www.scint-x.com Structured scintillators for digi tal X-Ray imaging Anna Sahlholm
Transcript
Page 1: M4D Det Structured Scintilla Tors for Digital x Ray Imaging

8/8/2019 M4D Det Structured Scintilla Tors for Digital x Ray Imaging

http://slidepdf.com/reader/full/m4d-det-structured-scintilla-tors-for-digital-x-ray-imaging 1/21

www.scint-x.com

Structured scintillators for digital X-Ray imaging

Anna Sahlholm

Page 2: M4D Det Structured Scintilla Tors for Digital x Ray Imaging

8/8/2019 M4D Det Structured Scintilla Tors for Digital x Ray Imaging

http://slidepdf.com/reader/full/m4d-det-structured-scintilla-tors-for-digital-x-ray-imaging 2/21

Per Wiklund, CEO. M.Sc., DIHM.R&D in VLSI process and device technology. International M&S at Ericsson.

Anna Sahlholm, Process Engineer. M.Sc.

Research experience in materials chemistry.

Olof Svenonius, CTO. M.Sc.Specialist in Optronics and sensors.

Carina Näsgårde, process technicianLong experience from semiconductor process development

Jan Linnros. Professor, KTH Stockholm20 years research in solid state physics. Patent holder. Headed research activities

Advisory Board:

Professors Ulf Welander, Sture Petersson, Christer Fröjd

Scint-X Team

Page 3: M4D Det Structured Scintilla Tors for Digital x Ray Imaging

8/8/2019 M4D Det Structured Scintilla Tors for Digital x Ray Imaging

http://slidepdf.com/reader/full/m4d-det-structured-scintilla-tors-for-digital-x-ray-imaging 3/21

Page 4: M4D Det Structured Scintilla Tors for Digital x Ray Imaging

8/8/2019 M4D Det Structured Scintilla Tors for Digital x Ray Imaging

http://slidepdf.com/reader/full/m4d-det-structured-scintilla-tors-for-digital-x-ray-imaging 4/21

Page 5: M4D Det Structured Scintilla Tors for Digital x Ray Imaging

8/8/2019 M4D Det Structured Scintilla Tors for Digital x Ray Imaging

http://slidepdf.com/reader/full/m4d-det-structured-scintilla-tors-for-digital-x-ray-imaging 5/21

Digital vs. Analog x-ray imaging:

+No chemicals involved. Image processing. Easy distribution

+Lower dose

−But lower resolution and contrast

Scint-X solves this problem !

Scintillator technology for imaging x-ray detectors

”state of the art” scintillator  Scint-X scintillator 

Page 6: M4D Det Structured Scintilla Tors for Digital x Ray Imaging

8/8/2019 M4D Det Structured Scintilla Tors for Digital x Ray Imaging

http://slidepdf.com/reader/full/m4d-det-structured-scintilla-tors-for-digital-x-ray-imaging 6/21

Scint-X technology

Conventional detector:

Trade-off between resolution andsensitivity

Scintillator (CsI:Tl) 

Pixellated sensor (CCD,CMOS) 

Scint-X patented solution –

Structured scintillator:

High resolution and high sensitivity

Etched channels (Si wafer) 

Page 7: M4D Det Structured Scintilla Tors for Digital x Ray Imaging

8/8/2019 M4D Det Structured Scintilla Tors for Digital x Ray Imaging

http://slidepdf.com/reader/full/m4d-det-structured-scintilla-tors-for-digital-x-ray-imaging 7/21

66CsI(Tl  ) (n=1,79) 

SiO 2 

(n=1,46) 

X-Ray photons(~0-70 keV) 

CCD pixel 

Silicon

Scintillating waveguide

CCD pixel 

θ c =35 ° 

Light (  λ=550 nm) 

Incident X-Rays generate light at 550 nm

wavelength 55 photons/X-ray photon/keV

Critical angle 35° (NA=0,57) 

Rays at larger angles absorbed in Si

Advantages:High absorption signal +Light confinement resolution +

Challenge:Waveguide loss signal -

Page 8: M4D Det Structured Scintilla Tors for Digital x Ray Imaging

8/8/2019 M4D Det Structured Scintilla Tors for Digital x Ray Imaging

http://slidepdf.com/reader/full/m4d-det-structured-scintilla-tors-for-digital-x-ray-imaging 8/21

Modulation Transfer Function (MTF)Ability to reproduce contrast vs. Spatial FrequencySlant edge analysis according to ISO12233- Low frequencies: coarse objects/large fluctuations- High frequencies: fine objects/pixel pitch

Characteristics

Ideal: MTF = sin(x) / x Slant edge applied at small angle

Page 9: M4D Det Structured Scintilla Tors for Digital x Ray Imaging

8/8/2019 M4D Det Structured Scintilla Tors for Digital x Ray Imaging

http://slidepdf.com/reader/full/m4d-det-structured-scintilla-tors-for-digital-x-ray-imaging 9/21

MTF Comparison:

Characteristics

0 5 10 15 20 250

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1

     M     T     F

Frequency [lp/mm]

30 µm

20 µm

40 µm

CsI needle

Page 10: M4D Det Structured Scintilla Tors for Digital x Ray Imaging

8/8/2019 M4D Det Structured Scintilla Tors for Digital x Ray Imaging

http://slidepdf.com/reader/full/m4d-det-structured-scintilla-tors-for-digital-x-ray-imaging 10/21

