+ All Categories
Home > Documents > INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same...

INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same...

Date post: 21-May-2020
Category:
Upload: others
View: 0 times
Download: 0 times
Share this document with a friend
102
INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN MEASUREMENTS OBTAINED FROM SELECTED CLINICAL LABORATORIES IN KENYA MANDANIA ESTHER WANGUI MASTER OF SCIENCE (Medical Laboratory Sciences) JOMO KENYATTA UNIVERSITY OF AGRICULTURE AND TECHNOLOGY 2016
Transcript
Page 1: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN

MEASUREMENTS OBTAINED FROM SELECTED CLINICAL

LABORATORIES IN KENYA

MANDANIA ESTHER WANGUI

MASTER OF SCIENCE

(Medical Laboratory Sciences)

JOMO KENYATTA UNIVERSITY OF

AGRICULTURE AND TECHNOLOGY

2016

Page 2: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

Inter- Laboratory Variability of Haemoglobin Measurements Obtained From

Selected Clinical Laboratories in Kenya

Mandania Esther Wangui

A Thesis Submitted in Partial Fulfilment for the Degree of Master of Medical

Laboratory Sciences in the Jomo Kenyatta University of Agriculture and

Technology

2016

Page 3: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

i i

DECLARATION

This thesis is my original work and has not been presented for a degree in any other

university.

Signature……………………………….. Date……………………………………

Esther Wangui Mandania

This thesis has been submitted for examination with our approval as university

supervisors.

Signature……………………………….. Date……………………………………

Dr. Juliette R. Ongus, Ph.D.

JKUAT, Kenya

Signature……………………………….. Date……………………………………

Serah N. Kaggia, MMed (Human Pathology)

JKUAT, Kenya

Signature……………………………….. Date……………………………………

Fatmah Abdallah MMed (Human Pathology)

UoN, Kenya

Page 4: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

i i i

DEDICATION

I dedicate this thesis to my family, my brothers and sisters for their moral support and

encouragement during the course of this study.

Page 5: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

iv

ACKNOWLEDGEMENT

My greatest thanksgiving goes to the Almighty God for seeing me through my

Masters programme.

I wish to express my profound gratitude to my supervisors Dr. Juliette Ongus, Dr.

Serah N. Kaggia both of the Department of Medical Laboratory Sciences, COHES,

JKUAT and Dr. Fatmah Abdalla of Unit of Haematology and Blood Transfusion,

School of Medicine, UON for their close supervision and the inspiration that made

this thesis a reality.

I am indebted to Dr. Jane Carter, Head of Clinical Services, AMREF for allowing part

of this research to be conducted in Central laboratory, AMREF, Nairobi and Mr.

Stephene Munene for technical support.

My sincere gratitude also goes to all my lecturers at the Department of Medical

Laboratory Sciences, COHES, JKUAT for their guidance and contribution to this

thesis either directly or indirectly. I am also grateful to Mr. Patrick Chiyo for his

statiscal support.

Finally, my heartfelt gratitude goes to my family and all my friends who gave me

moral support in one way or another.

Page 6: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

v

Table of Contents

DEDICATION ...................................................................................................................................................... III

ACKNOWLEDGEMENT .................................................................................................................................. IV

TABLE OF CONTENTS........................................................ ERROR! BOOKMARK NOT DEFINED.

LIST OF TABLES................................................................................................................................................ IX

LIST OF FIGURES................................................................................................................................................ X

OPERATIONAL DEFINATIONS .................................................................................................................. XI

LIST OF APPENDICES ............................................................................................................................... XVII

LIST OF ABBREVIATIONS AND ACRONYMS................................................................................XVIII

ABSTRACT .......................................................................................................................................................... XX

CHAPTER ONE............................................................................................................................................. 1

INTRODUCTION ......................................................................................................................................... 1

1.1. Background Information ................................................................................................................. 1

1.2. Statement of the Problem................................................................................................................ 3

1.3. Justification.......................................................................................................................................... 4

1.4. Research Questions ........................................................................................................................... 5

1.5. Hypothesis ............................................................................................................................................ 5

1.5.1 Null Hypothesis ............................................................................................................................. 5

1.5.2 Alternative Hypothesis ................................................................................................................ 6

1.6. Objectives............................................................................................................................................. 6

1.6.1. General Objectives .................................................................................................................. 6

1.6.2. Specific Objectives .................................................................................................................. 6

Page 7: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

vi

CHAPTER TWO…………………………………………………………………...7

LITERATURE REVIEW........................................................................................................................... 7

2.2. Variability of Hb measurements................................................................................................... 7

2.1. Impact of Physiologic factors on haemoglobin measurements .......................................... 8

2.1.1. Type of Blood Sample................................................................................................................. 8

2.1.2. Sample Site ..................................................................................................................................... 8

2.1.3. Time ................................................................................................................................................... 9

2.1.4. Body Position ................................................................................................................................. 9

2.2. Laboratory devices and their impact on variability of Hb measurements ....................10

2.3. Accuracy and precision of Hb measurements........................................................................12

2.4. Quality control in haemoglobinometry ....................................................................................14

2.5. External Quality Assessment .......................................................................................................15

CHAPTER THREE ....................................................................................................................................17

MATERIALS AND METHODS ...........................................................................................................17

3.1. Study design ......................................................................................................................................17

3.2. Study site ............................................................................................................................................17

3.3. Study Population..............................................................................................................................17

3.4. Sample size ........................................................................................................................................18

3.4.1. District hospital laboratories and Sub- district hospital laboratories....................19

3.4.2. Private laboratories ................................................................................................................20

3.4.3. Health centre laboratories ...................................................................................................20

3.4.4. Accredited laboratories ........................................................................................................20

3.4.5. National referral laboratories .............................................................................................20

Page 8: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

vii

3.5. Sampling Method ............................................................................................................................21

3.5.1. Inclusion Criteria....................................................................................................................21

3.5.2. Exclusion criteria ...................................................................................................................22

3.6. Laboratory procedures ...................................................................................................................22

3.6.1. Preparation of EQA samples ..............................................................................................22

3.6.2. Preparation of haemolysate ................................................................................................22

3.6.3. Assigning value of Hb concentration to the EQA samples (target values) ........23

3.6.4. Sample handling and storage .............................................................................................24

3.6.5. Sample packaging and transportation .............................................................................24

3.6.6. Sample processing .................................................................................................................25

3.7. Data Management and Analysis .................................................................................................26

3.8. Ethical Considerations ...................................................................................................................26

CHAPTER FOUR ............................................................................................................................................... 28

RESULTS .............................................................................................................................................................. 28

4.1. The performance of participating laboratories in differentiating, low, normal and

high Hb measurements .................................................................................................................................29

4.2. The inter-laboratory variability of the Hb measurements ..................................................30

4.3. Comparison of the variability of Hb measurements among the various Hb

measurement methods/equipment ............................................................................................................33

4.3.1. Variation of Hb measurements due to analysers .........................................................33

4.3.2. Variation of Hb measurements due to methods...........................................................39

CHAPTER FIVE ................................................................................................................................................. 42

DISCUSSION, CONCLUSIONS AND RECOMMENDATIONS.......................................................... 42

5.1. Discussion ................................................................................................................................................... 42

Page 9: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

viii

5.2. Conclusions................................................................................................................................................. 49

5.3. Recommendations................................................................................................................................... 50

REFERENCES ..................................................................................................................................................... 51

APPENDICES...................................................................................................................................................... 63

Page 10: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

ix

LIST OF TABLES

Table 3-1: Categories of Clinical laboratories in Kenya....................................................,....18

Table 3-2: Summary of the sampled laboratories for EQA.....................................................21

Table 4-3: The Different Analyzers and the Number of Laboratories Using Each Analyser

Type.........................................................................................................................................28

Table 4-4: The Number and Percentage of Laboratories that Performed Well (Passed) or

Failed according to CLIA’ 88 Test Performance Criteria........................................................30

Table 4-5: The Inter-Laboratory Variability of Hb Measurements for Each Type of Hb

Analyser...................................................................................................................................32

Table 4-6: The Accuracy of Analysers in estimating Hb values across all levels of

measurement in comparison to the expected or mean of all reference values (6.2, 13.6, and

18.1) combined (i.e. 12.633)....................................................................................................34

Table 4-7: The accuracy of the most commonly used analysers in Kenya in estimating low Hb

value (6.2 g/dl).........................................................................................................................36

Table 4-8: The accuracy of the most commonly used analysers in Kenya in estimating normal

Hb value (13.6 g/dl).................................................................................................................37

Table 4-9: The accuracy of the most commonly used analysers in Kenya in estimating high

Hb value (18.1 g/dl).................................................................................................................38

Table 4-10: The resuts for the questionnaire...........................................................................41

Page 11: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

x

LIST OF FIGURES

Figure 4-1: Coefficient of variation as a measure of inter-laboratory variation for the three Hb

test concentrations...................................................................................................................31

Figure 4-2: Percentage number of laboratories as per each method of equipment /analyser

operation..................................................................................................................................39

Figure 4-3: Coefficient of variation around the standard reference as a measure of reliability

of the different methods of Hb analyser operation...................................................................40

Page 12: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

xi

OPERATIONAL DEFINATIONS

Accuracy---------

Allowable Total

Error(TAE)-----------

Analytic Variability----

Bias( inaccuracy)------

Calibration-------------

Is the degree of closeness of measurements of a quantity to

that quantity’s actual (true) value.

A quality requirement that sets a limit for combined

imprecision (random error) and bias (inaccuracy, or

systematic error) that are tolerable in a single measurement

or single test result to insure clinical usefulness.

Random error in a laboratory result whose magnitude

depends on the methodology or analyser used.

Bias is the difference (generally unknown) between a

laboratory's average value (over time) for a test item and

the average that would be achieved by the reference

laboratory if it undertook the same measurements on the

same test item.

It’s a process of testing and adjusting an instrument or test

system, to establish a correlation between the measurement

response and the concentration of the substance that is

being measured by the test procedure.

Page 13: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

xii

Coefficient of

variation---------------

Commutability-------

Deviation index---------

External Quality

Assurance (EQA)------

A measure of variability or diversity associated with

random error or imprecision. SD shows how much

variation or dispersion there is from the mean (average or

other Expected value) during repeated measures. A small

SD indicates that data points tend to be very close to the

mean, whereas a large SD indicates that the data points are

spread over a wide range of values. SD is the square root of

a dataset’s variance.

Degree to which a material yields the same numerical

relationships between results of measurements by a given

set of measurement procedures, purporting to measure the

same quantity, as those between the expectations of the

relationships for the same procedures applied to those

types of material for which the procedures are intended.

A measure of how far a result differs from the mean value

as multiples of the standard deviation.

A programme which determines total testing performance

by comparing a laboratory or clinic’s test result (including

interpretation of results) to a known standard or to an

appropriate peer group mean generated from an inter-

Page 14: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

xii i

Imprecision(random

error) ------------

Inter-laboratory

Comparability-----------

Inter-Individual

Biologic Variation-----

Inter-laboratory

variability------------

laboratory comparison in which multiple laboratories

measure the same sample using the same test methods,

reagents and controls.

Lack of repeatability or reproducibility of the same result;

represented by the standard deviation (in units of the test)

or coefficient of variation, (in units of percent).

The organisation, performance and evaluation of

calibration/test results for the same or similar item by two

or more laboratories in accordance with predetermined

conditions

Variation in the mean values when a quantity is measured

in specimens from different individuals due to differences

in homeostatic set point of the individuals

It’s the variability of test results from different laboratories

using the same test method and analysing the same test

material. It includes both within-laboratory and between-

laboratory components of variance. Inter-laboratory

coefficients of variation (CVs) are CVs calculated based on

the total variance of results for a given method, endpoint,

Page 15: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

xiv

Internal

Quality Control---------

Intra-Individual

Biologic Variation-----

Intra-laboratory

Variability-----------

and sample type.

It involves the in-house procedures for continuous

monitoring of operations and systematic day-to-day

checking of the produced data to decide whether these are

reliable enough to be released. The procedures primarily

monitor the bias of data with the help of control samples

and the precision by means of duplicate analyses of test

samples and/or of control samples.

Variation in results when a quantity is measured in

different specimens from the same individual obtained over

a time span due to the imprecision of the measurement

procedure (metrological variability) as well as to the

rhythmic and random fluctuations of the quantity value

around a virtual homeostatic set point.

Is the variability of test results from the same laboratory

using the same test method and analysing the same test

material. Within-laboratory coefficients of variation (CVs)

are CVs calculated based on solely the within-laboratory

component of variance.

Page 16: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

xv

Intra-Subject

Variability ------------

Matrix-------------------

Matrix bias-------------

Precision-------------

Quality

Control (QC)----------

Includes differences in test results due to both subject

fluctuations and test performance fluctuations.

The matrix of the specimen is defined as the totality of

components of a material system except the analyte.

The component of the observed difference due to non-

commutability between a method/material combination.

Is the closeness of agreement between independent test

results obtained under stipulated conditions.

Procedures which monitor analytical performance of

instruments and detect analytical error. QC typically refers

to use of quality control materials and analysis of resulting

control data

Repeatability------------

Degree of consensus between successive measurements

which have been done on the same sample with very

similar conditions (same analyser, same user, same

laboratory, same methods, same lot of reagents) in a very

short time (e.g. same day). Often is symbolized as sr.

Page 17: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

xvi

Reproducibility---------

Standard

Deviation (SD)----------

Trueness------------------

Is the degree of consensus between successive

measurements achieved on the same sample with different

conditions (e.g. different analyzer, different user, different

lot of reagents) in a long time. Can be either intra-

laboratory or inter-laboratory, and is symbolized as sR.

A measurement of imprecision (random error), biologic

variation, or other variability in a population;

mathematically, CV is standard deviation divided by the

mean and expressed as a percentage.

