemental Specifications for Laboratory Information System Part No. J328622017-07-10
CXE4 Yes No <Dr05R4>
DXE4 No Yes <Dr05R4>
CDXE4 Yes Yes <Dr05R4>
XF1 No No <Dr06R1>
CXF1 Yes No <Dr06R1>
DXF1 No Yes <Dr06R1>
CDXF1 Yes Yes <Dr06R1>
XF2 No No <Dr06R2>
CXF2 Yes No <Dr06R2>
DXF2 No Yes <Dr06R2>
CDXF2 Yes Yes <Dr06R2>
XF3 No No <Dr06R3>
CXF3 Yes No <Dr06R3>
DXF3 No Yes <Dr06R3>
CDXF3 Yes Yes <Dr06R3>
XF4 No No <Dr06R4>
CXF4 Yes No <Dr06R4>
DXF4 No Yes <Dr06R4>
CDXF4 Yes Yes <Dr06R4>
XG1 No No <Dr07R1>
CXG1 Yes No <Dr07R1>
DXG1 No Yes <Dr07R1>
CDXG1 Yes Yes <Dr07R1>
XG2 No No <Dr07R2>
CXG2 Yes No <Dr07R2>
DXG2 No Yes <Dr07R2>
CDXG2 Yes Yes <Dr07R2>
XG3 No No <Dr07R3>
CXG3 Yes No <Dr07R3>
DXG3 No Yes <Dr07R3>
CDXG3 Yes Yes <Dr07R3>
XG4 No No <Dr07R4>
CXG4 Yes No <Dr07R4>
DXG4 No Yes <Dr07R4>
CDXG4 Yes Yes <Dr07R4>
XH1 No No <Dr08R1>
CXH1 Yes No <Dr08R1>
DXH1 No Yes <Dr08R1>
CDXH1 Yes Yes <Dr08R1>
XH2 No No <Dr08R2>
CXH2 Yes No <Dr08R2>
DXH2 No Yes <Dr08R2>
LIS Code Centrifuge Decapper First Exit
Part No. J32862 Instrument Manager™ Supplemental Specifications for Laboratory Information System 3–332017-07-10
CDXH2 Yes Yes <Dr08R2>
XH3 No No <Dr08R3>
CXH3 Yes No <Dr08R3>
DXH3 No Yes <Dr08R3>
CDXH3 Yes Yes <Dr08R3>
XH4 No No <Dr08R4>
CXH4 Yes No <Dr08R4>
DXH4 No Yes <Dr08R4>
CDXH4 Yes Yes <Dr08R4>
XI1 No No <Dr09R1>
CXI1 Yes No <Dr09R1>
DXI1 No Yes <Dr09R1>
CDXI1 Yes Yes <Dr09R1>
XI2 No No <Dr09R2>
CXI2 Yes No <Dr09R2>
DXI2 No Yes <Dr09R2>
CDXI2 Yes Yes <Dr09R2>
XI3 No No <Dr09R3>
CXI3 Yes No <Dr09R3>
DXI3 No Yes <Dr09R3>
CDXI3 Yes Yes <Dr09R3>
XI4 No No <Dr09R4>
CXI4 Yes No <Dr09R4>
DXI4 No Yes <Dr09R4>
CDXI4 Yes Yes <Dr09R4>
XJ1 No No <Dr10R1>
CXJ1 Yes No <Dr10R1>
DXJ1 No Yes <Dr10R1>
CDXJ1 Yes Yes <Dr10R1>
XJ2 No No <Dr10R2>
CXJ2 Yes No <Dr10R2>
DXJ2 No Yes <Dr10R2>
CDXJ2 Yes Yes <Dr10R2>
XJ3 No No <Dr10R3>
CXJ3 Yes No <Dr10R3>
DXJ3 No Yes <Dr10R3>
CDXJ3 Yes Yes <Dr10R3>
XJ4 No No <Dr10R4>
CXJ4 Yes No <Dr10R4>
DXJ4 No Yes <Dr10R4>
CDXJ4 Yes Yes <Dr10R4>
LIS Code Centrifuge Decapper First Exit
3–34 Instrument Manager™ Supplemental Specifications for Laboratory Information System Part No. J328622017-07-10
XK1 No No <Dr11R1>
CXK1 Yes No <Dr11R1>
DXK1 No Yes <Dr11R1>
CDXK1 Yes Yes <Dr11R1>
XK2 No No <Dr11R2>
CXK2 Yes No <Dr11R2>
DXK2 No Yes <Dr11R2>
CDXK2 Yes Yes <Dr11R2>
XK3 No No <Dr11R3>
CXK3 Yes No <Dr11R3>
DXK3 No Yes <Dr11R3>
CDXK3 Yes Yes <Dr11R3>
XK4 No No <Dr11R4>
CXK4 Yes No <Dr11R4>
DXK4 No Yes <Dr11R4>
CDXK4 Yes Yes <Dr11R4>
XL1 No No <Dr12R1>
