Date post: | 07-Aug-2015 |
Category: |
Documents |
Upload: | layal-jambi |
View: | 43 times |
Download: | 1 times |
www.le.ac.uk
Evaluation of a CdTe detector for medical imaging
Layal K. Jambi
Bioimaging UnitSpace Research CentreSupervisor: Dr John Lees and Professor Alan Perkins
Outline
• Introduction
• Main research area
• XRI-UNO CdTe detector
• Performance specification
• Further work
Radiation detectors
• Radiation detectors are the sensing element in nuclear measurement systems.
• Detection of radiation is related to the absorption of radiation and how it interacts with matter.
• There are several types of radiation detectors.
Scintillation detectors
• The main purpose of scintillation in detectors is that the scintillator material converts higher energy incident photons into several lower energy photons.
Figure - taken from the Idaho state university Radiation Information Network.
Scintillation detectors
• There are two types of solid state scintillators:– Inorganic scintillatorse.g. NaI and CsI.NaI(TI) is the most frequently used scintillation crystal.
– Organic scintillatorse.g. Plastics composed of aromatic hydrocarbons.
Semiconductor detectors
• Semiconductors are based on a more direct approach which converts photons directly into an electronic signal.
Figure - taken from NSEP Nuclear Safeguards Education Portal.
Small field of view (SFOV) cameras in medical imaging
• Pinhole collimator has been used with SFOV offering high spatial resolution.
• In medical imaging SFOV used for a small organ dedicated system.
CdTe X-ray and -ray detectors for imaging system
• Hybrid CdTe pixel detector arrays for breast screening to replace mammography.
• CdTe MediProbe for sentinel lymph nodes (SLNs) to replace the gamma probe.
XRI-UNO system
Physical specification
Dimensions (W x L x H)
138mm x 172mm x 34mm
Active area 14.08mm x 14.08mm
Pixel size / # Pixels 55 µm x 55 µm / 65.536 pixels
Semicoducting material
1mm Cadmium Telluride
Initial investigation• A 0.45mm diameter cannula tube, 18mm length
filled with 0.63 MBq 99mTc solution.
A. With high resolution parallel hole
collimator (1mm )
B. With low resolution parallel hole
collimator (2mm )
C. Without collimator
A B C
No. of frames 999 frame, each frames 100ms
Performance specification
1. Intrinsic spatial resolution
2. System spatial resolution
3. Intrinsic spatial uniformity
4. Intrinsic sensitivity
5. Count rate capability
1. Intrinsic spatial resolution
• It is the full width at half maximum (FWHM) of a line spread function (LSF) or of a point spread function (PSF) without a collimator.
1. Intrinsic spatial resolution
0 20 40 60 80 100 120
0
100
200
Cou
nts
Pixels
Edge Response Function (ERF) for Cd-109
1. Intrinsic spatial resolution
Modulus of LSF with fitted Gaussians
0 50 100
0
10
20
Counts Cumulative Fit Peak
Der
ivat
ive
of c
ount
s
Pixels
1. Intrinsic spatial resolution
Intrinsic spatial resolution
XRI-UNO CdTe 0.45 mm
CGC 0.63 mm
XRI-UNO is better in intrinsic spatial resolution
2. System spatial resolution
FWHM (black) and FWTM (red) for Cd-109 by using 0.5mm pinhole collimator
5 10 15 20 25 30 35 40 45 500
2
4
6
8
10
12
14
Cd-109 FWHM FWTM
Spa
tial R
esol
utio
n (m
m)
Perspex Thickness (mm)
2. System spatial resolution
System spatial resolution
XRI-UNO CdTe 1.61 mm
CGC 1.28 mm
XRI-UNO has poorer system spatial resolution
3. Intrinsic spatial uniformity
Intrinsic spatial uniformity
Co-efficient of variation XRI-UNO CdTe 0.38
CGC 1.58
Differential uniformity XRI-UNO CdTe 5.17
CGC 0.6
XRI-UNO is less uniform
4. Intrinsic sensitivity
• The proportion of photon flux incident on the detector that is recorded within the photopeak energy window being used.
