MW Group Pte Ltd Certificate No. : LA-1995-0097-C 196 Pandan Loop #02-21 Issue No. : 19 Pantech Business Hub Singapore 128384 Date : 13 May 2016 Page : 1 of 25 FIELD OF TESTING : Calibration and Measurement MEASURED QUANTITIES/ INSTRUMENTS/RANGE TO BE CALIBRATED METHOD CALIBRATION & MEASUREMENT CAPABILITY (CMC*) A. MECHANICAL 1. Pressure/Vacuum & Absolute Section 2.0 of quality procedure Measuring Devices Gauges manual MW 004 (Gauge & Absolute) - Gauges - Transducers - Transmitters - Digital Indicators - Pen Recorders - Manometers - Data Loggers - Barometers - Calibrators - Hook Gauges - Caisson -14.5 to 0 psi 0.11 psi 10 to 200 mbar 0.11 mbar 200 to 1000 mbar 0.11mbar 15 to 75 psi 0.06 psi 75 to 200 psi 0.11 psi 400 to 600 psi 0.08 psi 600 to 1000 psi 0.11 psi 1000 to 2000 psi 0.65 psi 2000 to 5000 psi 1.4 psi 5000 to 10000 psi 2.6 psi 10000 to 15000 psi 3.9 psi
Transcript
Page 1: LA-1995-0097-C - MW Group Pte Ltd - sac  · PDF fileMW Group Pte Ltd Certificate No. : LA ... Screwdriver Calibration Method In ... 600 ºC to 800 ºC Electrical Simulation

MW Group Pte Ltd Certificate No. : LA-1995-0097-C 196 Pandan Loop #02-21 Issue No. : 19 Pantech Business Hub Singapore 128384 Date : 13 May 2016

Page : 1 of 25

FIELD OF TESTING : Calibration and Measurement

MEASURED QUANTITIES/ INSTRUMENTS/RANGE TO BE

CALIBRATED METHOD

CALIBRATION & MEASUREMENT CAPABILITY

(CMC*)

A. MECHANICAL 1. Pressure/Vacuum & Absolute Section 2.0 of quality procedure Measuring Devices Gauges manual MW 004 (Gauge & Absolute)

- Gauges

- Transducers

- Transmitters

- Digital Indicators

- Pen Recorders

- Manometers

- Data Loggers

- Barometers

- Calibrators

- Hook Gauges

- Caisson

-14.5 to 0 psi 0.11 psi

10 to 200 mbar 0.11 mbar

200 to 1000 mbar 0.11mbar

15 to 75 psi 0.06 psi

75 to 200 psi 0.11 psi

400 to 600 psi 0.08 psi

600 to 1000 psi 0.11 psi

1000 to 2000 psi 0.65 psi

2000 to 5000 psi 1.4 psi

5000 to 10000 psi 2.6 psi

10000 to 15000 psi 3.9 psi

Page 2: LA-1995-0097-C - MW Group Pte Ltd - sac  · PDF fileMW Group Pte Ltd Certificate No. : LA ... Screwdriver Calibration Method In ... 600 ºC to 800 ºC Electrical Simulation

Certificate No. : LA-1995-0097-C Issue No. : 19 Date : 13 May 2016 Page : 2 of 25

MEASURED QUANTITIES/ INSTRUMENTS/RANGE TO BE

CALIBRATED METHOD

CALIBRATION & MEASUREMENT CAPABILITY

(CMC*)

15000 to 20000 psi 5.2 psi

20000 to 30000 psi 7.5 psi

30000 to 40000 psi 11 psi

2. On-site Pressure & Vacuum Section 2.0 and 3.0 of quality Measuring Devices procedure manual MW 004

− Gauges

− Transducers

− Transmitters

− Digital Indicators

− Pen Recorders (Digital Pressure)

− Manometers Indicator

− Data Loggers

− Calibrators

− Deadweight Testers

− (Verification)

− Relief Valve/Valves

-1 bar to 20 bar 0.13% of reading

0 bar to 400 bar 0.14% of reading

3. Dead Weight Testers (actual pressure values corresponding to specific masses)

-14.5 to 0 psi 0.11 psi

10 to 200 mbar 0.11 mbar

200 to 1000 mbar 0.11mbar

15 to 75 psi 0.06 psi

75 to 200 psi 0.11 psi

200 to 400 psi 0.07 psi

400 to 600 psi 0.08 psi

600 to 1000 psi 0.11 psi

1000 to 2000 psi 0.65 psi

2000 to 5000 psi 1.4 psi

5000 to 10000 psi 2.6 psi

10000 to 15000 psi 3.9 psi

15000 to 20000 psi 5.2 psi

20000 to 30000 psi 7.5 psi

30000 to 40000 psi 11 psi

Page 3: LA-1995-0097-C - MW Group Pte Ltd - sac  · PDF fileMW Group Pte Ltd Certificate No. : LA ... Screwdriver Calibration Method In ... 600 ºC to 800 ºC Electrical Simulation

Certificate No. : LA-1995-0097-C Issue No. : 19 Date : 13 May 2016 Page : 3 of 25

MEASURED QUANTITIES/ INSTRUMENTS/RANGE TO BE

CALIBRATED METHOD

CALIBRATION & MEASUREMENT CAPABILITY

(CMC*)

