+ All Categories
Home > Documents > K001EN-B

K001EN-B

Date post: 02-Jun-2018
Category:
Upload: m-abdullah-reza-dh
View: 213 times
Download: 0 times
Share this document with a friend

of 122

Transcript
  • 8/11/2019 K001EN-B

    1/122

    2003-03

    K001en-B

    MITRASTransmissometer

    Runway Visual Range

    Advanced Training Material

  • 8/11/2019 K001EN-B

    2/122

    2003-03

    K001en-B

    Contents

    Mitras Overview

    System description

    RVR assessmentFunctional Description

    Operating principle

    Mitras VisibilityCalculation

    Hardware

    LM11

    Installation

    Operation

    Communication

    Commands

    Troubleshooting

    Maintenance

    Periodic Maintenance

    Operational Check

    Calibration andLinearization Check

    Aligning Optics

    Replacing a Subassembly

    HANDS-ONHANDS-ON exercises

    Commands and communication

    Maintenance

    Troubleshooting and repair

  • 8/11/2019 K001EN-B

    3/122

    2003-03

    K001en-B

    MITRAS Transmissometer

  • 8/11/2019 K001EN-B

    4/122

    2003-03

    K001en-B

    MITRAS System

    Conforms to all ICAO and WMO requirements for RVRand Meteorological Optical Range (MOR)

    Accurate RVR assessment is a result of accurate MORmeasurement => reduces airport down-time

    Improved accuracy with fully automatic measurementand unique contamination control with compensation(longer maintenance interval)

  • 8/11/2019 K001EN-B

    5/122

    2003-03

    K001en-B

    Single Baseline System

    Double Baseline System

    LIGHT TRANSMITTERLIGHT RECEIVER

    LIGHT TRANSMITTER LIGHT RECEIVER

    LIGHT RECEIVER

    Light transmitter and Light receiver

  • 8/11/2019 K001EN-B

    6/122

    2003-03

    K001en-B

    Double baseline for CAT IIIB

    Double baseline for CAT IIIB is implemented by means oftwo different beams from one transmitter.

    The assessment of visibility ranges from 2/3 times theshort baseline up to 50 times long baseline.

    With 10m and 200m base lines visibility from 7m upto 10 000m.

  • 8/11/2019 K001EN-B

    7/122

    2003-03

    K001en-B

    Definition of MOR

    MOR can be measured by measuring the attenuation of light

    attenuation is mainly caused by scattering, to a smalldegree also by absorption (in smoke, dust, )

    100 %intensity 5 %intensity

    MOR

    MOR is defined as the distance where the intensity of a light

    beam has been attenuated to 5% of the original intensity

  • 8/11/2019 K001EN-B

    8/122

    2003-03

    K001en-B

    What is Runway Visual Range?

    Runway Visual Range (RVR) = Range over which the pilot on thecenter line of the runway can see:

    - Runway markings, or- Edge lights of the runway, or- Center line lights.

    RVR is not an observation nor a measurement.

    RVR assessment calculation follows Allards law which takes intoaccount:

    - Visibility (m)- Background luminance (cd/m2)- Airfield lighting intensity percentage- Airfield lighting characteristics (cd)

    For RVR calculations 5 m is used as the average eye level of a pilot inan aircraft.

    Traditionally RVR has been assessed by counting edge light lamps bythe runway. The lamps are at 60 m distance from each other. At CAT Iairport at least 14 lamps should be seen and 7 lamps at CAT II airport.

  • 8/11/2019 K001EN-B

    9/122

    2003-03

    K001en-B

    RVR Assessment

    MOR (Visibility)

    MITRAS

    Backgroundluminance (LM11)

    => Illuminationtreshold

    Runway lightsetting

    RVR Calculation

    MIDAS IVRVR value

  • 8/11/2019 K001EN-B

    10/122

    2003-03

    K001en-B

    GATEGORY DECISION HEIGHT RVR

    CAT I 60 m (200 ft) 550 m or vis. 800m

    CAT II 30 m (100 ft) 350 m

    CAT IIIA

  • 8/11/2019 K001EN-B

    11/122

    2003-03

    K001en-B

    Assessment of Runway Visual Range 2/3

    Take-off limits in all categories (from ICAO / DOC9365, 2nd edition 1991)

    Commonly acceptable take-off minima is 175m. When the facilities are: runway edge ligths, runway centre line

    ligths, centre line markings and touchdown, mid-point (ifmissing 350m) and stop-end RVR.

    Otherwise take-off minima is 500m. When the facilities are: runway edge ligths and either centre line

    ligths or centre line markings.

    VFR flights (VFR = Visual Flight Rules) require (Annex 2, 9thedition 1990)

    visibility > 5000 m

    clouds > 450 m

  • 8/11/2019 K001EN-B

    12/122

    2003-03

    K001en-B

    RVR / Visibility instrumentation for different airportcategories are summarized in table.

    At CAT II and III the RVR instrumentation isrequirement and at CAT I recommendation (Annex 3,14thEdition July2001).

    Assessment of Runway Visual Range 3/3

    CAT I CAT II CAT IIIA CAT IIIB CAT IIIC

    RV lights edge edge+center as CAT II as CAT II + TWY as CAT IIIB

    Instruments 1 2-3 3 3 3

  • 8/11/2019 K001EN-B

    13/122

    2003-03

    K001en-B

    Annex 3, 14thEdition July2001:

    Reporting steps of 25 m, when RVR < 400 m

    Reporting steps of 50 m, when 400 RVR 800 m

    Reporting steps of 100 m, when > 800 m

    Visibility rounded down to steps in METAR:

    50 m, when visibility < 800 m

    100 m, when visibility between 800 m and 5000 1000 m, when visibility between 5000 m and 10000

    reported as 10km when visibility is 10 km and over exceptwhen the conditions for the use of CAVOK apply.

    Reporting steps

  • 8/11/2019 K001EN-B

    14/122

    2003-03

    K001en-B

    R U N W A YR U N W A YR U N W A Y

    ATC/TOWER

    Digital display

    M

    M

    RVR Computer and

    accessories

    MIDAS IV: Small system

  • 8/11/2019 K001EN-B

    15/122

    2003-03

    K001en-B

    R U N W A Y

    CEILOMETER

    CT25K

    0...25000 ft

    Precipitation

    Geonor

    R U N W A Y

    RWY TEMPRWY TEMPRWY TEMP

    RWY TEMP RWY TEMP RWY TEMP

    WINDS

    PTU

    PTU

    WINDS

    MIDAS IV: Large system

  • 8/11/2019 K001EN-B

    16/122

    2003-03

    K001en-B

    MAINTENANCE

    WORKSTATION

    (RCM)

    OBSERVER

    WORKSTATION

    (OWS)

    CENTRAL DATA

    UNITS (CDUs)

    FIELD SENSORS

    FORECASTER

    WORKSTATION

    (FWS)

    WEATHER VIEW

    WORKSTATION

    (WV)

    SWITCH OVER

    UNIT

    DATABASE

    SERVER

    UPS

    DIGITAL

    DISPLAYS

    AFTN

    INTERFACE

    SERIAL

    OUTPUT

    RUNWAYLIGHT

    SETTING

    RVR_08

    ZCZC.. METAR ENGM ...

    1% 3% 10% 30%

    LOCAL AREA NETWORK

    ATIS

    MIDAS IV: Large system

  • 8/11/2019 K001EN-B

    17/122

    2003-03

    K001en-B

    MITRAS Operating Principle

  • 8/11/2019 K001EN-B

    18/122

    2003-03

    K001en-B

    Measurement principle (1)

    Light source

    - Xenon flashlamp

    - Represents the spectrums of sunlight and runway lights

    - Very stable, expected lifetime 55000 h

    - Light intensity can be controlled

    Benefits of reference measurement (transmitted intensity)- Short-term long-term variations are compensated

    very accurately

    - Transmissometer stability better than 0.3%

    - Unusual situation detected immediately

  • 8/11/2019 K001EN-B

    19/122

    2003-03

    K001en-B

    Optics of MITRAS Transmissometer

  • 8/11/2019 K001EN-B

    20/122

    2003-03

    K001en-B

    Measurement principle (2)

    Averaging time

    - Normal operation 30 s or 60 s

    - In alignment mode 15 s

    Pulse measurement principle

    - Light pulse duration 1.5 s (half-width)

    - Continuous light doesnt interfere: A 75000W lamp at 30 m

    distance has no influence Representative spectral response

    - Green optical filters are used to approximate the specralresponse of the human eye

    Economy mode The flashing interval automatically changes according to the

    visibility.

