http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
1/96
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
2/96
Sample Organic Laboratory Report 20
Characterization and Purification 23
EXPERIMENT 4: Extraction 48
EXPERIMENT 5: Distillation 53
EXPERIMENT 7: Identification of Organic Compounds by Qualitative
Analysis 62
EXPERIMENT 8A: Effect of Structure on the Free-radical Bromination
of Hydrocarbons 69
EXPERIMENT 8B: Relative Rates of Nucleophilic Substitution of
Reactions of Halides 72
EXPERIMENT 9: Medicinal Chemistry 75
EXPERIMENT 10: Nylon Experiment 78
Appendix A: Gas Chromatography 80
Appendix B: Infrared Spectroscopy 85
Appendix C: Nuclear Magnetic Resonance Spectroscopy 92
8/18/2019 Lab Manual CHML 2210 Fall 14
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
3/96
LABORATORY SAFETY RULES
Chemistry laboratories are places with unusual hazards, which for
the most part, are associated with the chemicals in them. Chemicals
are unforgiving: if you mishandle them, you can be seriously hurt.
TREAT THE LABORATORY AND EVERYTHING IN IT WITH RESPECT. The
following rules will minimize risks in the lab:
1. Safety goggles MUST be worn. When anyone is working, EVERYONE
must wear goggles. There are NO exceptions. Eyes are priceless and
irreplaceable. When goggles fog, step outside to mop them. Goggles
on top of the head do not protect the eyes.
2. An apron or lab coat MUST be worn when working. This adds an
extra layer of protection between chemicals and your body as well
as protects your clothes from discoloration or mutilation.
3. Shoes MUST cover the entire foot. NO exposed toes. NO flip
flops, sandals or extreme ballet flats that expose the entire top
of the foot. You may carry your shoes in a pack to lab.
4. Contact lenses tend to complicate eye injuries. The use of
contact lenses during lab periods is strongly discouraged.
5. Wearing long pants to lab is required. Tie back long hair to
keep it out of reactions. 6. Absolutely no smoking, eating or
drinking in the lab . 7. Locate the fire extinguishers, eye washes
and safety showers BEFORE you begin any
experimental work. 8. Do not run. Do not push. Do not work and
socialize. IF you’re sure yo ur sample can be
left safely you can talk. If you are not sure about the safety of
your sample or reaction — give it your attention and save the
conversation for later.
9. Do not alter the procedure for the experiment you are doing.
Follow all directions carefully.
10. Unauthorized experiments will guarantee an “F” for CHML 1045
for any student involved. (Note on academic honesty: failure to
perform analyses that you report, will result in a zero for the
experiment.)
11. Never taste or smell any chemical (to detect an odor, use your
hand to waft the chemical towards your nose). Do not touch
chemicals with your hands. At the least, you will get a skin rash.
Many chemicals are absorbed through the skin and some can cause
serious burns.
12. WARNING: All ethanol in this laboratory is denatured. Denatured
means that it is toxic as it contains poisons, which cannot be
removed by distillation. If you take any ethanol from the
laboratory for drinking purposes, you will kill or seriously injure
yourself or whomever you give it to.
13. Never heat a sealed apparatus —it WILL explode. The flying
glass will cause serious injuries.
14. Never heat a sample or reaction rapidly, always gradually.
NEVER leave a reaction unattended.
15. Your instructor will clean up broken glass (after you inform
the instructor); don’t pick it up yourself. Please treat glassware
with care.
8/18/2019 Lab Manual CHML 2210 Fall 14
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
4/96
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
5/96
LABORATORY SAFETY PROCEDURES
Safety is one of the major concerns in any lab. Safety for humans
is always a primary concern. The location of safety equipment
should be common knowledge to anyone working in the lab. In order
to become familiar with the safety equipment in this lab, make a
map on the reverse side of this sheet showing the location of the
following safety items. Indicate the fire escape with an
arrow.
1. Emergency shower and eye wash station
2. Fire extinguisher
3. Fire blanket
8. Electrical power cut off switch
9. First aid box
I have read and agree to follow ALL of the listed safety procedures
and precautions.
_________________________________________ ________________
Signature Date
_________________________________________ ________________
Instructor Section
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
6/96
6
Note: The next few pages are a general description of how to
prepare for lab, write up your experiments, write lab reports, etc
. Each lab instructor has personal preferences concerning these
topics. These preferences may differ somewhat from what is
presented here, but your instructor will tell you what their
preferences are. You should use the following information as a
guideline, but follow any specific directions your instructor gives
you.
THE LABORATORY NOTEBOOK Keeping organized, accurate records of
procedures, observations and data in a lab notebook is the
cornerstone of all kinds of scientific work. This includes organic
chemistry lab. A lab notebook is your record of procedures,
observations, amounts of chemicals and other information as you
complete experiments. Your lab notebook should give a step-by-step
record of what you do and observe and measure. Writing your lab
reports is easier for you to do with a well-organized lab notebook.
It also makes studying for the final exam easier.
Why bother with a lab notebook if I'm not majoring in chemistry? In
most scientific careers, graduate work, clinical studies, or
forensic analysis, etc., it is vital to properly record procedures,
data and results in a laboratory notebook. Improperly documented
procedures and data have resulted in patents being denied, research
findings being rejected or forensic tests being thrown out of
court. Maintaining a lab notebook takes some work, but many
instructors in upper-division science courses (and most supervisors
in the “real” world) will assume that you can properly use and
maintain a laboratory notebook.
What should be recorded in the lab notebook? Everything relating to
carrying out the experiment (title, objective(s), reaction, reagent
table, procedures, observations, raw data, etc.) should be written
directly into the notebook. At a minimum , all steps that must be
done in the lab (procedure, measuring mass, obtaining spectra,
etc.) must be recorded directly into the notebook.
Notebook guidelines Write all observations and data in ink directly
into your notebook as you carry out the experiment . (Yes, you have
to take your notebook to the balance or laboratory instrument to
record masses and/or measurements!) If you make an error in your
notebook, cross it out with a single line and re-enter your data
nearby. Do not obliterate your error with a large “blob” because if
the first entry was right, it is now unreadable. Do not write the
correction directly on top of the error; some notebooks use carbon
paper, and if the original copy is hard to read — then carbon copy
is usually impossible to read!) Write data only in your notebook.
Never write data on scratch paper or in the margins of the lab
manual! This very poor lab practice can (and does) lead to lost or
garbled data. The instructor reserves the right to confiscate
materials used to improperly record data and/or deduct points from
an offender. Please find below a summary of some guidelines for
keeping a laboratory notebook for organic chemistry lab courses.
These are taken from the UCL Journal of Technology (Jan
2000).
1. DO use bound books . Permanently bound books should be used.
They should be consecutively numbered and each page should be
dated, signed and witnessed.
2. Do use ink . Notebook entries should be made in ink and in
chronological order. Entries should not be erased or ‘whited out’.
If an entry contains an error, a line should be drawn through the
error and new text should continue in the next available
space.
3. Don’t leave blank spaces . Blank gaps between entries should be
avoided. If a blank space is left on a page, a line or cross should
be drawn through the blank space and the pages dated to prevent
subsequent entries.
8/18/2019 Lab Manual CHML 2210 Fall 14
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
7/96
7
4. Don’t modify . Prior entries should not be modified at a later
date. If data were omitted, the new data can be entered under a new
date and cross-referenced to the previous entry. Record experiments
when they are performed.
5. Do use past tense . Use the past tense (e.g., ‘was heated’) to
describe the experiments that were actually performed.
6. Do explain abbreviations and special terms . Explain all
abbreviations and terms that are nonstandard. Explain in context,
in table of abbreviations or in glossary.
7. Do staple attachments . Attachments such as graphs or computer
printouts should be permanently affixed in the notebook (by
stapling) and both the attachment and the notebook page signed and
dated.
8. Don’t remove originals . No original pages should be removed
from the notebook. 9. Do outline new experiments . When a new
project or experiment is started, the
objective and rationale should be briefly outlined e.g., in a short
paragraph or by providing a flowchart.
10. Do provide detail . Record experimental descriptions, including
operating conditions, test results, and all explanation of the
results, and thin layer chromatography sketches of the results. Any
conclusions should be short and supported by factual data. DO NOT
COPY THE DIRECTIONS IN THIS MANUAL. DO NOT wait until the end of
the experiment to write down observations – write them down as you
make them. Also, write directly into your notebook. NEVER take
notes on a paper towel or other loose sheet of paper.
If carbon copies are used, remember that your instructor will grade
the carbon copies so make sure they are readable! Whatever you
write in your notebook must give legible carbon copies. Make it
easy for the instructor to read what you write. Print in your
notebook; cursive writing often gives carbon copies that are very
difficult to read. Make sure that you press hard enough to produce
clear copies, and make sure the plastic shield is in place (for a
carbon copy notebook). Remember that tapping your pen, writing on
another paper resting on your notebook, or forgetting to put the
shield behind your carbon page will cause marks on the carbons that
will obscure what you
write. Periodically check your carbons to confirm that your writing
is clear and legible. Use several paragraphs instead of large
blocks of text because small paragraphs are easier to read . Keep
solvents and chemicals away from your notebook because they may
dissolve the ink or destroy the pages.
