+ All Categories
Home > Documents > man2550_23_04_2009

man2550_23_04_2009

Date post: 04-Apr-2018
Category:
Upload: hafisj
View: 214 times
Download: 0 times
Share this document with a friend

of 64

Transcript
  • 7/29/2019 man2550_23_04_2009

    1/64

    1

    www.hannainst .com

    MultiparameterpH/ORP/C

    EC/TDS/NaClBench Meter

    HI 2550

    Instruction Manual

  • 7/29/2019 man2550_23_04_2009

    2/64

    2

    TABLE OF CONTENTSTABLE OF CONTENTSTABLE OF CONTENTSTABLE OF CONTENTSTABLE OF CONTENTS

    WARRANTY .................................................................................................... 2PRELIMINARY EXAMINATION ............................................................................ 3GENERAL DESCRIPTION.................................................................................... 3FUNCTIONAL DESCRIPTION .............................................................................. 4SPECIFICATIONS ............................................................................................. 5OPERATIONAL GUIDE ....................................................................................... 7AUTO-RANGING .............................................................................................12pH CALIBRATION .......... ............ ............ ............ ............ ............ ............ .......... 13pH BUFFER TEMPERATURE DEPENDENCE ........... ............ ........... ............ ........... 18RELATIVE mV CALIBRATION. ............ ............ ........... ............ ............ ............ ...... 19ECS/TDS CALIBRATION ............ ............ ............ ............ ............ ............ ............ 19CONDUCTIVITY VERSUS TEMPERATURE CHART .......... ........... ............ ........... ....... 21NaCl CALIBRATION ......................................................................................... 22GOOD LABORATORY PRACTICE (GLP) ........... ............ ............ ........... ............ ....... 23

    LOGGING FUNCTION........... ........... ............ ............ ........... ............ ............ ...... 28SETUP ..........................................................................................................36TEMPERATURE CALIBRATION (FOR TECHNICAL PERSONNEL ONLY) ............ ........... 42mV CALIBRATION (FOR TECHNICAL PERSONNEL ONLY)........ ............ ........... ........ 44PC INTERFACE ...............................................................................................45ELECTRODE CONDITIONING & MAINTENANCE ............ ............ ............ ............ .... 51TROUBLESHOOTING GUIDE. ............ ............ ........... ............ ............ ............ ...... 54TEMPERATURE CORRELATION FOR pH SENSITIVE GLASS ........... ............ ............ .. 55ACCESSORIES ...............................................................................................56

    WARRANTY WARRANTY WARRANTY WARRANTY WARRANTY

    HI 2550 is guaranteed for two years against defects in workmanship and

    materials when used for their intended purpose and maintained accordingto instructions. Electrodes and probes are guaranteed for six months.Thiswarranty is limited to repair or replacement free of charge.

    Damage due to accidents, misuse, tampering or lack of prescribedmaintenance is not covered.

    If service is required, contact the dealer from whom you purchased theinstrument. If under warranty, report the model number, date ofpurchase, serial number and the nature of the problem. If the repair isnot covered by the warranty, you will be notified of the charges incurred.If the instrument is to be returned to Hanna Instruments, first obtaina Returned Goods Authorization number from the Technical Servicedepartment and then send it with shipping costs prepaid. When shippingany instrument, make sure it is properly packed for complete protection.

    Dear Customer,Thank you for choosing a Hanna Instruments product.Please read this instruction manual carefully before using this instrument.This manual will provide you with the necessary information for correctuse of this instrument, as well as a precise idea of its versatility.If you need additional technical information, do not hesitate to e-mail usat [email protected].

  • 7/29/2019 man2550_23_04_2009

    3/64

    3

    GENERAL DESCRIPT IONGENERAL DESCRIPT IONGENERAL DESCRIPT IONGENERAL DESCRIPT IONGENERAL DESCRIPT ION

    The HI 2550 is a microprocessor based pH, ORP, Conductivity (EC), TDS,NaCl, and Temperature Bench Meter. Relative mV feature is alsoprovided.pH measurements are compensated for temperature effect manually orautomatically with the HI 7662 temperature probe.Up to a five-point pH calibration can be performed using seven standardbuffers and two custom buffers.

    A calibration due alarm can be set to alert the user that too much timehas elapsed since the last pH calibration.

    Conductivity and TDS features auto-ranging, which automatically selectsthe scale with the highest resolution. Conductivity measurements arecompensated for temperature manually or automatically with atemperature sensor located inside the probe. The temperature coefficientis user selectable. Temperature compensation can be disabled to measureactual conductivity.

    The GLP feature provides data consistency.

    Data can be stored in the meters memory for later retrieval. The metersmemory can hold 200 manually logged points and 500 lot loggingpoints.

    An USB connection ensures communication with a PC.

    PREL IMINARY EXAMINATIONPREL IMINARY EXAMINATIONPREL IMINARY EXAMINATIONPREL IMINARY EXAMINATIONPREL IMINARY EXAMINATIONRemove the instrument from the packing material and examine it carefullyto make sure that no damage has occurred during shipping. If there is anydamage, notify your Dealer or the nearest Hanna Customer Service Center.Each instrument is supplied with:

    HI 1131B Glass-body Combination pH Electrode with 1 m (3.3')Cable

    HI 76310 Conductivity / TDS probe HI 7662 Temperature Probe

    HI 76404N Electrode Holder

    pH 4.01 & 7.01 Buffer Solutions (20 mL each) HI 7071 Electrolyte Solution

    12VDC Power Adapter

    Instruction Manual

    Note: Save all packing material until you are sure that the instrumentfunctions correctly. All defective items must be returned in theoriginal packing with the supplied accessories.

  • 7/29/2019 man2550_23_04_2009

    4/64

    4

    1) Liquid Crystal Display (LCD).2) CAL key, to enter and exit calibration mode.

    RCL key (alternate function), to enter and exit memory recall.

    3) CFM/GLP key, to confirm calibration selection, different setup valuesor to display Good Laboratory Practice information.

    4) C key, to manually increase temperature value or other parameters.TC key (alternate function), to view temperature coefficient value.

    5) C key, to manually decrease temperature value or other parameters.ATC key (alternate function), to select EC temperature compensation mode.

    6) SETUP key, to enter/exit SETUP mode.LOCK key (alternate function), to freeze current EC range on the LCD.

    7) RANGE key, to select measurement range (pH, mV, EC), switch tofocused data in SETUP or toggle between buffer value and temperatureduring calibration.MODE key (alternate function) to select mV or Rel mV on mV rangeor EC, TDS, NaCl on EC range.

    8) LOG/CLR key, to store a value into memory, to clear pH calibration,

    or to select to delete log records or lots.9) ALT key, to select alternate function.10) ON/OFF switch.11) Power supply socket.12) USB connector.13) EC electrode connector.14) BNC electrode connector.15) Temperature probe socket.16) Electrode reference socket.

    FUNCTIONAL DESCRIPT IONFUNCTIONAL DESCRIPT IONFUNCTIONAL DESCRIPT IONFUNCTIONAL DESCRIPT IONFUNCTIONAL DESCRIPT ION

    Front Panel

    Rear Panel

  • 7/29/2019 man2550_23_04_2009

    5/64

    5

    SPECIF ICAT IONSSPECIF ICAT IONSSPECIF ICAT IONSSPECIF ICAT IONSSPECIF ICAT IONS

    EGNAR

    Hp0.61ot0.2Hp00.61ot00.2

    Hp000.61ot000.2

    )PRO&ESI(Vm9.999)PRO&ESI(Vm0002

    mc/S99.92ot00.0mc/S9.992ot0.03

    mc/S9992ot003mc/Sm99.92ot00.3mc/Sm0.002ot0.03

    detasnepmocnumc/Sm0.005otpu )*(

    ytivitcudnoc

    mpp99.41ot00.0mpp9.941ot0.51

    mpp9941ot051l/g99.41ot05.1l/g0.001ot0.51

    l/g0.004otpu detasnepmocnu )*( SDT)rotcaf08.0htiw(

    lCaN%0.004ot0.0

    )egnarCE,Hp(C0.021ot0.02

    NOITULOSER

    Hp1.0Hp10.0Hp100.0

    )Vm0.0001(Vm1.0)Vm0002(Vm1

    10.0 mc/S

    1.0 mc/S1 mc/S

    mc/Sm10.0mc/Sm1.0

    mpp10.0mpp1.0mpp1

    l/g10.0l/g1.0

    lCaN%1.0

    C1.0

    YCARUCCAF86/C02@

    Hp10.0Hp200.0

    )Vm9.999(Vm2.0)Vm0002(Vm1

    50.0(gnidaer%1 1romc/S retaergrevehcihw,tigid )

    (gnidaerfo%1 1rompp30.0)retaergrevehcihw,tigid

    gnidaerfo%1

    )rorreeborpgnidulcxe(C4.0

    (*) Uncompensated conductivity (or TDS) is the conductivity (or TDS) value withouttemperature compensation.

  • 7/29/2019 man2550_23_04_2009

    6/64

    6

    egnartesffoVmleR Vm0002

    noitarbilaCHp

    ,noitarbilactniop5ro4,3,2,1elbaliavasreffubdradnats7

    ,)54.21,10.01,81.9,10.7,68.6,10.4,86.1(

    sreffubmotsuc2dna

    noitarbilaCCE

    ;noitarbilacepolstniop1:elbaliavasreffub6

    3141,0.48 mc/Smc/Sm8.111,0.08,88.21,00.5

    :tesffotniop100.0 mc/S

    noitarbilaClCaN htiwtniop1 L7307IH )lanoitpo(reffub

    erutarepmeTnoitasnepmoc

    :morfcitamotuArolaunaM)EGNARHp(C0.021ot0.02)EGNARCE(C0.021ot0.02

    otegnarytivitcudnocnodelbasidebnac(ytivitcudnoclautcaerusaem

    ytivitcudnoCtneiciffeocerutarepmet

    C/%00.6ot00.0)ylnoSDTdnaCErof(

    C/%09.1sieulavtluafed

    rotcafSDT08.0ot04.0

    )05.0sieulavtluafed(

    edortcelEHp B1311IH

    eborPCE01367IH

    eborperutarepmeT 2667IH

    ecnadepmitupnI)tupniCNB(

    01 21 smho

    erutaefdnamednogoL sdrocer002

    erutaeflavretnIgoL

    sdrocer005,)bAtS(gniggolytilibatS

    ces03,01,5nim081,021,06,03,51,01,5,2,1

    noitacinummocCP BSUdetalosiotpO

    ylppusrewoP retpadaCDV21

    snoisnemiD )3.4x7.8x2.9(mm901x222x532

    thgieW;)bl9.2(gK3.1

    )bl6.4(gK1.2redlohhtiwtik

    tnemnorivnE)F22123(C050

    gnisnednoc-nonHR%59.xam

    ytnarraW sraey2

  • 7/29/2019 man2550_23_04_2009

    7/64

    7

    3 cm(1")

    OPERATIONAL GUIDEOPERATIONAL GUIDEOPERATIONAL GUIDEOPERATIONAL GUIDEOPERATIONAL GUIDE

    POWER CONNECTIONPOWER CONNECTIONPOWER CONNECTIONPOWER CONNECTIONPOWER CONNECTION

    Plug the 12 VDC adapter into the power supply socket.