    M i l d

  • 8/11/2019 K001EN-B

    21/122

    2003-03

    K001en-B

    Measurement interval andnormal/economy mode

    Transmissometer measures the transmittance between theLight Transmitter LP11 and Light Receiver LR11. It calculatesthe moving average over 60 or 30 samples.

    In normal mode it measures transmittance every second. In economy mode it automatically changes the flashing

    interval according to the visibility.

    Visibility (MOR) Flashing interval

    under 2 km 1 s

    2 km

  • 8/11/2019 K001EN-B

    22/122

    2003-03

    K001en-B

    LP11 / LR11 Block Diagram

    Comparison of Performance

  • 8/11/2019 K001EN-B

    23/122

    2003-03

    K001en-B

    Comparison of PerformanceTransmissometer vs. Forward Scattering

    TRANSMISSOMETER

    - large sample volume- limited dynamic range

    - accuracy maintained in all types of precipitation- tedious installation- continuous maintenance and calibration

    FORWARD SCATTERING

    - small sample volume- large dynamic range

    - precipitation may cause inaccurate results (unless the measurement is compensated)- easy installation- less maintenance

  • 8/11/2019 K001EN-B

    24/122

    2003-03

    K001en-B

    MITRAS Visibility Calculations

  • 8/11/2019 K001EN-B

    25/122

    2003-03

    K001en-B

    Meteorological Optical Range (MOR) has been defined by

    the WMO as the basic parameter to express the optical state

    of the atmosphere

    MOR corresponds closely with human visibility

    observations (day observations)

    MOR has been defined as a purely physical quantity

    objective

    it can be measured

    Visibility & MOR

  • 8/11/2019 K001EN-B

    26/122

    2003-03

    K001en-B

    100 %intensity

    5 %intensity

    MOR

    MOR can be measured by measuring the attenuation of light attenuation is caused by scattering and absorption

    Definition of MOR

  • 8/11/2019 K001EN-B

    27/122

    2003-03

    K001en-B

    Attenuation of Light

    Law of Bouguer- Lambert:

    F F ex

    =

    0

    Definition of MOR by WMO

    MOR can be calculated from extinction coefficient

    F = flux of light (at distance x)

    Fo = original flux of light (x = 0)

    x = distance = extinction coefficient (= a + b)

    where a = absorption coefficient

    b = scattering coefficient

    F/F0 = 5% = 0.05

    F/F0 = e-x0.05 = e-x

    3ln(0.05)

    -=MORx =

  • 8/11/2019 K001EN-B

    28/122

    2003-03

    K001en-B

    Attenuation of Light

    Light fog

    Heavyfog (large )

    Clear (small )

    Fog

    Distance x

    Light

    flux

    F0= 100%

    F = 5%

    F= F0e- x

  • 8/11/2019 K001EN-B

    29/122

    2003-03

    K001en-B

    35 m 50 m 75 m 100 m

    Transmittance Visibilities (m) at different baselines

    0.995 20.947 29.925 44.887 59.8500.990 10.447 14.925 22.387 29.8500.950 2.047 2.924 4.387 5.8490.930 1.447 2.067 3.100 4.1340.900 997 1.424 2.135 2.8470.800 470 672 1.008 1.3440.600 206 294 440 5870.400 115 164 246 3270.200 65 93 140 1860.100 46 65 98 1300.015 25 36 54 71

    Meteorological Optical Range (MOR)

    MOR = - 3 x Baseline

    ln (Transmittance)

    Baseline

  • 8/11/2019 K001EN-B

    30/122

    2003-03

    K001en-B

    Visibility Calculation

    LPC11 software calculates visibility with Koschmieders law.

    ICAO has defined the extinction coefficient to be 0.05 which has beentaken into account in the following Koschmieders visibility (MOR)formula:

    Mitras transmissometer system uses LPI11 Light Intensity measurement units in

    LP11 and LR11 for measuring light.

    When calculating the visibility the software takes into account intensities with and

    without contamination calculation. If Contaminantion Compensation Mode is ON

    MITRAS transmits the compensated visibility to RVR Computer.

    If the LPI11 boards are unlinear the software takes that into account. (The

    linearization correction is made by techicians).

    Where Transmittance = K1 Received light intensity = K1 R (I)

    Transmitted light intensity TM (I)

    and K1 is set by technicians during good visibility

    Visibility = - 3 x Baseline ln (Transmittance)

  • 8/11/2019 K001EN-B

    31/122

    2003-03

    K001en-B

    Visibility Measuring Range

    Single Baseline Systems

    Baseline MOR range RVR range(at night)

    35 m 25....1500 m 100....1500 m

    50 m 40....2000 m 150....2000 m

    75 m 50....3000 m 200....3000 m

    100 m 70....4000 m 300....4000 m

    200 m 70..10000 m 300..10000 m

    Double Baseline Systems

    Baseline MOR range RVR range

    10 and 75 m 7....3000 m 40....3000 m

    10 and 200 m 7..10000 m 40..10000 m

  • 8/11/2019 K001EN-B

    32/122

    2003-03

    K001en-B

    Calculations

    Transmittance calculation

    Transmittance is the relationship between the received and transmitted light power. The transmitted light pulseintensity is measured for every pulse by the transmitter and the corresponding received power by the receiverelectronics. The transmitter asks for the received intensity and sends the flash intensity.Thus, both the receiver and the transmitter can perform the same calculations.

    Because of the electronics the scaling of the light intensity meter is not the same in the receiver and in thetransmitter. Therefore, at least linear scaling is needed to convert the intensity relation to a transmittance value.Besides linearization, scaling is sometimes needed to compensate the non-linearities in the measuringelectronics. These scaling factors are defined in the calibration procedure. Two calibration points are included inthe calculation. I.e. three linear scales are used to estimate the transmissometer response.An additional scale (K1) is used to adjust the original lines.The calibrated linear scales are called K2, K3 and K4.

    During every measurement cycle, the "raw transmittance" is calculated as follows:

    RAWinst = (R(I)-R(offset)) / (TM(I)-TM(offset)),

    T(RAW) = AVE(RAWinst(30)) or AVE(RAWinst(60))

  • 8/11/2019 K001EN-B

    33/122

    2003-03

    K001en-B

    Calculations

    Visibility calculation, non-compensated

    The Meteorological Optical Range, MOR or the visibility V is calculated using a contrast threshold 0.05recommended by the ICAO and the WMO.

    where B = the effective transmissometer baseline

    T ( NOC ) = non-compensated transmittance

    Contamination compensated visibility calculation

    For the contamination compensated transmittance correction factors TM(WT) (transmitter window) and R(WT)(receiver window) are needed.

    where R(WT) can be the transmittance correction of either long or short baseline receiver window.

    3 x B

    -ln T( NOC )V =

    T( NOC)

    TM( WT) x R( WT)T ( COM ) =

  • 8/11/2019 K001EN-B

    34/122

    2003-03

    K001en-B

    Calculations

    Transmittance calculation

    These instant values are collected in a 60 location rotating buffer.The latest 30 or 60 samples are used to calculate the average value T(RAW).The two lowest and two highest sample values are excluded from the average.The purpose of the correction is to filter out extreme instant transmittance values,

    for instance a bird's flight through the optical path.

    T(AVE) = K1*K1A*(K2,K3,K4*RAW)

    The autocalibration feature uses the scaling parameter K1A initiated to 1.00 by the CAL command.If then the visibility estimation has been too optimistic, i.e. one minute average of the measured transmittance

    exceeds 0.998, K1A reduces from 1.00. K1A is calculated as 0.998/measured transmittance.If K1A < 0.95, the CALIBRATION FAILURE bit is set in the status.However, the calculated correction is used.

    K2 is used below the lower calibration point (P1), K3 between the calibration points and K4 above the highercalibration point (P2).

  • 8/11/2019 K001EN-B

    35/122

    2003-03

    K001en-B

    Calculations

    Contamination compensated visibility calculation

    Contamination compensated visibility V ( COM ) is calculated by using the T ( COM ) :

    The contamination compensated visibility is sent to the data processing unit, when

    - the contamination compensation mode is ON

    - the window transmittance is better than 0,92 (user selectable) for light receiver window and 0,92 for light transmitter window (compared with clean windows). When transmittances of the windows are

    lower than these values, the windows are considered dirty.

    The non-compensated visibility value is sent to the data processing unit, when

    - the contamination compensation mode is OFF

    - the contamination compensated visibility value is not available.

    Contamination compensation calculations

    The contamination signal is measured every 10 seconds and average of six latest samples ( one minute ) is usedin the contamination calculation. The MES 4 tables are updated every minute with these one minute values.

    The scaling is initiated with the CLEAN command after cleaning the windows.

