    Operating Instructions

    Sartorius ME and SE Series

    ME and SE Models

    Electronic Analytical Balances and Semi-micro-, Micro- and Ultra-Microbalances

    98648-008-84

  • 7/25/2019 MAN-ME_SE-e

    2/173

    ME and SE models are high-resolutionbalances of special accuracy forextremely precise measurement of mass.

    These series cover a range from0.001 mg to 610 g.

    A broad range of special performancefeatures makes the ME and SE balancesideal for use as measuring and testequipment in ISO or GLP quality man-agement systems.These features include:

    The fully automatic self-calibratingand adjustment function, isoCAL(time- and temperature-dependent)

    reproTEST for quick determination ofthe standard deviation to check therepeatability of results

    ISO/GLP-compliant recording capabilityfor printouts

    Password-protected menu lock

    Display of maintenance |service intervals when due

    ME and SE balances meet the highestrequirements placed on the accuracyand reliability of weighing results

    through the following features: Efficient filtering-out of vibration

    Fully automatic draft shield with threemotorized, self-teaching draft shieldelements and soft-touch technology

    Stable and repeatable results

    Excellent readability under anylighting conditions

    Rugged design and durable weighingsystem

    ME215/235/254/414/614: Integrated static electricity eliminator

    feature to neutralize interfering electro-static charges (ionizer)

    ME and SE balances save work andspeed up both simple and complexroutine applications through:

    Ultrafast response times

    Built-in application programs;application level 1: Second weight unit Counting Weighing in percent Animal weighing Recalculation Calculation Density determination Differential weighing Air buoyancy correction Air density determination

    for ME5, SE2 Diameter determination

    Application level 2: Checkweighing Time-controlled functions

    Application level 3: Totalizing Formulation Statistics

    with the following additional functions: Second tare memory

    Identification codes Product data memory SQmin function Manual data storage in

    application level 3 DKD uncertainty of measurement

    Automatic initialization when youswitch on the balance

    Easy input of IDs forsamples or other weighed objects

    If requested: control using anexternal computer

    SymbolsThe following symbols are used inthese instructions:

    indicates steps you must perform

    $ indicates steps you must performonly under certain conditions

    > describes what happens after youhave performed a certain step

    indicates an item in a list

    ! indicates a hazard

    Conventions Used in These

    Operating Instructions: The pictures in these Operating

    Instructions are based on theME215S model. On other models,some display readouts and printoutsmay differ slightly from the onesshown. This will be explained incases where this is important foroperation of the balance.

    Hotline:For advice on the use of applica-tions, just call or fax your localSartorius office. For the address,please visit our Internet website at:www.sartorius.com

    2

    Intended Use

  • 7/25/2019 MAN-ME_SE-e

    3/173

    2 Intended Use

    4 Warning and Safety Instructions

    5 General Views of the Balances5 ME215/235/254/414/6146 ME36S7 ME5, SE2

    8 Operating Design

    12 Getting Started12 Storage and Shipping Conditions12 Unpacking the Balance12 Carrying the Balance14 Equipment Supplied15 Installation Instructions15 Remote Operation of the

    Display and Control Unit17 Connecting the Balance to AC Power18 Warmup Time19 Leveling the Balance20 Transporting the Balance

    23 Configuring the Balance23 Selecting the Language24 Navigating in the Setup Menu25 Entering the Time and Date26 Setting the Balance Functions29 Setting the Device Parameters29 Entering a Password34 Setting the Application Parameters

    42 Selecting the Printout Function44 Printout Configuration46 Device Information46 Factory Settings

    47 Operating the Balance47 Basic Weighing Function47 General Instructions for

    Analytical Weighing48 Below-Balance Weighing51 Device Parameters51 Opening and Closing the Draft Shield53 Static Electricity Eliminator (Ionizer)55 Calibration, Adjustment, Linearization65 Repeatability Test

    66 Application Programs67 Toggle between Two Weight Units69 Counting72 Weighing in Percent75 Calculation78 Density Determination83 Differential Weighing95 Air Buoyancy Correction

    102 Diameter Determination

    3

    Contents

    105 Time-Controlled Functions108 Statistics113 Extra Functions

    113 Second Tare Memory115 Individual Identification Codes119 Saving Values Manually in M+120 Changing the Resolution122 Product Data Memory124 SQmin Function126 DKD Uncertainty of Measurement128 Combining Applications129 Practical Combination of Several

    Applications (Example)131 Data Output Functions133 Interfaces136 Printouts140 Serial Communications Port145 Pin Assignment Charts146 Cabling Diagram

    147 Error Codes and Messages

    150 Care and Maintenance

    152 Recycling

    153 Overview153 Specifications159 Accessories (Options)160 Dimensions (Scale Drawings)163 Declarations of Conformity166 EC Type-Approval Certificate

    168 Plates and Markings171 Index

    AppendixEntering the General PasswordBrief Instructions

  • 7/25/2019 MAN-ME_SE-e

    4/173

    This balance complies with theEuropean Council Directives as well asinternational regulations and

    standards for electrical equipment,electromagnetic compatibility, andthe stipulated safety requirements.Improper use or handling, however,can result in damage and/or injury.

    Read these operating instructionsthoroughly before using yourbalance to prevent damage to theequipment. Keep these instructions ina safe place.

    Follow the instructions belowto ensure safe and trouble-freeoperation of your balance:

    ! Do not operate in a hazardousarea/location

    ! Make sure that the voltage ratingprinted on the AC adapter is identicalto your local line voltage

    ! If you use electrical equipment ininstallations and under ambientconditions requiring higher safetystandards, you must comply with theprovisions as specified in the appli-cable regulations for installation in

    your country.

    ! Exposure to excessive electromagnet-ic interference can cause the readoutvalue to change. Once the distur-bance has ceased, the instrument canbe used again in accordance with itsintended purpose.

    The only way to switch the power offcompletely is to disconnectthe AC adapter

    The balance housing is protectedagainst the penetration of solid objectswith a diameter of more than 2.5 mm(such as accumulated dust) anddripping water falling vertically (IP32) the housing is not completely dust-and leak-tight, however

    Protect the AC adapter from contactwith liquid

    Note on Installation:The operator shall be responsible forany modifications to Sartorius equip-ment and for any connections of cablesor equipment not supplied by Sartoriusand must check and, if necessary,correct these modifications and connec-tions. On request, Sartorius will provideinformation on the minimum operatingspecifications (in accordance with theStandards listed above for definedimmunity to interference).

    Connect only Sartorius accessories andoptions, as these are optimally designedfor use with your balance

    When cleaning your balance, makesure that no liquid enters the balancehousing; use only a slightly moistenedcloth to clean the balance.

    Do not open the balance housing.If the seal is broken, this will result inforfeiture of all claims under themanufacturers warranty.

    In case you have any trouble with yourbalance:

    $ contact your local Sartorius office,dealer or service center

    4

    Warning and Safety Instructions

  • 7/25/2019 MAN-ME_SE-e

    5/173

    5

    Models ME215/235/254/414/614

    General Views of the Balance

    Pos. Designation Spare Part Order No.

    17 On/off key18 Toggle key for alphabetic input19 Function keys20 Leveling foot 69MA009121 Lug for attaching an antitheft

    locking device22 Menu access switch23 Serial printer port (PRINTER)24 Serial communications port

    (PERIPHERALS)25 Terminal for connecting an

    equipotential bonding conductor

    26 DC jackNot shown:Set of dust covers 6960ME01Set of small parts (operating panel) 69ME0007Set of caps 69ME0008

    Pos. Designation Spare Part Order No.

    1 Draft shield door grips2 Weighing pan 69 ME00013 Shield disk (ME235S/P only)4 Shield plate 69 ME00025 Level indicator6 Operating panel7 Display8 Tare key9 Key for opening/closing 69ME0007

    draft shield (set of small parts)10 Decimal point key11 Print key

    12 Ionizer on/off key13 CF key (clear function)14 Numeric keys15 Toggle key for changing the application program16 Setup key for configuring the balance

    1

    2

    34

    5

    6

    7

    8

    910

    12

    13

    1415

    1617

    18

    8

    9

    19

    20

    21

    20

    22

    2324

    25

    26

    11

  • 7/25/2019 MAN-ME_SE-e

    6/173

    6

    General Views of the Balance

    Pos. Designation

    1 Draft shield door grips2 Weighing pan3 Shield disk4 Shield plate5 Level indicator6 Operating unit with control panel7 Serial communications port (PERIPHERALS)8 Serial printer port (PRINTER)9 Menu access switch

    10 DC jack11 Female connector for weigh cell

    12 Decimal point key13 Print key14 Info key for displaying device information

    ME36S

    Pos. Designation

    15 CF key (clear function)16 Toggle key for changing the application program17 Setup key for configuring the balance18 On/off key19 Toggle key for alphanumeric input20 Numeric keys21 Function keys22 Key for opening/closing draft shield23 Tare key24 Display25 Connector for plugging the weigh cell into the operating panel

    26 Leveling foot27 Lug for attaching an antitheft locking device28 Terminal for connecting an equipotential bonding conductor

    MES_06.cdr

    1817

    12

    13141516

    19

    20

    21

    28

    27

    26

    26

    25

    24

    2223

    2322

    1

    2

    3

    4

    5

    6

    789

    10

    11

  • 7/25/2019 MAN-ME_SE-e

    7/173

    7

    Pos. Designation

    1 Weighing pan2 Filter pan, 50 mmd3 Interior draft shield (only for SE2)4 Optional filter pan, 75 mmd5 Shield ring6 Optional filter pan, 90 mmd7 Shield disk8 Level indicator9 Leveling foot

    10 Draft shield cover11 Female connector for weigh cell

    12 DC jack for AC power13 Communications port (PERIPHERALS)14 Printer port (PRINTER)15 Tare key

    Pos. Designation

    16 Key for opening the draft shield counterclockwise17 Print key18 Info key for displaying device information19 CF key (clear function)20 Numeric keys21 Toggle key for changing to the next application program22 Key for accessing Setup mode (settings)23 On/off key24 Toggle key for alphanumeric input25 Key for opening the draft shield clockwise26 Function keys

    27 Display28 Lug for attaching an antitheft locking device29 Female connector for evaluation unit30 Terminal for connecting an equipotential bonding conductor31 Draft shield

    General Views of the Balance

    Models ME5 and SE2

    12

    3

    4

    5

    6

    7

    8

    9

    10

    11

    12

    13

    14

    15

    17

    16

    18

    19

    20

    23

    24

    25

    15

    28

    29

    30

    7

    1

    3

    31

    22

    21

    27

    26

  • 7/25/2019 MAN-ME_SE-e

    8/173

    8

    The balance consists of a weighing cell, adraft shield and a display and control unit.In addition to the choice of power supply,

    via AC adapter or external rechargeablebattery pack, your balance also has inter-face ports for connecting additionaldevices, such as a printer, computer, oruniversal remote control switch, etc.