Your notebook is not expected to be flawless, but it should be easy
to follow. Cross-outs and corrections are expected. Do not skip
pages, and do not tear original pages out of your notebook, even if
they are spoiled. If you run out of room on a page, do not try to
cram your writing into the available space, because this typically
leads to unreadable carbon copies. Instead, use “Continued on
page...” and “Continued from page…” to help the instructor follow
your report. Useful tip: Leave a few blank lines between steps of
your procedure, so you can insert any changes or forgotten steps.
Also, leave space between sections of your report in case
you need to add something you forgot to include or if you need to
correct or better explain something you previously mentioned.
GOAL: You are keeping a good laboratory notebook if a student who
has taken a similar organic lab course at another university could
carry out the experiment based on your notes. You should work
toward this goal.
8/18/2019 Lab Manual CHML 2210 Fall 14
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
8/96
8
These sections should be prepared and in your notebook before you
come to lab:
Title, lab meeting time, date of experiment: This is
self-explanatory.
Objective: Write one to three sentences about the goal(s) of the
experiment. This could be the identification of an unknown
compound, the preparation of a compound, isolating a pure compound
from an impure sample, etc.
Main reaction (if applicable): All preparative experiments should
include the balanced reaction. Include reaction conditions,
catalysts, etc.
Table of reagents, solvents and products: Prepare a table listing
the chemicals used in the experiment. List headings such as
compound names, molecular weights, densities, melting and boiling
points, refractive index, etc. Include the relevant properties in
the table. Irrelevant information such as the boiling point of a
solid or the density of a gas can be excluded. You can find
information in your textbook, online (chemfinder.com is a good
place to start), in chemical catalogs (Aldrich, ACROS, Lancaster,
etc.) or the CRC Handbook.
Procedure: Do not just copy out of the manual. Summarize the steps
you will use. Example: “A miniscale fractional distillation
apparatus was set up as in Figure X in the manual page X. Glass
beads were used as packing, as in Figure Y in the manual .” You do
not have to describe how you clamped each piece of glassware.
(Recipes in a cookbook do not explain how to turn on an oven or
boil water.) Similarly, you do not have to explain each step of
obtaining a refractive index or setting up a filtration because
this information is freely available in your manual. Leave one or
two blank lines between steps to give you room to insert any steps
you left out or to include any changes announced by the instructor.
Avoid using first person (I/me/my) in your report. Record the exact
amounts of chemicals that you use in the Observations
section.
A very convenient and effective way to organize your Procedure and
Observation sections is the “two column” method shown in the Sample
Laboratory Notebook on pages 14-20. You may
also want to create a flow chart if the procedure is complicated or
has multiple steps with several fractions.
These two sections must be also entered in your notebook as you
carry out the experiment:
Changes to the procedure: If you deviate from the written procedure
(even if the instructor announces the changes), you need to write
the changes in your notebook. Any changes such as using different
reagents, different amounts, using an additional portion of solvent
to wash a solution, heating for a different time, etc. must be
recorded in your notes. Even if you have to redo a step because of
a mistake, you still have to write the changes down. (This is why
leaving a few blank lines between steps is a good idea!)
Observations and Data: Record the exact amounts of chemicals
actually used in the experiment. (The procedure may say to use
0.0100 moles of cyclohexanol. This amount is 10.0 g of
cyclohexanol, but you need to record the amount you used (10.03 g,
9.86 g or whatever). Remember to base your yield calculations on
the amounts of reagents you actually used , not on the amounts
called for in the procedure. You should describe the colors and
appearances of the reagents (before and after mixing!), color
changes of solutions, temperature changes as a chemical is added to
a solution and so forth. Make as many observations as you can. A
surprisingly large number of students forget to record the
appearance of their products.
8/18/2019 Lab Manual CHML 2210 Fall 14
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
9/96
9
Sometimes, the cause of an unsuccessful experiment can be
pinpointed using observations. (True story: When half of the
students failed to get a reaction to occur, checking the notebooks
showed they used one bottle (with grey powder) and the others used
another bottle containing white powder. It became clear that the
first bottle had been mislabeled, so the students unknowingly used
the wrong material.
REMEMBER: Unwritten observations and data never happened!
Data Table: This is for “raw” numbers such as masses or volumes of
reagents, uncorrected melting points and uncorrected refractive
index values or masses of products. Important! All of this
information should have been previously recorded in the
Observations section. Remember to use significant figures
appropriately.
The following sections can be entered in your notebook outside of
lab time:
Interpretation of Spectra: Report the information (peak values,
intensity, interpretation, etc .) in tables in your notebook. See
the Sample report for examples. Do not write on the spectrum itself
except for your name and a title (e.g. “John Doe, IR Spectrum of
product of Exp. 13 , thin
film”). Be sure to note the presence (or absence) of significant
peaks. For example, if you are preparing an alcohol from a ketone,
you should look for peaks in your spectrum that indicate that could
show the presence of the alcohol product as well as any unreacted
ketone.
Calculations: This is where numerical data is processed.
Thermometer readings are corrected, measurements on a TLC plate are
converted into R f values, masses of reactants and products are
converted to moles and so forth. Use significant figures (sig
figs.) properly in your calculations! NOTE: In any preparative lab,
you need to determine the limiting reagent and calculate the
theoretical and actual yield —even if you obtain no product at all!
If you do not know or do not remember how to calculate yields,
consult your instructor.
Do not “create” or “discard” significant figures. If you measured
to only hundredths of grams,
your calculations must reflect this. Thus, if your product weighs
3.13 g, do not add a trailing zero (3.130 g). On the other hand, if
a beaker weighed 45.217 g full of reagent and 41.137 g after adding
the reagent to the flask, you added 4.080 g to the flask, not 4.08
grams. The zero in the thousandths place is a significant figure
here.
Discussion of Results: This is the most important section of the
laboratory report. Here is where you explain and interpret the
results of your experiment. The length and specific content will
depend on the experiment, but the Discussion section is usually at
least 10-15 sentences in length. The section will be somewhat
longer in preparative experiments. You should, of course, always
use correct grammar, spelling and terminology in the lab report,
but doing so is especially important in the Discussion section.
Why? Because the quality of your discussion indicates how clearly
you understand the objectives of the experiment and how you
interpreted
your results. If you make inaccurate or poorly supported claims,
use illogical reasoning, omit relevant information, or use the
wrong terminology, it gives the impression —whether true or not
—that you do not understand the experiment. This impression will
impact your score for that report. Likewise, unclear wording,
misspelled words and poor grammar give the impression of sloppiness
or that you did not proofread your work. Even if an experiment goes
poorly, a well- written, well-reasoned Discussion will often earn
more points than a badly written Discussion describing a successful
experiment. More information concerning what should be included in
the Discussion section is given on the next page. In contrast to
the Procedure, using first-person
8/18/2019 Lab Manual CHML 2210 Fall 14
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
10/96
10
pronouns (I/me/my) is more acceptable in the Discussion section.
See the following page for more information.
Summary of Experiment: The key word is “summary”. This section
should be brief (typically 3 - 5 sentences). The summary should
state the outcome of the experiment and a couple of important
details. You are expected to include specific information such as
vial numbers, measured properties, yields, etc. Do not introduce
any new information in this section. This section should only
include information that is mentioned elsewhere in the report. The
key topics from the Discussion section will probably be part of
your Summary, but remember that the Discussion and Summary are
separate sections of the report.
Example of a poorly-written Summary (The reader has no idea what
the result was or what the data means.):
I successfully accomplished the objective of this experiment. My
data was very close to the expected values, and there were no
significant problems in carrying out the experiment.
Example of a much better summary (A reader can clearly see what was
done and what the results meant.):
My unknown liquid (code B-23) was identified as tetralin. The
measured boiling point was about 5 C lower than the listed value
for tetralin, but the measured refractive index was very close to
the value in the manual. Tetralin was the only compound with a
boiling point and refractive index that were both fairly close to
the measured values, so I am confident that my identification is
correct.
Example of a summary for an experiment involving the preparation of
a compound : Benzhydrol was prepared by the reduction of
benzophenone by NaBH 4. The yield of purified product was 65.8%.
The IR spectrum closely matches that of a pure sample and the
melting point is only 1- 2 C below the reported value, indicating
that the product is proba bly quite pure. The loss of mass during
recrystallization shows that this is the step where the major loss
of product occurred.
REMEMBER! Focus on the highlights. Think: “What in terms of “What I
did in lab in 50 words or less.”
Answers to follow-up questions: Your instructor may assign some or
all of these questions.
Additional Guidelines for Writing Your Discussion Section The most
important rule is: do not make claims in unless they are backed up
with specific information! For example, if you state that the
sample you obtained was pure, you must support this by giving
specific evidence that you have recorded such as a narrow melting
range, a melting range or appearance that that matches an authentic
sample, etc. Use specific data, IR absorptions, etc . to support
your claims. See these three examples.
Good: “My sample melted at 113 -114 C, which is close to the m.p of
pure acetanilide (114 - 115 C).”
So- so: “My sample melted at 113 -114 C, which is close to the m.p
of acetanilide.”
8/18/2019 Lab Manual CHML 2210 Fall 14
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
11/96
11
Poor: “My sample had the same melting point as acetanilide.” (The
reader must then look up the values.)
Consider all relevant information, even some seems contradictory.