Closeness of agreement between the average value

obtained from a large set of test results and an accepted

reference value

Page 18: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

xvii

LIST OF APPENDICES

APPENDIX I: Procedure for Haemolysate Preparation..........................................56

APPENDIX II: Standard Operating Procedure for Triple Packaging

System………………………………………………………………………….…..58

APPENDIX III: Instructions for the Participating Laboratories................................60

APPENDIX IV: Questionnaire...................................................................................62

APPENDIX V: Worksheet..........................................................................................65

APPENDIX VI: Ethical Clearance.............................................................................73

APPENDIX VII: Informed Consent Form for Laboratory Management...................68

Page 19: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

xvii i

LIST OF ABBREVIATIONS AND ACRONYMS

AMREF African Medical Research Foundation

ANOVA Analysis Of Variance

BA Bland Altman

BGAs Blood gas analysers

CBC Complete blood count

CLIA Clinical Laboratory Improvement Amendments

CPD continuing professional development

CV Coefficient of Variation

DI Deviation Index

EA-REQAS East African Regional External Quality Assessment Scheme

EDTA Ethylene diamine tetra acetic acid

EQA External Quality Assurance

EQAS External Quality Assessment Schemes

ERC Ethical Review Committee

HA Haematology Analysers

Hb Haemoglobin

HICN Hemiglobincyanide

ICSH The International Committee for Standardisation in Haematology

IQA Internal Quality Assurance

IQR Inter-quartile Range

ISO International Organization for Standardization

KENAS Kenya National Assurance Schemes

Page 20: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

xix

Km Kilometre

KMLTTB Kenya Medical Laboratory Technicians and Technologists Board

KNBTS Kenya National Blood Transfusion Service

KNH Kenyatta National Hospital

Lab Laboratory

mL Millilitre

NaCl Sodium Chloride

NIQR Normalized Inter-quartile Range

PBS Peripheral blood smear

PCV Packed Cell Volume

POC Point- of- care

PT Proficiency Testing

RBC Red blood cells

RCC Regional Co-ordinating Centre

SD Standard Deviation

SOPs Standard Operating Procedures

TLC Total leukocyte count

tHb Total Haemoglobin

UON University of Nairobi

WBC White blood cells

WHO World Health Organisation

Page 21: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

xx

ABSTRACT

Accurate determination of haemoglobin (Hb) concentration is a common element in

assessing the extent of anaemia and an important variable in directing transfusion therapy in patients. Laboratory measurements should be reproducible and consistent

from day to day as well as between laboratories so that comparable results will be obtained when a given specimen is tested in different laboratories. Variability in Hb measurements is caused by many factors such as laboratory error, numerous

physiologic, temporal and methodologic factors. Therefore this study was aimed at determining the interlaboratory variability of Hb measurements obtained from

selected clinical laboratories across Kenya. A total of 188 public and 105 private laboratories selected from a total of 21 out of 47 Kenyan Counties received three specimens with low (A), normal (B) and high (C) Hb concentrations for analysis, after

which their results were compared and evaluated for accuracy. Laboratory performance was assessed using the Clinical Laboratory Improvement Amendments

of 1988 (CLIA’88) criteria. Coefficient of Variation (CV) was calculated as a measure of interlaboratory variability while the accuracy of the analysers was evaluated using one-way analysis of variance (ANOVA). A total of 67.98%, 64% and 50.6%

laboratories gave accurate results for samples A, B and C respectively. The results generated by the Celltac, Humalyzer Junior, Medonic, Mindray, Colourimeter,

Hemocontrol and Sysmex analysers were not significantly different (P>0.05) from the reference values. However, the Diaspect and Sahli analysers underestimated the Hb readings, while Hemocue, Urit and Mission overestimated the Hb readings (P<0.05).

Interlaboratory variation of 33.3%, 25.1% and 29.4% for samples A, B and C was recorded irrespective of the analyser a laboratory used. The CV for the automated,

semi-automated and manual methods was 7.08%, 7.04% and 34.26% respectively. The interlaboratory variation in Hb measurements resulted from variation in methodologies and types of analysers. Regular participation in External Quality

Assessment Schemes (EQAS) is essential in order to achieve interlaboratory comparability of Hb results. Laboratories should embrace automation and gradually

replace manual with automated methods, which are more accurate and reliable.

Page 22: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

1

CHAPTER ONE

INTRODUCTION

1.1. Background Information

Accurate determination of Hb concentration is a common element in assessing the

extent of anaemia and an important variable in directing transfusion therapy in

patients. This decision should be made based on accurate, reliable and timely

laboratory tests. The detection of anaemia based on Hb concentration in venous or

capillary blood is used to estimate the prevalence of anaemia in populations, allocate

resources to programmes, and target intervention programmes to vulnerable groups;

it is also used to screen individuals for participation in programmes and to evaluate

response to interventions (Morris et al., 1999). The Hb test is used to indirectly asses

the oxygen carrying capacity of blood thus making it an important aid in

establishing the presence of anaemia and treating anaemia. The measurement of

blood Hb is one of the most common routine clinical laboratory tests. Hb

determination may be requested as an individual test or as part of a complete blood

count (CBC), which may be performed using either capillary or venous blood. The

Hb test is precise, simple to perform, easily standardised and may be performed

either manually or by using automated Hb analysers (Barbara, Anna & Norma,

2000).

The gold standard for assessing Hb concentration is the direct cyanmethaemoglobin

method (Burger and Pierre-Louis, 2003) and the International Committee for

Standardisation in Haematology (ICSH) recommends the Drabkin’s method as the

Page 23: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

2

standard method for determining the Hb concentration of whole blood (WHO,

2008). Lately, new continuous, non-invasive methods of measuring Hb have been

introduced into the clinical environment whose accuracy may vary substantially

from invasive Hb measurements. This variability in measurements is caused by

many factors such as laboratory error, numerous physiologic, temporal and

methodologic factors (Lauren, 2013).

Despite the variability of Hb measurements, clinicians rely on accurate Hb values to

make decisions on transfusion and management of patients. Previous studies

conducted in other parts of the world have shown variation of Hb estimation

between laboratories (Blerk et al., 2007; Bilto, 1999; Renu et al., 2007).

The World Health Organisation (WHO) EQA for haematology assesses the

participating laboratories to correctly quantify the Hb level and number of white

blood cells and platelets (World Health Organisation, 2007). Patient Safety

Monitoring in International Laboratories (JHU-SMILE) remotely monitors

approximately 165 international laboratories and clinics in 22 countries, 84 of which

are in sub-saharan Africa. It is well known that laboratories in sub-saharan African

countries face many challenges to provision of quality results (WHO, 2008; Frean et

al., 2012). A 40 month study of the EQA performance of the Sub-Saharan African

JHU-SMILE laboratories reported that there was 0.6% failure rate for haemoglobin

in that cohort of laboratories (Amukele et al., 2012).

In East Africa, AMREF offers a number of programmes to address laboratory

quality, including continuing professional development, competency assessments,

blinded re-checking of test materials, regular structured support supervision and an

Page 24: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

3

East African Regional External Quality Assessment Scheme (EA-REQAS). EA-

REQAS was established in 2003 when the four Ministries of Health of East Africa

(Kenya, Uganda, Mainland Tanzania, and Zanzibar) agreed to share laboratory

standards and materials; develop standard documents targeting both clinicians and

laboratory staff; establish national quality assurance bodies; form an EA-REQAS

Committee to oversee activities; and appoint AMREF as the Regional Co-ordinating

Centre. Seven surveys have been submitted since 2007, with nearly 400

participating facilities currently. Reference laboratories were identified to produce

quality PT (Proficiency Testing) materials. There has been progressive improvement

in performance by a few facilities participating in all surveys and national policy on

laboratory methodologies has been influenced (3rd Regional Technical Meeting EA-

REQAS, 2010)

1.2. Statement of the Problem

Although quality assurance programmes play an integral part in clinical laboratories

of developed countries, these programmes have not been accorded the same degree

of importance in the laboratories of developing countries. In Kenya in particular,

only few laboratories have participated in such schemes implying that the Hb

measurements of majority of clinical laboratories may not be accurate and

comparable. In addition, the variability of the Hb measurements between clinical

laboratories of Kenya has not been quantified. In order to address this problem, this

study sought to asses the extent of inter-laboratory variations in Hb estimation in

Kenyan clinical laboratories and to determine if the discrepancies are potentially

large enough to affect decisions consequently impacting on patient treatment and

Page 25: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

4

management. Further, the study sought to evaluate the accuracy of the most

commonly used Hb analysers in Kenyan clinical laboratories.

1.3. Justification

Systematic error or bias may affect the measurement of a sample such that the result

obtained is not always a perfect value but may be some distance from the true value

(Kaplan, Pesce, & Kazmierczak, 2003). Use of different methodologies, reagents

and assay conditions (e.g. temperature, reagent concentrations, detection methods,

wash steps etc) are the reasons for the poor agreement between analysers. To

quantify these differences, methods need to be validated using reference

methodologies to ensure accuracy and comparability of results across laboratories

(Fierdoz, 2012).

Laboratories must continuously strive to maintain and improve the quality of

laboratory results, which is a never-ending process. Using statistical methods,

laboratories can continually improve their performance and the reliability of patient

results as well. Quality assurance in haematology is intended to ensure the reliability

of the laboratory results. A quality assurance programme has two main aspects,

namely, internal quality control (IQC) and external quality assessment (EQA) (Sah,

Raj & Prakash, 1999). IQC and EQA play very crucial roles in ensuring reliability

of analytical results (Westgard, 2008). EQAS of clinical laboratories ensures

between-laboratory comparability of results as well as detecting bias (systematic

errors) and overall review on the IQC programme (Rippey & Williamson, 1988;

Savage 1989).

Poor selection of techniques, lack of essential equipment, lack of quality control

materials and quality assurance systems, personnel issues and shortages of supplies

Page 26: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

5

are some of the challenges to providing quality services faced by laboratories in

developing countries (Carter et al., 2002). Significant exchange of information in

patient care, teaching and research is becoming more and more difficult due to the

steadily increasing variation in methods and equipment (Munich, 1982). It is

therefore important that interlaboratory variability of Hb measurements and

reliability of equipment/analysers and methods used for Hb measurements in

Kenyan clinical laboratories be assessed to enhance quality. Despite the existence of

the EQA programmes in Kenya, however, laboratory participation is still very low

and optional. Therefore, it is essential that studies such as this present one are

frequently carried out to help in the improvement of interlaboratory comparison.

1.4. Research Questions

1. Can the laboratories participating in EQA of Hb measurements accurately

differentiate normal, low and high Hb specimen?

2. Is the inter-laboratory variability of the Hb measurements in an external quality

assessment quantifiable?

3. Can the variability of Hb measurements among the various Hb measurement

methods/equipment be compared?

1.5. Hypothesis

1.5.1 Null Hypothesis

There is no difference in Hb measurements obtained from selected clinical

laboratories in Kenya.

Page 27: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

6

1.5.2 Alternative Hypothesis

There is difference in Hb measurements obtained from selected clinical laboratories

in Kenya.

1.6. Objectives

1.6.1. General Objectives

To determine the inter-laboratory variability of Hb measurements obtained from

selected clinical laboratories in Kenya.

1.6.2. Specific Objectives

1. To determine the ability of participating laboratories to accurately differentiate

normal, low and high Hb specimens.

2. To quantify the inter-laboratory variability of the Hb measurements.

3. To compare the variability of Hb measurements among the various Hb

measurement methods/equipment.

Page 28: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

7

CHAPTER TWO

LITERATURE REVIEW

2.2. Variability of Hb measurements

Both accuracy (how close the measurement is to the actual value) and precision

(how repeatable the measurement is) of “standard” laboratory measurements are

subject to numerous methodologic factors that affect them (Lauren, 2013). Various

methods of Hb determination have been used all through ((Barbara, Anna & Norma,

2000). In the 1950s numerous different methods were in use leading to variability of

the results. This led to several attempts being made to standardise the measurement

of the Hb concentration in human blood. The Haemiglobincyanide method was well

accepted, soon came into general use, and performed satisfactorily for many years.

Later, with automation of laboratory methods in laboratory medicine, various

methods came up. This led to the manual Haemiglobincyanide method being phased

out as a routine method and became the reference method with which the current

methods should agree (Zigilstra, 1997). Measurement of total haemoglobin (tHb)

concentration by an automated laboratory analyser or point-of-care (POC) analyser,

using a venous or capillary blood sample is the conventional method. ICSH has

recommended the Drabkin’s method as the standard method for determining the Hb

concentration of whole blood (Chen et al., 1992).

The clinical measurement of tHb has inherent variability. Inter-device variability is

one cause of variability, for example, CO-Oximeter and POC devices that are

commonly used to measure Hb have been shown to vary up to ±1.2 and ±1.3 g/dL,

Page 29: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

8

respectively. Additionally, there are a variety of physiologic and methodologic

factors that can significantly influence Hb levels in the body (Masimo, 2009).

2.1. Impact of Physiologic factors on haemoglobin measurements

A variety of factors influence Hb measurement and Hb levels in the body; these

include both the laboratory devices and physiologic factors. Sources of Hb variation

within the body include, the type of blood sample, sample site, time the sample is

taken, and body position.low Hb concentration readings: False high Hb

concentration readings:

2.1.1. Type of Blood Sample

Laboratory devices are designed to allow sampling of either venous or arterial

blood. Hb measurement may vary based on whether arterial or venous blood is used.

arterial Hb measurements can be expected to be, on average, 0.7 – 1.0 g/dL less than

the Hb measurements derived from venous blood (Mokken et al., 1996 & Yang ZW

et al., 2001) The percentage of plasma concentration can vary from the arterial to

venous blood based on a variety of physiologic factors, despite the total amount of

circulating red blood cells and Hb remaining relatively constant whether in arterial

or venous blood. The amount of plasma concentration can be higher in arterial

blood, potentially leading to lower concentration of Hb (Masimo, 2009)

2.1.2. Sample Site

The site on the body from where blood is drawn can also affect Hb measurements.

