www.kust.de | www.kust-elektronik.com KUST ELEKTRONIK: LM1010 Precission LCR Meter LM1010 is a precision LCR meter with high accuracy, good stability, and wide measurement range. Controlled by a 16 bits MPU, LM1010 can be used for evaluating LCR compo- nents, materials and semiconductor devices over a wide range of frequencies (50 Hz to 100kHZ) and test signal levels (50mV to 1V, with 10mV resolution ). With its powerful functions, excellent performance, easy to read LCD display and easy menu operation, LM1010 is suitable for high speed measurement needed on product line and high accuracy and stability measurement need in laboratory. By using the Handler interface, IEEE488 interface (optional), and RS232C interface, LM1010 ca be easily implemented to automatic test system or computer remote control. • AC Signal: 20mV - 1V • Up to 40 measurment/sec. • 4.3” LCD Display • Test Signal Level Monitor • 30Ω or 100Ω output Impedance • Parameters: L,C, R, |Z|, D, X, θ, Q, DCR • Basic Accuracy: 0.05% • Comms: RS232, HANDLER,USB host, GPIB(optional) • LED Indication of the result Green - PASS / Red - FAIL • Beeper - result sound indication • Trigger Function: Manual, Internal, External, BUS • Internal Memory for 100 Instrument Settings • Key-Lock with password function • Correction/Compensation Function
Transcript

www.kust .de | www.kust-e lektronik .com

KUST ELEKTRONIK:LM1010 Precission LCR Meter

LM1010 is a precision LCR meter with high accuracy, good stability, and wide measurement range. Controlled by a 16 bits MPU, LM1010 can be used for evaluating LCR compo-nents, materials and semiconductor devices over a wide range of frequencies (50 Hz to 100kHZ) and test signal levels (50mV to 1V, with 10mV resolution ). With its powerful functions, excellent performance, easy to read LCD display and easy menu operation, LM1010 is suitable for high speed measurement needed on product line and high accuracy and stability measurement need in laboratory. By using the Handler interface, IEEE488 interface (optional), and RS232C interface, LM1010 ca be easily implemented to automatic test system or computer remote control.• AC Signal: 20mV - 1V• Up to 40 measurment/sec.• 4.3” LCD Display• Test Signal Level Monitor• 30Ω or 100Ω output Impedance• Parameters: L,C, R, |Z|, D, X, θ, Q, DCR• Basic Accuracy: 0.05%• Comms: RS232, HANDLER,USB host, GPIB(optional)• LED Indication of the result Green - PASS / Red - FAIL• Beeper - result sound indication• Trigger Function: Manual, Internal, External, BUS• Internal Memory for 100 Instrument Settings• Key-Lock with password function• Correction/Compensation Function

2 KUST Elektronikwww.kust .de | www.kust-e lektronik .com

Measurement functionLCR test parameters | Z |, | Y |, C, L, X, B, R, G, D, Q, θ, DCR

Test frequency:

LM1010A50,60,80,100,120,150,200,250,300,400,500,600,800,1K, 1.2K, 1.5K, 2K, 2.5K, 3K, 4K, 5K, 6K, 8K, 10K, 12K, 15K, 20K, 25K, 30K, 40K, 50K, 60K, 70K, 80K, 90K, 95K 100K, (Hz), total of 37

Basic accuracy 0.05%

The equivalent circuit Series, parallel

Mathematical functions The absolute value of the deviation, percentage deviation

Range mode Automatic, hold, manually select

Trigger Internal, manual, external, bus

Measurement speed (≥1kHz) Fast: The fastest 40 times / sec speed: 11 times / sec, Slow: 3 times / sec.

The average number of 1-255

Delay time 0-60s, with 1ms step

Correction: Open / short / load

Test side configuration Five end

Display mode Direct reading, Δ, Δ%, V / I (measured voltage / current monitoring)

Display 5 resolution, 4.3-inch LCD Monitor

Test signal

Test frequency 50 Hz - 100 kHz

Output Impedance 30 Ω, 100Ω selectable

Test signal level 20mV – 1V Accuracy: 10%

Display range

| Z |, R, X 0.01mΩ - 99.999 MΩ

| Y |, G, B 0.0001μS - 99.999 S

C 0.0001 pF - 9.9999 F

L 0.0001 μH - 999.99 H

D 0.0001 - 9.9999

Q 0.0001 --99999

θ (DEG) -179.99º - 179.99 º

θ (RAD) 3.14159 - 3.14159

DCR 0.01mΩ - 99.999 MΩ

OtherOutput 4 Terminals – 4 wire Connection. Ground terminal.

Front USB Screen Shots, Set-up files storage, Firmware Update

Comparator function Sorting functionResult Indication Green / Red LED, soundMonitor Voltage & CurrentComparator function Sorting functionMemory 100 internal instrument settings for storage / recall, external 500 groups

Interface Standard RS232C, HANDLER, USB HOST, optional GPIB

LM1010 requires the fol-lowing power source:Voltage: 198 to 242 VacFrequency: 47 to 66 HzPower: 50 VA maximum

LM1010 must be oper-ated under within the following environment conditions:Temperature: 0 to 40 Humidity: less then 90% RH

KA1002 :4 terminal Kelvin test leads

KA1001: 4 terminal test fixture

KA1008: Gilded shorting plate

Every LM1010 LCR Meter is supplied with main’s lead and factory calibration.


Recommended