    The display and control unit is fastenedto the weighing cell. Operation of thebalance follows a uniform philosophy,which is described in this manual.

    Where not expressly indicated otherwise,the uses described in this manual apply toverified balance versions (indicated by thesuffix -.0CE in the model number), aswell as the standard version.

    Combination of Several ApplicationsYou can combine the use of variousapplication programs to meet your morecomplicated requirements.

    To select application programs one afterthe other, press d (toggle function).

    KeysYou can operate the balance either by

    using the keys on the display andcontrol unit or from an on-line PC.This manual describes operation usingthe balance keys.

    Labeled KeysThese keys always have the functionindicated by their label, but are notavailable at all times. Availability of theirfunctions depends on the currentoperating status of the balance and themenu settings.

    Meaning

    a Alphabetic keysPlease see section on Text Input

    e On | off keyTurns the balance on and offor switches it to the standby mode

    oMenu settingsAccesses and exits the Setup menu

    d Toggles to the nextapplication program

    c Clear functionDeletes keypad input

    Interrupts a calibration andadjustment routine in progressQuits application programs

    Operating Design

    ME215/235/254/414/614:J Turns the ionizer on and off

    ME36S, ME5, SE2:JDisplays device information

    rPrint keyOutputs displayed valuesor data logs to the serialcommunications and | orprinter port

    .Enters a decimal point

    1 ... 9 0 keysSee the section on Numeric Input

    w Tares the balance

    K, O, P Opens | closes the draft shield

    Numeric InputTo enter numbers: press1 9 0 .

    To store numbers entered: press the

    corresponding function key directlybelow the soft key label

    To delete an entire numeric input digitby digit: press the c key

    Text Input To enter numbers:

    see the section on Numeric Input

    To enter letters or characters:press the a key

    > Letters are displayed in the bottom linefor selection

    To select a different letter:

    press the corresponding soft keyto change the letter shown

    To select the letter | character shown:press the corresponding function keybelow the soft key label

    > The selected letter is shown onthe display

    $ Enter the next letter | character,if desired, as described above

    $ To exit the letter input mode (e.g., if thelast character entered is a letter): press

    the a key

    To store a word: press thecorresponding function key (soft key),such as ID

    To delete an input character by character:press the c key

    To delete user data: enter . ora space and save

  • 7/25/2019 MAN-ME_SE-e

    9/173

    9

    Function Keys (Soft Keys)The current function of soft keys isindicated in the bottom line of the display

    (footer).

    Texts (abbreviations) or symbols can bedisplayed.

    Texts (Examples)Cal: Start calibration | adjustmentS ID: Save ID

    The function keys are numbered from right(F1) to left (F6).

    SymbolsThe bottom line shows the followingsymbols:ooBack to the initial state

    (in the Setup menu: exit Setup)

    o Go to the higher selection level

    O Show sub-items under the active item

    Q Move upward in the input |output window

    q Move downward in the input |output window

    l Set the selected menu parameter

    F6 F5 F4 F3 F2 F1

    Operation

    Display for Weights andCalculated ValuesThis display is subdivided into 9 areas.

    Line for Metrological Data:When the balance is used in legal metrol-ogy, the following metrological specifi-cations of the balance are shown here:

    Max Maximum capacity (upper rangelimit) of the balance

    Min Minimum capacity (lowerrange limit) of the balance

    e Verification scale interval

    d Readability | scale interval

    On standard balances, onlyMax

    andd

    are displayed.

    Operating the Balance as a LegalMeasuring Instrument (Legal forTrade)

    Remove the cover plate from the backof the balance housing

    Move switch1 in the directionof the arrow

    > Switch left:Standard/not legal for tradeSwitch right:

    Legal for trade

    > Note:Do not move switch 2

    2

    1

    Line for metrological dataBar graph

    Measured value line

    Text lineSoft key labels

    Plus/minus sign UnitStability indicator Tare memory

    Calculated valueApplication pictograms

    Bar Graph:The bar graph indicates how much of thebalances capacity is used up by the

    current load; during checkweighing, itindicates the control limits.

    The following symbols may be displayed:

    0% Lower load limit

    100% Upper load limit

    Bar graph showing 10% intervals

    - Minimum for checkweighing

    = Target for checkweighing

    + Maximum for checkweighing

    Plus/Minus Sign, Stability Symbol:A plus or minus sign (F or H) is shownhere for a weight (or a calculated value,such as that for counting), or the Ssymbol indicating that a verified balance*has been zeroed or tared.

    Line for Measured Values:This area shows the weighed or calculatedvalue and the alphanumeric input.

    Unit and Stability:

    When the balance reaches stability, theweight unit or calculated unit is displayedhere.

    When the a symbol is displayed here,the value indicated in the readout cannotbe used in legal metrology.

    * = Verification scale interval e =scale interval d

    Operating Design

  • 7/25/2019 MAN-ME_SE-e

    10/173

    10

    Display for Menu ParameterSettings (Setup)This display is divided into three sections.

    Status Line:The status line of shows the functionof the display screen page. In the Setupmenu, the current menu path isshown here.

    Setup Menu Example: Balance/scale

    functions:

    Input and Output WindowThis window contains either detailedinformation (e.g., on the activeapplication) or a pick list. A selecteditem is displayed inversely (whitecharacters on a black background).You can also enter information in anactive field in this window using thealphabetic and numeric keys.

    Setup Menu Example,Device parameters, Adapt filter:

    The following symbol may be displayedin the input and output window:

    d this symbol marks the savedmenu setting

    Soft Key LabelsSee the description Function Keys(Soft Keys) on the previous page

    Minimum vibration

    o Normal vibration

    Strong vibration

    Extreme vibration

    SETUP BAL.FUNC.

    Line for Operating State

    Input and Output Window

    Soft Key Labels

    To set a parameter: Press the Q or q soft key repeatedly

    until the desired setting is selected

    (displayed inversely)

    Confirm your selection:press the l soft key

    To change the numeric value ofa parameter:

    Press the Q or q soft key repeatedly,if necessary, until the desired setting isselected (displayed inversely)

    Enter a new value or character:use the 0 1 9 . keysor the a key and enterthe desired letters

    Confirm your selection:press the l soft key

    To exit Setup: press the oo soft key

    Tare Memory, Calculated Values:The symbols displayed here indicate whenthere is a value in one of the tare

    memories or when the value shown is aresult of calculation rather than directmeasurement.

    These symbols are as follows:

    a Calculated value

    s1 Net value | tare memory

    s2 used by an application program(e.g., formulation, second tarememory)

    Application Pictograms:The pictograms displayed here indicatethe application(s) selected. The pictogramis displayed inversely (white on a blackbackground) when the correspondingapplication is active.

    For example, the following symbols may bedisplayed simultaneously:

    A The counting application is active

    H Checkweighing is also active

    S Print

    T Data record

    Text Line:Additional information is displayed here(e.g., operator guidance prompts, name ofthe active program, etc.)

    Soft Key Labels:The current functions of the soft keysabove the function keys (arrow keys) areindicated here; during calibration | adjust-ment, this line shows up- and down-arrows(Q and q) for selecting calibration andadjustment functions.

  • 7/25/2019 MAN-ME_SE-e

    11/173

    11

    Input

    Bar Code Scanner or Keyboard Input

    You can use a bar code scanner or anexternal keyboard to input alphanumericvalues. These inputs are processed inthe same manner as keypad inputs onthe display and control unit of the bal-ance. Bar code and keyboard inputs areonly displayed; they cannot activate anyfunction.

    To assign a bar code scanner orkeyboard input to a function, press oneof the following soft keys:

    Lot

    Samples Measured values Sample number Tare value Initial weight Backweighed value Sample ID

    Foot or Hand Switch InputYou can connect a foot switch ora hand switch to the balance to havethis device perform a keypadfunction (such as c or w).

    PC InputYou can use a computer to control thefunctions of the balance and displayand control unit via the communica-tions port (see the Data OutputFunction section in the chapterentitled Operating the Balance).

    Data Output

    The balance provides two interfaceports for outputting weights, calculatedvalues and parameter settings:

    Serial communications port(PERIPHERALS Serial I/O)

    Serial printer port(PRINTER Serial Out)

    Serial Printer PortIn addition to Sartorius printers (such asthe YDO03-0CE), you also have thechoice of connecting a remote displayor an external checkweighing display to

    the printer port.You can configure the data outputfunctions in the Setup menu to meetyour various requirements, includingISO | GLP requirements.