Large discrepancies were found between the values obtained from capillary blood

samples from the left and right hands of the same women, with intrasubject standard

deviation of 0.8 g/dL and correlation of 0.7 (Morris et al., 1999). The wide limits of

Page 30: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

9

agreement indicate that two samples from different fingers of the same person could

have Hb concentrations that differ by up to 2.0g/dL. Another study shows wide

variation in the Hb concentration of capillary blood samples obtained from different

fingers on the same individual at the same time. Intrapatient variability ranged as

high as 7% (Bouton et al., 1994)

2.1.3. Time

Even in stable patients, Hb measurement can vary significantly over time. In a study

of venous blood samples drawn from the same individuals on two different

occasions, within person variances could vary as much as 2.6 g/dL in males and 2.3

g/dL in females (Looker AC, et al., 1990 & Burger et al., 2004). In another study,

when Hb measurements were taken from the same individual on four (4) different

days consecutively, intrasubject variability was 7.0% and the standard deviation was

0.8 g/dL ((Bouton et al., 1994)

2.1.4. Body Position

The position of the body before and during the blood draw also affects Hb

measurements due to the normal composition of blood, interstitial fluid shifts, and

elevations of protein and white blood cells. Body position has a significant effect on

venous Hb measurements due to decreases in plasma volume on assuming an

upright position. Heart rate and blood pressure are higher when standing vs. sitting,

which induces the movement of intravascular fluid such as plasma into interstitial

compartments. This causes Hct and Hb levels to rise (haemoconcentration) and

plasma volume to decrease (Martin et al., 1997). Gore and colleagues showed a 6%

Page 31: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

10

reduction on plasma volume with standing, which changed Hb up to 2 g/dL (Gore

CJ, et al., 1992). Moving from seated to standing positions for 20 minutes may

result in a change in Hb concentration by >1.0 g/dL (Daniel-Johnson et al., 2007).

The converse is also true, indicating that if patients who are ambulatory change

body position prior to the blood draw, they may require a period of equilibration.

2.2. Laboratory devices and their impact on variability of Hb measurements

Historically venous, arterial, or capillary blood samples have been used for Hb

measurement; POC devices that use capillary blood samples. However, recently new

continuous, non-invasive methods of measuring Hb have been introduced into the

clinical environment whose accuracy can be variable compared to laboratory Hb

measures.

As clinicians interpret laboratory measurements, they expect that the values would

not change significantly if consecutive samples were measured repeatedly on the

same laboratory device or on different laboratory devices (Lauren, 2013). According

to the International Organization for Standardization (ISO), laboratory error is

“failure of a planned action to be completed as intended, or use of a wrong plan to

achieve an aim, occurring at any part of the laboratory cycle, from ordering,

examinations to reporting results and appropriately interpreting and reacting to

them” (Plebani & Lippi, 2010).

Care providers should take responsibility from the time of receipt of a request for a

pathology test/investigation, to the time outcomes are communicated to the requester

(The Royal College of Pathologists of Australasia, 2010). Therefore efforts to

Page 32: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

11

reduce errors should consider the total testing procedure and quality improvement

should be measurable in terms of improvement to patient safety and clinical care

outcome (Plebani, 2010; Plebani, 2009). The chance of inappropriate care due to

laboratory errors ranges from 6.4 % – 12 %, although up to 30 % of errors also

resulted in patient discomfort, an escalation of costs and unwarranted additional

testing (Lippi & Guidi, 2007). Laboratory error, physiologic, temporal and

methodologic factors are the causes of the variability in reported Haemoglobin

values (Lauren, 2013). Multiple variables, which include device calibration, sample

handling, and other sources of variation specific to the technology, affect the

accuracy of each of the different methodologies used for total Hb measurement

(Lauren, Stephanie & Erin, 2011).

Both intra-device and inter-device variability affect total Hb measurements. Intra-

device comparison is Hb variability from the same blood sample on the same device

while inter-device comparison is Hb variability from the same sample on different

devices (Gehring et al., 2007). Previous studies have shown that there is significant

inter-device and intra-device variation in the Hb measurements; in 2007, five (5)

different models of CO-Oximeters were used to evaluate both inter-device and intra-

device variation in Hb measurements. When the same blood sample was analysed on

2 identical devices to test intra-device variability, the standard deviation between

measurements ranged from 0.2 to 1.2 g/dL (Gehring et al., 2007). Both reference

devices and test devices produce and/or contain inherent errors (Bland and Altman,

1986). The tHb measurements using a venous sample can vary as much as 0.9 g/dL

Page 33: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

12

between different laboratory analysers across the normal tHb measurement range,

using a reference calibrator (RNA Medical, 2011).

Despite the use of POC haematology analysers becoming more frequent in the last

decade it has been shown that POC devices for Hb have reduced accuracy compared

to laboratory devices. Device methods, size of the blood sample, and strong

potential for confounding elements with capillary blood are factors that affect POC

device accuracy. At normal Hb levels of 13-15 g/ dL, the Clinical Laboratory

Improvement Amendments (CLIA) specification variance is approximately 1.0 g/dL

while in the anaemic range of 10 g/dL, the target variance is 0.7g/ Dl (Masimo,

2009). Previous studies reveal a significantly larger difference in haemoglobin

measurement between POC and laboratory devices; Hb measurement from capillary

blood in POC devices varies between 0.5 – 2.3 g/dL from reference standards

(Gomez-Simon et al., 2007; Patel et al., 2007; De Louw et al., 2007; Argawal &

Heinz, 2001).

2.3. Accuracy and precision of Hb measurements

Inter-laboratory imprecision as assessed by proficiency testing comprises

components of intra-laboratory imprecision and, more importantly, inter-laboratory

dissimilarities. Greater precision of techniques usually indicates a higher degree of

system stability among sites. Performance specifications for dispersion of results

and limits of acceptability are usually based on a percentage difference and /or

absolute difference from the target value (Rej & Jenny, 1992).

Page 34: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

13

The Clinical Laboratory Improvement Act of 1988 (CLIA, 1988) define fixed limit

goals in absolute terms or multiples of standard deviations for a particular analyte.

According to CLIA 88 proficiency testing criteria, acceptable analytical

performance for haemoglobin is target ± 7% (CLIA, 1988).

The desirable specifications for within-subject biological variation, between-subject

biological variation, Allowable Total Error, Imprecision, and Bias (inaccuracy), for

Hb derived from Intra- and Inter-Individual Biologic Variation are 2.8%, 6.6%,

4.1%, 1.4% and 1.8% respectively (Ricos et al., 1999).

Many studies that have been conducted in many parts of the world have quantified

the imprecision, and accuracy (bias) of Hb measurements using the standard

deviation, coefficient of variation and the deviation index. The EQAS in

haematology at National Institute of Health, Islamabad, Pakistan (IEQAS)

conducted a survey to assess individual values against consensus value (mean ±SD)

and deviation index (DI) from the mean, whereas coefficients of variation (CV)

were calculated for years 1996 to 2006. The results were expressed as percentage of

accurate versus inaccurate results, DI and CV. The laboratory achieved 87.74% of

values within acceptable limits for Hb, 72.03% for white blood count, 69.49% for

platelet and 77.03% for reticulocyte estimation. These results were satisfactory,

having DI values less than 3 for all four parameters. Results were varied among

individual components of the survey but the overall DI values lay fairly well within

the acceptable limits with a majority of the results having a DI value less than 3.

Those results having DI values less than 3 were further classified into three

categories; excellent (DI > -2.0 to < 1.0), satisfactory (DI 1.0 to < 2.0) and out of

acceptable limits (DI < -2.0 or > 2.0). Overall, the laboratory achieved 64.22 % of

Page 35: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

14

values with excellent results, 15.24% of results were within acceptable limits and

graded as satisfactory whereas 20.52% of values showed a high deviation from the

reference range. Mean DI values ± standard deviation and CV for Hb, were

calculated for each year. DI + standard deviation (SD) for Hb ranged between -

0.6±0.7 to 2.4±1.3.The study concluded that Participation in EQAS is extremely

beneficial for the improvement of laboratory performance and quality of care

(Birjees et al., 2009).

2.4. Quality control in haemoglobinometry

Regular calibration of spectrophotometers and Hb analysers used for Hb assays, as

specified by the manufacturer, must be done in order to achieve reliable Hb values.

Daily run of appropriate control solutions or when patient samples are run, recording

of control results and maintenance of records is a requirement. In addition each

instrument used must have its own particular checks that must be performed

(Barbara, Anna & Norma, 2000). Use of calibration specimens and high quality

control Hb solutions; checking and calibration of equipment are crucial for proper

determination of Hb in the blood. The results of Hb determination in control

solutions obtained in autoanalysers and manually, are comparable (Pupkova et al.,

2002).

Since 1996 the Hb standard (lot number BK-5182-2, later renamed lot 19-1-B518A)

has been in use till in 2008 when a new lot of the haemiglobincyanide (HiCN (Fe),

HiCN) standard was (#19-1-B806) was released by the ICSH in conjunction with

Eurotrol, B.V. The Haemoglobincyanide standard is used for the standardisation and

calibration of whole blood Hb measurements on most haemoglobinometers and

Page 36: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

15

automated blood cell counters. This new lot was produced following the same

methodology previously specified by ICSH. Later both the WHO Expert Committee

on Biological Standardisation (International Reference Reagent 98/708) and the

European Community Bureau of Reference endorsed this lot as an international

reference material for Hb (Davis & Jungerius, 2010).

2.5. External Quality Assessment

Currently, EQAS exist in the field of laboratory medicine in many countries. Most

of these are intended to assist individual laboratories to continuously monitor their

performance and to compare it with that of other laboratories, whereas others may

be mainly intended for accreditation or licensing purposes. Additionally, EQAS may

monitor the quality of the commercial analytical systems, reagents and test kits, and

they help manufacturers to achieve a better harmonization of results from the

different analytical techniques (Ricos et al., 1996).

The two main aims of EQAS are to set both the target values and limits for

acceptance.

Reference methods are used to assign the target values, but since only a few

schemes follow these principles, target values are derived from the statics of each

survey (Ricos et al., 1996). Continual participation in EQA is an effective means for

identifying and mitigating variables that influence the reliability of analytical assays

for predictive markers, thereby assisting in technical validation and standardization.

The concept of EQA for the national health laboratories network is useful, as it

identifies problems in the comparability of laboratory results and initiates a process

Page 37: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

16

towards solving these problems thus improving the quality of service at the level of

each individual laboratory and the network level (Olafsdottir et al., 1994).

The implementation of a quality assurance policy in a developing country requires a

commitment from the government, the professional societies and the laboratory

workers. It is important to recognize that a policy towards improving health care

should include an external surveillance system for health laboratories. An EQA

scheme will have the greatest impact when it is linked to a quality assurance

programme, which also includes internal quality control as an equally important

component (Deom et al., 1999).

Page 38: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

17

CHAPTER THREE

MATERIALS AND METHODS

3.1. Study design

This was a cross-sectional study using inter-laboratory survey to assess the extent of

inter-laboratory variability of Hb assays in the clinical laboratories of Kenya. The

study lasted a period of 6 months (Between February and July 2016).

3.2. Study site

This study covered all the different laboratory categories which included national

reference laboratories, accredited laboratories, district & sub-district laboratories,

health centre laboratories and private hospital laboratories (level A, B, C, D and E

respectively). The study covered a total of 21 out of 47 counties in Kenya namely;

Bomet, Embu, Isiolo, Kajiado, Kericho, Kiambu, Kirinyaga, Kisii, Kitui, Laikipia,

Machakos, Makueni, Meru, Muranga, Nairobi, Nakuru, Narok, Nyandarua, Nyeri,

Tharaka Nithi and Uasin Gishu. The main study site was the Central Laboratory,

African Medical and Research Foundation, Kenya Located in Nairobi, from where

the samples were prepared and distributed to all the participating laboratories.

3.3. Study Population

The study population consisted of both Public and private clinical laboratories. All

laboratory categories, that is, national reference laboratories, accredited, district &

sub-district, health centres and private hospital laboratories were considered. A total

of 827 privately-owned laboratories, 2 national reference laboratories, 4 accredited

Page 39: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

18

laboratories, 136 district hospitals, 130 sub-district and 1040 health centre

laboratories formed the study population from which the sample size was calculated.

3.4. Sample size

The sample size was calculated, as shown below, for the different levels of

laboratories using the population for each level. The number of laboratories

countrywide in each laboratory category represents the population from which a

representative sample was drawn. The following are the populations for the different

laboratory groups as indicated in the master facility list and as reported by the

KMLTTB (Kisabei, 2013, pers. comm).

Table 3-1: Categories of Clinical laboratories in Kenya

CATEGORY OF LABORATORY NUMBER

National reference laboratories 2

Accredited laboratories 4

District hospital laboratories 136

Sub- district hospital laboratories 130

Health centre laboratories 1,040

Private laboratories (Includes mission hospital laboratories) 827

Using Cochran’s sample size formula for continuous data (Cochran, 1977);

t)2 * (s)2

no= ----------------- (d)2

Where t = value for selected alpha level of 0.05(.025 in each tail) = 1.96

Where s = estimate of standard deviation in the population = 2.333

Page 40: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

19

(Estimate of variance deviation for 14 point scale calculated by using 14 [inclusive

range of scale] divided by 6 [number of standard deviations that include almost all

(approximately 98%) of the possible values in the range]).

Where d = acceptable margin of error for mean being estimated = 0.42 (Number of

points on primary scale * acceptable margin of error; points on primary scale = 14;

acceptable margin of error = 0.03 [error researcher is willing to accept]). Therefore d

= 14 * 0.03 = 0.42.

3.4.1. District hospital laboratories and Sub- district hospital

laboratories

Population = 136+130 =266 laboratories

t)2 * (s)2

no= ----------------- = (1.96)2 (2.333)2 (d)2 14 *0.03)2

=118

However, since this sample size exceeds 5% of the population (266 *.05=13.3),

Cochran’s (1977) correction formula was used to calculate the final sample size.

These calculations are as follows:

no (118) 118 n i = ---------------------------- = -------------------- = 81 (1 + no / Population) (1 + 118/266)

Where n0 = required return sample size according to Cochran’s formula= 118.

