M-1® Cot Fastener Maintenance Manual 6373 6376 6381 EN 6373-009-002 Rev C.0 2018/08
Transcript

MM--11®® CCoott FFaasstteenneerr

MMaaiinntteennaannccee MMaannuuaall

637363766381

EN

6373-009-002 Rev C.0 2018/08

SSyymmbboollss

Operating instructions/Consult instructions for use

General warning

Caution

Catalogue number

Lot (batch) code

Serial number

For US patents see www.stryker.com/patents

Manufacturer

Date of manufacture

Safe working load

6373-009-002 Rev C.0 EN

TTaabbllee ooff CCoonntteennttss

Warning/Caution/Note Definition ..............................................................................................................................2Summary of safety precautions ................................................................................................................................3Introduction ...............................................................................................................................................................4

Product description .................................................................................................................................................4Indications for use...................................................................................................................................................4Expected service life ...............................................................................................................................................4Contraindications ....................................................................................................................................................4Specifications .........................................................................................................................................................4Product illustration ..................................................................................................................................................5Contact information.................................................................................................................................................5Serial number location.............................................................................................................................................6Date of manufacture................................................................................................................................................6Cot compatibility .....................................................................................................................................................6

Cleaning....................................................................................................................................................................7Suggested cleaners ................................................................................................................................................7

Quick reference replacement parts ..........................................................................................................................8Center mount cot fastener, DIN guides ....................................................................................................................9Center mount cot fastener, straight guide without belts .........................................................................................10Center mount cot fastener, narrow guide ...............................................................................................................11Foot end latch assembly.........................................................................................................................................12Head end hook assembly, right ..............................................................................................................................13Head end hook assembly, left ................................................................................................................................14Pull pin assembly ....................................................................................................................................................15

6373-009-002 Rev C.0 1 EN

WWaarrnniinngg//CCaauuttiioonn//NNoottee DDeeffiinniittiioonnThe words WWAARRNNIINNGG, CCAAUUTTIIOONN, and NNOOTTEE carry special meanings and should be carefully reviewed.

WWAARRNNIINNGG -- Alerts the reader about a situation which, if not avoided, could result in death or serious injury. It may alsodescribe potential serious adverse reactions and safety hazards.

CCAAUUTTIIOONN -- Alerts the reader of a potentially hazardous situation which, if not avoided, may result in minor or moderateinjury to the user or patient or damage to the product or other property. This includes special care necessary for the safeand effective use of the device and the care necessary to avoid damage to a device that may occur as a result of use ormisuse.

NNoottee -- Provides special information to make maintenance easier or important instructions clearer.

EN 2 6373-009-002 Rev C.0

SSuummmmaarryy ooff ssaaffeettyy pprreeccaauuttiioonnssAlways read and strictly follow the warnings and cautions listed on this page. Service only by qualified personnel.

WWAARRNNIINNGG• Install the cot fastener by qualified personnel only. Improper installation could result in injury to the patient or operator.• Always make sure that all cots meet the installation specifications for the Stryker cot fastener system.• Always replace the cot fastener if it has been involved in an accident to avoid the risk of injury due to product damage.• Always use any appropriate personal protective equipment while power washing to avoid inhaling contagion. Power

washing equipment may aerate contamination.• Always wipe the product with clean water and dry after cleaning. Some cleaning products are corrosive in nature and

may cause damage to the product. Failure to properly rinse and dry the product leaves a corrosive residue on thesurface of the product and may cause premature corrosion of critical components.

CCAAUUTTIIOONN• Improper usage of the product can cause injury to the patient or operator. Operate the product only as described in this

manual.• Do not modify the product or any components of the product. Modifying the product can cause unpredictable operation

resulting in injury to patient or operator. Modifying the product also voids its warranty.

6373-009-002 Rev C.0 3 EN

IInnttrroodduuccttiioonnThis manual assists you with the operation or maintenance of your Stryker product. Read this manual before operating ormaintaining this product. Set methods and procedures to educate and train your staff on the safe operation or maintenanceof this product.

CCAAUUTTIIOONN• Improper usage of the product can cause injury to the patient or operator. Operate the product only as described in this

manual.• Do not modify the product or any components of the product. Modifying the product can cause unpredictable operation

resulting in injury to patient or operator. Modifying the product also voids its warranty.