    ISO: International Organizationfor Standardization

    GLP:Good Laboratory Practice

    You can have printouts generatedautomatically, or by pressing r;generation can be dependent on or

    independent of the stability or timeparameters.

    See the section on Data OutputFunctions in the chapter entitledOperating the Balance for a detaileddescription.

    Serial Communications PortYou can connect a PC, a remote display,an external checkweighing displayor a standard (non-verifiable) printerto this port.

    Request messages are sent via theinterface to initiate functions in theweighing cell and in the display andcontrol unit. Some of the functionsgenerate response messages.

    See the section on Data OutputFunctions in the chapter entitledOperating the Balance for a detaileddescription.

    Error CodesIf you press a key that has no function,or which is blocked at a certain point in

    an application program, this error isindicated as follows:

    a double beep is sounded as an acousticsignal if the key has no function

    a double beep is sounded and themessage No function is displayed inthe text line if the key function is notavailable at that time

    The response to an operator error isidentical in all operating modes. See thechapter entitled Error Codes for adetailed description.

    Storing SettingsSaving Parameter SettingsThe settings configured remain storedin the balances non-volatile memory. Inaddition, you can reload the factory set-tings.

    Saving SettingsUnder Setup > Device parameters >password you can assign passwords inorder to block access to:

    Balance | scale functions

    Device parameters

    Application parameters

    Printout

    Factory settings

  • 7/25/2019 MAN-ME_SE-e

    12/173

    Storage and Shipping ConditionsAllowable storage temperature: 5C to 40C (41F to 104F)The packaging has been designed to ensure that the A/D converter will not be

    damaged even if it is dropped from a height of 80 cm (approx. 31 inches). Do notexpose the balance to extreme temperatures, moisture, shocks, blows or vibration.

    Unpacking the Balance

    Lift the inner package containing the balance out of the outer packaging by the strap.

    Loosen and remove the strap.

    Remove the cardboard sleeve.

    12

    Getting Started

  • 7/25/2019 MAN-ME_SE-e

    13/173

    13

    Remove the following parts from the recessed spaces at the top of theinner packaging:

    AC adapter (in cardboard packaging)

    Weighing pan Shield plate

    Remove the two padding blocks that make up the inner packaging by pulling outward.

    Remove the shield disk (included with ME235S/P models only) and dust cover.

    Remove the retainer securing the front draft shield panel.

    Open the plastic wrapping.

    Place one hand under the display and control unit and the other under the back of thebalance and lift the balance out of the lower packaging.

  • 7/25/2019 MAN-ME_SE-e

    14/173

    14

    ! Do not lift the balance by the draft shield or the front panel, as this can resultin damage.

    Set up the balance at the place of installation. Open the draft shield doors.

    Remove the retainers (1)securing the display and control unit and remove the frontdraft shield panel.

    Remove the foam padding (2)from the draft shield.

    ! Save the box and all parts of the packaging in case it should become necessary totransport the balance over a long distance. Only the original packaging provides thebest protection for shipment (see also Transporting the Balance on page 20).Before packing the balance, unplug all connected cables to prevent damage.

    Equipment SuppliedThe following individual componentsare supplied:

    ME215/235/254/414/614 Balance AC adapter with power cord Weighing pan with hanger for below-

    balance weighing

    Shield disk Dust cover for the balance housing Dust cover for the display and control

    unit Instruction manual

    ME5, SE2 Weigh cell Draft shield Electronic evaluation unit Connecting cable AC adapter with power cord Accessories kit

    The accessories kit includes:

    Weighing pan Shield disk Interior draft shield (only for SE2) Brush Forceps Cloth

    ME36S Weigh cell Electronic evaluation unit AC adapter with power cord Weighing pan

    Shield disk

    ME5-F, SE2-F Weigh cell Draft shield cover Shield ring Electronic evaluation unit Connecting cable AC adapter with power cord Accessories kit

    The accessories kit includes: Filter pan, 50 mm d Weighing pan Shield disk

    Interior draft shield (only for SE2) Brush Forceps Cloth

    1

    1

    2

  • 7/25/2019 MAN-ME_SE-e

    15/173

    15

    Installation InstructionsThe ME/SE balances are designed to provide reliable weighing results under normal ambientconditions in the laboratory and in industry. Choose the right location to set up your balance

    by observing the following so that you will be able to work with added speed and accuracy: Set up the balance on a completely even surface on a low-vibration balance table or

    wall console Avoid placing the balance in close proximity to a heater or otherwise exposing the

    balance to heat or direct sunlight, as this can considerably increase the temperatureinside the draft shield (greenhouse effect), resulting in incorrect readouts dueto convection currents, turbulence and buoyancy effects.

    Protect the balance from drafts that come from open windows or doors Avoid brief fluctuations in room temperature Protect the balance from aggressive chemical vapors Do not expose the balance to extreme moisture

    Linearization after TransportAfter transporting the balance, its linearity may be outside the allowable tolerances(please refer to the Specifications in the Overview chapter). After transporting thebalance, be sure to perform internal linearization. Repeat this process to obtain optimalaccuracy. For directions on this procedure, please refer to the section on Linearization.

    Conditioning the Balance

    Moisture in the air can condense on the surfaces of a cold balance whenever it is broughtinto a substantially warmer place. If you transfer the balance to a warmer area, makesure to condition it for about 2 hours at room temperature, leaving it unplugged fromAC power. Afterwards, if you keep the balance connected to AC power, the continuouspositive difference in temperature between the inside of the balance and the outsidewill practically rule out the effects of moisture condensation.

    Setting Up the Balance ME215/235/254/414/614 Place the components listed below inside the weighing chamber in the order given: 1) Shield plate 2) Position the weighing pan and turn to the left or right until it snaps into place 3) Shield disk (ME235S/P only)

    $ Note: the shield disk minimizes the effects of drafts within the weighing chamber

    Remote Operation of the Display and Control Unit Unplug the cables, turn the balance on its side and lay it on a padded surface

    to avoid damaging the weighing system and draft shield

    Use an Allen wrench to remove the three fastening screws

    Remove the display unit and attach the connecting cable

    > Length of the connecting cable: 44 cm (17 inches)

    $ For information on longer cables, please see the Accessories section

    $ If you wish to use a longer cable, it must be installed by authorized

    Sartorius service technicians

    3

    2

    1

    ME215/235/254/414/614:

  • 7/25/2019 MAN-ME_SE-e

    16/173

    16

    Setting Up the ME36S Balance Place the components below in the weighing chamber in the order given:

    Shield plate

    Weighing pan

    Setting Up the ME5 or SE2 Balance Place the components below on the weigh cell base in the order given:

    Shield disk

    Weighing panPlease note: after placing the weighing pan on the base, turn the pan slightly towardsthe left then the right while pressing down on it to lock.

    Interior draft shield (only for SE2)

    Draft shield: center the hole over the pan (see arrows)

    Setting Up the ME5-F or SE2-F Balance Place the components below on the weigh cell base in the order given:

    Shield disk

    Interior draft shield ring

    50 mm dia. filter pan or weighing pan (or, optionally, the 75 or 90 mm dia. filter pan)Please note: after placing the pan on the base, turn the pan slightly towards the leftthen the right while pressing down on it to lock

    Interior draft shield (only for SE2)

    ! If the weighing pan is removed during operation, turn the balance off and then on againafter you return it to the balance.

    Draft shield cover

    Setting Up the Filter Balance for Left-handed Persons:

    Remove the draft shield cover

    Detach the pin on the right and re-attach on the left

    ME36S:

    ME5, SE2:

    ME5-F, SE2-F:

  • 7/25/2019 MAN-ME_SE-e

    17/173

    17

    Turn draft shield assembly by approx. 90 degrees toward the left(loosen knurled knob to turn)

    Connect the weigh cell to the evaluation unit Use a screwdriver to tighten the screws to the female connector on the weigh cell

    Connecting the Balance to AC PowerThe wide-range AC adapter is designed for 100 V to 240 V.

    Check the plug design of the power cord If it does not fit your wall outlet (mains supply), please contact your

    Sartorius office or dealer

    Use only Original Sartorius AC adapters and power cords

    AC adapters with a registered approval rating from a national testing laboratory

    $ To use a main feeder cable from the ceiling or to mount a CEE plug,have a certified electrician install it

    $ To use an external rechargeable battery pack, refer to the Accessories in theOverview chapter

    Insert the AC adapter plug with the angle facing downward into the jack on the balance

  • 7/25/2019 MAN-ME_SE-e

    18/173

    18

    Plug power cord into the AC adapter

    To power the balance with AC current, plug the power cord into

    a wall outlet (mains supply)

    Charging the Rechargeable Battery for Saving Data:All data is saved in the battery-backed memory. When initially operating the balance,leave it connected to AC power for one day to charge the battery. When the balanceis disconnected from AC power, the balance-generated data will remain stored forapproximately three months. In the standby mode, data is retained in the memory viathe power supply. Be sure to print out data before storing your balance for a relativelylong period.

    Safety PrecautionsThe AC adapter rated to Class 2 can be plugged into any wall outletwithout requiring any additional safety precautions. The ground or earth terminal isconnected to the scale housing, which can be additionally grounded, if required.The data interface is also electrically connected to the balance housing (ground).

    Note:This equipment has been tested and found to comply with the limits for a Class A digitaldevice, pursuant to Part 15 of the FCC rules. These limits are designed to providereasonable protection against harmful interference when the equipment is operated ina commercial environment. This equipment generates, uses and can radiate radiofrequency energy and, if not installed and used in accordance with the instructionmanual, may cause harmful interference to radio communications. Operation of thisequipment in a residential area is likely to cause harmful interference in which case theuser will be required to correct the interference at his own expense. Changes ormodifications not expressly approved by Sartorius AG could void the users authorityto operate the equipment.