Where n1 = required return sample size because sample > 5% of population.

Sample size = 81(40 district and 41 sub-district hospital laboratories)

Page 41: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

20

3.4.2. Private laboratories

Population -------------------------------------827

t)2 * (s)2

no= ----------------- = (1.96)2 (2.333)2 (d)2 14 *0.03)2

=118

Correction formula:

no (118) 118 n= --------------------------------------------- = ----------------------------- = 103

(1 + no / Population) (1 + 118/827)

3.4.3. Health centre laboratories

Population----------------------------1,040

t)2 * (s)2 no= ----------------- = (1.96)2 (2.333)2

(d)2 14 *0.03)2

=118

Correction formula:

no (118) 118 n= --------------------------------------------- = ----------------------------- = 105 (1 + no / Population) (1 + 118/1040)

3.4.4. Accredited laboratories

Population------------------4

Sample size-----------------3

3.4.5. National referral laboratories

Population ---------------2

Page 42: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

21

Sample size--------------1

The total sample size selected was … 81+103+105+3+1 =293 laboratories

Table 3-2: Summary of the sampled laboratories for EQA

CATEGORY OF LABORATORY NUMBER

National reference laboratories 1

Accredited laboratories 3

District hospital laboratories 40

Sub- district hospital laboratories 41

Health centre laboratories 105

Private laboratories 103

TOTAL 293

3.5. Sampling Method

Cluster random sampling of the laboratories from the 21 counties was used. To

arrive at the laboratory, a two-stage cluster sampling method was used. The counties

represented the clusters from which the laboratories were randomly selected. From

each county, a minimum of 7 public and 5 privately owned laboratories were

sampled. All laboratory categories were represented in all the counties except for

national reference and accredited laboratories which were sampled from the counties

they are located. Laboratories were selected from the rural, semi urban and urban

settings.

3.5.1. Inclusion Criteria

Both public and privately-owned laboratories that perform Hb measurements as one

of their routine laboratory tests and gave free informed consent for participation

were included in the study. All methods of Hb determination being used in the

laboratories were applicable.

Page 43: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

22

3.5.2. Exclusion criteria

Private laboratories, which are not registered by KMLTTB (Kenya Medical

Laboratory Technicians and Technologists Board), and all laboratories that are

further than 315 kilometres (km) from the Central Laboratory, AMREF, Kenya were

excluded from the study because of logistical challenges.

3.6. Laboratory procedures

3.6.1. Preparation of EQA samples

EQA samples with low, normal and high Hb values (samples A, B and C)

respectively were prepared at the Central Laboratory AMREF, Nairobi (EAREQAS,

operating under African Medical and Research Foundation, Kenya).

3.6.2. Preparation of haemolysate

Three EQA samples with low, normal and high Hb values (samples A, B and C)

respectively were prepared as follows:-

One pint (450 ml) of blood in a blood bag which had tested negative for Human

immunodeficiency virus, hepatitis B virus surface antigen, syphilis and hepatitis C

virus antibodies was obtained from the Kenya National Blood Transfusion Service

(KNBTS). Using a centrifuge tube, 40 ml of the blood was centrifuged to separate

plasma and Buffy coat aseptically. To the red blood cells deposit, 2-3 fold volume of

physiological saline (9 g/L Sodium Chloride)(NaCl) was added, mixed well and

centrifuged at 2000g for 5 minutes. The supernatant and any remaining buffy coat

were discarded completely. This saline wash was repeated 2 times to ensure

complete removal of plasma, white cells and platelets. To the washed cells, half its

Page 44: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

23

volume of carbon tetrachloride (neat), 99%, was added and the mixture shaken

vigorously in a mechanical shaker (Vibrofix VF1, Janke and Kunkel Model) at

2500g/minute for one hour. The mixture was then stored at 4oC overnight thus

forming a semi-solid interface of lipid/cell debris between carbon tetrachloride and

lysate. The mixture was centrifuged at 2500g for 20 minutes and the upper lysate

layer was carefully pipetted out into a clean Winchester bottle. Sterility and stability

of the haemolysate was maintained by the addition of preservatives and broad-

spectrum antibiotics as follows: To each 70 ml of lysate, 30 ml of glycerol was

added followed by addition of 25-50 mg of penicillin and 25-50 mg of gentamicin

per 500 mL of material. To make haemolysate with lower Hb concentration, an

appropriate volume of 30 % (v/v) glycerol in 9 g/L NaCl was added to the stock and

mixed well using a roller for one hour. While stirring continuously, 1 mL aliquots

were dispensed aseptically into 2 mL sterile vials, capped, sealed and labelled

appropriately. The samples were assigned unique codes A, B and C for sample with

low, normal and high Hb concentration respectively. The samples were preserved at

2-80 C in the refrigerator awaiting dispatch. An elaborate procedure of how the

samples were prepared is annexed (Appendix I).

3.6.3. Assigning value of Hb concentration to the EQA samples (target

values)

The Hb concentration for the three samples was assigned using the reference method

i.e. by use of a standard haematology analyser (Sysmex XS-800i, Sysmex

Corporation, Kobe, Japan) as the gold standard. These were used as the

target/reference values against which the results of the participating laboratories

Page 45: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

24

were compared. The target/reference values for the three samples were 6.2 g/dl, 13.6

g/dl and 18.1 g/dl for sample A, B and C respectively.

3.6.4. Sample handling and storage

In order to minimise pre-analytical errors before and after sending materials to the

laboratories strict measures regarding sample aliquoting and sample handling were

taken. When preparing aliquots only one vial was handled at a time to avoid

exposing the samples to room temperature for more than 10 minutes. Samples were

checked for any leakage, spillage or contamination and for correct labelling with

well sticking labels. They were then stored according to temperature requirements

until testing can be performed, that is at 2-80C.

3.6.5. Sample packaging and transportation

The sample package for each participating laboratory contained three samples

labelled sample A, sample B and sample C. The samples were placed in leak proof 2

ml plastic vials which were properly labelled with the unique code numbers. The

quality of the samples was maintained during transport by use of icepacks so as to

maintain temperatures of 2-80C. Samples were also secured during transport so that

there is no leakage, spillage or contamination. During transport to the laboratories a

triple packaging system was used which consisted of a leak-proof primary

receptacle, a leak-proof secondary packaging with sufficient absorbent material and

an outer packaging of adequate strength (Appendix II). An efficient and reliable

means of transport was used.

Page 46: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

25

In order to ensure that all the necessary requirements were followed by the

participating laboratories, the samples were accompanied by a complete set of

instructions with respect to storage, handling and deadline for analysis. The

instruction sheet contained information about the number of samples, type of

samples and details of how the samples are labelled, how to mix samples before

testing and a deadline for analysing the samples. Laboratories were informed to

perform checks for the samples before testing such as the storage conditions, code

numbers, breakage/ leakage, sign of contamination and the temperature condition at

which the samples were received (Appendix III).

3.6.6. Sample processing

The laboratories were instructed to process the samples within two days after

delivery using their current analytical procedures, to process the samples in the same

way as routine samples and record the results in the worksheet provided. Samples

were analysed in duplicate by performing two assays with a difference of not more

than six hours between the assays.

Alongside the study, each laboratory also received a questionnaire so as to collect

data on analytical aspects related to Hb determination. Data on analytic aspects of

Hb determination included, methods of analysis used, reference ranges used by each

laboratory, control materials and reagents used for analysis, use of Standard

Operating Procedures and previous participation in EQAS. Demographic data

included the code (assigned by Principal Investigator), location and mailing address

of the participating laboratories. The laboratories were also required to state clearly,

by filling the questionnaire, the date the samples were received, date of analysis and

Page 47: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

26

the type of equipment/method used. The results were collected from the

participating laboratories within one week after dispatch of the samples.

3.7. Data Management and Analysis

Data was entered in Ms excel worksheet, coded and edited using consistency

checks, checks for duplicate entries and range checks. Data was analysed using

XLSTAT statistical software (XLSTAT Version 2013.3.03). For inter-laboratory

variability, coefficient of variation (CV) was used. The higher the CV the greater the

dispersion or spread in values of the variable, whereas when the CV is lower, the

residuals relative to the predicted value are smaller. The variation was also assessed

using analysis of variance (ANOVA). Significant differences (P<0.001) between

means were assessed by ANOVA. For evaluation of performance of participating

laboratories, acceptable performance criteria given by the CLIA’ 88 was used.

Accuracy of results was analysed by calculating the difference (expressed as the

bias) between the Hb concentration provided by the participating laboratories and

target values. We tested for the effect of the Hb concentration on the proportion of

the laboratories that performed well based on CLIA’88 criterion using Chi-Square

analysis. Data presentation was done using tables and graphs.

3.8. Ethical Considerations

Approval to carry out the study was sought from the Jomo Kenyatta University of

Agriculture and Technology as well as Kenyatta National Hospital/University of

Nairobi Ethics and Research Committee (Ref: KNH – ERC/A/1) (Appendix VIII).

Consent was obtained from the laboratory managers of the participating laboratories

Page 48: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

27

prior to enrolment in the study. Confidentiality was maintained by coding the

participating laboratories rather than using their names. The questionnaires and

worksheets were kept under lock and key so that only the Principal Investigator

could access them.

Page 49: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

28

CHAPTER FOUR

RESULTS

A total of two hundred and ninety two laboratories responded. Twenty seven

different analysers were used across all the laboratories (Table 4-3).

Table 4-3: The Different Analyzers and the Number of Laboratories Using

Each Analyser Type

Analyzer Method of

Operation

Number of Hb

Measurements

Number of

Laboratories

ABX Micros

Automated

6

1

ACT Diff Beckman Coulter

Automated 18 3

BTS 305 Manual 12 2

Celltac Automated 174 29

Cera Check Manual 18 3

Colourimeter Semi- Automated 78 13

Coulter Counter Automated 24 4

Diaspect Manual 306 51

Drew Automated 6 1

Easy Mate Manual 12 2

Sahli Manual 144 24

Hb Meter Manual 12 2

Hemocontrol Manual 264 44

Hemocue Manual 384 64

Hichroma Automated 6 1

Humalyzer Junior Semi- Automated 30 5

Hybrid Automated 12 2

Kyrot Automated 6 1

Medonic Automated 30 5

Mindray Automated 60 10

Mission Manual 36 6

Pentra ES 60 Automated 6 1

RMS Automated 6 1

Stat Manual 12 2

Page 50: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

29

Sysmex Automated 48 8

Urit Manual 30 5

Erma Automated 12 2

4.1. The performance of participating laboratories in differentiating, low,

normal and high Hb measurements

Based on the CLIA’88 acceptable performance criteria for haemoglobin, which is,

target ± 7%, a total of 61% of laboratories had acceptable performance across all

measurements. The analyses shown in table 4-4 revealed that laboratory

performance using CLIA’88 criteria declined with increase in the concentration of

the target Hb value: 68%, 64% and 51% of laboratories passing for the sample with

low (6.2g/dl), normal (13.6g/dl) and high (18.1g/dl) Hb values respectively. These

differences were statistically significant (p <0.001) for the three Hb test

concentrations for both reading 1 and reading 2 (Table 4-4).

Page 51: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

30

Table 4-4: The Number and Percentage of Laboratories that Performed Well

(Passed) or Failed according to CLIA’ 88 Test Performance Criteria

Measurement Reading 1 Reading 2 Average

Passed Failed Percent

success

Passed Failed Percent

success

Passed Failed Percent

success

Low (A) 198 94 67.81 199 93 68.15 397 187 67.98

Normal (B) 191 101 65.41 183 109 62.67 374 210 64.04

High (C) 148 144 50.68 148 144 50.68 296 288 50.68

Chi square 21.16 19.50 40.30

P value P<0.001 P<0.001 P<0.001

4.2. The inter-laboratory variability of the Hb measurements

The overall inter-laboratory CV was 33.3% for sample A, 25.1% for sample B and

29.4% for sample C irrespective of the analyser a laboratory used (fig 4-1). When

inter-laboratory CV was calculated across laboratories using the same analyser, the

CV reduced to 5.1 % (Hemocontrol) to 41 % (Urit) for sample A, 2.2% (Celltac) to

35% (Diaspect) for sample B and 3.4 % (Medonic) to 42.6% (Diaspect) for sample

C (Table 4-5).

Page 52: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

31

Figure 4-1: Coefficient of variation as a measure of inter-laboratory variation

for the three Hb test concentrations

Page 53: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

32

Table 4-5: The Inter-Laboratory Variability of Hb Measurements for Each

Type of Hb Analyser

Mean ± 1 Standard Deviation CV (%)

Hb

Analyzer

n A B C A B C

Celltac 29 6.43±1.09 13.83±0.30 18.7±0.97 16.9 2.2 5.2

Colorimeter 13 6.64±1.18 13.57±1.79 17.39±1.72 17.8 13.2 9.9

Diaspect 51 4.06±0.68 9.61±3.36 9.37±3.99 16.8 35.0 42.6

Hemocontrol 44 6.25±0.32 13.96±0.51 18.86±0.84 5.1 3.6 4.4

Hemocue 64 8.37±2.37 16.16±3.09 21.14±4.10 28.3 19.1 19.4

Humalyzer

Junior

5 6.10±0.43 13.00±1.07 17.64±0.84 7.1 8.3 4.8

Medonic 5 6.84±1.56 14.22±1.85 18.54±0.63 22.9 13.0 3.4

Mindray 10 5.99±0.6 13.33±0.48 18.35±0.78 10.1 3.6 4.3

Mission 6 11.25±2.62 16.52±1.49 19.67±1.14 23.3 9.0 5.8

Sahli 24 5.80±1.83 10.44±2.55 13.2±2.89 31.6 24.4 21.9

Sysmex 8 6.23±0.52 13.43±0.73 18.25±1.07 8.3 5.5 5.8

Urit 5 8.44±3.47 17.36±4.98 21.76±3.07 41.1 28.7 14.1

Average 269 6.465+2.12 13.39+4.95 17.08+3.26 32.85 24.38 28.98

Page 54: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

33

4.3. Comparison of the variability of Hb measurements among the various

Hb measurement methods/equipment

4.3.1. Variation of Hb measurements due to analysers

Table 4-6 shows the accuracy of Hb analysers in estimating low, normal and high

Hb measurements. The values obtained by the different analysers and methods were

compared to the mean of the reference values. The mean Hb measurements from the

various analysers were significantly different from the mean of the reference values

(F (27) = 17.382, P<0.001,). In spite of this, seventy eight percent (n=21/27) of the Hb

analysers produced mean Hb values that were not significantly different from the

mean of the reference values and only 22% (n=6 of 27) had mean values deviating

significantly from that expected mean of the reference values. Most of the common

analysers produced estimates of Hb that were consistently different or similar to the

reference values for sample A, B and C (Tables 4-7, 4-8 & 4-9). The exception was

Sahli and Mission. Sahli gave Hb estimates that were similar to the reference value

for sample A but gave results that were significantly different than the reference

values for sample B and C. Mission gave estimates that were significantly different

from sample A and B reference values but not for sample C reference value.