For example, if a product looks identical to a known pure sample,
but the melting range data suggests the sample is impure, the
Discussion must address both pieces of evidence. Remember that
contradictory data may be due to misreading an instrument,
incorrect calculations, misinterpreting the instructions or many
other sources. If you find contradictions, double-check all of your
readings, calculations, etc. (Having a friend look at your data can
be helpful here.) If you cannot find a problem, but you still have
contradictions, then you must judge the importance or reliability
of each piece of data and the “weight” to give it as you draw
conclusions about the experiment. When you have contradictory data,
it is a good idea to consult your instructor before you begin to
write your Discussion.
Examples of appropriate topics for Discussion: Example #1. In
Experiment Y, you will purify a solid by recrystallization, and
identify it by melting point determination. The Discussion section
should address issues such as:
1) What is the identity of your solid, and how do you know that the
identification is correct? 2) Is the recrystallized material pure?
What evidence supports your claim that the product is
pure or not? 3) At which steps was product lost? If possible, point
out where major loss of product
occurred.
Example #2. In Experiment Z, you will prepare and purify an alkyl
halide and analyze it using refractometry and infrared (IR)
spectroscopy. Good topics for your Discussion include the
following:
1) Did you successfully prepare and isolate the desired product?
Was the product pure? 2) What does the IR spectrum indicate about
the identity of the product and/or the presence
of impurities? 3) What is the percent yield of isolated product? At
which steps were significant amounts of
product lost?
Three things you should n o t do in your Discussion 1. Do not
introduce new data, or observations. Never refer to anything
concerning the experiment that is not recorded in your notebook.
For example, your Discussion should not refer to the color of your
product unless it is recorded in your observations. Likewise, if
you spill a third of your product at the end of the experiment, but
you fail to document it in your observations, it is not appropriate
to cite spillage as a source of product loss in your Discussion.
Note: it is appropriate to refer to sources such as the Merck Index
or the CRC handbook for information about your product, but you
must properly cite the references you use.
Remember: Any unwritten observations or data never really happened!
Therefore, you cannot discuss something that never happened!
2. Do not re-state the procedure as part of the Discussion. Use the
Discussion section to explain and interpret your data and
results.
NOTE: It is okay to mention steps that affect the yield or purity
of your product. For example: “I started the lab late, so I was
only able to heat the mixture for 40 minutes, instead of 90.
The
8/18/2019 Lab Manual CHML 2210 Fall 14
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
12/96
12
reaction probably did not go to completion.” In short, discussing
how the procedure affects the results is fine; just restating the
procedure is not.
3. Do not combine your Discussion and Summary sections. The
Discussion section is where you evaluate and explain your results.
The Summary section is where you wrap up the experiment.
Keep this in mind when you are sick of preparing your notebook and
writing lab reports: Clear communication of information through
writing is the backbone of science. Strive to make your writing as
clear, complete, concise and correct as possible. Employers out in
the “Real World” place a high value on sound writing skills in
their employees . Also, many employers will expect new employees to
have experience in properly maintaining a laboratory notebook and
writing clear, concise lab reports.
Twelve Key Points to Remember as You Prepare for Lab and Write Your
Notebook 1. Preparing for lab is absolutely essential . Know what
you are supposed to do, what
chemicals you will use, what techniques you will employ and so
forth. Being prepared when you begin the experiment makes it much
easier to keep your observations, data, etc. organized. This, in
turn, makes it easier to find the information you need when you
write your Discussion section.
2. Summarize procedural steps when you write them. This saves time
for you and the reader as well. You may assume that the reader is
familiar with basic lab techniques, so you do not have to fully
explain them. Always include any additions or changes to the
procedure — including things like re-doing steps, performing
additional extraction, re-distilling your product or even
restarting the whole experiment.
3. The amounts of chemicals you list in the Procedure are
approximately what you will need. Therefore you will need to record
the exact masses and volumes in your Observations.
Always record the appearances of your solvents, reagents, the
colors of solutions, changes in appearance or texture during
heating or cooling, colors of the layers when performing
extractions, colors and appearance of products, etc.
4. When performing calculations, use significant figures
correctly.
5. Present data from IR and NMR spectra in tables. Try to interpret
all relevant signals. “Relevant” is loosely defined as a signal or
peak that is probably from either the product, starting material or
a solvent. You must develop your skills in interpreting spectra so
you can determine whether a given peak or signal is from the
product, starting material, contaminants, etc . Feel free to
consult your instructor if you need some help.
6. The Discussion section should focus on the results. (What did
you make/isolate? What is the purity? What is the % yield/recovery?
Where did losses occur? What problems did you have?) Support all
your claims.
7. Some material usually remains behind when pouring or
transferring chemicals. For example, scraping solid out of flasks
or off of a piece of filter paper usually fails to remove all of
it. Unavoidable losses caused by transferring materials from one
container to another are often
8/18/2019 Lab Manual CHML 2210 Fall 14
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
13/96
13
referred to as “mechanical losses.” Spills and similar mishaps are
not mechanical losses, and they should be addressed in the
Discussion section.
8. Use correct scientific terms. For example, The plural of
“spectrum” is spectra , not spectrums . Use “percent yield” and
“percent recovery” appropriately.
9. Carefully check your spelling and grammar to make sure what you
say makes sense. You may want to have a friend or classmate read
your report to help locate mistakes and parts that should be
revised. Terms that are in Mohrig or the manual should always be
correctly spelled.
10. Keep your Summary section brief, but remember to include
specifics such as % yield, code numbers of unknowns, etc.
11. Use common sense in your Discussion. For example, if your yield
of crude product was 52.4% and after recrystallization your
purified yield was 39.3%, you should not claim that the major
source of product loss was recrystallization. It is true that you
lost a quarter of the crude product’s ma ss during
recrystallization, but more remember that nearly half of the
theoretical yield was lost even before you isolated the crude
product! Other sources of loss are important and should be
considered. Similarly, if some product is spilled or left as
residue on glassware, and if the amount lost is small compared to
the amount you actually isolated, then the transfer losses are not
a major source of loss of product. Critically analyze your
Discussion and your reasoning to avoid making contradictory or
nonsensical claims. Reading it aloud —or even better —having
someone else read it aloud can help spot problems such as flawed
reasoning, contradictory sentences, poor grammar and so forth.
Hint: A clearly written report with sound reasoning and analysis
shows that the writer really understands the experiment. Similarly,
a report with contradictory reasoning, unreasonable (or
unsupported) claims or that ignores important data or results gives
the impression that the writer did not really understand what they
did.
12. Remember to make sure that your carbon copies are
legible!
A Sample Preparative Laboratory Notebook The following pages show
an example of how a preparative experiment should be written up in
a notebook. Remember that if your notebook is organized, your
carbon copies can serve as the majority of your report.
Ben Zeen Sunday, Feb. 30
Experiment 6: Preparation of Ethyl 4-Aminobenzoate
(Benzocaine)
Purpose: The goal of this experiment is to prepare benzocaine from
p-aminobenzoic acid. Theproduct will be purified by
recrystallization and the purity checked by m.p., IR, and
NMR.
Reaction:
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
14/96
Compound MW m.p., C
comments
p- Aminobenzoic acid (PABA) 137.136 187-9 white crystals* Ethanol
(EtOH) 46.068 78 0.789 flammable
Sulfuric acid, conc. (18 M) corrosive Ethyl p -aminobenzoate
165.189 92 Off-white crystals* Ether 35 0.71 very flammable
Other materials used: *11 th Ed. Merck Index
water 10% Na2CO3 solution saturated NaCl solution anhydrous MgSO
4
Procedure Observations 1. Add 0.020 mol of p -aminobenzoic acid
to
a 100-mL round-bottom flask Add 40 mL of 95% EtOH and swirl to mix.
Not all
solid will dissolve.
PABA is white fine crystals. 0.020 mol ~2.7 g. 2.67 g was actually
used. EtOH is clear liquid. Used 41 mL Swirling
dissolves most solid to give pale yellow sol ’n.
2. Cool flask in ice-water bath, then slowly add 2.5 mL of
concentrated sulfuric acid. Lots of precipitate will form, but it
will dissolve during reflux.
Acid was thick clear liquid. Adding acid caused some spattering.
Lots of white-yellow precipitate formed and sank to the bottom of
the flask. Solution turned pale yellow
3. Add a small magnetic stir bar as boiling stone. Assemble simple
reflux set-up with flask and reflux for 75-90 minutes. Stop the
reflux every 15 minutes to vigorously
swirl the flask to help solid to dissolve.
As solution boiled, it turned yellowish and solid began to
dissolve. Some solid was crust
just above liquid. Crust dissolved when flask was swirled after 15
minutes of reflux. All
solid dissolved after about 25 min. of reflux.Sol’n stayed yellow.
Reflux stopped at 75 min. 4. After reflux period, remove
heating
mantle and let mixture cool for 10 minutes, then cool with
cold-water bath.
No changes visible after initial cooling. No changes after cooling
in ice bath.
5. Pour contents of flask into 250-mL beaker. Rinse flask with 2-3
mL of ethanol
Had to use about 5 mL of ethanol to wash out the flask. Solution in
beaker is yellow.
6. Slowly add 10% Na2CO3 solution (with lots of stirring) until gas
evolution ceases and the
solution is strongly alkaline (pH of 9-10 asshown by test strips)
About 35 mL of base should be needed. Ppt. should form.
Na2CO3 solution is water-like. Lots of foaming as added to the
beaker. Some solid particles
deposited on sides of beaker. Fizzing stoppedafter about 30 mL of
solution added. Lots of ppt formed then.