    Connecting Electronic Peripheral Devices Make absolutely sure to unplug the balance from AC power before you connect or

    disconnect a peripheral device (printer or PC) to or from an interface port

    Warmup TimeEach time you move your balance to another location, you must condition it for at least12 hours to the new location. To deliver exact results, the balance must warm up for atleast 12 hours after initial connection to AC power. Only after this time will the balance

    have reached the required operating temperature.

    Using Balances Verified as Legal Measuring Instruments in the EU*:$ The balance must warm up for at least 24 hours after initial connection

    to AC power

    $ Warmup time each time power is turned on the ME614S-0CE, ME414S-0CE:at least 30 minutes

    $ Always wait for the power-on adjustment routine to be completed:for requirements see page 63.

    * including the Signatories of the Agreement on the European Economic Area

  • 7/25/2019 MAN-ME_SE-e

    19/173

    19

    Antitheft Locking DeviceTo fasten an antitheft locking device, use the lug located on the rear panelof the balance.

    Secure the balance at the place of installation, e.g., with a chainor a lock

    Leveling the BalancePurpose:

    To compensate for unevenness at the place of installation

    To achieve perfectly horizontal positioning of the balance for consistent repeatabilityof the weighing results

    Always level the balance again any time it has been moved.

    Only the 2 front feet are used for leveling.

    Turn the leveling feet as shown in the diagram until the air bubble is centered exactlywithin the circle of the level indicator

    > Several leveling steps are usually required

    Setting the Language> See the section on Setting the Language in the chapter Configuring the Balance

    Setting the Date and Time> See the example on page 25, in the chapter Configuring the Balance

  • 7/25/2019 MAN-ME_SE-e

    20/173

    $ The balance must be disconnected from power before preparing it for shipment.Disconnect the AC adapter and all interface cables from the balance.

    Transport Over Short Distances Place one hand under the display and control unit and the other under the back of the

    balance. Lift the balance carefully and carry it to the new location.

    $ Avoid subjecting the balance to vibration or shocks (impact).

    ! Do not lift the balance by the draft shield or the front panel, as this can result

    in damage.

    Transport or Shipping Over Long DistancesMake sure you use all components of the original packaging in the following cases:

    for transport or shipping over long distances. if it is not certain that the balance will remain upright during transport or shipping.

    Remove the following parts: Shield disk (included with ME235S/P models only) Weighing pan and shield plate:

    Reach beneath the shield plate and lift it carefully, together with the weighing pan,to avoid damaging the weighing system.

    Open the draft shield doors and carefully position the balance in the lower packaging. Press the foam padding against the housing.

    20

    Shipping the Balance

  • 7/25/2019 MAN-ME_SE-e

    21/173

    21

    Attach the retainers to secure the front draft shield panel and the display andcontrol unit.

    Place the dust cover on the display and control unit. Models ME235S/P only: Place the shield disk in a bag and lay it on the dust cover.

    Position the two padding blocks that make up the inner packaging and press inward.

    Place the following parts in the recessed spaces at the top of the inner packaging: Shield plate Weighing pan AC adapter (in cardboard packaging)

    Wrap the cardboard sleeve around the inner packaging.

  • 7/25/2019 MAN-ME_SE-e

    22/173

    Place the strap around the package and tighten it.

    Lift the packaged balance by the strap and place it in the shipping box, with the

    bottommost padding already in the box.

    Place the uppermost padding on top of the inner package.

    Close the shipping box and seal it appropriately for the intended transportor shipment.

    22

  • 7/25/2019 MAN-ME_SE-e

    23/173

    23

    Configuring the Balance

    PurposeYou can configure your ME/SE balanceto meet individual requirements by

    entering user data and settingparameters in the Setup menu.

    The Setup menu is divided into thefollowing items:

    Balance/scale functions Device parameters Application parameters Printout Device information Language Factory settings

    Setting the LanguageYou can choose from 5 languages forthe information display:

    German English (factory setting) English with U.S. date | time format French Italian Spanish

    Example: Selecting the Language U.S. Mode

    Step Press key(s) (or follow instructions) Display | Printout

    1. Select Setup menu o

    2. Select Language Repeatedlyand confirm press q soft key,

    thenO soft key

    3. Select U.S. mode q soft key

    4. Save language l soft key

    5. Exit the Setup menu oo soft key

    SETUP LANGUAGE

    Deutsch

    oEnglish

    U.S.-Mode

    Franais

    Italiano

    oo o q l

    SETUP LANGUAGE

    Deutsch

    English

    oU.S.-Mode

    Franais

    Italiano

    oo o q l

  • 7/25/2019 MAN-ME_SE-e

    24/173

    24

    Navigating in the Setup Menu (Examples):

    Example: Adapt the balance to Extreme vibration by selecting this setting

    Step Press key(s) (or follow instructions) Display | Printout

    1. Select Setup menu o

    2. Confirm Balance/scale functions O soft key

    3. Select menu item Adapt filter q soft key, thenand confirm O soft key

    4. Select menu item Extreme vibration q soft key

    5. Confirm menu item l soft key

    Extreme vibration

    6. If required, select further menu items q Q soft keys

    7. Save setting and exit Setup menu oo soft key

    Exiting the Setup MenuIf you use the oo soft key:

    The software will be restarted if you have changed a setting. The software will not be restarted if you have kept the same settings.

    In this case, the program will return to its initial state before you pressthe o key.

    If you press the o key: When you exit o, the software is generally restarted.

  • 7/25/2019 MAN-ME_SE-e

    25/173

    25

    Example: Entering the time and date

    Step Press key(s) (or follow instructions) Display | Printout

    1. Select Setup menu; o, thenselect Device parameters q and O soft keys

    2. Set clock press q repeatedly,then press O

    3. Enter the time 1 1 . 1 2

    . 3 0

    4. Set the time according toyour local clock l soft key

    5. Enter the date 1 3 . 0 3. 00

    6. Store the date l soft key

    7. Enter other data, if desired q Q soft keys

    8. Exit Setup menu oo soft key

    * = not on ME36S, ME5, SE2

  • 7/25/2019 MAN-ME_SE-e

    26/173

    26

    PurposeThis menu item enables you toconfigure the balance functions, i.e.,

    to meet individual requirements byselecting predefined parameters in theSetup menu. You can block access tothe menu by assigning a password.

    FeaturesThe balance functions are combined inthe following groups(1st menu level):

    Calibration | adjustment Adapt filter Application filter Stability range Taring Auto zero Weight unit 1 Display accuracy 1 Tare/zero with power on Factory settings: only wgh. param. (only

    the balance functions)For legal metrology, the selectionof individual parameters is limited.

    Factory SettingsParameters: The factory settings areidentified by the symbol o in the liststarting on page 27.

    PreparationShow available balance functions:

    Select Setup menu: press the o key

    > SETUP is displayed

    Select Balance functions:press the O soft key

    If you already assigned a password:

    > The password prompt is displayed

    $ If access is blocked by a password:enter the password using the numeric |alphabetic keys

    $ If the last character of the password isa letter: conclude input by pressing a

    Confirm your password and havethe balance functions displayed:Press the l soft key

    > Balance functions are displayed:

    $ To select the next group:press the q soft key (down arrow)

    $ To select the previous item ofa group: press the Q soft key(up arrow)

    $ To select the next sub-item withina group:press the O soft key (right arrow)

    $ To select the previous group:press the o soft key (left arrow)

    $ To confirm: press the l soft key

    Extra Functions Exit the Setup menu:

    press the oo soft key

    > Restart your application

    Print parameter settings: When the balance functions are

    displayed, press r> Printout (example)

    Texts with more than20 characters are cut off

    SETUPBAL. FUNC.

    --------------------Calibration/adjustmCAL/isoTST key fun

    Selection modeCal/adjustment seq

    Calibrate, then autoadjust

    isoCAL functionOn without resetting

    app.Start automatic ad

    isoCALPrint GLP/GMP adju

    Automatic if GLP isselected

    Parameter for exteWt. ID (W ID):

    Cal./adj. wt:200.00000 g

    Adapt filterNormal vibration

    Application filterFinal readout

    Stability range2 digits

    Auto zeroOn

    Weight unit 1Grams /g

    Display accuracy 1All digits

    --------------------

    Setting the Balance Functions (BAL.FUNC.)

  • 7/25/2019 MAN-ME_SE-e

    27/173

    27

    Balance Functions (Overview) factory setting user-defined setting(s)

    Setup Calibration | Cal | isoTST External calibration | adjustment** withBalance Adjustment key function factory-defined calibration weightfunctions External calibration | adjustment**;

    user-defined weightInternal calibration | adjustmentInternal linearizationCal key blockedreproTEST

    Selection mode for cal. | adjustment

    Calibration | Calibrate, then auto adjustadjustment sequence Calibrate, then manual adjust

    isoCAL function Off Only adjustment promptisoCAL and linearization onOn and reset application

    On without resetting application

    Start automatic User-defined adjustment Enter adjustment time 1: (hh.mm)adjustment times Enter adjustment time 2: (hh.mm)

    Enter adjustment time 3: (hh.mm) isoCAL

    Print GLP Automatic if GLP is selectedadjustment record On request,

    from record memory

    Parameter for Weight set ID (W ID): Enter 14 characters max.external weight Calibration |

    adjustment weight: Enter exact weight

    Adapt filter Minimum vibration(ambient Normal vibrationconditions) Strong vibration

    Extreme vibration

    Application filter Final readoutFilling mode

    Stability range 14 digit12 digit1 digit

    2 digits4 digits8 digits*

    * = not applicable to verified balances** = only external calibration is possible for verified balances

    Factor

    ysetting

    Factor

    ysetting

  • 7/25/2019 MAN-ME_SE-e

    28/173

  • 7/25/2019 MAN-ME_SE-e

    29/173

    29

    Setting the Device Parameters (Device)

    PurposeThis menu item enables you toconfigure the balance, i.e., to meet

    individual requirements by selectingpredefined menu parameters in theSetup menu. You can block access tothe menu by assigning a password.