4. Low Pressure Differential Measuring Devices (Gauge & Absolute)

− Gauges

− Transducers

− Transmitters

− Pen Recorders

− Digital Indicators

− Manometers

− Dataloggers

− Barometers

− Calibrators

− Hook Gauges

− Flow Meters

− Velocity Meters

− Caisson

-14.5 to 0 psi 0.11 psi

10 to 200 mbar 0.11 mbar

200 to 1000 mbar 0.11mbar

15 to 75 psi 0.06 psi

75 to 200 psi 0.11 psi

200 to 400 psi 0.07 psi

400 to 600 psi 0.08 psi

600 to 1000 psi 0.11 psi

5. Pressure Measuring Devices (Oil Free/Oxygen Use)

− Gauges

− Transducers

− Transmitters

− Pen Recorders

− Digital Indicators

− Manometers

− Dataloggers

− Barometers

− Calibrators

− Hook Gauges

− Caisson

Page 4: LA-1995-0097-C - MW Group Pte Ltd - sac  · PDF fileMW Group Pte Ltd Certificate No. : LA ... Screwdriver Calibration Method In ... 600 ºC to 800 ºC Electrical Simulation

Certificate No. : LA-1995-0097-C Issue No. : 19 Date : 13 May 2016 Page : 4 of 25

MEASURED QUANTITIES/ INSTRUMENTS/RANGE TO BE

CALIBRATED METHOD

CALIBRATION & MEASUREMENT CAPABILITY

(CMC*)

15 psi to 8500 psi Section 2.0 & 3.0 of quality 0.036% of reading

procedure manual MW 004

6. Pressure Relief Valve

(Gauge & Absolute)

− Pneumatic

− Oil

− Oxygen

-1 bar to 20 bar 0.19% of reading

0 bar to 70 bar 0.08% of reading

0 bar to 700 bar 0.09% of reading

7. Force Measuring Devices Section 2.0 & 3.0 of quality

4 kN to 250 kN procedure manual MW 004 0.93% of reading

8. 8. Vacuum Calibration Section 28 procedure manual 30% of reading

10 to 1 torr MW 004

1 to 1E-1

1E-1

to 1E-2

1E-2

to 1E-3

1E-3 to 1E

-4

1E-4 to 1E

-5

1E-5

to 1E-6

9. Flow Measuring Instruments

Positive displacement Method MW 0011 Mobile Flow 0.06 %

Coriolis Metering Laboratory facility

Ultrasonic

Vane

Paddle

Venturi

Orifices

Mass

Pitot

Electromagnetic

Turbine

Page 5: LA-1995-0097-C - MW Group Pte Ltd - sac  · PDF fileMW Group Pte Ltd Certificate No. : LA ... Screwdriver Calibration Method In ... 600 ºC to 800 ºC Electrical Simulation

Certificate No. : LA-1995-0097-C Issue No. : 19 Date : 13 May 2016 Page : 5 of 25

MEASURED QUANTITIES/ INSTRUMENTS/RANGE TO BE

CALIBRATED METHOD

CALIBRATION & MEASUREMENT CAPABILITY

(CMC*)

1000 litres

2000 litres

3000 litres

Flow rate : 500 litres/min (Min)

15000 litres/min (Max)

Meter Loop Proving Mobile Flow Metering 0.06 %

Laboratory facility

1000 litres

2000 litres API chapter 12, Part 4

3000 litres Water Draw

Flow rate : 500 litres/min (Min) Calibration of Meter Prover

15000 litres/min (Max)

1. Torque Measuring Instruments All type of Hand Torque Tools

− Round dial gauge Torque

− Preset Torque Wrench

− Torque Gauge (Digital / Analog)

− with torsion or flexion bar

− with rigid housing and electronic measurement

− with scale or dial or display

− with graduated or with display

− with fixed adjustment

− with adjustable non-graduated

Screwdriver Calibration Method In Accordance With In House Calibration Procedure MW004 section 10.0.0. and ISO 6789:2003 as a Guide.

− Rotating Torque Screw Driver

− with adjustable graduated or with display

Page 6: LA-1995-0097-C - MW Group Pte Ltd - sac  · PDF fileMW Group Pte Ltd Certificate No. : LA ... Screwdriver Calibration Method In ... 600 ºC to 800 ºC Electrical Simulation

Certificate No. : LA-1995-0097-C Issue No. : 19 Date : 13 May 2016 Page : 6 of 25

MEASURED QUANTITIES/ INSTRUMENTS/RANGE TO BE

CALIBRATED METHOD

CALIBRATION & MEASUREMENT CAPABILITY

(CMC*)

− with adjustable non-graduated

− with scale or dial or display with electronic measurement

− with electronic measurement

Range:

CW ACW

1.25 - 5 Nm 0.031 Nm 0.031 Nm

5 - 10 Nm 0.062 Nm 0.062 Nm

10 - 25 Nm 0.16 Nm 0.16 Nm

25 - 60 Nm 0.30 Nm 0.30 Nm

60 - 150 Nm 0.79 Nm 0.74 Nm

150 - 300 Nm 1.6 Nm 1.5 Nm

300 - 900 Nm 4.5 Nm 4.4 Nm

900 - 1500 Nm 7.5 Nm 7.6 Nm

Note:

1.CW- Clockwise

2.ACW- Anti-Clockwise

Page 7: LA-1995-0097-C - MW Group Pte Ltd - sac  · PDF fileMW Group Pte Ltd Certificate No. : LA ... Screwdriver Calibration Method In ... 600 ºC to 800 ºC Electrical Simulation

Certificate No. : LA-1995-0097-C Issue No. : 19 Date : 13 May 2016 Page : 7 of 25

MEASURED QUANTITIES/ INSTRUMENTS/RANGE TO BE

CALIBRATED METHOD

CALIBRATION & MEASUREMENT CAPABILITY

(CMC*)

B. TEMPERATURE 1. Temperature Measuring Devices Section 2.0 & 3.0 of quality procedure manual MW 005 PRT Sensor with indicator Liquid Bath & SPRT

-80 ºC to 420 ºC Fluidised & SPRT 0.082°C PRT Sensor without indicator 0.12°C -80 ºC to 420 ºC Thermocouple Sensor with Furnace & Std R/S type