    Notes: This instrument uses non volatile memory to retain the calibrationparameters and all other settings, even when unplugged.

    Make sure a fuse protects the main line.

    ELECTRODE AND PROBE CONNECTIONSELECTRODE AND PROBE CONNECTIONSELECTRODE AND PROBE CONNECTIONSELECTRODE AND PROBE CONNECTIONSELECTRODE AND PROBE CONNECTIONS

    For pH or ORP measurements connect an electrode with internal reference

    to the BNC connector on the back of the instrument.For electrodes with a separate reference connect the electrodes BNC to theBNC connector and the reference electrode plug to the reference socket.

    For temperature measurements and automatic temperature compensationconnect the temperature probe to the appropriate socket.

    For EC/TDS measurements connect the probe to the 7-pin connector.Make sure the probe sleeve is properly inserted.

    INSTRUMENT STINSTRUMENT STINSTRUMENT STINSTRUMENT STINSTRUMENT STAR TAR TAR TAR TAR T-UP-UP-UP-UP-UP

    Turn the instrument on by pressing the ON/OFF switch located onthe rear panel.

    All LCD tags are displayed and a beep is sounded while theinstruments perform a self test.

    The instrument will display LoAd message and blinking untilinitialization is complete.

    Notes: The instrument starts in the same range and mode as it wasat power off.

    The ALT&MODE keys change the measuring modes:

    - mV or Rel mV

    - EC or TDS or NaCl

    The RANGE key toggles between measurement ranges:

    - pH, mV or Rel mV, EC or TDS or NaCl.

    pH MEASUREMENTS

    Make sure the instrument has been calibrated beforetaking pH measurements.

    Submerse the electrode tip and the temperatureprobe approximately 3 cm (1) into the sampleto be tested and stir gently. Allow time for theelectrode to stabilize.

  • 7/29/2019 man2550_23_04_2009

    8/64

    8

    3 cm(1")

    The pH is displayed on the primary LCD and the temperature on the

    secondary LCD.

    If the reading is out of range, the closest full-scale value will be

    displayed blinking on the primary LCD.

    If measurements are taken successively in different samples, it is recommended

    to rinse the electrode thoroughly with deionized water or tap water and

    then with some of the next sample to prevent cross-contamination.

    The pH reading is affected by temperature. In order to measure the pH

    accurately, the temperature effect must be compensated for. To use the

    Automatic Temperature Compensation feature, connect and submerse

    the HI 7662 temperature probe into the sample as close as possible to

    the electrode and wait for a few seconds.

    If the temperature of the sample is known, manual temperature

    compensation can be used by disconnecting the temperature probe.The display will show the last temperature reading with

    the C tag blinking.

    The temperature can now be adjusted with the ARROW

    keys (from 20.0 C to 120.0 C).

    mV/ORP MEASUREMENTSmV/ORP MEASUREMENTSmV/ORP MEASUREMENTSmV/ORP MEASUREMENTSmV/ORP MEASUREMENTS

    An optional ORP electrode must be used to perform ORP measurements

    (see Accessories).

    Oxidation-Reduction Potential (REDOX) measurements provide the

    quantification of the oxidizing or reducing power of the tested sample.

    The surface of the ORP electrode must be clean and smooth in order to

    obtain an accurate measurement. Press RANGE to enter mV range.

    Submerse the tip of the ORP electrode (3 cm/1)

    into the sample to be tested and allow a few

    seconds for the reading to stabilize.

    pH

    pH

  • 7/29/2019 man2550_23_04_2009

    9/64

    9

    The instrument displays the mV reading on the primary LCD and thetemperature on the secondary LCD line.

    If the reading is out of range, the closest full-scale value will bedisplayed blinking on the primary LCD.

    RELATIVE mV MEASUREMENTSRELATIVE mV MEASUREMENTSRELATIVE mV MEASUREMENTSRELATIVE mV MEASUREMENTSRELATIVE mV MEASUREMENTS

    Press the ALT&MODE keys simultaneously while in mV range. TherEL and mV tags are displayed for about one second. After onesecond the temperature will be displayed on the secondary LCD andthe mV tag will blink.

    The reading displayed by the instrument is equal to the difference

    between the current mV input value and the relative mV offset establishedin the relative mV calibration.

    CONDUCTIVITY MEASUREMENTSCONDUCTIVITY MEASUREMENTSCONDUCTIVITY MEASUREMENTSCONDUCTIVITY MEASUREMENTSCONDUCTIVITY MEASUREMENTS

    Connect the conductivity probe to the instrument.

    Press the RANGE key to enter conductivity measurement range (EC).

    Submerse the probe into the solution to be tested. Thesleeve holes must be completely submersed. Tap theprobe repeatedly to remove any air bubbles that maybe trapped inside the sleeve.

    The conductivity value will be displayed on the primaryLCD and the temperature on the secondary LCD.

    If the reading is out of range, the full-scale value (200.0 for MTC/ATCmode or 500.0 for actual conductivity) will be displayed blinking.

    mV

    mV

    mV

  • 7/29/2019 man2550_23_04_2009

    10/64

    10

    IfLOCK key was pressed and the reading goes out of range, thefull-scale value of the frozen range will be displayed blinking.

    The conductivity reading is affected by temperature.

    Three options for temperature compensation are available in conductivity

    measurement mode.

    Automatic (ATC): The conductivity probe has a built-in temperaturesensor; the temperature value is used to automatically compensate the

    EC/TDS reading.

    Manual (MTC): The temperature value, shown on the secondary LCD,

    can be manually set with the ARROW keys. The C tag blinks when

    this option is active. This value will be used to compensate the EC/TDS

    reading.

    No Compensation (notc): The temperature value is displayed, but nottaken into account. When this option is selected the C tag will blink

    with slower frequency. The reading displayed on the primary LCD is the

    uncompensated EC or TDS value.

    To select the desired option, press the

    ALT&ATC keys until the option is displayedon the LCD.

    Notes: The default compensation mode is ATC.

    If no temperature probe is detected, ATC mode can not be

    selected and the instrument displays ---- on the secondary LCD.

    If temperature compensation is selected, measurements are compensated

    using the temperature coefficient (default value 1.90 %/C). To change

    the temperature coefficient, enter the setup mode and select tc (see

    SETUP for details, page 36). The current temperature coefficient can be

    quickly viewed by pressing the ALT&TC keys. The value is briefly

    displayed on the secondary LCD.

    If the temperature reading exceeds the limits of the meter (20.0 Cto 120.0 C), the C tag will blink and the closest full scale value

    will be displayed.

  • 7/29/2019 man2550_23_04_2009

    11/64

    11

    TDS MEASUREMENTSTDS MEASUREMENTSTDS MEASUREMENTSTDS MEASUREMENTSTDS MEASUREMENTS

    Press the ALT&MODE keys while in EC range. The instrument will switch

    to TDS measuring range. The TDS reading will be displayed on the

    primary LCD and the temperature reading on the secondary LCD.

    If the reading is out of range, the full-scale value (100.0 for MTC/ATC

    mode or 400.0 for uncompensated TDS) will be displayed blinking.

    IfLOCK was pressed and the reading goes out of range, the full-scalevalue of the frozen range will be displayed blinking.

    NaCl MEASUREMENTSNaCl MEASUREMENTSNaCl MEASUREMENTSNaCl MEASUREMENTSNaCl MEASUREMENTS

    Press the ALT&MODE keys while in EC range until NaCl is displayed onthe LCD. The instrument will display the NaCl reading on the primary LCD

    and the temperature reading on the secondary LCD line.

    If the reading is out of range, the full-scale value (400.0%) will be

    displayed blinking.

    TEMPERATURE MEASUREMENTSTEMPERATURE MEASUREMENTSTEMPERATURE MEASUREMENTSTEMPERATURE MEASUREMENTSTEMPERATURE MEASUREMENTS

    In pH and ORP mode, connect the HI 7662 temperature probe to the

    appropriate socket.

    Submerse the temperatureprobeinto the sample and allow the reading

    on the secondary LCD to stabilize.

    In EC/TDS/NaCl range, the HI 76310 probe has a built-in temperature

    sensor.

  • 7/29/2019 man2550_23_04_2009

    12/64

    12

    AUTO-RAN GINGAUTO-RANGINGAUTO-RAN GINGAUTO-RANGINGAUTO- RANGI NG

    The EC and TDS scales are auto-ranging. The meter automatically sets

    the scale with the highest possible resolution.

    By pressing ALT&LOCK, the auto-ranging feature is disabled and thecurrent range is frozen on the LCD. The Auto Off (auto-ranging

    disabled) tags will be displayed on the LCD for a few seconds. To restore

    the auto-ranging option, press ALT&LOCK again. The Auto On(auto-ranging enabled) tags will be displayed on the LCD for a few

    seconds.

    Note: Auto-ranging is automatically restored if the range is changed, if

    the setup or calibration modes are entered and if the meter is

    turned off and back on again.

  • 7/29/2019 man2550_23_04_2009

    13/64

    13

    ppppp H CALIBRATIONH CALIBRATIONH CALIBRATIONH CALIBRATIONH CALIBRATION

    Calibrate the instrument frequently, especially if high accuracy is required.

    The instrument should be recalibrated:

    Whenever the pH electrode is replaced.

    At least once a day.

    After testing aggressive chemicals.

    If CAL INTV tags are blinking during measurement.Every time you calibrate the instrument use fresh buffers and perform anelectrode Cleaning Procedure (see page 53).

    PREPARATIONPREPARATIONPREPARATIONPREPARATIONPREPARATIONPour small quantities of the buffer solutions into clean beakers. If possible,use plastic or glass beakers to minimize any EMC interferences.

    For accurate calibration and to minimize cross-contamination, use twobeakers for each buffer solution. One for rinsing the electrode and one forcalibration.

    If you are measuring in the acidic range, use pH 7.01 or 6.86 as first bufferand pH 4.01 as second buffer. If you are measuring in the alkaline range,

    use pH 7.01 or 6.86 as first buffer and pH 10.01 or 9.18 as second buffer.

    PROCEDUREPROCEDUREPROCEDUREPROCEDUREPROCEDURE

    Calibration can be performed at up to five-points.

    For accurate measurements, a three-point calibration is recommended.The calibration buffer can be selected from the calibration buffer list thatincludes the custom buffers and the memorized standard buffers:

    pH 1.68, 4.01, 6.86, 7.01, 9.18, 10.01 and 12.45

    The custom buffers allow the user to calibrate in a buffer solution differentfrom a standard one. Up to two custom buffers can be se t in SETUP menu(see page 36). Each custom buffer value can be changed in a 1.0 pHwindow around the set value, during calibration, when it is selected; theBUFFER pH tag will blink.

    The instruments will automatically skip the buffer used during calibrationand the buffers which are in a 0.2 pH window, around one of thecalibrated buffers.

    All new calibrations will override existing stored calibration data in a0.2 pH window. The slopes adjacent to the new points will bereevaluated.

  • 7/29/2019 man2550_23_04_2009

    14/64

    14

    3 cm(1")

    FIVEFIVEFIVEFIVEFIVE-----POINTPOINTPOINTPOINTPOINT CALIBRATIONCALIBRATIONCALIBRATIONCALIBRATIONCALIBRATION

    Submerse the pH electrode and the temperatureprobe approximately 3 cm (1) into a buffersolution and stir gently. The temperature probeshould be close to the pH electrode.

