SURGICAL TECHNIQUE Integra MBA Titanium Subtalar Implant PRODUCTS FOR SALE IN EUROPE, MIDDLE-EAST and AFRICA ONLY
Transcript
Page 1: Integra · integra™ MBa™ Titanium subtalar implant Surgical technique Integra™ MBA™4 PRODUCTS FOR SALE IN EUROPE, MIDDLE-EAST and AFRICA ONLY Surgical Technique 1. Incision

s u rg i c a l T ec h n i q u eIntegra™

MBA™ Titanium Subtalar Implant

PRODUCTS FOR SALE IN EUROPE, MIDDLE-EAST and AFRICA ONLY

Page 2: Integra · integra™ MBa™ Titanium subtalar implant Surgical technique Integra™ MBA™4 PRODUCTS FOR SALE IN EUROPE, MIDDLE-EAST and AFRICA ONLY Surgical Technique 1. Incision

integra™

MBa™ Titanium subtalar implant

Surgical technique Integra™ MBA™ PRODUCTS FOR SALE IN EUROPE, MIDDLE-EAST and AFRICA ONLY

2

Table of contents

Implant description �������������������������������������������������������������������������������������������������03The subtalar MBA™ system offers these unique features ����������������������������������03Indications ����������������������������������������������������������������������������������������������������������������03Contraindications ����������������������������������������������������������������������������������������������������03

Surgical technique 04

1� Incision and dissection �������������������������������������������������������������������������������������� 042� Probe insertion ��������������������������������������������������������������������������������������������������� 043� Guide pin insertion ��������������������������������������������������������������������������������������������� 054� Sizer insertion ����������������������������������������������������������������������������������������������������� 055� Trial implant insertion ���������������������������������������������������������������������������������������� 056� Intra-operative radiographs ����������������������������������������������������������������������������� 057� Implantation �������������������������������������������������������������������������������������������������������� 068� Guide pin removal and closure ������������������������������������������������������������������������� 06

Post-op/Follow-up 06

Instructions for use 07

References 08

Features 08

Component material 08

Page 3: Integra · integra™ MBa™ Titanium subtalar implant Surgical technique Integra™ MBA™4 PRODUCTS FOR SALE IN EUROPE, MIDDLE-EAST and AFRICA ONLY Surgical Technique 1. Incision

Pre-op lateral view Post-op lateral view Post-op ap view

3Surgical technique Integra™ MBA™ PRODUCTS FOR SALE IN EUROPE, MIDDLE-EAST and AFRICA ONLY

Indications

• Pediatric pes valgus�

• Adult posterior tibial tendon dysfunction�

Contraindications

• Vertical calcaneus�

• Rigid flatfoot (unreducible)�

• Rearfoot varus�

• Peroneal spasm�

• Ankle joint valgus�

• Excessive ligamentous laxity�

• Degenerative joint disease of the subtalar joint�

• Severe obesity�

• Asymptomatic flatfoot condition�

• Patients less than 3 years of age�

The subtalar MBA™ system offers these unique features

• Barrel-shaped implant provides optimal medial support�

• Uniform implant diameter minimizes lateral discomfort�

• Exclusive slotted design helps prevent implant extrusion�

• Cannulated, color-coded instrumentation simplifies surgical procedure�

Implant Description

The subtalar MBA™ titanium implant is the original, time-tested metallic arthrodesis implant for the correction of pediatric pes valgus and adult posterior tibial tendon dysfunction�

Page 4: Integra · integra™ MBa™ Titanium subtalar implant Surgical technique Integra™ MBA™4 PRODUCTS FOR SALE IN EUROPE, MIDDLE-EAST and AFRICA ONLY Surgical Technique 1. Incision

integra™

MBa™ Titanium subtalar implant

Surgical technique Integra™ MBA™ PRODUCTS FOR SALE IN EUROPE, MIDDLE-EAST and AFRICA ONLY