NNoottee• This manual is a permanent part of the product and should remain with the product even if the product is sold.• Stryker continually seeks advancements in product design and quality. This manual contains the most current product

information available at the time of printing. There may be minor discrepancies between your product and this manual. Ifyou have any questions, contact Stryker Customer Service or Technical Support at 1-800-327-0770.

PPrroodduucctt ddeessccrriippttiioonn

The Stryker Model 6373 ambulance cot fastener holds a compatible ambulance cot in an emergency vehicle for patienttransport purposes. The 6373 ambulance cot fastener performs this function after the user manually loads the compatiblecot into the emergency vehicle and guides it into the fastener.

IInnddiiccaattiioonnss ffoorr uussee

The Model 6373 manual cot fastener secures compatible ambulance cots during transport. The fastener is intended for usein pre-hospital environments, in emergency and non-emergency applications. Operate the product by trained professionalsonly. This includes emergency medical service and medical first responders.

EExxppeecctteedd sseerrvviiccee lliiffee

Cot fasteners have a seven year expected service life under normal use conditions and with appropriate periodicmaintenance.

CCoonnttrraaiinnddiiccaattiioonnss

None known.

SSppeecciiffiiccaattiioonnss

Models 6373 - Tray Mounted Cot Fastener (6373-000-000)

6376 - Floor Mounted Cot Fastener (6376-000-000)

6381 - Cot Fastener (6381-000-000)

Compatability See Cot compatibility (page 6)

The Model 6100 M-1 cot, in combination with a Model 6373, 6376, and 6381 cot fastener system, is designed inaccordance to BS EN 1789, Medical Vehicles and their Equipment - Road Ambulances. The cot fastener system must beinstalled by qualified personnel to be in conformance to this standard. 1/4 in. steel backing plates (or equivalent) are

EN 4 6373-009-002 Rev C.0

required beneath the mounting surface (ambulance floor or tray system) at all mounting points. Before putting the systeminto service, test the installation using BS EN 1789 as a minimum.

The Model 6100 M-1 cot, equipped with the AS/NZS 4535 upgrade, in combination with a Model 6373, 6376, or 6381 cotfastener system, is designed in accordance to AS/NZS 4535 Australian/New Zealand Standard - Ambulance RestraintSystems. The cot fastener system must be installed by qualified personnel to be in conformance to this standard. 1/4 in.steel backing plates (or equivalent) are required beneath the mounting surface (ambulance floor or tray system) at allmounting points. Before putting the system into service, test the installation using AS/NZS 4535 as a minimum.

Side restraint belts are required to conform to BS EN 1789 Medical Vehicles and their Equipment - Road Ambulances. Siderestraint belts are not required for applications where this standard does not apply.

PPrroodduucctt iilllluussttrraattiioonn

AB C

E

F G

B

D

A

FFiigguurree 11 –– RRuuggggeedd® CCoott FFaasstteenneerr

A Guide railB Front hookC BumperD Head endE Foot endF Red release buttonG Latch housing

CCoonnttaacctt iinnffoorrmmaattiioonn

Contact Stryker Customer Service or Technical Support at: 1-800-327-0770.

Stryker Medical3800 E. Centre AvenuePortage, MI 49002USA

To view your operations or maintenance manual online, see https://techweb.stryker.com/.

Have the serial number (A) of your Stryker product available when calling Stryker Customer Service or Technical Support.Include the serial number in all written communication.

6373-009-002 Rev C.0 5 EN

SSeerriiaall nnuummbbeerr llooccaattiioonn

A

FFiigguurree 22 –– SSeerriiaall nnuummbbeerr llooccaattiioonn

DDaattee ooff mmaannuuffaaccttuurree

The year of manufacture is the first 2 digits of the serial number.

CCoott ccoommppaattiibbiilliittyy

The Stryker Model 6373, 6376, and 6381 cot fastener systems are compatible only with cots that conform to the installationspecifications.

WWAARRNNIINNGG• Install the cot fastener by qualified personnel only. Improper installation could result in injury to the patient or operator.• Always make sure that all cots meet the installation specifications for the Stryker cot fastener system.• Always replace the cot fastener if it has been involved in an accident to avoid the risk of injury due to product damage.

