+ All Categories
Home > Documents > LightCycler ® 1536 Real-Time PCR System - LightCycler ... · J Place multiwell plate in a standard...

LightCycler ® 1536 Real-Time PCR System - LightCycler ... · J Place multiwell plate in a standard...

Date post: 25-Aug-2019
Category:
Upload: vunhan
View: 214 times
Download: 0 times
Share this document with a friend
12
Endpoint Genotyping This technical note describes how to set up, perform and analyze genotyping assays (allelic discrimination assays) to detect known single nucleotide poly- morphisms (SNPs) using the LightCycler® 1536 Real-Time PCR System and the LightCycler® 1536 Endpoint Genotyping Analysis Tool. Dual color hydrolysis probe assays Each hydrolysis probe contains two labels, a fluorescent reporter and a quencher, in close proximity to each other. In addition to the target specific primers, endpoint genotyping analysis uses two sequence-specific probes that are designed to detect allele X and allele Y, respectively and are labeled with different reporter dyes. Hydrolysis probes specificity: J FAM dye detects samples that are homozygous for allele X and heterozygous XY. J VIC (HEX) dye detects samples that are homozygous for allele Y and heterozygous XY. A fluorescence signal is generated when dyes, bound to allele-specific probes, are cleaved. The intensity of the signal depends on the amount of hybridization between probe and target, which will be greater for a perfect sequence match than for a mismatch. The endpoint fluorescent signal can be measured by the LightCycler® 1536 Instrument. LightCycler® 1536 Software automatically calculates endpoint fluorescence (EPF) values after the PCR run is finished. EPF values can be transferred, via export/ import commands, from the LightCycler® 1536 Software into the LightCycler® Endpoint Genotyping Analysis Tool. This tool can be used to group samples based on their endpoint fluorescence intensity and to classify these groups according to genotype (homozygous X, homozygous Y, heterozygous XY). Results can be displayed as a scatter plot, result table or plate view. Topic Purpose of this Technical Note Assay Principle LightCycler ® 1536 Real-Time PCR System LightCycler® 1536 Endpoint Genotyping Analysis Tool Technical Note Detection Format Assay Results
Transcript
Page 1: LightCycler ® 1536 Real-Time PCR System - LightCycler ... · J Place multiwell plate in a standard swing-bucket centrifuge and centrifuge for 2 minutes at 1,500 x g. J Load the multiwell

Endpoint Genotyping

This technical note describes how to set up, perform and analyze genotyping assays (allelic discrimination assays) to detect known single nucleotide poly-morphisms (SNPs) using the LightCycler® 1536 Real-Time PCR System and the LightCycler® 1536 Endpoint Genotyping Analysis Tool.

Dual color hydrolysis probe assays Each hydrolysis probe contains two labels, a fluorescent reporter and a quencher, in close proximity to each other. In addition to the target specific primers, endpoint genotyping analysis uses two sequence-specific probes that are designed to detect allele X and allele Y, respectively and are labeled with different reporter dyes.

Hydrolysis probes specificity:J FAM dye detects samples that are homozygous for allele X and heterozygous XY. J VIC (HEX) dye detects samples that are homozygous for allele Y and heterozygous XY.

A fluorescence signal is generated when dyes, bound to allele-specific probes, are cleaved. The intensity of the signal depends on the amount of hybridization between probe and target, which will be greater for a perfect sequence match than for a mismatch. The endpoint fluorescent signal can be measured by the LightCycler® 1536 Instrument.

LightCycler® 1536 Software automatically calculates endpoint fluorescence (EPF) values after the PCR run is finished. EPF values can be transferred, via export/ import commands, from the LightCycler® 1536 Software into the LightCycler® Endpoint Genotyping Analysis Tool. This tool can be used to group samples based on their endpoint fluorescence intensity and to classify these groups according to genotype (homozygous X, homozygous Y, heterozygous XY). Results can be displayed as a scatter plot, result table or plate view.