Indicator -80 ºC to 250 ºC 0.13°C

100°C to 1250°C 1.7°C Thermocouple Sensor without

Indicator Furnace & Std R/S type

-80 ºC to 250 ºC 0.51°C 100°C to 1250°C 1.9°C

2. Ice Point of Water 0 ºC 16mK

3. Temperature Radiation Devices

Section 7.0.0 quality procedure

MW 005

0 ºC to 550 ºC 7.6 ºC

4. Thermocouples(Input Mode) Section 2.0 & 3.0 of quality

a. Type B procedure manual MW 005-

600 ºC to 800 ºC Electrical Simulation

800 ºC to 1000 ºC 0.84°C

1000 ºC to 1550 ºC 0.78°C

1550 ºC to 1820 ºC 0.72°C

0.74°C

b. Type C

0 ºC to 150 ºC 0.75°C

150 ºC to 650 ºC 0.73°C

650 ºC to 1000 ºC 0.76°C

1000 ºC to 1800 ºC 1.2°C

1800 ºC to 2 316 ºC 1.2°C

Page 8: LA-1995-0097-C - MW Group Pte Ltd - sac  · PDF fileMW Group Pte Ltd Certificate No. : LA ... Screwdriver Calibration Method In ... 600 ºC to 800 ºC Electrical Simulation

Certificate No. : LA-1995-0097-C Issue No. : 19 Date : 13 May 2016 Page : 8 of 25

MEASURED QUANTITIES/ INSTRUMENTS/RANGE TO BE

CALIBRATED METHOD

CALIBRATION & MEASUREMENT CAPABILITY

(CMC*)

c. Type E

-250 ºC to -100 ºC 0.57°C -100ºC to 300 ºC 0.18°C 300ºC to 600 ºC 0.20°C

d. Type J -210 ºC to -100 ºC 0.37°C -100 ºC to 100 ºC 0.26°C -100 ºC to 700 ºC 0.28°C 700 ºC to 1200 ºC 0.33°C

e. Type K -200 ºC to -100 ºC Section 2.0 & 3.0 of quality 0.42°C -100 ºC to 100 ºC procedure manual MW 005- 0.25°C 100 ºC to 900 ºC Electrical Simulation 0.34°C 900 ºC to 1372 ºC 0.49°C

f. Type N -200 ºC to -100 ºC 0.47°C -100 ºC to 100 ºC 0.24°C 100 ºC to 400 ºC 0.23°C 400 ºC to 1300 ºC 0.33°C

g. Type R 0°C to 200 °C 0.68°C

200°C to 400 °C 0.45°C

400°C to 900 °C 0.43°C

900°C to 1700 °C 0.50°C

h. Type S 0 ºC to 250 ºC 0.72°C 250 ºC to 1000 ºC Section 2.0 & 3.0 of quality 0.63°C 1000 ºC to 1400 ºC Electrical Simulation 0.63°C 1400 ºC to 1767 ºC 0.63°C

i. Type T -250 ºC to -150 ºC 0.74°C

-150 ºC to 0 ºC 0.33°C

0 ºC to 100 ºC 0.26°C

100 ºC to 400 ºC 0.24°C

Page 9: LA-1995-0097-C - MW Group Pte Ltd - sac  · PDF fileMW Group Pte Ltd Certificate No. : LA ... Screwdriver Calibration Method In ... 600 ºC to 800 ºC Electrical Simulation

Certificate No. : LA-1995-0097-C Issue No. : 19 Date : 13 May 2016 Page : 9 of 25

MEASURED QUANTITIES/ INSTRUMENTS/RANGE TO BE

CALIBRATED METHOD

CALIBRATION & MEASUREMENT CAPABILITY

(CMC*)

4.1 Thermocouples(Output Mode)

a. Type E

-250 ºC to -100 ºC 0.57°C

-100ºC to 300 ºC 0.18°C

300ºC to 600 ºC 0.20°C

600ºC to 1000 ºC 0.25°C

b. Type J -210 ºC to -100 ºC 0.37°C

-100 ºC to 100 ºC 0.27°C

-100 ºC to 700 ºC 0.29°C

700 ºC to 1200 ºC 0.34°C

c. Type K -200 ºC to -100 ºC 0.42°C

-100 ºC to 100 ºC 0.26°C

100 ºC to 900 ºC 0.35°C

900 ºC to 1372 ºC 0.49°C

d. Type N -200 ºC to -100 ºC 0.47°C

-100 ºC to 100 ºC 0.24°C

100 ºC to 400 ºC 0.23°C

400 ºC to 1300 ºC 0.33°C

e. Type R

0°C to 200 °C 0.68°C

200°C to 400 °C 0.45°C

400°C to 900 °C 0.43°C

900°C to 1700 °C 0.50°C

f. Type T

-250 ºC to -150 ºC Section 2.0 & 3.0 of quality 0.74°C

-150 ºC to 0 ºC procedure manual MW 005- 0.33°C

0 ºC to 100 ºC Electrical Simulation 0.26°C

100 ºC to 400 ºC 0.25°C

Page 10: LA-1995-0097-C - MW Group Pte Ltd - sac  · PDF fileMW Group Pte Ltd Certificate No. : LA ... Screwdriver Calibration Method In ... 600 ºC to 800 ºC Electrical Simulation

Certificate No. : LA-1995-0097-C Issue No. : 19 Date : 13 May 2016 Page : 10 of 25

MEASURED QUANTITIES/ INSTRUMENTS/RANGE TO BE

CALIBRATED METHOD

CALIBRATION & MEASUREMENT CAPABILITY

(CMC*)