    Press CAL. The CAL and BUF tags willappear and the 7.01 buffer will be displayedon the secondary LCD.

    If necessary, press the ARROW keys to select a different buffer value.

    The tag will blink on the LCD until the reading is stable.

    When the reading is stable and close to the selectedbuffer, the READY tag will be displayed and theCFM tag will blink.

    Press CFM to confirm calibration.

    The calibrated value is then displayed on the primaryLCD and the second expected buffer value on thesecondary LCD.

    After the first calibration point is confirmed, submerse the pHelectrode and the temperature probe approximately 3 cm (1)

    into the second buffer solution and stir gently. The temperatureprobe should be close to the pH electrode.

    If necessary, press the ARROW keys to select a different buffer value.

    The tag will blink on the LCD until the reading is stable.

    When the reading is stable and close to the selectedbuffer, the READY tag will be displayed and theCFM tag will blink.

    Press CFM to confirm calibration.

    The calibrated value is then displayed on the primaryLCD and the third expected buffer value on thesecondary LCD.

    After the second calibration point is confirmed, submerse the pHelectrode and the temperature probe approximately 3 cm (1) intothe third buffer solution and stir gently. The temperature probeshould be close to the pH electrode.

  • 7/29/2019 man2550_23_04_2009

    15/64

    15

    If necessary, press the ARROW keys to select a different buffer value.

    The tag will blink on the LCD until the reading isstable.

    When the reading is stable and close to the selectedbuffer, the READY tag will be displayed and theCFM tag will blink.

    Press CFM to confirm calibration.

    The calibrated value is then displayed on the primary LCD and thefourth expected buffer value on the secondary LCD.

    After the third calibration point is confirmed, submerse the pH electrode

    and the temperature probe approximately 3 cm (1) into the fourthbuffer solution and stir gently. The temperature probe should be closeto the pH electrode.

    If necessary, press the ARROW keys to select a different buffer value.

    The tag will blink on the LCD until the reading isstable.

    When the reading is stable and close to the selectedbuffer, the READY tag will be displayed and theCFM tag will blink.

    Press CFM to confirm calibration.

    The calibrated value is then displayed on the primary LCD and thefifth expected buffer value on the secondary LCD.

    After the fourth calibration point is confirmed, submerse the pHelectrode and the temperature probe approximately 3 cm (1) intothe fifth buffer solution and stir gently. The temperature probeshould be close to the pH electrode.

    If necessary, press the ARROW keys to select a different buffer value.

    The tag will blink on the LCD until the reading isstable.

    When the reading is stable and close to the selectedbuffer, the READY tag will be displayed and theCFM tag will blink.

    Press CFM to confirm the fifth calibration point.

    The instrument stores the calibration value and returns to normalmeasurement mode.

    FOUR, THREE OR TWO-POINT CALIBRATIONFOUR, THREE OR TWO-POINT CALIBRATIONFOUR, THREE OR TWO-POINT CALIBRATIONFOUR, THREE OR TWO-POINT CALIBRATIONFOUR, THREE OR TWO-POINT CALIBRATION

    Proceed as described in FIVE-POINT CALIBRATION section.

    Press CAL after the fourth, third or second calibration point wasconfirmed. The instruments will memorize the calibration data andreturn to measurement mode.

  • 7/29/2019 man2550_23_04_2009

    16/64

    16

    ONE-POINT CALIBRATIONONE-POINT CALIBRATIONONE-POINT CALIBRATIONONE-POINT CALIBRATIONONE-POINT CALIBRATION

    Two SETUP selectable options are available: Pnt and OFFS.

    If the Pnt option is selected, the new calibration point overrides anexisting one. The adjacent slopes will be reevaluated.

    If the OFFS option is selected, an electrode offset correction is performed.

    The adjacent slopes will remain unchanged.

    Proceed as described in FOUR, THREE or TWO-POINT CALIBRATIONsection.

    Press CAL after the first calibration point was confirmed. Theinstruments will memorize the one-point calibration data andreturn to measurement mode.

    Notes: If the value measured by the instrument is not close to the

    selected buffer, WRONG and WRONG tagswill blink alternately. Check if the correct buffer has beenused, or clean the electrode by following the CleaningProcedure (see page 53). If necessary, change the buffer orthe electrode.

    When a custom buffer is displayed, the BUFFER pH tagblinks. To change the custom buffer value in accordance withthe buffer temperature proceed as described in WORKINGWITH CUSTOM BUFFERS (see page 17).

    If the buffer temperature or the manual temperatureexceeds the temperature limits of the buffer, WRONG tagand temperature reading will blink.

    If WRONG, BUFFER pH tags and OLD message aredisplayed blinking on the secondary LCD line, an inconsistencybetween new and previous (old) calibration is detected.Clear calibration parameters and proceed with calibrationfrom the current calibration point. The instrument will keepall confirmed values during current calibration.

    To clear calibration parameters for all uncalibrated buffersstarting with current buffer, press CLR. The calibration willcontinue from the current point. If this procedure is

    performed while calibrating in the first calibration point,the instrument returns to measurement mode.

    Press RANGE to toggle between pH buffer,calibration buffer number and temperaturereading.

  • 7/29/2019 man2550_23_04_2009

    17/64

    17

    Each time a buffer is confirmed, the new calibration datareplaces the old calibration data for the corresponding buffer.If current buffer has no previous data stored and the cali-bration is not full (five buffers), the current buffer is addedto the existing calibration. If the existing calibration is full,the instrument asks which buffer to replace.

    Press the ARROW keys to select another buffer to bereplaced.Press CFM to confirm the buffer that will be replaced.Press CAL to leave calibration without replacing.

    Note: If the replaced buffer is outside the 0.2 pH window,around each of the calibrated buffers, it is possible toselect this buffer for next calibration during currentcalibration.

    WORKING WITH CUSTOM BUFFERSWORKING WITH CUSTOM BUFFERSWORKING WITH CUSTOM BUFFERSWORKING WITH CUSTOM BUFFERSWORKING WITH CUSTOM BUFFERS

    If a custom buffer was set in SETUP menu, it can be selected duringcalibration by pressing the ARROW keys. The BUFFER pH tag willblink.

    Press SETUP if you want to adjust the buffer value. The buffer value will

    start blinking.

    Use the ARROW keys to change the buffer value.

    After 5 seconds, the buffer value is updated. Press SETUP if you want tochange it again.

    Note: Custom buffer value can be adjusted in a 1.00 pH window,around the set value.

  • 7/29/2019 man2550_23_04_2009

    18/64

    18

    The temperature has an effect on pH. The calibration buffer solutions areaffected by temperature changes to a lesser degree than normal solutions.During calibration the instrument will automatically calibrate to the pHvalue corresponding to the measured or set temperature.

    During calibration the instrument will display the pH buffer value at 25 C.

    pppppH BUFFER TEMPERATUREH BUFFER TEMPERATUREH BUFFER TEMPERATUREH BUFFER TEMPERATUREH BUFFER TEMPERATUREDEP EN DEN C EDEP EN DEN C EDEP EN DEN C EDEP EN DEN C EDEP EN DEN C E

    PMET SREFFUBHp

    C F 86.1 10.4 68.6 10.7 81.9 10.01 54.21

    0 23 76.1 10.4 89.6 31.7 64.9 23.01 83.31

    5 14 76.1 00.4 59.6 01.7 93.9 42.01 81.31

    01 05 76.1 00.4 29.6 70.7 33.9 81.01 99.21

    51 95 76.1 00.4 09.6 50.7 72.9 21.01 08.21

    02 86 86.1 00.4 88.6 30.7 22.9 60.01 26.21

    52 77 86.1 10.4 68.6 10.7 81.9 10.01 54.21

    03 68 86.1 20.4 58.6 00.7 41.9 69.9 92.21

    53 59 96.1 30.4 48.6 99.6 11.9 29.9 31.21

    04 401 96.1 40.4 48.6 89.6 70.9 88.9 89.11

    54 311 07.1 50.4 38.6 89.6 40.9 58.9 38.11

    05 221 17.1 60.4 38.6 89.6 10.9 28.9 07.11

    55 131 27.1 80.4 48.6 89.6 99.8 97.9 75.11

    06 041 27.1 90.4 48.6 89.6 79.8 77.9 44.11

    56 941 37.1 11.4 48.6 99.6 59.8 67.9 23.11

    07 851 47.1 21.4 58.6 99.6 39.8 57.9 12.11

    57 761 67.1 41.4 68.6 00.7 19.8 47.9 01.11

    08 671 77.1 61.4 78.6 10.7 98.8 47.9 00.11

    58 581 87.1 71.4 78.6 20.7 78.8 47.9 19.01

    09 491 97.1 91.4 88.6 30.7 58.8 57.9 28.01

    59 302 18.1 02.4 98.6 40.7 38.8 67.9 37.01

  • 7/29/2019 man2550_23_04_2009

    19/64

    19

    RELATIVERELATIVERELATIVERELATIVERELATIVE mmmmmV CAL IBRATIONV CALIBRATIONV CAL IBRATIONV CALIBRATIONV CALIBRATION

    Press CAL when the instrument is in RELATIVE mV measurementmode. The mV and tags will be displayed. Absolute mV isdisplayed on the primary LCD and AbS message is displayed on thesecondary LCD.

    When the absolute reading is stable and in measurement range, theinstrument asks for confirmation.

    If the reading is out of range, WRONG tag will be displayed.

    Press CFM to confirm the absolute value. The instrument will display

    0.0 mV on the primary LCD and rEL message on the secondaryLCD. In this moment the relative mV offset is equal to absolute mVreading.

    Use the ARROW keys if you want to change the displayed relativemV value.

    Press CFM to confirm the relative mV value. The instrument returnsto measurement mode.

    Note: The relative mV value can be changed only inside the relative mVoffset window ( 2000 mV).

    EC/TDS CAL IBRATIONEC/TDS CAL IBRATIONEC/TDS CAL IBRATIONEC/TDS CAL IBRATIONEC/TDS CAL IBRATION

    Selectable calibration points for conductivity are 0.00 S for offset and84.0S, 1413S, 5.00 mS, 12.88 mS, 80.0 mS, 111.8 mS for slope.

    Rinse the probe with calibration solution or deionizedwater. Submerse the probe into the solution. The sleeveholes must be completely submersed. Tap the proberepeatedly to remove any air bubbles that may be trappedinside the sleeve.To enter EC calibration, select the EC range and pressCAL.

    The BUF and CAL tags are displayed. The primaryLCD will display the EC reading. The secondary LCD will display the value.The and ~ tags will blink.

    Note: The TDS reading is automatically derived from the EC reading andno specific calibration for TDS is needed. Pressing CAL when TDSrange is selected has no effect.

    For zero calibration, just leave the dry probe in the air. This calibrationis performed in order to correct the reading at 0.00 S. The slope isevaluated when the calibration is performed at any other point.

  • 7/29/2019 man2550_23_04_2009

    20/64

    20

    Select the desired value with the ARROW keys, if necessary.

    When the reading is stable, READY tag is displayed and CFM tag

    starts blinking on the LCD, asking for confirmation.

    Press CFM to confirm calibration.

    The instrument stores the calibration value and returnsto measurement mode.

    Notes: If the reading is too far from the expected value, the

    WRONG and tags will blink. Calibration can not beconfirmed.