Topic

Purpose of this Technical Note

Assay Principle

LightCycler® 1536 Real-Time PCR SystemLightCycler® 1536 Endpoint Genotyping Analysis Tool

Technical Note

Detection Format

Assay Results

Page 2: LightCycler ® 1536 Real-Time PCR System - LightCycler ... · J Place multiwell plate in a standard swing-bucket centrifuge and centrifuge for 2 minutes at 1,500 x g. J Load the multiwell

2

Required Materials

Instruments / Disposables LightCycler® 1536 Real-Time PCR System1)

LightCycler® 1536 Endpoint Genotyping Analysis Tool2)

LightCycler® 1536 Multiwell Plate1)

RealTime ready DNA Probes Master1,3)

Sealing foil4)

Thermal plate sealer5)

Low-volume dispenser6)

Reagents / Samples J Sample material (1- 5 ng purified genomic DNA in 2 µl total PCR reaction volume)

J Negative template control (NTC)

Detection Assay(s) SNP Genotyping Assay(s) / Allelic Discrimination Assay(s) (two sequence-specific primers, two sequence-specific hydrolysis probes 1 FAM-labeled, VIC (HEX)-labeled

Products Catalog Number Pack Size

LightCycler® 1536 Instrument 05 334 276 001 1 Instrument

LightCycler® 1536 Multiwell Plate 05 358 639 001 10 x 10 plates

LightCycler® 1536 Endpoint Genotyping Analysis Tool2)

RealTime ready DNA Probes Master 05 502 381 0015x conc., 5 x 1 ml (12,500 x 2 µl reactions)

Roche Applied Science - Ordering Information

Products from Roche Applied Science: 1) For detailed information using the LightCycler® 1536 System, please refer to the LightCycler® 1536 Operator’s Guide.

2) For further details on the LightCycler® 1536 Endpoint Genotyping Analysis Tool, please contact your local Roche representative.

3) For detailed information using the RealTime ready DNA Probes Master, please refer to the package insert for this product.

For detailed information regarding the LightCycler® 1536 System, please visit: www.lightcycler1536.com.

Other products which were used for the setup of the LightCycler® 1536 Multiwell Plate in the sample experiment described in this Technical Note:

4) Sealing foil: Clear Weld Seal Mark II, 4titude®, Ltd., UK

5) Semi-automated heat sealing: Agilent PlateLoc Thermal Microplate Sealer, Agilent Technologies, CA, USA

6) Low-volume dispenser used for this example: InnovadyneTM NanodropTM Express (1x 16 head), IDEX Health & Science LLC., WA, USA

Page 3: LightCycler ® 1536 Real-Time PCR System - LightCycler ... · J Place multiwell plate in a standard swing-bucket centrifuge and centrifuge for 2 minutes at 1,500 x g. J Load the multiwell

3

Required Materials

Instruments / Disposables LightCycler® 1536 Real-Time PCR System1)

LightCycler® 1536 Endpoint Genotyping Analysis Tool2)

LightCycler® 1536 Multiwell Plate1)

RealTime ready DNA Probes Master1,3)

Sealing foil4)

Thermal plate sealer5)

Low-volume dispenser6)

Reagents / Samples J Sample material (1- 5 ng purified genomic DNA in 2 µl total PCR reaction volume)

J Negative template control (NTC)

Detection Assay(s) SNP Genotyping Assay(s) / Allelic Discrimination Assay(s) (two sequence-specific primers, two sequence-specific hydrolysis probes 1 FAM-labeled, VIC (HEX)-labeled

Products Catalog Number Pack Size

LightCycler® 1536 Instrument 05 334 276 001 1 Instrument

LightCycler® 1536 Multiwell Plate 05 358 639 001 10 x 10 plates

LightCycler® 1536 Endpoint Genotyping Analysis Tool2)

RealTime ready DNA Probes Master 05 502 381 0015x conc., 5 x 1 ml (12,500 x 2 µl reactions)

Figure 1: Endpoint Genotyping Analysis – Workflow Overview.

The workflow times listed in this example refer to an experiment that analyzes 16 SNPs for 96 different genomic DNA samples. The calculated total

time (~ 155 minutes) is valid only for that particular plate layout and workflow, as performed with the equipment listed above.

I. Manual preparation of Source Plates (96-well format):

Source Plate I: PCR reaction mix (RealTime ready DNA Probes Master, primers, probes) Source Plate II: Genomic DNA samples (with optional addition of Setup Control)

II. Automated preparation of Destination Plate (LightCycler® 1536 Multiwell Plate): Automated dispensing of equal volumes of PCR reaction mix (from Source Plate I, 1 µl) and DNA samples (from Source Plate II, 1 µl) into each destination plate well, using a low-volume liquid dispenser.