7. Pour mixture into separatory funnel. Add 40 mL of ether to
dissolve as much of the solid benzocaine as possible. Stopper and
shake funnel. VENT! Separate layers.
Ether is clear liquid. Ether solution after shaking is slightly
yellow. Aqueous layer (bottom) is sort of yellow, but cloudy,
too.
8/18/2019 Lab Manual CHML 2210 Fall 14
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
15/96
15
8. Pour aqueous layer back into sep. funnel and extract with 20 mL
of ether in same way as previous step. Separate layers and add
ether layer to earlier ether solution.
Ether solution after second extraction is nearly colorless.
Combined ether solution is pale yellow.
9. Wash ether layer with 10 mL of sat’dNaCl solution, then dry
using MgSO 4 in Erlenmeyer.
No change in appearance of organic solution.MgSO4 made the “snow
globe” when swirled.
10. Gravity filter solution into 24/40 joint round-bottom flask.
Rinse flask and filter paper with a little ether.
Used 5 mL of ether to rinse flask and filter paper.
11. Add a glass bead as boiling stone and remove ether and ethanol
using rotary evaporator. Scrape solid from flask. Weigh and obtain
crude melting point.
Removal of solvent gave pale yellow residue stuck to sides of
flask. Easy to scrape from sides but some left. Yellowish powder.
2.55 g 2.05 g crude product (damp). m.p = 86- 90 C.
12. Purify solid by mixed-solvent recrystallization using just
enough boiling- hot ethanol to dissolve product. (10-13 mL)
Used 11 mL of hot ethanol. Gave yellowish solution.
13. After solid dissolves, add warm water drop-wise with stirring
until the cloudiness does not disappear. Cool on wood block 10 min.
Then chill in ice bath for 15 min.
Cloudiness formed after adding about 1 mL of water. More solid
formed as solution cooled. Solid was whiter in color than the crude
solid.
14. Vacuum filter, rinsing w/ ethanol. Obtain
final mass and m.p. Obtain IR, NMR, and submit sample in labeled
vial.
Expect yield of purified product of 40-60%
Solid was creamy white. Filtrate was
yellowish. 1.39 g of dry pdt. obtained. m.p. was 90- 91 C. IR, NMR
spectra obtained using CDCl3 solvent and attached to report.
The IR was obtained by allowing a CH 2Cl2 solution to evaporate on
a salt plate.
8/18/2019 Lab Manual CHML 2210 Fall 14
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
16/96
16
Ben Zeen IR Spectrum of Benzocaine (CH 2Cl2 solution evaporated on
a salt plate)
Analysis of the IR spectrum of the product gave the following
information:
(The absorptions that were intense or which were easily identified
are listed.)
Absorption, cm -1 intensity / shape Interpretation
3424 medium / fairly sharp Overtone of C=O 3346 + 3225 2 med-strong
bands N —H stretches of NH 2
3047 sharp / weak aromatic C –H stretch 2986-2900 several weak
bands C-H stretches in alkyl groups 1687 very strong / sharp C=O
stretch of ester 1634 very strong N —H bending (probably) 1601
strong / sharp C=C stretch of aromatic ring
1518 strong / sharp C=C stretch of arom. ring (probably)1476 weak /
sharp (unsure) 1450 weak / sharp C-H bend of alkyl groups 1367 weak
/ sharp C-H bend of CH 3 group 1312, 1174 very strong C —O of ester
849 medium / sharp o.o.p. C-H bends in 1,4-disub arom. ring
Summary of IR interpretation I expected benzocaine to show two
bands from the NH 2 group at around 3300 cm -1, plus absorptions
from the aromatic ring at near 3050, 1600, 1500 and around 835 cm
-1. This is basically what was seen. The absorptions at 3000 and at
1450 and 1380 cm -1 were from the alkyl groups. The carbonyl group
was not at 1735 cm -1, but instead it was at 1687 cm -1. Two strong
absorptions at 1312 and 1174 cm-1 are probably the C —O stretches
from the ester. There also is an unexpected, intense absorption at
3424 cm -1. I did not observe any strong, broad O —H near 3400 that
would have been visible if water or residual alcohol had been
present. The 3424 band is fairly sharp, so I believe it is from
something else. It is almost exactly twice the frequency of the C=O
band, so I think it is an overtone of the C=O stretch, but this is
far more intense than any I have seen one before. Maybe the amine
group affects the C=O group somehow.
8/18/2019 Lab Manual CHML 2210 Fall 14
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
17/96
17
I believe that the C=O is the band at 1687 cm -1, but I do not know
why it is below 1700. Wade, chapter 21.4 says the C=O for
conjugated esters is around 1710-1725 cm -1, and the book shows C=O
of methyl benzoate at 1723 cm -1. For now, I am assuming that the
band at 1687 is the C=O, because there is nothing else that seems
to fit. Maybe the amine group lowers the stretching below 1700. I
was unable to find any information in Wade or Mohrig that explained
the result.
NMR SPECTRUM Ben Zeen 1H NMR spectrum of my Benzocaine. Solvent is
CDCl 3.
3H, triplet
Table of NMR Data Chemical Integration Splitting Assignment Shift,
ppm
7.8 2H doublet A 6.6 2H doublet B 4.3 2H quartet C 4.1 2H br.
singlet D 1.3 3H doublet E
Data TableMass of 4-aminobenzoic acid (PABA): 2.67 g Volume of
ethanol: 41 mL Mass of crude product: 2.05 g Mass of recrystallized
product: 1.39 g
Calculation of amounts of reagents used and determining the
limiting reagent:
H 2 N O
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
18/96
18
Moles of PABA: 2.67 g of PABA x 1 mole PABA = 0.0195 moles PABA
137.136 g PABA
Moles of EtOH
41 mL of EtOH x 0.95 x 0.789 g EtOH x 1 mole EtOH = 0.70 moles EtOH
1 mL of EtOH 46.068 g EtOH
The “0.95” corrects for the fact that the ethanol used contained
95% EtOH, 5% H 2O.
PABA is the limiting reagent.
Calculation of the theoretical yield of Benzocaine:
0.0195 moles PABA x 1 mol of Benzocaine = 0.0195 moles
benzocaine
1 mol of PABA = 0.0195 moles benzocaine x 165.189 g benzocaine =
3.22 g benzocaine
1 mole benzocaine
Calculation of crude yield of benzocaine (assuming it is pure,
which it isn’t):
2.05 g crude yield x 100% = 63.7 % crude yield 3.22 g theoretical
yield
Calculation of the actual yield of purified Benzocaine:
1.39 g actual yield x 100% = 43.1 % yield of purified product 3.22
g theoretical yield
Discussion of results: I believe that I successfully prepared a
sample of fairly pure benzocaine. The melting range of my product
was 90- 1 C, which is narrow and close to the accepted value for
pure benzocaine (92 C). This is one indication that my product is
rather pure. The physical appearance of my product also fits the
description in the Merck Index.
The IR spectrum fits what would be expected for benzocaine. The NH
2 group is visible as two absorptions at 3346 and 3225 cm -1. The
ester group is shown by the strong C=O band at 1687 cm -1 and the C
—O bands at 1312 and 1174 cm -1. The aromatic ring is shown by the
C-H stretching band at 3047, the C=C stretches at 1601 and
(probably) 1518 and the o.o.p C-H bend at 849 cm -1. The o.o.p bend
also showed that the compound was 1,4- disubstituted, as would be
expected. There were no broad O-H stretches near 3300 cm -1
that would be present if any unreacted PABA, residual ethanol or
water were present, so I know that these are not in my
product.
8/18/2019 Lab Manual CHML 2210 Fall 14
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
19/96
19
I do not know why the C=O stretching frequency is below 1700 cm -1,
but my spectrum does closely match the one the IR spectrum of
benzocaine found at the Sigma-Aldrich website
(www.sigmaaldrich.com/spectra/ftir/FTIR000308.PDF) . The online IR
spectrum also has the band at 1690 cm -1, and nearly matches my IR
spectrum. Therefore, I believe that the IR suggests that my product
is quite pure.
The NMR spectrum is what would be expected for benzocaine. The
ethyl group is shown by the 3H triplet and 2H quartet. The broad
singlet at 4.1 ppm is the amine hydrogens. The aromatic H’s are the
two 2H doublets at 6.6 and 7.8 ppm. The pattern indicates a 1,4 -
disubstituted aromatic ring. The carbonyl group deshields the H’s
ort ho to it, so they are the ones that are further downfield.
There were no unexplained peaks present, so the NMR spectrum also
indicates that my product is pure.
My 43.1 % yield of purified product is within the 40-60% range
suggested by the lab manual, but this means that over half of the
potential amount of product was lost. The
most obvious source of loss was the recrystallization step. Some
benzocaine remained dissolved in the ethanol, so it was lost in
this step. About 0.66 g of mass was lost, but the crude product was
still damp (probably ether and/or ethanol), so the amount of solid
lost here is less than 0.66 g, but it is not possible to know how
much it was.
Also, the reaction was probably incomplete. I stopped the reflux at
75 minutes, instead of letting it go 90 minutes. The Introduction
in the Lab Manual says the reaction was a Fischer esterification
reaction, and these reactions are equilibria that do not go to
completion. If I stopped too soon, then the reaction may not even
have reached equilibrium. The crude yield was less than 64 %, so
about a third of the potential yield of
product was lost even before isolating the crude product. However,
there is no way tocalculate how much of the product never was
formed by stopping the reflux so soon.