    In this case check if the calibration solution has been used or

    clean the probe by following the Cleaning Procedure (seepage 53).

    If the meter is in ATC mode and the buffer temperature is

    out of the 0.0 C to 60.0 C interval, WRONG C tagsand the temperature will blink.

    For best results choose an EC buffer value close to the sample

    to be measured.

    In order to minimize any EMC interference, use plastic or

    glass beakers.

    It is possible to set the cell constant value directly, without

    following the calibration procedure. To set the cell constant,enter SETUP mode and select CEL (see SETUP for details,page 36).

    SBUF

    CAL

    SBUF

    CAL

    READY

    CFM

    WRONG

    CAL

  • 7/29/2019 man2550_23_04_2009

    21/64

    21

    CONDUCTIVITY VERSUSCONDUCTIVITY VERSUSCONDUCTIVITY VERSUSCONDUCTIVITY VERSUSCONDUCTIVITY VERSUSTEMPERATURE CHARTTEMPERATURE CHARTTEMPERATURE CHARTTEMPERATURE CHARTTEMPERATURE CHART

    The conductivity of an aqueous solution is a measure of its ability to carry

    an electrical current by means of ionic motion.

    The conductivity invariably increases with increasing temperature.

    It is affected by the type and number of ions in the solutions and by

    the viscosity of the solution itself. Both parameters are temperature

    dependent. The dependency of conductivity on temperature is

    expressed as a relative change per Celsius degrees at a particulartemperature, commonly as %/C.

    The following table lists the temperature dependence of HANNA EC

    calibration buffers.

    C F

    0307IH

    0308IH

    ( )mc/S

    1307IH

    1308IH

    ( )mc/S

    3307IH

    3308IH

    ( )mc/S

    4307IH

    4308IH

    ( )mc/S

    5307IH

    5308IH

    ( )mc/S

    9307IH

    9308IH

    ( )mc/S

    0 23 0517 677 46 00384 00456 0672

    5 14 0228 698 56 00535 00147 0813

    01 05 0339 0201 76 00695 00238 5163

    51 95 08401 7411 86 00456 00529 360461 8.06 02701 3711 07 00276 00449 5514

    71 6.26 05901 9911 17 00586 00369 5424

    81 4.46 09111 5221 37 00896 00289 7334

    91 2.66 03411 1521 47 00317 002001 9244

    02 86 07611 8721 67 00427 001201 3254

    12 8.96 01911 5031 87 00047 000401 7164

    22 6.17 05121 2331 97 00257 009501 1174

    32 4.37 09321 9531 18 00567 009701 5084

    42 2.57 04621 6831 28 00387 008901 2094

    52 77 08821 3141 48 00008 008111 0005

    62 8.87 03131 0441 68 00318 008311 6905

    72 6.08 07331 7641 78 00038 007511 0915

    82 4.28 02631 4941 98 00948 007711 682592 2.48 07831 1251 09 00368 007911 3835

    03 68 02141 8451 29 00288 008121 9745

    13 8.78 07341 5751 49 00009 009321 5755

  • 7/29/2019 man2550_23_04_2009

    22/64

    22

    NaCl CAL IBRATIONNaCl CAL IBRATIONNaCl CAL IBRATIONNaCl CAL IBRATIONNaCl CAL IBRATIONNaCl calibration is a one-point calibration at 100.0% NaCl. Use the

    HI 7037L calibration solution (sea water solution) as a 100% NaCl

    calibration solution.

    Rinse the probe with some of the calibration solution or

    deionized water. Submerse the probe into HI 7037Lsolution. The sleeve holes must be completely submersed.

    Tap the probe repeatedly to remove any air bubbles that

    may be trapped inside the sleeve.

    To enter NaCl calibration select the NaCl range and pressCAL.

    The BUF and CAL tags are displayed. The primary LCD will display

    the NaCl reading in percentage. The secondary LCD will display 100.

    The and ~ tags will blink.

    When the reading is stable, the READY tag will be displayed and the

    CFM tag starts blinking on the LCD, asking for confirmation.

    Press CFM to confirm calibration.

    The instrument stores the calibration value and returns to

    measurement mode.

    Notes: If the reading is too far from the expected value, WRONG

    tags will blink. Calibration cannot be confirmed.

    If the temperature of the buffer is out of the 0.0 C to 60.0 C

    temperature interval, the WRONG and C tags andthe temperature will blink.

    If a new EC calibration is performed, the NaCl calibration isautomatically cleared. A new NaCl calibration is required.

    BUF

    CAL %

    NaCl

    %

    BUF

    READY

    CFM

    CAL

    NaCl

  • 7/29/2019 man2550_23_04_2009

    23/64

    23

    GOOD LABORATORY PRACTICE (GLP)GOOD LABORATORY PRACTICE (GLP)GOOD LABORATORY PRACTICE (GLP)GOOD LABORATORY PRACTICE (GLP)GOOD LABORATORY PRACTICE (GLP)

    GLP is a set of functions that allows storage and retrieval of data

    regarding the maintenance and status of the electrode.

    All data regarding pH, Rel mV, EC and NaCl calibration is stored for the

    user to review when necessary.

    EXPIRED CALIBRATIONEXPIRED CALIBRATIONEXPIRED CALIBRATIONEXPIRED CALIBRATIONEXPIRED CALIBRATION

    For pH calibration, this instrument allows the user to set the number of

    days before the next required pH calibration. This value can be set from

    1 to 7 days. The default setting is OFF (disabled).When calibration has expired the CAL and INTV tags will blink to

    warn the user that the instrument should be recalibrated.

    For example, if a 4 days time out has been selected, the instrument will

    issue the alarm exactly 4 days after the last calibration.

    If the expiration value is changed (e.g. to 5 days), then the alarm will

    be immediately recalculated and appear 5 days after the last calibration.

    Note: If the instrument was not calibrated, the CAL INTV tags will

    be displayed even if the feature is disabled in SETUP menu.

    pHpHpHpHpH CALIBRATION DATACALIBRATION DATACALIBRATION DATACALIBRATION DATACALIBRATION DATA

    Calibration data is stored automatically after a successful calibration.

    To view the pH calibration data, press GLP when the instrument is in pH(mV) measurement mode.

    The instrument will display the time (hh:mm) of the last calibration.

    Use the ARROW keys to scroll through the calibration data:

    The date (yyyy.mm.dd).

    The pH calibration offset.

    mVCAL

  • 7/29/2019 man2550_23_04_2009

    24/64

    24

    The pH calibration slope (the GLP slope is the average of the calibrationslopes; the percentage is referred to the ideal value of 59.16 mV/pH).

    The calibration buffers in order.

    The first pH calibration buffer:

    The second pH calibration buffer:

    The third pH calibration buffer:

    The fourth pH calibration buffer:

    The fifth pH calibration buffer:

    Notes: The OLd message displayed beside the pH value meansthat this buffer was not used during last calibration. Pressand hold down the SETUP key if you want to see calibrationdate (or time if old calibration was performed on the sameday as the current calibration).

    If no bUF message appears on the LCD, the instrumentinforms you that calibration was performed in less thanthree points.

    CAL pH

    BUF

    CAL pH

    BUF

    CAL pH

    BUF

    BUF

    CAL

    Calibration Expiration status:

    - if disabled.

  • 7/29/2019 man2550_23_04_2009

    25/64

    25

    RRRRRelative mVelative mVelative mVelative mVelative mV CALIBRATION DATACALIBRATION DATACALIBRATION DATACALIBRATION DATACALIBRATION DATA

    Relative mV calibration data is stored automatically after a successfulcalibration.

    To view the Relative mV calibration data, press GLP when the instrumentis in Relative mV measurement mode.

    The instrument will display the time (hh:mm) of the last calibration.

    Use the ARROW keys to scroll through the calibration data:

    The date (yyyy:mm:dd).

    The Relative mV calibration offset.

    The instrument ID.

    INTVCAL

    TIMECAL

    - or the number of days until the calibration alarm will be displayed.

    - or if expired (7 days ago).

    The instrument ID.

  • 7/29/2019 man2550_23_04_2009

    26/64

    26

    EC CALIBRATION DATAEC CALIBRATION DATAEC CALIBRATION DATAEC CALIBRATION DATAEC CALIBRATION DATA

    EC calibration data is stored automatically after a successful calibration.

    To view the EC calibration data, press GLP when the instrument is in ECmeasurement mode.The instrument will display the time (hh:mm) of the last calibration.

    Use the ARROW key to scroll through the calibration data: The date (yyyy:mm:dd).

    The EC calibration buffer.

    The cell constant.

    The calibration offset factor.

    The reference temperature.

    The temperature coefficient.

  • 7/29/2019 man2550_23_04_2009

    27/64

    27

    The temperature compensation mode.

    The instrument ID.

    NNNNNaClaC laClaC laCl CALIBRATION DATACALIBRATION DATACALIBRATION DATACALIBRATION DATACALIBRATION DATA

    NaCl calibration data is stored automatically after a successful calibration.To view the NaCl calibration data, press GLP when the instrument is inNaCl measurement mode.The instrument will display the time (hh:mm) of the last calibration.

    Use the ARROW key to scroll through the calibration data:

    The date (yyyy:mm:dd).

    The salinity coefficient.

    The cell constant.

    The reference temperature.

  • 7/29/2019 man2550_23_04_2009

    28/64

    28

    The temperature compensation mode.

    The instrument ID.

    Notes: If no temperature compensation is selected during calibration,the temperature coefficient is not displayed in GLP.

    Press GLP at any moment and the instrument will return tomeasurement mode.

    If calibration has not been performed on the selected range,the instrument displays no CAL message blinking.

    CAL

    LOGGING FUNCTIONLOGGING FUNCTIONLOGGING FUNCTIONLOGGING FUNCTIONLOGGING FUNCTION

    Up to 700 logged samples can be stored into memory.

    200 manually logged records and 500 lot logging records can be storedin the memory. To select logging type enter SETUP menu.

    LOGGING THE CURRENT DATALOGGING THE CURRENT DATALOGGING THE CURRENT DATALOGGING THE CURRENT DATALOGGING THE CURRENT DATA (manual logging)(manual logging)(manual logging)(manual logging)(manual logging)

    Select the manual logging mode in SETUP menu.

    To store the current reading into memory press LOG while the instrumentis in measurement mode.

    The instrument will display MAn on the primary LCD, the record

    number on the secondary LCD and LOG tag for a few seconds (seeexample below: record No. 15):

  • 7/29/2019 man2550_23_04_2009

    29/64

    29

    If stability logging is selected, a complete set of data is memorized everytime the reading becomes stable after an unstable condition.

    To stop interval logging press LOG key again. The LOG tag will becleared.

    Note: When pressing any key that is not active, while lot logging isrunning, the following message is displayed for a few seconds.

    followed by the number of free records:

    If there are less than 6 memory locations remaining, the record numberand Lo message will be displayed to alert the user.

    If the log space is full (200 records), FULL LOG message will be

    displayed and no more data will be saved.

    When LOG is pressed, a complete set of information is stored: date, time,pH, mV, EC, TDS, NaCl temperature and calibration data.

    LOT LOGGINGLOT LOGGINGLOT LOGGINGLOT LOGGINGLOT LOGGING

    Select StAb (stability logging) or the desired time interval.