Stryker

• Model 6100 MM--11 Roll-in system

Ferno-Washington

• Model X-2*

NNoottee -- * 2000 model year or earlier. Stryker is not responsible for changes in specifications to other manufacturer’s cots.

EN 6 6373-009-002 Rev C.0

CClleeaanniinnggWWAARRNNIINNGG -- Always use any appropriate personal protective equipment while power washing to avoid inhaling contagion.Power washing equipment may aerate contamination.

CCAAUUTTIIOONN• Do not steam clean or ultrasonically clean the product.• Do not exceed 180°F (82°C) as the maximum water temperature.• Do not exceed 240°F (115°C) as the maximum air dry temperature (cart washers).• Do not exceed 1500 psi (130.5 bar) as the maximum water pressure. If you are using a hand held wand to wash the

product, keep the pressure nozzle at a minimum of 24 in. (61 cm) from the product.

The product is power washable. The product may show some signs of oxidation or discoloration from continuous washing.No degradation of the product’s performance will occur from power washing if you follow proper procedures.

• Follow the cleaning solution manufacturer’s dilution recommendations exactly.

• Power wash with recommended cleaners.

• Power wash the rails and interface plates with a hand held wand unit or wipe the product with a clean cloth andrecommended cleaners.

SSuuggggeesstteedd cclleeaanneerrss

In general, when used in concentrations recommended by the manufacturer, either phenolic type or quaternary (excludingVViirreexx® TB) type disinfectants can be used. Iodophor type disinfectants are not recommended for use because stainingmay occur.

WWAARRNNIINNGG -- Always wipe the product with clean water and dry after cleaning. Some cleaning products are corrosive innature and may cause damage to the product. Failure to properly rinse and dry the product leaves a corrosive residue onthe surface of the product and may cause premature corrosion of critical components.

Suggested cleaners include:

• Quaternary cleaners (active ingredient - ammonium chloride)

• Phenolic cleaners (active ingredient - o-phenylphenol)

• Chlorinated bleach solution (5.25% - less than 1 part bleach to 100 parts water)

Avoid over saturation. Do not allow the product to stay wet longer than the chemical manufacturer's guidelines for properdisinfecting.

NNoottee -- Failure to follow the above directions when using these types of cleaners may void this product’s warranty.

6373-009-002 Rev C.0 7 EN

QQuuiicckk rreeffeerreennccee rreeppllaacceemmeenntt ppaarrttssThese parts are currently available for purchase. Call Stryker Customer Service: 1-800-327-0770 for availability and pricing.

NNaammee NNuummbbeerr

Rail assembly 6360-020-000

Rail assembly, short 6362-020-000

Rail assembly, long ** 6362-021-000

NNoottee -- TThhee SSttrryykkeerr MMooddeell 66337711 ccoott ffaasstteenneerr wwiitthh tthhee lloonngg rraaiill aasssseemmbbllyy ooppttiioonn ((66336622--002211--000000)) iiss oonnllyy ccoommppaattiibblleewwiitthh tthhee eexxtteennddeedd hheeaadd sseeccttiioonn aasssseemmbbllyy ((99999966--339900--220000))..

EN 8 6373-009-002 Rev C.0

CCeenntteerr mmoouunntt ccoott ffaasstteenneerr,, DDIINN gguuiiddeess6373-010-010 Rev K (Reference only)

IItteemm NNuummbbeerr NNaammee QQuuaannttiittyyA 0007-086-000 Truss head torx screw 4

C 0037-235-000 Square plug 2D 6372-010-011 Foot end latch assembly (page 12) 1E 6372-010-014 Head end hook assembly, right (page

13)1

F 6372-010-015 Head end hook assembly, left (page 14) 1G 6060-090-002 Serial number tag 1H 0025-079-000 Blind rivet 4

K 6372-090-002 Housing label 1L 6373-090-001 Specification label 1M 6373-010-051 DIN rail weldment 1

T 6372-010-020 M-1 fastener spacer 1

6373-009-002 Rev C.0 9 EN

CCeenntteerr mmoouunntt ccoott ffaasstteenneerr,, ssttrraaiigghhtt gguuiiddee wwiitthhoouutt bbeellttss6376-010-010 Rev F (Reference only)