    3x B

    - lnT (COM)V ( COM ) =

  • 8/11/2019 K001EN-B

    36/122

    2003-03

    K001en-B

    RVR Computer Calculations

    RVR Computer calculates RVR data every 15 s. It pollsMITRAS transmissometers for visibility, backgroundluminance and status data. It receives runway light

    percentage data from the interface unit or uses manuallyinput value. It has the characteristics of airfield lights inits memory.

    When calculating the RVR the RVR Computer takes into

    account marked discontinuity. It calculates instant RVRand following 10 minute values: average, one-minuteminimum and one-minute maximum.

    It compares last 5 minute RVR values to those for 5 to 10

    minutes earlier and calculates tendency (U=up, D=downand N=neutral).

    RVR C C l l i

  • 8/11/2019 K001EN-B

    37/122

    2003-03

    K001en-B

    RVR Computer Calculations

    MIDAS IV RVR compoter transmits RVR values to theDD50 digital displays, printers and operators display.

    The transmission to the DD50 includes instant RVR dataand the 1 and 10 minute RVR values with following rules:

    - average if the one-minute minimum and maximum doesdiffers from the average RVR less than 20% or is

  • 8/11/2019 K001EN-B

    38/122

    2003-03

    K001en-B

    Hardware

    LP11 Bl k Di

  • 8/11/2019 K001EN-B

    39/122

    2003-03

    K001en-B

    LP11 Block Diagram

    LR11 Bl k Di

  • 8/11/2019 K001EN-B

    40/122

    2003-03

    K001en-B

    LR11 Block Diagram

    LPC11 Transmissometer CPU

  • 8/11/2019 K001EN-B

    41/122

    2003-03

    K001en-B

    LPC11 Transmissometer CPU

    LPC11 has

    - Intel 8031 CPU chip with 7 MHz clock, 512 kb EPROM, 64 kb RAMmaintenance terminal line.

    - DUART for modem line (DMX21), current loop (LPT11), frequencycounting (LMB11), peak reset and power error control.

    - Analog multiplexer and 12 A/D converter.

    - D/A converter.

    LPC11 is used in Light Transmitter LP11 andLight Receiver LR11

    - same software

    - different jumper setting

    Blinking LEDs indicate as follows:- Green LED blinking at 1 s interval (normal).

    - Red LED indicates Reset, or an error found by star up self-test, orwatchdog circuit has given a hardware reset.

    - Yellow LED blinks when LPC11 triggers the flash lamp to flash.

    LPC11 Processor Block Diagram

  • 8/11/2019 K001EN-B

    42/122

    2003-03

    K001en-B

    LPC11 Processor Block Diagram

    LPC11 Jumper Setting

  • 8/11/2019 K001EN-B

    43/122

    2003-03

    K001en-B

    LPC11 Jumper Setting

    LPC11 Software v3 27 operation

  • 8/11/2019 K001EN-B

    44/122

    2003-03

    K001en-B

    LPC11 Software v3.27 operation

    LPC11 Software v3.27 is used in Light Transmitter LP11 andLight Receiver LR11.

    - same software operates for single and double baseline systems- different jumper setting at LP11 and LR11 CPU boards

    Software has following modules:- Built in startup tests

    - Measurement of transmitted and received light intensity

    - Visibility calculation and contamination compensation calculation

    - Message generation

    - Communication between LP11 and LR11, LP11 and RVRComputer, LP11/LR11 and maintenance terminal

    - Contamination signal measurement, calculation and warning

    - Monitoring temperature, flash lamp intensity and contaminationsignals

    - Warning generation if monitored signals are not within limits

    - Setting reference signal levels when the optics has been cleaned

    LPF11 Flash Control Unit

  • 8/11/2019 K001EN-B

    45/122

    2003-03

    K001en-B

    LPF11 Flash Control Unit

    The inputs are High voltage (750 VAC) and logic signals:

    - triggering pulse for Xenon flash lamp

    - flash voltage signal for controlling high voltage

    Intensity control with Flash Gain, charged energyE = CU where C = capacitance and

    U = charging high voltage (300-800 V)

    Xenon flash lamp is a stable light source. Light pulse duration

    is 1.5 micro sec. The arc is bored between 2 electrodes mounted inside a glass

    bulb.

    Triggering socket is a potted assembly. It is connected to the

    Flash Control Unit with 4 leads. Flashing interval is 1 s normal mode and in economy mode

    1 s to 10 s.

    LPF11 Circuit Diagram

  • 8/11/2019 K001EN-B

    46/122

    2003-03

    K001en-B

    LPF11 Circuit Diagram

    LPH11 Heating Control Board

  • 8/11/2019 K001EN-B

    47/122

    2003-03

    K001en-B

    LPH11 Heating Control Board

    Controls the heating of the internal optics and protectionwindow.

    The electronics of optics operates best in stable

    temperature. The operating temperature range is definedwith a jumper on the LPH11.

    The heating of protection windows prevents surfacesfrom condensating water.

    The green led light shows that window heating is on.

    LPH11 Circuit Diagram

  • 8/11/2019 K001EN-B

    48/122

    2003-03

    K001en-B

    LPH11 Circuit Diagram

    LPH 11 Jumper Setting

  • 8/11/2019 K001EN-B

    49/122

    2003-03

    K001en-B

    LPH 11 Jumper Setting

    Contamination Detector

  • 8/11/2019 K001EN-B

    50/122

    2003-03

    K001en-B

    Contamination Detector

    DMX21 FSK Modem

  • 8/11/2019 K001EN-B

    51/122

    2003-03

    K001en-B

    DMX21 FSK Modem

    MITRAS communicates to the RVR Computer via 300 bpsleased modem line. Line speed 300 bps, Even parity, 7 data bits and 1 stop bit

    The DMX21 operates in ANSWER mode and its switchesshould be Up, Middle and Down.

    The LEDs indicate polling (RXD), answering (TXD), and Carrierdetected.

    Troubleshooting - Cover theLEDs with hands in order to see how the LEDs are blinking.

    - Check that RXD LEDblinks every 15 s when RVRComputer polls

    it and the MITRASanswers with longer message (TXD LEDs blink).- Check that the green LED on the LPC11 CPU board blinks every

    second. If not reset the MITRAS light transmitter LP11.

    - Turn off power. Remowe the DMX21 and check the jumper

    setting. Restart the LP11 again.- Check cables and their connections.

    13100LP Cabling Diagram

  • 8/11/2019 K001EN-B

    52/122

    2003-03

    K001en-B

    13100LP Cabling Diagram

    12930LP Wiring Diagram

  • 8/11/2019 K001EN-B

    53/122

    2003-03

    K001en-B

    g g

  • 8/11/2019 K001EN-B

    54/122

    2003-03

    K001en-B

    LM11 Background Luminance Meter

    LM11 Background luminance meter

  • 8/11/2019 K001EN-B

    55/122

    2003-03

    K001en-B

    g

    Connected to the MITRAS LP11 transmitter

    Measurement range: 4 30.000 cd/m2

    Spectral response range: 300700 nm, peak = 550nmInternal heating to avoid condensation

    Cable totransmitter

    Mechanicalinterface

    Tilt adustment

    Optics

    LM11 Background luminance meter

  • 8/11/2019 K001EN-B

    56/122

    2003-03

    K001en-B

    g

    MEASUREMENT PRINCIPLE

    LM11 measures the amount of ligth coming from a 7ocone to

    the 46 mm lens.incoming ligth is focused on the matte glass

    matte glass will prevent from increasing the luminance

    values caused by bright spotsmatte glass provides a smooth average response

    The ligth spectrum is detected by PIN photodiodegreen filtered

    photoelectric current is converted into frequency which

    linearly corresponds to cd/m2divided by 10.

    LM11 Background luminance meter

  • 8/11/2019 K001EN-B

    57/122

    2003-03

    K001en-B

    g

    Internal heating to avoid condensation

    body and lens heated

    monitored by two thermistors

    INSTALLATION

    The meter should be directed towards the North sky

    Direct sunligth into meter should be avoided

    The meter is normally tilted upwards to the horizontal level

    at angle of 20oto 50o

    Recommended installation angle of 30o

    Directed sligthly above the runway ligths

    Background Luminance Meter LM11

  • 8/11/2019 K001EN-B

    58/122

    2003-03

    K001en-B

    Side view: 1 Housing2 Back plate3 Aluminium body4 Hood5 Power transistor6 Cable with connector7 Matte glass

    8 Green filter9 PIN photodiode10 Lens11 Front plate12 O-ring13 Rubber gasket14 M4 screws

  • 8/11/2019 K001EN-B

    59/122

    2003-03

    K001en-B

    Installation

    Selecting location

  • 8/11/2019 K001EN-B

    60/122

    2003-03

    K001en-B

    Transmissometers should be located in position whichensure representative measurements for intended purpose(see references from Users Guide).