CXL1 Yes No <Dr12R1>
DXL1 No Yes <Dr12R1>
CDXL1 Yes Yes <Dr12R1>
XL2 No No <Dr12R2>
CXL2 Yes No <Dr12R2>
DXL2 No Yes <Dr12R2>
CDXL2 Yes Yes <Dr12R2>
XL3 No No <Dr12R3>
CXL3 Yes No <Dr12R3>
DXL3 No Yes <Dr12R3>
CDXL3 Yes Yes <Dr12R3>
XL4 No No <Dr12R4>
CXL4 Yes No <Dr12R4>
DXL4 No Yes <Dr12R4>
CDXL4 Yes Yes <Dr12R4>
XM1 No No <Dr13R1>
CXM1 Yes No <Dr13R1>
DXM1 No Yes <Dr13R1>
CDXM1 Yes Yes <Dr13R1>
XM2 No No <Dr13R2>
CXM2 Yes No <Dr13R2>
DXM2 No Yes <Dr13R2>
CDXM2 Yes Yes <Dr13R2>
XM3 No No <Dr13R3>
LIS Code Centrifuge Decapper First Exit
Part No. J32862 Instrument Manager™ Supplemental Specifications for Laboratory Information System 3–352017-07-10
CXM3 Yes No <Dr13R3>
DXM3 No Yes <Dr13R3>
CDXM3 Yes Yes <Dr13R3>
XM4 No No <Dr13R4>
CXM4 Yes No <Dr13R4>
DXM4 No Yes <Dr13R4>
CDXM4 Yes Yes <Dr13R4>
XN1 No No <Dr14R1>
CXN1 Yes No <Dr14R1>
DXN1 No Yes <Dr14R1>
CDXN1 Yes Yes <Dr14R1>
XN2 No No <Dr14R2>
CXN2 Yes No <Dr14R2>
DXN2 No Yes <Dr14R2>
CDXN2 Yes Yes <Dr14R2>
XN3 No No <Dr14R3>
CXN3 Yes No <Dr14R3>
DXN3 No Yes <Dr14R3>
CDXN3 Yes Yes <Dr14R3>
XN4 No No <Dr14R4>
CXN4 Yes No <Dr14R4>
DXN4 No Yes <Dr14R4>
CDXN4 Yes Yes <Dr14R4>
XO1 No No <Dr15R1>
CXO1 Yes No <Dr15R1>
DXO1 No Yes <Dr15R1>
CDXO1 Yes Yes <Dr15R1>
XO2 No No <Dr15R2>
CXO2 Yes No <Dr15R2>
DXO2 No Yes <Dr15R2>
CDXO2 Yes Yes <Dr15R2>
XO3 No No <Dr15R3>
CXO3 Yes No <Dr15R3>
DXO3 No Yes <Dr15R3>
CDXO3 Yes Yes <Dr15R3>
XO4 No No <Dr15R4>
CXO4 Yes No <Dr15R4>
DXO4 No Yes <Dr15R4>
CDXO4 Yes Yes <Dr15R4>
XP1 No No <Dr16R1>
CXP1 Yes No <Dr16R1>
LIS Code Centrifuge Decapper First Exit
3–36 Instrument Manager™ Supplemental Specifications for Laboratory Information System Part No. J328622017-07-10
DXP1 No Yes <Dr16R1>
CDXP1 Yes Yes <Dr16R1>
XP2 No No <Dr16R2>
CXP2 Yes No <Dr16R2>
DXP2 No Yes <Dr16R2>
CDXP2 Yes Yes <Dr16R2>
XP3 No No <Dr16R3>
CXP3 Yes No <Dr16R3>
DXP3 No Yes <Dr16R3>
CDXP3 Yes Yes <Dr16R3>
XP4 No No <Dr16R4>
CXP4 Yes No <Dr16R4>
DXP4 No Yes <Dr16R4>
CDXP4 Yes Yes <Dr16R4>
XQ1 No No <Dr17R1>
CXQ1 Yes No <Dr17R1>
DXQ1 No Yes <Dr17R1>
CDXQ1 Yes Yes <Dr17R1>
XQ2 No No <Dr17R2>
CXQ2 Yes No <Dr17R2>
DXQ2 No Yes <Dr17R2>
CDXQ2 Yes Yes <Dr17R2>
XQ3 No No <Dr17R3>
CXQ3 Yes No <Dr17R3>
DXQ3 No Yes <Dr17R3>
CDXQ3 Yes Yes <Dr17R3>
XQ4 No No <Dr17R4>
CXQ4 Yes No <Dr17R4>
DXQ4 No Yes <Dr17R4>
CDXQ4 Yes Yes <Dr17R4>
XR1 No No <Dr18R1>
CXR1 Yes No <Dr18R1>
DXR1 No Yes <Dr18R1>
CDXR1 Yes Yes <Dr18R1>
XR2 No No <Dr18R2>
CXR2 Yes No <Dr18R2>
DXR2 No Yes <Dr18R2>
CDXR2 Yes Yes <Dr18R2>
XR3 No No <Dr18R3>
CXR3 Yes No <Dr18R3>