Page 55: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

34

Table 4-6: The Accuracy of Analysers in estimating Hb values across all levels

of measurement in comparison to the expected or mean of all reference values

(6.2, 13.6, and 18.1) combined (i.e. 12.633)

Analyzer N Mean

Hb

Value Standard.

Error

T. Statistic P value

Intercept 162 12.63 12.633 0.403 31.353 < 0.001

ABX Micros 6 11.63 1.000 2.132 -0.469 0.639

Beckman

coulterAct Diff

18 12.92 0.283 1.274 0.222 0.824

BTS 305 12 10.53 2.100 1.534 -1.369 0.171

Celltac 174 12.99 0.356 0.560 0.636 0.525

Cera Check 18 10.89 1.739 1.274 -1.365 0.173

Colourimeter 78 12.6 -0.035 0.707 -0.049 0.961

Coulter

Counter

24 12.52 -0.117 1.122 -0.104 0.917

Diaspect 306 7.68 -4.951 0.498 -9.935 < 0.001

Drew 6 13.45 0.817 2.132 0.383 0.702

Easy mate 12 15.57 2.933 1.534 1.912 0.056

Hb Meter 12 16.76 4.125 1.534 2.688 0.007

Hemocontrol 264 13.05 0.416 0.512 0.812 0.417

Hemocue 384 15.23 2.593 0.480 5.398 < 0.001

Hichroma 6 12.75 0.117 2.132 0.055 0.956

Humalyzer

Junior

30 12.09 -0.540 1.019 -0.530 0.596

Hybrid 12 11.84 -0.792 1.534 -0.516 0.606

Kyrot 6 13.5 0.867 2.132 0.406 0.684

Medonic 30 13.13 0.497 1.019 0.487 0.626

Mindray 60 12.54 -0.090 0.775 -0.116 0.908

Mission 36 15.73 3.092 0.945 3.272 0.001

Pentra Es 60 6 12.77 0.133 2.132 0.063 0.950

Page 56: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

35

RMS 6 12.45 -0.183 2.132 -0.086 0.931

Sahli 144 9.81 -2.826 0.587 -4.812 < 0.001

Stat 12 14.96 2.325 1.534 1.515 0.130

Sysmex 48 12.64 0.010 0.843 0.012 0.990

Urit 30 15.85 3.217 1.019 3.156 0.002

Erma 12 12.51 -0.125 1.534 -0.081 0.935

Page 57: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

36

Table 4-7: The accuracy of the most commonly used analysers in Kenya in

estimating low Hb value (6.2 g/dl)

Analyzer Method

of

operation

N Mean Value Standard.

error

T

statistic

P value

Intercept 6.200 0.284 21.823 < 0.001

Celltac A 29 6.43 0.234 0.395 0.594 0.553

Colourimeter SA 13 6.64 0.438 0.498 0.880 0.380

Diaspect M 51 4.06 -2.137 0.351 -6.083 < 0.001

Hemocontrol M 44 6.25 0.048 0.361 0.132 0.895

Hemocue M 64 8.37 2.167 0.339 6.397 < 0.001

Humalyzer

Junior

SA 5 6.1 -0.100 0.719 -0.139 0.889

Medonic A 5 6.84 0.640 0.719 0.890 0.374

Mindray A 10 5.99 -0.210 0.546 -0.384 0.701

Mission M 6 11.25 5.050 0.666 7.579 < 0.001

Sahli M 24 5.8 -0.396 0.414 -0.956 0.340

Sysmex A 8 6.23 0.025 0.594 0.042 0.966

Urit M 5 8.44 2.240 0.719 3.117 0.002

KEY:

A – Automated

SA – Semi Automated

M - Manual

Page 58: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

37

Table 4-8: The accuracy of the most commonly used analysers in Kenya in

estimating normal Hb value (13.6 g/dl)

Analyzer Method

of

operation

N Mean Value Standard

error

T

statistic

P value

Intercept 13.6 0.447 30.418 < 0.0001

Celltac A 29 13.83 0.231 0.621 0.372 0.71

Colourimeter SA 13 13.57 -0.031 0.784 -0.039 0.969

Diaspect M 51 9.61 -3.988 0.553 -7.213 < 0.0001

Hemocontrol M 44 13.96 0.357 0.568 0.628 0.53

Hemocue M 64 16.16 2.556 0.533 4.795 < 0.0001

Humalyzer SA 5 13 -0.6 1.131 -0.53 0.596

Medonic A 5 14.22 0.62 1.131 0.548 0.584

Mindray A 10 13.33 -0.27 0.86 -0.314 0.754

Mission M 6 16.52 2.917 1.049 2.782 0.006

Sahli M 24 10.44 -3.158 0.652 -4.846 < 0.0001

Sysmex A 8 13.43 -0.175 0.935 -0.187 0.852

Urit M 4 17.36 3.76 1.131 3.324 0.001

KEY:

A – Automated

SA – Semi Automated

M - Manual

Page 59: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

38

Table 4-9: The accuracy of the most commonly used analysers in Kenya in

estimating high Hb value (18.1 g/dl)

Analyser Method

of

operation

N Mean Value Standard

error

T

statistic

P value

Intercept 18.100 0.540 33.498 < 0.001

Celltac A 29 18.7 0.600 0.751 0.799 0.425

Colourimeter SA 13 17.39 -0.708 0.948 -0.747 0.456

Diaspect M 51 9.37 -8.733 0.668 -13.069 < 0.001

Hemocontrol M 44 18.86 0.759 0.686 1.106 0.270

Hemocue M 64 21.14 3.038 0.644 4.714 < 0.001

Humalyzer Junior

SA 5 17.64 -0.460 1.367 -0.337 0.737

Medonic A 5 18.54 0.440 1.367 0.322 0.748

Mindray A 10

18.35 0.250 1.039 0.241 0.810

Mission M 6 19.67 1.567 1.267 1.236 0.217

Sahli M 24 13.2 -4.900 0.788 -6.221 < 0.001

Sysmex A 8 18.25 0.150 1.130 0.133 0.895

Urit M 5 22.63 4.525 1.504 3.008 0.003

KEY:

A – Automated

SA – Semi Automated

M - Manual

Page 60: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

39

4.3.2. Variation of Hb measurements due to methods

A total of 74.32%, (n=217) of the laboratories used analysers/machines that

employed a manual method, 24% used an automated methods (n=70) and 1.7%,

(n=5) used semi-automated methods (fig 4-2). ANOVA comparisons revealed that

the mean Hb across the three methods of analyser/machine operation were not

significantly different from the mean of the reference values (F(3) =1.333, P=0.262,

mean of reference value =12.633). When we calculated the CV using the mean of

reference values, the CV for the automated and semi automated methods were

similar but was large for analysers using the manual method (CV for automated

analyzers =7.08%, CV for semi automated analyzers = 7.04% and CV for manual

analyzers =34.26% (fig 4-3).

Figure 4-2: Percentage number of laboratories as per each method of

equipment /analyser operation

Page 61: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

40

Figure 4-3: Coefficient of variation around the standard reference as a measure

of reliability of the different methods of Hb analyser operation

4.4. Resuts for the questionnaire

Important analytical data regarding the practice of internal quality control in the

laboratories was as shown in table 4-10.

Page 62: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

41

Table 4-10: The resuts for the questionnaire

S/No Question Number Of

Laboratories

Percentage

1. Used SOPs during testing 200 68.5 %

2. Previous participation in EQA 44 15%

3. Used quality control materials 131 45%

4. Reported the correct reference values for Hb 254 87%

5. Used manual methods 217 74.3%

6. Used automated methods 70 24%

7. Used semi-automated methods 5 1.7%

Page 63: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

42

CHAPTER FIVE

DISCUSSION, CONCLUSIONS AND RECOMMENDATIONS

5.1. Discussion

As demonstrated in this study, 61% of laboratories met the CLIA’88 criterion for

samples A, B and C (Table 4-4). The ability of a laboratory to accurately determine

the Hb values of a sample depended on the type of analyser/method used and the

application and adherence to quality control practices by the laboratory. The results

in this study have demonstrated variation of Hb results when laboratories analyse

the same sample. Such results will have potential implications in the classification of

anaemic patients. For instance, assuming sample A, B and C was representative of

an anaemic subject, a healthy normal subject and a subject with higher Hb

concentration respectively. A total of 1.7 % (5 out of 292), 19.8 % (58 out of 292)

and 37.3 % (109 out of 292) laboratories respectively, would have misclassified

these cases.

The present study demonstrates inter-laboratory variation in Hb measurements for

the three Hb samples, which was slightly higher at low Hb concentration. Patients

usually obtain health care services in different clinical settings and therefore may

have their Hb levels measured in different laboratories using different methods that

may not be comparable. In this study the range of Hb results obtained for each

analyser/method varied widely. Confusion may arise when laboratories give

different results for the same sample. For instance, in the present study there was a

considerable high overall interlaboratory variation. However, when the analysers

Page 64: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

43

were grouped according to the type of analyser, the variation of Hb results reduced

to a great extent. This suggests that both method/analyser bias and laboratory-

specific bias are the cause of the overall variability of interlaboratory Hb results.

Unfortunately, the interlaboratory variation may have consequences for clinical

practice depending on the case at hand. The results obtained in the present study are

similar to those found in a survey of Belgian hospitals to assess the reliability of Hb

measurements which reported interlaboratory variation ranged from 0.6 % (Roche)

to 6.7 % (IL) for sample 1 and from 2.0 % (Radiometer) to 4.5 % (IL) for sample 2

( Blerk et al., 2007). Also, a study carried out to assess the analytical quality of

Jordanian haematology laboratories in the routine haematological parameters

reported that there was considerable inter-laboratory variation in laboratories using

manual methods (Bilto, 1999).

In this study, some analysers gave results that are comparable to the reference values

while as others overestimated and others underestimated the values. An overall

negative bias was observed for Hb results obtained using Diaspect and Sahli.

Analytic bias can directly affect patient classification and clinical decision making

(Klee, 1995). In addition, underestimation and overestimation of Hb concentrations

could lead to unnecessary clinical interventions in patients. Some analysers gave

results that are comparable to the reference values while as others overestimated and

others underestimated the values.

The underestimation of Hb values by both Diaspect and Sahli appears to increase

with an increase in Hb concentration. These results supports findings from other

Page 65: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

44

studies, which reported that Sahli method underestimated Hb values. According to

Patil, Thakare and Patil, (2013) Sahli underestimates the Hb in capillary blood by

0.62gm/dl and 1.12g/dl in venous blood. Kapil et al., (2002) compared HemoCue®

analyser with Sahli’s method and they found that Sahli’s method underestimates Hb

by 1.06 g/dl compared to HemoCue® analyser. Barduagni et al., (2003) compared

the performance of the Sahli and the colour scale methods in diagnosing anaemia in

school children and found that Sahli method gave low mean Hb level (11.4g/dl) and

had low specificity (39.0%), thus giving a high rate of false positive results. The

considerable variability in the Hb values obtained with the Sahli could be due to its

inbuilt errors, subjective visual colour comparison, inaccuracy in pipetting of blood,

fading of comparator after prolonged use and poor sensitivity and reliability.

Contrary to the results presented in this study, a study conducted to evaluate the

utility of the Diaspect as a point of care analyser reported that the Hb values

obtained with Diaspect system compared well with the reference analyser

(Robertson, Lewis, & Osei-Bimpong 2011). It is evident from the findings of this

study that both Diaspect and Sahli methods may classify healthy individuals as

anaemic.

On the other hand, Hemocue, Urit and Mission were consistently overestimating the

Hb values when compared to the reference. Overestimation of Hb values by

Hemocue and Urit appears to increase with an increase in Hb concentration with an

exception of Mission where the overestimation decreases with an increase in Hb

concentration. Hemocue, Urit, Mission showed a positive bias in that order of

increasing bias.

Page 66: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

45

The reliability, precision and accuracy of hemocue is controversial. The results of

this study showed that hemocue overestimated Hb values by more than 2g/dl from

the reference and this overestimate appears to increase with increase in the actual Hb

value. The present study findings agree with Mohanram, Ramana-Rao & Sastry,

(2002) who observed that hemocue method overestimated Hb by more than 2g/dl.

Prakash et al., (1999), in their study on utility of hemocue in estimation of Hb

against standard blood cell counter reported that hemocue produced consistently

higher values of Hb than the blood cell counter and suggested that a constant

correction factor of 0.5g/dl should be subtracted from Hb estimates of hemocue.

Bhaskaram et al., (2003) also found that hemocue overestimated Hb values by 10%

- 15% when they compared it to cyanmethaemoglobin method in their study on

validation of Hb estimation using hemocue and this overestimate appeared to

increase with increase in the actual Hb value. However other studies such as those

conducted by Sari et al., (2001) to estimate the prevalence of anaemia among

Indonesian mothers, Von Schenck, Falkensson and Lundergerg, (1986) in evaluation

of hemocue and Rechner et al., (2002) on evaluation of hemocue compared with the

coulter STKS have found hemocue to be reliable and accurate and have supported

its use. Yet, a study conducted in Mexico reported that hemocue underestimated Hb

values (Neufeld et al., 2002). However, due to these contradicting findings on the

accuracy and reliability of hemocue, it is suggested that further studies need to be

done to validate the hemocue haemoglobinometer.