    FeaturesThe device parameters are combined inthe following groups(1st menu level):

    Draft shield

    Ionizer*

    Password

    User ID

    Clock

    Interfaces

    Display

    Keys

    Extra functions

    Factory settings: only device parameters

    Factory SettingsParameters: The factory settings areidentified by the symbol in the liststarting on page 31.

    * = not on ME36S, ME5, SE2

    PreparationDisplay available device parameters

    Select the Setup menu:press o

    > SETUP is displayed:

    Select Device parameters:use the q and O soft keys

    If no password has been assigned,anyone can access the Setup menudevice parameters

    If a password has already been assigned:

    > The password prompt is displayed

    $ If access is blocked by a password: enterthe password using the numeric and |or alphabetic keys

    $ If the last character of the passwordis a letter:conclude input by pressingthe a key

    Press l to confirm the password

    > Device parameters are now displayed:

    $ To select the next group:press the q soft key (down arrow)

    $ To select the previous menu itemof a group: press the Q soft key(up arrow)

    $ To select the next sub-item within

    a group: press the O soft key(right arrow)

    $ To select the previous group:press the o soft key (left arrow)

    $ Press l to confirm the selectedmenu item

    Entering or Changing a Password

    Lets assume that a password with 8

    characters max. has already beenassigned to access the Setup deviceparameters

    Select the Setup menu:press o

    > SETUP is displayed

    Select device parameters:Use the q and O soft keys

    If you have already assigneda password:

    > The password prompt is displayed:

    $ Enter the password

    $ Press the l soft key to confirmyour password and view the deviceparameters

    Write down your password here for easyreference:

    Password = .............................

    If you assign a password and then for-get what the word is:

    $ Enter the General Password(see Appendix)

    $ Press the l soft key to confirm and dis-play the password

    > The device parameters are displayed

    Select the device parameter Password:If necessary, repeatedly press q or Qand O, until you see

    > Password: and any existing password

  • 7/25/2019 MAN-ME_SE-e

    30/173

    30

    New password: Enter the numbersand/or letters for the new password(8 characters max.)

    If none is displayed, this meansno password has been assignedTo delete the user password:Press . or c and confirm

    To confirm:press the l soft key

    Exit the Setup menu:press the oo soft key

    > Restart the application

    Extra Functions

    Exit the Setup menu:

    press the oo soft key

    > Restart the application

    Print the parameter settings: If the device parameters

    are displayed: press r

    > Printout (example)

    --------------------SETUP

    DEVICE--------------------Draft shieldLeft/right keySame function

    Automatic modeOff

    Weight resolutionShow all decimal pla

    cesIonizerOnAuto-off time:

    10 secUser IDUser ID:

    InterfacesSerial communicatiSBIBaudrate

    1200 baudNumber of data b

    7 data bitsParity

    OddNumber of stop b

    1 stop bitHandshake mode

    Hardware handshakeafter 1 char

    Serial printer (PR

    YDP03Baudrate1200 baud

    ParityOdd

    Handshake modeHardware handshake

    after 1 charFunction: external

    Print keyFunction: control

    OutputDisplayContrast

    2

    etc.

  • 7/25/2019 MAN-ME_SE-e

    31/173

    31

    Device Parameters (Overview) factory setting user-defined setting(s)

    Setup Deviceparameters

    Draft shield Keys Same functionLeft/right key Separate function

    Off

    Automatic mode OffClose >function > openClose >

    perform function

    Weight Show all decimal placesresolution Do not show last decimal placewith door open

    Ionizer* Off On Auto-off time: 10 s 1 .. 99 seconds

    Password: None Enter 8 charactersmax.

    User ID: None Enter 20 charactersmax.

    Clock Time: Enter

    hh.mm.ss

    Date: Enter dd.mm.yyor mm.dd.yy (U.S. mode only)(01.01.97)

    Interfaces Serial SBI Baud rate 150 baudcommunication 300 baud(PERIPHERALS) 600 baud

    1200 baud2400 baud4800 baud9600 baud19,200 baud

    Number of data bits 7 bit 1)

    8 bit

    Parity Space 2)

    o OddEvenNone3)

    Number of stop bits 1 stop bit2 stop bits

    Handshake mode Software handshake Hardware handshake,

    1 character after CTSxBPI Network address: Enter any number from 0 to 31

    For the display, keys and extra functions, see next pages

    * = not on ME36S, ME5, SE21) not if None parity is selected2) only if 7 data bits selected3) only if 8 data bits selected

    Factor

    ysetting

    Factor

    ysetting

    Factor

    ysetting

    Factor

    ysetting

  • 7/25/2019 MAN-ME_SE-e

    32/173

    32

    Setup Deviceparameters

    Interfaces Serial YDP01ISprinter(PRINTER) YDP02 see YDP03 (not with 19,200 baud)

    YDP03 Baud rate 1200 baud2400 baud4800 baud9600 baud19,200 baud

    Parity Space Odd

    Even

    Handshake Software handshakemode Hardware handshake,

    1 character after CTS

    YDP01IS Label(label printer)

    YDP02IS

    YDP02IS Label(label printer)

    Universal Baud rate 150 baud300 baud

    600 baud1200 baud2400 baud4800 baud

    9600 baud19,200 baud

    Number of data bits 7 bits 1)

    8 bits

    Parity OddEven

    None 2)

    Number of stop bits 1 stop bit

    2 stop bits

    Handshake Software handshake modeHardware handshake,1 character after CTS

    YDP04IS

    YDP04IS Label

    For the display, keys and extra functions, see next page

    1) not if None parity is selected2) only if 8 data bits selected

    Factor

    ysetting

    Factor

    ysetting

  • 7/25/2019 MAN-ME_SE-e

    33/173

    33

    Setup Deviceparameters

    Interfaces Function: Print key rexternal switch Tare key

    Calibration keyF1 function key(Right soft key)Clear function key cF2 function key(2nd soft key from the right)Bar code scanner/extra keyboardIonizer keyRight draft shield keyLeft draft shield key

    Function: control ports Input Output

    Display Contrast Select contrast (0 to 4): 01

    o 234

    Background WhiteBlack

    Digit size 10mm + bar graph + text display13mm + bar graph

    13mm + text display13mm

    Application Offsymbols On

    Keys CF function Clear all applicationsin application Clear only selected applications

    CF function Delete entire inputfor input Delete last character

    Block key All keys unblockedfunctions All blocked except

    for o, e

    Alphanumerickeys blocked

    Extra functions Acoustic signal OnOff

    Power-on Off/on/standbymode Auto on

    Factory settings: Noonly Yesdevice parameters

    Factor

    ysetting

    Factor

    ysetting

  • 7/25/2019 MAN-ME_SE-e

    34/173

  • 7/25/2019 MAN-ME_SE-e

    35/173

    35

    Application Parameters (Overview) factory settingsuser-defined setting(s)

    Setup Application 1 Off (weighing only)Application (Basic settings)parameters Toggle wt. units Weight unit 1 Grams /g

    Kilograms /kg*Carats /ctPounds /lb*Ounces /oz*Troy ounces /ozt*Hong Kong taels /tlh*Singapore taels /tls*Taiwanese taels /tlt*Grains /GN*Pennyweights /dwt*

    # Milligrams /mgParts per pound //lb*Chinese taels /tlc*Momme /mom*Austrian carats /K*Tola /tol*Baht /bat*Mesghal /MS*

    Display accuracy 1 All digitsFewer for weightchangesLast digit off*

    Weight unit 2 # Grams /gKilograms /kg*Carat s/ctPounds /lb*Ounces /oz*Troy ounces /ozt*Hong Kong taels /tlh*Singapore taels /tls*Taiwanese taels /tlt*Grains /GN*

    Pennyweights /dwt* Milligrams /mgParts per pound //lb*Chinese taels /tlc*Momme /mom*Austrian carats /K*Tola /tol*Baht /bat*Mesghal /MS*

    Display accuracy 2 All digitsFewer digits forweight changeLast digit off*

    For counting through air buoyancy correction, see the following pages

    Applications 2 and 3 see following pages

    * not applicable to verified balances# = factory setting on ME36S, ME5, SE2

    Factor

    ysetting

    Factor

    ysetting

  • 7/25/2019 MAN-ME_SE-e

    36/173

    36

    Setup Application 1 Counting Accuracy Average Display accuracyApplication (Basic settings) piece weight calculation + 1 decimal placeparameters (resolution) + 2 decimal places

    Average piece Off weight updating Manual

    Automatic

    Percent weighing Weight storage accuracy Display accuracy(resolution) + 1 decimal places

    + 2 decimal places

    Decimal places for Nonereadout in percent 1 decimal place

    2 decimal places3 decimal places4 decimal places5 decimal places6 decimal places7 decimal places

    Display calculated value ResidueLossRatio 1 (DR)Ratio 2 (OR)

    Animal weighing* Animal activity Calm(averaging) Normal

    Active0.1% of the animal/object0.2% of the animal/object0.5% of the animal/object1% of the animal/object2% of the animal/object

    5% of the animal/object10% of the animal/object20% of the animal/object50% of the animal/object100% of the animal/object

    Start Manual mode Automatic mode

    Minimum load for automatic Nonestorage 10 digits

    20 digits50 digits

    100 digits200 digits500 digits

    1000 digits

    Decimal places in Noneresult display 1 decimal place

    2 decimal place3 decimal places4 decimal places5 decimal places6 decimal places7 decimal places

    Printout None Average weight only

    Average and calculated value

    Recalculation*

    For calculation through air buoyancy correction, see following pages

    Applications 2 and 3 see following pages

    * = How to run this application is described in detail in our Masterpro LA. Installation and Operating Instructions.Request your copy directly from Sartorius or download it from the Internet (www.sartorius.com; see download)

    Factor

    ysetting

  • 7/25/2019 MAN-ME_SE-e

    37/173

    37

    Setup Application 1 Calculation Decimal places in NoneApplication (Basic settings) calculated result 1 decimal placeparameters 2 decimal places

    3 decimal places4 decimal places5 decimal places6 decimal places7 decimal places

    Density Method Density of liquid Buoyancy

    DisplacementPycnometer

    Liquid causing Waterbuoyancy Ethanol

    User-definable

    Decimal places Nonefor disp. of vol./ 1 decimal placedensity 2 decimal places

    3 decimal places4 decimal places

    Printout NoneAll data

    Differential weighing Weighing sequence* Individual weighingConsecutive individual weighing

    Combined weighingSerial weighing

    Tare weighing No Yes

    Result with Nonedecimal point 1 decimal place

    2 decimal places3 decimal places4 decimal places5 decimal places6 decimal places7 decimal places

    Autosave Offvalues On, first value at stability

    On, last value at stability

    On, value bet.70 130% at stabil.