DXR3 No Yes <Dr18R3>
LIS Code Centrifuge Decapper First Exit
Part No. J32862 Instrument Manager™ Supplemental Specifications for Laboratory Information System 3–372017-07-10
CDXR3 Yes Yes <Dr18R3>
XR4 No No <Dr18R4>
CXR4 Yes No <Dr18R4>
DXR4 No Yes <Dr18R4>
CDXR4 Yes Yes <Dr18R4>
XS1 No No <Dr19R1>
CXS1 Yes No <Dr19R1>
DXS1 No Yes <Dr19R1>
CDXS1 Yes Yes <Dr19R1>
XS2 No No <Dr19R2>
CXS2 Yes No <Dr19R2>
DXS2 No Yes <Dr19R2>
CDXS2 Yes Yes <Dr19R2>
XS3 No No <Dr19R3>
CXS3 Yes No <Dr19R3>
DXS3 No Yes <Dr19R3>
CDXS3 Yes Yes <Dr19R3>
XS4 No No <Dr19R4>
CXS4 Yes No <Dr19R4>
DXS4 No Yes <Dr19R4>
CDXS4 Yes Yes <Dr19R4>
XT1 No No <Dr20R1>
CXT1 Yes No <Dr20R1>
DXT1 No Yes <Dr20R1>
CDXT1 Yes Yes <Dr20R1>
XT2 No No <Dr20R2>
CXT2 Yes No <Dr20R2>
DXT2 No Yes <Dr20R2>
CDXT2 Yes Yes <Dr20R2>
XT3 No No <Dr20R3>
CXT3 Yes No <Dr20R3>
DXT3 No Yes <Dr20R3>
CDXT3 Yes Yes <Dr20R3>
XT4 No No <Dr20R4>
CXT4 Yes No <Dr20R4>
DXT4 No Yes <Dr20R4>
CDXT4 Yes Yes <Dr20R4>
XU1 No No <Dr21R1>
CXU1 Yes No <Dr21R1>
DXU1 No Yes <Dr21R1>
CDXU1 Yes Yes <Dr21R1>
LIS Code Centrifuge Decapper First Exit
3–38 Instrument Manager™ Supplemental Specifications for Laboratory Information System Part No. J328622017-07-10
XU2 No No <Dr21R2>
CXU2 Yes No <Dr21R2>
DXU2 No Yes <Dr21R2>
CDXU2 Yes Yes <Dr21R2>
XU3 No No <Dr21R3>
CXU3 Yes No <Dr21R3>
DXU3 No Yes <Dr21R3>
CDXU3 Yes Yes <Dr21R3>
XU4 No No <Dr21R4>
CXU4 Yes No <Dr21R4>
DXU4 No Yes <Dr21R4>
CDXU4 Yes Yes <Dr21R4>
XV1 No No <Dr22R1>
CXV1 Yes No <Dr22R1>
DXV1 No Yes <Dr22R1>
CDXV1 Yes Yes <Dr22R1>
XV2 No No <Dr22R2>
CXV2 Yes No <Dr22R2>
DXV2 No Yes <Dr22R2>
CDXV2 Yes Yes <Dr22R2>
XV3 No No <Dr22R3>
CXV3 Yes No <Dr22R3>
DXV3 No Yes <Dr22R3>
CDXV3 Yes Yes <Dr22R3>
XV4 No No <Dr22R4>
CXV4 Yes No <Dr22R4>
DXV4 No Yes <Dr22R4>
CDXV4 Yes Yes <Dr22R4>
XW1 No No <Dr23R1>
CXW1 Yes No <Dr23R1>
DXW1 No Yes <Dr23R1>
CDXW1 Yes Yes <Dr23R1>
XW2 No No <Dr23R2>
CXW2 Yes No <Dr23R2>
DXW2 No Yes <Dr23R2>
CDXW2 Yes Yes <Dr23R2>
XW3 No No <Dr23R3>
CXW3 Yes No <Dr23R3>
DXW3 No Yes <Dr23R3>
CDXW3 Yes Yes <Dr23R3>
XW4 No No <Dr23R4>
LIS Code Centrifuge Decapper First Exit
Part No. J32862 Instrument Manager™ Supplemental Specifications for Laboratory Information System 3–392017-07-10
CXW4 Yes No <Dr23R4>
DXW4 No Yes <Dr23R4>
CDXW4 Yes Yes <Dr23R4>
XX1 No No <Dr24R1>
CXX1 Yes No <Dr24R1>
DXX1 No Yes <Dr24R1>
CDXX1 Yes Yes <Dr24R1>
XX2 No No <Dr24R2>
CXX2 Yes No <Dr24R2>
DXX2 No Yes <Dr24R2>
CDXX2 Yes Yes <Dr24R2>
XX3 No No <Dr24R3>
CXX3 Yes No <Dr24R3>
DXX3 No Yes <Dr24R3>