The findings in this study about Urit agree well with those of Jitthai (2012) who

reported that Urit gave significantly higher Hb values than the automated blood

Page 67: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

46

analyser. Generally, the variation of Hb values was slightly higher at low and high

Hb concentrations and least at normal Hb concentration suggesting that majority of

the laboratories may be using calibrators for the normal Hb concentration only while

calibrating the analysers and fail to include calibrators for low and high Hb

concentrations. The results presented in the current study demonstrate that the

problem of inter-laboratory variability even between those laboratories that use the

same brand of analyser is still big suggesting that there are inconsistent calibrations

of instruments at the laboratory level. It is apparent that the variation of Hb

measurements varies across the type of analyser used and the methods.

One possible explanation for deviant results from the different laboratories could

have been due to “matrix effects”. Processed blood haemolysate is prepared in a

matrix that is less complete than patient’s specimens and may have properties that

differ from pure blood. For example, addition of stabilizers or other additives that

are used in the preparation of processed blood samples are known to alter the

properties of the specimen (Rej, Jenny and Bretaudire, 1984., Fasce et al., 1973,

Breutaudire et al., 1981, Ulda, 1984, Fraser & Peake, 1980, Rej & Drake, 1991).

Such anomalies are considered to be “matrix effects” (Clark, Kricka & Whitehead,

1981). Although matrix effects might explain some of the evident method- specific

biases reported in this study, and which is expected to be more or less constant

across all analyzers (Miller & Kaufman, 1993; Power, 1995). The widely variant

results obtained with Sahli, Diaspect, Hemocue, Mission and Urit are less likely to

be caused by matrix effects but may rather reflect serious calibration problems with

these instruments in the field. However, other causes of bias that can result in false

Page 68: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

47

test results include instrument calibration errors, reagent lot differences, inaccurate

dilutions, personal performance error and other factors (Klee, 1999; Jenny &

Jackson, 2000; Miller, Levinson & Elin, 2003.

The present study has demonstrated that there is considerable high variation of Hb

values when using manual methods when compared with the automated methods.

All the analysers that produced Hb values that deviated significantly from the

expected mean values employed the manual method of operation. Also as is evident

in this study, the manual methods showed higher CVs than the automated methods

indicating that the automated methods generally have higher precision than the

manual methods. These results agree with those of a previous study conducted by

Fink et al., (1997) where manual methods produced greater CVs than the automated

methods. Also, as reported by Bilto (1999), in his study carried out to assess the

analytical quality of Jordanian haematology laboratories, the cell counter methods

showed better inter-laboratory agreement than the manual methods. Inherent errors

in the manual methods of Hb value estimation on a sample and errors caused by the

observer would be the causes of this great variation seen in the manual methods. Use

of calibration specimens and high quality control Hb solutions; checking and

calibration of equipment are crucial for proper determination of Hb in the blood

(Pupkova et al., 2002). These apparent differences in Hb results obtained by

different laboratories using different analysers/methods (inter-device) and in

different laboratories using the same analysers/methods (intra- device) can be

addressed by harmonization of procedures for Hb measurement in all the

laboratories.

Page 69: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

48

The results obtained from the questionnaire indicate that very few (15%)

laboratories participate in External Quality Assessment Schemes/PT programmes

(Table 4-10). A possible explanation to this may be due to lack of funds since EQA

is expensive, lack of awareness by some of the laboratories and lack of commitment

by the laboratory staff and management as well. Regular participation in EQA

assists individual laboratories to continuously monitor their performance and to

compare it with that of other laboratories. Also, it is an effective means for

identifying problems that cause interlaboratory variation of laboratory results and

initiates a process towards solving these problems thus improving the quality of

service at the level of each individual laboratory. Lack of participation in EQA is

one of the factors among others, which cause interlaboratory variation of laboratory

results.

Different laboratories used different reagents and control materials depending on the

method of analysis/analysers used in their laboratory. In order to achieve

standardisation and interlaboratory comparability of Hb results, manufacturers

should consider coming up with a single common type of calibration material that

can be used for all analytical methods and analysers. A total of 55% of laboratories

reported not using quality control/reference materials when performing Hb test. This

could be due to the fact that quality control materials are expensive and are not

readily available to most laboratories. A close observation of the results indicate that

even those laboratories that use quality control materials do not use all the three

levels but only the normal Hb concentration.

Page 70: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

49

A good number (31.5%) of the laboratories did not use SOPs when performing Hb

tests, however, most laboratories reported not experiencing any challenges in the

analysis of the samples. These poor quality control practices further increase the

problem of interlaboratory variation of Hb measurements. Recent recommendations

from the European External Quality Assessment–Organizers Working Groups A and

B ( Libeer et al., 1996; Thienpont et al., 1995), the European Union directive for in

vitro diagnostic medical devices (Directive 98/79/EC 1998) the International

Organization for Standardization (ISO 17511:2003, 2007) and the Clinical and

Laboratory Standards Institute (CLSI Document X5-R, 2006) have recognized the

objective of harmonized results and the importance of using commutable reference

materials for method calibration traceability and for proficiency testing (PT)

programs. To ensure standardisation of Hb measurements throughout the world, The

International Council for Standardization in Haematology (ICSH) in conjunction

with Eurotrol, B.V. (Ede, NL; http://www.eurotrol.com) released a new lot of the

haemiglobincyanide (HiCN(Fe), HiCN) standard in 2008 for use in the

standardization and calibration of whole blood Hb measurements on most

haemoglobinometers and automated blood cell counters (Davis & Jungerius, 2010).

However this reference standard is not readily available in the developing countries

laboratories. There was a slight variation in the reference ranges being used in the

laboratories which can have impacts on patient classification.

5.2. Conclusions

Laboratories participating in EQAs of Hb measurements are able to accurately

differentiate low and normal Hb specimens and to a lesser extent, high Hb specimen.

Page 71: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

50

Inter-laboratory variability of Hb measurements in an external quality assessment is

quantifiable. Inter-laboratory variation of Hb results for samples with low, normal

and high Hb concentrations was 33.3%, 25.1% and 29.4% respectively.

The results have shown that Celltac, Humalyzer Junior, Medonic, Mindray,

Colourimeter, Hemocontrol and Sysmex produce accurate and reliable Hb results

than Diaspect, Sahli, Hemocue, Urit and Mission.

The manual methods produced higher CVs than the automated methods thus

indicating that manual methods generally have lower precision.

5.3. Recommendations

Clinical laboratories need to regularly participate in continuous EQAS in

order to achieve interlaboratory comparability of Hb results.

There is need for an oversight body and policies to be established by the

government that will oversee the implementation of method comparison

studies in both private and public hospital laboratories in order to ensure

interlaboratory and inter-method comparison of analytical results for Hb.

This study has established the need for validation and standardisation of all

analysers, methods and procedures of Hb measurement being used in the

Kenyan clinical laboratories in order to achieve interlaboratory

comparability.

Laboratories should embrace automation and gradually replace manual

methods with automated methods of Hb measurement which are more

accurate and reliable.

Page 72: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

51

REFERENCES

Amukele, T., Michael, K., Hanes, M., Miller, R. & Jackson J. (2012). External quality

assurance performance of clinical research laboratories in Sub-Saharan Africa.

American Journal of Clinical Pathology, 138 (5), 720-723.

Argawal, R. & Heinz, T. (2001). Bedside haemoglobinometry in hemodialysis patients:

Lessons from point of care testing. ASAIO Journal, 47(3), 240-243.

Barbara, H., Anna, P. & Norma, J. (2000). Basic Medical Laboratory Techniques. 4th

Ed. Albany, NY. Delmar Publlishers.

Barduagni, P., Ahmed, A.S., Curtale, F., Raafat, M. & Soliman, L. (2003). Performance

of Sahli and colour scale methods in diagnosing anaemia among school children in

low prevalence areas. Trop Med Int Health, 8 (7), 615 – 618.

Bhaskaram, P., Balakrishna, N., Radhakrishna, K.V. & Krishnaswamy, K. (2003).

Validation of haemoglobin estimation using hemocue. Indian J Pediat, 70, 25-28.

Bilto, Y. Y. (1999). Consensus and accuracy in haematology laboratories of developing

countries: the Jordanian experience. Clin Lab Haematol, 21 (1), 11–15.

Birjees, M.K., Nazish, G.I., Nadira, J., Nadia, N. & Masood, A. (2009). External Quality

Assessment Scheme (IEQAS) in Haematology at National Institute of Health,

Islamabad, Pakistan. J Coll Physicians Surg Pak, 19 (12), 786-790.

Bland, J.M. and Altman, D.G. (1986). Statistical-methods for assessing agreement

between two methods of clinical measurement. The Lancet, 1, 307–310.

Blerk, M.V., Coucke, W., Chatelain, B., Goossens, W., Jochmans, K., Meeus, P… &

Libeer, J-C. (2007). External quality assessment in the measurement of

Page 73: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

52

haemoglobin by blood gas analysers in Belgium. Scand J Clin Lab Invest, 67, 735–

740.

Boulton, F., Nightingale, M. & Reynolds, W. (1994). Improved strategy for prospective

blood donors for anemia. Transfusion Medicine, 4, 221-225.

Bretaudire, J-P., Dumont, G., Rej, R., & Bailly, M. (1981). Suitability of control

materials. General principles and methods of investigation. Clin Chem, 27, 798-805.

Burger, S. & Pierre-Louis, J. (2003). A Procedure to Estimate the Accuracy and

Reliability of HemoCue™ Measurementsof Survey Workers. Retrieved from:

[http://www.idpas.org/pdf/2362ProcedureToEstimate.pdf] website [accessed 12th

July 2013].

Carter, J.Y., Lema, O.E., Adhiambo, C.G. & Materu, S.F. (2002). Developing external

quality assessment programmes for primary health: care level in resource

constrained countries. Accredit Qual Assur, 7, 345-350.

Chen, P., Short, T., Leung, D. & Oh, T. (1992). A clinical evaluation of the Hemocue

haemoglobinometer using capillary, venous, and arterial samples. Anaesth Intensive

Care, 20, 497-503.

Clark, P.M., Kricka, L.J. & Whitehead, T.P. (1981). Matrix effects in clinical analysis:

commutability of control materials between Ektachem, Beckman and SMA 12/60

glucose and urea methods. Clin Chim Acta, 113, 293-303.

Clinical and Laboratory Standards Institute. (2006). Metrological traceability and its

implementation: a report, Joint IFCC-CLSI project, CLSI document, EP32-R

Page 74: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

53

(Formerly X05-R). CLSI, Wayne, Pennsylvania, USA, 2006. (Sample retrieved

from www.ifcc.org or www.clsi.org)

Clinical Laboratory Improvement Act. (1988). Federal Register, Feb 28, 1992, 57 (40),

7002- 7186.

Cochran, W. G. (1977). Sampling techniques (3rd ed.). New York: John Wiley & Sons.

Croll, M.H., Chesler, R., and Elin, R.J. (1989). Effect of lyophilization on results of five

enzymatic methods for cholesterol. Clin Chem, 35, 1523-1526.

Davis, B.H. & Jungerius, B. (2010). International Council for Standardization in

Haematology technical report 1-2009: new reference material for

haemiglobincyanide for use in standardization of blood haemoglobin measurements.

Int J Lab Hematol, (32), 139–141.

De Louw, A. V., Lasserre, N., Drouhin, F., Thierry, S., Lecuyer, L., Caen, D. &

Tenaillon, A. (2007). Reliability of HemoCue in patients with gastrointestinal

bleeding. Intensive Care Med, 33 (2), 355-358.

Deom, R., El Aouad, C.C., Heuck, S., Kumara, S.M., Lewis, A. & Wardle, J. (1999).

Requirements and Guidance for External quality Assessment Schemes for Health

Laboratories. WHO/DILJLAB / 99.2.

Directive 98/79/EC of the European Parliament and of the Council of 27 October 1998

(1998). In vitro diagnostic medical devices. Off J Eur Communities, L 331. (41), 1–

37.

EAREQAS. ( 2010). Report of the East African Intercountry Meeting on Strengthening

Capacity for Post Market Surveillance of Diagnostics & third regional meeting of

the East African Regional External Quality Assessment Scheme. Nairobi, Kenya.

March 15-17, 2010.

Page 75: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

54

Fasce, C.F., Rej, R., Copeland, W.H., & Vanderlinde, R.E. (1973). A discussion of

enzyme reference materials: applications and specifications. Clin Chem, 19, 5-9.

Fierdoz, O. (2012). Essential laboratory knowledge for the clinician. Laboratory testing

forms an integral part of patient management. Contin Med Educ, 30 (7), 244-248.

Fink, N.E., Fernandez Alberti, A. & Mazziota, D. (1997). External assessment of analytic

quality in haematology: a necessity in Latin America. Rev Panam Salud Publica,

2(3); 181-188.

Fraser, C.G. &Peake, M.J. (1980). Problems associated with clinical chemistry quality

control materials. Crit Rev Clin Lab Sci, (12), 59-86.

Frean, J., Perovic, O., Fensham, V., McCarthy, K., Poonsamy, B., Dini, L . . . &

Ndihokubya, J. (2012). External quality assurance of national public health

laboratories in Africa, 2002-2009 [abstract]. Bull World Health Organ, 90, 191-199.

Gehring, H., Duembgen, L., Peterlein, M., Hagelberg, S. & Dibbelt, L. (2007).

Hemoximetry as the “gold standard”? Error assessment based on differences among

identical blood gas analyzer devices of five manufacturers. Anesth Analg, 105, 24–

30.

Gomez, S.A., Navarro, N.L., Pérez, C.E., Lozano, M.L., Candela, J.M., Cascales, A . . .

& Rivera, J. (2007). Evaluation of four rapid methods for hemoglobin screening of

whole blood donors in mobile collection settings. Transfus Apher Sci, 36 (3), 235-

242.