    Minimum load Nonefor autosave 10 digits

    20 digits50 digits100 digits200 digits500 digits1000 digits

    Save statistics NoYes

    Air buoyancy correction: See next page

    Applications 2 and 3: see following pages

    * = Setting can only be changed when the program is initially run and when the Wg.seq. key option is set to No

    Factor

    ysetting

  • 7/25/2019 MAN-ME_SE-e

    38/173

    38

    Setup Application 1 Differential weighing Generate printout NoneApplication (Basic settings) Automatic after backweighingparameters Auto. after initial weigh.

    and backweighingAutomatic after tare-, initial-and backweigh.

    Include sample ID in text line NoYes

    Wg.seq key No Yes

    Clear sample after individual Noweight, result + and unload Yes

    Last residual weight is stored No

    as initial weight (ashing) Yes

    Air buoyancy correction Air density determination NoYes

    Change steel/aluminum Noreferences Yes

    Diameter determination

    Application 2 Off(Control functions) Checkweighing* Activation of port lines Within checkweighing range

    Always onStability andcheckweighing rangeAt stabilityStability + checkweigh.range on

    Type of checkweighing input Target, min., max. weightMinimum, maximum weightTarget, min. in %, max. in %

    Weight display mode Absolute valueDifference from the target

    Automatic printout Yesof OK values No

    Time-controlled Function after time interval Beep (acoustic signal)functions Lock in readout Automatic printout of values

    Store value in applicat. 3 memory(totalizing, formulation, statistics)

    Automatic function restart OnOff

    Storage mode Without stabilityAfter stabilityAfter higher stability

    Print then tare Off On

    * = How to run this application is described in detail in our Masterpro LA. Installation and Operating Instructions.Request your copy directly from Sartorius or download it from the Internet (www.sartorius.com; see download)

    Factor

    ysetting

    Factor

    ysetting

  • 7/25/2019 MAN-ME_SE-e

    39/173

    39

    Setup Application 3 OffApplication (Data records)parameters Totalizing* Automatic storage Off

    On, first value at stabilityOn, last value at stabilityOn, value bet. 70% 130%at stability

    Minimum load for Noneautomatic storage 10 digits

    20 digits50 digits100 digits200 digits500 digits1000 digits

    Source of data for auto storage Application 1

    Application 2

    Evaluated values NetCalculatedNet + calculated

    Evaluation mode, MR function Intermediate evaluation, printFinal evaluation, printIntermediate evaluation,display+printFinal evaluation,display+print

    M+/M function, then tare OffOn

    Printout of individual Nocomponents Yes

    Formulation* Automatic storage OffOn, first value at stability

    Minimum load for Noneautomatic storage 10 digits

    20 digits50 digits100 digits200 digits500 digits

    1000 digits

    Source of data for Application 1automatic storage Application 2

    Evaluated values NetCalculatedNet + calculated

    Evaluation mode,MR function Intermediate evaluation, printFinal evaluation, print

    Printout of individual Nocomponents Yes

    * = How to run this application is described in detail in our Masterpro LA. Installation and Operating Instructions.Request your copy directly from Sartorius or download it from the Internet (www.sartorius.com; see download)

    Factor

    ysetting

    Factor

    ysetting

  • 7/25/2019 MAN-ME_SE-e

    40/173

    40

    Setup Application 3 Statistics Automatic storage o Off Application (Data records) On, first value at stabilityparameters On, last value at stability

    On, value 70% 130%at stability

    Minimum load for Noneautomatic storage 10 digits

    o 20 digits50 digits100 digits200 digits500 digits1000 digits

    Source of data for o Application 1automatic storage Application 2

    Evaluated values o NetCalculatedNet + calculated

    Evaluation mode, o Intermediate evaluation, printMR function Final evaluation, print

    Intermediate evaluation,display + printFinal evaluation,display + print

    M+/M function, then tare o OffOn

    Printout of individual Nocomponents o Yes

    Extra Functions for other extra functions, see next page

    Factor

    ysetting

    Factor

    ysetting

  • 7/25/2019 MAN-ME_SE-e

    41/173

    41

    Setup Extra Functions OffApplication F4 keyparameters Extra Functions 2nd tare memory Container tare weight No

    F5 key Yes

    Automatic printout Net valueTare/preset tare

    Off

    Identification Printout Automatic, if configuredcodes Once after pressing r(IDs) if configured

    Each time the print key(r) is pressedOnce for M+ function(application 3 memory)

    Manual storagein app. 3memory M+ (totalizing,formulation, statistics)

    Changing the resolution

    Product data memory

    SQmin Function Display o Text displayBargraph

    Print in o OffGLP Header On

    DKD uncertainty Display

    1)

    U (absolute)of measurement U* (relative)PG (process accuracy)

    Print1) U (absolute)U* (relative)PA (process accuracy)

    Process accuracy Input: 1100,5 decimal places

    Auto-start Onapplication when Offpower goes on

    Factory settings Noonly for Yesapplication-parameters

    1) an asterisk (*) indicates an activated menu item. You can select up to 3 items.

    Factor

    ysetting

  • 7/25/2019 MAN-ME_SE-e

    42/173

    Selecting the Printout Function (Printout)

    42

    PurposeThis menu item enables you toconfigure the printout to meet your

    individual requirements by selectingpredefined menu parameters in theSetup menu. Printouts of weights andother measured or calculated values andIDs enable you to document your data.You can select the particular data youwish to print. To prevent changes toyour settings, you can block access tothe menu by assigning a password.

    FeaturesThe device parameters are combined inthe following groups(1st menu level):

    Application-defined output Automatic output of displayed values Output to interface ports Line format ISO/GLP printout Identification # (identifier) Factory settings only printout

    Factory SettingsParameters: The factory settingsare identified by the symbol inthe list on the next page.

    PreparationDisplay available printout parameters

    Select the Setup menu:press o

    > SETUP is displayed:

    Select Printout:use the q and O soft keysIf no password has been assigned,anyone can access the printoutparameters in the Setup menu

    If a password has already been assigned:> The password prompt is displayed

    $ If access is blocked by a password: enterthe password using the numeric and/oralphabetic keys

    $ If the last character of the passwordis a letter: conclude input by pressingthe a key

    Press l to confirm the password

    > Printout parameters are now displayed:

    $ To select the next group:press the q soft key (down arrow)

    $ To select the previous item ofa group: press the Q soft key(up arrow)

    $ To select the next sub-item withina group: press the O soft key (rightarrow)

    $ To select the previous group:press the o soft key (left arrow)

    $ To confirm: press the l soft key

    Extra Functions

    Exit the Setup menu:

    press the oo soft key> Restart your application

    Print parameter settings: When the printout parameters

    are displayed, press r

    > Printout (Example)

    --------------------SETUP

    PRINTOUT--------------------Application-definedStability paramete

    With stabilityPrint on request t

    OffAuto print upon in

    All valuesConfigured printouIndiv.: Printout

    Automatic output ofStability parameteWitout stability

    Stop auto printNot possible

    Time-dependent aut1 display update

    Output to interfaceSerial communicati

    Application-definedoutput

    Serial printer (PRApplication-defined

    outputLine format

    For other apps/GLP (22 characters)

    ISO/GLP/GMP printouOff

    Identification #Lot (L ID):

    ID1:ID1

    etc.

  • 7/25/2019 MAN-ME_SE-e

    43/173

    43

    Printout Parameters (Overview) factory settinguser-defined setting(s)

    SETUP Application- Stability parameter Without stability 1)

    Printout defined output o With stability

    Print on request, then tare o OffOn

    Auto print upon Off initialization o All values

    (of the application) Only main values

    Configured printout See own chapter startingon the next page

    Automatic Stability parameter o Without stabilityoutput At stabilityof displayed Automatic aftervalue weight change 2)

    Stop auto print Use print key ro Not possible

    Time-dependent o 1 display updateauto print 2 display updates

    10 display updates *100 display updates *

    Output to Serial communication Offinterface ports (PERIPHERALS) o Application-defined output

    Automatic output of displayed value

    Serial printer) Off (PRINTER) o Application-defined output

    Line format For raw data (16 characters)o For other apps/

    GLP (22 characters)

    ISO/GLP o Offprintout Only for calibration/adjustmentAlways on

    Identification # Lot ID (L ID): Enter 20 characters max.ID1: ID1ID2: ID2ID3: ID3ID4: ID4

    Factory settings: Noprintout only Yes

    * = changing settings not applicable to verified balances1)

    = Information on use in legal metrology: Only permitted for control purposes; printouts are not allowed2) = auto print when load change is > 10 d and stability is reached: no printout until residual difference in load value is < 5 d

    Factor

    ysetting

    Factor

    ysetting

  • 7/25/2019 MAN-ME_SE-e

    44/173

    Printout Configuration

    44

    PurposeThis menu item enables you toconfigure individual printout formats.