CDXX3 Yes Yes <Dr24R3>
XX4 No No <Dr24R4>
CXX4 Yes No <Dr24R4>
DXX4 No Yes <Dr24R4>
CDXX4 Yes Yes <Dr24R4>
X1 No No <SORT01>
CX1 Yes No <SORT01>
DX1 No Yes <SORT01>
CDX1 Yes Yes <SORT01>
X2 No No <SORT02>
CX2 Yes No <SORT02>
DX2 No Yes <SORT02>
CDX2 Yes Yes <SORT02>
X3 No No <SORT03>
CX3 Yes No <SORT03>
DX3 No Yes <SORT03>
CDX3 Yes Yes <SORT03>
X4 No No <SORT04>
CX4 Yes No <SORT04>
DX4 No Yes <SORT04>
CDX4 Yes Yes <SORT04>
X5 No No <SORT05>
CX5 Yes No <SORT05>
DX5 No Yes <SORT05>
CDX5 Yes Yes <SORT05>
X6 No No <SORT06>
CX6 Yes No <SORT06>
LIS Code Centrifuge Decapper First Exit
3–40 Instrument Manager™ Supplemental Specifications for Laboratory Information System Part No. J328622017-07-10
DX6 No Yes <SORT06>
CDX6 Yes Yes <SORT06>
X7 No No <SORT07>
CX7 Yes No <SORT07>
DX7 No Yes <SORT07>
CDX7 Yes Yes <SORT07>
X8 No No <SORT08>
CX8 Yes No <SORT08>
DX8 No Yes <SORT08>
CDX8 Yes Yes <SORT08>
X9 No No <SORT09>
CX9 Yes No <SORT09>
DX9 No Yes <SORT09>
CDX9 Yes Yes <SORT09>
X10 No No <SORT10>
CX10 Yes No <SORT10>
DX10 No Yes <SORT10>
CDX10 Yes Yes <SORT10>
X11 No No <SORT11>
CX11 Yes No <SORT11>
DX11 No Yes <SORT11>
CDX11 Yes Yes <SORT11>
X12 No No <SORT12>
CX12 Yes No <SORT12>
DX12 No Yes <SORT12>
CDX12 Yes Yes <SORT12>
X13 No No <SORT13>
CX13 Yes No <SORT13>
DX13 No Yes <SORT13>
CDX13 Yes Yes <SORT13>
X14 No No <SORT14>
CX14 Yes No <SORT14>
DX14 No Yes <SORT14>
CDX14 Yes Yes <SORT14>
X15 No No <SORT15>
CX15 Yes No <SORT15>
DX15 No Yes <SORT15>
CDX15 Yes Yes <SORT15>
X16 No No <SORT16>
CX16 Yes No <SORT16>
DX16 No Yes <SORT16>
LIS Code Centrifuge Decapper First Exit
Part No. J32862 Instrument Manager™ Supplemental Specifications for Laboratory Information System 3–412017-07-10
CDX16 Yes Yes <SORT16>
X17 No No <SORT17>
CX17 Yes No <SORT17>
DX17 No Yes <SORT17>
CDX17 Yes Yes <SORT17>
X18 No No <SORT18>
CX18 Yes No <SORT18>
DX18 No Yes <SORT18>
CDX18 Yes Yes <SORT18>
X19 No No <SORT19>
CX19 Yes No <SORT19>
DX19 No Yes <SORT19>
CDX19 Yes Yes <SORT19>
X20 No No <SORT20>
CX20 Yes No <SORT20>
DX20 No Yes <SORT20>
CDX20 Yes Yes <SORT20>
X21 No No <SORT21>
CX21 Yes No <SORT21>
DX21 No Yes <SORT21>
CDX21 Yes Yes <SORT21>
X22 No No <SORT22>
CX22 Yes No <SORT22>
DX22 No Yes <SORT22>
CDX22 Yes Yes <SORT22>
X23 No No <SORT23>
CX23 Yes No <SORT23>
DX23 No Yes <SORT23>
CDX23 Yes Yes <SORT23>
X24 No No <SORT24>
CX24 Yes No <SORT24>
DX24 No Yes <SORT24>
CDX24 Yes Yes <SORT24>
X25 No No <SORT25>
CX25 Yes No <SORT25>
DX25 No Yes <SORT25>
CDX25 Yes Yes <SORT25>
X26 No No <SORT26>