International Committee for Standardization in Haematology—Expert Panel on

Haemoglobinometry. (1987). Recommendations for reference method for

haemoglobinometry in human blood (ICSH standard 1986) and specifications for

Page 76: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

55

international haemoglobincyanide reference preparation (3rd Ed.). Clin Lab

Haematol, 9, 73–79.

ISO 17511:2003. In Vitro Diagnostic Medical Devices—Measurement of Quantities in

Biological Samples—Metrological Traceability of Values Assigned to Calibrators

and Control Materials. Geneva, Switzerland: International Organization for

Standardization. Available at: http://www.iso.org/. [accessed October 28, 2007].

Jenny, R.W., & Jackson-Tarentino, K.Y. (2000). Causes of unsatisfactory performance in

proficiency testing. Clin Chem, 46 (1), 89-99.

Jitthai, S. (2012). A comparison of haemoglobin measurement between

haemoglobinometer (POCT) and automated blood cell analyzer in hemodialysis

patient. J Med Technol Assoc of Thai, 40 (3), 4332 - 4337.

Kapil, U., Tandon, M. Pathak, P. & Dwivedi, S.N. (2002). Comparison of haemoglobin

values obtained by Haemocue and Sahli’s methods. Ind J Public Health, 46 (1), 28 -

30.

Klee, G.G. (1995). Analytic performance goals based on direct effect of analytic bias on

medical classification decisions. CDC, 1995 Institute: Frontiers in Laboratory

Practice Research, 219–226.

Klee, G.G. (1999). Maximizing efficacy of endocrine tests: importance of decision-focused

testing strategies and appropriate patient preparation. Clin Chem, 45 (8(B)), 1323-

1330.

Kaplan, L.A., Pesce, A.J. & Kazmierczak, S.C. (ed.) (2003). Clinical Chemistry: Theory,

Analysis, Correlation. 4th Ed. St Louis: CV Mosby.

Kisabei, P. (2013). KMLTTB registered laboratories. [E-mail]. Message to: Mandania, E.

13th June, 2013.

Page 77: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

56

Lauren, B. (2013). Factors affecting haemoglobin measurement. [Online] DOI

10.1007/s10877-013-9456-3.Retrieved from:

http://link.springer.com/article/10.1007%2Fs10877-013-9456-3. [Accessed 12th

may 2013].

Lauren, B., Stephanie, R. & Erin, M. (2011). Continuous noninvasive haemoglobin

monitoring during complex spine surgery: IARS, 113 (6), 1396-1402.

Lewis, S.M., (1998). Quality assurance in haematology. World Health Organization.

(WHO/lab/98.4), 1998: p.104 published online at:

http://whqlibdoc.who.int/hq/1998/WHO_LAB_98.4.pdf. [accessed 12th July 2013].

Libeer, J.C., Baadenhuijsen, H., Fraser, C.G., Petersen, P.H., Ricoss, C., Stock, D. &

Thienpont, L. (1996). Characterization and classification of external quality

assessment schemes (EQA) according to objectives such as evaluation of method

and participant bias and standard deviation. Eur J Clin Chem Clin Biochem, 34,

665–678.

Lippi, G. & Guidi, G.C. (2007). Risk management in the preanalytical phase of laboratory

testing. Clin Chem Lab Med, 45, 720 – 727.

Lippi, G., Banfi, G., Buttarello, M., Ceriotti, F., Daves, M. & Dolci, A. (2007).

Recommendations for detection and management of unsuitable samples in clinical

laboratories. Clin Chem Lab Med, 45, 728 – 736.

Looker, A. C., Sempos, CT., Liu, K. A., Johnson, C. L. & Gunter, E. (1990). Within

person variance in biochemical indicators of iron status: Effects on prevalence

indicators. Amer J Clin Nutr, 52, 541-547.

Martin, D. T., Ashenden, M., Parisotto, R., Pyne, D. & Hahn, A. G. (1997). Blood

testing for professional cyclists: what's a fair hematocrit limit? Sportscience News.

Page 78: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

57

[Retrieved from] http://www.sportsci.org/news/news9703/AISblood.html. [accessed

27th March 2013].

Masimo Cooperation: Technical Bulletin. (2009). Total Haemoglobin Measurements:

Accuracy of Laboratory Devices and Impact of Physiologic Variation. Supplement

S26-030F 7799-5447A-0109. Retrieved from:

[http://www.masimo.com/pdf/SpHb/LAB5447A.pdf] [accessed 12th April 2013].

Miller, W.G. & Kaufman, H.W. (1993). College of American Pathologists Conference

XXIII on matrix effects and accuracy assessment in clinical chemistry: introduction.

Arch Pathol Lab Med, 17, 343-344.

Miller, J.J., Levinson, S.S., Elin, R.J. (2003). Interferences in laboratory tests. In: Clinical

diagnostic technology: the total testing process. Volume 2: the analytical phase.

Washington, DC: American Association of Clinical Chemistry, 2003.

Mokken, F. C., Van der, F. J., Henny, C. P., Goe dhart, P. T. & Gelb, A. W. (1996).

Differences in peripheral arterial and venous hemorheologic parameters. Annals of

Hematology, 73 (3), 135-137.

Mohanram, M., Ramana-Rao G.V. & Sastry J.G. (2002). A comparative study on

prevalence of anaemia in women by cyanmethemoglobin and hemocue methods.

Indian J Community Med, 27 (2), 58-61.

Morris, S., Ruel, M., Cohen, R., Dewey, K., Brière, B. D. L. & Hassan, M. (1999).

Precision, accuracy, and reliability of haemoglobin assessment with use of capillary

blood. Am J Clin Nutr, 69 (6), 1243-1248.

Munich, D.S. (1982). A New Concept for Quality Control of Clinical Laboratory

Investigations in the Light of Clinical Requirements and Based on Reference

Method Values. J. Clin. Chem. Clin. Biochem, 20, 817-824.

Page 79: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

58

Neufeld, L., Garcia-Guerra, A., Sanchez-Francia, D., Newton-Sanchez, O., Ramirez-

Villalobos, M.D. & Rivera-Dommarco, J. (2002). Haemoglobin measured by

hemocue and a reference method in venous and capillary blood: a validation study.

Salud Publica Mex, 44, 219 – 227.

Olafsdottir, E., Hellsing, K., Steensland, H., Tenhunen, R. & Uldall, A. (1994). Adding

new scopes of traditional EQA schemes emphasizing quality improvement. Upsala J

Med Sci, 99; 377-384.

Patel, K.P., Hay, G.W., Cheteri, M.K. & Holt, D.W. (2007). Haemoglobin test results

variability and cost analysis of eight different analyzers during open heart surgery. J

Extra Corpor Technol, 39, 10-17.

Patil, P., Thakare, G. & Patil, S. (2013). Variability and Accuracy of Sahli’s Method in

Estimation of Haemoglobin Concentration. Natl J Integr Res Med, 4 (1), 38 – 44.

Plebani, M. (2009). Exploring the iceberg of errors in laboratory medicine. Clin Chim Acta,

404; 16 – 23.

Plebani, M. (2010). The detection and prevention of errors in laboratory medicine. Ann

Clin Biochem, 47, 101 – 110.

Plebani, M. & Lippi, G. (2010). To err is human. To misdiagnose might be deadly. Clin

Biochem, 43, 1 – 3.

Power, D.M. (1995). Overcoming the limitations of artificial clinical specimens: linkage to

performance of patients’’ specimens [editorial]. Clin Chem, 41, 1689-1692.

Prakash, S., Kapil, U., Singh, G., Dwivedi, S.N & Tandon, M. (1999). Utility of

hemocue in estimation of haemoglobin against standard Blood Cell Counter method.

J Assoc Physicians India, 47 (10), 995-997.

Page 80: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

59

Pupkova, V.I., Zhdanova, V.V., Stepanova, E.G., Leonova, T.I. & Palamarchuk,

M.B. (2002). Quality problems in quantitative determination of Hb in blood. Klin.

Lab. Diagn, 12, 36-39.

Rechner, I.J., Twigg, A., Davies, A.F. & Imong, S. (2002). Evaluation of the hemocue

compared with the coulter STKS for measurement of neonatal haemoglobin. Arch

Dis Child Fetal Neonatal Ed, 86; F188 – 189.

Rej, R. & Drake, P. (1991). The nature of calibrators in immunoassays: are they

commutable with test samples? Must they be? Scand J Clin Lab Invest, 51 (205),

47-54.

Rej, R. & Jenny, R.W. (1992). How good are clinical laboratories? An assessment of

current performance. Clin Chem, 38(7), 1210-1217.

Rej, R., Jenny, R.W & Bretaudire, J-P. (1984). Quality control in clinical chemistry:

characterization of reference materials. Talanta, 31, 851-862.

Renu, S., Katoch, S. C., Upendra, S., Seema, R. & Hema, A. (2007). Impact of external

haematology proficiency testing programme on quality of laboratories. Indian J Med

Res, 126, 428-432.

Ricos, C., Alvarez, V., Cava, F., Garcia-Lario, J. V., Hernandez, A., Jimenez, C. V . . .

& Simon, M. (1999). Desirable specifications for imprecision, inaccuracy, and total

allowable error, calculated from data on within-subject and between-subject

biologic variation. Available from: [http://www.westgard.com/biodatabase1.htm]

[accessed 19th July 2013].

Ricos, C., Baadenhuijsen, H., Libeer, C.J., Petersen, P.H., Stockl, D., Thienpont, L. &

Fraser, C.C. (1996). External quality assessment: currently used criteria for

Page 81: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

60

evaluating performance in European countries, and criteria for future harmonization.

Eur J Clin Chem Clin Biochem, 34, 159-165.

Rippey, J.H. & Williamson, W.E. (1988). The overall role of a proficiency testing

programme. Arch Pathol Lab Med, 112, 340–342.

RNA Medical. (2011). CVC 223 CO-Oximeter Calibration Verification Controls. RNA

Medical, Division of Bionostics. Devens, MA 01434. [Available from]:

http://www.rnamedical.com/pdf/datasheets/cvc223_rev_c.pdf. [accessed 10th July

2013].

Robertson, L., Lewis, D. & Osei-Bimpong, A. (2011). Utility of the Diaspect

Haemoglobinometry System as Point- Of- Care Analyzer. J Near Patient Test

Technol, 10 (1), 17 – 21.

Sah, S.P., Raj, G.A. & Prakash, M.B. (1999). Quality assurance programme in

haematology at a teaching hospital in the eastern region of Nepal. Indian J Pathol

Microbiol, 42 (2), 145-9.

Sari, M., De Pee, S., Martini, E., Herman, S., Sugiatmi. & Bloem M.W, (2001).

Estimating the prevalence of anaemia: a comparison of three methods. Bull World

Health Organ, 79, 506 – 511.

Savage, R. A. (1989). Proficiency testing and laboratory quality. Arch Pathol Lab Med,

113, 983–984.

The Royal College of Pathologists of Australasia. (2010). Chain of Information Custody

for the Pathology Request-Test- Report Cycle in Australia (Guidelines for Pathology

Requesters and Pathology Providers). K:\2002 onward\Operations\Policy Manual

\Guidelines\Chain of Information Custody.DOC. Retieved from:

Page 82: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

61

http://www.health.gov.au/internet/main/publishing.nsf/Content/FDD654D0C46DC3

74CA2573A0000A9C1A/$File/ChainInfoCustody.pdf [accessed 17th July 2013].

Thienpont, L., Franzini, C., Kratochvila, J., Middle, J., Ricos, C., Siehnann, L. &

Stockt, D. (1995). Analytical quality specifications for reference methods and

operating specifications for networks of reference laboratories. Eur J Clin Chem

Clin Biochem, (33), 949–957.

Uldal, A. (1984). Apparent interlaboratory variation using different types of reference sera.

Scand J Clin Lab Invest, 44(172), 163-171.

Von Schenck, H., Falkensson, M. & Lundergerg, B. (1986). Evaluation of “hemocue,” a

new device for determining haemoglobin. Clin Chem, 32, 526 – 529.

Westgard J.O. (2008). Use and interpretation of common statistical tests in method

comparison studies. Clin Chem, 54, 612.

World Health Organization. (2008). Worldwide prevalence of anaemia 1993-2005: WHO

global database on anaemia. [Online]. Retrieved from:

[http://whqlibdoc.who.int/publications/2008/9789241596657_eng.pdf] website

[accessed 12th July 2013].

WHO Regional Office for Africa. (2008). Strengthening public health laboratories in the

WHO African region: a critical need for disease control. Retrieved from:

http://www.afro.who.int/en/fifty-eight-session.html. Accessed september 6, 2012.

World Health Organization. (2007). External quality assessment scheme for haematology.

[Internet]. [Cited 2007]. Retrived from:

http://www.who.int/diagnostics_laboratory/quality/haematology/en/

Page 83: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

62

Zigilstra, W.G. (1997). Standardization of haemoglobinometry: History and new

challenge. Comp Haematol Int, 1, 125-132.

Page 84: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

63

APPENDICES

Appendix I: Procedure for Haemolysate Preparation

STANDARD OPERATING PROCEDURE FOR PREPARATION OF

HAEMOLYSATE SAMPLE

1. Centrifuge anticoagulated blood from a blood bank, with negative serology

for the human immunodeficiency (HIV) hepatitis B (HBsAg), hepatitis C

(HCV) virus in screw capped bottles of appropriate size. Aseptically separate

the plasma from the white layer (layer of leukocytes or buffy coat).

2. Add to each red cell deposit an excess of physiological saline 9g/L (NaCl),

mix well and centrifuge. Discard the supernatant and any remaining buffy

coat.

3. Repeat saline wash two times with sterile physiological saline solution to the

complete elimination of the plasma, leukocytes, and platelets, each time

removing the top layer of the package of erythrocytes in each wash.

4. Add to the washed cells an equal volume of water and 0.5 volumes of

carbon tetrachloride (CCl3). Cap the containers and shake them vigorously in

a mechanical Shaker for 1 hour. Refrigerate overnight to allow the lipids/cell

debris to form a semi-solid interface between carbon tetrachloride and lysate.