Some product was probably lost in the extractions, drying and
filtering, but I took care to rinse my glassware to try to keep
this to a minimum. Mechanical losses, such as not being able to
scrape all of the product from the flask and traces left on the
glassware and filter and filter paper, probably added up to a few
percent of the loss. I saw no large deposits of residue on any
glassware that would suggest greater losses.
Summary:
Benzocaine was successfully prepared in 43.1 % yield. The product
appeared to be quitepure, as judged by the narrow melting point
range, appearance, as well as the IR and NMR spectra.
Recrystallization was a major source of product loss, but it was
clear that the reaction did not proceed to completion.
[Note: Any additional spectra (such as those used for comparison or
provided by the instructor), along with any assigned homework
problems, are attached to the lab and turned in as well.]
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
20/96
WRITING LABORATORY REPORTS
The cornerstone of science is communicating experimental results.
For each experiment, you will submit a written report that
describes the reaction you carried out, the procedures you
followed, the observations you made and the data you obtained and
your interpretation of the results. Lab reports are made up of
several sections as described below. Your instructor will provide
you specific information about what is to be included for
particular experiments. If you are unsure what to include something
in your report, just ask your instructor.
The following is an example of an excellent organic lab report. Use
it as a guideline for your own reports, but don’t plagiarize!
Abstract Observing the separation made by the spinach extract on
the thin layer chromatography plates, a solution of 7:3
toluene:acetone was identified as the best eluent for calculating
retardation factors. Compounds and Rf values were experimentally
calculated as follows. The following compounds separated; B-
carotenes, pheophytin a, pheophytin b, chlorophyll a, and
chlorophyll b, along with their respective Rf values; .98, .90.
.68, .61, and .53. Two fractions were collected from the column
chromatography experiment, along with a sample of the original
spinach extract. Fraction one resulted in a single compound with Rf
value of .98. Fraction two resulted in two compounds with Rf values
of .45 and .10. The original spinach extract resulted in four
compounds with Rf values of .98, .63, .45 and .22. TLC was used for
the identification of compounds in an unknown pain reliever based
on Rf values for Tylenol and Bayer. The unknown resulted in three
compounds with Rf values of .64, .46, and .24. Tylenol was
experimentally found to have a Rf value of .45. Bayer was
experimentally found to have an Rf of .5.
Introduction Thin-layer chromatography is a technique used to
separate compounds with in mixtures. This is useful for evaluating
the purity of compounds and monitoring the progress of a reaction.
TLC consists of two
phases, a mobile phase and a stationary phase. TLC functions by the
differences in polarity of the compounds with the solvent used,
result in varying migration in well defined and separated
spots.
Reagents and Equipment Reagents used in this experiment include
methanol, hexanes, water, sodium sulfate, toluene, acetone,
spinach, spinach extract from column chromatography experiment, and
pain relievers. The equipment used includes a mortar and pestle,
beakers with glass watches, glass funnel, cotton plug, 125 mL
separatory funnel, Erlenmeyer flask, Pasteur pipets, and test
tubes.
Procedure and Observations Ten grams of spinach was cut, dried and
drained before weighing. The spinach was added to the mortar with
twelve milliliters of methanol and crushed for three minutes
resulting in a bright green mixture. The methanol was then
discarded. Fifteen milliliters of hexanes and five milliliters of
methanol was added to the mortar with the leaves and crushed for
five minutes resulting in a dark green mixture. The liquid
was
then transferred to a beaker and filtered with a cotton plug into
the separatory funnel. Three layers formed. A dark green layer
formed on top followed, by a lighter green layer followed, by an
almost clear layer. Five milliliters of water was added to the
separatory funnel. Funnel was inverted and periodicall y vented.
The dark green organic layer stayed on top. The lower aqueous layer
was extracted and discarded. The organic layer in the separatory
funnel was then rinsed three more times, resulting in the aqueous
layer becoming more clear with each rinse. The organic layer was
then separated and transferred to an Erlenmeyer flask. Anhydrous
sodium sulfate was added to absorb remaining water. Solution was
then filtered with a cotton plug and funnel into a test tube. The
test tube was sealed for one week. With the test tube concentrated
down to one to three milliliters the experiment was continued one
week later.
8/18/2019 Lab Manual CHML 2210 Fall 14
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
21/96
21
Five development chambers were prepared in beakers. Solvents used
included hexanes, toluene, 9:1 toulene-acetone, 7:3
toulene-acetone, and acetone. TLC plates were prepared for each
development chamber. The 9:1 toulene-acetone was used to determine
the correct size of compound to be used. The 7:3 toulene-acetone
was found to have the most separation. The spinach extract from the
column chromatography experiment was then dissolved in a small
amount of the solvent used to collect the fraction. Each fraction
and remaining extract was placed on a TLC plate in the 7:3
toulene-acetone developing chamber. Tylenol, Bayer, and an unknown
pain reliever was respectively crushed to a fine
powder and placed in separate test tubes. Each compound was
dissolved with a 1:1 mixture of methylene chloride and ethanol
solution. Using .5% acetic acid in ethyl acetate as the solvent,
TLC plates were labeled and ran. The unknown was paired with each
comparative pain reliever on a separate TLC plate. The Unknown
resulted in three separate compounds. The Tylenol and Bayer each
resulted in one compound. Ultraviolet light was used to visually
see the compounds and mark the TLC plate appropriately.
Data The following is a table of the spinach extract with 7:3 v/v
toluene:acetone used as the eluent. TABLE 1 Compound Rf B-carotenes
.98 Pheophytin a .90 Pheophytin b .68 Chlorophyll a .61 Chlorophyll
b .53
The following is a table of the fractions and extract from column
chromatography experiment with 7:3 v/v toluene:acetone used as the
eluent.
TABLE 2 Compound Rf
.10
.63
.45
.22 The following is a table representing the unknown pain
reliever, Tylenol, and Bayer with .5% acetic acid in ethyl acetate
used as the eluent. Ultraviolet light was used to visually see the
compounds. TABLE 3 Compound Rf Unknown .64
.46
8/18/2019 Lab Manual CHML 2210 Fall 14
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
22/96
22
Results Based on the separation of the spinach extract using
hexanes, toluene, 9:1 toluene-acetone, 7:3 toluene acetone and
acetone as the eluents, 7:3 toluene acetone was found to have the
best separation of compounds. Acetone was found to not be a good
solvent to separate the pigments due low polarity. Taking the Rf
values of each compound and comparing to the polarity of each
pigment, the correlation is a direct relation. As the polarity of
the compound decreases so does the Rf value of the compound. Using
the spinach extract from part one as a basis of comparison for the
spinach extract from the column chromatography experiment, fraction
1 was found to contain B-carotene. Based on the polarity of the
carotenes and the pigment separations of the other eluents tested,
a mixture of hexanes and toluene would
provide the best separation. Hexanes provided a minor movement of
the carotenes with little movement of the rest of the pigments due
to the differences in polarity. Toluene provided much more movement
up the TLC plate of the carotenes and remaining pigments but not
enough separation. To achieve an acceptable Rf value, close to but
not greater than .75, a mixture of the two solvents is necessary.
Fraction 2 was found to contain chlorophyll b. Chlorophyll a has a
methyl group in the position chlorophyll b has an aldehyde group,
making chlorophyll b more polar. Based on this assessment, the
original appearance of the fraction and the proximity of Rf values,
chlorophyll b was determined as the compound. The additional Rf
compound found in fraction 2 is too far from the range of
experimental Rf from any data collected in table 1 to be
determined. The column chromatography spinach extract was found to
contain B-carotenes, chlorophyll a chlorophyll b and an additional
compound that was too far from the range of experimental Rf values
in table 1 to be determined. The unknown pain reliever was found to
have three separate compounds with similar components of both
Tylenol and Bayer. The Rf values for Tylenol and Bayer were very
close in proximity, along with a component in the unknown. This
indicates that the unknown contains a component of both the Tylenol
and the Bayer. The overall TLC plates of the pain relievers were
not as clean in separation as the spinach plates. The compounds
came out as large smears indicating overloading of the TLC plate.
For future experiments, a technique of using less compound on the
TLC plate will prove to provide better data.
Follow up questions (deleted so that you can come up with your own
answers)
References Techniques for Characterizing and Purifying Organic
Compounds. CHML 2210-70B Brevard Community, Palm Bay Campus. Fall
2013 Edition.
8/18/2019 Lab Manual CHML 2210 Fall 14
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
23/96
Characterization and Purification
There are four problems that confront organic chemists on a daily
basis: 1. is the sample I have pure? 2. if it is not pure, what can
I do to remove the contaminants? 3. if it is pure, what properties
does the sample have which will enable me to recognize the
same compound if I obtain it again from a different reaction
source? 4. what is this stuff anyway?
Virtually any method, which can be used to characterize a compound,
will also serve as another check of the purity of the sample in
question. No single means of characterization is sufficient to
establish the identity of a previously known compound and no single
means for indicating purity is sufficient to establish purity. Good
experimental work in the characterization and purification area
systematically includes several indicators of good purity and an
array of sample properties to characterize the compound in
question. Multiple redundancies in information make
characterization possible.
In this course, the first half of organic chemistry, you will learn
many basic techniques. All of these techniques have application
beyond the chemistry laboratory (biology uses chromatography
extensively).