CX26 Yes No <SORT26>
DX26 No Yes <SORT26>
CDX26 Yes Yes <SORT26>
LIS Code Centrifuge Decapper First Exit
3–42 Instrument Manager™ Supplemental Specifications for Laboratory Information System Part No. J328622017-07-10
X27 No No <SORT27>
CX27 Yes No <SORT27>
DX27 No Yes <SORT27>
CDX27 Yes Yes <SORT27>
X28 No No <SORT28>
CX28 Yes No <SORT28>
DX28 No Yes <SORT28>
CDX28 Yes Yes <SORT28>
X29 No No <SORT29>
CX29 Yes No <SORT29>
DX29 No Yes <SORT29>
CDX29 Yes Yes <SORT29>
X30 No No <SORT30>
CX30 Yes No <SORT30>
DX30 No Yes <SORT30>
CDX30 Yes Yes <SORT30>
LIS Code Centrifuge Decapper First Exit
Part No. J32862 Instrument Manager™ Supplemental Specifications for Laboratory Information System 3–432017-07-10
3–44 Instrument Manager™ Supplemental Specifications for Laboratory Information System Part No. J328622017-07-10
4 Non-OCD Instrument
4.1 IntroductionThis section applies only when a non-OCD Instrument is connected to the enGen™ Laboratory Automation System.
NOTE: Ortho Clinical Diagnostics makes no claims regarding the ability to interface non-OCD analyzers nor to the performance of non-OCD analyzers that are interfaced to the enGen™ Laboratory Automation System.
4.2 Non-OCD Instrument Interface Options
4.2.1 Option 1: Non-OCD Instrument is NOT interfaced to OCD-IM but directly to the LIS
This is the preferred way of interfacing non-OCD instruments to the enGen™ Laboratory Automation System and implies that the non-OCD instrument is LAS interfaced to the enGen™ Laboratory Automation System (On-Track) but the instrument LIS interface is directly connected to the LIS. In this case, OCD will provide the supplementary document “LIS Instructions for non-OCD Instrument Routing connected to the enGen Track” which will be used as a reference for routing purposes. However, all test order and result messages must be handled between the LIS and the non-OCD Instrument according to the LIS specifications from the non-OCD Instrument vender.
4.2.2 Option 2: Non-OCD Instrument is interfaced to OCD-IM
In case a non-OCD instrument is connected to the enGen™ Laboratory Automation System and the instrument is LIS interfaced to the OCD-IM, the customer or the LIS vendor should contact OCD for additional instructions.
Part No. J32862 Instrument Manager™ Supplemental Specifications for Laboratory Information System 4–12017-07-10
4–2 Instrument Manager™ Supplemental Specifications for Laboratory Information System Part No. J328622017-07-10
This page is intentionally left blank.