    To start interval logging press LOG key while the instrument is inmeasurement mode.

    When the selected interval is reached or when the reading is stable (forlog on stability), the instrument will display the current lot number onthe primary LCD line, the record number on the secondary LCD line andthe LOG tag (see example below: Lot 5 record 7)

    followed by the number of free records on the corresponding memoryspace.

  • 7/29/2019 man2550_23_04_2009

    30/64

    30

    VIEW LOGGED DATAVIEW LOGGED DATAVIEW LOGGED DATAVIEW LOGGED DATAVIEW LOGGED DATA

    Press the ALT&RCLkeys while in measurement mode to retrieve thestored information.

    If no data was logged for the current selected measurement range andno lots are memorized, one of the next messages will be displayed:

    No pH measurements records:

    No Relative mV and mV records:

    No EC records:

    No TDS records:

    No NaCl records:

    Otherwise, the instrument will display the lot number on the primary LCDline, the number of records on the secondary, LOG tag and CFMblinking. If samples were logged on demand MAn will be displayed onthe primary LCD and the number of samples logged on the secondary(see example below: manual log, 15 samples logged).

    Press ARROW keys to select different lot.

    L OG mV

    LOG

    LOG

    LOG

    LOG

    LOG

  • 7/29/2019 man2550_23_04_2009

    31/64

    31

    In TDS measurement range:

    In NaCl measurement range:

    Use the ARROW keys to scroll through the records.

    Press CFM to view record information.

    IfRCL was entered while in pH measurement range:

    In Rel mV range:

    In EC measurement range:

    pHLOG

  • 7/29/2019 man2550_23_04_2009

    32/64

    32

    Note: The instrument will automatically skip log records from othermeasurement ranges.

    To view additional information press RANGE:

    For pH

    The mV value on the primary LCD and the temperature value on thesecondary LCD.

    TIMELOG

    The time on the primary LCD, along with TIME tag and the recordnumber on the secondary LCD.

    The date on the primary LCD, along with DATE tag.

    The calibration offset on the primary LCD and OFS message onthe secondary LCD.

    The calibration slope on the primary LCD and SLP message onthe secondary LCD.

    The interval for lot logging.

    To delete manual logged records: press CFM while MAn is displayed toview manually logged records. Press the CLR key, dEL and the recordnumber will be displayed. Press CFM to delete. Use the ARROW keys tochange the record number.

    To delete a lot, use the ARROW keys to select the desired lot. Press CLRkey, dEL LOt will appear on the display. Press CFM key to delete.

    mVLOG

    %LOG

  • 7/29/2019 man2550_23_04_2009

    33/64

    33

    The dEL message is displayed on the primary LCD and the selected recordon the secondary LCD, along with LOG tag.

    The ARROW keys can be used to change the record or lot number.

    Press SETUP to delete all records/lots. The display will show dEL inthe primary LCD and ALL in the secondary LCD.

    Press CFM to confirm delete. While deleting the tag will blink.

    Press CAL or RANGE or CLR to escape and return to the RCL screen.

    If del ALL option was selected, all the log on demand records or lotsare deleted. While deleting the tag is displayed blinking.

    Press ALT&RCL exit record information and enter lot information.

    Press ALT&RCL again to return to measurement mode.

    If one or more records/lots were deleted the tag blinks until thelog memory space is reorganized.

    Note: If at least one record was deleted the log memory space will berecognized in about 4-5 seconds. During this period the willblink.

    For Relative mV and mV Range

    The temperature value, the mV absolute value, the time and thedate as described above.

    The Relative mV offset.

    For EC Range

    The time and date as described on pH Range. The EC on primary LCD and temperature value on the secondary LCD.

    The offset factor on the primary LCD and OFS message on thesecondary LCD.

    mVLOG

  • 7/29/2019 man2550_23_04_2009

    34/64

    34

    The cell constant on the primary LCD and CEL message on thesecondary LCD line.

    The reference temperature on the primary LCD and rEFmessage on the secondary LCD.

    The temperature coefficient on the primary LCD and tc messageon the secondary LCD.

    The temperature compensation mode.

    For TDS Range The temperature reading as described in pH range.

    The conductivity value on the primary LCD and the temperaturevalue on the secondary LCD.

    The time and the date as described in pH Range.

    The TDS factor on the primary LCD and cF message on thesecondary LCD.

    The reference temperature, the temperature coefficient,the temperature compensation mode and the cell constant asdescribed in EC Range.

  • 7/29/2019 man2550_23_04_2009

    35/64

    35

    For NaCl Range

    The conductivity and temperature reading value as described in TDSRange.

    The time and date as described in pH Range.

    The salinity factor on the primary LCD and cF message on thesecondary LCD, with LOG and NaCl tags displayed.

    The reference temperature, the temperature compensation modeand the cell constant message as described above.

    Note: After LOG is pressed or dEL is confirmed, the instrument willdisplay the amount of free log space for about one second(example 25 records free).

    Press the ALT&RCL keys to leave RECALL mode at any time.

    NaClLOG

    Note: When an information that does not display the record numberis selected, pressing the SETUP key will display the recordnumber on the secondary LCD line.

  • 7/29/2019 man2550_23_04_2009

    36/64

    36

    SET UPSET UPSET UPSET UPSET UP

    Setup mode allows viewing and modifying the following instrumentparameters.

    In according with the selected range, SETUP menu allows the possibilityto view and /or change specific range parameters and common parameters(for all the ranges).

    The common parameters are:

    Log interval

    Current Time (hour & minute)

    Current Date (year, month & day)

    Beep Status

    Instrument Id

    Temperature Unit

    The specific parameters are:

    In pH range

    Expired Calibration Alarm

    First Custom Buffer

    Second Custom Buffer

    One-point Calibration Behavior

    pH Resolution

    In EC/TDS/NaCl range

    Cell Constant

    TDS Factor

    Temperature Compensation Coefficient

    Reference Temperature

    To enter SETUP mode press SETUPwhile the instrument is in measurementmode.

    Select a parameter with the ARROW keys.

    Press CAL to change a parameter value. The selected parameter willstart blinking.

    Press RANGE to toggle between displayed parameters.

    Press the ARROW keys to increase or decrease the displayed value.

    Press CFM to save the modified value or CAL to escape without saving.

  • 7/29/2019 man2550_23_04_2009

    37/64

    37

    EXPIRED CALIBRATION ALARMEXPIRED CALIBRATION ALARMEXPIRED CALIBRATION ALARMEXPIRED CALIBRATION ALARMEXPIRED CALIBRATION ALARM

    Press CAL when the calibration time-out is displayed. Calibrationtime-out (OFF or 1 to 7 days) will start blinking.

    Press the ARROWkeys to change the calibration time-out value.

    Press CFM to save the modified calibration time-out value.

    Press CAL to escape without saving.

    LOG INTERVALLOG INTERVALLOG INTERVALLOG INTERVALLOG INTERVAL

    Press CAL when log interval is displayed. The log interval is selectedblinking (MAn for log on demand, StAb log on stability, interval inseconds or minutes).

    Press the ARROW keys to change the custom buffer value.

    Press CFM to confirm the selection.

    Press CAL to escape without saving.

    FIRST CUSTOM BUFFERFIRST CUSTOM BUFFERFIRST CUSTOM BUFFERFIRST CUSTOM BUFFERFIRST CUSTOM BUFFER

    Press CAL when cb1 is displayed. The custom buffer (disabled noor 0 to 16 pH) will start blinking.

    Press the ARROW keys to change the custom buffer value.

    Press CFM to save the modified custom buffer value.

    Press CAL to escape without saving.

    SECOND CUSTOM BUFFERSECOND CUSTOM BUFFERSECOND CUSTOM BUFFERSECOND CUSTOM BUFFERSECOND CUSTOM BUFFER

    Press CAL when cb2 is displayed. The custom buffer (disabled no

    or 0 to 16 pH) will start blinking.

    Press the ARROW keys to change the custom buffer value.

    Press CFM to save the modified custom buffer value.

    Press CAL to escape without saving.

  • 7/29/2019 man2550_23_04_2009

    38/64

    38

    pH RESOLUTIONpH RESOLUTIONpH RESOLUTIONpH RESOLUTIONpH RESOLUTION

    Press CAL when rES message is displayed on the secondary LCD. Theset resolution (0.1, 0.01 or 0.001) will start blinking.

    Press the ARROW keys to toggle between 0.1, 0.01 and 0.001 options.Press CFM to save the modified value.

    Press CAL to escape without saving.

    CURRENT TIMECURRENT TIMECURRENT TIMECURRENT TIMECURRENT TIME

    Press CAL when the current time is displayed. The hour will start blinking.

    Press the ARROWkeys to change the hour.

    Press RANGE. The minutes will start blinking.

    Press the ARROW keys to change the minutes.

    Press CFM to save the modified value.

    Press CAL to escape without saving.

    ONE-POINT CALIBRATION BEHAVIORONE-POINT CALIBRATION BEHAVIORONE-POINT CALIBRATION BEHAVIORONE-POINT CALIBRATION BEHAVIORONE-POINT CALIBRATION BEHAVIOR

    Press CAL when 1 Pnt message is displayed on the secondary LCD.One of the two options (Pnt or OFFS) will start blinking (see pHCALIBRATION PROCEDURE for details, page 13).

    Press the ARROW keys to toggle between Pnt and OFFS options.

    Press CFM to save the behavior for one-point calibration.

    Press CAL to escape without saving.

    Note: To remove a custom buffer from the calibration list enter custombuffer press CAL key then press CLR key. The nocb1 or cb2message will be displayed and the instrument return for theSETUP parameter scroll mode.

  • 7/29/2019 man2550_23_04_2009

    39/64

    39

    BEEP STATUSBEEP STATUSBEEP STATUSBEEP STATUSBEEP STATUS

    Press CAL when the beep status is displayed. Beep status (On orOFF) will start blinking.

    Press the ARROWkeys to change the beep status (On or OFF).

    Press CFM to save the modified beep status.

    Press CAL to escape without saving.

    When enable, beep sounds as a short beep every time a key is pressed orwhen the calibration can be confirmed.

    A long beep alert that the pressed key is not active or a wrong conditionis detected while in calibration.

    CURRENT DATECURRENT DATECURRENT DATECURRENT DATECURRENT DATE

    Press CAL when the current date is displayed. The year will start blinking.

    Press the ARROWkeys to change the year.

    Press RANGE. The month will start blinking.

    Press the ARROWkeys to change the month.

    Press RANGE. The day will start blinking.

    Press the ARROW keys to change the day.

    Press CFM to save the modified value.

    Press CAL to escape without saving.

  • 7/29/2019 man2550_23_04_2009

    40/64

    40

    CELL CONSTANTCELL CONSTANTCELL CONSTANTCELL CONSTANTCELL CONSTANT

    Press CAL when the cell constant is displayed. The cell constant will startblinking.

    Press the ARROW keys to change the cell constant (0.500 to 1.700).

    Press CFM to save the modified cell constant.

    Press CAL to escape without saving.

    IIIIINSTRUMENT IDNSTRUMENT IDNSTRUMENT IDNSTRUMENT IDNSTRUMENT ID

    Press CAL when InId is displayed. The instrument ID (0000 to9999) will start blinking.

    Press the ARROWkeys to change the instrument ID value.

    Press CFM to save the modified instrument ID value.