4

Surgical Technique

1. Incision and dissection

Make a 1-3 cm incision over the sinus tarsi along the relaxed skin tension lines (Fig. 1)�

• The intermediate dorsal cutaneous nerves should course superior to the incision�

• The sural nerve should course inferior to the incision�

• Incise the deep fascia�

The anterior lateral edge of the posterior facet of the calcaneus should be palpable with instrumentation upon completing the dissection into the tarsal canal (Fig. 2)�

2. Probe insertion

• Insert the yellow probe instrument through the sinus tarsi into the sinus canalis from lateral to medial until tenting is noted on the medial aspect of the foot (Fig. 3)�

• The probe should be positioned perpendicular to the lateral wall of the calcaneus, angled slightly posterior and superior�

• Move the probe in a clockwise and counter clockwise direction to slightly dilate the tarsal canal�

• Remove the probe�

Fig. 1

Fig. 2 Fig. 3

incision

Probe instrument

05-1002 Yellow probe

Page 5: Integra · integra™ MBa™ Titanium subtalar implant Surgical technique Integra™ MBA™4 PRODUCTS FOR SALE IN EUROPE, MIDDLE-EAST and AFRICA ONLY Surgical Technique 1. Incision

05-0017 Guide pin

05-100605-100805-100905-101005-1012 Sizers

05-1020 3�5 mm hex

5Surgical technique Integra™ MBA™ PRODUCTS FOR SALE IN EUROPE, MIDDLE-EAST and AFRICA ONLY

5. Trial implant insertion

• After the appropriate size is determined, remove the sizer�

• Insert the corresponding trial implant using the cannulated 3.5 mm hex insertion tool (Fig. 6)�

• Asses the range of motion of the subtalar joint�

• Use intra-operative imaging to evaluate the degree of correction and placement of the trial implant�

3. Guide pin insertion

• Insert the guide pin and make a small incision medially to allow the guide pin to exit through the medial aspect of the foot (Fig. 4)�

• The distal aspect of the pin should exit the skin just inferior to the tibialis posterior tendon and anterior and slightly inferior to the medial malleolus�

4. Sizer insertion

• Place the cannulated 6 mm sizer over the guide pin, and insert it through the sinus tarsi into the sinus canalis from lateral to medial (Fig. 5)�

• Assess the range of motion of the subtalar joint�

• Continue to insert the remaining sizers (8, 9, 10, 12 mm) until proper correction is achieved�

• The appropriate size will allow approximately 2-4 degrees of subtalar joint eversion�

6. Intra-operative radiographs

• To determine the correct position on AP (anteroposterior) view, the leading edge of the trial implant should approach, but not cross, the longitudinal bisection of the talus (Fig. 7)�

• The trailing edge of the implant should be at least 5 mm medial to the lateral wall of the calcaneus�

• Examining the lateral view, the trial implant should be angled posterior, and the implant should not be sitting on the floor of the calcaneus�

Fig. 4

15°

45°

05-1002 Yellow probe

guide pin

Fig. 5

6 mm sizer

3.5 mm hex

Fig. 6

longitudinal bisection of the talus

guide pin

5 mm

Fig. 7

Trial

Page 6: Integra · integra™ MBa™ Titanium subtalar implant Surgical technique Integra™ MBA™4 PRODUCTS FOR SALE IN EUROPE, MIDDLE-EAST and AFRICA ONLY Surgical Technique 1. Incision

Surgical technique Integra™ MBA™ PRODUCTS FOR SALE IN EUROPE, MIDDLE-EAST and AFRICA ONLY

6

7. Implantation

• Once the appropriate sized trial implant and placement are determined, remove the trial implant�

• The equivalent size sterile implant is placed onto the insertion tool over the guide pin and threaded in a clockwise direction until clinical correction is noted (Fig. 9)�

• Intra-operative imaging is essential to verify proper positioning of the implant (Refer to Step 6)�

• Once the implant has been properly positioned, assess the range of motion of the subtalar joint�

• The appropriate size will allow 2-4 degrees of subtalar joint eversion.

insertion tool

8. Guide pin removal and closure

• Remove the guide pin medially when satisfactory positioning of the MBA™ implant is obtained (Fig. 10)�

• Copiously irrigate the area with saline and reevaluate subtalar joint motion�

• Close the deep tissue, fascia, and subcutaneous and skin layers�

• Place the foot in a mildly compressive dressing�

Post-op / Follow-up

• Restrict ambulation for the first 48 hours followed by protective weight-bearing in a removable, below-the-knee, walking cast for two weeks�

• Allow a gradual return to activity over the course of the next month�

• Typically, adjunctive procedures are performed, so the appropriate post-operative care should be followed for these procedures�

Fig. 10

Fig. 9

05-1002 Probe instrument

05-1020 Insertion Device

05-0017Guide Pin 2 mm

Page 7: Integra · integra™ MBa™ Titanium subtalar implant Surgical technique Integra™ MBA™4 PRODUCTS FOR SALE IN EUROPE, MIDDLE-EAST and AFRICA ONLY Surgical Technique 1. Incision

7Surgical technique Integra™ MBA™ PRODUCTS FOR SALE IN EUROPE, MIDDLE-EAST and AFRICA ONLY

1. Description:

The Subtalar MBA System consists of a soft threaded im-plant, designed to be inserted between the posterior and middle facets of the subtalar joint and corresponding in-strumentation to facilitate insertion. It is important that the instruments and trial implants used are those specifically designed for this device to ensure accurate insertion.The Subtalar MBA implant is cylindrical in shape and in-corporates a center cannula designed for use with a guide wire to facilitate proper placement of the implant. An internal hex-head allows for maximum torque with minimal risk of stripping. External rounded threads increase ease of inser-tion, while slotting of the implant absorbs peak shock and allows fibrous in-growth.

2. Indications:The Subtalar MBA System is indicated for use in treatment of the hyperpronated foot and stabilization of the subtalar joint. It is designed to block the posterior and inferior dis-placement of the talus, thus allowing normal subtalar joint motion while blocking excessive pronation and the resulting sequela.

- Severely pronated foot - Walking intemperance - Calcaneal stance position greater than 5° - Manually correctable deformities - Mid-tarsal breech (arch pain) - Forefoot varus greater than 10°

3. Contraindications:The Subtalar MBA implant is contraindicated for use in pa-tients with the following conditions:

- Active local infection. (Any evidence of infection) - Metal sensitivity or allergic reaction to foreign bodies - Poor or insufficient bone stock - The presence of any clinical or functional abnormalities that would preclude the potential of achieving a good result for the patient - Other conditions that may place the patient at risk (Physiologically)

4. Warnings:For safe and effective use of this implant system, the sur-geon should be familiar with the recommended surgical pro-cedure for this device (see the Surgical Protocol, 81-0006). Improper selection, placement, positioning, or seating of the implant may result in unusual loading conditions which could affect the long-term service life of the implant.In every case, accepted surgical practices should be fol-lowed in post-operative care. The patient should be made aware of the limitations of the subtalar implant and that physical activity and full weight bearing have been impli-cated in premature failure of similar devices.Patient sensitivity to implant materials should be considered and assessed prior to surgery.

5. Adverse effects:The following are specific adverse effects, which should be understood by the surgeon and explained to the patient. These do not include all adverse effects, which can occur with surgery in general, but are important considerations specific to metallic internal stabilization devices. General surgical risks should be explained to the patient prior to surgery.

- Infection - Pain, discomfort, or abnormal sensations due to presence of the implant - Metal sensitivity, or allergic reaction to a foreign body - Migration of the implant, loosening of the implant - Delayed correction in alignment - Decrease in bone density due to stress shielding - Bursitis

6. Implant materials:The Subtalar MBA implant is manufactured from titanium alloy (Ti-6Al-4V ELI, ASTM F 136, ISO 5832-3).