III. Processing of the loaded Destination Plate (LightCycler® 1536 Multiwell Plate): - Heat sealing - Centrifugation

IV. Performing the LightCycler® 1536 Real-Time PCR run (45 cycles)

V. Exporting the LightCycler® 1536 result table (*.txt) from the LightCycler® 1536 Software to a specified directory

VI. Importing the *.txt file into the LightCycler® 1536 Endpoint Genotyping Analysis Tool

VII. Performing a manual endpoint genotyping analysis within the LightCycler® 1536 Endpoint Genotyping Analysis Tool

VIII. Optional: Exporting the results of the endpoint genotyping analysis to a dedicated folder in a specified directory for storage

~ 20

~ 60

~ 5

Time(min.)

~ 15

~ 50

~ 5

1536 Endpoint Genotyping

Semi-A

utomated P

CR

Plate-SetupP

CR

Run / A

nalysisD

ownstream

Analysis

Page 4: LightCycler ® 1536 Real-Time PCR System - LightCycler ... · J Place multiwell plate in a standard swing-bucket centrifuge and centrifuge for 2 minutes at 1,500 x g. J Load the multiwell

4

Different source plates must be prepared, depending on the specific automatedlaboratory workflow used, to set up the LightCycler® 1536 Multiwell Plate (Destination Plate).

In the example described below, two source plates (96-well format) are used forthe final setup of the LightCycler® 1536 Multiwell Plate:

1. Semi-Automated Setup of the PCR Plate

Figure 2: Setup Scheme: Sample experiment for endpoint genotyping analysis of 16 SNPs for 96 genomic DNA samples,

including no template controls (NTCs).

Preparation of Source Plates Preparation of Destination Plate

JSource Plates can be filled manually or set up using automation.

JTo simplify the experimental workflow, in most routine ex- periments, you can pre-plate either the genomic DNA samples or SNP Genotyping Assays (whichever of these components has the higher complexity), and dry down that component on the plate.

JAutomated setup of the Destination Plate using a low-volume dispenser.

Destination Plate:

(1536-well format)

2 µl/well

(1 µl)

(1 µl)

(96-well format)

96 samples

Source Plate II:

JGenomic DNA samples

JSetup Control (optional)

Source Plate I:

JRealTime ready DNA Probes Master

JSNP Genotyping Assays (primers, probes)

JWater, PCR Grade

16 SNPs

(96-well format)

Page 5: LightCycler ® 1536 Real-Time PCR System - LightCycler ... · J Place multiwell plate in a standard swing-bucket centrifuge and centrifuge for 2 minutes at 1,500 x g. J Load the multiwell

5

J Pipette the appropriate volumes of the different components into Source Plate I.

Source Plate II: Genomic DNA (optional with Setup Control)

J Prepare 2x concentrated aliquots of the genomic sample DNAs. The final DNA concentration in the aliquots should be 2 to 10 ng / 1 µl. J Pipette the 2x concentrated sample DNAs and an aliquot of the 20x concentrated Setup Control into Source Plate II:

Automated Dispensing of the Destination Plate and Plate Processing:

J Dispense equal volumes (1 µl) of the PCR reaction mix (Source Plate I) and DNA samples (Source Plate II) into wells of the Destination Plate (LightCycler® 1536 Multiwell Plate) using a low-volume liquid dispenser.J Heat seal the multiwell plate with adequate sealing film.J Place multiwell plate in a standard swing-bucket centrifuge and centrifuge for 2 minutes at 1,500 x g.J Load the multiwell plate into the LightCycler® 1536 Instrument.

J Calculate the total volume of a 2x concentrated PCR reaction mixture containing the 5x Master Mix and the specific PCR primers and probes.

J Prepare a 20x conc. SNP Genotyping Assay Mix that contains all specific PCR primers (18 µM each) and both hydrolysis probes (FAM-labeled and VIC (HEX)-labeled, 5 µM each).

ComponentsVolume (µl)for One PCR Reaction (2 µl)

Sample experiment: Volume (µl) for 96 PCR Reactions*

RealTime ready DNA Probes Master,Master Mix, 5x conc.

0.4 µl 38.4 µ + x** µl

SNP Genotyping Assay Mix, 20x conc. 0.1 µl 9.6 µl + y** µl

Water, PCR Grade 0.5 µl 48 µl + z** µl

Final Volume 1.0 µl

Source Plate I: PCR Reaction Mix (without sample DNA)

Components Volume for One PCR Reaction (µl)

Sample DNA, 2x conc. or Water, PCR Grade

0.9 µl

Setup Control, 20x conc. 0.1 µl

Final Volume 1.0 µl

NOTE* Sample experiment: Endpoint genotyping analysis for 96 genomic DNA samples per SNP.