    With the formulation, totalizing andstatistics application, you can alsodefine the values to be included onthe total printout when the MR keyis pressed.

    Under Setup > Printout > Application-defined output > Configured printout,you can configure individual, compo-nent or total data records that containthe items in each application that areavailable for printouts. Configure theseprintouts after you have configured theapplications, because some entries inthe data record depend on the particu-lar application.

    Features Maximum items in a data record: 60

    Separate configuration of printoutformats for individual weights,components, total, backweighing andstatistics

    Individual printout generation:press the r key

    Automatic printout of application data:e.g., results from animal weighing ordensity application (Setup menu: Appli-cation 1: Density: Printout: All data) OKvalues from checkweighing application,time-controlled printouts, 2nd tarememory

    Component printout:For results from totalizing, formulationor statistics applications, pressM+ or M- (Setup: Application 3: ...,Printout of individual components: On)

    Total printout: For totalizing,formulation or statistics applications,press MR

    Backweighing printouts or records:automatically generated after back-weighing or manually by pressing ther

    key when the result is displayedat the end of backweighing

    Statistics printout or output:To generate, press the r key whenthe statistics are displayed

    Printouts for Differential Weighing:These printouts can be generated asstandard or configured (user-defined)reports.

    You can configure the followingprintouts:

    Individual printout

    Backweighing printout

    Statistics printouts

    Printouts are generated in one oftwo ways:

    at the request of the user by pressingthe r key (print on request)

    automatically, if configuredin the Setup menu [Applicationparameters: Application 1: Differentialweighing: Generate printout: Auto]

    You can turn off automatic printoutgeneration in the Setup menu[Application parameters:

    Application 1: Differential weighing:Generate printout: None]

    Data records are deleted after you havedswitched to a different application oractivated or de-activated an extrafunction in the application parametersof the Setup menu

    A new pick list for a data record iscreated based on the currently activeapplication programs and extrafunctions

    Printout items can be deletedindividually

    No printout is generated when thefollowing setting is configured:Setup: Printout: Line format:For raw data (16 characters)

    Print item Form feed for footer:Advance to beginning of next label inthe YDP01IS-Label and YDP02IS-Label [printer] interface mode

    Extra Functions

    Exit printout configuration:press oo soft key

    > Restart application

    Printing Select and List Settings

    LIST: print the currently selected listSELECT: printout items that can stillbe selected

    When the select bar is onLIST or SELECT:press the r key

    > Printout (Example)

    BACKW. PRINT.LIST====================Sample dateNet initial wt.Backweighed resLoss in %====================etc.

  • 7/25/2019 MAN-ME_SE-e

    45/173

    45

    Example:Configure an Individual Printout for Counting Application to Include Dotted Line, Date/Time, Piece Count and Net Weight

    Settings (changes in the factory settings required for this example):Setup: Application parameters: Application 1: CountingExit the Setup menu: press the oo soft keyThen call Setup again: Printout: Application-defined output: Configured printout

    Step Press key(s) (or follow instructions) Display/Output

    1. Select Setup menu, o, thenthen Printout q repeatedly and

    O soft key

    2. Confirm Application-defined O soft keyoutput

    3. Select and confirm q soft key 3x andConfigured printout O soft key

    4. Confirm Indiv. printout O soft key

    5. Select Blank line O, q, l soft keys

    6. Select Date/time q soft key twice, thenl soft key

    7. Select Piece count q soft key repeatedly,then l soft key

    8. Select net weight Q soft key repeatedly,then l soft key

    9. Exit Printout configuration oo soft key

    10. Perform weighing operations, rthen print

    --------------------14.01.2000 09:19Qnt + 598 pcs

    N + 2003.13 g

  • 7/25/2019 MAN-ME_SE-e

    46/173

    Device Information

    46

    PurposeThis menu item enables you to haveinformation displayed about the specific

    balance (device).

    Display Device Information onME215/235/254/414/614

    Select the Setup menu:press the o key

    > SETUP is displayed:

    Select Info:Repeatedly press the q soft key, thenpress the O soft key

    Select Device information:press the O soft key

    > Device information is displayed:

    Display Device Informationon ME36S, ME5 or SE2

    Select Device Information:Press the J key

    > Device information is displayed

    oo o q

    SETUP INFO

    Version no: 01-41-05

    Wgh.sys. ver. #: 00-21-09

    Draft sh. ver.#: 05-01-03

    Model: ME215S

    Serial no: 91205355

    Print device information:Press the r key

    > Printout (Example)

    --------------------23.12.2001 13:02Model ME215SSer. no. 91205355Ver. no. 01-41-05(Version of the operating program)ID BECKER123(User ID)--------------------L ID LOT 23(Lot ID)SETUP

    INFO

    --------------------Version no.:

    01-41-05(Version of the operating program)Wgh. sys. ver. #:

    00-21-09(Version no. of the weighing cell)Dft. shield v. no.:

    05-01-03(Program version no. of thedraft shield)Model:

    ME215SSerial no.:

    91205355

    Next mainten.:01.01.2003

    Service phone:00495513080

    SQmin:0.0300 g

    --------------------

    Return to SETUP overview:press the o soft key

    Exit Setup menu:Press the oo soft key

    > Original settings are restored

    Factory SettingsEach parameter category has a factory

    setting. In the Setup menu, you canrestore all factory settings by confirmingthe selection YES.

    The following settings are not restored: Language Password Display contrast Time (clock)

  • 7/25/2019 MAN-ME_SE-e

    47/173

    47

    Basic Weighing Function

    PurposeThe basic weighing function is alwaysaccessible and can be used aloneor in combination with an applicationprogram (Toggle between Weight Units,Counting, Weighing in Percent, etc.).

    Features Taring the balance Assigning IDs to weights Printing weights Printing ID codes for weights

    Soft Key FunctionsCal Start calibration/adjustmentisoCAL

    Press when necessary tostart calibration andadjustment

    S ID Stored ID entered

    General Instructionsfor Analytical Weighing

    Handling Samples and ContainersThe sample should be conditionedto the temperature inside the balance.This is the only way to avoid errorscaused by air buoyancy and deviations

    caused by convection currents at thesurface of the sample.

    Since these effects increase proportion-ally to the volume and surface of thesample, make sure that the size of thetare vessel selected is appropriate forthe initial sample.

    Never use your bare hands to touchsamples to be weighed. In addition tothe effect of the temperature, theextremely hygroscopic behaviorof fingerprints left on the sample willotherwise cause considerable inter-

    ference during weight measurement.

    Use forceps or a similar utensil that isappropriate to place your sample care-fully on the pan. Working with yourbalance requires a steady hand and asmooth, uninterrupted technique.

    If the weighing chamber has not beenopened for a relatively long period,it may have a temperature differentfrom that of the balances surroundingenvironment. When you open theweighing chamber, a change in temp-erature will inevitably occur, due to the

    laws of physics, and may show up asa change in the weight readout.

    Operating the Balance

    Therefore, we recommend that beforeyou begin the actual weighing series youopen and close the weighing chamber at

    the same rate as you will be doing duringweighing. After the weighing chamberhas been closed, the weight readout willusually stabilize after about 8 seconds.The accuracy of the weight readouts willincrease as you continue weighing withgreater consistency.

    Weighing Electrostatically ChargedSamples and ContainersMajor measuring errors can occur whenelectrostatically charged samples andcontainers are weighed. This problemparticularly involves samples that haveextremely poor conductivity (glass, plastic,filters) since they can discharge electro-static i.e., friction-induced charges onlyover a relatively long period of time. Theresult is an interaction of forces among thecharges adhering to the sample and thestationary components of the balance(weighing chamber base plate, draft shieldconstruction, balance housing). This isnoticeable when the weight readout drifts.At high humidity, this effect is notvery pronounced or may not occur at all,due to the thin layer of water thatcondenses on the sample and, through

    conductive discharge, counteractsinterfering static electricity.

    In addition to taking purely mechanicalcounteractive measures (protecting thesample using a special antistatic weighingpan see the Accessories), you canneutralize the surface charges bybombarding them with ions of theopposite polarity (see page 53 forinstructions on activating the ionizer).This is a highly effective procedurefor eliminating static electricity.

    The balances environment, including the

    operator, can considerably interfere withweighing results, due to static electricity.The balances of the ME/SE series havebeen designed to counteract this phenom-enon: the glass surfaces of the draft shieldhave a special metallic coating.

    The rear panel of the balance has aterminal for connecting an equipotentialgrounding conductor. It is used foradditional grounding of a peripheraldevice (for example, a vibrating spatula).This terminal is designed for singlegrounding wires up to .25 standardgauge or 6 mm2 stranded wires and for .18

    standard gauge or 4mm2

    stranded wires.

    Weighing Magnetic or MagnetizableSamples

    It is technically impossible to avoid the useof magnetizable materials in the manufac-ture of balances. Ultimately, the operatingprinciple of high-resolution balances isbased on electromagnetic force compensa-tion of the load placed on the weighingpan.

    When magnetic or magnetizable samplesor containers (e.g., a beaker with a stirrer)are weighed, interactions among theabove-mentioned components of thebalance may occur, distorting weightreadouts.

    Unlike deviations caused by electrostaticcharges, magnetic interference is usuallyconstant over time. However, it is sensitiveto and dependent on the position of thesample container on the weighing pan andis also characterized by poor repeatability.