7. Packaging and sterility:All implants have been sterilized with a minimum dose of 2.5 Mrads of gamma irradiation, sterility assurance level 10 (E-6). Prior to use, inspect package for damage which may compromise sterility. If damaged, the product must be as-sumed to be non-sterile. If a sterile metal implant package is opened but the device is not used, it should be resterilized according to the hospital’s standard practice. The Subtalar

MBA implants and instruments may be steam sterilized us-ing the following process parameters:

Method Steam SteamCycle Gravity VacuumTemperature 270°F (132°C) 270°F (132°C)Exposure time 30 Minutes 4 Minutes

Surgical implants should not be reused. Any implant once used should be discarded. Even though it may appear un-damaged, it may have small defects or internal stress pat-terns which may lead to failure.

8. Instructions for reprocessing:The instrument tray and its contents are provided non-sterile and must be sterilized prior to surgery. The following instructions should be used for cleaning and decontaminat-ing non-sterile product.All products should be cleaned, decontaminated and steri-lized before use. Always immediately clean and decontami-nate all devices that have been soiled.

Manual Cleaning Procedure: - Prepare a neutral pH enzymatic detergent. Enzol®, as per the manufacturer’s recommendation at 1 oz. per gallon using lukewarm tap water. - Fully immerse each device in the prepared detergent and allow them to soak for a minimum of two minutes. - After soaking the devices, scrub them using a soft bristle brush and circular strokes to remove any visible soil. Pay particular attention to all the areas where the soil could be imbedded (i.e. grooves, crevices, lumens, blind holes.) Use a syringe to flush lumens and a pipe cleaner to clean lumens and holes. Perform cleaning under the water surface to limit aerosolization of the cleaning fluid and soil, as well as for worker and environmental safety. - Rinse devices in lukewarm deionized water for a minimum of one minute to remove any detergent residuals. - Prepare a neutral pH enzymatic detergent, Enzol®, in a sonicator as per the manufacturer’s recommendation at 1 oz. per gallon using lukewarm tap water. Fully immerse the devices in the detergent and sonicate for 10 minutes. - After sonication, rinse the devices with lukewarm reverse osmosis/deionized water for one minute. - Dry the devices using a clean cloth.

Automated Cleaning Procedure: - Prepare an enzymatic detergent (Klenzyme®) using lukewarm tap water as per the manufacturer’s recommendation. - Fully immerse the devices and allow to soak for a minimum of two minutes. - Following the soak time, flush any lumens of the device using a syringe. - Rinse the devices under lukewarm running tap water for a minimum of one minute. - Place the devices into an automated washer (Steris 444 or equivalent). The washer cycle parameters are:

Phase Recirculation Time (Min.)

WaterTemperature Detergent

Pre-Wash 1 02:00 Cold Tap Water NA

Enzyme Wash 01:00 Hot Tap WaterKlen-zyme®, 1 oz/gal

Wash 1 02:00 60°CRenu-Klenz™ 1/2 oz/gal

Rinse 1 05:00 Hot Tap Water NA

- Dry the devices with a clean cloth and visually examine to determine if all adherent visible soil has been removed.

Thermal Disinfection Cycle for Automated Cleaning Procedure: Thermal disinfection rinse/cycle should be used at 82.2°C for 1 minute.

9. Product handling:Store implants unopened in their respective protective pack-ages until use. When removing the implant from its pack-aging, observe all relevant aseptic instructions. Protect the prosthesis from contact with objects, which may damage the surface finish. Inspect each implant prior to use and

dispose of implants that exhibit surface or configuration damage.Contouring or clamping of implants should be avoided if possible. It is recommended that implants should not be cut, sharply bent or re-bent, notched, or scratched. These alterations can produce defects or stresses, which may lead to failure of the implant.