    Press CAL to escape without saving.

    Note: The instrument ID is downloaded to a PC as part of a loggeddata, set to identify its origin.

    TEMPERATURE UNITTEMPERATURE UNITTEMPERATURE UNITTEMPERATURE UNITTEMPERATURE UNIT

    Press CAL when tnP is displayed. The temperature unit will startblinking.

    Press the ARROW keys to change the option.

    Press CFM to save the modified temperature unit.

    Press CAL to escape without saving.

  • 7/29/2019 man2550_23_04_2009

    41/64

    41

    TDS FACTORTDS FACTORTDS FACTORTDS FACTORTDS FACTOR

    Press CAL when tdS is displayed. The TDS factor will start blinking.

    Press the ARROW keys to change the TDS factor (0.40 to 0.80).Press CFM to save the modified TDS factor.Press CAL to escape without saving.

    TEMPERATURE COMPENSATION COEFFICIENTTEMPERATURE COMPENSATION COEFFICIENTTEMPERATURE COMPENSATION COEFFICIENTTEMPERATURE COMPENSATION COEFFICIENTTEMPERATURE COMPENSATION COEFFICIENT

    Press CAL when the temperature compensation coefficient is displayed.The temperature compensation coefficient will start blinking.

    Press the ARROW keys to change the temperature compensationcoefficient. (0.00 to 6.00 %/C).Press CFM to save the modified temperature compensation coefficient.

    Press CAL to escape without saving.

    REFERENCE TEMPERATUREREFERENCE TEMPERATUREREFERENCE TEMPERATUREREFERENCE TEMPERATUREREFERENCE TEMPERATURE

    Press CAL when the reference temperature is displayed. The referencetemperature will start blinking.

    Press the ARROW keys to toggle between 20.0 C and 25.0 C referencetemperature value.

    Press CFM to save the modified reference temperature value.

    Press CAL to escape without saving.

  • 7/29/2019 man2550_23_04_2009

    42/64

    42

    The instrument has two temperature channels: one that measures thetemperature from the HI 7662 probe while the instrument is in pH/mVrange and the other that measures temperature from the EC probe whilethe instrument is in EC/TDS/NaCl range.

    All the instruments are factory calibrated for temperature on both channels.

    Hannas temperature probes are interchangeable and no temperaturecalibration is needed when they are replaced.

    If the temperature measurements are inaccurate, temperature recalibrationshould be performed.

    For an accurate recalibration, contact your dealer or the nearest HannaCustomer Service Center, or follow the instructions below.

    Prepare a vessel containing ice and water and another one containinghot water (around 50 C). Place insulation material around the vesselsto minimize temperature changes.

    Use a calibrated thermometer with a resolution of 0.1 C as areference thermometer. Connect the HI 7662 probe to the appropriatesocket for the pH temperature channel or the HI 76310 probe for theEC temperature channel.

    With the instrument off, press and hold down the CFM & SETUPkeys, then power on the instrument to calibrate the pH temperature

    channel or &RANGE keys and then power on the instrument tocalibrate the EC temperature channel. The CAL tag will appearand the secondary LCD will show 0.0 C.

    Submerse the temperature probe (or EC probe) in the vessel with iceand water as near as possible to the calibrated thermometer. Allowa few seconds for the probe to stabilize.

    Use the ARROW keys to set the reading on the secondary LCD to thatof ice and water, measured by the calibrated thermometer. Whenthe reading is stable and close to the selected calibration point,READY tag will appear and CFM tag will blink.

    Press CFM to confirm. The secondary LCD will show 50.0 C.

    TEMPERATURE CALIBRATIONTEMPERATURE CALIBRATIONTEMPERATURE CALIBRATIONTEMPERATURE CALIBRATIONTEMPERATURE CALIBRATION((((( fo r techn i cal personnel on l y fo r techn i cal personnel on l y fo r techn i cal personnel on l y fo r techn i cal personnel on l y fo r techn i cal personnel on l y )))))

    CAL

    CAL

  • 7/29/2019 man2550_23_04_2009

    43/64

    43

    Submerse the temperature probe (or EC probe) in the second vesselas near as possible to the calibrated thermometer. Allow a fewseconds for the probe to stabilize.

    Use the ARROW keys to set the reading on the secondary LCD to thatof the hot water.

    When the reading is stable and close to the selected

    calibration point, READY tag will appear andCFM tag will blink.

    Press CFM to confirm. The instrument memorizecalibration and restart in measurement mode.

    Note: If the reading is not close to the selected calibration point,WRONG tag will blink. Change the temperature probe (or ECprobe) and restart calibration.

  • 7/29/2019 man2550_23_04_2009

    44/64

    44

    mmmmmV CALIBRATIONV CALIBRATIONV CALIBRATIONV CALIBRATIONV CALIBRATION((((( fo r techn i cal personnel on l y fo r techn i cal personnel on l y fo r techn i cal personnel on l y fo r techn i cal personnel on l y fo r techn i cal personnel on l y )))))

    All the instruments are factory calibrated for mV.

    Hannas ORP electrodes are interchangeable and no mV calibration is

    needed when they are replaced.

    If the mV measurements are inaccurate, mV recalibration should be

    performed.

    For an accurate recalibration, contact your dealer or the nearest Hanna

    Customer Service Center or follow the instructions below.

    A two-point calibration can be performed at 0.0 mV and 1800.0 mV.

    Attach to the BNC connector a mV simulator with an accuracy of

    0.1 mV.

    With the instrument off, press and hold down the CAL & keys,then power on the instrument. The CAL tag will appear and thesecondary LCD will show 0.0 mV.

    Set 0.0 mV on the simulator.

    When the reading is stable and close to the selected calibration

    point, READY tag will appear and CFM tag will blink.

    Press CFM to confirm. The secondary LCD will display 1800 mV.

    Set 1800.0 mV on the simulator.

    When the reading is stable and close to the selected calibrationpoint, READY tag will appear and CFM tag will blink.

    Press CFM to confirm. The instrumentmemorize calibration and

    restart to measurement mode.

    Notes: If the reading is not close to the selected calibration point,WRONG tag will blink. Verify calibration condition orcontact your vendor if you can not calibrate.

    Pressing CAL key during calibration process the instrumentquit calibration mode and restart to measurement mode

    without memorizing calibration.

  • 7/29/2019 man2550_23_04_2009

    45/64

    45

    Data transmission from the instrument to the PC can be done withthe HI 92000 Windows compatible software (optional). HI 92000also offers graphing and on-line help feature.

    Data can be exported to the most popular spreadsheet programs forfurther analysis.

    To connect your instrument to a PC, use a standard USB cable connector.Make sure that your instrument is switched off and plug one connector tothe instrument USB socket and the other to the USB port of your PC.

    Note: If you are not using Hanna Instruments HI 92000 software,please see the following instructions.

    SENDING COMMANDS FROM PCSENDING COMMANDS FROM PCSENDING COMMANDS FROM PCSENDING COMMANDS FROM PCSENDING COMMANDS FROM PC

    It is also possible to remotely control the instrument with any terminalprogram. Use a standard USB cable to connect the instrument to a PC,start the terminal program and set the communication options as follows:8, N, 1, no flow control, 9600 baud rate.

    COMMAND TYPESCOMMAND TYPESCOMMAND TYPESCOMMAND TYPESCOMMAND TYPES

    To send a command to the instrument the scheme is:

    where: is a selectable ASCII characterbetween 0 and 47.

    is the command code (3 characters).

    Note: Either small or capital letters can be used.

    SIMPLE COMMANDSSIMPLE COMMANDSSIMPLE COMMANDSSIMPLE COMMANDSSIMPLE COMMANDS

    RNG Is equivalent to pressing RANGE

    CAL Is equivalent to pressing CAL

    CFM Is equivalent to pressing CFM

    UPC Is equivalent to pressing the UP arrow key

    DWC Is equivalent to pressing the DOWN arrow key

    SET Is equivalent to pressing SETUP

    LOG Is equivalent to pressing LOGMOD Is equivalent to pressing MODE

    GLP Is equivalent to pressing GLP

    CLR Is equivalent to pressing CLR

    RCL Is equivalent to pressing RCL

    PC INTERFACEPC INTERFACEPC INTERFACEPC INTERFACEPC INTERFACE

  • 7/29/2019 man2550_23_04_2009

    46/64

    46

    CHRxx Change the instrument range according with the parametervalue (xx):

    xx=00 pH range/0.001 resolution

    xx=01 pH range/0.01 resolution

    xx=02 pH range/0.1 resolution

    xx=03 mV range

    xx=04 Relative mV range

    xx=06 EC range

    xx=07 TDS range

    xx=08 NaCl rangeThe instrument sends the ACK (6) character every time a command isrecognized and a NAK (21) character for invalid commands.

    COMMANDS REQUIRING AN ANSWERCOMMANDS REQUIRING AN ANSWERCOMMANDS REQUIRING AN ANSWERCOMMANDS REQUIRING AN ANSWERCOMMANDS REQUIRING AN ANSWER

    RAS Causes the instrument to send a complete set of readings inaccording with the current range:

    pH, mV and temperature reading on pH range.

    mV and temperature reading on mV range.

    Rel mV, absolute mV and temperature reading onRel mV range.

    Conductivity and temperature reading on EC range.

    TDS and temperature reading on TDS range.

    NaCl and temperature reading on NaCl range.

    The answer string contains:

    Meter mode (2 chars):

    00 - pH range (0.001 resolution)

    01 - pH range (0.01 resolution)

    02 - pH range (0.1 resolution)

    03 - mV range

    04 - Rel mV range

    06 - EC range

    07 - TDS range

    08 - NaCl range

    Meter status (2 chars of status byte): represents a 8bit hexadecimal encoding.

    0x10 - temperature probe is connected

    0x01 - new GLP data available

    0x02 - new SETUP parameter

  • 7/29/2019 man2550_23_04_2009

    47/64

    47

    Reading status (2 chars): R - in range, O - overrange, U - under range. First character correspondsto the appropriate range reading. Second charactercorresponds to mV reading (if exist).

    Primary reading (corresponding to the selected range)- 7 ASCII chars, including sign and decimal point.

    Secondary reading (only when primary reading is not

    mV, EC, NaCl, TDS) - 7 ASCII chars, including sign and

    decimal point.

    Temperature reading - 8 ASCII chars, with sign and

    two decimal points, always in C.MDR Requests the instrument model name and firmware code.

    GLP Requests the calibration data record.

    The answer string contains:

    GLP status (1 char): represents a 4 bit hexadecimal

    encoding.

    0x01 - pH calibration available

    0x02 - Rel mV calibration available

    0x04 - EC calibration available

    0x08 - NaCl calibration available

    pH calibration data (if available), which contains:

    the number of calibrated buffers (1 char) the offset, with sign and decimal point (7 chars)

    the average of slopes, with sign and decimal point

    (7 chars)

    the calibration time, yymmddhhmmss (12 chars)

    buffers information (for each buffer)

    type (1 char): 0 - standard, 1 - custom

    status (1 char): N (new) - calibrated in last

    calibration; O (old) - from an old calibration.

    warnings during calibration (2 chars): 00 - no

    warning, 04 - Clean Electrode warning.

    buffer value, with sign and decimal point(7 chars).

    the calibration time,yymmddhhmmss (12 chars).