** The total required volume of each reagent must include sufficient extra volume to offset (a) reagent pipetting losses and (b) other losses (e.g., dead volume) that depend on the specific low-volume liquid dispenser used to transfer reaction mix to the Destination Plate.

Page 6: LightCycler ® 1536 Real-Time PCR System - LightCycler ... · J Place multiwell plate in a standard swing-bucket centrifuge and centrifuge for 2 minutes at 1,500 x g. J Load the multiwell

6

2. LightCycler® 1536 Real-Time PCR Run Generating Endpoint Fluorescence (EPF) Values

NOTE* For the subsequent target-specific endpoint genotyping analysis to operate properly, each item in the sample list must contain both a sample name and a target name.

To program a new experiment for endpoint genotyping analysis:

J Create a new experiment.J Set the detection format: Dual Color Hydrolysis Probes / UPL Probes.J Define programs: Preincubation, 2 Step Amplification, Cooling (Keep default parameters for all programs).J Select the pipetting control mode according to your chosen experimental setup: Setup Control or Master Control.

J Import a sample list containing sample names and target names*.J Add notes regarding the experiment (optional). J Start the PCR run on the LightCycler® 1536 Instrument.J Save the experiment. J At the end of a successful run, the results are automatically displayed in two formats (result table, amplification charts).J Export and save the Result Table.txt file from the run to a dedicated folder.

Page 7: LightCycler ® 1536 Real-Time PCR System - LightCycler ... · J Place multiwell plate in a standard swing-bucket centrifuge and centrifuge for 2 minutes at 1,500 x g. J Load the multiwell

7

2. LightCycler® 1536 Real-Time PCR Run Generating Endpoint Fluorescence (EPF) Values

3. LightCycler® 1536 Endpoint Genotyping Analysis Analysis of EPF Values using the LightCycler® 1536 Endpoint Genotyping Analysis Tool

NOTERequirements for using the LightCycler® 1536 Endpoint Genotyping Analysis Tool:J The tool is an Excel template (.xlt file) that requires Microsoft® Excel 2003 or later version. J LightCycler® 1536 Software (version 1.0) expresses floating point numbers in the English (USA) format. Therefore you must use the English version of Microsoft Excel and set the Microsoft Windows Regional and Language Options to English (USA).J Recommended screen resolution for the computer monitor: 1236 x 1024.

To analyze the EPF values generated with the LightCycler® 1536 Instrument / Software:

J Open the LightCycler® 1536 Endpoint Genotyping Analysis Tool (LightCycler® 1536 Endpoint Genotyping Analysis Tool.xlt file) in Microsoft Excel.

J If using Microsoft Excel, version 2003: select Activate macros in the Excel wizard that appears or J If using Microsoft Excel, version 2007: choose Options under “Global Selection”, then select Security Warning - Macro and check Enable this content.

NOTEThe Excel Analysis sheet is protected: Sorting and writing is disabled, filtering is enabled.

Page 8: LightCycler ® 1536 Real-Time PCR System - LightCycler ... · J Place multiwell plate in a standard swing-bucket centrifuge and centrifuge for 2 minutes at 1,500 x g. J Load the multiwell

8

A) Symmetric Auto-Scaling

(both axes have the same scaling).

B) Axis Specific Auto-Scaling

(scale of each axis was set according

to the maximum fluorescence value

seen in each channel).

A B

Figure 3: Description of the Analysis Screen and its adjustable functions.

Angles for FAM and VIC Zone, respectively:Determines areas of the scatter plot within which samples are called Homozygote FAM or Homozygote VIC. The size of the area in which samples are called Homozygous for the allele that binds the FAM or VIC probe can be manually adjusted using the corresponding up/down buttons. Threshold Pos/Neg: Determines the area of the plot within which samples are called Negative. The size of the area in which samples are called Negative can be manually adjusted using the corresponding up/down size tabs. Angle Unknown Zone: Determines the area of the plot within which samples are called Unknown. The size of the area that defines samples of Unknown genotype can be manually optimized (using the corresponding up/down size tabs) to meet specific experimental requirements.