Washing Dishes
In order to characterize and purify compounds successfully and to
carry out actual reactions, one must start each experiment with
clean glassware. The easiest way to obtain this is to do your
dishes at the end of the previous experiment since many procedures
need to be done under dry conditions. This isn’t trivial because
many organic compounds are not very soluble in water. Suggested
steps are as follows:
1. Remove grease (if any) from ground-glass joints. If stopcock
grease has been used, it must be removed before washing is
attempted, otherwise grease distributes itself over everything. To
de-grease joints, moisten a paper towel with ether or
dichloromethane and wipe the grease off the joint. Ether is
marginally better but is flammable.
2. Scrub the glassware with soap and water. Use a brush. 3. Rinse
the soap off with water. 4. Visually inspect the glassware to see
if it’s clean. If so, you’re done. If not, a little more
work is needed. 5. If the glass isn’t clean, get it soapy again;
skirt acetone onto the s tain and scrub again.
Use a brush. Wet, soapy acetone is a pretty good general solvent
for slime. 6. Continue steps 3-5 until glass is clean.
8/18/2019 Lab Manual CHML 2210 Fall 14
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
24/96
C. Heath and Co. Lexington, MA.
COOH
napthalene cinnamic acid
Introduction: The melting point and boiling points of a pure solid
organic compound are characteristic physical properties, along with
molecular weight, refractive index and density. Both melting point
and boiling point are affected by the forces that attract one
molecule to another: ionic attraction, van der Waals forces,
dipole-dipole interactions and hydrogen bonding.
A pure solid will melt reproductively over a narrow range of
temperatures, less than 1 0C. Using less than 1 mg of material, a
melting point can be determined. The melting point apparatus is
very simple, consisting of a thermometer, a capillary tube to hold
the sample and a heating bath. Melting points are determined for
three reasons: 1) if the compound is known, the melting points will
help characterize the sample in hand 2) if the compound is new then
the recording of the melting point is done to allow for further
characterization by others 3) The range of the melting point is
indicative of the purity of the compound; an impure compound will
melt over a wide range of temperatures. Boiling point is used to
characterize a new organic liquid and knowledge of the boiling
point helps to compare one organic liquid with another. Comparison
of boiling points with melting points is instructive but process
for determining boiling points is more complex. Boiling point
determination requires more material and is affected less by
impurities, it is not as good an indicator of purity as melting
point.
Procedure: Part I: Calibration of the thermometer: Determination of
the melting point of standard substances over the range of interest
gives the difference between the values found and those expected.
This difference constitutes the correction that must be applied to
future temperature readings.
IMPORTANT! The liquid in the thermometers is very thin and can
collect in the top part of the thermometer, resulting in inaccurate
readings. Always store your thermometer with the bulb end
lowermost. Also, check your thermometer before use to make sure
that the thread of liquid is continuous and that the top is free of
liquid.
Thermometer Calibration (Your instructor may tell you to skip this
part; check your syllabus.) The thermometers used in this course
are fairly accurate (±1 C) and usually do not nee d to be
calibrated, so your instructor may have you skip this step. If you
are instructed to calibrate your thermometer, follow the
instructions given below.
Why carry out a thermometer calibration at all?
8/18/2019 Lab Manual CHML 2210 Fall 14
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
25/96
25
When measuring temperatures, the upper part of the thermometer is
cooler than the bulb, and the liquid in the upper part of the
thermometer contracts slightly. The resulting reading can be 1-3
degrees below the temperature of the bulb. Many thermometers are
accurate enough to need no correction, but a calibration curve
corrects for less-accurate thermometers.
To prepare a calibration curve, record the identities and true
melting ranges of three or four compounds provided by the
instructor. Obtain reliable melting ranges for these solids and
record them in your notebook. Determine the average values for both
the observed and the true melting ranges for each compound. .
To prepare your calibration curve, label the X- axis as “Observed
Temperature” and the Y -axis as the “Corrected” or “True”
Temperature. Start numbering your graph at 30 or 40, using 1 per
division on the
paper. Next, plot the melting points of your known compounds on the
graph. Use small dots for the data points. Draw a straight line
which best fits these points (use pencil first). Your line will
probably not pass through the origin. Show your calibration curve
to the instructor for checking. Save your approved calibration
curve for use throughout the semester, and label your thermometer
with your name and locker number. If you break or lose your
thermometer, you will need to prepare a new calibration curve for
your new thermometer.
Converting measured temperatures to corrected values is simple: The
corrected temperature (Y-axis) is where the measured temperature
(X-axis) intersects with the calibration line. You need to do this
with
both the lower and upper values of a temperature range. Note: Do
not use averages of temperatures in Part B or any future
experiments; average temperatures are only used when constructing a
calibration curve!
Take a 10 mg sample of trans -cinnamic acid and prepare a melting
point tube by pushing a melting point capillary into the powder and
force the powder down in the capillary by tapping the capillary or
by dropping it through a long glass tube held vertically and
resting on a hard surface. The column of solid should be no more
than 2-3 mm in height and should be tightly packed. Take a melting
point range of this compound using the Mel-Temp apparatus in the
lab. The temperature can be raised rapidly until the temperature is
about 20 0C below this point. Then slow down heating considerably
so that rate of increase is no more than 1 0C per minute while the
sample is melting. Compare the melting point range with the
published melting point of t-cinnamic acid. Record the difference
between published MP and the MP you obtained. Repeat the same
procedure using urea. Record the differences. If the differences
recorded for t-cinnamic acid and urea agree within 1 0C, use this
correction for every melting point you perform in this laboratory,
provided that the same thermometer is always used. Write this
correction in your lab report.
Part II: Melting Points of urea and cinnamic acid mixtures. Make
mixtures of urea and cinnamic acid in the approximate proportions
1:4, 1:1, 4:1 by putting side by side the correct number of
equal-sized small piles of the two and then mixing them. Grind the
mixtures thoroughly for at least a minute with a mortar and pestle.
Not the ranges of the three mixtures and use the temperatures of
complete liquefaction to construct a rough diagram of mp versus
composition.
8/18/2019 Lab Manual CHML 2210 Fall 14
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
26/96
Record this information in your notebook:
Unknown Number: __________ Appearance of Unknown:
___________________ m.p. of Unknown (uncorrected):__________C (If
applicable): Corrected m.p. value: __________C
Determine the melting point of the unknown given by your instructor
and on the basis of that melting point, identify that substance.
Run two capillaries of the unknown. Run a very fast determination
on the first to obtain approximate melting point and then use the
second to do a slow, careful determination. Record your results and
identify the compound you tested in your lab report.
Table 1.1 Melting Points of Unknowns Compound Melting Point (
C)
Benzophenone 49-51 Naphthalene 80-82 Benzoic Acid 121.5-122 Urea
132.5-133 Salicylic Acid 158.5-159
Clean Up: Place all leftover solids in the solid waste bottles
provided. Capillary tubes can be placed in the broken glass box in
the laboratory.
Suggested topics for Discussion What is the identity of your
unknown? What is your rationale for your identification and how
confident are you in your identification of your unknown? What
problems (if any) did you have during the experiment? Feel free to
include additional topics as appropriate.
Follow-up Questions (It is usually helpful to consult your lecture
text when answering these.) 1. A student prepared a mixture of two
compounds. The mixture showed a narrow melting
range, but the value is considerably lower than the melting range
of either component in the mixture. Explain.
2. A student attempts to obtain a melting range, but he filled the
capillary with five times as much sample as necessary. How might
this affect the observed melting point? Explain.
3. A student was trying to obtain the melting point of borneol.
While heating the sample, he stepped away from the apparatus for a
few minutes. He saw that the capillary tube was empty, even though
the temperature was well below the listed boiling point of borneol.
What probably happened? What should the student do to obtain the
melting range of his solid?
8/18/2019 Lab Manual CHML 2210 Fall 14
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
27/96
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
28/96
Chlorophyll a Chorophyll b Blue-green, polar Green, polar C55H72MgN
4O5 C55H70MgN 4O6 M.W. 893.5026 M.W. 907.4862
-carotene Yellow, nonpolar C40H56 M.W. 536.8824
CH3 CH 3 CH 3 CH 3
CH 3 R
Objective(s): To separate the color pigments in spinach by column
chromatography.
8/18/2019 Lab Manual CHML 2210 Fall 14
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
29/96
Technique(s) used: Isolation of a natural product; extraction;
preparation and use of column chromatography.
Zubrick 7 th Ed. – Column Chromatography and TLC page 221 and page
239 Microscale column chromatography Zubrick 7 th page 243
CAUTIONS : Avoid inhaling the solvent fumes; keep the stoppers in
the test tubes.
Procedure:
At the beginning of the lab set up a hot water bath on a hot plate
and watch the water level so it doesn’t get too low in the beaker
(hot water bath).
Extraction of the pigments:
Weigh out 5 - 6 grams of spinach. Remove as many of the stems as
possible, then put the leaves in a mortar with a scoop of
sand and 5 mL of methanol. Grind the mixture with a pestle for 10
minutes. Pour the ground mixture into a large test
tube and add about 10 mL of hexane. With a spatula or stirring rod,
mix the hexane with the ground spinach.