4.2 Thermocouples (Output Mode- Onsite) a. Type E

-250 ºC to -50 ºC 0.97°C -50ºC to 100 ºC 0.42°C 0ºC to 600 ºC 0.37°C 600ºC to 1000 ºC 0.39°C

b. Type J

-200 ºC to -50 ºC 0.85°C -50 ºC to 0 ºC 0.48°C 0 ºC to 400 ºC 0.47°C 400 ºC to 800 ºC 0.44°C 800 ºC to 1200 ºC 0.50°C

c. Type K

-200 ºC to -50 ºC 0.56°C -50 ºC to 0 ºC 0.45°C 0 ºC to 400 ºC 0.41°C 400 ºC to 800ºC 0.50°C 800 ºC to 1372ºC 0.51°C

d. Type N

-200 ºC to -50 ºC 0.80°C -50 ºC to 0 ºC 0.38°C 0 ºC to 800 ºC 0.39°C 800 ºC to 1300 ºC 0.44°C

e. Type R 0°C to 100 °C 0.62°C 100°C to 1200 °C 0.61°C 1200°C to 1700 °C 0.63°C

5. PRT- Electrical Simulation a. Pt 385, 100 ohm Section 2.0 & 3.0 of quality -200 ºC to -80 ºC procedure manual MW 005 - 0.12°C -80 ºC to 0 ºC Electrical Simulation 0.13°C 0 ºC to 100 ºC 0.14°C 100 ºC to 300 ºC 0.15°C 300 ºC to 400 ºC 0.15°C 400 ºC to 630 ºC 0.17°C 630 ºC to 800 ºC 0.25°C

Page 11: LA-1995-0097-C - MW Group Pte Ltd - sac  · PDF fileMW Group Pte Ltd Certificate No. : LA ... Screwdriver Calibration Method In ... 600 ºC to 800 ºC Electrical Simulation

Certificate No. : LA-1995-0097-C Issue No. : 19 Date : 13 May 2016 Page : 11 of 25

MEASURED QUANTITIES/ INSTRUMENTS/RANGE TO BE

CALIBRATED METHOD

CALIBRATION & MEASUREMENT CAPABILITY

(CMC*)

b. b Pt 3926, 100 ohm

-200 ºC to -80 ºC 0.13°C

-80 ºC to 0 ºC 0.17°C

0 ºC to 100 ºC 0.14°C

100 ºC to 300 ºC 0.15°C

300 ºC to 400 ºC 0.15°C

400 ºC to 630 ºC 0.16°C

c. cPtNi 385, 120 ohm (Ni120)

-80 ºC to 0 ºC 0.14°C

0 ºC to 100 ºC 0.18°C

100 ºC to 260 ºC 0.18°C

6. On-site temperature measuring Dry Block

devices Bi-metal thermometer

Analog gauges

Thermocouples

PRT thermometers

Temperature pen recorders

Temperature chart recorders

Digital temperature indicators

Temperature switches

Temperature transducer/ transmitter

Thermistor probe

-40 ºC to 600 ºC 1.2°C

7. Humidity/Temperature Measuring Devices 5% to 60% (at 30 °C) Humidity/Temperature 60% to 95% (at 30 °C) Chamber 3 % RH 25°C to 50°C 0.33°C

8. Controlled Temperature Enclosure Section 2.0 of quality

-80°C to 200°C procedure manual MW 005 1.7°C 201°C to 1000°C (Computerised Data Loggers 2.5°C

Page 12: LA-1995-0097-C - MW Group Pte Ltd - sac  · PDF fileMW Group Pte Ltd Certificate No. : LA ... Screwdriver Calibration Method In ... 600 ºC to 800 ºC Electrical Simulation

Certificate No. : LA-1995-0097-C Issue No. : 19 Date : 13 May 2016 Page : 12 of 25

MEASURED QUANTITIES/ INSTRUMENTS/RANGE TO BE

CALIBRATED METHOD

CALIBRATION & MEASUREMENT CAPABILITY

(CMC*)

C. ELECTRICAL

1. DC Voltage Measuring Instruments

0 to 329.9999 mV 0.026 mV 0 to 3.299999 V 0.00019 V 0 to 32.99999 V 0.0019 V 30 to 329.9999 V 0.022 V 100 to 1000.000 V 0.065 V

2. DC Current Measuring Instruments

0 to 3.29999 mA 0.0005 mA

0 to 32.9999 mA 0.004 mA

0 to 329.999 mA 0.042 mA

0 to 2.19999 A 0.0008 A

0 to 10.9999 A 0.0079 A

1 to 550 A 1.8 A

3. Resistance Measuring Instruments

0 to 10.99 Ω 0.011Ω

11 to 32.999 Ω 0.022Ω

33 to 109.999 Ω 0.029Ω

110 to 329.999 Ω 0.051Ω

330 to 1.09999 kΩ 0.0002kΩ

1.1 k to 3.29999 kΩ Section 2.0, 3.0 & 4.0 Quality 0.0004kΩ

3.3 k to 10.9999 kΩ Procedure Manual MW 007 0.0018kΩ

11 k to 32.9999 kΩ 0.0041kΩ

33 k to 109.999 kΩ 0.021kΩ

110 k to 329.999 kΩ 0.053kΩ

330 k to 1.09999 MΩ 0.0012MΩ

1.1 M to 3.29999 MΩ 0.0013MΩ

3.3 M to 10.9999 MΩ 0.0081MΩ

11 M to 32.9999 MΩ 0.045MΩ

33 M to 109.999 MΩ 0.64MΩ

110 M to 330 MΩ 1.9 MΩ

Page 13: LA-1995-0097-C - MW Group Pte Ltd - sac  · PDF fileMW Group Pte Ltd Certificate No. : LA ... Screwdriver Calibration Method In ... 600 ºC to 800 ºC Electrical Simulation

Certificate No. : LA-1995-0097-C Issue No. : 19 Date : 13 May 2016 Page : 13 of 25