5. On the following day, centrifuge at 2 500 g for 20 minutes; transfer the

upper layers into universal containers and centrifuge at about 3000g for one

hour. Collect the upper 95% and pool into a clean tube.

6. To each 70 mL of lysate, add 30 mL of glycerol. Then add antibiotics (25-

50 mg of penicillin) and 25-50 mg of gentamicin per 500 mL of material.

Page 85: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

64

This material is stored at 4 ° c until use for short periods or frozen at -20 for

longer period until required for dispensing.

7. If a lower concentration is required, add an appropriate volume of 30% (v/v)

glycerol in a solution of 9 g/l of sodium chloride (NaCl) to the stock. Mix

well.

8. With continuous stirring distribute the mixture aseptically into sterile

containers. Cover and seal.

9. Assign the haemoglobin value. Preserved at 4 ° C, the product should keep

the value that has been assigned for several months.

Page 86: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

65

Appendix II: Standard Operating Procedure for Triple Packaging System

STANDARD OPERATING PROCEDURE ON TRIPLE PACKAGING

SYSTEM

The triple packaging system consists of three layers: the primary receptacle, the

secondary packaging and the outer packaging. This packaging system is meant to

withstand leakage of contents, shocks, temperature, pressure changes and other

conditions that can occur during ordinary handling in transportation.

1) Put the sample in a primary receptacle (tubes/vials) and ensure that all the

tubes/vials that are being sent are tightly closed to avoid any leakage of the

sample. The primary receptacle containing the specimen must be watertight,

leak proof and appropriately labelled as to content.

2) Wrap every tube/vial in enough absorbent material like paper towels or

cotton wool to absorb all fluid in case of breakage or leakage.

3) Place the samples in a second watertight, leak proof packaging such as a zip

lock bag so as to enclose and protect the primary receptacle(s). Several

wrapped primary receptacles may be placed in a single secondary packaging.

4) Get a plastic container and place the samples in the zip lock bag into it and

close it.

5) Place four frozen ice packs from the -20 degrees Celsius compartment of the

refrigerator, on the sides of a small cool box ensuring that the secondary

container is in the middle of the ice packs.

6) Place absorbent material on top and if necessary between the ice packs.

Page 87: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

66

7) Place the case investigation forms or appropriate laboratory request forms in

a plastic bag to keep from becoming contaminated or destroyed by the wet

ice packs.

Note: An itemized list of package contents must be included between the

outer and secondary container.

8) The cool box is then sealed and addressed to the receiver.

9) Label the box with biohazard symbol and with upward arrows

Note: The outer package must be of adequate strength for its capacity, mass, and

intended use. It must be must be rigid and have one side that is at least 100 mm

X 100 mm, in order for required markings and labels to fit.

Page 88: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

67

Appendix III: Instructions for the Participating Laboratories

INSTRUCTIONS FOR HAEMOGLOBIN TEST SAMPLE HANDLING FOR

EQAS IN HB MEASUREMENT

Inter-laboratory variability of haemoglobin measurements obtained from

selected clinical laboratories in Kenya

Introduction

This is a set of instructions for the above study that should be followed by every

participating laboratory so as to ensure proper sample analysis is done.

You are kindly asked to read keenly the instructions below before performing any

analysis of the samples. Feel free to seek any clarification from the Principle

Investigator.

Instructions to be followed before and during sample analysis:

1) Always keep the samples refrigerated at 2 – 6°c when not in use. Do not

freeze the samples.

2) Perform sample analysis within two days after the delivery of the samples to

your laboratory.

3) Verify that the samples are in good condition and well labeled before the

analysis. There are three samples labeled sample A, Sample B & Sample C.

Each vial contains exactly 1ml of sample.

4) Let the sample attain room temperature before carrying out the test.

5) Mix the samples well before the test by gently inverting the vial about 3 – 4

times.

Page 89: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

68

6) Make sure that your machine/ equipment is in good working condition

before the analysis.

7) Perform a single haemoglobin test of each of the samples in the morning

hours using the analytical method that is currently being used in your

laboratory. Record the results in the worksheet provided.

8) Perform the second haemoglobin test of each of the samples in the afternoon

hours. Record the results in the worksheet provided.

NOTE: The time difference between the morning and afternoon readings

should not exceed 6 hours.

9) Lastly fill the questionnaire provided.

10) The results and the questionnaire will be collected from the laboratory within

one week of sample delivery.

Page 90: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

69

Appendix IV: Questionnaire

QUESTIONNAIRE FOR INTER-LABORATORY VARIABILITY OF

HAEMOGLOBIN MEASUREMENTS STUDY

I am Esther Wangui Mandania, a master’s student at Jomo Kenyatta University of

Agriculture and Technology, faculty of COHES, Department of Medical Laboratory

Sciences. I am conducting research on interlaboratory variability of haemoglobin

measurements obtained from selected clinical laboratories in Kenya.

This is a qualitative questionnaire that is intended to provide the researcher with

important information regarding haemoglobin measurement in the clinical

laboratories of Kenya in order to assess the inter-laboratory variability of

haemoglobin measurements.

Kindly provide the researcher with the following information;

1. Details for the laboratory:

Code assigned to the laboratory (by PI) ------------------------------------------

Mobile Phone No: --------------------------------------------------------------------

Physical Address: P.O BOX--------------------------------------------------------

Email Address------------------------------------------------------------------------

2. Which date and time were the samples received in your laboratory?

Date--------------------------------------Time-------------------------------------

3. In which condition were the samples received by your laboratory?

□ Intact (No damage/breakage)

□ Damaged and leaking

□ Labelled

□ Unlabeled

Page 91: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

70

4. What volume of each EQA sample did you receive?

□Less than 0.5mL

□ 0.5mL

□ More than 0.5mL

5. Which equipment and method of haemoglobin estimation was used to

measure haemoglobin of the EQA samples in your laboratory?

Method?--------------------------------------------------------------------------------

Equipment? --------------------------------------------------------------------- ---

6. Please specify the reference levels for haemoglobin used in your laboratory

for the following;

Adult women---------------------------------------------

Adult men-------------------------------------------------

Children----------------------------------------------------

7. Which reagents did your laboratory use for haemoglobin estimation for the

samples? ---------------------------------------------------------------------------------

---------------------------------------------------------------------------------------------

What challenges and difficulties did your laboratory experience in the

analysis of the samples?---------------------------------------------------------------

---------------------------------------------------------------------------------------------

---------------------------------------------------------------------------------------------

State the SOP that your laboratory used during samples analysis? -------------

---------------------------------------------------------------------------------------------

Page 92: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

71

8. Please state the control materials that were used during samples analysis -----

---------------------------------------------------------------------------------------------

------------------------------------------------------------- ---------------------------

9. Which EQA scheme for haemoglobin has your laboratory participated in

before? -----------------------------------------------------------------------------------

---------------------------------------------------------------------------------------------

Page 93: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

72

Appendix V: Worksheet

Worksheet for inter-laboratory variability of Hb measurements study

Code of laboratory:

Date & time of analysis:

Results (g/dL)

Type of EQA Sample Reading 1 Reading 2

EQA Sample A

EQA Sample B

EQA Sample C

Page 94: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

73

Appendix VI: Ethical Clearance

Page 95: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

74

Page 96: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

75

Appendix VII: Informed Consent Form for Laboratory Management

INFORMED CONSENT FORM

Inter-laboratory variability of haemoglobin measurements obtained from

selected clinical laboratories in Kenya

Introduction

This informed consent form is for the laboratory in-charges, supervisors or

managers of clinical laboratories that i am inviting to participate in research on

haemoglobin measurement. The title of the research is “Inter-laboratory

variability of haemoglobin measurements obtained from selected clinical

laboratories in Kenya”

Principal investigator:

Name: Esther Wangui Mandania

Study Authority: Jomo Kenyatta University of Agriculture and Technology

Faculty: College of Health Sciences

Department: Medical Laboratory Sciences

Address: P.O Box 2028, Thika

Phone No: 0727906351

E-mail: [email protected]

Authorizing body: KNH/UON-ERC

This Informed Consent Form has two parts:

Information sheet (to share information about research with you)

Certificate of consent (for signatures if you agree to take part)

Page 97: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

76

You will be given a copy of the full informed consent form

PART I: Information Sheet

I am Esther Wangui Mandania, a master’s student at Jomo Kenyatta University of

Agriculture and Technology, faculty of COHES, Department of Medical Laboratory

Sciences.

I am doing research on variability of haemoglobin measurements between clinical

laboratories in Kenya. I am going to give you information and invite you to

participate in this research. Before you decide to participate in this research, you can

talk to anyone you feel comfortable with about the research. Also ask any questions

that you may have about the research. You do not have to decide today whether or

not you will participate in the research, sufficient time will be given to you to make

this decision.

If there is anything or any word that you do not understand, please ask me to stop as

we go through the information and i will take time to explain.

Purpose of the research

Comparability of haemoglobin test results is difficult due to variability of

haemoglobin measurements between laboratories. Laboratories should strive to

produce reliable results that are consistent between laboratories so that comparable

results for the same test are obtained in all the laboratories. The reason i am doing

this research is to find out if laboratories can give the same results for the same test

and to determine the variability of haemoglobin measurements between clinical

laboratories in Kenya. The information obtained from the research will help to

identify areas of improvement in haemoglobin measurements in the laboratories.

Page 98: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

77

Procedures

This research will involve analysis of three EQA samples only (A, B and C)

which will be analysed the same day and filling the results of analysis in a

worksheet.

If your laboratory agrees to participate in this research you will receive three

EQA samples labelled A, B and C accompanied by a questionnaire and a

worksheet.

I will request you to analyse the samples within two days in the routine way

as patient samples are analysed, record the results in the worksheet provided

and fill the questionnaire.

You will then send the results and the questionnaire to the principal

investigator or they will be picked from the laboratory whichever you prefer.

Duration

Your expected time commitment for this study will not be more than four days in

total. However the investigator will take about 6 months to complete the whole

research study.

Risks

There are no any risks for your participation in this study.

Benefits

If you (your Laboratory) participates in this research, you will receive a summary of

your(laboratory’s) performance which will include the results from the laboratory,

the expected values (targets) and the potential causes of errors and remedial actions(

for laboratories that will not be in a position to give the target values). This will

provide an opportunity to the laboratory to identify the areas of improvement in the

Page 99: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

78

measurement of haemoglobin as well as in general areas of your laboratory, which

the laboratory can work on to improve the quality of the results.

It is hoped that your participation will help the researcher determine and learn more

about variability of haemoglobin measurements between clinical laboratories in

Kenya, how they impact on patient management and how inter-laboratory

comparability of results can be achieved.

Confidentiality

The information that will be collected from this research project will be kept

confidential. The results from your laboratory and all information about the

laboratory will not be disclosed to any other laboratory or party. The results from

your laboratory and any information about your laboratory will have a number on it

instead of the name of the laboratory. Only the researcher will know what your

laboratory number is. During dissemination of the research findings, this

information will only be reported as group data with no identifying information. All

data, including questionnaires and worksheets for the study will be kept in a secure

location under lock and key and no one will have access to them except the

researcher. After the research is complete, all the raw data that is, the questionnaires

and worksheets will be destroyed by shredding followed by burning.

Compensation

You will not be given any compensation to take part in this research.

Voluntary Participation

Your Participation in this research study is entirely voluntary. It is your choice to

decide whether to participate or not. Whether you choose to participate or not, this

will not affect the relationship you have with the researcher. You may change your

Page 100: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

79

mind later and stop participating even if you agreed earlier. Also you have the right

to withdraw at anytime or refuse to participate entirely in the research study without

any consequences. If you do decide to take part in this study, you will be asked to

sign the consent form.

Dissemination of findings

Research findings will be disseminated through publications, conferences and

through other scientific reports however, no identifying information will be used so

as to maintain confidentiality and privacy. Also you will receive a summarised

report of your laboratory’s performance.

Who to contact

If you have any questions regarding this study, you may ask them now or later, even

after the study has started. If you wish to ask questions later, you may contact any of

the following:

Principal Investigator:

Name: Esther Wangui Mandania

P.O BOX 2028, Thika

Mobile telephone: 0727-906351 or 0789-641355

E- Mail address: [email protected].

Supervisor:

Dr. Juliette R. Ongus

Jomo Kenyattta University of Agriculture and Technology

Department of Medical Laboratory Sciences

Mobile Telephone: +254 722339682

Page 101: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

80

E- Mail address: [email protected]

KNH/UON-ERC Secretariat:

PROF. M L. CHINDIA

SECRETARY, KNH/UON-ERC

Tel: 726300-9 Ext. 44102 OR +2542726300-19 Ext.44102

Post address: P O BOX 20723-00202, Nairobi, Kenya

Physical Address: School Of Pharmacy, UON behind KNH Dental clinic

E-mail: [email protected]

Website: www.ounbi.ac.ke

Link: www.uonbi.ac.ke/activities/KNHUoN

This proposal has been reviewed and approved by the KNH/UON-ERC. If you wish

to find out more, contact the KNH/UON-ERC secretariat using the contact provided

above.

PART II: Certificate of Consent

I have read the foregoing information and understood, i have had the opportunity to

ask questions about it and any questions that i have asked have been answered to my

satisfaction. I understand that my participation is voluntary and that i am free to

withdraw at any time without giving any reason and without cost. I have received a

copy of this consent form. I voluntarily agree to participate in this study.

Laboratory Code: --------------------------------------------------------------------------------

Page 102: INTER- LABORATORY VARIABILITY OF HAEMOGLOBIN …€¦ · measure the same sample using the same test methods, reagents and controls. Lack of repeatability or reproducibility of the

81

Name of laboratory in-charge/supervisor/manager: ----------------------------------------

Signature: -------------------------------------------------------------Date: --------------------

Name of PI: ----------------------------------------------------------- ID No: -----------------

Signature: ------------------------------------------------------------- Date: -------------------


Recommended