    Locations at airport (according to ICAO Annex 3):

    CAT I runways

    One transmissometer near the touchdown zone, 50mbaseline recommended

    CAT II runways

    Less than 2400m long: Two transmissometers

    More than 2400m long: Three transmissometers

    CAT III runways

    Three transmissometers, double baseline systems with 10m

    and 75m baselines recommended

    Sensor Locations at Airport

  • 8/11/2019 K001EN-B

    61/122

    2003-03

    K001en-B

    Transmissometer foundation recommendation

  • 8/11/2019 K001EN-B

    62/122

    2003-03

    K001en-B

    Base plate set assembly & installation

  • 8/11/2019 K001EN-B

    63/122

    2003-03

    K001en-B

    LP11 Cabling and earthing recommendation

  • 8/11/2019 K001EN-B

    64/122

    2003-03

    K001en-B

    Cabling diagram for single baseline

  • 8/11/2019 K001EN-B

    65/122

    2003-03

    K001en-B

    Service platform principal dimensions

  • 8/11/2019 K001EN-B

    66/122

    2003-03

    K001en-B

  • 8/11/2019 K001EN-B

    67/122

    2003-03

    K001en-B

    Communication

    I/O Lines

  • 8/11/2019 K001EN-B

    68/122

    2003-03

    K001en-B

    Output interfacesBoth the light transmitter and the light receiver have three separate serial lines which are used as follows:

    Light transmitter

    (Line 0) Maintenance line (CPU) RS-232C

    (Line 1) to RVR computer (DUART B) modem or MITIF

    (Line 2) to light receiver (DUART A) current loop

    Light receiver

    (Line 0) Maintenance line (CPU) RS-232C

    (Line 1) Auxiliary (DUART B) (RS-232C)

    (Line 2) to light transmitter (DUART A) current loop

    Maintenance line

    The maintenance line is used for monitoring, testing and calibration purposes. Maintenance line commands aredescribed in title "Commands".

    Line parameters:

    300 baud, 7 data bits, 1 stop bit, even parity

    Communication protocol:

    - Conversation mode, character by character

    - XON/XOFF handshake

    - Timeout 60 seconds

    Transmissometer Message

  • 8/11/2019 K001EN-B

    69/122

    2003-03

    K001en-B

    where

    P, ID, V, B, S constant characters

    X variable value

    _ ASCII character 20 hex (Space)

    ASCII character 0C hex (Carriage return)

    ASCII character 02 hex (Start of text)

    ASCII character 03 hex (End of text)

    ASCII character 0A hex (Line feed)

    Status of receiver 2 is set spaces if a single baseline system is used.

    Status of each equipment consists of two HEX-ASCII characters.

    ID_X_V_XXXXXX_B_XXXXX_SXXXXXX_

    Status of receiver 2 (VI)

    Status of receiver 2 (V)

    Status of receiver 1 (IV)

    Status of receiver 1 (III)

    Status of transmitter (II)Status of transmitter (I)

    Background luminance

    Visibility

    Transmitter ID (0...7)

    Status

  • 8/11/2019 K001EN-B

    70/122

    2003-03

    K001en-B

    Transmitter status:

    BIT.0 = 1 Meas mode (0 = OFF) (1 = ON)

    BIT.1 = 2 Cont/other (0 = OFF) (1 = ON)

    BIT.2 = 4 Optical surface (0 = OK) (1 = DIRTY)

    BIT.3 = 8 Power supply (0 = OK) (1 = FAIL)

    BIT.4 = 1 Heating (0 = OK) (1 = FAIL)

    BIT.5 = 2 Flash lamp (0 = OK) (1 = WEAK)

    BIT.6 = 4 BL meter (0 = OFF) (1 = ON) 1)

    BIT.7 = 8 Measurement loop (0 = OK) (1 = FAIL)

    Receiver status:

    BIT.0 = 1 Meas mode (0 = OFF) (1 = ON)

    BIT.1 = 2 Cont/other (0 = OFF) (1 = FAIL)

    BIT.2 = 4 Optical surface (0 = OK) (1 = DIRTY)

    BIT.3 = 8 Power supply (0 = OK) (1 = FAIL)

    BIT.4 = 1 Heating (0 = OK) (1 = FAIL)

    BIT.5 = 2 Calibration (0 = OK) (1 = FAIL)

    BIT.6 = 4 Test (0 = OK) (1 = FAIL)

    BIT.7 = 8 Consistency (0 = OK) (1 = FAIL) 2)

    1)ON if the background luminance (BL) meter is set on with the command MODE2)Only in double baseline system

    I

    II

    IV

    (VI)

    III

    (V)

    Operation and Maintenance Connection

  • 8/11/2019 K001EN-B

    71/122

    2003-03

    K001en-B

    Modem line to data processing unitNormally, the data processing unit uses this line for message polling from the MITRAS light transmitter(s) but onecan also use it for maintenance purposes similar to the MITRAS maintenance line.

    Line parameters:

    - 300 baud, 7 data bits, 1 stop bit, even parity

    - Optional 1200 baud for RS-232 interface (Not used)

    Communication protocol:

    - Data packet protocol

    - Data validity confirmed with character parity check

    - Timeout 60 seconds

    Commands

    The MITRAS Transmissometer software has operator's commands for monitoring, testing and calibrating the unit.Commands can be given via both the maintenance line and the modem line,i.e. computer/auxiliary line.

    NOTE: The line is automatically closed after 60 seconds, if no commands are given.

    If you want to keep the maintenance line open, give a space as input. After a space mark, you can later give thecommand you want.

    All commands or inputs are entered for processing by pressing ENTER or CARRIAGE RETURN. In some cases thecommand can be followed by a space and the desired parameter value(s).

    OPEN Opens the line for operator commands.

    OPEN for modem line.

    HELP Outputs a list of commands.

  • 8/11/2019 K001EN-B

    72/122

    2003-03

    K001en-B

    Commands

    Operation and Maintenance Commands

  • 8/11/2019 K001EN-B

    73/122

    2003-03

    K001en-B

    The transmissometer command set consists of six different types of commands:

    1. Transmissometer initializing commands for start-up

    - not normally used after start-up period

    2. Transmissometer initializing commands for operational use

    - can be used during operational use

    3. Message commands

    - can be used for maintenance and test purposes and also for data output

    4. Mode selecting commands

    - aimed at mode altering during operational use

    - effects of mode selecting commands must be familiar to the operator

    before use

    5. Routine maintenance commands

    - used to maintain accurate operation of the transmissometer

    - some of these commands require that the user is authorized to changeparameters, e.g. to calibrate according to visual observation

    6. Special maintenance commands

    - needed only when troubleshooting failures or when testing specialfunctions of the transmissometer.

    Initializing Commands for Startup

  • 8/11/2019 K001EN-B

    74/122

    2003-03

    K001en-B

    CONF With this command you may initialize (and set) the system parameters.Usually this command is used during installation only.

    The system response to CONF command is presented below:

    >CONF

    CONFIGURATION STARTED 0000-00-00 00:44:51

    MITRAS SOFTWARE VERSION V3.25 2000-04-19

    TRANSMITTER ID 1

    SINGLE BASELINE

    MEAS MODE ON

    ECOM MODE ON

    AUTO CALIBRATION ON

    JUMPER FIELD 00100000

    SELECT OPERATION MODES

    MEAS MODE (1=ON, 0=OFF) 1

    ECON MODE (1=ON, 0=OFF) 1

    CONT MODE (1=ON, 0=OFF) 1

    AUTO CALIBRATION (1=ON, 0=OFF) 1

    BL METER (1=ON, 0=OFF) 1

    AVE COUNT (1=60, 0=30) 1

    Initializing Commands for Startup cont.

  • 8/11/2019 K001EN-B

    75/122

    2003-03

    K001en-B

    SET THE BASELINE

    BASELINE 49.40

    INITIALIAZE SCALING FACTORS

    K1 0.5807

    K2 1.0000

    K3 1.0000K4 1.0000

    TRANSMITTER OFFSET (mV) 0.0

    RECEIVER OFFSET (mV) 0.0

    BACKGROUND LUMINANCE SCALE KBL 10.000

    INITIALIZE CALIBRATION POINTSP 1 0.5000

    P 2 0.9900

    CURRENT FLASH GAIN 127

    AP 24.0000

    DIRTY LIMIT 0.920

    CONFIGURATION END

    Initializing Commands for Startup cont.