1 2

3

4

Spin buttons Scatter plot Filter function

Sample list

Negative

Detector

Unknown

Detector

Homozygote

Unknown

Homozygote

Drop-down-buttonfor target selection

Heterozygote

1

2

3

4

Display options for the scatter plot (drop-down menu, upper right corner of screen):

Page 9: LightCycler ® 1536 Real-Time PCR System - LightCycler ... · J Place multiwell plate in a standard swing-bucket centrifuge and centrifuge for 2 minutes at 1,500 x g. J Load the multiwell

9

J If multiple genotype targets are analyzed in a single LightCycler® 1536 Instrument run, as in the sample experiment described here, you must manually perform the following analysis steps for each geno- type target (e.g., ADD1, TGF beta):

J Select the first genotype target via the drop-down tab in the Target column.

J Using the relevant spin buttons, adjust the settings for Angle FAM Zone, Angle VIC Zone, Angle of Unknown Zone, and Threshold Pos/Neg.

J Click the “Submit to Result Tab” button to copy the calculated genotypes of the currently selected target into the result table (Excel Result sheet).

J Repeat the above three steps for each genotype target in the experiment.

J To view the genotyping calls for all targets, select the Result tab. Genotypes displayed on the result table (Excel Result sheet) are based on the individual angle and threshold settings that were manually selected for each target.

NOTEThe Excel Result sheet is not protected; both read and write functions are enabled.

Figure 3: Description of the Analysis Screen and its adjustable functions.

Page 10: LightCycler ® 1536 Real-Time PCR System - LightCycler ... · J Place multiwell plate in a standard swing-bucket centrifuge and centrifuge for 2 minutes at 1,500 x g. J Load the multiwell

10

J Click the “Plate View” button to display the genotype calls arranged on a schematic of the LightCycler® 1536 Multiwell Plate. If desired, click the “Print” button, and the plate view will be printed to the default printer.

J Click the “Export Result.txt” button to save the data displayed on the Excel Result sheet as a .txt file.

J In both versions, Microsoft Excel 2003 and Microsoft Excel 2007, save the entire file (i.e., both the Analysis and Result sheets) as an Excel 97-2003 workbook (*.xls).

NOTEA complete overview of all the target-specific results obtained for the analyzed samples is shown only on the Excel Result sheet; the data displayed on the Excel Analysis sheet is only for one genotype target at a time.

NOTEWhen working with Microsoft Excel 2007, do not save the file as an Excel 2007 workbook with macros (*.xlsm) or as an Excel 2007 template with macros (*.xltm). When the file is saved in either of these formats, its macros may be permanently disabled, especially if no antivirus protection is installed on your PC.

Page 11: LightCycler ® 1536 Real-Time PCR System - LightCycler ... · J Place multiwell plate in a standard swing-bucket centrifuge and centrifuge for 2 minutes at 1,500 x g. J Load the multiwell

11

DisclaimerThis Light Cycler® 1536 Real-Time PCR System is a real-time thermal cycler

licensed for research use only under U.S. Patent No. 6,814,934 and correspond-

ing claims in its non-U.S. counterparts, and under one or more of U.S. Patents

Nos. 5,475,610, 5,602,756, 6,703,236, 7,238,517, 7,504,241, 7,537,377 or correspond-

ing claims in their non-U.S. counterparts, owned by Applied Biosystems, LLC.

No rights are conveyed expressly, by implication or estoppels to any other

patent claims. The LightCycler® 1536 Real-Time PCR System is not cleared for

in vitro diagnostic use in the U.S. For further information on purchasing addition-

al rights, contact the Director of Licensing at Applied Biosystems, 850 Lincoln

Centre Drive, Foster City, California, 94404, USA.

The LightCycler® 1536 Multiwell Plate is manufactured under license from IT-IS

International Limited and under Patent number US 60/970401.

No right is conveyed expressly, by implication, or by estoppels under any other

patent claim, such as claims to apparatus, reagents, kits, or methods such as 5'

nuclease methods. License rights to practice PCR methods under the foregoing

Roche patents for research and other non-in vitro diagnostic applications may

be purchased from Applied Biosystems or may be obtained by purchasing

licensed reagents from Roche, Applied Biosystems, or other Authorized third

party. For information on purchasing licenses for research and other non-

in vitro diagnostic applications, contact the Director of Licensing at Applied

Biosystems, 850 Lincoln Centre Drive, Foster City, California 94404, USA.