MEASURED QUANTITIES/ INSTRUMENTS/ RANGE TO BE

CALIBRATED METHOD

CALIBRATION & MEASUREMENT CAPABILITY

(CMC*)

4. AC Voltage (Sinewave)

Measuring Instruments 1 m to 32.999 mV

10 to 45 Hz Section 2.0, 3.0 & 4.0 Quality 0.16mV

45 to 10 kHz Procedure Manual MW 007 0.079mV

10 k to 20 kHz 0.098 mV

20 k to 50 kHz 0.12 mV

50 k to 100 kHz 0.17 mV

100 k to 500 kHz 0.45 mV

33 m to 329.999 mV

10 to 45 Hz 0.99 mV

45 to 10 kHz 0.21 mV

10 k to 20 kHz 0.4 mV

20 k to 50 kHz 0.65 mV

50 k to 100 kHz 1.1 mV

100 k to 500 kHz 3.0 mV

0.33 to 3.29999 V 10 to 45 Hz 0.0059 V 45 to 10 kHz 0.0012 V 10 k to 20 kHz 0.012 V 20 k to 50 kHz 0.056 V 50 k to 100 kHz 0.011 V

100 k to 500 kHz 0.023 V

3.3 to 32.9999 V

10 to 45 Hz Section 2.0, 3.0 & 4.0 Quality 0.059 V 45 to 10 kHz Procedure Manual MW 007 0.016 V

10 k to 20 kHz 0.033 V 20 k to 50 kHz 0.077 V 50 k to 100 kHz 0.11 V

33 to 329.999V

Page 14: LA-1995-0097-C - MW Group Pte Ltd - sac  · PDF fileMW Group Pte Ltd Certificate No. : LA ... Screwdriver Calibration Method In ... 600 ºC to 800 ºC Electrical Simulation

Certificate No. : LA-1995-0097-C Issue No. : 19 Date : 13 May 2016 Page : 14 of 25

MEASURED QUANTITIES/ INSTRUMENTS/ RANGE TO BE

CALIBRATED METHOD

CALIBRATION & MEASUREMENT CAPABILITY

(CMC*)

45 to 1 kHz 0.2V 1 k to 10 kHz 0.32V 10 k to 20 kHz 0.38 V

330 to 1020 V

45 to 1 kHz 0.67 V

1 k to 5 kHz 2.5 V

5 k to 10 kHz 2.9 V

5. AC Current (Sinewave) Measuring Instruments

0.029 m to 0.32999 mA 10 to 20 Hz Section 2.0, 3.0 & 4.0 Quality 0.0016 mA 20 to 45 Hz Procedure Manual MW 007 0.0013 mA 45 to 1 kHz 0.0014 mA 1 k to 5 kHz 0.0020 mA 5 k to 10 kHz 0.0050 mA 0.33 m to 3.2999 mA 10 to 20 Hz 0.0079 mA 20 to 45 Hz 0.0043 mA 45 to 1 kHz 0.0043 mA 1 k to 5 kHz 0.0079 mA 5 k to 10 kHz 0.023 mA

3.3 to 32.9999 mA Section 2.0, 3.0 & 4.0 Quality

Procedure Manual MW 007 10 to 20 Hz 0.079 mA 20 to 45 Hz 0.042 mA 45 to 1 kHz 0.038 mA 1 k to 5 kHz 0.079 mA 5 k to 10 kHz 0.23 mA

33 to 329.99 mA

10 to 20 Hz 0.79 mA 20 to 45 Hz 0.42 mA 45 to 1 kHz 0.38 mA 1 k to 5 kHz 0.79 mA 5 k to 10 kHz 2.3 mA

Page 15: LA-1995-0097-C - MW Group Pte Ltd - sac  · PDF fileMW Group Pte Ltd Certificate No. : LA ... Screwdriver Calibration Method In ... 600 ºC to 800 ºC Electrical Simulation

Certificate No. : LA-1995-0097-C Issue No. : 19 Date : 13 May 2016 Page : 15 of 25

MEASURED QUANTITIES/ INSTRUMENTS/ RANGE TO BE

CALIBRATED METHOD

CALIBRATION & MEASUREMENT CAPABILITY

(CMC*)

0.33 to 2.19999 A

10 to 45 Hz 0.0053 A 45 to 1 kHz 0.0006 A 1 k to 5 kHz 0.019 A

2.2 to 11 A

45 to 65 Hz 0.01 A 65 to 500 Hz 0.015 A 500 to 1 kHz 0.044 A

Toroidal – Wound Current Clamps

10 to 550 A

DC, 0 - 11 A 1.8 A

45 to 65 Hz, 2.2 11A 1.9 A 65 to 440 Hz, <2.2A 0.55A

6. Capacitance Measuring

Instruments Section 2.0, 3.0 & 4.0 Quality

Procedure Manual MW 007

0.33 n to 0.4999 nF ) 50 to 1000 Hz 0.015 nF

0.50 n to 1.0999 nF ) 0.021 nF

1.10 n to 3.2999 nF ) 0.033 nF

3.30 n to 10.999 nF ) 0.14 nF

11.0 n to 32.999 nF ) 0.24 nF

33.0 n to 109.99 nF ) 0.45 nF

110 n to 329.99 nF ) 1.8 nF

0.33 µ to 1.0999 µF ) 0.0042 µF

1.10 µ to 3.2999 µF ) 50 to 400 Hz 0.02 µF

3.30 µ to 10.999 µF ) 0.054 µF

11.0 µ to 32.999 µF ) 50 to 1000 Hz 0.22 µF

33.0 µ to 109.99 µF ) 50 to 200 Hz 0.75 µF

110 µ to 329.00 µF ) 50 to 100 Hz 3.0 µF

330 µ to 1.1000 mF ) 0.018 mF

Page 16: LA-1995-0097-C - MW Group Pte Ltd - sac  · PDF fileMW Group Pte Ltd Certificate No. : LA ... Screwdriver Calibration Method In ... 600 ºC to 800 ºC Electrical Simulation