• Tested by placing various width of scattering medium (Perspex) between the source and the detector.
4. Intrinsic sensitivity
0 10 20 30 40 50 60 70
0.020
0.025
0.030
0.035
0.040
0.045
Co
un
ts p
er
seco
nd
pe
r in
cid
en
t co
un
ts
Perspex Thickness (cm)
Cd-109 placed at 420mm away from the detector with increasing layers of Perspex
4. Intrinsic sensitivity
Intrinsic sensitivity
XRI-UNO CdTe 28839
CGC 62300
XRI-UNO is less sensitive
5. Count rate capability
176 MBq of 99mTc placed directly on top of the un-collimated detector
0 2000 4000 6000 80000.0
2.0x106
4.0x106
6.0x106
8.0x106
1.0x107
1.2x107
1.4x107
Me
asu
red
co
un
ts
Incident counts
5. Count rate capability
Count rate capability
XRI-UNO CdTe 8134
CGC 1200
XRI-UNO has higher count rate capability
Summary
Performance specification XRI-UNO CdTe CGC
Intrinsic spatial resolution Better
System spatial resolution poorer ✓Intrinsic spatial uniformity Less uniform ✓Intrinsic sensitivity Less sensitive ✓Count rate capability Higher
Conclusion
• The XRI-UNO CdTe detector exceeds the CGC in areas such as intrinsic resolution and count rate capability.
• The XRI-UNO CdTe detector would not be able to replace the CGC due to low sensitivity.
Further work
• No more semiconductors.
• Try to use different type of scintillators such as Gadolinium Oxysulfide (GOS) ceramic scintillator.
Acknowledgement • University of Leicester
Dr. John Lees, Sarah Bugby, Mohammed Alqahtani, Dr. Simon Lindsay, Bahadar Bhatia and William R McKnight
• University of Liverpool
Sean Tipper
• University Hospitals Nottingham
Prof. Alan Perkins and A K Ng
• Leicester Royal Infirmary
Helen Hill and David Monk
Further information
• Bugby, S.L., J.E. Lees, B.S. Bhatia, and A.C. Perkins, Characterisation of a high resolution small field of view portable gamma camera. Phys Med, 2014. 30(3): p. 331-9.
• Bhatia, B.S., S.L. Bugby, J.E. Lees, and A.C. Perkins, A scheme for assessing the performance characteristics of small field-of-view gamma cameras. Phys Med, 2015. 31(1): p. 98-103.
Radioactive Sources used
Source Activity (MBq) Energy (keV) Diameter (mm)
Cadmium-109 345 22 8
Cobalt-57 2 122 6
Technetium-99m 229 140 NA
Comparison TableDetector CGC CdTe
Field of view (mm) 40 x 40 14.08 x 14.08
Intrinsic spatial resolution FWHM (mm) 0.63 0.4555
FWTM (mm) 1.06 0.7985
System spatial resolution FWHM (mm) 1.28 1.61
FWTM (mm) 2.35 8.00
Intrinsic spatial uniformity Integral uniformity (%) 8.5 100
Spread of differential uniformity (%) 0.6 5.174
Co-efficient of variation (%) 1.58 0.38
Intrinsic Sensitivity Counts per second 62300 28839
Count rate capability Maximum counts (incident counts per MBq)
1200 1668
Contrast to noise ratio CNR No data 12
GOS Advantages
• High sensitivity, short decay time, short afterglow.
• Eco-friendly with no hazardous materials contained.
• Resistant to humidity.
• High x-ray shielding capability to protect light receiving element.
• Toshiba also has line-up of green and red light emitting high sensitivity scintillators, although with inferior decay time, afterglow characteristics.
http://www.toshiba-tmat.co.jp/eng/list/sc_cera.htm