Medical X-ray

Intra-oral sensorsThinner detector Improved image resolution and contrastModular scintillator – lower cost

Industrial RadiographyAXI, X-ray microscopy of materials

High resolutionModular scintillator – lower cost

Market for imaging x-ray detectors

Page 11: M4D Det Structured Scintilla Tors for Digital x Ray Imaging

8/8/2019 M4D Det Structured Scintilla Tors for Digital x Ray Imaging

http://slidepdf.com/reader/full/m4d-det-structured-scintilla-tors-for-digital-x-ray-imaging 11/21

Process

Etched wafers

Deposition of CsI(Tl) 

Melting and filling of CsI(Tl), annealing

Polishing

Polymer spinning

Dicing

Page 12: M4D Det Structured Scintilla Tors for Digital x Ray Imaging

8/8/2019 M4D Det Structured Scintilla Tors for Digital x Ray Imaging

http://slidepdf.com/reader/full/m4d-det-structured-scintilla-tors-for-digital-x-ray-imaging 12/21

Page 13: M4D Det Structured Scintilla Tors for Digital x Ray Imaging

8/8/2019 M4D Det Structured Scintilla Tors for Digital x Ray Imaging

http://slidepdf.com/reader/full/m4d-det-structured-scintilla-tors-for-digital-x-ray-imaging 13/21

X-Ray source:Planmeca Intra, commercial dental sourceE = 60..70 kVpt = 20..800 ms

Sensor:Hamamatsu S8981, commercial oral sensor Scintillator removed – bare FOP1000x1500 pixels, 20x20 µm

20x30 mm active area

Measurement setup

Page 14: M4D Det Structured Scintilla Tors for Digital x Ray Imaging

8/8/2019 M4D Det Structured Scintilla Tors for Digital x Ray Imaging

http://slidepdf.com/reader/full/m4d-det-structured-scintilla-tors-for-digital-x-ray-imaging 14/21

Results - Comparison with ”state of the art” competing technology

Example: Integrated Circuit X-ray inspection

State of the art scintillator Scint-X scintillator (40 um pitch)

Page 15: M4D Det Structured Scintilla Tors for Digital x Ray Imaging

8/8/2019 M4D Det Structured Scintilla Tors for Digital x Ray Imaging

http://slidepdf.com/reader/full/m4d-det-structured-scintilla-tors-for-digital-x-ray-imaging 15/21

Results - Comparison with ”state of the art” competing technology

Page 16: M4D Det Structured Scintilla Tors for Digital x Ray Imaging

8/8/2019 M4D Det Structured Scintilla Tors for Digital x Ray Imaging

http://slidepdf.com/reader/full/m4d-det-structured-scintilla-tors-for-digital-x-ray-imaging 16/21

Research at KTH: publications, PhD thesis, patent

Scint-X AB founded in 2007

Four full time employees

Funding: 10 MSEK (3 MSEK venture capital)

Prototypes tested – anticipated function verified

Two ”Letter of Intent” signed (dental x-ray)

Two orders from Swedish Space Corporation (Material analysis)

Project for industrialization started 2009

Where are we now?

Page 17: M4D Det Structured Scintilla Tors for Digital x Ray Imaging

8/8/2019 M4D Det Structured Scintilla Tors for Digital x Ray Imaging

http://slidepdf.com/reader/full/m4d-det-structured-scintilla-tors-for-digital-x-ray-imaging 17/21

Thank you!

Questions? 

Page 18: M4D Det Structured Scintilla Tors for Digital x Ray Imaging

8/8/2019 M4D Det Structured Scintilla Tors for Digital x Ray Imaging

http://slidepdf.com/reader/full/m4d-det-structured-scintilla-tors-for-digital-x-ray-imaging 18/21

Results - Comparison with ”state of the art” competing technology

Page 19: M4D Det Structured Scintilla Tors for Digital x Ray Imaging

8/8/2019 M4D Det Structured Scintilla Tors for Digital x Ray Imaging

http://slidepdf.com/reader/full/m4d-det-structured-scintilla-tors-for-digital-x-ray-imaging 19/21

Conversion Function (CF)Light output vs. Exposure time (dose)- X-Ray to photon conversion- Waveguide efficiency

Sample S14: 365x365 pixels

Characteristics

Page 20: M4D Det Structured Scintilla Tors for Digital x Ray Imaging

8/8/2019 M4D Det Structured Scintilla Tors for Digital x Ray Imaging

http://slidepdf.com/reader/full/m4d-det-structured-scintilla-tors-for-digital-x-ray-imaging 20/21

VarianceVariance vs. Exposure time (dose)- Total noise: X-Ray induced & other 

Sample S14: Gain-compensatedIdeal: Variance = k x Dose

Characteristics

Page 21: M4D Det Structured Scintilla Tors for Digital x Ray Imaging

8/8/2019 M4D Det Structured Scintilla Tors for Digital x Ray Imaging

http://slidepdf.com/reader/full/m4d-det-structured-scintilla-tors-for-digital-x-ray-imaging 21/21

Signal to Noise Ratio (SNR)Signal (CF) / Noise (√variance) vs. Exposure time (dose)

- Total noise: X-Ray induced & other 

Ideal: SNR = k x √Dose

Characteristics


Recommended