Certificate No. : LA-1995-0097-C Issue No. : 19 Date : 13 May 2016 Page : 16 of 25

MEASURED QUANTITIES/ INSTRUMENTS/RANGE TO BE

CALIBRATED METHOD

CALIBRATION & MEASUREMENT CAPABILITY

(CMC*)

7. DC Power Measuring Instruments

33 mV to 1000 V

3.3 mA to 8.9990 mA 0.91 W

9.0 mA to 32.999 mA 0.20 W

33 mA to 89.990 mA 0.10 W

90 mA to 329.99 mA 0.22 W

0.33A to 0.8999 A 1.5 W

0.90A to 2.1999 A 3.2 W

2.20A to 4.4999 A 6.1 W

4.50A to 11.000 A 12 W

8. AC Power (45Hz to 65Hz) PF=1

Measuring Instruments

330 mV to 1000V

9.0 mA to 32.999 mA Section 2.0, 3.0 & 4.0 0.33 W

33 mA to 89.990 mA Quality Procedure Manual 0.11 W

90 mA to 329.99 mA MW 007 0.50 W

33 mV to 329.999 mV

330 mV to 1020 V

0.33 A to 0.8999 A 2.3 W

0.90 A to 2.1999 A 3.5 W

2.20 A to 4.4999 A 7.8 W

4.50 A to 11.000 A 17 W

Page 17: LA-1995-0097-C - MW Group Pte Ltd - sac  · PDF fileMW Group Pte Ltd Certificate No. : LA ... Screwdriver Calibration Method In ... 600 ºC to 800 ºC Electrical Simulation

Certificate No. : LA-1995-0097-C Issue No. : 19 Date : 13 May 2016 Page : 17 of 25

MEASURED QUANTITIES/ INSTRUMENTS/RANGE TO BE

CALIBRATED METHOD

CALIBRATION & MEASUREMENT CAPABILITY

(CMC*)

9. Frequency Measuring Instruments

0.01 to 119.99 Hz 0.008 Hz

120.0 to 1199.9 Hz 0.063 Hz

1.200k to 11.999 k Hz 0.28 Hz

12.00 k to 119.99 k Hz 2.9 Hz

1.200 M to 2.000 M Hz 0.058 kHz

10. . AC Voltage DC Offset (Sinewaves)

Measuring Instruments

Range: 3.3 mV to 32.999 mV Offset: 0 to 50 mV 0.45 %

Range: 33 to 329.999 mV

Offset: 0 to 500 mV 0.45 %

Range: 0.33 to 3.29999 V

Offset: 0 to 5 V 0.45 %

Range: 3.3 to 32.9999 V

Offset: 0 to 50 V 0.45 %

(Trianglewaves ) Section 2.0, 3.0 & 4.0 Quality

Procedure Manual MW 007

Range: 9.3 mV to 92.999 V p-p Offset: 0 to 50 mV 0.45 %

93 to 929.999mV p-p Offset: 0 to 500 mV 0.45 %

Range: 0.93 to 9.29999V p-p Offset: 0 to 5 V 0.45 % Range: 9.3 to 92.9999 V p-p Offset: 0 to 50 V 0.45 % Truncated sine waves Range: 9.3 mV to 92.999 V p-p Offset: 0 to 50 mV 0.45 %

Page 18: LA-1995-0097-C - MW Group Pte Ltd - sac  · PDF fileMW Group Pte Ltd Certificate No. : LA ... Screwdriver Calibration Method In ... 600 ºC to 800 ºC Electrical Simulation

Certificate No. : LA-1995-0097-C Issue No. : 19 Date : 13 May 2016 Page : 18 of 25

MEASURED QUANTITIES/ INSTRUMENTS/RANGE TO BE

CALIBRATED METHOD

CALIBRATION & MEASUREMENT CAPABILITY

(CMC*)

Range: 93 to 929.999mV p-p Offset: 0 to 500 mV 0.45 % Range: 0.93 to 9.29999V p-p Offset: 0 to 5 V 0.45 % Range: 9.3 to 92.9999 V p-p Offset: 0 to 50 V 0.45 % (Squarewaves) Range: 6.6 mV to 65.9999 V p-to-p Offset: 0 to 50 mV 0.45 %

Range: 66 to 659.999 mV p-p

0 t Offset: 0 to 500 mV 0.45 %

0.6 Range: 0.66 to 6.59999 V p-p

Offset: 0 to 5 V 0.45 %

Range: 6.6 to 65.9999 V p-p

Offset: 0 to 50 V 0.45 %

11. 1AC Current (Non-Sinewaves)

Measuring Instruments Triangle &

Truncated Sinewave

2.9 mA to 92.999 mA

0.01 to 10 Hz 5.7 %

10 to 45 Hz Section 2.0, 3.0 & 4.0 Quality 0.75 %

45 to 1 k Hz Procedure Manual MW 007 0.75 %

1 k to 10 kHz 5.7 %

93 mA to 929.999 mA

0.01 to 10 Hz 5.7 %

10 to 45 Hz 1.8 %

45 to 1 k Hz 1.8 %

0.93 A to 6.19999A

Page 19: LA-1995-0097-C - MW Group Pte Ltd - sac  · PDF fileMW Group Pte Ltd Certificate No. : LA ... Screwdriver Calibration Method In ... 600 ºC to 800 ºC Electrical Simulation

Certificate No. : LA-1995-0097-C Issue No. : 19 Date : 13 May 2016 Page : 19 of 25