    Rel mV calibration data (if available), which contains:

    the calibration offset, with sign (7 chars)

    the calibration time,yymmddhhmmss (12 chars).

  • 7/29/2019 man2550_23_04_2009

    48/64

    48

    EC calibration data (if available), which contains:

    the number of calibrated standards (1 char)

    the offset factor, with sign and decimal point

    (7 chars)

    the cell constant, with sign and decimal point

    (7 chars)

    the calibration time, yymmddhhmmss (12 chars)

    standards information (for each standard)

    standard value, with sign and decimal point

    (7 chars). buffer unit (2 chars; 00-S; 01-mS)

    Reference Temperature with and decimal point

    (4 chars)

    Temperature Compensation mode (2 chars)

    00 - no temperature compensation

    01 - automatic temperature compensation

    00 - manual temperature compensation

    TC coeficient with sign and decimal point (4 chars)

    calibration time,yymmddhhmmss(12 chars).

    Na Cl Calibration data

    the number of calibrated data (1 char) salinity coeficient, with sign and decimal point

    (7 chars)

    Cell constant, with sign and decimal point(2 chars)

    the calibration time (2 chars)

    buffer information, for each buffer:

    Reference Temperature with and decimal point

    (4 chars)

    Temperature Compensation mode (2 chars)

    00 - no temperature compensation

    01 - automatic temperature compensation

    00 - manual temperature compensation

    TC coeficient with sign and decimal point (4 chars)

    calibration time,yymmddhhmmss(12 chars).

    PAR Requests the setup parameters setting.

    The answer string contains:

    Instrument ID (4 chars)

  • 7/29/2019 man2550_23_04_2009

    49/64

    49

    Calibration alarm time out for pH range (2 chars)

    SETUP information (2 chars): 8 bit hexadecimalencoding.

    0x01 - beep ON (else OFF)

    0x04 - degrees Celsius (else degrees Fahrenheit)

    0x08 - Offset calibration (else Point calibration)

    The number of custom buffers (1 char)

    The custom buffer values, with sign and decimalpoint, for each defined custom buffer (7 chars)

    Log type 2 chars

    01 - manual log

    02 - stability lot log

    03 to 14 - the coresponding interval for lot log (5 sto 180 min)

    cell constant, with sign and decimal point (6 chars)

    TDS factor, with sign and decimal point (5 chars)

    TC coef, with sign and decimal point (5 chars)

    Reference Temperature, with sign and decimal point(5 chars)

    Temperature Compensation mode (1 char)

    NSLx Requests the number of logged samples.

    x = P - request for pH range

    M - request for mV range

    E - request for EC range

    N - request for NaCl range

    T - request for TDS range

    LODxnnn: request thennn record of the manual log on the x range

    LODxALLff: groups all Log On Demand Records in frames of 8 recordseach for the selected range

    Command Parameters:

    x - range (see Note)

    ALL - download all records for the selected range

    ff - requested frame number - first frame is labeled 01LLsxff: requests information about all lots on the specified range, it sendsthe information in frames of 10 lots each (a frame contains informationabout 10 lots)

    Command Parameters:

    x - range (see Note)

    ff - requested frame number - first frame is labeled 01

  • 7/29/2019 man2550_23_04_2009

    50/64

    50

    GLDxxxff: Requests the records of the xxx lot number. The records aresent in frames of 10 records; ff is the frame number (01 firstframe). (Example: Lot 13 has 53 records. The records will besent in 6 frame, 5 with10 records and 1 with 3 records.)

    Command Parameters:

    xxx - Lot number (eq: for lot number 1 xxx = 001)

    ff - requested frame number - first frame is labeled 01

    Errors: Err3 log on demand empty.

    Err4 requested set parameter is not available.

    Err5 command argument is wrong.

    Err6 requested range not available.

    Err7 meter in log mode.

    Err8 is sent if instrument is not in measurement mode.

    NAK (21) character is sent when the instrument receivesan unknown or a corrupted command.

    Note: P - request for pH range.

    M - request for mV range.

    E - request for EC range.

    T - request for TDS range.

    N - request for NaCl range.

  • 7/29/2019 man2550_23_04_2009

    51/64

    51

    PREPARATION PROCEDUREPREPARATION PROCEDUREPREPARATION PROCEDUREPREPARATION PROCEDUREPREPARATION PROCEDURE

    Remove the protective cap of the pH electrode.

    DO NOT BE ALARMED IF SALT DEPOSITS ARE PRESENT. This is normalwith electrodes. They will disappear when rinsed with water.

    During transport, tiny bubbles of air may form inside the glass bulb

    affecting proper functioning of the electrode. These bubbles can beremoved by shaking down the electrode as you would do with a glassthermometer.

    If the bulb and/or junction is dry, soak the electrode in HI 70300 orHI 80300Storage Solution for at least one hour.

    ELECTRODE CONDIT IONINGELECTRODE CONDIT IONINGELECTRODE CONDIT IONINGELECTRODE CONDIT IONINGELECTRODE CONDIT IONING& MAINTENANCE& MAINTENANCE& MAINTENANCE& MAINTENANCE& MAINTENANCE

  • 7/29/2019 man2550_23_04_2009

    52/64

    52

    For refillable electrodes:For refillable electrodes:For refillable electrodes:For refillable electrodes:For refillable electrodes:

    If the filling solution (electrolyte) is more than 2 cm (1) below the fillhole, add HI 7082 or HI 8082 3.5M KCl Electrolyte Solution for doublejunction or HI 7071 or HI 8071 3.5M KCl+AgCl Electrolyte Solution forsingle junction electrodes.

    For faster response, unscrew the fill hole screw during measurements.

    For AMPHELFor AMPHELFor AMPHELFor AMPHELFor AMPHEL electrodes:electrodes:electrodes:electrodes:electrodes:

    If the electrode does not respond to pH changes, the battery is run downand the electrode should be replaced.

    MEASUREMENTMEASUREMENTMEASUREMENTMEASUREMENTMEASUREMENT

    Rinse the pH electrode tip with distilled water. Submerse the tip (3 cm /1)in the sample and stir gently for a few seconds.

    For a faster response and to avoid cross-contamination of the samples,rinse the electrode tip with a few drops of the solution to be tested, beforetaking measurements.Take care that the sleeve holes of the EC probe are completely submersed.Tap the probe repeatedly to remove air bubbles that may be trappedinside the sleeve.

    STORAGE PROCEDURESTORAGE PROCEDURESTORAGE PROCEDURESTORAGE PROCEDURESTORAGE PROCEDURE

    To minimize clogging and assure a quick response time, the glass bulband the junction of pH electrode should be kept moist and not allowed todry out.

    Replace the solution in the protective cap with a few drops ofHI 70300or HI 80300 Storage Solution or, in its absence, Filling Solution(HI 7071 or HI 8071 for single junction and HI 7082 or HI 8082 fordouble junction electrodes). Follow the Preparation Procedure on page53 before taking measurements.

    Note: NEVER STORE THE ELECTRODE IN DISTILLED OR DEIONIZED WATER.

    PERIODIC MAINTENANCEPERIODIC MAINTENANCEPERIODIC MAINTENANCEPERIODIC MAINTENANCEPERIODIC MAINTENANCE

    Inspect the electrode and the cable. The cable used for connection to theinstrument must be intact and there must be no points of brokeninsulation on the cable or cracks on the electrode stem or bulb. Connectorsmust be perfectly clean and dry. If any scratches or cracks are present,replace the electrode. Rinse off any salt deposits with water.

    pH Electrode MaintenanceFor refillable electrodes:For refillable electrodes:For refillable electrodes:For refillable electrodes:For refillable electrodes:

    Refill the reference chamber with fresh electrolyte (HI 7071 or HI 8071 forsingle junction and HI 7082 or HI 8082 for double junction electrodes).

    Allow the electrode to stand upright for 1 hour.

    Follow the Storage Procedure above.

  • 7/29/2019 man2550_23_04_2009

    53/64

    53

    pHpHpHpHpH ELECTRODE CLEANING PROCEDUREELECTRODE CLEANING PROCEDUREELECTRODE CLEANING PROCEDUREELECTRODE CLEANING PROCEDUREELECTRODE CLEANING PROCEDURE

    General Soak in Hanna HI 7061 or HI 8061 General CleaningSolution for approximately hour.

    Protein Soak in Hanna HI 7073 or HI 8073 Protein CleaningSolution for 15 minutes.

    Inorganic Soak in Hanna HI 7074 Inorganic Cleaning Solutionfor 15 minutes.

    Oil/grease Rinse with Hanna HI 7077 or HI 8077 Oil and FatCleaning Solution.

    IMPORTANT: After performing any of the cleaning procedures, rinse theelectrode thoroughly with distilled water, refill the reference chamberwith fresh electrolyte (not necessary for gel-filled electrodes) and soak theelectrode in HI 70300 or HI 80300 Storage Solution for at least 1 hourbefore taking measurements.

    EC Probe MaintenanceEC Probe MaintenanceEC Probe MaintenanceEC Probe MaintenanceEC Probe Maintenance

    Rinse the probe with clean water after measurements. If a more thoroughcleaning is required, remove the probe sleeve and clean the probe witha cloth or a nonabrasive detergent. Make sure to reinsert the sleeve ontothe probe properly and in the right direction. After cleaning the probe,recalibrate the instrument.

    The platinum rings support is made of glass. Take great care whilehandling the probe.

  • 7/29/2019 man2550_23_04_2009

    54/64

    54

    TROUBLESHOOTING GUIDETROUBLESHOOTING GUIDETROUBLESHOOTING GUIDETROUBLESHOOTING GUIDETROUBLESHOOTING GUIDE

    SMOTPMYS MELBORP NOITULOS

    tfirdevissecxe/esnopserwolS

    .Hpgnirusaem

    .edortceleHpytriD nipitedortceleehtkaoS

    1607IH ro 1608IH 03rof

    ehtnaelcnehtdnasetunim

    .edortcele

    dnapuetautculfsgnidaeR

    .)esion(nwod

    .noitcnujytrid/deggolC

    leveletylortcelewoL

    sedortceleHpelballifer(

    .)ylno

    ylreporptoneveelseborpCEedisniselbbubria;detresni

    .eveels

    .edortceleehtnaelC

    rof(noituloshserfhtiwllifeR

    .)ylnosedortceleHpelballifer

    ehtpaT.eveelsehttresnI

    .selbbubriaevomeroteborp

    tpeccatonseodretemehT

    rofnoitulosnoitarbilaceht

    .noitarbilac

    .edortceleHpredrofotuO

    roedortceleytriD

    noitarbilacdetanimatnoc

    .noitulos

    .erudecorpgninaelcehtwolloF

    ehtecalperstluseronllitsfI

    .edortcele

    ":swohsyalpsidehtfI Hp "

    "dna 00.2- "ro" 0.61 "0

    .gniknilb

    .elacsHpehtniegnarfotuO a .retemehtetarbilaceR)

    elpmasHpehterusekaM)b

    .egnardeificepsehtnisi

    leveletylortceleehtkcehC)c

    ehtfoetatslarenegehtdna

    .edortcele

    ":swohsyalpsidehtfI Vm "

    "dna 0002- "ro" 0002 ".gniknilb

    .elacsVmehtniegnarfotuO yfireV.detcennoctonedortcelE

    .tcennocsiedortceleehttaht

    SDT,CEswohsyalpsidehT

    .gniknilbgnidaerlCaNro

    roSDT,CEniegnarfotuO

    .elacslCaN

    ekaM.retemehtetarbilaceR

    deificepsnisinoitulosehterus

    .egnar

    ehterusekaM KCOL sawyek

    .desserpton

    krowtonseodretemehT

    .eborperutarepmetehthtiw

    roeborperutarepmetnekorB

    eborperutarepmetgnorw

    .desu

    erutarepmetehtecalpeR

    .eborp

    etarbilacotsliafretemehT

    .sgnidaerytluafsevigro

    .edortcelenekorB .edortceleehtecalpeR

    etarbilacotsliafretemehT

    .lCaN

    .noitarbilacCEtcerrocnI CEniretemehtetarbilaceR

    .egnar .1ottnatsnocllecteS

    retemehtputratstA

    sgatDCLllasyalpsid

    .yltnenamrep

    .dekcolbsisyekehtfoenO tcatnocrodraobyekehtkcehC

    .rodneveht

    " xxrrE egassemrorre"

    .deyalpsid

    .rorrelanretnI nehtdnaretemehtfforewoP

    rorreehtfI.notirewop

    .rodnevehttcatnoc,stsisrep

  • 7/29/2019 man2550_23_04_2009

    55/64

    55

    Typical Electrode Life

    Ambient Temperature 1 3 years

    90 C Less than 4 months

    120 C Less than 1 month

    Alkaline Error

    High concentrations of sodium ions interfere with readings in alkalinesolutions. The pH at which the interference starts to be significantdepends upon the composition of the glass. This interference is calledalkaline error and causes the pH to be underestimated. Hannas glassformulations have the indicated characteristics.

