+ All Categories
Home > Documents > Mahatma Gandhi Cancer Hospital and Research Institute Dr P. S. Bhattacharyya, MD Radiation...

Mahatma Gandhi Cancer Hospital and Research Institute Dr P. S. Bhattacharyya, MD Radiation...

Date post: 05-Jan-2016
Category:
Upload: rachel-wells
View: 214 times
Download: 0 times
Share this document with a friend
Popular Tags:
17
Mahatma Gandhi Cancer Hospital and Research Institute Dr P. S. Bhattacharyya, MD Radiation Oncologist. Elekta Synergy CT Simulator Flexitron HDR
Transcript
Page 1: Mahatma Gandhi Cancer Hospital and Research Institute Dr P. S. Bhattacharyya, MD Radiation Oncologist. Elekta Synergy CT Simulator Flexitron HDR.

Mahatma Gandhi Cancer Hospital and Research Institute

Dr P. S. Bhattacharyya, MDRadiation Oncologist.

Elekta Synergy

CT Simulator

Flexitron HDR

Page 2: Mahatma Gandhi Cancer Hospital and Research Institute Dr P. S. Bhattacharyya, MD Radiation Oncologist. Elekta Synergy CT Simulator Flexitron HDR.

Computed tomography assisted High Dose Rate Interstitial Brachytherapy in Gynaecologic Malignancies,

with Inverse Planning Simulated Annealing ; An Institutional experience.

Author-Dr P S Bhattacharyya, MD Co-author-Mr A C prabhu, Med. Phy.

Page 3: Mahatma Gandhi Cancer Hospital and Research Institute Dr P. S. Bhattacharyya, MD Radiation Oncologist. Elekta Synergy CT Simulator Flexitron HDR.

Introduction

Ca. Cx - 40% of all cases treated at MGCH EBRT + CDDP 40 mg/m ICRT / IBT GEC Estro guidelines for 3D ICRT MUPIT

2

Page 4: Mahatma Gandhi Cancer Hospital and Research Institute Dr P. S. Bhattacharyya, MD Radiation Oncologist. Elekta Synergy CT Simulator Flexitron HDR.

Aims of the study

To report IPSA based 3D HDR IBT in cancer Cervix and Vagina

DVH analysis of CTV and OAR

Page 5: Mahatma Gandhi Cancer Hospital and Research Institute Dr P. S. Bhattacharyya, MD Radiation Oncologist. Elekta Synergy CT Simulator Flexitron HDR.

Materials and Methods

May 2012 to Dec 2012 Retrospective analysis 34 IBT applications in 17, Ca.Cx / Vagina patients Duration < 3 hrs Median 19 (17-23) needles used 8 – 9.5 Gy X 2 prescibed dose IPSA

Page 6: Mahatma Gandhi Cancer Hospital and Research Institute Dr P. S. Bhattacharyya, MD Radiation Oncologist. Elekta Synergy CT Simulator Flexitron HDR.

Clinical and Histopathological features PRESENTATIONBleeding P/VWhite dischargePelvic Pain

NO. OF PATIENTS 14 12 06

PERCENTAGE 82 70 35

HPESq. Cell Ca.Adeno Ca.

16 01

94 06

SITEIntact CervixVaultVagina

10 05 02

59 29 12

FIGO STAGE IIBIIIBIIIB with L/3 Vagina involved

07 10 03

41 59 30

Total patients studied-17 Median age -56 Yrs.(18-68 yr)

Page 7: Mahatma Gandhi Cancer Hospital and Research Institute Dr P. S. Bhattacharyya, MD Radiation Oncologist. Elekta Synergy CT Simulator Flexitron HDR.

Perineal Template Indications

Bad Geometry 47% (08) Ca. Vault area 29% (05) Vaginal and para vaginal involvement 12% (02) Non negotiable cervical OS 06% (01) ICRT high dose to bladder 06% (01)

Total No. of patients studied-17

Page 8: Mahatma Gandhi Cancer Hospital and Research Institute Dr P. S. Bhattacharyya, MD Radiation Oncologist. Elekta Synergy CT Simulator Flexitron HDR.