MEASURED QUANTITIES/ INSTRUMENTS/RANGE TO BE

CALIBRATED METHOD

CALIBRATION & MEASUREMENT

CAPABILITY (CMC*)

10 to 45 Hz 6.5 %

45 to 1 k Hz 1.7 %

1 k to 5 k Hz 6.5 %

6.2A to 11A

45 to 500 Hz 3.5 %

500 to 1 kHz 6.5 %

12. AC Current (Non-Sinewaves)

Measuring Instruments Squarewave

2.9 mA to 65.999 mA

0.01 to 10 Hz Section 2.0, 3.0 & 4.0 5.7 %

10 to 45 Hz Quality Procedure Manual 1.7 %

45 to 1 k Hz MW 007 1.5 %

66 mA to 659.999 mA

0.01 to 10 Hz 5.7 %

10 to 45 Hz 1.5 %

45 to 1 k Hz 1.7 %

0.66 A to 4.39999 A 10 to 45 Hz 6.5 % 45 to 1 k Hz 1.7 % 1 k to 5 k Hz 6.5 % 4.4A to 11 A 45 to 500 Hz 3.5 % 500 to 1 kHz 6.5 %

13. Oscilloscope Function

VOLT FUNCTION Measuring 1 mV to 10 V/Div ( 87.5% coverage of grid)

Page 20: LA-1995-0097-C - MW Group Pte Ltd - sac  · PDF fileMW Group Pte Ltd Certificate No. : LA ... Screwdriver Calibration Method In ... 600 ºC to 800 ºC Electrical Simulation

Certificate No. : LA-1995-0097-C Issue No. : 19 Date : 13 May 2016 Page : 20 of 25

MEASURED QUANTITIES/ INSTRUMENTS/RANGE TO BE

CALIBRATED METHOD

CALIBRATION & MEASUREMENT

CAPABILITY (CMC*)

7mV to 7 V Into 50 Ω 0.15% 7mV to 70 V Into 1MΩ 0.15% EDGE FUNCTION Measuring Rise Time 250mVpp @1MHz < 400 ps 1.1% LEVELED SINE WAVE FUNCTION Measuring Flatness (Bandwidth) Calibration Procedure MW007 , Section -6.1.0 2.1% 50 kHz to 100 MHz 2.6% 100 MHz to 300 MHz FREQUENCY FUNCTION Measuring @1.3Vpp 50kHz 29 ppm 50 kHz to 100 MHz 29 ppm 100 MHz to 300 MHz 29 ppm TIME MARKER FUNCTION Measuring 0.5 ns to 2ns/div 0.68 % Range : 10 ns to 2 ns 5ns to 200ns/div Range : 1µs to 20 ns 0.14 % 0.5 µs to 10 µs/div Range : 2 µs to 50 µs 1.2 % 20 µs to 1s/div Range : 100 µs to 5s 0.59 % TRIGGER SYSTEM Amplitude Sensitivity Internal Triggering DC to 50 MHz @ 0.35 div 0.06 div 50 MHz to 300 MHz @ 1 div 0.06 div

Page 21: LA-1995-0097-C - MW Group Pte Ltd - sac  · PDF fileMW Group Pte Ltd Certificate No. : LA ... Screwdriver Calibration Method In ... 600 ºC to 800 ºC Electrical Simulation

Certificate No. : LA-1995-0097-C Issue No. : 19 Date : 13 May 2016 Page : 21 of 25

MEASURED QUANTITIES/ INSTRUMENTS/RANGE TO BE

CALIBRATED METHOD

CALIBRATION & MEASUREMENT

CAPABILITY (CMC*)

External Triggering DC to 50 MHz @ 250 mVpp 1.3 % 50 MHz to 300 MHz @ 500 mVpp 1.3 %

14. DC Voltage Source Instruments

0 mV to 100.0000 mV 0.0097mV 0.1 V to 1.00000 V 0.0001V 1 V to 10.00000 V 0.0005V 10 V to 100.0000 V Section 2.0, 3.0 & 4.0 0.0058V 100 V to 1000.000 V Quality Procedure Manual 0.063V

MW 007

15. Resistance Source Instruments

Test Current

0 -100.0000 Ω 1 mA 0.017 Ω

0.1-1.000000 kΩ 1 mA 0.00013 kΩ

1.1-10.00000 kΩ 100 µA 0.0013 kΩ

10-100.0000 kΩ 10 µA 0.013 kΩ

0.1-1.000000 MΩ 5 µA 0.00013 MΩ

1.1-10.00000 MΩ 500 nA 0.0046 MΩ

10.1-100.0000 MΩ 500 nA // 10 MΩ 0.92 MΩ

16. DC Current Source Instruments

Burden Voltage

0-10.00000mA < 0.1 V 0.0079 mA

10-100.000mA < 0.6 V 0.063 mA

0-1.000000 A < 1 V 0.0013 A

1-3.000000 A < 2 V 0.0048 A

17. Continuity Measuring Instruments

18. True RMS AC Voltage Source

Instruments

10 mV to 100.000 mV

3 Hz to 5 Hz 1.2 mV

5 Hz to 10 Hz 0.45 mV

10 Hz to 20 Hz 0.12 mV

Page 22: LA-1995-0097-C - MW Group Pte Ltd - sac  · PDF fileMW Group Pte Ltd Certificate No. : LA ... Screwdriver Calibration Method In ... 600 ºC to 800 ºC Electrical Simulation

Certificate No. : LA-1995-0097-C Issue No. : 19 Date : 13 May 2016 Page : 22 of 25

MEASURED QUANTITIES/ INSTRUMENTS/RANGE TO BE

CALIBRATED METHOD

CALIBRATION & MEASUREMENT CAPABILITY

(CMC*)