  • 8/11/2019 K001EN-B

    76/122

    2003-03

    K001en-B

    PAR With this command all the necessary parameters, such as transmissometer scaling

    factors, calibration points, baseline etc. can be reviewed at one glance by the operator.> PAR

    SYSTEM PARAMETERS

    TRANSMITTANCE SCALING FACTORS

    K1 0.5807

    K1A 1.0000

    K2 1.0000

    K3 1.0000

    K4 1.0000

    RECEIVER OFFSET (mV) 0.0

    CALIBRATION POINTS P1 0.50 P2 0.99

    AVERAGING 60S, CONT ON

    TRANSMITTER OFFSET (mV) 0.0

    BACKGROUND LUMINANCE SCALE KBL 10.000

    BASELINE 49.40

    FLASH GAIN 127

    CONTAMINATION SCALES AP 24.0 DIRTY LIMIT 0.920

    WINDOW SIGNAL SCALE 1.092

    SETF Sets the flashlamp intensity, may be used for calibration and maintenance purposes.

    Initializing Commands for Operational Use

  • 8/11/2019 K001EN-B

    77/122

    2003-03

    K001en-B

    OPEN Reserves the communication line and serves as password to the system.

    OPEN

    In modem line multiple transmissometers are separately connected to the line, when

    the operator is using the OPEN mode. Only the unit with correct identification (ID)

    switches the carrier on.

    CLOSE Is the counterpart of the OPEN command switching the carrier off and releasing theline. Must be used as the last command when maintenance session is finished.

    DATE With this command you can ask for the date or change it. To change the date,input a new date after the command as follows:

    >DATE 1988 02 31.

    TIME With this command you can ask for the time or change it. To change the time,

    input a new time after the command: >TIME 23 45.

    NOTE: The time and date are mainly for test purposes and not accurate. There is no battery

    back-up either.

    Initializing Commands for Operational Use cont.

  • 8/11/2019 K001EN-B

    78/122

    2003-03

    K001en-B

    HELP Displays the transmissometer commands. Some of them are not effective in the receiver.>HELP

    MITRAS TRANSMISSOMETER COMMAND SET

    MES DISPLAY MESSAGE NUMBER N

    PAR DISPLAY SYSTEM PARAMETERS

    STATUS DISPLAY STATUS MESSAGE

    CLEAN INIT CONTAMINATION MEASUREMENT

    CONF INIT CONFIGURATION

    CAL CALIBRATION ADJUST

    FCAL TWO POINT FILTER CALIBRATION

    CHECK CALIBRATION CHECK

    MODE SET OPERATION MODESSETF SET FLASH LAMP GAIN

    DMES RAW DATA MESSAGE

    OPEN RESERVE THE COMMUNICATION LINE

    CLOSE RELEASE THE COMMUNICATION LINE

    ALIG ALIGNMENT HELP

    TCON CONTAMINATION TEST

    AMES AUTOMATIC MESSAGE

    FRE BL FREQUENCY INPUT TEST

    AN N ANALOG INPUT TEST, CHANNEL N

    RESET RESET COMMAND

    TIME

  • 8/11/2019 K001EN-B

    79/122

    2003-03

    K001en-B

    The operator may give message commands in the form MES or in the form MES N C T, where

    N is the message number; values for N are 0...5.

    C is used when continuous output is desired

    T is the nominal time interval between output, when certain output

    interval is desired in seconds. T may range from 1 to 59 seconds.

    MES 1 C 10 Gives message 1 with a 10 seconds delay between messages.

    MES 0 (or MES) Output message consists of date, time, transmissometer ID, visibility,

    contamination compensated visibility, background luminance and

    transmissometer status:

    1988-12-31 23:59:59 ID 1 V2450 CV2870 B1200 S4101

    Message Commands cont.

  • 8/11/2019 K001EN-B

    80/122

    2003-03

    K001en-B

    MES 1 Output message consists of date, time, TM(I) = light transmitter flashlamp intensity

    R(I) = light intensity from long baseline receiver

    T(RAW) = averaged raw transmittance value

    T(NOC) = non-compensated transmittance value

    V(NOC) = non-compensated visibility value

    T(COM) = contamination compensated transmittance value

    V(COM) = contamination compensated visibility value

    DATE TIME TM(I) R(I) T(RAW) T(NOC) V(NOC) T(COM) V(COM)

    1988-12-31 23:59:59 2543 3211 1.0500 0.8500 2450 0.8720 2870

    MES 2 Output message consists of date, time,

    TM(T) = light transmitter optics temperature

    TM(C) = average of scaled transmitter contamination measurements

    TM(WT) = light transmitter window transmittance

    R(T) = light receiver optics temperature

    R(C) = long baseline receiver contamination measurement

    R(WT) = light receiver window transmittance

    DATE TIME TM(T) TM(C) TM(WTP) R(T) R(C) R(WT)

    1988-02-04 12:30:59 30.8 0.9500 0.9992 31.2 0.9400 0.9985

    Message Commands cont.

  • 8/11/2019 K001EN-B

    81/122

    2003-03

    K001en-B

    MES 3 Output message consists of the transmittance measurement table,

    which includes the last 60 instant RAW values.The averaged output is a mean of the last 30 or 60 valuesof this table depending on the selected mode.

    MES 4 Output message consists of scaled contamination measurement table(s)of both the light transmitter and the light receiver(s).The last 60 one minute instant values are displayed.

    MES 5 Output message consists of visibility, compensated visibility,

    background luminance and transmissometer status.

    Message Commands cont.

  • 8/11/2019 K001EN-B

    82/122

    2003-03

    K001en-B

    Certain airport systems require an automatic message sending mode. If the automatic messagemode is not used, the RVR system polls a message from Mitras.

    AMES N T The AMES command defines the data message, which is sent as the

    automatic message or as the default polled message. The output of the

    automatic message can be seen only in normal data output (modem/MITIF),not via a maintenance line connection.

    The software versions 3.22 or later include AMES command.

    N Message number (0...6). Message -1 disables the automatic message.

    T Message time interval (1...255 seconds). If value is not given the previous

    interval setting is used.

    NOTE: The default automatic message mode setting in MIDAS RVR system is

    AMES 6 15

    Mode Selecting Commands

  • 8/11/2019 K001EN-B

    83/122

    2003-03

    K001en-B

    MODE With this command the operator can change the functional modes of the transmissometer.These functions are also included in the CONF command (configuration session).

    The mode in question is selected simply by inputting 0 or 1.

    The following modes and alternatives are available:

    MEAS MODE 1) ON - visibility measurements are made

    2) OFF - stand-by, flashlamp is not used

    ECON MODE 1) ON - during good visibility flashing rate is reduced, for low visibility

    automatically increased

    - purpose: to save flashlamp and make continuous measurement possible

    2) OFF - economy mode is not applied, flashing rate is fixed to 1 flash/second.

    CONT MODE 1) ON - contamination compensation is used

    2) OFF - no compensation is made

    BL METER 1) ON - background luminance measurement is done

    2) OFF - no background luminance measurement

    AVE COUNT 1) 60 - 60 values used for transmittance averaging

    2) 30 - 30 values used for transmittance averaging

    AUTOCALIBRATION 1) ON - switches autocalibration on

    2) OFF - no autocalibration

    Routine Maintenance Commands

  • 8/11/2019 K001EN-B

    84/122

    2003-03

    K001en-B

    HELP Displays the transmissometer commands. All of them are not effective in thelight receiver unit

    STATUS Outputs both light transmitter and light receiver(s) status information. Use

    only via data processing unit or at the light transmitter.

    DMES DMES displays the raw data message in double baseline format.

    DMES C Continues the display until ESC is pressed.

    DMES C 30 Continues the display until ESC is pressed keeping a 30 seconds' delay

    between messages.

    DMES command output

  • 8/11/2019 K001EN-B

    85/122

    2003-03

    K001en-B

    TM(I) RL(I) RS(I) TL(AVE) TS(AVE) V(L) V(S) BL C(TM) C(L) C(S)1317 1464 //// 0.9524 ////// 4615 ///// 840 0.9998 0.9983 /////

    TM(I) = flash light intensity from the A/D converter

    RL(I) = received intensity from the long baseline receiver

    RS(I) = received intensity from the short baseline receiver in double baseline systems

    TL(AVE) = scaled transmission without contamination compensation

    TS(AVE) = as TL(AVE) but from the short baseline data

    V(L) = visibility calculated from the upper limit corrected TL(NOC)

    V(S) = visibility calculated from the upper limit corrected TS(NOC)

    BL = background luminance value

    C(TM) = scaled (clean = 1) contamination value of the transmitter window

    C(L) = scaled contamination value of the long baseline receiver window

    C(S) = scaled contamination value of the short baseline receiver window

    Routine Maintenance Commands cont.

