INTRAORAL CAMERA WITH MULTIPLE IMAGE MEMORY
Transcript
Page 1: Intra Oral Camera Camrex 192

INTRAORAL CAMERA WITH MULTIPLE IMAGE MEMORY

Page 2: Intra Oral Camera Camrex 192

SpecificationAC 100~240V

VGA out 1024x768

Resolution480 TV Line

DimensionsIntraoral Camera L227 x W27 x H32.9 mmWireless Receiver L172 x W108 x H31 mm

Weight89 g/110 g

Page 3: Intra Oral Camera Camrex 192

ACCESSORIESACCESSORIES

Page 4: Intra Oral Camera Camrex 192

Wireless Camera Handpiece Handpiece Charging Station

Wireless Receiver Antenna

Page 5: Intra Oral Camera Camrex 192
Page 6: Intra Oral Camera Camrex 192

FEATURESFEATURES

Page 7: Intra Oral Camera Camrex 192

Screw the antenna onto the rear of the receiver and set upright.

Plug in the receiver using either of the AC adaptors supplied.

The On/Off switch is at the front

right of the unit.

Page 8: Intra Oral Camera Camrex 192

Use the "FREQUENCY" button on the front of the receiver to select frequency A. LED light should indicate the selected frequency.

NOTE: The handpiece has been preset to transmit on frequency A. If more than one CAMREX wireless camera is in range, or if there is excessive interference, see below under "Changing Handpiece Frequency".

Page 9: Intra Oral Camera Camrex 192

The receiver can be connected to a monitor/projector/TV in two ways:

1) Monitor with RCA video input jack: Using the cable provided, connect it to CAMREX's RCA video output at the rear of the receiver. Set the monitor's input source to "VIDEO“.

RCA Video Cable

Page 10: Intra Oral Camera Camrex 192

The receiver can be connected to a monitor/projector/TV in two ways:

2) Monitor with a VGA connector input: Using the cable provided, connect it to the VGA output at the rear of the receiver. Set the monitor's input source to "VGA".

VGA Cable

Page 11: Intra Oral Camera Camrex 192

Handpiece setup The handpiece requires no special setup other than charging

the battery. Remember to use a sterile disposable sleeve for each patient.

NOTE: Before using CAMREX for the first time, charge the

handpiece for at least 6 hours.

Page 12: Intra Oral Camera Camrex 192

Charging station setup Plug in the charger using either of the AC

adaptors supplied. Place the handpiece in the charger with

control buttons facing forward, making sure the contacts are engaged and power is off.

The charging light should glow orange, turning green when

the handpiece is fully charged. When not in use, the CAMREX wireless

camera should be placed in the charging station.The handpiece is equipped with a Low Battery LED.

Charge the handpiece when the Low Battery lights up.

Page 13: Intra Oral Camera Camrex 192

Changing handpiece frquency The handpiece has been preset to transmit on frequency A. Ordinarily you will not need to change this unless:--There is more than one CAMREX wireless camera in range.--There is excessive interference. If so:Near the bottom of the handpiece on the camera side is a smallopening exposing two small switches. Using apin, lightly and carefully arrange the switches according to the desired frequency as shown:

Page 14: Intra Oral Camera Camrex 192
Page 15: Intra Oral Camera Camrex 192

Fit a disposable sleeve over the camera handpiece for each new patient.

Press the handpiece POWER button. The LEDs around the lens should light and live video should

appear on the monitor.

Page 16: Intra Oral Camera Camrex 192

To freeze an image, press the FREEZE button. The monitor will show a still image and a white storage channel number.

To store the image and continue on another channel with live video, press the CHANNEL button.

Up to four images can be stored. If you do not wish to store the frozen image, press FREEZE

again to return to live video.

Page 17: Intra Oral Camera Camrex 192

To view stored images press CHANNEL to cycle throughthe four channels and a fifth screen showing all four at once.

Stored images are marked with a yellow channel number. To replace a stored image, advance to the image you no longer need and

capture a new image as in steps 3, 4 and 5. To delete all tored images, press and hold CHANNEL for two full seconds. Replacing or deleting images requires that the handpiece power be ON.

Page 18: Intra Oral Camera Camrex 192

1. Built-in wireless receiver with 4 separate frequencies for optimal signal reception.

2. Video output connects directly to TV or computer monitor through AV or VGA input.

3. CCD technology. Four white LED to provide maximum realistic color reproduction.

4. Mini-head design allows for easy access to the third molar.

5. Simultaneously view all 4 stored images in quadrant format or individual image full screen.

Page 19: Intra Oral Camera Camrex 192

6. Easy and convenient battery replacement.7. Large capacity lithium battery provides 3 hours of

continuous use on a full charge for 6 hours.8. Low battery warning light reminder.9. Different frequency setting allows to operate multiple

sets in the same clinic without interference. 10. Image transmission up to 5M diameter range without

blockage.

Page 20: Intra Oral Camera Camrex 192

Quadrant “Freeze” images can be stored in memory.(One image per channel.)

To simultaneously view all four store images in quadrant format, press the channel to display.

Page 21: Intra Oral Camera Camrex 192

To View the Stored Images: Depress the Channel Select Button. The first Image stored will appear on screen with a yellow channel number that Indicates it is a “Stored Image”.

To Delete all Stored Images: Depress the Channel Channel Select Button Select Button and hold for 22 seconds.

The camera will return to “Live” Mode when all stored images have been erased.

Page 22: Intra Oral Camera Camrex 192

NOTE: If the camera is turned “On” while viewing and a

“Live” image appears on screen, the channel is free for use.

If the camera is turned “Off” while viewing stored images and no image appears on screen, the channel is free for use.

Page 23: Intra Oral Camera Camrex 192

FAQFAQ

Page 24: Intra Oral Camera Camrex 192

Q1Q1

What’s the zoom lens’ function and how to clean it?

Page 25: Intra Oral Camera Camrex 192

A1A1 The zoom lens is special designed for Dentamerica

intraoral cameras and slides easily on and off the camera.

50x magnifying power. Use it to magnifying and examine caries, root canal access etc. You can also attach the surface viewer ring to examine occlusal surfaces.

It can be chemically sterilized not autoclavable.

Page 26: Intra Oral Camera Camrex 192

Q2Q2

How can I store images to a computer?

Page 27: Intra Oral Camera Camrex 192

A2A2 Optional USB video grabber enables the Camrex

to connect to the computer via the USB port and allow you to operate in the professional Camrex imaging software.

CAMREX 195 USB1.1 VISTA compatible

CAMREX 196 USB2.0 Window XP or below compatible

Page 28: Intra Oral Camera Camrex 192
Page 29: Intra Oral Camera Camrex 192
Page 30: Intra Oral Camera Camrex 192

Q3Q3 What to do if the image suddenly disappears?

Page 31: Intra Oral Camera Camrex 192

Q3Q3 Make sure the handpiece and the receiver are set to

the same frequency. Adjust the antenna. Change the receiver box location and adjust the

reception angle.

Page 32: Intra Oral Camera Camrex 192

For More Details :

Call On : +91-9867496337

Email : [email protected]

Visit : www.chromadent.com