10. CautionFederal Law (USA) restricts this device to sale by or on the order of a physician.

©2007 *Patented. Integra and the wave logo are trade-marks of Integra LifeSciences Corporation or its subsidi-aries. ©2007 Integra LifeSciences Corporation. All rights reserved.

82-0003J

Subtalar implant MBA™

Instruction for Use

05-1002 Probe instrument

05-1020 Insertion Device

05-0017Guide Pin 2 mm

Page 8: Integra · integra™ MBa™ Titanium subtalar implant Surgical technique Integra™ MBA™4 PRODUCTS FOR SALE IN EUROPE, MIDDLE-EAST and AFRICA ONLY Surgical Technique 1. Incision

©2011 Integra LifeSciences Corporation� All rights reserved� ILS 08-07-089-02-11 PRODUCTS FOR SALE IN EUROPE, MIDDLE-EAST and AFRICA ONLY�

Integra LifeSciences Services (France) SASSales & Marketing EMEA Immeuble Séquoia 2 97 allée Alexandre Borodine Parc technologique de la Porte des Alpes 69800 Saint Priest FRANCE +33 (0)4 37 47 59 00 fax +33 (0)4 37 47 59 99 emea�info@integralife�com integralife.com

Customer ServicesInternational: +33 (0)4 37 47 59 50 +33 (0)4 37 47 59 25 (Fax) csemea@integralife�comUnited Kingdom: csuk�ortho@integralife�comFrance: +33 (0)4 37 47 59 10 +33 (0)4 37 47 59 29 (Fax) cs-ortho@integralife�comBenelux: +32 (0)2 257 4130 +32 (0)2 253 2466 (Fax) custsvcbenelux@integralife�com Switzerland: +41 (0)2 27 21 23 30 +41 (0)2 27 21 23 99 (Fax) custsvcsuisse@integralife�com

Distributed by

Availability of these products might vary from a given country or region to another, as a result of specific local regulatory approval or clearance requirements for sale in such country or region� ▪ Always refer to the appropriate instructions for use for complete clinical instructions� ▪ Non contractual document� The manufacturer reserves the right, without prior notice, to modify the products in order to improve their quality� ▪ WARNING: Applicable laws restrict these products to sale by or on the order of a physician� ▪ Klenzyme, Enzol and Renuklenz are trademarks or registered trademarks of their respective owners� MBA, Subtalar MBA, Integra and the Integra logo are trademarks of Integra LifeSciences Corporation�

Integra LifeSciences Corp (Integra Ohio) 4900 Charlemar Drive Cincinnati OH 45227 USA

integra™

MBa™ Titanium subtalar implant

Features

• The Subtalar MBA™ implant system has the longest history of successful clinical outcomes�*

• Barrel-shaped implant provides optimal medial support�

• Uniform implant diameter minimizes lateral discomfort�

• Slotted design helps prevent extrusion�

• Simple, minimally-invasive surgical procedure�

* Data available upon request

Component material

• Subtalar implant: Titanium alloy�

1

2

3 6

4

5

Instrument tray

# Reference Description

Instruments

3 05-0016 Guide Pin Insert Holder

4 05-0206 Trial 6 mm

4 05-0208 Trial 8 mm

4 05-0209 Trial 9 mm

4 05-0210 Trial 10 mm

4 05-0212 Trial 12 mm

1 05-1002 Yellow Probe

2 05-1006 6 mm Sizer

2 05-1008 8 mm Sizer

2 05-1009 9 mm Sizer

2 05-1010 10 mm Sizer

2 05-1012 12 mm Sizer

5 05-1020 Insertion Device

Instrument tray

# Reference Description

Implants

05-0106 MBA™ 6 mm Implant

05-0108 MBA™ 8 mm Implant

05-0109 MBA™ 9 mm Implant

05-0110 MBA™ 10 mm Implant

05-0112 MBA™ 12 mm Implant

Disposables