Decant the hexane into another test tube. Be sure to press the
spinach to get out as much of the hexane as possible. To the hexane
in the second tube, add about 10 mL of DI water, stopper the tube
and shake, but not too hard. Remove the lower aqueous layer with a
pipet and repeat the extraction. This will remove most of the
methanol from the hexane.
Dry the hexane by doing the following: set up a vacuum filtration
with a Buchner funnel and filter paper. Place a scoop of drying
agent, the anhydrous sodium sulfate, on the filter paper. Turn on
the water to the aspirator to do a vacuum filtration. Carefully
pour the green hexane solution into the Buchner funnel over the
drying agent and filter it. Alternate procedure: Put a scoop of
anhydrous sodium sulfate drying agent in the flask. Swirl and let
sit 5 minutes. If the drying agent is still clumping, add more
until it moves like sand in bottle. Using a small funnel, put a
small piece of a cotton ball loosely over the hole to the stem.
Gravity filter the solution through the cotton ball to separate the
extract from the drying agent. Save the green hexane spinach
extract solution in a clean test tube.
Add a boiling chip to the test tube and evaporate the solution to
about 0.5-mL in a hot water bath on a hot plate. This is the green
hexane spinach extract that you will use with your column.
8/18/2019 Lab Manual CHML 2210 Fall 14
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
30/96
Figure 1. Pasteur pipet wet-column chromatography
Set up a microscale column as shown in Figure 1. Obtain a
chromatography tube such as a Pasteur pipet. Shorten the tip of the
Pasteur pipet; if it is not already shorten. Gently press a small
glass wool plug into the tip of the tube - being careful not to
pack it too
tight. Add about 50 mg of sand on top of the glass wool.
Weigh about 500 mg of alumina and break up any clumps with the
stirring rod (if necessary,use a mortar and pestle to make a fine
powder of alumina). Slowly add the alumina to the column. Every so
often, thump the column with your finger to aid in settling the
alumina.
After all the alumina has been added, place about 50 mg of sand on
top of the alumina.
The chromatography separation: Clamp the column on a ring stand. In
four large test tubes add the following elution solvents:
1 st test tube 10 mL of hexane 2 nd test tube 7 mL hexane + 3 mL
toluene (7 / 3 hexane/toluene) 3 rd test tube 3 mL toluene + 7 mL
ethyl acetate (3/7 toluene/ethyl acetate) 4 th test tube 10 mL of
methanol.
Take two pipets; one for the green hexane spinach extract solution
and one for solvents. Prepare to run the column. The microscale
column (Pasteur pipets) doesn’t have a stopcock. Once it starts
running, it goes until it runs out. You can’t let the column
dry
out. Have all your elution solvents and the green hexane spinach
extract solution ready and be prepared to run the column once you
start it.
Now slowly add some hexane into the column and let it wet the
entire column. From this point on, it is extremely important that
the liquid level never be allowed
to get below the top of the sand.
8/18/2019 Lab Manual CHML 2210 Fall 14
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
31/96
31
DO NOT EVER LET THE COLUMN RUN DRY! NEVER LET THE SOLVENT LEVEL
FALL BELOW THE LEVEL OF THE ALUMINA!
Carefully pipet the green hexane spinach extract (from the
extraction of the pigments section above) to the top of your
column. Don’t fill the column with liquid; drip the solution onto
the sand. Leave a small amount for the TLC experiment.
When the green hexane spinach extract is all on the column, just
below the sand, add a small amount of hexane, the first elution
solvent.
Watch your column! Do not ever let the column run dry. Collect the
drops off the end of the column in a small beaker. As long as the
first color band is seen moving down the column, continue with
hexane. As the color band reaches the glass wool plug, place a
small test tube in place of the
beaker to catch the colored solution in the smallest possible
volume. Note in your lab book, the color, solvent, and approximate
volume.
When no color bands are seen moving with your present elution
solvent, change to the next more polar elution solvent.
Continue the process until all the solvents have been used. Collect
each color band in small test tubes as a series of fractions and
label all test
tubes. Any color left at the top of the column cannot be removed
.
Store the reserved extract and fractions collected off the column
in OPEN test tubes to be used next week in the TLC experiment
.
Notebook Report: Keep a detailed record of all your observations
during the extraction of the pigment and the chromatographic
separation. In tabular format record the number of the fractions
being collected, the time over which the fractions were collected,
the concentration of the solvent (10 mL hexane; 7/3 hexane to
toluene…..) on the column, the appearance of colors on the column,
and the color of the solution in the fraction being collected.
Based upon color and polarity, identify each fraction either as
ß-carotene, chlorophyll a , chlorophyll b, or xanthophyll.
Number
of Fraction
Color of Fractions in Test Tube
1. 2. 3. 4. 5. 6.
Under optimum separation, the fractions collected show the
following pigments in spinach: ß-Carotenes (yellow-orange) nonpolar
Pheophytin a (gray, may be nearly as intense as chlorophyll b)
Pheophytin b ( gray, may not be visible) Chlorophyll a (blue-green,
more intense than chlorophyll b) Chlorophyll b ( green) more polar
than chlorophyll a Xanthophylls (yellow)
8/18/2019 Lab Manual CHML 2210 Fall 14
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
32/96
A characterization method and criterion for purity
Thin layer chromatography (TLC) uses glass, aluminum or plastic
plates covered with a layer of
adsorbent. Common adsorbents are silica gel, alumina and cellulose.
The sample is applied asa dilute solution to a position near the
bottom of the plate. The plate is stood on end in a shallow pool of
solvent (called the eluant), which will run up the plate.
As the eluant passes over the applied sample, the molecules, which
are adsorbed on (stuck to) the adsorbent, will tend to dissolve in
the eluant, move ahead in solution, and re-adsorb further up the
plate. The sample will constantly re-dissolve and re-adsorb while
the plate is running (being developed). The more strongly the
sample adsorbs on the adsorbent, the less it will tend to dissolve,
and the more slowly it will move up the plate. Different compounds
will, in general, tend to dissolve, and the more slowly it will
move up the plate. Different compounds will, in general, adsorb
with different strengths, so that different compounds usually move
different distances up a give plate.
Spotting and Developing TLC Plates: To apply a sample to a TLC
plate, it is dissolved in a suitable solvent (try CH 2Cl2 first) at
a concentration of 100 mg per mL of solvent. The solution is
applied to the plate with a double- open ended capillary tube,
~3/8” from the bottom of the plate so that a spot of ~1/8” diameter
is formed (Figure 1).
Figure 1. TLC plate with starting material product and a mixture of
both spotted on the plate.
The appropriate solvent (eluant) for development is poured into a
TLC chamber (either a special chamber or a beaker). The solvent
should be 1/8” deep, so that the spot will wash the sample off the
chamber (otherwise, solvent will wash the sample off the plate).
The plates are then stood in the chamber (or beaker) and the
chamber is covered (with a watch glass). Once the eluant reaches
the top of the plate, the plate is removed from the chamber. Upon
removal, mark, in pencil, where the solvent stopped on the plate.
Lay the plate flat to dry.
Silica gel is very polar. Therefore, polar compounds tend to stay
near the origin, while non-polar compounds tend to move much faster
up the plate. A suitable eluant is chosen by trial and error, and
often differs from the solvent used to spot the sample. If a
compound runs too fast
8/18/2019 Lab Manual CHML 2210 Fall 14
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
33/96
33
(too far up the plate) your next plate should be run with a less
polar eluant. If the compound runs too slowly, your next plate
should be run with a more polar eluant. See Table 1.
Table I. Comparative Polarities of Solvents LEAST POLAR
hexane * cyclcohexane
pyridine acetone
1-propanol ethanol
methanol* water
MOST POLAR
*indicates the “big four” solvents which can be used in mixtures to
separate most organic compounds.
Visualizing Spots . A TLC plate, which has just been developed,
looks (unless the sample is colored) just like an undeveloped
plate: a plain white surface. To locate the spots of compounds,
they must be visualized. Visualization can be done either
destructively or non- destructively.
a) Non-destructive Visualization . The most common non-destructive
technique requires that the TLC plate be prepared with an
adsorbent, which contains a fluorescent indicator (a compound that
fluoresces when you shine UV light on it. The TLC plates that we
will be using have this indicator, which fluoresces green under
254-nm light. If such a plate is spotted and developed, UV light
lets you see the spot(s). The beauty of this kind of visualization
is that you can wash your sample back off the plate and recover it
unchanged; for the closely related techniques thick-layer
chromatography, this advantage is very important.
b) Destructive Visualization . The usual techniques for destructive
visualization involve spraying the plate with a reagent which
immediately reacts with adsorbed sample to give a product which
either I) differs in color from the background color of the sprayed
plates or II) has low volatility, so that vigorous heating will
lead to charring of the sample, giving black spots on a white
background. A wide variety of general-use and special-purpose
sprays exist. Spray visualization is destructive because the sample
is irreversibly converted into another compound. Every method is
unable to detect some samples (for example, 254- nm UV
visualization can’t detect a sample that doesn’t absorb 254
-nm
8/18/2019 Lab Manual CHML 2210 Fall 14
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
34/96
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
35/96
35
still “clean”, you can have good evidence, but not proof, that the
two compounds are the same (Figure 5). It the middle spot splits in
two, you have proved that the compounds are different.
A Measure of Impurity . If your sample is a mixture, and if the
components have different R f values under your TLC conditions, you
will see several spots on the plate after visualization. TLC alone
never tells you how many components a mixture has---it tells you
the MINIMUM number of compounds, which must be present. It is a
real possibility that your visualization method may not show all
the spots present, or that one spot may represent a mixture of
compounds, which have the same R f value (co-elution).
