Date post: | 18-Jul-2015 |
Category: |
Education |
Upload: | educacionanestesia |
View: | 975 times |
Download: | 1 times |
WHAT IS IT???
• One corrugate tube
• 22mm diameter
• Septum/wall extruded down its middle
• Inspiratory and expiratory channels are completely divided
• Made of three different resins
• Area of each distinct channel = 15.5mm
PATENTS
The Limbo has two patents:– Process = 6,431,218D1. Method of
manufacture.
– Design = 6,394,442B1. Single limb circle breathing system.
CHARACTERISTICS OF LIMBO
• Less compliant - the septum along with the stiffness of the tubing decreases compliance thereby increasing total volume delivered to the patient. “What you set is what you get.”
• Less heat loss - the septum permits the transfer of heat from the patient’s warm exhaled breath. Tests reveal an increase of 6.2°C @ 4L/min fresh gas flow rates.
CHARACTERISTICS OF LIMBO
• Less weight - less plastic results in less weight and less torque on the patient’s artificial airway device. Limbo weighs 34% less than coaxial circuits and 25% less than traditional circle circuits.
• Less inventory - One circuit for pediatrics and adults! Use Limbo with pediatric patients weighing 4.5kg or greater.
LIMBO WEIGHT PARAMETERS
• Minimum tidal volume requirements to avoid rebreathing CO2:– 3 x amount of mechanical dead space
• Dead space of the Limbo is ~15cc– 3 x 15cc = 45cc tidal volume minimum
• 10cc per kg of body weight is the usual estimate of VT– 45cc minimum VT/10cc body weight = 4.5kg
or 9.9lbs
4300MDE
• Adapter for administering an MDI
• Disposable• For use in respiratory
care or anesthesia• 4300MDI also
available - fits onto inspiratory limb
Multi-Access™ Adapter
• Used for measuring proximal airway pressure
• For Bear 1000 or any other vent requiring proximal pressure port
• 22mm I.D./ 22mm O.D.
• Product code: 174024
RESPIRATORY CARE
• For use with but not limited to Servo 900c (pictured), 300
• Temperature port adapter P/N 174020 if using heater
• Possible future design with heated wire
RESPIRATORY CARE
• For use with PB7200 (pictured), PB840
• Not limited to these vents
• May be used with any volume ventilator that uses a circle breathing circuit
QUALITY ASSURANCE• Every two hours, 10 hose segments are
verified in order to validate the extrusion process
• Every circuit is 100 % leak tested with a flow detector – A circle is created with the tubing and is
pressurized according to ISO standards– If a leak is detected the circuit is considered a
failure and is discarded– Actions are taken to modify the materials
SALES STRATEGIES
• Inventory consolidation = one circuit for peds and adults
• Three applications = OR, TXP, RT
• Mapleson D adapter
• No risk of increased dead space (King F2)
• Thermally efficient
• Flexible configurations
• Unique!