  • 8/11/2019 K001EN-B

    86/122

    2003-03

    K001en-B

    CAL The final calibration and routine corrections can be done using this command. Software readjusts the K1 scale and initiates the autocalibration factor K1A. The new K1, K1A and the corresponding transmittance are displayed. K1 should now be close to value 1.0 (if FCAL is first carried out).

    The system response to the CAL command is presented below.

    >CAL 10000

    K1(L) T(L)

    0.756 0.9778

    CALIBRATION DONE

    CLEAN Give the CLEAN command only when it is necessary. The CLEAN command sets new

    reference values for contamination measurement. The system response to the CLEAN

    command is presented below.

    >CLEAN

    CONTAMINATION SIGNAL CALIBRATION IS DONE IN FEW MINUTES

    Routine Maintenance Commands cont.

    CHECK filt l

  • 8/11/2019 K001EN-B

    87/122

    2003-03

    K001en-B

    CHECKThe calibration quality is checked with test filters and the CHECK command. The

    software displays the 'target visibility' calculated from the current visibility taking into

    account the attenuation caused by the filter in use. The software outputs the reached

    visibility (VIS) value of each second. It calculates and displays the difference from the

    filter visibility (VER) and the filtered reference transmittance (TER).

    The system response to the CHECK command is presented below:

    >CHECK 0.6

    CALIBRATION CHECK

    REF TRANSMITTANCE(S) AND CORRECT FILTER VISIBILITIES

    CURRENT T, FILTER VISIBILITY

    0.9766 421

    VIS VER TER

    436 15 0.005

    CHECK END

    Special Maintenance Commands

    FCAL Th FCAL (filt lib ti ) d i d t d t i th li f t K2 K3 d K4 d i

  • 8/11/2019 K001EN-B

    88/122

    2003-03

    K001en-B

    FCAL The FCAL (filter calibration) command is used to determine the scaling factors K2, K3, and K4 used inthe transmittance scaling. These factors are automatically recalculated after each filter calibration

    procedure. Visibility must be stable during the sequence, which takes about five minutes. Windows

    should be clean. The factors K2, K3, and K4 can be initialized to the default value 1.0 with the CONF

    command. The command sequence for a double baseline system is presented below.

    >FCAL

    FILTER CALIBRATION

    GIVE CORRECT VISIBILITY VALUE

    GIVE FILTER P1 (LOW) VALUE

    SET FILTER AND TYPE ESC WHEN VALUE IS STABLE

    INST(L) AVE(L) INST(S) AVE(S)

    0.5510 0.5511 0.5510 0.5511

    GIVE FILTER P2 (HIGH) VALUE

    SET FILTER AND TYPE ESC WHEN VALUE IS STABLE

    INST(L) AVE(L) INST(S) AVE(S)

    0.9310 0.9320 0.9110 0.9107

    NEW SCALING FACTORS K2 K3 K40.9078 1.1070 0.9042

    ARE THE SCALES OK ? (Y/N) Y

    NEW SCALES ARE UPDATED TO EEPROM

    FILTER CALIBRATION DONE

    Special Maintenance Commands cont.

  • 8/11/2019 K001EN-B

    89/122

    2003-03

    K001en-B

    TCON This command is for contamination measurement test purposes. The output includes a

    direct contamination signal (stray light included), reference intensity from

    contamination detector light source and scaled contamination measurement value.

    > TCON CONTAMINATION MEASUREMENT TEST

    SIGNAL : REF =

    3908 3908 1.0000

    3909 3909 1.0000

    ALIG This command helps one in the alignment process during the transmissometer

    installation. It displays continuously (until ESC is pressed) the maximum transmissionsince start and the difference of the latest value from the maximum. Bel-character

    "beep" is echoed when a new maximum is reached.

    Special Maintenance Commands cont.

    The normalizing scale for this result value is calculated by the clean command

  • 8/11/2019 K001EN-B

    90/122

    2003-03

    K001en-B

    The normalizing scale for this result value is calculated by the clean command.AN N Analog channel input test. Outputs A/D conversion result from selected channel

    number N.

    AN 1 Outputs continuously A/D conversion results of the light intensity measurement

    channel.

    AN 2 Outputs continuously A/D conversion results of the temperature measurement channel.AN 3 Outputs continuously A/D conversion results of the contamination measurement

    channel.

    AN 4 Outputs continuously A/D conversion results of the contamination

    measurement reference channel (contamination lamp intensity).

    FRE Frequency input test for background luminance meter interface. Result is output in Hz.

    >FREQ

    FREQUENCY INPUT

    INST(1S), AVE(10S)

    1020 1020

    1019 1020

  • 8/11/2019 K001EN-B

    91/122

    2003-03

    K001en-B

    Troubleshooting

    Status

    MES includes status bits in 4 character hexadecimal

  • 8/11/2019 K001EN-B

    92/122

    2003-03

    K001en-B

    MES includes status bits in 4 character hexadecimalnumber for single baseline system and 6 character fordouble baseline.

    Status 4101 is OK for transmissometer with Backgroundluminance meter and 0101 for others (double baseline410101 and 010101).

    STA command gives status in plain language.

    Status Information

    The transmissometer output for STATUS command is:

  • 8/11/2019 K001EN-B

    93/122

    2003-03

    K001en-B

    The transmissometer output for STATUS command is:

    TRANSMISSOMETER STATUS

    TRANSMITTER REC(L) REC(S)

    MEAS MODE ON ON ON

    CONT/OTHER STATUS OK OK OK

    OPTICAL SURFACE OK OK OK

    POWER SUPPLY OK OK OK

    HEATING OK OK OK

    FLASH LAMP OK

    BL METER ON

    MEASUREMENT LOOP OK

    CONTAMIN. LAMP OK OK OK

    CALIBRATION OK OK

    TEMPERATURES 30.3 29.5 31.2 CONTAMINATION 0.0 0.0 0.0

    MEASURED DATA

    T(LONG) T(SHORT) V(L) V(S) BL C(T) C(L) C(S)

    0.9523 0.9840 4608 3769 333.0 0.9998 0.9983 1.0044

    Alarms Presented in Central Computer

    Alarms on operator's terminal of data processing unit

  • 8/11/2019 K001EN-B

    94/122

    2003-03

    K001en-B

    Alarms on operator s terminal of data processing unit

    BLANK = Transmissometer operation is OK.

    Transmissometer status

    FAIL = Any of transmissometer status alarm bits may be set on, i.e. maintenance

    needed.

    OFF = Transmissometer measurement mode is OFF; the measurement mode can be set ON by maintenance commands.

    Optical surface

    DIRTY = The protective windows of the transmissometer should be cleaned.

    Cleaning of the background luminance meter lens should be done at the

    same time.

    If the status DIRTY remains on after cleaning the windows,

    check the operation of contamination compensation.

    Status Information

    The first eight status bits of the status message are included in the standard message sent to thedata processing unit The standard message includes 8 status bits for the light transmitter and 8

  • 8/11/2019 K001EN-B

    95/122

    2003-03

    K001en-B

    The first eight status bits of the status message are included in the standard message sent to thedata processing unit. The standard message includes 8 status bits for the light transmitter and 8status bits for the light receiver. The CONT/OTHER FAIL is used as a common failure indicationfor status bits 9, 10, 11 etc. Only bit 9 is operative and informs on contamination lamp brightness.Status bits 10, 11, etc. are not presented, they are reserved for future use.

    The status information can be analyzed by means of the following alternatives:

    MEAS MODE OFF - Transmissometer is in stand-by mode.

    MEAS MODE ON - Normal visibility measurement is performed.

    CONT/OTHER STATUS OK - Contamination lamp intensity high enough, no failures detected.

    CONT/OTHER STATUS FAIL - Contamination lamp intensity below 1000 (too low) or some other failure detected.

    OPTICAL SURFACE OK - No cleaning of protective windows is needed.

    OPTICAL SURFACE DIRTY - Protective windows should be cleaned.

    - Transmittance is decreased with four percent per window and the visibility output is slashed.

    Status Information

    Transmissometer status information in detail

  • 8/11/2019 K001EN-B

    96/122

    2003-03

    K001en-B

    The corresponding status bits in the light receiver(s) operate in the same way.There are some special status bits available only at the light receiver side:

    CALIBRATION OK - No calibration error detected.

    CALIBRATION FAILURE - The transmittance data is continuously 0.998 which is an

    unusual situation.

    - The autocalibration constant K1A is less than 0.95.

    - The cause can be in the light transmitter or receiver. - If this status occurs from time to time, the optical alignment

    should be checked.

    Troubleshooting Hints

    Hints for troubleshooting:

  • 8/11/2019 K001EN-B

    97/122

    2003-03

    K001en-B

    POWER SUPPLY FAIL * Check the fuses on the mother board.