Interstitial applicators

Vienna CT-MRI Ring/ Ovoid Applicator

Syed Neblett Template

Page 9: Mahatma Gandhi Cancer Hospital and Research Institute Dr P. S. Bhattacharyya, MD Radiation Oncologist. Elekta Synergy CT Simulator Flexitron HDR.

MUPIT Applicator

Page 10: Mahatma Gandhi Cancer Hospital and Research Institute Dr P. S. Bhattacharyya, MD Radiation Oncologist. Elekta Synergy CT Simulator Flexitron HDR.

BED and DEQ Equations

BEDExt+HDR= (nd[1+(---)]) + (nd[1+(---)])

DEQ = BEDExt+HDR

d d

d ref 1+ ( ---- )

n = no. Of fractionsd = dose per fraction in Gya/b = alpha / beta, 10 for tumor, 3 for late effects

BED=biologically equivalent doseDEq= Equivalent dose (Gy)d ref= reference dose, 2 Gy

a/b a/b

a/b

Page 11: Mahatma Gandhi Cancer Hospital and Research Institute Dr P. S. Bhattacharyya, MD Radiation Oncologist. Elekta Synergy CT Simulator Flexitron HDR.

IPSA Planning

Page 12: Mahatma Gandhi Cancer Hospital and Research Institute Dr P. S. Bhattacharyya, MD Radiation Oncologist. Elekta Synergy CT Simulator Flexitron HDR.

CTV DVH Interpretations

CTV Range Median BED 10Tumor effects

Dose Eq.Gy

D100 4.87 – 8.49 Gy 5.82 Gy

D95 7.26 – 9.17 Gy 8.42 Gy 84.08 70.06

D90 8.14– 9.60 Gy 8.74 Gy 85.80 71.50

V100 85 – 96.77 % 91.13%

Median Prescribed dose- 8.8 Gy

Page 13: Mahatma Gandhi Cancer Hospital and Research Institute Dr P. S. Bhattacharyya, MD Radiation Oncologist. Elekta Synergy CT Simulator Flexitron HDR.

OAR DVH Interpretation

BladderVolume cc

RangeGy

Median dose Gy

BED 3Late effects

Dose Eq.Gy

0.1 6.25-9.85

8.01 130 78.31

1 5.52-7.90

6.75 115.87 69.80

2 4.98-7.13

6.29 110.95 66.84

5 3.82-6.47

5.63 105.36 63.47

Rectal Volume cc

RangeGy

Median dose Gy

BED 3Late effects

Dose Eq.Gy

0.1 6.47-8.49

7.36 122 73.49

1 5.70-7.27

6.49 113.06 68.11

2 5.30-6.01

6.02 108.20 65.18

5 4.23-6.46

5.40 102 61.44

URINARY BLADDER RECTUM

Median Prescribed Dose- 8.8 Gy

Page 14: Mahatma Gandhi Cancer Hospital and Research Institute Dr P. S. Bhattacharyya, MD Radiation Oncologist. Elekta Synergy CT Simulator Flexitron HDR.

IPSA isodose distribution

Page 15: Mahatma Gandhi Cancer Hospital and Research Institute Dr P. S. Bhattacharyya, MD Radiation Oncologist. Elekta Synergy CT Simulator Flexitron HDR.

Comparision of bladder dose

0.1 cc bladder dose 12 Gy with ICRT 0.1 cc bladder dose 8.1 Gy with IBT

Page 16: Mahatma Gandhi Cancer Hospital and Research Institute Dr P. S. Bhattacharyya, MD Radiation Oncologist. Elekta Synergy CT Simulator Flexitron HDR.

Limitations of the study

All the patients did not recieve 6 cycles of CDDP Total duration of treatment more than 8 weeks Applicator volume in CTV No subdivisions in CTV

Page 17: Mahatma Gandhi Cancer Hospital and Research Institute Dr P. S. Bhattacharyya, MD Radiation Oncologist. Elekta Synergy CT Simulator Flexitron HDR.

Thank you