    To reduce the effect described above,we recommend increasing the distancebetween the sample and the weighing panby inserting a non-magnetizable materialbetween them (the reduction in force isproportional to the quadrate of the distance).In special cases, soft-magnetic plates

    should be used to shield against interferingmagnetic effects.

    In the presence of extremely strongmagnetic fields for instance, whenmeasuring the susceptibility of a samplein an electromagnet you should usethe below-balance weighing port, whichcomes standard on your balance.

  • 7/25/2019 MAN-ME_SE-e

    48/173

  • 7/25/2019 MAN-ME_SE-e

    49/173

  • 7/25/2019 MAN-ME_SE-e

    50/173

    50

    Example W2

    Enter ABC123 as the sample ID

    Note: The sample ID generally applies to one weighing operation only The ID is deleted after data output

    Step Press key(s) (or follow instructions) Display/Output

    Initial state (balance unloaded)(ID can also be entered whilebalance is loaded)

    1. Select alphabetic input a

    2. Select the required letter group A ABCDEF soft key

    3. Enter the letter A A soft key

    (To delete a letter:c

    ,c

    )

    4. Select the letter group and ABCDEF soft keyenter B B soft key

    5. Select the letter group and ABCDEF soft keyenter C C soft key

    (If you enter only letters,conclude input a)

    6. Enter the numbers 1, 2 and 3 1 2 3

    7. Store the ID S ID soft key(20 characters max.)

    The next printout will include

    the sample ID

    g

    Max 210 g

    0%

    d=0.01mg

    100%

    000000SCal

    Max 210 g

    0% AH

    L

    S

    T

    d=0.01mg

    100%

    FS

    as

    ABCDEFGHIJKLMNOPQRSTUVWXYZ/=-?:#*"&U

    OOOg000000

    Max 210 g

    0% A

    H

    L

    S

    T

    d=0.01mg

    100%

    FS

    as

    A B C D E F

    OOOg000000

    Max 210 g

    0% A

    HL

    S

    T

    d=0.01mg

    100%

    FS

    as

    ABCDEFGHIJKLMNOPQRSTUVWXYZ/=-?:#*"&U

    A

    Max 210 g

    0% A

    H

    L

    S

    T

    d=0.01mg

    100%

    F

    S

    a

    s

    ABCDEFGHIJKLMNOPQRSTUVWXYZ/=-?:#*"&U

    AB

    Max 210 g

    0% A

    H

    LS

    T

    d=0.01mg

    100%

    F

    S

    a

    s

    ABCDEFGHIJKLMNOPQRSTUVWXYZ/=-?:#*"&UABC

    Max 210 g

    0% A

    H

    L

    S

    T

    d=0.01mg

    100%

    FS

    as

    S ID

    ABC123

    g

    Max 210 g

    0%

    d=0.01mg

    100%

    000000

    S

    Cal

  • 7/25/2019 MAN-ME_SE-e

    51/173

    51

    Device Parameters

    This function can be deactivated.If not, the draft shield doors will closeautomatically if the balance has not

    been used for 2 minutes (protectschamber from dust).

    The function Close draft shield auto-matically when function is activatedcan be combined with functions andapplications that require the withstability parameter in order for weightsto be accepted: Turn on the balance (tare when the

    power goes on) Tare after stability Print on request after stability Start all adjustment functions 2nd tare memory Manual weight storage mode for the

    following functions: Counting, weighing in percent,

    recalculation, density determination,differential weighing

    Checkweighing, time-controlledfunctions with storage mode

    Totalizing, formulation, statistics

    A lower weight resolution is possiblewhen the draft shield doors are open.

    The left and right K keys for operatingthe draft shield doors can have the same function have separate functions be switched off.

    Draft Shield onME215|235|414|614|36S:

    You can define which draft shielddoor(s) will open and close when youpress the left or right K key (the Geniusbalance has a self-teaching capability)

    If a door encounters an obstacle whilemoving, the following will happen: While opening: the door will stop While closing: the door will re-open

    Factory Settings of the ParametersLeft/right draft shield keys:Same function

    Automatic mode: Off

    Weight resolution when door is open:Show all decimal

    places

    Opening and Closingthe Draft Shield

    PurposeThe ME/SE balance is a high-resolutionweighing instrument. It has a draftshield so that convection currents can-not affect the weighing result. To load asample on and remove it from theweighing pan, a draft shield door mustbe opened and closed. You can do thisin various ways, depending on themenu setting you select.

    Features The draft shield doors can be opened

    and closed at any time, regardless of the

    application used

    The draft shield doors can be openedand closed by pressing the respectivekeys, by activating an external switch orby sending a command to the balancesinterface port

    The draft shield doors can be set so thatthey operate automatically in one of thefollowing ways when a specific balancefunction is performed, such as taring(zeroing): Close automatically Close automatically, then open

    Draft Shield on ME5 | SE2

    You can define the function of the draft shield

    Setup, draft shield keys:Keys Same function Different (separate) function

    O , P key Opens according to Opens 100 clockwise predefined Closes opening position Numeric input of the aperture Closes

    Numeric keys + O , P key Enter and save value for No function aperture 44 181: Opens counter- clockwise 182 316: Opens clockwise 0 43: Deletes stored value

    Numeric keys + Counterclockwise opening No functionP key aperture between 46 and 180

    Learning mode Yes; define desired opening Noposition manually

  • 7/25/2019 MAN-ME_SE-e

    52/173

  • 7/25/2019 MAN-ME_SE-e

    53/173

    53

    Interfaces

    Purpose

    This item enables you to set theparameters for the following interfaces:

    Serial communications port Serial printer port External switch function Control port function

    Serial Communications PortYou can set the serial communicationsport to use for the following modes:

    SBI XBPI Sartonet

    Serial Printer PortYou can set the serial printer port to usefor the following printers:

    YDPO1IS YDP02 YDP03 YDPO1IS Label YDPO2IS YDP02IS Label Universal YDP04IS YDP04IS Label

    External Universal Remote SwitchYou can connect an external universal

    remote switch (foot switch or bar codescanner or an external keyboard) to oneof the two serial ports. Then you canassign one of the following functions tobe performed whenthe switch is activated:

    Print key

    Tare key

    Cal key

    F1 function key

    CF key

    F2 function key

    Bar code scanner/extrakeyboard (requires a specialconnecting cable)

    Ionizer key

    Right draft shield key

    Left draft shield key

    ME215/235/414/614 Models:Static Electricity Eliminator(Ionizer) J

    PurposeIonization of air (i.e., charging the airwith ions of both positive and negativecharges) causes the air surrounding asample to become electrically conductive.Charges are neutralized in the air streamor grounded.

    Be especially careful when weighingelectrostatically charged samples.

    Features The ionizer can be turned on and off

    independently of the position of thedraft shield doors

    When the draft shield doors are open,the static electricity eliminator (ionizer)will shut off after the preset time haselapsed. However, the ionizing time willstart over if the static electricityeliminator is still running when a doorcloses.

    The static electricity eliminator (ionizer)can be configured in the Setup menu sothat it will stay on only for a certaintime.

    When the ionizer is active, this isindicated by the flashing symbol J(flashes from the outside toward theinside and vice versa)

    Factory Setting for the Parameters

    Static electricity eliminator (ionizer):On; Auto off after 10 sec.

    Configuring the Static Electricity Elimi-nator (Ionizer)

    Ionizer Off

    On Auto-off time:1 to 99 seconds

    Function keyJ Ionizer key

    PasswordYou can enter a password to block accessto the menu parameter settings and to ID

    code and exact calibration weight inputs.See the detailed description in the chapteron Configuring the Balance: Setting theDevice Parameters.

    User IDYou can enter your own personal password(20 characters max.)

    ClockISO/GLP printouts in particular must begenerated with the date and time stampof the specific measurement. This date andtime stamp is optional on other printouts.

    See the chapter on Configuring theBalance: Setting the Device Parameters.

  • 7/25/2019 MAN-ME_SE-e

    54/173

    54

    KeysYou can assign different functions tothe c key for deleting input and

    applications.

    When you delete applications, you candelete either the data stored for allapplications or just selected data.CF function

    in application

    When you delete input, you can eitherdelete all the data input in a field, oronly the last character entered.CF function for inputs

    You can block key functions; you canchoose whether to block all keys (excepte, o, draft shield left/right andionizer) or just the alphanumeric keys.Block key functions

    Extra Functions

    Acoustic SignalAn acoustic signal is emitted whenyou press a key. When the key pressed isallowed, the signal is a single beep-tone; when it is not allowed, this is sig-naled by a double-beep (key does notinitiate a function). In the Setup menu,you can configure whether

    the acoustic signal should sound (On

    ) the acoustic signal should not sound(Off)

    Power-On ModeYou can configure the balance so thatwhen a power supply is connected,

    the balance is off(Off/on/standby orOff/on)

    the balance switches on automatically(Auto on)

    You can also set the configurations so

    that when the balance is turned offafter use, it is off (Off/on)

    in the standby mode(Off/on/standby)

    After you turn on the balance,a self-test of the functions is run(TEST is displayed in the text line;the bar graph is shown)

    Control Port FunctionYou can connect either a checkweighingdisplay or an external universal switch to

    the serial communications port on thebalance (factory setting).

    To do so, you need to configure theinterface for input or output.

    Pin Assignment Chart of the FemaleInterface Connector

    Pin Function: Input15 r key; see

    Universal switch16 Left K key17 Soft key 6 (Cal)fgh Q18 Soft key 1 (F)19 w key

    Pin Function: Output15 External switch

    (see above)16 Control port 1: lighter17 Control port 2: equal18 Control port 3: heavier19 Control port 4: set

    For further information on thepin assignment chart, see the section onPin Assignment Charts in the chapter