    * Replace the LPP11 power supply unit.

    HEATING FAILURE 1. Check internal temperature by the STATUS command.

    2. Check the heating fuses, heating supply switch and heating voltage on the connector X7 of LPT11 (a.110 VAC).

    3. Observe the length of the heating duty cycle indicated by the green LEDs on the LPH11 unit: is it reasonable concerning ambient temperature?

    * If not, replace LPH11 and check the length of the heating duty cycle again. If this did not help, check the temperature voltage signal from the LPI11 output TEa.

    4. If the heating duty cycle is reasonable and replacement of LPP11 (or LPI11) did not help:

    * Check whether the bimetallic switch in the electronics box is conducting ( turn mains heating and supply power off and measure resistance over it, should be lower than 1 ohm ).

    * Check the cables, connectors and heating elements.

    Troubleshooting, cont.

    Hints for troubleshooting :

  • 8/11/2019 K001EN-B

    98/122

    2003-03

    K001en-B

    FLASHLAMP FAILURE 1. Check the TM(I), i.e. intensity of the flashlamp by the DMES or MES 1 commands,should be over 1000.

    2. Check the flashlamp intensity setting values.

    * Try to give a bigger value for SETF and check whether the brightness of the flashlamp increases to a sufficiently high level, for instance over 1500 is acceptable.

    * If the cause for reduced intensity was aging of flashlamp, check the stability of it after giving a new SETF value, and taking TM(I) values in one hour's intervals,

    at least 10 values each time.

    CURRENT LOOP FAILURE 1. Give reset to the light transmitter, check the operation of the current loop by means of the green LED on the LPT11 unit.

    2. Give reset to the light receiver(s).

    3. Check the current loop supply voltage on the LPT11.

    NOTE: More hints for troubleshooting can be found from Users Guide.

  • 8/11/2019 K001EN-B

    99/122

    2003-03

    K001en-B

    Maintenance

    Periodic Maintenance

    The periodic maintenance of the transmissometers guarantees that the MITRASsystem operates within the range of its characteristics and the ICAO

  • 8/11/2019 K001EN-B

    100/122

    2003-03

    K001en-B

    p gsystem operates within the range of its characteristics and the ICAOrecommendations.

    The following table presents the recommended interval for periodic maintenance.

    Maintenance Action Recommendedinterval

    MaximumInterval

    Operational check of

    transmissometers via data

    processing unit or PC

    1 week 2 weeks

    Cleaning of windows 1 month or asindicated

    3 months

    Cleaning of windows if no

    contamination compensation

    1 week or as

    indicated

    2 weeks

    Alignment and calibration check 6 months (or after

    wet or frostyseason)

    12 months

    General checking of cables,

    connectors, and parts in

    electronics box and optics heads,

    and Backround Luminance Meter

    LM11

    6 months 12 months

    Check After 1 Month of Operation

    Stability troubleshooting during start-up period, first 6 months:At start-up after installation

  • 8/11/2019 K001EN-B

    101/122

    2003-03

    K001en-B

    p

    - Make sure that the transmissometer is not in the contamination compensation mode.

    - Clean the windows.

    - Check the calibration of the transmissometer.

    - With PAR check thar the K1A equals 1.00.

    Alignment check after one month's operation

    (The main purpose is to check that the transmissometer alignment is stable.)

    - Change ECONOMY mode OFF.

    - Clean the windows.

    - Check the upper-range value(s) by giving MES 1 C or DMES. If it has decreased 3 percent or more:

    * Check that anchor bolts and frangible bolts are fixed properly.

    * Realign the light transmitter and the light receiver(s).

    * Check that the fixing nuts of the alignment adjustment screws are locked properly.

    * Recalibrate the transmissometer.

    - If the upper-range value has changed less than 3 percent, the installation of the transmissometer can be considered stable.

    - Check K1A by giving PAR.

    - Give CAL to rescale the transmissometer, if the upper-range value has changed more than 1 percent.

    Checks After 2 Months of Operation

    Stability check 2 months after installation

  • 8/11/2019 K001EN-B

    102/122

    2003-03

    K001en-B

    - Clean the windows.

    - Compare the transmissometer output transmittance values with the values you checked one month ago.

    - If changes are less than 1 %, stability of the transmissometer is OK.

    NOTE: Changes in visibility conditions may increase fluctuation in transmittance values. With a 75 m baseline transmissometer the output changes 1.1 % when visibility changes from 15 km up to 50 km.

    - With MES 1 C check the upper range value. If changes are in the range of 1.5 %...3 %, the obtained differencies in the upper-range values may be due to visibility variations, which must be first considered. However, the checkpoint at 50 % should not change more than 1.5 %, and the checkpoints 10 %, 25 % should not change more than 1 %.

    - Check that K1A is between 0.95 and 1.00.

    * In case change of upper-range values is 1.5 %...3 %, repeat the stability check on the next clear day.

    - If the above mentioned limits are exceeded, go through the linearization check.

    Operation and Maintenance

    Contamination compensation operational check

  • 8/11/2019 K001EN-B

    103/122

    2003-03

    K001en-B

    - When stability check is passed, change the transmissometer to operate in the contamination compensation mode.

    - Make a calibration check after one month's operation of contamination compensation.

    - Do not clean the windows.

    By STA command you can read the contamination percentage, signals and actual transmittance values.

    - Save or write down the values.

    - Clean now the transmitter window and see by DMES that C(T) is 0.9...1.1 and stable.

    - Recheck by STA and save the values. The change of T(LONG) and T(SHORT) compared to

    the change in the percentage value shall be approximately the same. - Repeat the steps for the receiver window(s).

    - If the differencies are more than 1.5 %, you may calculate a new compensation power Ap:

    where C = contamination signal before cleaning

    T = T before/T after

    NOTE: Ap = 24 is recommended to be used.

    In( C)

    ln (T )

    Ap =

  • 8/11/2019 K001EN-B

    104/122

    2003-03

    K001en-B

    Calibration and Linearization check

    Calibration CHECK-1

    1) Collect tools, filters, laptop terminal, cable, manual, cleaning liquid, unlintedcloth and ladder.

  • 8/11/2019 K001EN-B

    105/122

    2003-03

    K001en-B

    cloth and ladder.

    2) Take a copy of Calibration Sheets, visibility over 10 km and no wind overthe runway to the transmissometers.

    3) Connect the Laptop PC to LP11 and start Terminal emulation-program at PC.

    Check STA for status and DMES C for the outputted visibility.4) Clean optical windows of LP11, LR11 and LM11.

    5) Check the current settings of the jumper (X2) from Heating Control BoardLPH11.

    6) Switch the AC and heating power on both the transmitter LP11 and receiverunit(s) LR11. Inside temperature need about 30 to 60 minutes to stabilize.

    7) Use STATUS command to check that transmissometer status does notindicate FAIL.

    Calibration CHECK-2

    8) Use MODE command to select the correct functional modes.>MODE

  • 8/11/2019 K001EN-B

    106/122

    2003-03

    K001en-B

    MODE

    SELECT OPERATION MODES

    MEAS MODE (1=ON, 0=OFF) 1 1

    ECON MODE (1=ON, 0=OFF) 1 0

    CONT MODE (1=ON, 0=OFF) 1 0

    AUTO CALIBRATION (1=ON, 0=OFF) 1 0

    BL METER (1=ON, 0=OFF) 1 X

    AVE COUNT (1=ON, 0=OFF) 1 1

    9) Test the contamination measurement of the transmitter and receiver unit(s)with the TCON command. The SIGNAL and REF parameters must be at range1000...3500 when protective windows are clean. TCON command must begiven separately for both transmitter and receiver(s).

    10) Use DMES command to verify that the scaled contamination values C(TM)and C(L) (and C(S) for short baseline) are close to 1 (0.96...1.04) when thewindows are clean. If the value is below 0.96, clean the windows againg andrepeat test. If the value is below 0.96 after several cleanings, give CLEANcommand. If the value is above 1.04, give CLEAN command.

    Calibration CHECK-3

    11) Use the CONF command to check that the baseline length is correct.

  • 8/11/2019 K001EN-B

    107/122

    2003-03

    K001en-B

    ) g

    12) In the case the Ligth Intensity Measurement Board LPI11 of the receiver isreplaced, check that the scaling factors K1, K1A, K2, K3 and K4 are initializedcorrectly before the calibration and linearization checks are performed. Use

    the CONF command to write down the previous set values. After this initialize all the scaling factors K1, K1A, K2, K3 and K4 of the

    corresponding baseline(s) to 1.

    13) Use DMES C command to check current visibility value V(L) (and V(S)) andcompare the value with reference visibility value. If the reference visibility

    value and V(L) (and V(S)) correspond calibration is not needed.