A license to practice PCR methods with real-time detection under patents of

F. Hoffmann-La Roche Ltd, and Roche Molecular Systems ("Roche") for in vitro

diagnostic applications may be purchased from Roche or may be obtained by

purchasing in vitro diagnostic reagents from Roche or any other Authorized

third party. The Light Cycler® 1536 Real-Time PCR System is not cleared for in

vitro diagnostic use in the U.S.

A license to perform the patented 5' Nuclease Process for research is obtained

by the purchase of (i) both Authorized 5' Nuclease Core Kit and Licensed Probe,

(ii) a Licensed 5' Nuclease Kit, or (iii) license rights from Applied Biosystems.

This product is an Authorized 5' Nuclease Core Kit. Use of this product is

covered by one or more of the following US patents and corresponding patent

claims outside the US: 5,079,352, 5,789,224, 5,618,711, 6,127,155, 5,677,152,

5,773,258, 5,407,800, 5,322,770, 5,310,652, 5,210,015, 5,487,972, and claims out-

side the US corresponding to US Patent No. 4,889,818. The purchase of this pro-

duct includes a limited, non-transferable immunity from suit under the foregoing

patent claims for using only this amount of product for the purchaser's own

internal research. Separate purchase of a Licensed Probe would convey rights

under the applicable claims of US Patents Nos. 5,538,848, 5,723,591, 5,876,930,

6,030,787, 6,258,569, 5,804,375 (claims 1-12 only), and 6,214,979, and correspond-

ing claims outside the United States. No right under any other patent claim and

no right to perform commercial services of any kind, including without limitation

reporting the results of purchaser's activities for a fee or other commercial con-

sideration, is conveyed expressly, by implication, or by estoppel. This product is

for research use only. Diagnostic uses under Roche patents require a separate

license from Roche. Further information on purchasing licenses may be obtained

from the Director of Licensing, Applied Biosystems, 850 Lincoln Centre Drive,

Foster City, California 94404, USA. Under license from Gilead Sciences Inc.

Trademarks / Regulatory Disclaimer

For life science research only. Not for use in diagnostic procedures.

LIGHTCYCLER and REALTIME READY are trademarks of Roche.

Other brands or product names are trademarks of their respective holders.

Limitation of Liability THE PRODUCT IS NOT A FULLY TESTED PRODUCT. CUSTOMER'S USE OF THE

PRODUCT IS AT CUSTOMER'S SOLE RISK. THE PRODUCT MAY CONTAIN

DEFECTS, FAIL TO COMPLY WITH APPLICABLE SPECIFICATIONS AND/OR

PRODUCE UNINTENDED OR ERRONEOUS RESULTS. ROCHE DISCLAIMS

AND EXCLUDES ALL REPRESENTATIONS AND WARRANTIES OF ANY KIND

OR CHARACTER, EXPRESS OR IMPLIED, WITH RESPECT TO THE PRODUCT,

AND CUSTOMER ACCEPTS THE PRODUCT "AS IS," WITH ALL

FAULTS. WITHOUT LIMITING THE FOREGOING, ROCHE DISCLAIMS AND

EXCLUDES ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS

FOR PARTICULAR PURPOSE, ANY WARRANTIES ARISING FROM COURSE OF

DEALING OR USAGE OF TRADE, AND ANY WARRANTIES THAT THE PRO-

DUCT IS FREE FROM THE RIGHTFUL CLAIM OF ANY THIRD PARTY BY WAY

OF INFRINGEMENT OR THE LIKE.

ROCHE SHALL NOT BE SUBJECT TO, AND CUSTOMER RELEASES ROCHE

AND ITS AFFILIATES FROM, ANY LIABILITY OR OBLIGATION OF ANY KIND

ARISING FROM OR IN CONNECTION WITH THE PRODUCT, ANY ACTS OR

OMISSIONS RELATING TO THE PRODUCT OR ANY USE OR PERFORMANCE

OF THE PRODUCT.

Page 12: LightCycler ® 1536 Real-Time PCR System - LightCycler ... · J Place multiwell plate in a standard swing-bucket centrifuge and centrifuge for 2 minutes at 1,500 x g. J Load the multiwell

Published by:

Roche Diagnostics GmbH Roche Applied ScienceWerk Penzberg 82372 PenzbergGermany © 2009 Roche Diagnostics GmbH All rights reserved.1s

t Ver

sion

(O

ctob

er 2

009)


Recommended