TLC Procedure (http://orgchem.colorado.edu/hndbksupport/TLC/TLC
procedure.html)
1. Prepare the developing container.
The developing container for TLC can be a specially designed
chamber, a jar with a lid, or a beaker with a watch glass on the
top:
In the teaching labs, we use a beaker with a watch glass on
top.
Pour solvent into the beaker to a depth of ust less than 0.5
cm.
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
36/96
To aid in the saturation of the
TLC chamber with solvent vapors,line part of the inside of the
beaker with filter paper.
Cover the beaker with a watch glass, swirl it gently, and allow it
to stand while you
repare your TLC plate.
2. Prepare the TLC plate.
TLC plates used in the organic chem teaching labs are purchased as
5 cm x 20 cm sheets. Each large sheet is cut horizontally into
plates which are 5 cm tall by various widths; the more samples you
plan to run on a plate, the wider it needs to be.
8/18/2019 Lab Manual CHML 2210 Fall 14
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
37/96
37
Shown in the photo to the left is a box of TLC plates, a large
un-cut TLC sheet, and a small TLC plate which has been cut to a
convenient size.
Plates will usually be cut and ready
for you when you come to lab.
Handle the plates carefully so that
you do not disturb the coating of
adsorbent or get them dirty.
easure 0.5 cm from the bottom of the late. Take care not to press
so hard
with the pencil that you disturb the adsorbent.
Using a pencil, draw a line across the plate at the 0.5 cm mark.
This is the origin : the line on which you will "spot" the
plate.
It's kind of hard to see the pencil line in the above photos, so
here is a close-up of how the plate looks after the line has been
drawn.
8/18/2019 Lab Manual CHML 2210 Fall 14
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
38/96
38
Under the line, mark lightly the name of the samples you will spot
on the plate, or mark numbers for time points. Leave enough space
between the samples so that they do not run together, about 4
samples on a 5 cm wide plate is advised.
Use a pencil and do not press down so hard that you disturb the
surface of the plate. A close-up of a
plate labeled "1 2 3" is shown to the right.
3. Spot the TLC plate
The sample to be analyzed is added to the plate in a process called
"spotting". If the sample is
not already in solution, dissolve about 1 mg in a few drops of a
volatile solvent such as hexanes,ethyl acetate, or methylene
chloride. As a rule of thumb, a concentration of "1%" or "1 gram in
100 mL" usually works well for TLC analysis. If the sample is too
concentrated, it will run as a smear or streak ; if it is not
concentrated enough, you will see nothing on the plate. The "rule
of thumb" above is usually a good estimate, however, sometimes only
a process trial and error (as in, do it over ) will result in
well-sized, easy to read spots.
add a few drops of solvent . . .
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
39/96
39
The solution is applied to the TLC plate with a capillary tube or a
microcap, as seen in the photos.
Take a tube and dip it into the solution of the sample to be
spotted. Then, touch the end of the tube gently to the adsorbent on
the origin in the place which you have marked for the sample. Let
all of the contents of the tube run onto the plate. Be careful not
to disturb the coating of adsorbent.
dip the microcap into solution - the arrow points to the microcap,
it is tiny and hard to see
make sure it is filled - hold it up to the light if necessary
touch the filled microcap to TLC plate to spot it - make sure you
watch to
see that all the liquid has drained from the microcap
do this rinse
process 3 times!
rinse the microcap with clean solvent by first illing it . .
.
. . . and then draining it by touching it to a paper towel
8/18/2019 Lab Manual CHML 2210 Fall 14
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
40/96
40
If the microcapillary tube breaks or clogs, you may obtain a new
one.
here's the TLC plate, spotted and ready to be developed
4. Develop the plate.
Place the prepared TLC plate in the developing beaker, cover the
beaker with the watch glass, and leave it undisturbed on your bench
top. Run until the solvent is about half a centimeter below the top
of the plate (see photos below).
place the TLC plate in the developing container - make sure the
solvent is not too deep
The solvent will rise up the TLC plate by capillary action. In
this
photo, it is not quite halfway up the plate.
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
41/96
41
In this photo, it is about 3/4 of the way up the plate.
The solvent front is about half a cm below the top of the plate -
it is now ready to be
removed.
Remove the plate from the beaker.
quickly mark a line across the plate at the solvent front
with a pencil
Allow the solvent to evaporate completely from the plate. If the
spots are
colored, simply mark them with a pencil.
5. Visualize the spots
If your samples are colored, mark them before they fade by circling
them lightly with a pencil (see photo above).
Most samples are not colored and need to be visualized with a UV
lamp. Hold a UV lamp over the plate and mark any spots which you
see lightly with a pencil.
Beware! UV light is damaging both to your eyes and to your skin!
Make
sure you are wearing your goggles and do not look directly into the
lamp.Protect your skin by wearing gloves .
If the TLC plate runs samples which are too concentrated, the spots
will be streaked and/or run together. If this happens, you will
have to start over with a more dilute sample to spot and run on a
TLC plate.
8/18/2019 Lab Manual CHML 2210 Fall 14
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
42/96
42
this is a UV lamp here are two proper sized spots, viewed under a
UV lamp
(you would circle these while
viewing them) The plate to the left shows three compounds run at
three different concentrations. The middle and right plate show
reasonable
spots; the left plate is run too concentrated and the spots are
running together, making it difficult to get a good and accurate R
f reading.
Here's what overloaded plates look like compared to well-spotted
plates. The plate on the left has a large yellow smear; this smear
contains the same two compounds which are nicely resolved on the
plate next to it. The plate to the far right is a UV visualization
of the same overloaded plate.
NOTE: Your instructor will indicate which parts of the experiment
are to be performed during the laboratory period.
PART ONE: Identification of Components in Spinach by Thin-layer
Chromatography (Week 1- follow Instructor’s directions)
PROCEDURE:
Reference: Palleros, D. A. Experimental Organic Chemistry . 2001,
J. Wiley and Sons, New York, pgs. 193-195.
8/18/2019 Lab Manual CHML 2210 Fall 14
http://slidepdf.com/reader/full/lab-manual-chml-2210-fall-14
43/96
43
Vegetable Extract Weigh out 10 g of spinach leaves and chop them up
using scissors. If they were previously frozen, drain as much water
out of the leaves as possible before weighing.
Place the spinach in a large mortar, add 12 mL of methanol, and
crush the leaves with a pestle for about 3 minutes. With the aid of
the pestle or spatula, squeeze the spinach against the side wall of
the mortar to remove as much methanol as possible. Transfer the
liquid to a 250-mL Erlenmeyer flask, labeled “methanol -water” and
set aside. The contents of the flask will eventually be
discarded.
Extract the remains of the leaves with a mixture of 15 mL hexanes
and 5 mL methanol, crushing the tissue with the pestle for about 5
minutes. The extract should be deep green. Leaving behind as much
solid as possible, transfer the liquid (a mixture of hexanes and
methanol) to a 100-mL beaker. Filter it using a glass funnel (4-6
cm diameter) with a cotton plug. Collect the filtrate directly into
a 125-mL separatory funnel supported on a ring stand (make sure the
stopcock is closed).
Add 5 mL of water to the separatory funnel, shake, vent and allow
the layers to separate. Which layer is the aqueous layer? Collect
the aqueous layer in the Erlenmeyer flask labeled “methanol
-water.” Extract the remaining hexane layer with 5 mL of water in
the separatory funnel. Collec t the aqueous layer along with any
emulsion present in the “methanol -water” flask.
Transfer the organic layer to a clean and dry 50-mL Erlenmeyer
flask; add a small spatulaful of granular sodium sulfate (Na 2SO 4)
to dry the organic layer. Cap the flask with a stopper and swirl
occasionally. After about 5 minutes, filter the suspension by using
a clean and dry micro- funnel (about 2.5 cm in diameter) and a
cotton plug. Collect the filtrate in a dry 50-mL round- bottom
flask. Using the rota-vap, evaporate the solvent until the volume
is approximately 2-3 mL. The color of the final extract should be
deep green. With the aid of a Pasteur pipet, transfer the liquid to
a labeled test tube.
TLC Analysis
Clean and thoroughly dry five 150-mL beakers and five watch glass
covers. Label them with the names of the solvents to be used: 1)
hexanes 2) toluene 3) toluene-acetone (9:1) 4) toluene- acetone
(7:3) 5) acetone.
Determining the Optimum Number of Applications. Take the plate
labeled “toluene - acetone” (9:1). This plate will be used to
determine the optimum number of applications needed to visualize
the separation of pigments. On this plate, the sample will be
applied on three different spots, each with a different number of
applications (1, 4 and 7). Immerse a double open-ended capillary
tube in the extract; the liquid will rise through capillary action.
Apply the
liquid to the plate by touching the plate with the tip of the
capillary tube and applying slight pressure. If necessary, gently
wiggle the capillary to make the liquid flow. The spot should be
about 0.5 cm from the side edge of the plate. Raise the capillary
tube to stop the flow of liquid when the diameter of the spot is
about 3 mm. On the same plate, apply a second and a third spot at a
distance of about 0.5 cm from each other. The number of
applications should be 4 and for the third spot, 7. Allow the
solvent to evaporate between applications. Failure