20 kHz to 50 kHz 0.2 mV

50 kHz to 100 kHz 0.77 mV

100 kHz to 300 kHz 5.1 mV

1.00000 V to 750 V

3 Hz to 5 Hz Section 2.0, 3.0 & 4.0 Quality 8.8 V

5 Hz to 10 Hz Procedure Manual MW 007 3.3 V

10 Hz to 20 Hz 0.77 V

20 kHz to 50 kHz 1.5 V

50 kHz to 100 kHz 5.8 V

100 kHz to 300 kHz 38 V

19. True RMS AC Current Source

Instruments

0 A to 1.00000 A

3 Hz to 5 Hz 0.012 A

5 Hz to 10 Hz 0.0039 A

10 Hz to 5 kHz 0.0016 A

1.1A to 3.00000 A

3 Hz to 5 Hz 0.04 A

5 Hz to 10 Hz 0.014 A

10 Hz to 5 kHz 0.0072 A

20. Frequency Source Instruments

100 mv to 750 V

3 Hz to 5 Hz 0.013 Hz

5 Hz to 10 Hz 0.013 Hz

10 Hz to 40 kHz 0.018 kHz

40 kHz to 300 kHz 0.036 kHz

21. DC Voltage Source(On-site) Section 3.4.0 Quality

Procedure Manual MW 007 0V to 30 V (Upper Display) 0.015 V 0V to 20 V (Lower Display) 0.014 V 0mV to 90 mV 0.13 mV

Page 23: LA-1995-0097-C - MW Group Pte Ltd - sac  · PDF fileMW Group Pte Ltd Certificate No. : LA ... Screwdriver Calibration Method In ... 600 ºC to 800 ºC Electrical Simulation

Certificate No. : LA-1995-0097-C Issue No. : 19 Date : 13 May 2016 Page : 23 of 25

MEASURED QUANTITIES/ INSTRUMENTS/RANGE TO BE

CALIBRATED METHOD

CALIBRATION & MEASUREMENT CAPABILITY

(CMC*)

22. DC Voltage Measurement (On-site)

0mV to 100 mV 0.13 mV 0 V To 10 V 0.013 V

23. DC mA Measurement and Source

(On-site)

0mA to 24 mA 0.12 mA

24. mV Measurement and Source

(On-site)

-10 to 75 mV 0.12 mV

25. Resistance Source(On-site) Section 3.4.0 Quality

Procedure Manual MW 007 15 Ω to 400 Ω 0.21 Ω 400 to 1.5 kΩ 1.7 kΩ 1.5 to 3.2 kΩ 2.1 kΩ

26. Resistance Measurement (On-site)

15 to 400 Ω 0.3 Ω 400 to 1.5 kΩ 1.3 kΩ 1.5 to 3.2 kΩ 1.7 kΩ

27. Frequency Source(On-site)

1 to 1100 Hz 0.76 Hz

1.0 to 10.0 kHz 0.12 kHz

28. Frequency Measurement (On-site)

1 to 1100 Hz 0.66 Hz

1.0 to 10.0 kHz 0.13 kHz

Page 24: LA-1995-0097-C - MW Group Pte Ltd - sac  · PDF fileMW Group Pte Ltd Certificate No. : LA ... Screwdriver Calibration Method In ... 600 ºC to 800 ºC Electrical Simulation

Certificate No. : LA-1995-0097-C Issue No. : 19 Date : 13 May 2016 Page : 24 of 25

MEASURED QUANTITIES/ INSTRUMENTS/RANGE TO BE

CALIBRATED METHOD

CALIBRATION & MEASUREMENT

CAPABILITY (CMC*)

D. DIMENSION

1. Caliper JIS B 7507:1993 Vernier, dial Vernier, dial and digital calipers and digital calipers a. Outside Measuring Face Range : Up to 300 mm 0-20mm 9µm 25-50mm 9µm 50-100mm 9µm

100-150mm 10µm

150-200mm 10µm

200-250mm 11µm 250-300mm 12µm b. Inside Measuring Face Range : Up to 300 mm 0-20mm 9µm 25-50mm 9µm 50-100mm 9µm 100-150mm 10µm 150-200mm 10µm 200-250mm 11µm 250-300mm 12µm 2. External Micrometer JIS B 7502:1994 Micrometer Range : Up to 100 mm callipers Flatness 3µm Parallelism 4µm Instrumental Error : 0-25mm 1µm 25-50mm 2µm 50-75mm 2µm 75-100mm 2µm 3. Dial Gauge a. Dial Gauge / Indicator JIS B 7503:2011 Mechanical

0- 25 mm dial gauges and 2µm BS EN ISO 463 : 2006

b. Dial Test Indicator JIS B 7533:1990 Mechanical

0-1.5 mm dial test gauges and 2µm BS EN ISO 9493 : 2010

* CMC is expressed as an expanded uncertainty estimated at a level of confidence of approximately 95%.

Page 25: LA-1995-0097-C - MW Group Pte Ltd - sac  · PDF fileMW Group Pte Ltd Certificate No. : LA ... Screwdriver Calibration Method In ... 600 ºC to 800 ºC Electrical Simulation

Certificate No. : LA-1995-0097-C Issue No. : 19 Date : 13 May 2016 Page : 25 of 25

Approved signatories

Mr Mark Wheeler - For all tests

Mr V ArunKumar - For all tests (except A3 and A9)

Note : This laboratory is accredited in accordance with the recognised International Standard ISO/IEC 17025:2005. A laboratory's fulfilment of the requirements of ISO/IEC 17025:2005 means the laboratory meets both the technical competence requirements and management system requirements that are necessary for it to consistently deliver technically valid calibration results. The management system requirements in ISO/IEC 17025:2005 (Section 4) are written in language relevant to laboratory operations and meet the principles of ISO 9001:2008 Quality Management Systems — Requirements and are aligned with its pertinent requirements.