    If the V(L) is not close to reference value, give the correct MOR value with CALcommand

    >CAL ie. > CAL 10000

    14) Use MODE command to retreive earlier functional modes (see step 8).

    Linearization CHECK-1

    1) Check the calibration and visibility. If calibration needs adjustemnts do thecalibration procedure.

  • 8/11/2019 K001EN-B

    108/122

    2003-03

    K001en-B

    2) Use MODE command to select the correct functional modes.

    >MODE

    3) Clean all optical surfaces as described in calibration section.

    4) Use the Mitras calibrator filter set. Use the test filters 0.90, 0.50 and 0.25 andthe combination of 0.50 and 0.25 as 0.125. Clean the filters with a milddetergent and a soft, lint-free cloth.

    5) Give command

    > CHECK 6) Insert the test filter into the transmitter hood in such a way the filter value can

    be seen from front. Wait until the VER and TER values have settled. VER is thedifference from the filter determined visibility and TER is the difference fromthe filter determined transmittance. Stop the output on the screen by pressingthe ESC key.

    7) Write the FILTER, CURRENT T, VISIBILITY, VIS, VER, and TER values down tothe Calibration Sheet.

    Linearization CHECK-2

    8) Remove the test filter. Give the DMES C command and wait until the V(L) (andV(S)) value stabilizes back to the current visibility value without the filters.Stop the output on the screen by pressing ESC

  • 8/11/2019 K001EN-B

    109/122

    2003-03

    K001en-B

    Stop the output on the screen by pressing ESC.

    9) Repeat the steps from a to e with all the filters and filter combinations.

    10) Check from the Calibration Sheet that the TER values are within the tolerance

    |TER| < 0.02. If a test filter is out of tolerance, perform the check again. Smallamounts of pollution, dust, sand, or other fluctuations may cause erroneousresults. Also check that the test filters are clean and undamaged. In casefurther problems arise, contact Vaisala.

    NOTE:The FCAL command should not be used without permission from Vaisala.

    11) Use MODE command to retreive earlier functional modes (see step 2).

    Calibration and linerization check sheet

    Filter Calibration with FCAL CommandDATE INITIALS RVR ID

    CORRECT VIS.

  • 8/11/2019 K001EN-B

    110/122

    2003-03

    K001en-B

    DATE INITIALS RVR ID

    FILTER CURRENT T VISIBILITY VIS VER TER OK

    DATE INITIALS RVR ID

    FILTER CURRENT T VISIBILITY VIS VER TER OK

    DATE INITIALS RVR ID

    FILTER CURRENT T VISIBILITY VIS VER TER OK

    DATE INITIALS RVR ID

    FILTER CURRENT T VISIBILITY VIS VER TER OK

    VALUE INST (L) AVE (L) INST (S) AVE (S)

    FILTER P1 (LOW)

    FILTER P2 (HIGH)

    NEW SCALING FACTORS K2: K3: K4:

    Linearization Check with CHECK CommandThe recommended visibility tolerances are

    given in the table below

    MOR Visibility ( metres ) Tolerance ( metres )

    < 150 25

    150...500 50

    500...10000 100

    1000...1200 200

    >1200 20 % of the reference visibility

    Select filters for FCAL

  • 8/11/2019 K001EN-B

    111/122

    2003-03

    K001en-B

    Linearize the T(NOC) with parameters K1, K2, K3 and K4.At the end K1=K1A=1 and K2, K3, K4 have new values.

    Choose the two (2) filter values where the LPI11 are most unlinear. Draw Tfilter/ Toutput graphics indicating TER values

    Select the filters where TER are the biggest or on the different side of linar line. If the TER of very short visibility is over 0.015 consider how many meters the visibility differs from correct- often it could be only couple of

    meters.

    FCAL command

    FCAL Used to determine the scaling factors K2, K3 and K4 used in the transmittance scaling. These factors are recalculated automatically after FCAL procedure. See chapter 4, "Periodic maintenance", for further information. The command sequence is as follows:

    FCAL

  • 8/11/2019 K001EN-B

    112/122

    2003-03

    K001en-B

    >FCAL

    FILTER CALIBRATION

    GIVE CORRECT VISIBILITY VALUE

    GIVE FILTER P1 (LOW) VALUE

    SET FILTER AND TYPE ESC WHEN VALUE IS STABLE

    INST(L) AVE(L) INST(S) AVE(S)

    0.5510 0.5511 0.5510 0.5511

    GIVE FILTER P2 (HIGH) VALUE

    SET FILTER AND TYPE ESC WHEN VALUE IS STABLEINST(L) AVE(L) INST(S) AVE(S)

    0.9310 0.9320 0.9110 0.9107

    NEW SCALING FACTORS K2 K3 K4

    0.9078 1.1070 0.9042

    ARE THE SCALES OK ? ( Y / N ) Y

    NEW SCALES ARE UPDATED TO EEPROM

    FILTER CALIBRATION DONE

    FCAL Linearization -1

    1) Check that there is no filter in the Transmitter hood.

    If there is remove the filter and wait one minute for stabilizing

  • 8/11/2019 K001EN-B

    113/122

    2003-03

    K001en-B

    If there is remove the filter and wait one minute for stabilizing.

    Follow the stabilizing with MES 1 C command

    2) Give FCAL-command with your maintenance terminal.

    3) MITRAS ask for following inputs:

    VISIBILITY

    LOW FILTER VALUE

    4) Insert the low filter into the Transmitter hood. The label showing the transmittance of filter shall be away from coming light direction.

    5) Hit key when the difference between INSTANT and AVE values

    has settled. The difference should be less than 0.01.If not restart the FCAL-command.

    FCAL Linearization -2

    6) MITRAS asks HIGH FILTER VALUE

    7) Write down INST and AVE values of low filters

  • 8/11/2019 K001EN-B

    114/122

    2003-03

    K001en-B

    7) Write down INST and AVE values of low filters.

    8) Insert the low filter into the Transmitter hood. The label showing

    the transmittance of filter shall look at the receiver LR11.

    9) Hit key when the difference between INSTANT and AVE

    values has settled. The difference should be less than 0.01. If not restart the FCAL-command.

    10) MITRAS shows the new K2, K3, K4 and asks: DO YOU WANT TO STORE THE NEW VALUES INTO EEPROM?

    (Y=YES)

    11) Write down INST and AVE values of high filter.

    12) Write down the new K2, K3, K4 values.

    13) Perform CHECK-command and check the results.

    14) Update your CALIBRATION SHEET for History File.

  • 8/11/2019 K001EN-B

    115/122

    2003-03

    K001en-B

    Aligning Optics

    Aligning Optics

    The long baseline is always aligned first.

  • 8/11/2019 K001EN-B

    116/122

    2003-03

    K001en-B

    The alignment of the optics starts with aligning the lighttransmitter LP11.

    Next go to the LR11.

    The horizontal alignment:

    The flashing light should be seen at an equal distance ateither side of the receiver unit.

    The vertical alignment

    The flashing light should be seen (50m baseline and 2,5m

    masts) at the height of 0,5m and at the distance of 2m eachsides of the receiver unit.

    Aligning Optics

    Remove the LPI11.Align the receiver unit in such a way that the flashing light

  • 8/11/2019 K001EN-B

    117/122

    2003-03

    K001en-B

    g y g gpoints roughly to the center of optics when you look throughthe filters.

    Connect the LPI11 back and monitor the R(I) values withterminal.

    When the R(I) is within the limits of 1500 to 3500 the

    alignment is in order.

    Perform a calibration check.

    Optics Head Alignment

  • 8/11/2019 K001EN-B

    118/122

    2003-03

    K001en-B

    1 = Allen screw forvertical adjustment

    2 = Allen screws for horizontal adjustment

    3 = Set screws for horizontal adjustment4 = Allen screws for

    horizontal adjustment

    Double baseline optics adjustment

  • 8/11/2019 K001EN-B

    119/122

    2003-03

    K001en-B

    1 Knob for horizontal adjusting

    Direction A

    Direction A

    1

    1

    2

    2

    9606-005

    3

    3

    4

    4 4 Knob for vertical locking

    3 Knob for vertical adjusting

    2 Knob for horizontal locking

  • 8/11/2019 K001EN-B

    120/122

    2003-03

    K001en-B

    Replacing a Subassembly

    Replacement of contamination measurement lamp

  • 8/11/2019 K001EN-B

    121/122

    2003-03

    K001en-B

    4

    5

    3

    7

    C

    6

    A

    9606-006

    2

    1

    B

    Replacement of flashlamp

  • 8/11/2019 K001EN-B

    122/122

    2003-03

    K001en-B


Recommended