    The resistance of glass electrodes partially depends on the temperature.The lower the temperature, the higher the resistance. It takes more timefor the reading to stabilize if the resistance is higher. In addition, theresponse time will suffer to a greater degree at temperatures below 25 C.

    Since the resistance of the pH electrode is in the range of 50 200 Mohm,the current across the membrane is in the pico Ampere range. Largecurrents can disturb the calibration of the electrode for many hours.

    For these reasons high humidity environments, short circuits and staticdischarges are detrimental to a stable pH reading.

    The pH electrodes life also depends on the temperature. If constantlyused at high temperatures, the electrode life is drastically reduced.

    TEMPERATURE CORRELATIONTEMPERATURE CORRELATIONTEMPERATURE CORRELATIONTEMPERATURE CORRELATIONTEMPERATURE CORRELATIONFORFORFORFORFOR ppppp H SENSITIVE GLASSH SENSITIVE GLASSH SENSITIVE GLASSH SENSITIVE GLASSH SENSITIVE GLASS

    1.0 Mol L-1 Na+

    0.1 Mol L-1 Na+

    Sodium Ion Correction for the Glass at 20-25 CConcentration pH Error

    13.0013.5014.0012.5013.0013.5014.00

    0.100.140.200.100.180.290.40

  • 7/29/2019 man2550_23_04_2009

    56/64

    56

    pH BUFFER SOLUTIONS

    HI 70004P pH 4.01 Buffer Sachets, 20 mL, 25 pcsHI 70007P pH 7.01 Buffer Sachets, 20 mL, 25 pcsHI 70010P pH 10.01 Buffer Sachets, 20 mL, 25 pcsHI 7001L pH 1.68 Buffer Solution, 500 mLHI 7004L pH 4.01 Buffer Solution, 500 mLHI 7006L pH 6.86 Buffer Solution, 500 mLHI 7007L pH 7.01 Buffer Solution, 500 mL

    HI 7009L pH 9.18 Buffer Solution, 500 mLHI 7010L pH 10.01 Buffer Solution, 500 mLHI 8004L pH 4.01 Buffer Solution in FDA approved bottle, 500 mLHI 8006L pH 6.86 Buffer Solution in FDA approved bottle, 500 mLHI 8007L pH 7.01 Buffer Solution in FDA approved bottle, 500 mLHI 8009L pH 9.18 Buffer Solution in FDA approved bottle, 500 mLHI 8010L pH 10.01 Buffer Solution in FDA approved bottle, 500 mL

    pH ELECTRODE STORAGE SOLUTIONS

    HI 70300L Storage Solution, 500 mLHI 80300L Storage Solution in FDA approved bottle, 500 mL

    CLEANING SOLUTIONS

    HI 7061M General Cleaning Solution, 250 mL bottleHI 7061L General Cleaning Solution, 500 mL bottleHI 8061M General Cleaning Solution, 250 mL bottle FDA approved bottleHI 8061L General Cleaning Solution, 500mL bottle FDA approved bottleHI 70000P Electrode Rinse Sachets, 20 mL, 25 pcsHI 7073L Protein Cleaning Solution, 500 mLHI 7074L Inorganic Cleaning Solution, 500 mLHI 7077L Oil & Fat Cleaning Solution,500 mLHI 8073L Protein Cleaning Solution in FDA approved bottle, 500 mLHI 8077L Oil & Fat Cleaning Solution in FDA approved bottle, 500 mL

    pH ELECTRODE REFILL ELECTROLYTE SOLUTIONS

    HI 7071 3.5M KCl + AgCl Electrolyte, 4x30 mL, for single junction

    electrodesHI 7072 1M KNO3

    Electrolyte, 4x30 mLHI 7082 3.5M KCl Electrolyte, 4x30 mL, for double junction electrodesHI 8071 3.5M KCl + AgCl Electrolyte in FDA approved bottle, 4x30 mL,

    for single junction electrodesHI 8072 1M KNO

    3Electrolyte in FDA approved bottle, 4x30 mL

    HI 8082 3.5M KCl Electrolyte in FDA approved bottle, 4x30 mL, fordouble junction electrodes

    ACCESSORIESACCESSORIESACCESSORIESACCESSORIESACCESSORIES

  • 7/29/2019 man2550_23_04_2009

    57/64

    57

    ORP PRETREATMENT SOLUTIONS

    HI 7091L Reducing Pretreatment Solution, 500 mLHI 7092L Oxidizing Pretreatment Solution, 500 mL

    CONDUCTIVITY SOLUTIONS

    HI 70030P 12880 S/cm (mho/cm), 20 mL sachets (25 pcs.)HI 70031P 1413 S/cm (mho/cm), 20 mL sachets (25 pcs.)HI 70033P 84 S/cm (mho/cm), 20 mL sachets (25 pcs.)HI 70039P 5000 S/cm (mho/cm), 20 mL sachets (25 pcs.)HI 7030M 12880 S/cm (mho/cm), 250 mL bottleHI 7031M 1413 S/cm (mho/cm), 250 mL bottleHI 7033M 84 S/cm (mho/cm), 250 mL bottleHI 7030M 12880 S/cm (mho/cm), 250 mL bottleHI 7034M 80000 S/cm (mho/cm), 250 mL bottleHI 7035M 111800 S/cm (mho/cm), 250 mL bottleHI 7039M 5000 S/cm (mho/cm), 250 mL bottleHI 7030L 12880 S/cm (mho/cm), 500 mL bottleHI 7031L 1413 S/cm (mho/cm), 500 mL bottleHI 7033L 84 S/cm (mho/cm), 500 mL bottleHI 7034L 80000 S/cm (mho/cm), 500 mL bottleHI 7035L 111800 S/cm (mho/cm), 500 mL bottleHI 7039L 5000 S/cm (mho/cm), 500 mL bottleHI 7037L 100% NaCl sea water standard solution, 500 mL bottle

    HI 8030L 12880 S/cm (mho/cm), 500 mL FDA approved bottleHI 8031L 1413 S/cm (mho/cm), 500 mL FDA approved bottleHI 8033L 84 S/cm (mho/cm), 500 mL FDA approved bottleHI 8034L 80000 S/cm (mho/cm), 500 mL FDA approved bottleHI 8035L 111800 S/cm (mho/cm), 500 mL FDA approved bottleHI 8039L 5000 S/cm (mho/cm), 500 mL FDA approved bottle

    TDS SOLUTIONS

    HI 70080C 800 ppm (mg/L), 20 mL (25 pcs.)HI 70080P 800 ppm (mg/L), 20 mL (25 pcs.)HI 70032C 1382 ppm (mg/L), 20 mL (25 pcs.)HI 70032P 1382 ppm (mg/L), 20 mL (25 pcs.)HI 77300C 1382 ppm (mg/L) & pH 7.01, 20 mL (10 pcs.)

    HI 77300P 1382 ppm (mg/L) & pH 7.01 20 mL (10 pcs.)HI 70442C* 1500 ppm (mg/L) , 20 mL (25 pcs.)HI 70442P* 1500 ppm (mg/L) , 20 mL (25 pcs.)HI 77200C* 1500 ppm (mg/L) & pH 7.01, 20 mL (10 pcs.)HI 77200P* 1500 ppm (mg/L) & pH 7.01, 20 mL (10 pcs.)HI 7032M 1382 ppm (mg/L) , 250 mLHI 7032L 1382 ppm (mg/L) , 500 mL

  • 7/29/2019 man2550_23_04_2009

    58/64

    58

    HI 70442M* 1500 ppm (mg/L), 250 mLHI 70442L* 1500 ppm (mg/L), 500 mL* 1500 ppm TDS have an approximate conversion factor of:

    0.65 ppm = 1 S/cm conversion rate.

    pH ELECTRODES

    All electrodes part numbers ending in B are supplied with a BNC connectorand 1 m (3.3') cable, as shown below :

    HI 1043BGlass-body, double junction, refillable, combination pH electrode.

    Use: strong acid/alkali.

    HI 1053B

    Glass-body, triple ceramic, conic shape, refillable, combination pHelectrode.

    Use: emulsions.

    HI 1083B

    Glass-body, micro, Viscolene, non-refillable, combination pH electrode.

    Use: biotechnology, micro titration.

    HI 1131B

    Glass-body, single junction, refillable, combination pH electrode.

    Use: general purpose.

  • 7/29/2019 man2550_23_04_2009

    59/64

    59

    HI 1330B

    Glass-body, semimicro, single junction, refillable, combination pH electrode.

    Use: laboratory, vials.

    HI 1331B

    Glass-body, semimicro, single junction, refillable, combination pH electrode.

    Use: flasks.

    HI 1230B

    Plastic-body (PES), double junction, gel-filled, combinationpH electrode.

    Use: general, field.

    HI 2031B

    Glass-body, semimicro, conic, refillable, combination pH electrode.

    Use: semisolid products.

    HI 1332B

    Plastic-body (PES), double junction, refillable, combination pH electrode.Use: general purpose.

  • 7/29/2019 man2550_23_04_2009

    60/64

    60

    FC 100B

    Plastic-body (PVDF), double junction, refillable, combination pH electrode.

    Use: general purpose for food industry.

    FC 200B

    Plastic-body (PVDF), open junction, conic, Viscolene, non-refillable,

    combination pH electrode. Use: meat & cheese.

    FC 210B

    Glass-body, double junction, conic, Viscolene, non-refillable, combinationpH electrode. Use: milk, yogurt.

    FC 220B

    Glass-body, triple-ceramic, single junction, refillable, combination pHelectrode. Use: food processing.

    FC 911B

    Plastic-body (PVDF), double junction, refillable with built-in amplifier,combination pH electrode. Use: very high humi