Date post: | 18-Jan-2018 |
Category: |
Documents |
Upload: | anthony-bryant |
View: | 217 times |
Download: | 0 times |
Location of the LW detector- Simulation of the LW signals
Lawrence Deacon RHUL ATF2 meeting August 23rd 2006 KEK
Possible Locations
● Between BH5 and QD6● Between QD6 and QF5● Further downstream
BDSIM screenshots● Green line is a photon. It passes
through QD6 and QF5 and hits SF5.
● BH5 must have a window
● Holes would need to be drilled in yoke of SF5 for detection further downstream. This would affect the magnetic fields
BH5
QD6
QF5
SF5
Possible Detector Locations
< QD6● Space is 0.55m
● Could fit aerogel Cerenkhov (detector we are currently using at ATF) if extra mirrors are used to reflect the cerenkhov light hoizontally before it is reflected down into the PMT
● Very difficult to fit current calorimeter in.
● Could limit detector choices.
● Laserwire signal between 2cm and 5cm from beamline
QD6 > QF5
● Laserwire signal would pass through gap between pole tips
● Bolt heads need to be shortened
QD6 > QF5
● Space is 1.6m – space required for BPM
● Can put larger detectors at this location
● Laserwire signal approx. 15cm from beam line
Simulations
● ATF2 laser wire simulated using BDSIM
Quadrupole Model
● Modelled in BDSIM as shown
● Red is iron● Green is copper (part
of coils)● Dimensions match
technical drawing
Simulation materials/ apertures● Beam pipe radius= 2.14cm (except in apertures)
● Beam pipe material is aluminium throughout (should be steel in drifts and copper in magnets?)
● Magnet material is Iron (small part of quad is copper)● BH5 is vacuum (window to allow photons through)
● Apertures between pole tips and coils in quarupoles
Simulation Apertures
Simulation Beam Parameters at Start of EXT
● Energy cutoff = 1MeV● Energy= 1.28GeV● SigmaE= 0.8 e-3● Horizontal emittance= 2.0 e-9● Normalized vertical emittance= 3.0e-9
Simulation Beam Parameters at Start of EXT
● Beta x = 7.328756752062● Beta y = 3.138666109153 ● Alpha x = 1.182561568783● Alpha y = -1.770354080712● From mad files (Mark Woodley, SLAC)
Laser Wire Parameters
● Position: extraction line at MW0X (between QD18X and QD19X)
● Wavelength= 532nm● Spotsize= 15 microns
Simulated e- Beam at Start of EXT/Laserwire
● Top: y ● Bottom: x● Left: start ● Right:at laserwire● Units: microns
Simulated Compton Scattered Photon Spectrum
● Photon spectum at laserwire IP
● Units: GeV● High energy cutoff
agrees with theory● Artificial cutoff below
1MeV
Laserwire Photons at IP
● Left: xp/yp● Right: x/y
Comparison of Detector Locations
● Particle distribution x y
● Left: <QD6● Right: QD6>QF5
Comparison of Detector Locations
● Left: primary particles Right:not primary particles
● Top: <QD6 Bottom: QD6>QF5
Comparison of Detector Locations● Secondary particles were scattered over large angles.
● Secondary particle number and energy could fluctuate more than that of primary particles
● Compared energy for different sampler (detector) sizes at both locations and
● Standard deviation in the energy/energy for different sampler sizes at both locations
● Statistics are from 10 shots with 1000 scattered photons each (will be higher will shorter wavelength laser).
● Assumed that detector is uniformly sensitive to edge. Further simulation required to include effects of mirrors etc.
Comparison Of Detector Locations● Energy per shot/ GeV● One shot scatters 1000
photons from e- beam. Increases with shorter laser wavelength. May go to shorter wavelength in future.
● Top: at < QD6● Bottom: at QD6<QF5
Comparison of Detector Locations● standard devation in
energy/energy
● Around 2.5% fluctuations and 9.5 GeV at <QD6 and 3.5% fluctuations and 8GeV at QD6<QFF
● Detector area should be 5m by 5cm to maximise signal?
Best Location● The difference in signal between the two locations is not
great
● More space in the second location provides more flexibility
● Laser wire signal is further from beam line at second location so more space is available
● The background at one location could be higher than at the other
● Will bolt heads may need to be shortened in QD6 for second location?
Future Work
● Simulate background● Compare simulations with ATF results● To do this the current detector must be calibrated● Test new detector setup● Try calorimeter● Try simulations with different e- beam optics● Simulate detector: lead, aerogel, mirrors.