IItteemm NNuummbbeerr NNaammee QQuuaannttiittyyA 0007-086-000 Truss head torx screw 4B 0037-235-000 Square plug 2C 6372-010-011 Foot end latch assembly (page 12) 1D 6372-010-014 Head end hook assembly, right (page

13)1

E 6372-010-015 Head end hook assembly, left (page 14) 1F 6372-010-051 Straight rail weldment 1G 6376-090-001 Specification label 1H 6372-090-002 Housing label 1

M 6060-090-002 Serial number tag 1N 0025-079-000 Blind rivet 6P 6372-010-020 M-1 fastener spacer 2

EN 10 6373-009-002 Rev C.0

CCeenntteerr mmoouunntt ccoott ffaasstteenneerr,, nnaarrrrooww gguuiiddee6381-010-010 Rev G (Reference only)

IItteemm NNuummbbeerr NNaammee QQuuaannttiittyyA 0007-086-000 Truss head torx screw 4B 0037-235-000 Square plug 2C 6372-010-011 Foot end latch assembly (page 12) 1D 6372-010-014 Head end hook assembly, right (page

13)1

E 6372-010-015 Head end hook assembly, left (page 14) 1F 6381-010-051 DIN rail weldment 1G 6381-090-001 Specification label 1H 6372-090-002 Housing label 1

M 6060-090-002 Serial number tag 1N 0025-079-000 Blind rivet 6P 6372-010-020 M-1 fastener spacer 2

6373-009-002 Rev C.0 11 EN

FFoooott eenndd llaattcchh aasssseemmbbllyy6372-010-011 Rev B (Reference only)

IItteemm NNuummbbeerr NNaammee QQuuaannttiittyyA 0001-030-000 Flat head cap screw 4B 0001-140-000 Flat head cap screw 2C 0011-065-000 Washer 1

E 0028-303-000 Truarc ring 1

G 6372-010-012 Rotary latch assembly, right 1H 6372-010-050 Bottom plate weldment 1J 6372-110-066 Casting foot latch housing 1K 6372-010-077 Release button 1M 6372-010-080 Clevis pin 1N 6372-010-081 Wear block 2

EN 12 6373-009-002 Rev C.0

HHeeaadd eenndd hhooookk aasssseemmbbllyy,, rriigghhtt6372-010-014 Rev C (Reference only)

IItteemm NNuummbbeerr NNaammee QQuuaannttiittyyA 0004-460-000 Button head cap screw 1B 0014-020-000 Washer 1C 0014-021-000 Nylon washer 2D 0028-181-000 Truarc ring 1E 0056-026-000 Bumper 1F 6372-010-016 Pull pin assembly (page 15) 1G 6372-010-087 Flip-up hook 1H 6372-010-091 Clevis pin 1J 6372-210-065 Casting head retaining hook 1

6373-009-002 Rev C.0 13 EN

HHeeaadd eenndd hhooookk aasssseemmbbllyy,, lleefftt6372-010-015 Rev C (Reference only)

IItteemm NNuummbbeerr NNaammee QQuuaannttiittyyA 0004-460-000 Button head cap screw 1B 0014-020-000 Washer 1C 0014-021-000 Nylon washer 2D 0028-181-000 Truarc ring 1E 0056-026-000 Bumper 1F 6372-010-016 Pull pin assembly (page 15) 1G 6372-010-087 Flip-up hook 1H 6372-010-091 Clevis pin 1J 6372-210-066 Casting head retaining hook 1

EN 14 6373-009-002 Rev C.0

PPuullll ppiinn aasssseemmbbllyy6372-010-016 Rev A (Reference only)

IItteemm NNuummbbeerr NNaammee QQuuaannttiittyyA 0038-469-000 Compression spring 1B 6372-010-088 Pull pin housing 1C 6372-010-089 Ball knob 1D 6372-010-090 Pin 1

6373-009-002 Rev C.0 15 EN

Stryker Medical3800 E. Centre AvenuePortage, MI 49002USA

6373-009-002 Rev C.0WCR: AA.8

2018/08


Recommended